Vol. 98-B, Issue SUPP_4 | Pages 113 - 113
1 Jan 2016
Munir S Molloy D Hasted T Jack CM Shimmin A Walter W
Full Access

Sagittal stability of the knee is believed to be of significant importance following total knee arthroplasty. We examine four different knee designs at a minimum of twenty-four months postoperatively. Sagittal stability was measured at four degrees of flexion: 0°; 30°; 60°; and 90°, to examine the effect of design on mid-flexion stability.

The knee designs included were: the rotating platform LCS design (DePuy); the cruciate sparing Triathlon system (Stryker); SAIPH system (Matortho, UK); and the medial rotating knee design, MRK (Matortho, UK).

Following ethical approval, 64 cases were enrolled into the study, 22 male and 42 female. Inclusion criteria included: a minimum of 18 months from surgery; ability to flex beyond 90 degrees; and have no postoperative complications. 18 LCS, 18 MRK, 14 SAIPH and 14 Triathlon knee designs were analysed. Sagittal stability was measured using the KT1000 device. Active range of movement was measured using a hand held goniometer and recorded as was Oxford knee score, WOMAC knee score, SF12 and Kujala patellofemoral knee score.

Mean follow-up was 33.7 months postoperative, with a mean age of 72 years. Mean weight was 82.7kgs and height 164cms. There was no significant difference in preoperative demographics between the groups. Mean active post-operative range of motion of the knee was from 2–113° with no significant difference between groups.

Sagittal stability was similar in all four groups in full extension; however the MRK and SAIPH designs showed improved stability in the mid-range of flexion (30–90°). Patient satisfaction also showed a similar trend with MRK achieving better patient reported functional outcomes and satisfaction than that of the SAIPH, LCS and Triathlon systems.

All four knee designs demonstrated good post-operative range of movement with comparative improvement of patient scores to other reported studies. The MRK and SAIPH knee design showed an improved mid-flexion sagittal stability with better patient reported satisfaction and functional scores.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 20 - 20
1 Jan 2016
Marel E Walter L Solomon M Shimmin A Pierrepont J
Full Access

Malorientation of the acetabular cup in Total Hip replacement (THR) may contribute to premature failure of the joint through instability (impingement, subluxation or dislocation), runaway wear in metal-metal bearings when the edge of the contact patch encroaches on the edge of the bearing surface, squeaking of ceramic-ceramic bearings and excess wear of polyethylene bearing surfaces leading to osteolysis.

However as component malorientation often only occurs in functional positions it has been difficult to demonstrate and often is unremarkable on standard (usually supine) pelvic radiographs. The effects of spinal pathology as well as hip pathology can cause large rotations of the pelvis in the sagittal plane, again usually not recognized on standard pelvic views. While Posterior pelvic rotation with sitting increases the functional arc of the hip and is protective of a THR in regards to both edge loading and risk of dislocation, conversely Anterior rotation with sitting is potentially hazardous.

We developed a protocol using three functional positions – standing, supine and flexed seated (posture at “seat-off” from a standard chair). Lateral radiographs were used to define the pelvic tilt in the standing and flexed seated positions. Pelvic tilt was defined as the angle between a vertical reference line and the anterior pelvic plane. Supine pelvic tilt was measured from computed tomography.

Proprietary software (Optimized Ortho, Sydney) based on Rigid Body Dynamics then modelled the patients’ dynamics through their functional range producing a patient-specific simulation which also calculates the magnitude and direction of the dynamic force at the hip and traces the contact area between prosthetic head/liner onto a polar plot of the articulating surface. Given prosthesis specific information edge-loading can then be predicted based on the measured distance of the edge of the contact patch to the edge of the acetabular bearing.

Results and conclusions

The position of the pelvis in the sagittal plane changes significantly between functional activities. The extent of change is specific to each patient.

Spinal pathology can be an insidious “driver” of pelvic rotation, in some cases causing sagittal plane spinal imbalance or changes in orientation of previously well oriented acetabular components.

Squeaking of ceramic on ceramic bearings appears to be multi factorial, usually involving some damage to the bearing but also usually occurring in the presence of anterior or posterior edge loading. Often these components will appear well oriented on standard views [Fig 1].

Runaway wear in hip resurfacing or large head metal-metal THR may be caused by poor component design or manufacture or component malorientation. Again we have seen multiple cases where no such malorientation can be seen on standard pelvic radiographs but functional studies demonstrate edge loading which is likely to be the cause of failure [Fig 2].

Clinical examples of all of these will be shown.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 19 - 19
1 Jan 2016
Marel E Walter L Solomon M Shimmin A Pierrepont J
Full Access

Achieving optimal acetabular cup orientation in Total Hip Replacement (THR) remains one of the most difficult challenges in THR surgery (AAOR 2013) but very little has been added to useful understanding since Lewinnek published recommendations in 1978. This is largely due to difficulties of analysis in functional positions. The pelvis is not a static reference but rotates especially in the sagittal plane depending upon the activity being performed. These dynamic changes in pelvic rotation have a substantial effect on the functional orientation of the acetabulum, not appreciated on standard radiographs [Fig1].

Studies of groups of individuals have found the mean pelvic rotation in the sagittal plane is small but large individual variations commonly occur. Posterior rotation, with sitting, increases the functional arc of the hip and is protective of a THR in regards to both edge loading and risk of dislocation. Conversely Anterior rotation, with sitting, is potentially hazardous.

We developed a protocol using three functional positions – standing, supine and flexed seated (posture at “seat-off” from a standard chair). Lateral radiographs were used to define the pelvic tilt in the standing and flexed seated positions. Pelvic tilt was defined as the angle between a vertical reference line and the anterior pelvic plane (defined by the line joining both anterior superior iliac spines and the pubic symphysis). In the supine position pelvic tilt was defined as the angle between a horizontal reference line and the anterior pelvic plane. Supine pelvic tilt was measured from computed tomography.

Proprietary software (Optimized Ortho, Sydney) based on Rigid Body Dynamics then modelled the patients’ dynamics through their functional range producing a patient-specific simulation which also calculates the magnitude and direction of the dynamic force at the hip and traces the contact area between prosthetic head/liner onto a polar plot of the articulating surface, Fig 2. Given prosthesis specific information edge-loading can then be predicted based on the measured distance of the contact patch to the edge of the acetabular liner.

Delivery of desired orientation at surgery is facilitated by use of a solid 3D printed model of the acetabulum along with a patient specific guide which fits the model and the intra-operative acetabulum (with cartilage but not osteophytes removed) - an incorporated laser pointer then marks a reference point for the reamer and cup inserter to replicate the chosen orientation.

Results and conclusions

The position of the pelvis in the sagittal plane changes significantly between functional activities. The extent of change is specific to each patient. Spinal pathology is a potent “driver” of pelvic sagittal rotation, usually unrecognised on standard radiographs. Pre-operative patient assessment can identify potential orientation problems and even suitability for hard on hard bearings.

Optimal cup orientation is likely patient-specific and requires an evaluation of functional pelvic dynamics to pre-operatively determine the target angles.

Post-operatively this technique can identify patient and implant factors likely to be causing edge loading leading to early failure in metal on metal bearings or squeaking in ceramic on ceramic bearings.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 171 - 171
1 Dec 2013
Shimmin A Martos SM Owens J Iorgulescu A
Full Access

Introduction

The SAIPH™ (MatOrtho, UK) total knee replacement is a new fixed-bearing prosthesis design having attributes of a mobile bearing and the posterior stabilised categories for knee arthroplasties. The implant design goal is an articulation that provides definitive anteroposterior stability to beneficially control tibiofemoral translation, the ability for the tibia to axially rotate to accommodate various lifestyle activities, and to maintain a relatively posterior femoral position on the tibia to facilitate range of motion. This study aims to analyze knee kinematics of the SAIPH™ total knee arthroplasty (TKA) by videofluroscopy during four different weightbearing activities.

Method

Fourteen consecutive patients operated on by a single surgeon, with a minimum follow up of 24 months were included in this IRB-approved study. A medially conforming knee was implanted in all cases. Participants in the study were asked to perform weightbearing kneeling, lunging, step-up/down and pivoting activities while their knee motions were recorded by videofluoroscopy. Three-dimensional (3D) joint kinematics were determined using model-image registration. The 3D orientation of each TKA component was expressed using standard joint angle conventions, and the anterior/posterior location of each condyle was expressed relative to the deepest part of the tibial sulcus.