The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1011 - 1016
1 Aug 2011
Langton DJ Jameson SS Joyce TJ Gandhi JN Sidaginamale R Mereddy P Lord J Nargol AVF

There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent National Joint Registry data have shown clear differences in the rates of failure of different designs of hip resurfacing. Our aim was to update the failure rates related to metal debris for the Articular Surface Replacement (ASR). A total of 505 of these were implanted. Kaplan-Meier analysis showed a failure rate of 25% at six years for the ASR resurfacing and of 48.8% for the ASR total hip replacement (THR). Of 257 patients with a minimum follow-up of two years, 67 (26.1%) had a serum cobalt concentration which was greater than 7 μg/l. Co-ordinate measuring machine analysis of revised components showed that all patients suffering adverse tissue reactions in the resurfacing group had abnormal wear of the bearing surfaces. Six THR patients had relatively low rates of articular wear, but were found to have considerable damage at the trunion-taper interface. Our results suggest that wear at the modular junction is an important factor in the development of adverse tissue reactions after implantation of a large-diameter MoM THR


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 4 | Pages 504 - 509
1 May 2004
Lie SA Havelin LI Furnes ON Engesæter LB Vollset SE

We present the results for 4762 revision total hip arthroplasties with no previous infection in the hip, which were reported to the Norwegian Arthroplasty Register between 1987 and 2003. The ten-year failure rate for revised prostheses was 26% (95% CI 25 to 26). Cox regression analyses were undertaken separately for acetabular and femoral revision components. Cemented revision components without allograft was the reference category. For acetabular components, we found a significantly reduced risk of failure for uncemented revisions both with (relative risk (RR) = 0.66; 95% CI 0.43 to 0.99) and without (RR = 0.37; 95% CI 0.22 to 0.61) allograft. For femoral components, we found a significantly reduced risk of failure for uncemented revisions, both with (RR = 0.27; 95% CI 0.16 to 0.46) and without (RR = 0.22; 95% CI 0.11 to 0.46) unimpacted allograft. This reduced risk of failure also applied to cemented revision components with allograft (RR = 0.53; 95% CI 0.33 to 0.84) and with impaction bone grafting (RR = 0.34; 95% CI 0.19 to 0.62). Revision prostheses have generally inferior results when compared with primary prostheses. Recementation without allograft, and uncemented revision with bone impaction, were associated with worse results than the other revision techniques which we studied


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 608 - 615
1 May 2011
Bolland BJRF Culliford DJ Langton DJ Millington JPS Arden NK Latham JM

This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


Bone & Joint Open
Vol. 4, Issue 5 | Pages 329 - 337
8 May 2023
Khan AQ Chowdhry M Sherwani MKA McPherson EJ

Aims. Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). Methods. In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years. Results. Mean follow-up was 4.6 years (4.1 to 5.0) in the CCS group and 5.5 years (5.25 to 5.75) in the VOOF group. The mean Harris Hip Score at two-year follow-up was 83.85 in the CCS group versus 88.00 in the VOOF group (p < 0.001). At the latest follow-up, all-cause failure rate was 29.1% in the CCS group and 11.7% in the VOOF group (p = 0.003). The total cost of the VOOF technique was 7.2% of a THA, and total cost of the CCS technique was 6.3% of a THA. Conclusion. The VOOF technique decreased all-cause failure rate compared to CCS. The total cost of VOOF was 13.5% greater than CCS, but 92.8% less than a THA. Increased cost of VOOF was considered acceptable to all patients in this series. VOOF technique provides a reasonable alternative to THA in patients who cannot afford a THA procedure. Cite this article: Bone Jt Open 2023;4(5):329–337


Bone & Joint Open
Vol. 3, Issue 6 | Pages 485 - 494
13 Jun 2022
Jaubert M Le Baron M Jacquet C Couvreur A Fabre-Aubrespy M Flecher X Ollivier M Argenson J

Aims. Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. Methods. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test. Results. Our staged exchange practice shows poor results, which means a 42% mechanical spacer failure rate, and a 20% recurrent infection rate over the two years average follow-up period. Moreover, we found a high rate of spacer dislocation (23%) and a low rate of spacer fracture (8%) compared to the previous literature. Our findings stress that the majority of spacer complications and failures is reflecting a population with high comorbid burden, highlighted by the American Society of Anesthesiology grade, Charlson Comorbidity Index, and Lee score associations, as well as the cardiac, pulmonary, kidney, or hepatic chronic conditions. Conclusion. Our experience of a two-stage hip exchange revision noted important complication rates associated with high failure rates of polymethylmethacrylate spacers. These findings must be interpreted in the light of the patient’s comorbidity profiles, as the elective population for staged exchange has an increasing comorbid burden leading to poor results. In order to provide better results for this specific population, our conclusion suggests that comparative strategy studies are required to improve our therapeutic indication. Cite this article: Bone Jt Open 2022;3(6):485–494


Bone & Joint Open
Vol. 4, Issue 3 | Pages 182 - 187
14 Mar 2023
Sheridan GA Hanlon M Welch-Phillips A Spratt K Hagan R O'Byrne JM Kenny PJ Kurmis AP Masri BA Garbuz DS Hurson CJ

Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings. Methods. A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size. Results. Using a previously-reported ALVAL classification, the majority of pseudotumour-positive caseswere found to have intermediate-grade group 2 (n = 10; 63%) or group 3 (n = 4; 25%) histological findings. Two further patients (13%) had high-grade group 4 lesions. HLA-DQB1*05:03:01 (p = 0.0676) and HLA-DRB1*14:54:01 (p = 0.0676) alleles were significantly associated with a higher risk of pseudotumour formation, while HLA-DQA1*03:01:01 (p = 0.0240), HLA-DRB1*04:04:01 (p = 0.0453), HLA-C*01:02:01 (p = 0.0453), and HLA-B*27:05:02 (p = 0.0855) were noted to confer risk reduction. Conclusion. These findings confirm the association between specific HLA genotypes and the risk of pseudotumour development around MoM hip resurfacings. Specifically, the two ‘at risk’ alleles (DQB1*05:03:01 and DRB1*14:54:01) may hold clinical value in preoperative screening and prospective surgical decision-making. Cite this article: Bone Jt Open 2023;4(3):182–187


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 859 - 866
1 Jul 2022
Innocenti M Smulders K Willems JH Goosen JHM van Hellemondt G

Aims. The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). Methods. We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups. Results. The indication for revision influenced PROMs improvement over time. This finding mainly reflected preoperative differences between the groups, but diminished between the first and second postoperative years. Preoperatively, patients revised due to infection and aseptic loosening had a lower mOHS than patients with other indications for revision. Pain scores at baseline were highest in patients being revised for dislocation. Infection and aseptic loosening groups showed marked changes over time in both mOHS and EQ-5D-3L. Overall complications and re-revision rates were 35.4% and 9.7% respectively, with no differences between the groups (p = 0.351 and p = 0.470, respectively). Conclusion. Good outcomes were generally obtained regardless of the reason for revision, with patients having the poorest preoperative scores exhibiting the greatest improvement in PROMs. Furthermore, overall complication and reoperation rates were in line with previous reports and did not differ between different indications for rTHA. Cite this article: Bone Joint J 2022;104-B(7):859–866


Bone & Joint Open
Vol. 3, Issue 3 | Pages 182 - 188
1 Mar 2022
Boktor J Badurudeen A Rijab Agha M Lewis PM Roberts G Hills R Johansen A White S

Aims. In UK there are around 76,000 hip fractures occur each year 10% to 15% of which are undisplaced intracapsular. There is considerable debate whether internal fixation is the most appropriate treatment for undisplaced fractures in older patients. This study describes cannulated hip screws survivorship analysis for patients aged ≥ 60 years with undisplaced intra-capsular fractures. Methods. This was a retrospective cohort study of consecutive patients aged ≥ 60 years who had cannulated screws fixation for Garden I and II fractures in a teaching hospital between March 2013 and March 2016. The primary outcome was further same-side hip surgery. Descriptive statistics were used and Kaplan-Meier estimates calculated for implant survival. Results. A total of 114 operations were performed on 112 patients with a mean age of 80.2 years (SD 8.9). The 30-day and one-year mortality were 1% (n = 1) and 13% (n = 15), respectively. Median follow-up was 6.6 years (interquartile range 6.0 to 7.3). Kaplan-Meier estimates showed a survivorship of 95% at one year and 90% at five years (95% confidence interval 84% to 95%) for cannulated screws. Nine patients underwent further hip surgery: four revision to total hip arthroplasty, one revision to hemiarthroplasty, three removals of screws, and one haematoma washout. Posterior tilt was assessable in 106 patients; subsequent surgery was required in two of the six patients identified with a posterior angle > 20° (p = 0.035 vs angle < 20°). Of the 100 patients with angle < 20°, five-year survivorship was 91%, with seven patients requiring further surgery. Conclusion. This study of cannulated hip screw fixation for undisplaced fractures in patients aged ≥ 60 years reveals a construct survivorship without further operation of 90% at five years. Cannulated screws can be considered a safe reliable treatment option for Garden I and II fractures. Caution should be taken if posterior tilt angle on lateral view exceeds 20°, due to a higher failure rate and reoperation, and considered for similar management to Garden III and IV injuries. Cite this article: Bone Jt Open 2022;3(3):182–188


Bone & Joint Open
Vol. 3, Issue 1 | Pages 61 - 67
18 Jan 2022
van Lingen CP Ettema HB Bosker BH Verheyen CCPM

Aims. Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. Methods. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up. Results. Mean follow-up was 10.9 years (10 to 12), with a cumulative survival rate of 82.4%. Reason for revision was predominantly pseudotumour formation (68%), apart from loosening, pain, infection, and osteolysis. The prevalence of pseudotumour formation around the prostheses was 41%, while our previous report of this cohort (with a mean follow-up of 3.6 years) revealed a 39% prevalence. The ten-year revision-free survival with pseudotumour was 66.7% and without pseudotumour 92.4% (p < 0.05). There was poor discriminatory ability for cobalt for pseudotumour formation. Conclusion. This prospective study reports a minimum ten-year follow-up of large-head MoM THA. Revision rates are high, with the main reason being the sequelae of pseudotumour formation, which were rarely observed after five years of implantation. Blood ion measurements show limited discriminatory capacity in diagnosing pseudotumour formation. Our results evidence that an early comprehensive follow-up strategy is essential for MoM THA to promptly identify and manage early complications and revise on time. After ten years follow-up, we do not recommend continuing routine CT scanning or whole cobalt blood measurements, but instead enrolling these patients in routine follow-up protocols for THA. Cite this article: Bone Jt Open 2022;3(1):61–67


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1247 - 1253
1 Jul 2021
Slullitel PA Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims. There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA. Methods. We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality. Results. The cumulative incidence of septic failure was 8% (95% confidence interval (CI) 3.5 to 15) at one year, 13.8% (95% CI 7.6 to 22) at two years, and 19.7% (95% CI 12 to 28.6) at five and ten years of follow-up. A femoral bone defect worse than Paprosky IIIA (hazard ratio (HR) 13.58 (95% CI 4.86 to 37.93); p < 0.001) and obesity (BMI ≥ 30 kg/m. 2. ; HR 3.88 (95% CI 1.49 to 10.09); p = 0.005) were significantly associated with septic failure. Instability and periprosthetic fracture were the most common reasons for mechanical failure (5.7% and 4.5%, respectively). The cumulative incidence of aseptic failure was 2% (95% CI 0.4 to 7) at two years, 9% (95% CI 4 to 17) at five years, and 12% (95% CI 5 to 22) at ten years. A previous revision to treat PJI was significantly associated with non-septic failure (HR 9.93 (95% CI 1.77 to 55.46); p = 0.009). At the five-year timepoint, 93% of the patients were alive (95% CI 84% to 96%), which fell to 86% (95% CI 75% to 92%) at ten-year follow-up. Conclusion. Massive femoral bone loss was associated with greater chances of developing a further septic failure. All septic failures occurred within the first five years following the one-stage exchange. Surgeons should be aware of instability and periprosthetic fracture being potential causes of further aseptic revision surgery. Cite this article: Bone Joint J 2021;103-B(7):1247–1253


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 592 - 600
1 May 2017
Matharu GS Nandra RS Berryman F Judge A Pynsent PB Dunlop DJ

Aims. To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure. Patients and Methods. We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression. Results. Failure occurred in 97 hips (17.0%). The ten-year cumulative failure rate was 27.1% (95% confidence interval (CI) 21.6 to 33.7). Primary implantation from 2006 onwards (hazard ratio (HR) 4.30; 95% CI 1.82 to 10.1; p = 0.001) and bilateral MoM hip arthroplasty (HR 1.59; 95% CI 1.03 to 2.46; p = 0.037) predicted failure. The effect of implantation year on failure varied over time. From four years onwards following surgery, hips implanted since 2006 had significantly higher failure rates (eight years 28.3%; 95% CI 23.1 to 34.5) compared with hips implanted before 2006 (eight years 6.3%; 95% CI 2.4 to 15.8) (HR 15.2; 95% CI 2.11 to 110.4; p = 0.007). Conclusion. We observed that 36 mm MoM Pinnacle THAs have an unacceptably high ten-year failure rate, especially if implanted from 2006 onwards or in bilateral MoM hip patients. Our findings regarding implantation year and failure support recent concerns about the device manufacturing process. We recommend all patients undergoing implantation since 2006 and those with bilateral MoM hips undergo regular investigation, regardless of symptoms. Cite this article: Bone Joint J 2017;99-B:592–600


Bone & Joint Open
Vol. 4, Issue 2 | Pages 53 - 61
1 Feb 2023
Faraj S de Windt TS van Hooff ML van Hellemondt GG Spruit M

Aims

The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA).

Methods

Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 801 - 807
23 Oct 2023
Walter N Szymski D Kurtz SM Lowenberg DW Alt V Lau EC Rupp M

Aims

This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures, intertrochanteric fractures, and subtrochanteric fractures in the elderly USA population? and 2) Which factors influence adverse outcomes?

Methods

Proximal femoral fractures occurred between 1 January 2009 and 31 December 2019 were identified from the Medicare Physician Service Records Data Base. The Kaplan-Meier method with Fine and Gray sub-distribution adaptation was used to determine rates for nonunion, infection, and mechanical complications. Semiparametric Cox regression model was applied incorporating 23 measures as covariates to identify risk factors.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 775 - 782
1 Jul 2023
Koper MC Spek RWA Reijman M van Es EM Baart SJ Verhaar JAN Bos PK

Aims

The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate.

Methods

A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.


Bone & Joint Open
Vol. 5, Issue 11 | Pages 971 - 976
5 Nov 2024
Baker G Hill J O'Neill F McChesney J Stevenson M Beverland D

Aims

In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes.

Methods

A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 946 - 952
1 Sep 2023
Dhawan R Young DA Van Eemeren A Shimmin A

Aims

The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit.

Methods

Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.


Bone & Joint Open
Vol. 5, Issue 1 | Pages 28 - 36
18 Jan 2024
Selmene MA Moreau PE Zaraa M Upex P Jouffroy P Riouallon G

Aims

Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications.

Methods

Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims

Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up.

Methods

This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 511 - 517
1 May 2023
Petrie MJ Panchani S Al-Einzy M Partridge D Harrison TP Stockley I

Aims

The duration of systemic antibiotic treatment following first-stage revision surgery for periprosthetic joint infection (PJI) after total hip arthroplasty (THA) is contentious. Our philosophy is to perform an aggressive debridement, and to use a high local concentration of targeted antibiotics in cement beads and systemic prophylactic antibiotics alone. The aim of this study was to assess the success of this philosophy in the management of PJI of the hip using our two-stage protocol.

Methods

The study involved a retrospective review of our prospectively collected database from which we identified all patients who underwent an intended two-stage revision for PJI of the hip. All patients had a diagnosis of PJI according to the major criteria of the Musculoskeletal Infection Society (MSIS) 2013, a minimum five-year follow-up, and were assessed using the MSIS working group outcome-reporting tool. The outcomes were grouped into ‘successful’ or ‘unsuccessful’.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 136 - 143
1 Feb 2024
van der Lelij TJN Marang-van de Mheen PJ Kaptein BL Koster LA Ljung P Nelissen RGHH Toksvig-Larsen S

Aims

The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface.

Methods

Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 435 - 441
1 May 2024
Angelomenos V Mohaddes M Kärrholm J Malchau H Shareghi B Itayem R

Aims

Refobacin Bone Cement R and Palacos R + G bone cement were introduced to replace the original cement Refobacin Palacos R in 2005. Both cements were assumed to behave in a biomechanically similar fashion to the original cement. The primary aim of this study was to compare the migration of a polished triple-tapered femoral stem fixed with either Refobacin Bone Cement R or Palacos R + G bone cement. Repeated radiostereometric analysis was used to measure migration of the femoral head centre. The secondary aims were evaluation of cement mantle, stem positioning, and patient-reported outcome measures.

Methods

Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.


Bone & Joint Open
Vol. 5, Issue 6 | Pages 514 - 523
24 Jun 2024
Fishley W Nandra R Carluke I Partington PF Reed MR Kramer DJ Wilson MJ Hubble MJW Howell JR Whitehouse SL Petheram TG Kassam AM

Aims

In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.

Methods

Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).


Bone & Joint Open
Vol. 3, Issue 11 | Pages 867 - 876
10 Nov 2022
Winther SS Petersen M Yilmaz M Kaltoft NS Stürup J Winther NS

Aims

Pelvic discontinuity is a rare but increasingly common complication of total hip arthroplasty (THA). This single-centre study evaluated the performance of custom-made triflange acetabular components in acetabular reconstruction with pelvic discontinuity by determining: 1) revision and overall implant survival rates; 2) discontinuity healing rate; and 3) Harris Hip Score (HHS).

Methods

Retrospectively collected data of 38 patients (39 hips) with pelvic discontinuity treated with revision THA using a custom-made triflange acetabular component were analyzed. Minimum follow-up was two years (mean 5.1 years (2 to 11)).


Bone & Joint Research
Vol. 11, Issue 12 | Pages 881 - 889
1 Dec 2022
Gómez-Barrena E Padilla-Eguiluz N López-Marfil M Ruiz de la Reina R

Aims

Successful cell therapy in hip osteonecrosis (ON) may help to avoid ON progression or total hip arthroplasty (THA), but the achieved bone regeneration is unclear. The aim of this study was to evaluate amount and location of bone regeneration obtained after surgical injection of expanded autologous mesenchymal stromal cells from the bone marrow (BM-hMSCs).

Methods

A total of 20 patients with small and medium-size symptomatic stage II femoral head ON treated with 140 million BM-hMSCs through percutaneous forage in the EudraCT 2012-002010-39 clinical trial were retrospectively evaluated through preoperative and postoperative (three and 12 months) MRI. Then, 3D reconstruction of the original lesion and the observed postoperative residual damage after bone regeneration were analyzed and compared per group based on treatment efficacy.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 443 - 449
1 Apr 2018
Kalsbeek JH van Walsum ADP Vroemen JPAM Janzing HMJ Winkelhorst JT Bertelink BP Roerdink WH

Aims. The objective of this study was to investigate bone healing after internal fixation of displaced femoral neck fractures (FNFs) with the Dynamic Locking Blade Plate (DLBP) in a young patient population treated by various orthopaedic (trauma) surgeons. Patients and Methods. We present a multicentre prospective case series with a follow-up of one year. All patients aged ≤ 60 years with a displaced FNF treated with the DLBP between 1st August 2010 and December 2014 were included. Patients with pathological fractures, concomitant fractures of the lower limb, symptomatic arthritis, local infection or inflammation, inadequate local tissue coverage, or any mental or neuromuscular disorder were excluded. Primary outcome measure was failure in fracture healing due to nonunion, avascular necrosis, or implant failure requiring revision surgery. Results. In total, 106 consecutive patients (mean age 52 years, range 23 to 60; 46% (49/106) female) were included. The failure rate was 14 of 106 patients (13.2%, 95% confidence interval (CI) 7.1 to 19.9). Avascular necrosis occurred in 11 patients (10.4%), nonunion in six (5.6%), and loss of fixation in two (1.9%). Conclusion. The rate of fracture healing after DLBP fixation of displaced femoral neck fracture in young patients is promising and warrants further investigation by a randomized trial to compare the performance against other contemporary methods of fixation. Cite this article: Bone Joint J 2018;100-B:443–9


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1032 - 1038
1 Sep 2022
García-Rey E Cruz-Pardos A Saldaña L

Aims

A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade.

Methods

Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1039 - 1046
1 Sep 2022
Özdemir E Kuijpers MFL Visser J Schreurs BW Rijnen WHC

Aims

The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique.

Methods

We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 867 - 874
1 Jul 2022
Ji B Li G Zhang X Xu B Wang Y Chen Y Cao L

Aims

Periprosthetic joint infections (PJIs) with prior multiple failed surgery for reinfection represent a huge challenge for surgeons because of poor vascular supply and biofilm formation. This study aims to determine the results of single-stage revision using intra-articular antibiotic infusion in treating this condition.

Methods

A retrospective analysis included 78 PJI patients (29 hips; 49 knees) who had undergone multiple prior surgical interventions. Our cohort was treated with single-stage revision using a supplementary intra-articular antibiotic infusion. Of these 78 patients, 59 had undergone more than two prior failed debridement and implant retentions, 12 patients had a failed arthroplasty resection, three hips had previously undergone failed two-stage revision, and four had a failed one-stage revision before their single-stage revision. Previous failure was defined as infection recurrence requiring surgical intervention. Besides intravenous pathogen-sensitive agents, an intra-articular infusion of vancomycin, imipenem, or voriconazole was performed postoperatively. The antibiotic solution was soaked into the joint for 24 hours for a mean of 16 days (12 to 21), then extracted before next injection. Recurrence of infection and clinical outcomes were evaluated.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1344 - 1350
1 Oct 2012
Penny JO Ding M Varmarken JE Ovesen O Overgaard S

Radiostereometric analysis (RSA) can detect early micromovement in unstable implant designs which are likely subsequently to have a high failure rate. In 2010, the Articular Surface Replacement (ASR) was withdrawn because of a high failure rate. In 19 ASR femoral components, the mean micromovement over the first two years after implantation was 0.107 mm (. sd. 0.513) laterally, 0.055 mm (. sd. 0.204) distally and 0.150 mm (. sd. 0.413) anteriorly. The mean backward tilt around the x-axis was -0.08° (. sd. 1.088), mean internal rotation was 0.165° (. sd. 0.924) and mean varus tilt 0.238° (. sd. 0.420). The baseline to two-year varus tilt was statistically significant from zero movement, but there was no significant movement from one year onwards. We conclude that the ASR femoral component achieves initial stability and that early migration is not the mode of failure for this resurfacing arthroplasty


Bone & Joint Research
Vol. 7, Issue 11 | Pages 595 - 600
1 Nov 2018
Bergiers S Hothi HS Henckel J Eskelinen A Skinner J Hart A

Objectives. Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. Methods. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors. Results. There was no significant difference between the taper and bearing wear rates of the ‘pre-2007’ and ‘2007 onwards’ groups (p = 0.67 and p = 0.39, respectively). Pinnacles implanted from 2007 onwards were revised after a mean time of 50 months, which was significantly earlier than the ‘pre-2007’ hips (96 months) (p < 0.001). A reduction in the time to revision was present year on year from 2003 to 2011. Conclusion. We found no difference in the wear rate of these implants based on the year of implantation. The ‘pre-2007’ hips had a two-fold greater time to revision than those implanted after 2007; this may be due to the increased surveillance of MoM hips following UK regulatory advice and several high-profile failures. Interestingly, we observed a decreasing trend in the mean time to revision every year from 2003 onwards. Cite this article: S. Bergiers, H. S. Hothi, J. Henckel, A. Eskelinen, J. Skinner, A. Hart. Wear performance of retrieved metal-on-metal Pinnacle hip arthroplasties implanted before and after 2007. Bone Joint Res 2018;7:595–600. DOI: 10.1302/2046-3758.711.BJR-2018-0143.R1


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1045 - 1051
1 Aug 2013
Arthroplasty Society C

The purpose of this study was twofold: first, to determine whether the five-year results of hip resurfacing arthroplasty (HRA) in Canada justified the continued use of HRA; and second, to identify whether greater refinement of patient selection was warranted. This was a retrospective cohort study that involved a review of 2773 HRAs performed between January 2001 and December 2008 at 11 Canadian centres. Cox’s proportional hazards models were used to analyse the predictors of failure of HRA. Kaplan–Meier survival analysis was performed to predict the cumulative survival rate at five years. The factors analysed included age, gender, body mass index, pre-operative hip pathology, surgeon’s experience, surgical approach, implant sizes and implant types. The most common modes of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and 646 in women. The mean age at operation was 50.5 years (. sd. 8.72; 18 to 82) and mean follow-up was 3.4 years (. sd. 2.1; 2.0 to 10.1). At the last follow-up a total of 101 HRAs (3.6%) required revision. Using revision for all causes of failure as the endpoint, Kaplan–Meier survival analysis showed a cumulative survival of 96.4% (95% confidence interval (CI) 96.1 to 96.9) at five years. With regard to gender, the five-year overall survival was 97.4% in men (95% CI 97.1 to 97.7) and 93.6% in women (95% CI 92.6 to 94.6). Female gender, smaller femoral components, specific implant types and a diagnosis of childhood hip problems were associated with higher rates of failure. The most common cause of failure was fracture of the femoral neck, followed by loosening of the femoral component. The failure rates of HRA at five years justify the ongoing use of this technique in men. Female gender is an independent predictor of failure, and a higher failure rate at five years in women leads the authors to recommend this technique only in exceptional circumstances for women. . Cite this article: Bone Joint J 2013;95-B:1044–50


Bone & Joint Open
Vol. 3, Issue 7 | Pages 557 - 565
11 Jul 2022
Meier MK Reche J Schmaranzer F von Tengg-Kobligk H Steppacher SD Tannast M Novais EN Lerch TD

Aims

The frequency of severe femoral retroversion is unclear in patients with femoroacetabular impingement (FAI). This study aimed to investigate mean femoral version (FV), the frequency of absolute femoral retroversion, and the combination of decreased FV and acetabular retroversion (AR) in symptomatic patients with FAI subtypes.

Methods

A retrospective institutional review board-approved observational study was performed with 333 symptomatic patients (384 hips) with hip pain due to FAI evaluated for hip preservation surgery. Overall, 142 patients (165 hips) had cam-type FAI, while 118 patients (137 hips) had mixed-type FAI. The allocation to each subgroup was based on reference values calculated on anteroposterior radiographs. CT/MRI-based measurement of FV (Murphy method) and AV were retrospectively compared among five FAI subgroups. Frequency of decreased FV < 10°, severely decreased FV < 5°, and absolute femoral retroversion (FV < 0°) was analyzed.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 767 - 774
1 Jul 2022
Nakashima Y Ishibashi S Kitamura K Yamate S Motomura G Hamai S Ikemura S Fujii M

Aims

Although periacetabular osteotomies are widely used for the treatment of symptomatic dysplastic hips, long-term surgical outcomes and patient-reported outcome measures (PROMs) are still unclear. Accordingly, we assessed hip survival and PROMs at 20 years after transpositional osteotomy of the acetabulum (TOA).

Methods

A total of 172 hips in 159 patients who underwent TOA were followed up at a mean of 21.02 years (16.6 to 24.6) postoperatively. Kaplan-Meier analysis was used to assess survivorship with an endpoint of total hip arthroplasty (THA). PROMs included the visual analogue scale (VAS) Satisfaction, VAS Pain, Oxford Hip Score (OHS), and Forgotten Joint Score-12 (FJS-12). Thresholds for favourable outcomes for OHS (≥ 42) and FJS-12 (≥ 51) were obtained using the receiver operating characteristic curve with VAS Satisfaction ≥ 50 and VAS Pain < 20 as anchors.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 368 - 375
1 Mar 2022
Kuijpers MFL Colo E Schmitz MWJL Hannink G Rijnen WHC Schreurs BW

Aims

The aim of this study was to determine the outcome of all primary total hip arthroplasties (THAs) and their subsequent revision procedures in patients aged under 50 years performed at our institution.

Methods

All 1,049 primary THAs which were undertaken in 860 patients aged under 50 years between 1988 and 2018 in our tertiary care institution were included. We used cemented implants in both primary and revision surgery. Impaction bone grafting was used in patients with acetabular or femoral bone defects. Kaplan-Meier analyses were used to determine the survival of primary and revision THA with the endpoint of revision for any reason, and of revision for aseptic loosening.


Aims

To establish the survivorship, function, and metal ion levels in an unselected series of metal-on-metal hip resurfacing arthroplasties (HRAs) performed by a non-designer surgeon.

Methods

We reviewed 105 consecutive HRAs in 83 patients, performed by a single surgeon, at a mean follow-up of 14.9 years (9.3 to 19.1). The cohort included 45 male and 38 female patients, with a mean age of 49.5 years (SD 12.5)


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 359 - 367
1 Mar 2022
Deere K Matharu GS Ben-Shlomo Y Wilkinson JM Blom AW Sayers A Whitehouse MR

Aims

A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database.

Methods

Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1175 - 1179
1 Sep 2016
Olsen M Lewis PM Morrison Z McKee MD Waddell JP Schemitsch EH

Aims. One method of femoral head preservation following avascular necrosis (AVN) is core decompression and insertion of a tantalum rod. However, there may be a high failure rate associated with this procedure. The purpose of this study was to document the clinical and radiological outcomes following total hip arthroplasty (THA) subsequent to failed tantalum rod insertion. Patients and Methods. A total of 37 failed tantalum rods requiring total hip arthroplasty were identified from a prospective database. There were 21 hips in 21 patients (12 men and nine women, mean age 37 years, 18 to 53) meeting minimum two year clinical and radiographic follow-up whose THAs were carried out between November 2002 and April 2013 (mean time between tantalum rod implantation and conversion to a THA was 26 months, 6 to 72). These were matched by age and gender to individuals (12 men, nine women, mean age 40 years, 18 to 58) receiving THA for AVN without prior tantalum rod insertion. Results. There were no functional outcome differences between the two groups. Tantalum residue was identified on all post-operative radiographs in the tantalum group. Linear wear rates were comparable between groups with no evidence of catastrophic wear in either group. Conclusion. In the short term, tantalum rod implantation does not demonstrate an adverse effect on subsequent total joint replacement surgery. There is however, a high rate of retained tantalum debris on post-operative radiographs and thus there is an unknown risk of accelerated articular wear necessitating longer term study. Cite this article: Bone Joint J 2016;98-B:1175–9


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 341 - 351
1 Mar 2022
Fowler TJ Aquilina AL Reed MR Blom AW Sayers A Whitehouse MR

Aims

Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA.

Methods

We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors.


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 212 - 220
1 Feb 2022
Fishley WG Selvaratnam V Whitehouse SL Kassam AM Petheram TG

Aims

Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique.

Methods

We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 303 - 309
1 Mar 2017
Beaulé PE Bleeker H Singh A Dobransky J

Aims. Joint-preserving surgery of the hip (JPSH) has evolved considerably and now includes a number of procedures, including arthroscopy, surgical dislocation, and redirectional osteotomies of the femur and acetabulum. There are a number of different factors which lead to failure of JPSH. Consequently, it is of interest to assess the various modes of failure in order to continue to identify best practice and the indications for these procedures. . Patients and Methods. Using a retrospective observational study design, we reviewed 1013 patients who had undergone JPSH by a single surgeon between 2005 and 2015. There were 509 men and 504 women with a mean age of 39 years (16 to 78). Of the 1013 operations, 783 were arthroscopies, 122 surgical dislocations, and 108 peri-acetabular osteotomies (PAO). We analysed the overall failure rates and modes of failure. Re-operations were categorised into four groups: Mode 1 was arthritis progression or organ failure leading to total hip arthroplasty (THA); Mode 2 was an Incorrect diagnosis/procedure; Mode 3 resulted from malcorrection of femur (type A), acetabulum (type B), or labrum (type C) and Mode 4 resulted from an unintended consequence of the initial surgical intervention. Results. At a mean follow-up of 2.5 years, there had been 104 re-operations (10.2%) with a mean patient age of 35.5 years (17 to 64). There were 64 Mode 1 failures (6.3%) at a mean of 3.2 years following JPSH with a mean patient age of 46.8 years (18 to 64). There were 17 Mode 2 failures (1.7%) at a mean of 2.2 years post-JPSH with a mean patient age of 28.9 years (17 to 42) (2% scopes; 1% surgical dislocations). There were 19 Mode 3 failures (1.9%) at a mean of 2.0 years post-JPSH, with a mean patient age of 29.9 years (18 to 51) (2% scopes; 2% surgical dislocations; 5% PAO). There were 4 Mode 4 failures (0.4%) at a mean of 1.8 years post-JPSH with a mean patient age of 31.5 years (15 to 43). Using the modified Dindo-Clavien classification system, the overall complication rate among JPSHs was 4.2%. Conclusion. While defining the overall re-operation and complication rates, it is important to define the safety and effectiveness of JPSH. Standardisation of the modes of failure may help identify the best practice. Application of these modes to large clinical series, such as registries, will assist in further establishing how to improve the efficacy of JPSH. Cite this article: Bone Joint J 2017;99-B:303–9


Bone & Joint Open
Vol. 2, Issue 11 | Pages 926 - 931
9 Nov 2021
Houdek MT Wyles CC Smith JH Terzic A Behfar A Sierra RJ

Aims

Bone marrow-derived mesenchymal stem cells obtained from bone marrow aspirate concentrate (BMAC) with platelet-rich plasma (PRP), has been used as an adjuvant to hip decompression. Early results have shown promise for hip preservation in patients with osteonecrosis (ON) of the femoral head. The purpose of the current study is to examine the mid-term outcome of this treatment in patients with precollapse corticosteroid-induced ON of the femoral head.

Methods

In all, 22 patients (35 hips; 11 males and 11 females) with precollapse corticosteroid-induced ON of the femoral head underwent hip decompression combined with BMAC and PRP. Mean age and BMI were 43 years (SD 12) and 31 kg/m² (SD 6), respectively, at the time of surgery. Survivorship free from femoral head collapse and total hip arthroplasty (THA) and risk factors for progression were evaluated at minimum five-years of clinical follow-up with a mean follow-up of seven years (5 to 8).


The Bone & Joint Journal
Vol. 97-B, Issue 10 | Pages 1338 - 1344
1 Oct 2015
te Stroet MAJ Keurentjes JC Rijnen WHC Gardeniers JWM Verdonschot N Slooff TJJH Schreurs BW

We present the results of 62 consecutive acetabular revisions using impaction bone grafting and a cemented polyethylene acetabular component in 58 patients (13 men and 45 women) after a mean follow-up of 27 years (25 to 30). All patients were prospectively followed. The mean age at revision was 59.2 years (23 to 82). . We performed Kaplan–Meier (KM) analysis and also a Competing Risk (CR) analysis because with long-term follow-up, the presence of a competing event (i.e. death) prevents the occurrence of the endpoint of re-revision. . A total of 48 patients (52 hips) had died or had been re-revised at final review in March 2011. None of the deaths were related to the surgery. The mean Harris hip score of the ten surviving hips in ten patients was 76 points (45 to 99). The KM survivorship at 25 years for the endpoint ‘re-revision for any reason’ was 58.0% (95% confidence interval (CI) 38 to 73) and for ‘re-revision for aseptic loosening’ 72.1% (95% CI 51 to 85). With the CR analysis we calculated the KM analysis overestimates the failure rate with respectively 74% and 93% for these endpoints. The current study shows that acetabular impaction bone grafting revisions provide good clinical results at over 25 years. Cite this article: Bone Joint J 2015;97-B:1338–44


Bone & Joint Research
Vol. 5, Issue 9 | Pages 379 - 386
1 Sep 2016
Pahuta M Smolders JM van Susante JL Peck J Kim PR Beaule PE

Objectives. Alarm over the reported high failure rates for metal-on-metal (MoM) hip implants as well as their potential for locally aggressive Adverse Reactions to Metal Debris (ARMDs) has prompted government agencies, internationally, to recommend the monitoring of patients with MoM hip implants. Some have advised that a blood ion level >7 µg/L indicates potential for ARMDs. We report a systematic review and meta-analysis of the performance of metal ion testing for ARMDs. Methods. We searched MEDLINE and EMBASE to identify articles from which it was possible to reconstruct a 2 × 2 table. Two readers independently reviewed all articles and extracted data using explicit criteria. We computed a summary receiver operating curve using a Bayesian random-effects hierarchical model. Results. Our literature search returned 575 unique articles; only six met inclusion criteria defined a priori. The discriminative capacity of ion tests was homogeneous across studies but that there was substantial cut-point heterogeneity. Our best estimate of the “true” area under curve (AUC) for metal ion testing is 0.615, with a 95% credible interval of 0.480 to 0.735, thus we can state that the probability that metal ion testing is actually clinically useful with an AUC ≥ 0.75 is 1.7%. Conclusion. Metal ion levels are not useful as a screening test for identifying high risk patients because ion testing will either lead to a large burden of false positive patients, or otherwise marginally modify the pre-test probability. With the availability of more accurate non-invasive tests, we did not find any evidence for using blood ion levels to diagnose symptomatic patients. Cite this article: M. Pahuta, J. M. Smolders, J. L. van Susante, J. Peck, P. R. Kim, P. E. Beaule. Blood metal ion levels are not a useful test for adverse reactions to metal debris: a systematic review and meta-analysis. Bone Joint Res 2016;5:379–386. DOI: 10.1302/2046-3758.59.BJR-2016-0027.R1


Bone & Joint Research
Vol. 5, Issue 8 | Pages 338 - 346
1 Aug 2016
MacLeod AR Sullivan NPT Whitehouse MR Gill HS

Objectives. Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Materials and Methods. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences. Results. 36 mm diameter heads had significantly lower pull-off forces than 28 mm heads when impacted at 4 kN and 5 kN (p < 0.001; p < 0.001), but not at 6 kN (p = 0.21). Mean pull-off forces at 4 kN and 5 kN impaction forces were approximately 20% larger for 28 mm heads compared with 36 mm heads. Finite element and analytical models demonstrate that the differences in pull-off strength can be explained by differences in structural rigidity and the resulting interface pressures. Conclusion. This is the first study to show that 36 mm Co-Cr heads have up to 20% lower pull-off connection strength compared with 28 mm heads for equivalent assembly forces. This effect is likely to play a role in the high failure rates of large diameter MoM hips. Cite this article: A. R. MacLeod, N. P. T. Sullivan, M. R. Whitehouse, H. S. Gill. Large-diameter total hip arthroplasty modular heads require greater assembly forces for initial stability. Bone Joint Res 2016;5:338–346. DOI: 10.1302/2046-3758.58.BJR-2016-0044.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 6 | Pages 832 - 838
1 Aug 2002
Espehaug B Furnes O Havelin LI Engesæter LB Vollset SE

Using data from the Norwegian Arthroplasty Register, we have the assessed survival of 17 323 primary Charnley hip prostheses in patients with osteoarthritis based upon the type of cement used for the fixation of the implant. Overall, 9.2% had been revised after follow-up for ten years; 71% of the failures involved aseptic loosening of the femoral component. We observed significantly increased rates of failure for prostheses inserted with CMW1 and CMW3 cements. Using implants fixed with gentamicin-containing Palacos cement as the reference, the adjusted Cox regression failure rate ratios were 1.1 (95% CI 0.9 to 1.4) for implants cemented with plain Palacos, 1.1 (95% CI 0.7 to 1.6) for Simplex, 2.1 (95% 1.5 to 2.9) for gentamicin-containing CMW1, 2.0 (95% CI 1.6 to 2.4) for plain CMW1 and 3.0 (95% CI 2.3 to 3.9) for implants fixed with CMW3 cement. The adjusted failure rate at ten years varied from 5.9% for implants fixed with gentamicin-containing Palacos to 17% for those fixed with CMW3


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 452 - 460
1 Apr 2016
Mahmoud SSS Pearse EO Smith TO Hing CB

Aims. The optimal management of intracapsular fractures of the femoral neck in independently mobile patients remains open to debate. Successful fixation obviates the limitations of arthroplasty for this group of patients. However, with fixation failure rates as high as 30%, the outcome of revision surgery to salvage total hip arthroplasty (THA) must be considered. We carried out a systematic review to compare the outcomes of salvage THA and primary THA for intracapsular fractures of the femoral neck. . Patients and Methods. We performed a Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) compliant systematic review, using the PubMed, EMBASE and Cochrane libraries databases. A meta-analysis was performed where possible, and a narrative synthesis when a meta-analysis was not possible. Results. Our analyses revealed a significantly increased risk of complications including deep infection, early dislocation and peri-prosthetic fracture with salvage THA when compared with primary THA for an intracapsular fracture of the femoral neck (overall risk ratio of 3.15). Functional outcomes assessment using EuroQoL (EQ)-5D were not significantly different (p = 0.3). Conclusion. Salvage THA carries a significantly higher risk of complications than primary THA for intracapsular fractured neck of femur. Current literature is still lacking well designed studies to provide a full answer to the question. Take home message: Salvage THA is associated with more complications than primary THA for intracapsular neck of femur fractures. . Cite this article: Bone Joint J 2016;98-B:452–60


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 129 - 134
1 Jul 2021
Ayekoloye CI Abu Qa'oud M Radi M Leon SA Kuzyk P Safir O Gross AE

Aims

Improvements in functional results and long-term survival are variable following conversion of hip fusion to total hip arthroplasty (THA) and complications are high. The aim of the study was to analyze the clinical and functional results in patients who underwent conversion of hip fusion to THA using a consistent technique and uncemented implants.

Methods

A total of 39 hip fusion conversions to THA were undertaken in 38 patients by a single surgeon employing a consistent surgical technique and uncemented implants. Parameters assessed included Harris Hip Score (HHS) for function, range of motion (ROM), leg length discrepancy (LLD), satisfaction, and use of walking aid. Radiographs were reviewed for loosening, subsidence, and heterotopic ossification (HO). Postoperative complications and implant survival were assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1222 - 1230
1 Jul 2021
Slullitel PA Garcia-Barreiro GG Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims

We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty.

Methods

We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.


Bone & Joint Open
Vol. 2, Issue 7 | Pages 535 - 539
15 Jul 2021
Zak SG Lygrisse K Tang A Meftah M Long WJ Schwarzkopf R

Aims

As our population ages, the number of octogenarians who will require a total hip arthroplasty (THA) rises. In a value-based system where operative outcomes are linked to hospital payments, it is necessary to assess the outcomes in this population. The purpose of this study was to compare outcomes of elective, primary THA in patients ≥ 80 years old to those aged < 80.

Methods

A retrospective review of 10,251 consecutive THA cases from 2011 to 2019 was conducted. Patient-reported outcome (PRO) scores (Hip disability and Osteoarthritis Outcome Score (HOOS)), as well as demographic, readmission, and complication data, were collected.