Advertisement for orthosearch.org.uk
Results 1 - 20 of 29
Results per page:

Aims

For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis.

Methods

We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 1000 - 1007
1 Sep 2024
Gong T Lu M Sheng H Li Z Zhou Y Luo Y Min L Tu C

Aims

Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes.

Methods

This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.


Aims

To evaluate mid-to long-term patient-reported outcome measures (PROMs) of endoprosthetic reconstruction after resection of malignant tumours arising around the knee, and to investigate the risk factors for unfavourable PROMs.

Methods

The medical records of 75 patients who underwent surgery between 2000 and 2020 were retrospectively reviewed, and 44 patients who were alive and available for follow-up (at a mean of 9.7 years postoperatively) were included in the study. Leg length discrepancy was measured on whole-leg radiographs, and functional assessment was performed with PROMs (Toronto Extremity Salvage Score (TESS) and Comprehensive Outcome Measure for Musculoskeletal Oncology Lower Extremity (COMMON-LE)) with two different aspects. The thresholds for unfavourable PROMs were determined using anchor questions regarding satisfaction, and the risk factors for unfavourable PROMs were investigated.


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1180 - 1188
1 Oct 2022
Qu H Mou H Wang K Tao H Huang X Yan X Lin N Ye Z

Aims

Dislocation of the hip remains a major complication after periacetabular tumour resection and endoprosthetic reconstruction. The position of the acetabular component is an important modifiable factor for surgeons in determining the risk of postoperative dislocation. We investigated the significance of horizontal, vertical, and sagittal displacement of the hip centre of rotation (COR) on postoperative dislocation using a CT-based 3D model, as well as other potential risk factors for dislocation.

Methods

A total of 122 patients who underwent reconstruction following resection of periacetabular tumour between January 2011 and January 2020 were studied. The risk factors for dislocation were investigated with univariate and multivariate logistic regression analysis on patient-specific, resection-specific, and reconstruction-specific variables.


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 504 - 509
1 Apr 2022
Kennedy JW Farhan-Alanie OM Young D Kelly MP Young PS

Aims

The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases.

Methods

This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1405 - 1413
1 Aug 2021
Ogura K Fujiwara T Morris CD Boland PJ Healey JH

Aims

Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model.

Methods

We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 170 - 176
1 Feb 2020
Bernthal NM Burke ZDC Hegde V Upfill-Brown A Chen CJ Hwang R Eckardt JJ

Aims

We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction.

Methods

This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1084 - 1089
1 Aug 2008
Guo W Ji T Yang R Tang X Yang Y

In developing countries locally-made low-cost prostheses are mainly used in limb-salvage surgery to alleviate the economic burden. We retrospectively collected data on 104 patients treated by limb-salvage surgery between July 1997 and July 2005. We used a locally-designed and fabricated stainless-steel endoprosthesis in each case. Oncological and functional outcomes were evaluated at a mean follow-up of 47 months (12 to 118). A total of 73 patients (70.2%) were free from disease, nine (8.7%) were alive with disease, 19 (18.2%) had died from their disease and three (2.9%) from unrelated causes. According to the Musculoskeletal Tumor Society scoring system, the mean functional score was 76.3% (SD 17.8). The five-year survival for the implant was 70.5%. There were nine cases (8.7%) of infection, seven early and two late, seven (6.7%) of breakage of the prosthesis, three (2.9%) of aseptic loosening and two (1.9%) of failure of the polyethylene bushing. Multivariate analysis showed that a proximal tibial prosthesis and a resection length of 14 cm or more were significant negative prognostic factors. Our survival rates and Musculoskeletal Tumor Society functional scores are similar to those reported in the literature. Although longer follow-up is needed to confirm our results, we believe that a low-cost custom-made endoprosthesis is a cost-effective and reliable reconstructive option for limb salvage in developing countries


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 522 - 528
1 May 2019
Medellin MR Fujiwara T Clark R Stevenson JD Parry M Jeys L

Aims

The aim of this study was to evaluate the prosthesis characteristics and associated conditions that may modify the survival of total femoral endoprosthetic replacements (TFEPR).

Patients and Methods

In all, 81 patients treated with TFEPR from 1976 to 2017 were retrospectively evaluated and failures were categorized according to the Henderson classification. There were 38 female patients (47%) and 43 male patients (53%) with a mean age at diagnosis of 43 years (12 to 86). The mean follow-up time was 10.3 years (0 to 31.7). A survival analysis was performed followed by univariate and multivariate Cox regression to identify independent implant survival factors.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1640 - 1646
1 Dec 2018
Medellin MR Fujiwara T Clark R Jeys LM

Aims

The aim of this study was to describe, analyze, and compare the survival, functional outcome, and complications of minimally invasive (MI) and non-invasive (NI) lengthening total femoral prostheses.

Patients and Methods

A total of 24 lengthening total femoral prostheses, 11 MI and 13 NI, were implanted between 1991 and 2016. The characteristics, complications, and functional results were recorded. There were ten female patients and ten male patients. Their mean age at the time of surgery was 11 years (2 to 41). The mean follow-up was 13.2 years (seven months to 29.25 years). A survival analysis was performed, and the failures were classified according to the Modified Henderson System.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 857 - 864
1 Jun 2016
Grimer RJ Aydin BK Wafa H Carter SR Jeys L Abudu A Parry M

Aims

The aim of this study was to establish what happens to patients in the long term after endoprosthetic replacement for a primary malignant tumour of bone.

Patients and Methods

We conducted a retrospective analysis of a prospectively maintained database to identify all patients who had undergone an endoprosthetic replacement more than 25 years ago and who were still alive. Their outcomes were investigated with reference to their complications and need for further surgery. A total of 230 patients were identified. Their mean age at diagnosis was 20.7 years (five to 62). The most common diagnosis was osteosarcoma (132). The most common site was the distal femur (102).


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1706 - 1712
1 Dec 2014
Bus MPA Boerhout EJ Bramer JAM Dijkstra PDS

Peri-acetabular tumour resections and their subsequent reconstruction are among the most challenging procedures in orthopaedic oncology. Despite the fact that a number of different pelvic endoprostheses have been introduced, rates of complication remain high and long-term results are mostly lacking.

In this retrospective study, we aimed to evaluate the outcome of reconstructing a peri-acetabular defect with a pedestal cup endoprosthesis after a type 2 or type 2/3 internal hemipelvectomy.

A total of 19 patients (11M:8F) with a mean age of 48 years (14 to 72) were included, most of whom had been treated for a primary bone tumour (n = 16) between 2003 and 2009. After a mean follow-up of 39 months (28 days to 8.7 years) seven patients had died. After a mean follow-up of 7.9 years (4.3 to 10.5), 12 patients were alive, of whom 11 were disease-free. Complications occurred in 15 patients. Three had recurrent dislocations and three experienced aseptic loosening. There were no mechanical failures. Infection occurred in nine patients, six of whom required removal of the prosthesis. Two patients underwent hindquarter amputation for local recurrence.

The implant survival rate at five years was 50% for all reasons, and 61% for non-oncological reasons. The mean Musculoskeletal Tumor Society score at final follow-up was 49% (13 to 87).

Based on these poor results, we advise caution if using the pedestal cup for reconstruction of a peri-acetabular tumour resection.

Cite this article: Bone Joint J 2014;96-B:1706–12.


Bone & Joint Research
Vol. 4, Issue 9 | Pages 154 - 162
1 Sep 2015

Objective

Clinical studies of patients with bone sarcomas have been challenged by insufficient numbers at individual centres to draw valid conclusions. Our objective was to assess the feasibility of conducting a definitive multi-centre randomised controlled trial (RCT) to determine whether a five-day regimen of post-operative antibiotics, in comparison to a 24-hour regimen, decreases surgical site infections in patients undergoing endoprosthetic reconstruction for lower extremity primary bone tumours.

Methods

We performed a pilot international multi-centre RCT. We used central randomisation to conceal treatment allocation and sham antibiotics to blind participants, surgeons, and data collectors. We determined feasibility by measuring patient enrolment, completeness of follow-up, and protocol deviations for the antibiotic regimens.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1425 - 1431
1 Oct 2013
Hardes J Henrichs MP Gosheger G Gebert C Höll S Dieckmann R Hauschild G Streitbürger A

We evaluated the clinical results and complications after extra-articular resection of the distal femur and/or proximal tibia and reconstruction with a tumour endoprosthesis (MUTARS) in 59 patients (mean age 33 years (11 to 74)) with malignant bone or soft-tissue tumours. According to a Kaplan–Meier analysis, limb survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection was the most common indication for subsequent amputation (eight patients). Survival of the prosthesis without revision was 48% (95% CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years post-operatively. Failure of the prosthesis was due to deep infection in 22 patients (37%), aseptic loosening in ten patients (17%), and peri-prosthetic fracture in six patients (10%). Wear of the bearings made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal Tumor Society score was 23 (10 to 29). An extensor lag > 10° was noted in ten patients (17%).

These results suggest that limb salvage after extra-articular resection with a tumour prosthesis can achieve good functional results in most patients, although the rates of complications and subsequent amputation are higher than in patients treated with intra-articular resection.

Cite this article: Bone Joint J 2013;95-B:1425–31.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 399 - 405
1 Mar 2014
Zang J Guo W Yang Y Xie L

We report our early experience with the use of a new prosthesis, the Modular Hemipelvic Prosthesis II, for reconstruction of the hemipelvis after resection of a primary malignant peri-acetabular tumour involving the sacroiliac joint.

We retrospectively reviewed the outcome of 17 patients who had undergone resection of a pelvic tumour and reconstruction with this prosthesis between July 2002 and July 2010.

One patient had a type I+II+III+IV resection (ilium + peri-acetabulum + pubis/ischium + sacrum) and 16 had a type I+II+IV resection (ilium + acetabulum + sacrum). The outcome was assessed at a mean follow-up of 33 months (15 to 59). One patient was alive with disease, 11 were alive without disease and five had died of disease. The overall five-year survival rate was 62.4%. Six patients had a local recurrence. The mean Musculoskeletal Tumour Society score was 58% (33 to 77). Deep infection occurred in two patients, problems with wound healing in five and dislocation in one.

For patients with a primary malignant peri-acetabular sarcoma involving the sacroiliac joint, we believe that this new prosthesis is a viable option for reconstruction of the bony defect left following resection of the tumour. It results in a satisfactory functional outcome with an acceptable rate of complications.

Cite this article: Bone Joint J 2014;96-B:399–405.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 265 - 269
1 Feb 2012
Hwang N Grimer RJ Carter SR Tillman RM Abudu A Jeys LM

We reviewed our initial seven-year experience with a non-invasive extendible prosthesis in 34 children with primary bone tumours. The distal femur was replaced in 25 cases, total femur in five, proximal femur in one and proximal tibia in three. The mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain alive. The prostheses were lengthened by an electromagnetic induction mechanism in an outpatient setting and a mean extension of 32 mm (4 to 80) was achieved without anaesthesia. There were lengthening complications in two children: failed lengthening in one and the formation of scar tissue in the other. Deep infection developed in six patients (18%) and local recurrence in three. A total of 11 patients required further surgery to the leg. Amputation was necessary in five patients (20%) and a two-stage revision in another. There were no cases of loosening, but two patients had implant breakage and required revision. The mean Musculoskeletal Tumor Society functional score was 85% (60% to 100%) at last known follow-up. These early results demonstrate that the non-invasive extendible prosthesis allows successful lengthening without surgical intervention, but the high incidence of infection is a cause for concern.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 425 - 430
1 Mar 2012
Picardo NE Blunn GW Shekkeris AS Meswania J Aston WJ Pollock RC Skinner JA Cannon SR Briggs TW

In skeletally immature patients, resection of bone tumours and reconstruction of the lower limb often results in leg-length discrepancy. The Stanmore non-invasive extendible endoprosthesis, which uses electromagnetic induction, allows post-operative lengthening without anaesthesia. Between 2002 and 2009, 55 children with a mean age of 11.4 years (5 to 16) underwent reconstruction with this prosthesis; ten patients (18.2%) died of disseminated disease and one child underwent amputation due to infection. We reviewed 44 patients after a mean follow-up of 41.2 months (22 to 104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to 30) and the Toronto Extremity Salvage score was 92.3% (55.2% to 99.0%). There was no local recurrence of tumour. Complications developed in 16 patients (29.1%) and ten (18.2%) underwent revision.

The mean length gained per patient was 38.6 mm (3.5 to 161.5), requiring a mean of 11.3 extensions (1 to 40), and ten component exchanges were performed in nine patients (16.4%) after attaining the maximum lengthening capacity of the implant. There were 11 patients (20%) who were skeletally mature at follow-up, ten of whom had equal leg lengths and nine had a full range of movement of the hip and knee.

This is the largest reported series using non-invasive extendible endoprostheses after excision of primary bone tumours in skeletally immature patients. The technique produces a good functional outcome, with prevention of limb-length discrepancy at skeletal maturity.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1378 - 1382
1 Oct 2009
Shekkeris AS Hanna SA Sewell MD Spiegelberg BGI Aston WJS Blunn GW Cannon SR Briggs TWR

Endoprosthetic replacement of the distal tibia and ankle joint for a primary bone tumour is a rarely attempted and technically challenging procedure. We report the outcome of six patients treated between 1981 and 2007. There were four males and two females, with a mean age of 43.5 years (15 to 75), and a mean follow-up of 9.6 years (1 to 27). No patient developed a local recurrence or metastasis. Two of the six went on to have a below-knee amputation for persistent infection after a mean 16 months (1 to 31). The four patients who retained their endoprosthesis had a mean musculoskeletal tumour society score of 70% and a mean Toronto extremity salvage score of 71%. All were pain free and able to perform most activities of daily living in comfort.

A custom-made endoprosthetic replacement of the distal tibia and ankle joint is a viable treatment option for carefully selected patients with a primary bone tumour. Patients should, however, be informed of the risk of infection and the potential need for amputation if this cannot be controlled.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1632 - 1637
1 Dec 2007
Myers GJC Abudu AT Carter SR Tillman RM Grimer RJ

We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database.

A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic loosening was the most common reason for revision in the fixed-hinge knees, fracture of the implant in the early design of rotating hinges and infection in the current version. The risk of revision for aseptic loosening in the fixed-hinge knees was 46% at ten years. This was reduced to 3% in the rotating-hinge knee with a hydroxyapatite collar. The cemented, rotating hinge design currently offers the best chance of long-term survival of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 982 - 988
1 Jul 2012
Puri A Gulia A Jambhekar N Laskar S

We analysed the outcome of patients with primary non-metastatic diaphyseal sarcomas who had en bloc resection with preservation of the adjoining joints and reconstruction with re-implantation of sterilised tumour bone after extracorporeal radiation (50 Gy). Between March 2005 and September 2009, 32 patients (16 Ewing’s sarcoma and 16 osteogenic sarcoma) with a mean age of 15 years (2 to 35) underwent this procedure. The femur was the most common site in 17 patients, followed by the tibia in 11, humerus in three and ulna in one. The mean resected length of bone was 19 cm (10 to 26). A total of 31 patients were available at a mean follow-up of 34 months (12 to 74). The mean time to union for all osteotomy sites was 7.3 months (3 to 28): metaphyseal osteotomy sites united quicker than diaphyseal osteotomy sites (5.8 months (3 to 10) versus 9.5 months (4 to 28)). There were three local recurrences, all in soft-tissue away from irradiated graft. At the time of final follow-up, 19 patients were free of disease, one was alive with disease and 11 had died of disease. The mean Musculoskeletal Tumor Society Score for 29 patients evaluated at the last follow-up was 26 (9 to 30).

Extracorporeal irradiation is an oncologically safe and inexpensive technique for limb salvage in diaphyseal sarcomas and has good functional results.