Abstract
We report our results using three different threaded acetabular components (Mecring A, Mecring B and Weill) in 715 hips with a follow-up of between one and ten years (median: 99.1, 56.5, 38.3 months, respectively). All cups were implanted with one type of cementless stem.
The clinical results were good or acceptable in about 70% of the hips, but signs of loosening with radiolucency and/or migration were found in 10.1%. Radiological evidence of loosening did not correlate significantly with the clinical outcome. Pain was not a reliable indicator of loosening and its absence sometimes allowed severe osteolysis to develop. Twenty-five hips were revised (3.5%) for aseptic loosening of the acetabular component. Kaplan-Meier estimates of the cumulative rate of failure showed a rapid increase five years after the initial operation, but no significant correlation with gender, age or weight.
The high rate of failure indicates that further use of these acetabular components cannot be recommended. Annual radiographs are required to assess osteolysis even if the patients are free from pain.