The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 2 | Pages 209 - 214
1 Mar 2003
Aldinger PR Breusch SJ Lukoschek M Mau H Ewerbeck V Thomsen M

We followed the first 354 consecutive implantations of a cementless, double-tapered straight femoral stem in 326 patients. Follow-up was at a mean of 12 years (10 to 15). The mean age of the patients was 57 years (13 to 81). At follow-up, 56 patients (59 hips) had died, and eight (eight hips) had been lost to follow-up. Twenty-five hips underwent femoral revision, eight for infection, three for periprosthetic fracture and 14 for aseptic loosening.

The overall survival was 92% at 12 years (95% CI 88 to 95). Survival with femoral revision for aseptic loosening as an endpoint was 95% (95% CI 92 to 98). The median Harris hip score at follow-up was 84 points (23 to 100). Radiolucent lines (< 2 mm) in Gruen zones 1 and 7 were present in 38 (16%) and 34 hips (14%), respectively. Radiolucencies in zones 2 to 6 were found in five hips (2%).

The results for mid- to long-term survival with this femoral component are encouraging and compare with those achieved in primary cemented total hip arthroplasty. The high rate of loosening of the cup and the high rate of pain are, however, a source of concern.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 3 | Pages 366 - 370
1 May 1997
Simank H Brocai DRC Reiser D Thomsen M Sabo D Lukoschek M

We report our results using three different threaded acetabular components (Mecring A, Mecring B and Weill) in 715 hips with a follow-up of between one and ten years (median: 99.1, 56.5, 38.3 months, respectively). All cups were implanted with one type of cementless stem.

The clinical results were good or acceptable in about 70% of the hips, but signs of loosening with radiolucency and/or migration were found in 10.1%. Radiological evidence of loosening did not correlate significantly with the clinical outcome. Pain was not a reliable indicator of loosening and its absence sometimes allowed severe osteolysis to develop. Twenty-five hips were revised (3.5%) for aseptic loosening of the acetabular component. Kaplan-Meier estimates of the cumulative rate of failure showed a rapid increase five years after the initial operation, but no significant correlation with gender, age or weight.

The high rate of failure indicates that further use of these acetabular components cannot be recommended. Annual radiographs are required to assess osteolysis even if the patients are free from pain.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 2 | Pages 203 - 206
1 Mar 1991
Bernd L Blasius K Lukoschek M Lucke R

Autologous stump capping is a procedure designed to prevent bony overgrowth in skeletally immature amputation stumps. All 19 capping procedures in the lower extremities were successful after an average follow-up of 7.3 years. All patients use their prostheses, and no secondary operations have been needed for stump problems. Of the 31 cap-plasties of the humerus, six required re-operation. The overall failure rate of 12% is low compared with the failure rate of re-amputation.