In a single-blind, randomised series of knee replacements in 116 patients, we used radiostereometric analysis (RSA) to measure micromotion in three types of tibial implant fixation for two years after knee replacement. We compared hydroxyapatite-augmented porous coating, porous coating, and cemented fixation of the same design of tibial component. At one to two years, porous-coated implants migrated at a statistically significantly higher rate than hydroxyapatite-augmented or cemented implants. There was no significant difference between hydroxyapatite-coated and cemented implants. We conclude that
We have carried out a radiostereometric study of 50 patients (54 knees) with osteoarthritis of the knee who were randomly allocated to receive a cemented or a hydroxyapatite-coated femoral component for total knee replacement. The patients were also stratified to receive one of three types of articulating surface (standard, rotating platform, Freeman-Samuelson (FS)1000) all based on the Freeman-Samuelson design. The tibial components were cemented in all cases. Radiostereometry was performed post-operatively and at 3, 12 and 24 months. The analysis was restricted to rotation of the femoral component over time. After two years, rotation of the femoral components in the transverse, longitudinal and sagittal planes did not differ between the cemented and the hydroxyapatite-coated implants (p = 0.2 to 0.9). In total knee replacements with a rotating platform, the femoral component tended to tilt more posteriorly than in the other two designs, regardless of the choice of fixation (cemented or hydroxyapatite-coated, p = 0.04). The standard version of the femoral component, whether cemented or hydroxyapatite-coated, rotated more into valgus than was observed with the rotating-platform and FS1000 designs (p = 0.005). The increased constraint provided by the FS1000 component did not appear to have any adverse effect on fixation of the femoral component.
This study describes 146 primary total knee replacements, either fully or partially coated with
Aims.
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
Aims. The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA).
This prospective study compares the outcome of
157
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of
Endoprosthetic reconstruction with a distal femoral arthroplasty (DFA) can be used to treat distal femoral bone loss from oncological and non-oncological causes. This study reports the short-term implant survivorship, complications, and risk factors for patients who underwent DFA for non-neoplastic indications. We performed a retrospective review of 75 patients from a single institution who underwent DFA for non-neoplastic indications, including aseptic loosening or mechanical failure of a previous prosthesis (n = 25), periprosthetic joint infection (PJI) (n = 23), and native or periprosthetic distal femur fracture or nonunion (n = 27). Patients with less than 24 months’ follow-up were excluded. We collected patient demographic data, complications, and reoperations. Reoperation for implant failure was used to calculate implant survivorship.Aims
Methods
This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.Aims
Methods
We randomised 62 knees to receive either cemented or cementless versions of the Oxford unicompartmental knee replacement. The implants used in both arms of the study were similar, except that the cementless components were coated with porous titanium and
We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article:
Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.Aims
Methods
Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term migration with one- and two-year migration to evaluate the predictive ability of short-term migration data and to compare migration and inducible displacement between cemented and cementless (porous metal monoblock) components at least ten years postoperatively. Patients who had participated in RSA migration studies with two-year follow-up were recruited to return for a long-term follow-up, at least ten years from surgery. Two cemented tibial designs from two manufacturers and one porous metal monoblock cementless tibial design were studied. At the long-term follow-up, patients had supine RSA examinations to determine migration and loaded examinations (single leg stance) to determine inducible displacement. In total, 79 patients (54 female) returned, with mean time since surgery of 12 years (10 to 14). There were 58 cemented and 21 cementless tibial components.Aims
Patients and Methods
Bone void fillers are increasingly being used for dead space management in arthroplasty revision surgery. The aim of this study was to investigate the influence of calcium sulphate bone void filler (CS-BVF) on the damage and wear of total knee arthroplasty using experimental wear simulation. A total of 18 fixed-bearing U2 total knee arthroplasty system implants (United Orthopedic Corp., Hsinchu, Taiwan) were used. Implants challenged with CS-BVF were compared with new implants (negative controls) and those intentionally scratched with a diamond stylus (positive controls) representative of severe surface damage (n = 6 for each experimental group). Three million cycles (MC) of experimental simulation were carried out to simulate a walking gait cycle. Wear of the ultra-high-molecular-weight polyethylene (UHMWPE) tibial inserts was measured gravimetrically, and damage to articulating surfaces was assessed using profilometry.Objectives
Methods
The lack of effective treatment for cartilage defects has prompted investigations using tissue engineering techniques for their regeneration and repair. The success of tissue-engineered repair of cartilage may depend on the rapid and efficient adhesion of transplanted cells to a scaffold. Our aim in this study was to repair full-thickness defects in articular cartilage in the weight-bearing area of a porcine model, and to investigate whether the CD44 monoclonal antibody biotin-avidin (CBA) binding technique could provide satisfactory tissue-engineered cartilage. Cartilage defects were created in the load-bearing region of the lateral femoral condyle of mini-type pigs. The defects were repaired with traditional tissue-engineered cartilage, tissue-engineered cartilage constructed with the biotin-avidin (BA) technique, tissue-engineered cartilage constructed with the CBA technique and with autologous cartilage. The biomechanical properties, Western blot assay, histological findings and immunohistochemical staining were explored.Objectives
Methods
Fractures around total knee arthroplasties pose
a significant surgical challenge. Most can be managed with osteosynthesis
and salvage of the replacement. The techniques of fixation of these
fractures and revision surgery have evolved and so has the assessment
of outcome. This specialty update summarises the current evidence
for the classification, methods of fixation, revision surgery and
outcomes of the management of periprosthetic fractures associated
with total knee arthroplasty. Cite this article:
The treatment of osteochondral lesions is of
great interest to orthopaedic surgeons because most lesions do not heal
spontaneously. We present the short-term clinical outcome and MRI
findings of a cell-free scaffold used for the treatment of these
lesions in the knee. A total of 38 patients were prospectively evaluated
clinically for two years following treatment with an osteochondral
nanostructured biomimetic scaffold. There were 23 men and 15 women; the
mean age of the patients was 30.5 years (15 to 64). Clinical outcome
was assessed using the Knee Injury and Osteoarthritis Outcome Score
(KOOS), the Tegner activity scale and a Visual Analgue scale for
pain. MRI data were analysed based on the Magnetic Resonance Observation
of Cartilage Repair Tissue (MOCART) scoring system at three, 12
and 24 months post-operatively. There was a continuous significant
clinical improvement after surgery. In two patients, the scaffold
treatment failed (5.3%) There was a statistically significant improvement
in the MOCART precentage scores. The repair tissue filled most of
the defect sufficiently. We found subchondral laminar changes in all
patients. Intralesional osteophytes were found in two patients (5.3%).
We conclude that this one-step scaffold-based technique can be used
for osteochondral repair. The surgical technique is straightforward,
and the clinical results are promising. The MRI aspects of the repair
tissue continue to evolve during the first two years after surgery.
However, the subchondral laminar and bone changes are a concern. Cite this article:
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
Between 2003 and 2007, 99 knees in 77 patients
underwent opening wedge high tibial osteotomy. We evaluated the effect
of initial stable fixation combined with an artificial bone substitute
on the mid- to long-term outcome after medial opening-wedge high
tibial osteotomy (HTO) for medial compartmental osteoarthritis or
spontaneous osteonecrosis of the knee in 78 knees in 64 patients
available for review at a minimum of five years (mean age 68 years;
49 to 82). The mean follow-up was 6.5 years (5 to 10). The mean
Knee Society knee score and function score improved from 49.6 ( Opening-wedge HTO using a stable plate fixation system combined
with a bone substitute is a reliable procedure that provides excellent
results. Although this treatment might seem challenging for older
patients, our results strongly suggest that the results are equally
good. Cite this article:
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients
with end-stage anteromedial osteoarthritis of the knee, with radiological
evidence of improved fixation compared with previous reports using
cemented fixation. Cite this article:
The most common reasons for revision of unicompartmental
knee arthroplasty (UKA) are loosening and pain. Cementless components
may reduce the revision rate. The aim of this study was to compare
the fixation and clinical outcome of cementless and cemented Oxford
UKAs. A total of 43 patients were randomised to receive either a cemented
or a cementless Oxford UKA and were followed for two years with
radiostereometric analysis (RSA), radiographs aligned with the bone–implant
interfaces and clinical scores. The femoral components migrated significantly during the first
year (mean 0.2 mm) but not during the second. There was no significant
difference in the extent of migration between cemented and cementless
femoral components in either the first or the second year. In the
first year the cementless tibial components subsided significantly
more than the cemented components (mean 0.28 mm ( As second-year migration is predictive of subsequent loosening,
and as radiolucency is suggestive of reduced implant–bone contact,
these data suggest that fixation of the cementless components is
at least as good as, if not better than, that of cemented devices. Cite this article:
The cementless Oxford unicompartmental knee replacement
has been demonstrated to have superior fixation on radiographs and
a similar early complication rate compared with the cemented version.
However, a small number of cases have come to our attention where,
after an apparently successful procedure, the tibial component subsides into
a valgus position with an increased posterior slope, before becoming
well-fixed. We present the clinical and radiological findings of
these six patients and describe their natural history and the likely
causes. Two underwent revision in the early post-operative period,
and in four the implant stabilised and became well-fixed radiologically with
a good functional outcome. This situation appears to be avoidable by minor modifications
to the operative technique, and it appears that it can be treated
conservatively in most patients. Cite this article:
We investigated whether strontium-enriched calcium
phosphate cement (Sr-CPC)-treated soft-tissue tendon graft results
in accelerated healing within the bone tunnel in reconstruction
of the anterior cruciate ligament (ACL). A total of 30 single-bundle
ACL reconstructions using tendo Achillis allograft were performed
in 15 rabbits. The graft on the tested limb was treated with Sr-CPC,
whereas that on the contralateral limb was untreated and served
as a control. At timepoints three, six, nine, 12 and 24 weeks after
surgery, three animals were killed for histological examination.
At six weeks, the graft–bone interface in the control group was
filled in with fibrovascular tissue. However, the gap in the Sr-CPC
group had already been completely filled in with new bone, and there
was evidence of the early formation of Sharpey fibres. At 24 weeks,
remodelling into a normal ACL–bone-like insertion was found in the
Sr-CPC group. Coating of Sr-CPC on soft tissue tendon allograft
leads to accelerated graft healing within the bone tunnel in a rabbit
model of ACL reconstruction using Achilles tendon allograft. Cite this article:
A consecutive series of patients with a hydroxyapatite-coated
uncemented total knee replacement (TKR) performed by a single surgeon
between 1992 and 1995 was analysed. All patients were invited for
clinical review and radiological assessment. Revision for aseptic
loosening was the primary outcome. Assessment was based on the Knee
Society clinical score (KSS) and an independent radiological analysis.
Of 471 TKRs performed in 356 patients, 432 TKRs in 325 patients
were followed for a mean of 16.4 years (15 to 18). The 39 TKRs in
31 patients lost to follow-up had a mean KSS of 176 (148 to 198)
at a mean of ten years. There were revisions in 26 TKRs (5.5%),
of which 11 (2.3%) were for aseptic loosening. Other further surgery
was carried out on 49 TKRs (10.4%) including patellar resurfacing
in 20, arthrolysis in 19, manipulation under anaesthetic in nine
and extensor mechanism reconstruction in one. Survivorship at up to 18 years without aseptic loosening was
96% (95% confidence interval 91.9 to 98.1), at which point the mean
KSS was 176 (134 to 200). Of 110 knees that underwent radiological
evaluation, osteolysis was observed in five (4.5%), one of which
was revised. These data indicate that uncemented hydroxyapatite-coated TKR
can achieve favourable long-term survivorship, at least as good
as that of cemented designs.
Matrix-assisted autologous chondrocyte transplantation (MACT)
has been developed and applied in the clinical practice in the last
decade to overcome most of the disadvantages of the first generation
procedures. The purpose of this systematic review is to document
and analyse the available literature on the results of MACT in the
treatment of chondral and osteochondral lesions of the knee. All studies published in English addressing MACT procedures were
identified, including those that fulfilled the following criteria:
1) level I-IV evidence, 2) measures of functional or clinical outcome,
3) outcome related to cartilage lesions of the knee cartilage.Objectives
Methods
Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.
This review considers the surgical treatment
of displaced fractures involving the knee in elderly, osteoporotic patients.
The goals of treatment include pain control, early mobilisation,
avoidance of complications and minimising the need for further surgery.
Open reduction and internal fixation (ORIF) frequently results in
loss of reduction, which can result in post-traumatic arthritis
and the occasional conversion to total knee replacement (TKR). TKR
after failed internal fixation is challenging, with modest functional
outcomes and high complication rates. TKR undertaken as treatment
of the initial fracture has better results to late TKR, but does
not match the outcome of primary TKR without complications. Given
the relatively infrequent need for late TKR following failed fixation,
ORIF is the preferred management for most cases. Early TKR can be
considered for those patients with pre-existing arthritis, bicondylar
femoral fractures, those who would be unable to comply with weight-bearing restrictions,
or where a single definitive procedure is required.
In a series of 1304 patients (1867 knees), the results of simultaneous and staged bilateral total knee arthroplasty were compared with each other and with unilateral total knee arthroplasty. The bilateral procedures had a significantly higher rate of complications than unilateral procedures, almost entirely because of thromboembolic problems. However, this did not correspond to an increase in mortality. If a bilateral procedure was indicated, then a simultaneous procedure had no increased risk over a staged procedure. There was no increase in cardiovascular complications, the rate of deep-vein thrombosis or pulmonary embolism or mortality. The rate of infection was lower with a bilateral procedure and the overall revision rate was less than 1% in all groups. The prosthesis functioned as well in all groups in the medium and longer term periods. We feel that simultaneous bilateral total knee arthroplasty is a safe and successful procedure when compared with a staged bilateral procedure. It also has the added benefit of single anaesthetic, reduced costs and decreased total recovery time when compared to a staged bilateral procedure. For these reasons it should be considered as an option in the presence of bilateral knee joint disease.
Using meta-analysis we compared the survival and clinical outcomes of cemented and uncemented techniques in primary total knee replacement. We reviewed randomised controlled trials and observational studies comparing cemented and uncemented fixation. Our primary outcome was survival of the implant free of aseptic loosening. Our secondary outcome was joint function as measured by the Knee Society score. We identified 15 studies that met our final eligibility criteria. The combined odds ratio for failure of the implant due to aseptic loosening for the uncemented group was 4.2 (95% confidence interval (CI) 2.7 to 6.5) (p <
0.0001). Subgroup analysis of data only from randomised controlled trials showed no differences between the groups for odds of aseptic loosening (odds ratio 1.9, 95% CI 0.55 to 6.40, p = 0.314). The weighted mean difference for the Knee Society score was 0.005 (95% CI −0.26 to 0.26) (p = 0.972). There was improved survival of the cemented compared to uncemented implants, with no statistically significant difference in the mean Knee Society score between groups for all pooled data.
We report the long-term survival of a prospective randomised consecutive series of 501 primary knee replacements using the press-fit condylar posterior cruciate ligament-retaining prosthesis. Patients received either cemented (219 patients, 277 implants) or cementless (177 patients, 224 implants) fixation. Altogether, 44 of 501 knees (8.8%) underwent revision surgery (24 cemented This single-surgeon series, with no loss to follow-up, provides reliable data of the revision rates of one of the most commonly-used total knee replacements. The survival of the press-fit condylar total knee replacement remained good at 15 years, irrespective of the method of fixation.
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation.
We report the clinical and radiographic outcome of a consecutive series of 138 hydroxyapatite-coated total knee replacements with a mean follow-up of 11 years (10 to 13). The patients were entered into a prospective study and all living patients (76 knees) were evaluated. The Hospital for Special Surgery knee score was obtained for comparison with the pre-operative situation. No patient was lost to follow-up. Radiographic assessment revealed no loosening. Seven prostheses have been revised, giving a cumulative survival rate of 93% at 13 years. We believe this to be the longest follow-up report available for an hydroxyapatite-coated knee replacement and the first for this design of Insall-Burstein II knee.
We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up.
Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray. We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement. Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The migration of short-stemmed trays correlated inversely with the density of the impacted groups. That of impaction-grafted tibial trays was in the range reported for uncemented primary trays. Movements of short-stemmed trays without cortical support were largest and sensitive to the degree of compaction of the graft. If support of the rim was sufficient or a long stem was used, impacted morsellised bone graft achieved adequate initial stability.