Advertisement for orthosearch.org.uk
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Arthroplasty

MIDDLE-TERM RESULTS OF THREADED ACETABULAR CUPS

HIGH FAILURE RATES FIVE YEARS AFTER SURGERY



Download PDF

Abstract

We report our results using three different threaded acetabular components (Mecring A, Mecring B and Weill) in 715 hips with a follow-up of between one and ten years (median: 99.1, 56.5, 38.3 months, respectively). All cups were implanted with one type of cementless stem.

The clinical results were good or acceptable in about 70% of the hips, but signs of loosening with radiolucency and/or migration were found in 10.1%. Radiological evidence of loosening did not correlate significantly with the clinical outcome. Pain was not a reliable indicator of loosening and its absence sometimes allowed severe osteolysis to develop. Twenty-five hips were revised (3.5%) for aseptic loosening of the acetabular component. Kaplan-Meier estimates of the cumulative rate of failure showed a rapid increase five years after the initial operation, but no significant correlation with gender, age or weight.

The high rate of failure indicates that further use of these acetabular components cannot be recommended. Annual radiographs are required to assess osteolysis even if the patients are free from pain.


Correspondence should be sent to Dr H.-G. Simank.

For access options please click here