Advertisement for orthosearch.org.uk
Results 1 - 50 of 63
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 308 - 313
1 Mar 2008
Murtha PE Hafez MA Jaramaz B DiGioia AM

Three-dimensional surface models of the normal hemipelvis derived from volumetric CT data on 42 patients were used to determine the radius, depth and orientation of the native acetabulum. A sphere fitted to the lunate surface and a plane matched to the acetabular rim were used to calculate the radius, depth and anatomical orientation of the acetabulum. For the 22 females the mean acetabular abduction, anteversion, radius and normalised depth were 57.1° (50.7° to 66.8°), 24.1° (14.0° to 33.3°), 25 mm (21.7 to 30.3) and 0.79 mm (0.56 to 1.04), respectively. The same parameters for the 20 males were 55.5° (47.7° to 65.9°), 19.3° (8.5° to 32.3°), 26.7 mm (24.5 to 28.7) and 0.85 mm (0.65 to 0.99), respectively. The orientation of the native acetabulum did not match the safe zone for acetabular component placement described by Lewinnek. During total hip replacement surgeons should be aware that the average abduction angle of the native acetabulum exceeds that of the safe zone angle. If the concept of the safe zone angle is followed, abduction of the acetabular component should be less than the abduction of the native acetabulum by approximately 10°


Bone & Joint Open
Vol. 5, Issue 5 | Pages 394 - 400
15 May 2024
Nishi M Atsumi T Yoshikawa Y Okano I Nakanishi R Watanabe M Usui Y Kudo Y

Aims

The localization of necrotic areas has been reported to impact the prognosis and treatment strategy for osteonecrosis of the femoral head (ONFH). Anteroposterior localization of the necrotic area after a femoral neck fracture (FNF) has not been properly investigated. We hypothesize that the change of the weight loading direction on the femoral head due to residual posterior tilt caused by malunited FNF may affect the location of ONFH. We investigate the relationship between the posterior tilt angle (PTA) and anteroposterior localization of osteonecrosis using lateral hip radiographs.

Methods

Patients aged younger than 55 years diagnosed with ONFH after FNF were retrospectively reviewed. Overall, 65 hips (38 males and 27 females; mean age 32.6 years (SD 12.2)) met the inclusion criteria. Patients with stage 1 or 4 ONFH, as per the Association Research Circulation Osseous classification, were excluded. The ratios of anterior and posterior viable areas and necrotic areas of the femoral head to the articular surface were calculated by setting the femoral head centre as the reference point. The PTA was measured using Palm’s method. The association between the PTA and viable or necrotic areas of the femoral head was assessed using Spearman’s rank correlation analysis (median PTA 6.0° (interquartile range 3 to 11.5)).


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 855 - 858
1 Jul 2009
Wroblewski BM Siney PD Fleming PA

The design of the Charnley total hip replacement follows the principle of low frictional torque. It is based on the largest possible difference between the radius of the femoral head and that of the outer aspect of the acetabular component. The aim is to protect the bone-cement interface by movement taking place at the smaller radius, the articulation. This is achieved in clinical practice by a 22.225 mm diameter head articulating with a 40 mm or 43 mm diameter acetabular component of ultra-high molecular weight polyethylene. We compared the incidence of aseptic loosening of acetabular components with an outer diameter of 40 mm and 43 mm at comparable depths of penetration with a mean follow-up of 17 years (1 to 40). In cases with no measurable wear none of the acetabular components were loose. With increasing acetabular penetration there was an increased incidence of aseptic loosening which reflected the difference in the external radii, with 1.5% at 1 mm, 8.8% at 2 mm, 9.7% at 3 mm and 9.6% at 4 mm of penetration in favour of the larger 43 mm acetabular component. Our findings support the Charnley principle of low frictional torque. The level of the benefit is in keeping with the predicted values


Bone & Joint Open
Vol. 5, Issue 1 | Pages 28 - 36
18 Jan 2024
Selmene MA Moreau PE Zaraa M Upex P Jouffroy P Riouallon G

Aims

Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications.

Methods

Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 783 - 791
1 Aug 2024
Tanaka S Fujii M Kawano S Ueno M Nagamine S Mawatari M

Aims

The aim of this study was to determine the clinical outcomes and factors contributing to failure of transposition osteotomy of the acetabulum (TOA), a type of spherical periacetabular osteotomy, for advanced osteoarthritis secondary to hip dysplasia.

Methods

We reviewed patients with Tönnis grade 2 osteoarthritis secondary to hip dysplasia who underwent TOA between November 1998 and December 2019. Patient demographic details, osteotomy-related complications, and the modified Harris Hip Score (mHHS) were obtained via medical notes review. Radiological indicators of hip dysplasia were assessed using preoperative and postoperative radiographs. The cumulative probability of TOA failure (progression to Tönnis grade 3 or conversion to total hip arthroplasty) was estimated using the Kaplan-Meier product-limited method. A multivariate Cox proportional hazards model was used to identify predictors of failure.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 336 - 343
1 Apr 2024
Haertlé M Becker N Windhagen H Ahmad SS

Aims

Periacetabular osteotomy (PAO) is widely recognized as a demanding surgical procedure for acetabular reorientation. Reports about the learning curve have primarily focused on complication rates during the initial learning phase. Therefore, our aim was to assess the PAO learning curve from an analytical perspective by determining the number of PAOs required for the duration of surgery to plateau and the accuracy to improve.

Methods

The study included 118 consecutive PAOs in 106 patients. Of these, 28 were male (23.7%) and 90 were female (76.3%). The primary endpoint was surgical time. Secondary outcome measures included radiological parameters. Cumulative summation analysis was used to determine changes in surgical duration. A multivariate linear regression model was used to identify independent factors influencing surgical time.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 140 - 147
1 Feb 2023
Fu Z Zhang Z Deng S Yang J Li B Zhang H Liu J

Aims

Eccentric reductions may become concentric through femoral head ‘docking’ (FHD) following closed reduction (CR) for developmental dysplasia of the hip (DDH). However, changes regarding position and morphology through FHD are not well understood. We aimed to assess these changes using serial MRI.

Methods

We reviewed 103 patients with DDH successfully treated by CR and spica casting in a single institution between January 2016 and December 2020. MRI was routinely performed immediately after CR and at the end of each cast. Using MRI, we described the labrum-acetabular cartilage complex (LACC) morphology, and measured the femoral head to triradiate cartilage distance (FTD) on the midcoronal section. A total of 13 hips with initial complete reduction (i.e. FTD < 1 mm) and ten hips with incomplete MRI follow-up were excluded. A total of 86 patients (92 hips) with a FTD > 1 mm were included in the analysis.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 310 - 316
1 Mar 2017
Hothi H Henckel J Shearing P Holme T Cerquiglini A Laura AD Atrey A Skinner J Hart A

Aims. The aim of this study was to compare the design of the generic OptiStem XTR femoral stem with the established Exeter femoral stem. Materials and Methods. We obtained five boxed, as manufactured, implants of both designs at random (ten in total). Two examiners were blinded to the implant design and independently measured the mass, volume, trunnion surface topography, trunnion roughness, trunnion cone angle, Caput-Collum-Diaphyseal (CCD) angle, femoral offset, stem length, neck length, and the width and roughness of the polished stem shaft using peer-reviewed methods. We then compared the stems using these parameters. Results. We found that the OptiStems were lighter (p < 0.001), had a rougher trunnion surface (p <  0.001) with a greater spacing and depth of the machined threads (p < 0.001), had greater trunnion cone angles (p = 0.007), and a smaller radius at the top of the trunnion (p = 0.007). There was no difference in stem volume (p = 0.643), CCD angle (p = 0.788), offset (p = 0.993), neck length (p = 0.344), stem length (p = 0.808), shaft width (p = 0.058 to 0.720) or roughness of the polished surface (p = 0.536). Conclusion. This preliminary investigation found that whilst there were similarities between the two designs, the generic OptiStem is different to the branded Exeter design. Cite this article: Bone Joint J 2017;99-B:310–16


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


Bone & Joint Open
Vol. 3, Issue 12 | Pages 991 - 997
23 Dec 2022
McPherson EJ Stavrakis AI Chowdhry M Curtin NL Dipane MV Crawford BM

Aims

Large acetabular bone defects encountered in revision total hip arthroplasty (THA) are challenging to restore. Metal constructs for structural support are combined with bone graft materials for restoration. Autograft is restricted due to limited volume, and allogenic grafts have downsides including cost, availability, and operative processing. Bone graft substitutes (BGS) are an attractive alternative if they can demonstrate positive remodelling. One potential product is a biphasic injectable mixture (Cerament) that combines a fast-resorbing material (calcium sulphate) with the highly osteoconductive material hydroxyapatite. This study reviews the application of this biomaterial in large acetabular defects.

Methods

We performed a retrospective review at a single institution of patients undergoing revision THA by a single surgeon. We identified 49 consecutive patients with large acetabular defects where the biphasic BGS was applied, with no other products added to the BGS. After placement of metallic acetabular implants, the BGS was injected into the remaining bone defects surrounding the new implants. Patients were followed and monitored for functional outcome scores, implant fixation, radiological graft site remodelling, and revision failures.


Bone & Joint Open
Vol. 3, Issue 10 | Pages 795 - 803
12 Oct 2022
Liechti EF Attinger MC Hecker A Kuonen K Michel A Klenke FM

Aims

Traditionally, total hip arthroplasty (THA) templating has been performed on anteroposterior (AP) pelvis radiographs. Recently, additional AP hip radiographs have been recommended for accurate measurement of the femoral offset (FO). To verify this claim, this study aimed to establish quantitative data of the measurement error of the FO in relation to leg position and X-ray source position using a newly developed geometric model and clinical data.

Methods

We analyzed the FOs measured on AP hip and pelvis radiographs in a prospective consecutive series of 55 patients undergoing unilateral primary THA for hip osteoarthritis. To determine sample size, a power analysis was performed. Patients’ position and X-ray beam setting followed a standardized protocol to achieve reproducible projections. All images were calibrated with the KingMark calibration system. In addition, a geometric model was created to evaluate both the effects of leg position (rotation and abduction/adduction) and the effects of X-ray source position on FO measurement.


Bone & Joint Research
Vol. 10, Issue 12 | Pages 830 - 839
15 Dec 2021
Robertson G Wallace R Simpson AHRW Dawson SP

Aims

Assessment of bone mineral density (BMD) with dual-energy X-ray absorptiometry (DXA) is a well-established clinical technique, but it is not available in the acute trauma setting. Thus, it cannot provide a preoperative estimation of BMD to help guide the technique of fracture fixation. Alternative methods that have been suggested for assessing BMD include: 1) cortical measures, such as cortical ratios and combined cortical scores; and 2) aluminium grading systems from preoperative digital radiographs. However, limited research has been performed in this area to validate the different methods. The aim of this study was to investigate the evaluation of BMD from digital radiographs by comparing various methods against DXA scanning.

Methods

A total of 54 patients with distal radial fractures were included in the study. Each underwent posteroanterior (PA) and lateral radiographs of the injured wrist with an aluminium step wedge. Overall 27 patients underwent routine DXA scanning of the hip and lumbar spine, with 13 undergoing additional DXA scanning of the uninjured forearm. Analysis of radiographs was performed on ImageJ and Matlab with calculations of cortical measures, cortical indices, combined cortical scores, and aluminium equivalent grading.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 767 - 774
1 Jul 2022
Nakashima Y Ishibashi S Kitamura K Yamate S Motomura G Hamai S Ikemura S Fujii M

Aims

Although periacetabular osteotomies are widely used for the treatment of symptomatic dysplastic hips, long-term surgical outcomes and patient-reported outcome measures (PROMs) are still unclear. Accordingly, we assessed hip survival and PROMs at 20 years after transpositional osteotomy of the acetabulum (TOA).

Methods

A total of 172 hips in 159 patients who underwent TOA were followed up at a mean of 21.02 years (16.6 to 24.6) postoperatively. Kaplan-Meier analysis was used to assess survivorship with an endpoint of total hip arthroplasty (THA). PROMs included the visual analogue scale (VAS) Satisfaction, VAS Pain, Oxford Hip Score (OHS), and Forgotten Joint Score-12 (FJS-12). Thresholds for favourable outcomes for OHS (≥ 42) and FJS-12 (≥ 51) were obtained using the receiver operating characteristic curve with VAS Satisfaction ≥ 50 and VAS Pain < 20 as anchors.


Bone & Joint Open
Vol. 3, Issue 1 | Pages 12 - 19
3 Jan 2022
Salih S Grammatopoulos G Burns S Hall-Craggs M Witt J

Aims

The lateral centre-edge angle (LCEA) is a plain radiological measure of superolateral cover of the femoral head. This study aims to establish the correlation between 2D radiological and 3D CT measurements of acetabular morphology, and to describe the relationship between LCEA and femoral head cover (FHC).

Methods

This retrospective study included 353 periacetabular osteotomies (PAOs) performed between January 2014 and December 2017. Overall, 97 hips in 75 patients had 3D analysis by Clinical Graphics, giving measurements for LCEA, acetabular index (AI), and FHC. Roentgenographical LCEA, AI, posterior wall index (PWI), and anterior wall index (AWI) were measured from supine AP pelvis radiographs. The correlation between CT and roentgenographical measurements was calculated. Sequential multiple linear regression was performed to determine the relationship between roentgenographical measurements and CT FHC.


Bone & Joint Open
Vol. 3, Issue 3 | Pages 182 - 188
1 Mar 2022
Boktor J Badurudeen A Rijab Agha M Lewis PM Roberts G Hills R Johansen A White S

Aims

In UK there are around 76,000 hip fractures occur each year 10% to 15% of which are undisplaced intracapsular. There is considerable debate whether internal fixation is the most appropriate treatment for undisplaced fractures in older patients. This study describes cannulated hip screws survivorship analysis for patients aged ≥ 60 years with undisplaced intra-capsular fractures.

Methods

This was a retrospective cohort study of consecutive patients aged ≥ 60 years who had cannulated screws fixation for Garden I and II fractures in a teaching hospital between March 2013 and March 2016. The primary outcome was further same-side hip surgery. Descriptive statistics were used and Kaplan-Meier estimates calculated for implant survival.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 839 - 844
1 Jul 2018
Ollivier M Laumonerie P LiArno S Faizan A Cavaignac E Argenson J

Aims

In patients where the proximal femur shows gross deformity due to degenerative changes or fracture, the contralateral femur is often used to perform preoperative templating for hip arthroplasty. However, femurs may not be symmetrical: the aim of this study was to determine the degree of variation between hips in healthy individuals and to determine whether it is affected by demographic parameters.

Materials and Methods

CT-scan based modelling was used to examine the pelvis and bilateral femurs of 345 patients (211 males, 134 women; mean age 62 years (standard deviation (sd) 17), mean body mass index 27 kg/m2 (sd 5)) representing a range of ethnicities. The femoral neck-shaft angle (NSA), femoral offset (FO), femoral neck version (FNV), femoral length (FL), femoral canal flare index (fCFI), and femoral head radius (FHr) were then determined for each patient. All measurements were constructed using algorithm-calculated landmarks, resulting in reproducible and consistent constructs for each specimen. We then analyzed femoral symmetry based on absolute differences (AD) and percentage asymmetry (%AS) following a previously validated method.


Bone & Joint Research
Vol. 8, Issue 6 | Pages 246 - 252
1 Jun 2019
Liddle A Webb M Clement N Green S Liddle J German M Holland J

Objectives

Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds.

Methods

A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 41 - 45
1 Jan 2019
Jones CW De Martino I D’Apolito R Nocon AA Sculco PK Sculco TP

Aims

Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years.

Materials and Methods

We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum.


The Bone & Joint Journal
Vol. 100-B, Issue 5 | Pages 570 - 578
1 May 2018
Gollwitzer H Suren C Strüwind C Gottschling H Schröder M Gerdesmeyer L Prodinger PM Burgkart R

Aims

Asphericity of the femoral head-neck junction is common in cam-type femoroacetabular impingement (FAI) and usually quantified using the alpha angle on radiographs or MRI. The aim of this study was to determine the natural alpha angle in a large cohort of patients by continuous circumferential analysis with CT.

Methods

CT scans of 1312 femurs of 656 patients were analyzed in this cross-sectional study. There were 362 men and 294 women. Their mean age was 61.2 years (18 to 93). All scans had been performed for reasons other than hip disease. Digital circumferential analysis allowed continuous determination of the alpha angle around the entire head-neck junction. All statistical tests were conducted two-sided; a p-value < 0.05 was considered statistically significant.


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1434 - 1441
1 Nov 2018
Blakeney WG Beaulieu Y Puliero B Lavigne M Roy A Massé V Vendittoli P

Aims

This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing.

Patients and Methods

Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.


Bone & Joint Research
Vol. 6, Issue 2 | Pages 113 - 122
1 Feb 2017
Scholes SC Hunt BJ Richardson VM Langton DJ Smith E Joyce TJ

Objectives

The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces.

Methods

In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1347 - 1354
1 Oct 2016
Palan J Smith MC Gregg P Mellon S Kulkarni A Tucker K Blom AW Murray DW Pandit H

Aims

Periprosthetic fracture (PF) after primary total hip arthroplasty (THA) is an uncommon but potentially devastating complication. This study aims to investigate the influence of cemented stem designs on the risk of needing a revision for a PF.

Patients and Methods

We analysed data on 257 202 primary THAs with cemented stems and 390 linked first revisions for PF recorded in the National Joint Registry (NJR) of England, Wales and Northern Ireland to determine if a cemented femoral stem brand was associated with the risk of having revision for a PF after primary THA. All cemented femoral stem brands with more than 10 000 primary operations recorded in the NJR were identified. The four most commonly used cemented femoral stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem (n = 15 113) and Charnley (n = 20 182).

We compared the revision risk ratios due to PF amongst the stems using a Poisson regression model adjusting for patient factors. Compared with the Exeter V40, the age, gender and ASA grade adjusted revision rate ratio was 3.89 for the cemented CPT stem (95% confidence interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to 1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1290 - 1297
1 Oct 2017
Devane PA Horne JG Foley G Stanley J

Aims

This paper describes the methodology, validation and reliability of a new computer-assisted method which uses models of the patient’s bones and the components to measure their migration and polyethylene wear from radiographs after total hip arthroplasty (THA).

Materials and Methods

Models of the patient’s acetabular and femoral component obtained from the manufacturer and models of the patient’s pelvis and femur built from a single computed tomography (CT) scan, are used by a computer program to measure the migration of the components and the penetration of the femoral head from anteroposterior and lateral radiographs taken at follow-up visits. The program simulates the radiographic setup and matches the position and orientation of the models to outlines of the pelvis, the acetabular and femoral component, and femur on radiographs. Changes in position and orientation reflect the migration of the components and the penetration of the femoral head. Validation was performed using radiographs of phantoms simulating known migration and penetration, and the clinical feasibility of measuring migration was assessed in two patients.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 865 - 871
1 Jul 2017
Amstutz HC Le Duff MJ Bhaurla SK

Aims

A contact patch to rim (CPR) distance of < 10 mm has been associated with edge-loading and excessive wear. However, not all arthroplasties with a low CPR distance show problems with wear. Therefore, CPR distance may not be the only variable affecting the post-operative metal ion concentrations.

Patients and Methods

We used multiple logistic regression to determine what variables differed between the patients who had high and low cobalt (CoS) and chromium (CrS) serum ion concentrations within a cohort of patients with low (< 10 mm) CPR distances. A total of 56 patients treated with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS ion studies performed more than one year after surgery. The mean age of the patients at the time of surgery was 51.7 years (29 to 70), with 38 women (68%) and 18 men (32%).


Bone & Joint Research
Vol. 6, Issue 5 | Pages 351 - 357
1 May 2017
Takahashi E Kaneuji A Tsuda R Numata Y Ichiseki T Fukui K Kawahara N

Objectives

Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep.

Methods

We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 445 - 450
1 Apr 2017
Marsh AG Nisar A El Refai M Patil S Meek RMD

Aims

The purpose of this study was to evaluate whether an innovative templating technique could predict the need for acetabular augmentation during primary total hip arthroplasty for patients with dysplastic hips.

Patients and Methods

We developed a simple templating technique to estimate acetabular component coverage at total hip arthroplasty, the True Cup: False Cup (TC:FC) ratio. We reviewed all patients with dysplastic hips who underwent primary total hip arthroplasty between 2005 and 2012. Traditional radiological methods of assessing the degree of acetabular dysplasia (Sharp’s angle, Tönnis angle, centre-edge angle) as well as the TC:FC ratio were measured from the pre-operative radiographs. A comparison of augmented and non-augmented hips was undertaken to determine any difference in pre-operative radiological indices between the two cohorts. The intra- and inter-observer reliability for all radiological indices used in the study were also calculated.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 749 - 758
1 Jun 2017
García-Rey E Cruz-Pardos A García-Cimbrelo E

Aims

To determine the effect of a change in design of a cementless ceramic acetabular component in fixation and clinical outcome after total hip arthroplasty

Patients and Methods

We compared 342 hips (302 patients) operated between 1999 and 2005 with a relatively smooth hydroxyapatite coated acetabular component (group 1), and 337 hips (310 patients) operated between 2006 and 2011 using a similar acetabular component with a macrotexture on the entire outer surface of the component (group 2). The mean age of the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70) in group 2. The mean follow-up was 12.7 years (10 to 17) for group 1 and 7.2 years (4 to 10) for group 2.


Bone & Joint Research
Vol. 6, Issue 1 | Pages 8 - 13
1 Jan 2017
Acklin YP Zderic I Grechenig S Richards RG Schmitz P Gueorguiev B

Objectives

Osteosynthesis of anterior pubic ramus fractures using one large-diameter screw can be challenging in terms of both surgical procedure and fixation stability. Small-fragment screws have the advantage of following the pelvic cortex and being more flexible.

The aim of the present study was to biomechanically compare retrograde intramedullary fixation of the superior pubic ramus using either one large- or two small-diameter screws.

Materials and Methods

A total of 12 human cadaveric hemipelvises were analysed in a matched pair study design. Bone mineral density of the specimens was 68 mgHA/cm3 (standard deviation (sd) 52). The anterior pelvic ring fracture was fixed with either one 7.3 mm cannulated screw (Group 1) or two 3.5 mm pelvic cortex screws (Group 2). Progressively increasing cyclic axial loading was applied through the acetabulum. Relative movements in terms of interfragmentary displacement and gap angle at the fracture site were evaluated by means of optical movement tracking. The Wilcoxon signed-rank test was applied to identify significant differences between the groups


Bone & Joint Research
Vol. 5, Issue 8 | Pages 338 - 346
1 Aug 2016
MacLeod AR Sullivan NPT Whitehouse MR Gill HS

Objectives

Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection.

Materials and Methods

Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 307 - 312
1 Mar 2016
Maggs JL Smeatham A Whitehouse SL Charity J Timperley AJ Gie GA

Aims

We report on the outcome of the Exeter Contemporary flanged cemented all-polyethylene acetabular component with a mean follow-up of 12 years (10 to 13.9). This study reviewed 203 hips in 194 patients. 129 hips in 122 patients are still in situ; 66 hips in 64 patients were in patients who died before ten years, and eight hips (eight patients) were revised. Clinical outcome scores were available for 108 hips (104 patients) and radiographs for 103 hips (100 patients).

Patients and Methods

A retrospective review was undertaken of a consecutive series of 203 routine primary cemented total hip arthroplasties (THA) in 194 patients.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 21 - 27
1 Jan 2016
Basheer SZ Cooper AP Maheshwari R Balakumar B Madan S

Slipped capital femoral epiphysis (SCFE) may lead to symptomatic femoroacetabular impingement (FAI). We report our experience of arthroscopic treatment, including osteochondroplasty, for the sequelae of SCFE.

Data were prospectively collected on patients undergoing arthroscopy of the hip for the sequelae of SCFE between March 2007 and February 2013, including demographic data, radiological assessment of the deformity and other factors that may influence outcome, such as the presence of established avascular necrosis. Patients completed the modified Harris hip score (mHHS) and the non-arthritic hip score (NAHS) before and after surgery.

In total, 18 patients with a mean age of 19 years (13 to 42), were included in the study. All patients presented with pain in the hip and mechanical symptoms, and had evidence of FAI (cam or mixed impingement) on plain radiographs.

The patients underwent arthroscopic osteoplasty of the femoral neck. The mean follow-up was 29 months (23 to 56).

The mean mHHS and NAHS scores improved from 56.2 (27.5 to 100.1) and 52.1 (12.5 to 97.5) pre-operatively to 75.1 (33.8 to 96.8, p = 0.01) and 73.6 (18.8 to 100, p = 0.02) at final follow-up, respectively. Linear regression analysis demonstrated a significant association between poorer outcome scores and increased time to surgery following SCFE (p < 0.05 for all parameters except baseline MHHS).

Symptomatic FAI following (SCFE) may be addressed using arthroscopic techniques, and should be treated promptly to minimise progressive functional impairment and chondrolabral degeneration.

Take home message: Arthroscopy of the hip can be used to treat femoroacetabular impingement successfully following SCFE. However, this should be performed promptly after presentation in order to prevent irreversible progression and poorer clinical outcomes.

Cite this article: Bone Joint J 2016;98-B:21–7.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 36 - 42
1 Nov 2014
Sheth NP Melnic CM Paprosky WG

Acetabular bone loss is a challenging problem facing the revision total hip replacement surgeon. Reconstruction of the acetabulum depends on the presence of anterosuperior and posteroinferior pelvic column support for component fixation and stability. The Paprosky classification is most commonly used when determining the location and degree of acetabular bone loss. Augments serve the function of either providing primary construct stability or supplementary fixation.

When a pelvic discontinuity is encountered we advocate the use of an acetabular distraction technique with a jumbo cup and modular porous metal acetabular augments for the treatment of severe acetabular bone loss and associated chronic pelvic discontinuity.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):36–42.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 47 - 51
1 Nov 2012
Gustke K

The use of short stems has become more popular with the increased interest in less invasive approaches for total hip arthroplasty. The curved broaches and stem can be inserted along a curved track to avoid the abductor attachments. Short stems have the potential of being more bone conserving by allowing for higher neck retention, maintenance of the medial greater trochanter, and preferential stress transfer to the proximal femur.

An initial experience with 500 new short stems (Fitmore, Zimmer Orthopedics) stems used for total hip arthroplasty is reported. No stems have been revised for aseptic loosening.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 917 - 923
1 Jul 2015
Singh G Nuechtern JV Meyer H Fiedler GM Awiszus F Junk-Jantsch S Bruegel M Pflueger G Lohmann CH

The peri-prosthetic tissue response to wear debris is complex and influenced by various factors including the size, area and number of particles. We hypothesised that the ‘biologically active area’ of all metal wear particles may predict the type of peri-prosthetic tissue response.

Peri-prosthetic tissue was sampled from 21 patients undergoing revision of a small diameter metal-on-metal (MoM) total hip arthroplasty (THA) for aseptic loosening. An enzymatic protocol was used for tissue digestion and scanning electron microscope was used to characterise particles. Equivalent circle diameters and particle areas were calculated. Histomorphometric analyses were performed on all tissue specimens. Aspirates of synovial fluid were collected for analysis of the cytokine profile analysis, and compared with a control group of patients undergoing primary THA (n = 11) and revision of a failed ceramic-on-polyethylene arthroplasty (n = 6).

The overall distribution of the size and area of the particles in both lymphocyte and non-lymphocyte-dominated responses were similar; however, the subgroup with lymphocyte-dominated peri-prosthetic tissue responses had a significantly larger total number of particles.

14 cytokines (interleukin (IL)-1ß, IL-2, IL-4, IL-5, IL-6, IL-10, IL-13, IL-17, interferon (IFN)-γ, and IFN-gamma-inducible protein 10), chemokines (macrophage inflammatory protein (MIP)-1α and MIP-1ß), and growth factors (granulocyte macrophage colony stimulating factor (GM-CSF) and platelet derived growth factor) were detected at significantly higher levels in patients with metal wear debris compared with the control group.

Significantly higher levels for IL-1ß, IL-5, IL-10 and GM-CSF were found in the subgroup of tissues from failed MoM THAs with a lymphocyte-dominated peri-prosthetic response compared with those without this response.

These results suggest that the ‘biologically active area’ predicts the type of peri-prosthetic tissue response. The cytokines IL-1ß, IL-5, IL-10, and GM-CSF are associated with lymphocyte-dominated tissue responses from failed small-diameter MoM THA.

Cite this article: Bone Joint J 2015;97-B:917–23.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1169 - 1177
1 Sep 2011
Underwood R Matthies A Cann P Skinner JA Hart AJ

The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component.

There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p < 0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the in vivo wear rates are discussed: these may provide an explanation as to why the ASR is more sensitive to suboptimal positioning than the BHR.


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 492 - 497
1 Apr 2015
Ike H Inaba Y Kobayashi N Yukizawa Y Hirata Y Tomioka M Saito T

In this study we used subject-specific finite element analysis to investigate the mechanical effects of rotational acetabular osteotomy (RAO) on the hip joint and analysed the correlation between various radiological measurements and mechanical stress in the hip joint.

We evaluated 13 hips in 12 patients (two men and ten women, mean age at surgery 32.0 years; 19 to 46) with developmental dysplasia of the hip (DDH) who were treated by RAO.

Subject-specific finite element models were constructed from CT data. The centre–edge (CE) angle, acetabular head index (AHI), acetabular angle and acetabular roof angle (ARA) were measured on anteroposterior pelvic radiographs taken before and after RAO. The relationship between equivalent stress in the hip joint and radiological measurements was analysed.

The equivalent stress in the acetabulum decreased from 4.1 MPa (2.7 to 6.5) pre-operatively to 2.8 MPa (1.8 to 3.6) post-operatively (p < 0.01). There was a moderate correlation between equivalent stress in the acetabulum and the radiological measurements: CE angle (R = –0.645, p < 0.01); AHI (R = –0.603, p < 0.01); acetabular angle (R = 0.484, p = 0.02); and ARA (R = 0.572, p < 0.01).

The equivalent stress in the acetabulum of patients with DDH decreased after RAO. Correction of the CE angle, AHI and ARA was considered to be important in reducing the mechanical stress in the hip joint.

Cite this article: Bone Joint J 2015;97-B:492–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1309 - 1316
1 Oct 2007
Gosvig KK Jacobsen S Palm H Sonne-Holm S Magnusson E

Femoroacetabular cam impingement is thought to be a cause of premature osteoarthritis of the hip.

The presence of cam malformation was determined in 2803 standardised anteroposterior (AP) pelvic radiographs from the Copenhagen Osteoarthritis Study by measuring the alpha (α) angle and the triangular index, a new measure of asphericity of the femoral head. In addition, the α-angle and the triangular index were assessed on the AP and lateral hip radiographs of 82 men and 82 women randomly selected from patients scheduled for total hip replacement (THR). The influence of varying femoral rotation on the α angle and the triangular index was also determined in femoral specimens under experimental conditions.

From the 2803 radiographs the mean AP α-angle was 55° (30° to 100°) in men and 45° (34° to 108°) in women. Approximately 6% of men and 2% of women had cam malformation. The α-angle and triangular index were highly inter-related. Of those patients scheduled for THR, 36 men (44%) and 28 women (35%) had cam malformation identifiable on the AP radiographs. The triangular index proved to be more reliable in detecting cam malformation when the hip was held in varying degrees of rotation.

The combination of the α-angle and the triangular index will allow examination of historical radiographs for epidemiological purposes in following the natural history of the cam deformity.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1441 - 1448
1 Nov 2014
Bali K Railton P Kiefer GN Powell JN

We report the clinical and radiological outcome of subcapital osteotomy of the femoral neck in the management of symptomatic femoroacetabular impingement (FAI) resulting from a healed slipped capital femoral epiphysis (SCFE). We believe this is only the second such study in the literature.

We studied eight patients (eight hips) with symptomatic FAI after a moderate to severe healed SCFE. There were six male and two female patients, with a mean age of 17.8 years (13 to 29).

All patients underwent a subcapital intracapsular osteotomy of the femoral neck after surgical hip dislocation and creation of an extended retinacular soft-tissue flap. The mean follow-up was 41 months (20 to 84). Clinical assessment included measurement of range of movement, Harris Hip Score (HHS) and Western Ontario and McMaster Universities Osteoarthritis score (WOMAC). Radiological assessment included pre- and post-operative calculation of the anterior slip angle (ASA) and lateral slip angle (LSA), the anterior offset angle (AOA) and centre head–trochanteric distance (CTD). The mean HHS at final follow-up was 92.5 (85 to 100), and the mean WOMAC scores for pain, stiffness and function were 1.3 (0 to 4), 1.4 (0 to 6) and 3.6 (0 to 19) respectively. There was a statistically significant improvement in all the radiological measurements post-operatively. The mean ASA improved from 36.6° (29° to 44°) to 10.3° (5° to 17°) (p <  0.01). The mean LSA improved from 36.6° (31° to 43°) to 15.4° (8° to 21°) (p < 0.01). The mean AOA decreased from 64.4° (50° to 78°) 32.0° (25° to 39°) post-operatively (p < 0.01). The mean CTD improved from -8.2 mm (-13.8 to +3.1) to +2.8 mm (-7.6 to +11.0) (p < 0.01). Two patients underwent further surgery for nonunion. No patient suffered avascular necrosis of the femoral head.

Subcapital osteotomy for patients with a healed SCFE is more challenging than subcapital re-orientation in those with an acute or sub-acute SCFE and an open physis. An effective correction of the deformity, however, can be achieved with relief of symptoms related to impingement.

Cite this article: Bone Joint J 2014;96-B:1441–8.


Bone & Joint Research
Vol. 1, Issue 3 | Pages 25 - 30
1 Mar 2012
Wroblewski BM Siney PD Fleming PA

Objectives

Metal-on-metal (MoM) hip resurfacing was introduced into clinical practice because it was perceived to be a better alternative to conventional total hip replacement for young and active patients. However, an increasing number of reports of complications have arisen focusing on design and orientation of the components, the generation of metallic wear particles and serum levels of metallic ions. The procedure introduced a combination of two elements: large-dimension components and hard abrasive particles of metal wear. The objective of our study was to investigate the theory that microseparation of the articular surfaces draws in a high volume of bursal fluid and its contents into the articulation, and at relocation under load would generate high pressures of fluid ejection, resulting in an abrasive water jet.

Methods

This theoretical concept using MoM resurfacing components (head diameter 55 mm) was modelled mathematically and confirmed experimentally using a material-testing machine that pushed the head into the cup at a rate of 1000 mm/min until fully engaged.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 312 - 318
1 Mar 2014
Meermans G Van Doorn WJ Koenraadt K Kats J

The orientation of the acetabular component can influence both the short- and long-term outcomes of total hip replacement (THR). We performed a prospective, randomised, controlled trial of two groups, comprising of 40 patients each, in order to compare freehand introduction of the component with introduction using the transverse acetabular ligament (TAL) as a reference for anteversion. Anteversion and inclination were measured on pelvic radiographs.

With respect to anteversion, in the freehand group 22.5% of the components were outside the safe zone versus 0% in the transverse acetabular ligament group (p = 0.002). The mean angle of anteversion in the freehand group was 21° (2° to 35°) which was significantly higher compared with 17° (2° to 25°) in the TAL group (p = 0.004). There was a significant difference comparing the variations of both groups (p = 0.008).

With respect to inclination, in the freehand group 37.5% of the components were outside the safe zone versus 20% in the TAL group (p = 0.14). There was no significant difference regarding the accuracy or variation of the angle of inclination when comparing the two groups.

The transverse acetabular ligament may be used to obtain the appropriate anteversion when introducing the acetabular component during THR, but not acetabular component inclination.

Cite this article: Bone Joint J 2014;96-B:312–18.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 580 - 589
1 May 2014
Nakahara I Takao M Sakai T Miki H Nishii T Sugano N

To confirm whether developmental dysplasia of the hip has a risk of hip impingement, we analysed maximum ranges of movement to the point of bony impingement, and impingement location using three-dimensional (3D) surface models of the pelvis and femur in combination with 3D morphology of the hip joint using computer-assisted methods. Results of computed tomography were examined for 52 hip joints with DDH and 73 normal healthy hip joints. DDH shows larger maximum extension (p = 0.001) and internal rotation at 90° flexion (p < 0.001). Similar maximum flexion (p = 0.835) and external rotation (p = 0.713) were observed between groups, while high rates of extra-articular impingement were noticed in these directions in DDH (p < 0.001). Smaller cranial acetabular anteversion (p = 0.048), centre-edge angles (p < 0.001), a circumferentially shallower acetabulum, larger femoral neck anteversion (p < 0.001), and larger alpha angle were identified in DDH. Risk of anterior impingement in retroverted DDH hips is similar to that in retroverted normal hips in excessive adduction but minimal in less adduction. These findings might be borne in mind when considering the possibility of extra-articular posterior impingement in DDH being a source of pain, particularly for patients with a highly anteverted femoral neck.

Cite this article: Bone Joint J 2014;96-B:580–9.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 449 - 454
1 Apr 2014
Laborie LB Lehmann TG Engesæter IØ Sera F Engesæter LB Rosendahl K

We report on gender-specific reference intervals of the alpha angle and its association with other qualitative cam-type findings in femoroacetabular impingement at the hip, according to a population-based cohort of 2038 19-year-olds, 1186 of which were women (58%). The alpha angle was measured on standardised frog-leg lateral and anteroposterior (AP) views using digital measurement software, and qualitative cam-type findings were assessed subjectively on both views by independent observers. In all, 2005 participants (837 men, 1168 women, mean age 18.6 years (17.2 to 20.1) were included in the analysis. For the frog-leg view, the mean alpha angle (right hip) was 47° (26 to 79) in men and 42° (29 to 76) in women, with 97.5 percentiles of 68° and 56°, respectively. For the AP view, the mean values were 62° (40 to 105) and 52° (36 to 103) for men and women, respectively, with 97.5 percentiles of 93° and 94°. Associations between higher alpha angles and all qualitative cam-type findings were seen for both genders on both views. The reference intervals presented for the alpha angle in this cross-sectional study are wide, especially for the AP view, with higher mean values for men than women on both views.

Cite this article: Bone Joint J 2014;96-B:449–54.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 881 - 885
1 Jul 2011
Cobb JP Davda K Ahmad A Harris SJ Masjedi M Hart AJ

Large-head metal-on-metal total hip replacement has a failure rate of almost 8% at five years, three times the revision rate of conventional hip replacement. Unexplained pain remains a feature of this type of arthroplasty.

All designs of the femoral component of large-head metal-on-metal total hip replacements share a unique characteristic: a subtended angle of 120° defining the proportion of a sphere that the head represents. Using MRI, we measured the contact area of the iliopsoas tendon on the femoral head in sagittal reconstruction of 20 hips of patients with symptomatic femoroacetabular impingement. We also measured the articular extent of the femoral head on 40 normal hips and ten with cam-type deformities. Finally, we performed virtual hip resurfacing on normal and cam-type hips, avoiding overhang of the metal rim inferomedially.

The articular surface of the femoral head has a subtended angle of 120° anteriorly and posteriorly, but only 100° medially. Virtual surgery in a normally shaped femoral head showed a 20° skirt of metal protruding medially where iliopsoas articulates.

The excessive extent of the large-diameter femoral components may cause iliopsoas impingement independently of the acetabular component. This may be the cause of postoperative pain with these implants.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 147 - 156
1 Feb 2014
Rajpura A Kendoff D Board TN

We reviewed the literature on the currently available choices of bearing surface in total hip replacement (THR). We present a detailed description of the properties of articulating surfaces review the understanding of the advantages and disadvantages of existing bearing couples. Recent technological developments in the field of polyethylene and ceramics have altered the risk of fracture and the rate of wear, although the use of metal-on-metal bearings has largely fallen out of favour, owing to concerns about reactions to metal debris. As expected, all bearing surface combinations have advantages and disadvantages. A patient-based approach is recommended, balancing the risks of different options against an individual’s functional demands.

Cite this article: Bone Joint J 2014;96-B:147–56.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1303 - 1308
1 Oct 2007
Nixon M Taylor G Sheldon P Iqbal SJ Harper W

We matched 78 patients with a loose cemented Charnley Elite Plus total hip replacement (THR) by age, gender, race, prosthesis and time from surgery with 49 patients with a well-fixed stable hip replacement, to determine if poor bone quality predisposes to loosening. Clinical, radiological, biomechanical and bone mineral density indicators of bone quality were assessed.

Patients with loose replacements had more pain, were more likely to have presented with atrophic arthritis and to have a history of fragility fracture, narrower femoral cortices and lower peri-prosthetic or lumbar spine bone mineral density (all t-test, p < 0.01). They also tended to be smokers (chi-squared test, p = 0.08). Vitamin-D deficiency was common, but not significantly different between the two groups (t-test, p = 0.31)

In this series of cemented hip replacements performed between 1994 and 1998, aseptic loosening was associated with poor bone quality. Patients with a THR should be screened for osteoporosis and have regular radiological surveillance.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1203 - 1208
1 Sep 2010
Brunner A Hamers AT Fitze M Herzog RF

The β-angle is a radiological tool for measuring the distance between the pathological head-neck junction and the acetabular rim with the hip in 90° of flexion in patients with femoroacetabular impingement. Initially it was measured using an open-chamber MRI. We have developed a technique to measure this angle on plain radiographs. Correlation analysis was undertaken to determine the relationship between the range of movement and the β-angle in 50 patients with femoroacetabular impingement and 50 asymptomatic control subjects. Inter- and intra-observer reliability of the β-angle was also evaluated.

Patients with femoroacetabular impingement had a significantly smaller (p < 0.001) mean β-angle (15.6°, 95% confidence interval (CI) 13.3 to 17.7) compared with the asymptomatic group (38.7°, 95% CI 36.5 to 41.0). Correlation between internal rotation and the β-angle was high in the impingement group and moderate in the asymptomatic group. The β-angle had excellent inter- and intra-observer reliability in both groups.

Our findings suggest that the measurement of the β-angle on plain radiography may represent a valid, reproducible and cost-effective alternative to open MRI in the assessment of the pathological bony anatomy in patients with cam, pincer and mixed femoroacetabular impingement.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1487 - 1493
1 Nov 2012
Wensaas A Gunderson RB Svenningsen S Terjesen T

Deformity after slipped upper femoral epiphysis (SUFE) can cause cam-type femoroacetabular impingement (FAI) and subsequent osteoarthritis (OA). However, there is little information regarding the radiological assessment and clinical consequences at long-term follow-up. We reviewed 36 patients (43 hips) previously treated by in situ fixation for SUFE with a mean follow-up of 37 years (21 to 50). Three observers measured the femoral head ratio (FHR), lateral femoral head ratio (LFHR), α-angle on anteroposterior (AP) and frog-leg lateral views, and anterior femoral head–neck offset ratio (OSR). A Harris hip score < 85 and/or radiologically diagnosed osteoarthritis (OA) was classified as a poor outcome. Patients with SUFE had significantly higher FHR, LFHR and α-angles and lower OSR than a control group of 22 subjects (35 hips) with radiologically normal hips. The interobserver agreement was less, with wider limits of agreement (LOA), in hips with previous SUFE than the control group. At long-term follow-up abnormal α-angles correlated with poor outcome, whereas FHR, LFHR and OSR did not.

We conclude that persistent deformity with radiological cam FAI after SUFE is associated with poorer clinical and radiological long-term outcome. Although the radiological measurements had quite wide limits of agreement, they are useful for the diagnosis of post-slip deformities in clinical practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 901 - 907
1 Jul 2012
Esposito CI Walter WL Roques A Tuke MA Zicat BA Walsh WR Walter WK

We analysed 54 alumina ceramic-on-ceramic bearings from total hip replacements retrieved at one centre after a mean duration of 3.5 years (0.2 to 10.6) in situ. These implants were obtained from 54 patients (16 men and 38 women) with a mean age of 67 years (33 to 88) who underwent revision for a variety of reasons. Posterior edge loading was found in the majority of these retrievals (32 out of 54). Anterosuperior edge loading occurred less often but produced a higher rate of wear. Stripe wear on the femoral heads had a median volumetric wear rate of 0.2 mm3/year (0 to 7.2). The wear volume on the femoral heads corresponded to the width of edge wear on the matching liner. Anteversion of the acetabular component was found to be a more important determinant than inclination for wear in ceramic bearings. Posterior edge loading may be considered to be a normal occurrence in ceramic-on-ceramic bearings, with minimal clinical consequences. Edge loading should be defined as either anterosuperior or posterior, as each edge loading mechanism may result in different clinical implications.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 27 - 33
1 Jan 2011
Treacy RBC McBryde CW Shears E Pynsent PB

We report the survival, radiological and functional outcomes of a single surgeon series of his first 144 consecutive Birmingham hip resurfacing procedures (130 patients) at a minimum of ten years. There were ten revisions during this time. Although no patients were lost to follow-up some did not complete the scoring assessment or undergo radiological assessment at ten years.

The ten-year survival for male patients was 98.0% (95% confidence interval 95.2 to 100). The ten-year survival for the total cohort with aseptic revision as the endpoint was 95.5% (95% confidence interval 91.8 to 99.0) and including revisions for sepsis was 93.5% (95% confidence interval 89.2 to 97.6). The median modified Oxford hip score at ten years was 4.2% (interquartile range 0 to 19) and the median University of California, Los Angeles score was 7.0 (interquartile range 5.0 to 8.0).

This study confirms the midterm reports that metal-on-metal hip resurfacing using the Birmingham Hip provides a durable alternative to total hip replacement, particularly in younger male patients wishing to maintain a high level of function, with low risk of revision for at least ten years.