We report the follow-up at 12 years of the use of the Elite Plus total hip replacement (THR). We have previously reported the results at a mean of 6.4 years. Of the 217 patients (234 THRs), 83 had died and nine had been lost to follow-up. The patients were reviewed radiologically and clinically using the Oxford hip score. Of the 234 THRs, 19 (8.1%) had required a revision by the final follow-up in all but one for aseptic loosening. Survivorship analysis for revision showed a survival of 93.9% (95% confidence interval (CI) 89.2 to 96.5) at ten years, and of 88.0% (95% CI 81.8 to 92.3) at 12 years. At the final follow-up survival analysis showed that 37% (95% CI 37.3 to 44.7) of the prostheses had either failed radiologically or had been revised. Patients with a radiologically loose femoral component had a significantly poorer Oxford hip score than those with a well-fixed component (p = 0.03). Radiological loosening at 6.4 years was predictive of failure at 12 years. Medium-term radiographs and clinical scores should be included in the
Cardiac magnetic resonance (CMR) was used to assess whether cardiac function or tissue composition was affected in patients with well-functioning metal-on-metal hip resurfacing arthroplasties (MoMHRA) when compared with a group of controls, and to assess if metal ion levels correlated with any of the functional or structural parameters studied. In all, 30 participants with no significant cardiac history were enrolled: 20 patients with well-functioning MoMHRA at mean follow-up of 8.3 years post-procedure (ten unilateral, ten bilateral; 17 men, three women) and a case-matched control group of ten non-MoM total hip arthroplasty patients (six men, four women). The mean age of the whole cohort (study group and controls) at the time of surgery was 50.6 years (41.0 to 64.0). Serum levels of cobalt and chromium were measured, and all patients underwent CMR imaging, including cine, T2* measurements, T1 and T2 mapping, late gadolinium enhancement, and strain measurements.Aims
Patients and Methods
Ceramic-on-metal (CoM) is a relatively new bearing
combination for total hip arthroplasty (THA) with few reported outcomes.
A total of 287 CoM THAs were carried out in 271 patients (mean age
55.6 years (20 to 77), 150 THAs in female patients, 137 in male)
under the care of a single surgeon between October 2007 and October
2009. With the issues surrounding metal-on-metal bearings the decision
was taken to review these patients between March and November 2011,
at a mean follow-up of 34 months (23 to 45) and to record pain,
outcome scores, radiological analysis and blood ion levels. The
mean Oxford Hip Score was 19.2 (12 to 53), 254 patients with 268
hips (95%) had mild/very mild/no pain, the mean angle of inclination
of the acetabular component was 44.8o (28o to
63o), 82 stems (29%) had evidence of radiolucent lines
of >
1 mm in at least one Gruen zone and the median levels of cobalt
and chromium ions in the blood were 0.83 μg/L (0.24 μg/L to 27.56 μg/L)
and 0.78 μg/L (0.21 μg/L to 8.84 μg/L), respectively. The five-year
survival rate is 96.9% (95% confidence interval 94.7% to 99%). Due to the presence of radiolucent lines and the higher than
expected levels of metal ions in the blood, we would not recommend
the use of CoM THA without further long-term follow-up. We plan
to monitor all these patients regularly. Cite this article:
Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up.Aims
Methods
Aims. Routine
Aims. Periprosthetic proximal femoral fractures (PFFs) are a major complication after total hip arthroplasty (THA). Health status after PFF is not specifically investigated. The aim of this study is to evaluate the health status pattern over two years after sustaining a PFF. Methods. A cohort of patients with PFF after THA was derived from the Brabant Injury Outcomes
Objectives. Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. Methods. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors. Results. There was no significant difference between the taper and bearing wear rates of the ‘pre-2007’ and ‘2007 onwards’ groups (p = 0.67 and p = 0.39, respectively). Pinnacles implanted from 2007 onwards were revised after a mean time of 50 months, which was significantly earlier than the ‘pre-2007’ hips (96 months) (p < 0.001). A reduction in the time to revision was present year on year from 2003 to 2011. Conclusion. We found no difference in the wear rate of these implants based on the year of implantation. The ‘pre-2007’ hips had a two-fold greater time to revision than those implanted after 2007; this may be due to the increased
Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (. sd. 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis. Results. Mean Harris Hip Scores improved from 54.1 (. sd. 20.5) to 91.2 (. sd. 10.8) at two years postoperatively (p < 0.001). All patients had radiologically well-fixed components, no patients experienced any instability, and no patients required any further intervention. Mean cobalt levels increased from 0.065 ug/l (. sd. 0.03) preoperatively to 0.30 ug/l (. sd. 0.51) at one year postoperatively (p = 0.01) but decreased at two years postoperatively to 0.16 ug/l (. sd. 0.23; p = 0.2). Four patients (9.3%) had a cobalt level outside the reference range (0.03 ug/l to 0.29 ug/l) at two years postoperatively, with values from 0.32 ug/l to 0.94 ug/l. The mean femoral BMD ratio was maintained in Gruen zones 2 to 7 at both one and two years postoperatively using this stem design. At two years postoperatively, mean BMD in the medial calcar was 101.5% of the baseline value. Conclusion. Use of a modular DM prosthesis and cementless, tapered femoral stem has shown encouraging results in young, active patients undergoing primary THA. Elevation in mean cobalt levels and the presence of four patients outside the reference range at two years postoperatively demonstrates the necessity of continued
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
Objectives. High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA). Methods. We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015. Results. Our search strategy identified 1082 studies of which 115 studies were included for full article review. Following review, 17 articles were found that investigated surrogate markers of long-term outcome. These included one systematic review, one randomised control trial (RCT), one case control study and 13 case series. Validated surrogate markers included Radiostereometric Analysis (RSA) and Einzel-Bild-Röntgen-Analyse (EBRA), each measuring implant migration and wear. We identified five RSA studies (one systematic review and four case series) and four EBRA studies (one RCT and three case series). Patient Reported Outcome Measures (PROMs) at six months have been investigated but have not been validated against long-term outcomes. Conclusions. This systematic review identified two validated surrogate markers of long-term primary THA outcome: RSA and EBRA, each measuring implant migration and wear. We recommend the consideration of RSA in the pre-market testing of new implants. EBRA can be used to investigate acetabular wear but not femoral migration. Further studies are needed to validate the use of PROMs for post-market
In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes. A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone.Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).Aims
Methods
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the
We examined patient and surgical factors associated
with deep surgical site infection (SSI) following total hip replacement
(THR) in a large integrated healthcare system. A retrospective review
of a cohort of primary THRs performed between 2001 and 2009 was
conducted. Patient characteristics, surgical details, surgeon and
hospital volumes, and SSIs were identified using the Kaiser Permanente
Total Joint Replacement Registry (TJRR). Proportional-hazard regression
models were used to assess risk factors for SSI. The study cohort
consisted of 30 491 THRs, of which 17 474 (57%) were performed on
women. The mean age of the patients in the whole series was 65.5
years (13 to 97; . sd. 11.8) and the mean body mass index
was 29.3 kg/m. 2. (15 to 67; . sd. 5.9). The incidence
of SSI was 0.51% (155 of 30 491). Patient factors associated with
SSI included female gender, obesity, and American Society of Anesthesiologists
(ASA) score ≥ 3. Age, diagnosis, diabetes and race
were not associated with SSI. The only surgical factor associated
with SSI was a bilateral procedure. Surgeon and hospital volumes,
use of antibiotic-laden cement, fixation method, laminar flow, body
exhaust suits, surgical approach and fellowship training were not
associated with risk of SSI. A comprehensive infection
We prospectively studied 217 patients who underwent 234 Elite Plus total hip arthroplasties. At a mean of 6.4 (SD 0.7) years post-operatively, 39 patients had died and 22 were either lost to follow-up or had no radiographs available. Clinical (Oxford hip score) and radiological assessments were performed on 156 patients (168 hip arthroplasties) who had a mean age of 67.7 (SD 9.7) years at operation. In the assessed group, 26 of 159 (16.4%) of femoral stems which had not already been revised and 19 of 159 (11.9%) of acetabular cups were definitely loose. In total, 52 of 168 (31%) of hips had either been revised or had definite evidence of loosening of a component. We could not establish any relationship between clinical and radiological outcomes. Despite the fact that the clinical outcome and rate of revision for the Elite Plus appeared to meet international standards, our findings give us cause for concern. We believe that joint registries should include radiological
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
We matched 78 patients with a loose cemented Charnley Elite Plus total hip replacement (THR) by age, gender, race, prosthesis and time from surgery with 49 patients with a well-fixed stable hip replacement, to determine if poor bone quality predisposes to loosening. Clinical, radiological, biomechanical and bone mineral density indicators of bone quality were assessed. Patients with loose replacements had more pain, were more likely to have presented with atrophic arthritis and to have a history of fragility fracture, narrower femoral cortices and lower peri-prosthetic or lumbar spine bone mineral density (all t-test, p <
0.01). They also tended to be smokers (chi-squared test, p = 0.08). Vitamin-D deficiency was common, but not significantly different between the two groups (t-test, p = 0.31). In this series of cemented hip replacements performed between 1994 and 1998, aseptic loosening was associated with poor bone quality. Patients with a THR should be screened for osteoporosis and have regular radiological
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term
To establish the survivorship, function, and metal ion levels in an unselected series of metal-on-metal hip resurfacing arthroplasties (HRAs) performed by a non-designer surgeon. We reviewed 105 consecutive HRAs in 83 patients, performed by a single surgeon, at a mean follow-up of 14.9 years (9.3 to 19.1). The cohort included 45 male and 38 female patients, with a mean age of 49.5 years (SD 12.5)Aims
Methods
We retrospectively examined the long-term outcome of 96 asymptomatic hips in 96 patients with a mean age of 49.3 years (16 to 65) who had radiological evidence of femoroacetabular impingement. When
Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years. Using the database of
a single academic centre, 126 revision THAs in 117 patients using a single
design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.Aims
Methods
The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported.Aims
Methods
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
Using a regional arthroplasty register we assessed the outcome at five years of 1198 primary Charnley total hip replacements (THRs) carried out in 1152 patients across a single UK health region in 1990. Information regarding outcome was available for 1080 hips (90%) and 499 had an independent clinical and radiological assessment. By five years the known rate of aseptic loosening was 2.3%, of deep infection 1.4%, of dislocation 5.0% and of revision 3.2%. The radiological assessment of 499 THRs revealed gross failure in a further 5.2%, which had been previously unrecognised. The combined rate of failure of nearly 9% is higher than those published from specialist centres and surgeons, but is probably more representative of the norm. Our study supports the need for a national register and
We aimed to report the mid- to long-term rates of septic and aseptic failure after two-stage revision surgery for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We retrospectively reviewed 96 cases which met the Musculoskeletal Infection Society criteria for PJI. The mean follow-up was 90 months (SD 32). Septic failure was assessed using a Delphi-based consensus definition. Any further surgery undertaken for aseptic mechanical causes was considered as aseptic failure. The cumulative incidence with competing risk analysis was used to predict the risk of septic failure. A regression model was used to evaluate factors associated with septic failure. The cumulative incidence of aseptic failure was also analyzed.Aims
Methods
The complex relationship between acetabular component position and spinopelvic mobility in patients following total hip arthroplasty (THA) renders it difficult to optimize acetabular component positioning. Mobility of the normal lumbar spine during postural changes results in alterations in pelvic tilt (PT) to maintain the sagittal balance in each posture and, as a consequence, markedly changes the functional component anteversion (FCA). This study aimed to investigate the in vivo association of lumbar degenerative disc disease (DDD) with the PT angle and with FCA during postural changes in THA patients. A total of 50 patients with unilateral THA underwent CT imaging for radiological evaluation of presence and severity of lumbar DDD. In all, 18 patients with lumbar DDD were compared to 32 patients without lumbar DDD. In vivo PT and FCA, and the magnitudes of changes (ΔPT; ΔFCA) during supine, standing, swing-phase, and stance-phase positions were measured using a validated dual fluoroscopic imaging system.Aims
Methods
There are few medium- and long-term data on the outcome of the use of proximal femoral structural allografts in revision hip arthroplasty. This is a study of a consecutive series of 40 proximal femoral allografts performed for failed total hip replacements using the same technique with a minimum follow-up of five years (mean 8.8 years; range 5 to 11.5 years). In all cases the stem was cemented into both the allograft and the host femur. The proximal femur of the host was resected in 37 cases. There were four early revisions (10%), two for infection, one for nonunion of the allograft-host junction, and one for allograft resorption noted at the time of revision of a failed acetabular reconstruction. Junctional nonunion was seen in three patients (8%), two of whom were managed successfully by bone grafting, and bone grafting and plating respectively. Instability was observed in four (10%). Trochanteric nonunion was seen in 18 patients (46%) and trochanteric escape in ten of these (27%). The mean Harris hip score improved from 39 to 79. Severe resorption involving the full thickness of the allograft was seen in seven patients (17.5%). This progressed rapidly and silently, but has yet to cause failure of any of the reconstructions. Profound resorption of the allograft may be related to a combination of factors, including a slow form of immune rejection, stress shielding and resorption due to mechanical disuse with solid cemented distal fixation, and the absence of any masking or protective effect which may be provided by the retention of the bivalved host bone as a vascularised onlay autograft. Although continued
To compare the in vivo long-term fixation achieved by two acetabular components with different porous ingrowth surfaces using radiostereometric analysis (RSA). This was a minimum ten-year follow-up of a prospective randomized trial of 62 hips with two different porous ingrowth acetabular components. RSA exams had previously been acquired through two years of follow-up. Patients returned for RSA examination at a minimum of ten years. In addition, radiological appearance of these acetabular components was analyzed, and patient-reported outcome measures (PROMs) obtained.Aims
Methods
Tranexamic acid (TXA) has been shown to reduce blood loss and transfusion requirements in patients undergoing orthopaedic surgery. There remains a lack of prospective evidence for the use of TXA in patients undergoing periacetabular osteotomy (PAO). The purpose of this study was to determine if intravenous (IV) TXA is effective in reducing calculated blood loss and transfusions after PAO. This was a single-centre prospective double-blind placebo-controlled randomized trial of 81 patients aged 12 to 45 years undergoing elective PAO by a single surgeon. The intervention group (n = 40) received two doses of IV TXA of a maximum 1 g in each dose; the control group (n = 41) received two doses of 50 ml 0.9% saline IV. The primary outcome was perioperative calculated blood loss. Secondary outcomes included allogenic transfusions and six-week postoperative complications.Aims
Methods
Short, bone-conserving femoral components are increasingly used in total hip arthroplasty (THA). They are expected to allow tissue-conserving implantation and to render future revision surgery more straightforward but the long-term data on such components is limited. One such component is the global tissue-sparing (GTS) stem. Following the model for stepwise introduction of new orthopaedic implants, we evaluated early implant fixation and clinical outcome of this novel short-stem THA and compared it to that of a component with established good long-term clinical outcome. In total, 50 consecutive patients ≤ 70 years old with end-stage symptomatic osteo-arthritis were randomized to receive THA with the GTS stem or the conventional Taperloc stem using the anterior supine intermuscular approach by two experienced hip surgeons in two hospitals in the Netherlands. Primary outcome was implant migration. Patients were followed using routine clinical examination, patient reported outcome using Harris Hip Score (HHS), Hip Disability And Osteoarthritis Outcome Score (HOOS), EuroQol five-dimension questionnaire (EQ5D), and Roentgen Stereophotogrammetric Analysis (RSA) at three, six, 12, and 24 months. This study evaluated the two-year follow-up results.Aims
Methods
The present study investigated the five-year interval changes in pseudotumours and measured serum metal ions at long-term follow-up of a previous report of 28 mm diameter metal-on-metal (MoM) total hip arthroplasty (THA). A total of 72 patients (mean age 46.6 years (37 to 55); 43 men, 29 women; 91 hips) who underwent cementless primary MoM THA with a 28 mm modular head were included. The mean follow-up duration was 20.3 years (18 to 24). All patients had CT scans at a mean 15.1 years (13 to 19) after the index operation and subsequent follow-up at a mean of 20.2 years (18 to 24). Pseudotumour volume, type of mass, and new-onset pseudotumours were evaluated using CT scanning. Clinical outcomes were assessed by Harris Hip Score (HHS) and the presence of groin pain. Serum metal ion (cobalt (Co) and chromium (Cr)) levels were measured at the latest follow-up.Aims
Patients and Methods
The aim of this randomized trial was to compare the functional outcome of two different surgical approaches to the hip in patients with a femoral neck fracture treated with a hemiarthroplasty. A total of 150 patients who were treated between February 2014 and July 2017 were included. Patients were allocated to undergo hemiarthroplasty using either an anterolateral or a direct lateral approach, and were followed for 12 months. The mean age of the patients was 81 years (69 to 90), and 109 were women (73%). Functional outcome measures, assessed by a physiotherapist blinded to allocation, and patient-reported outcome measures (PROMs) were collected postoperatively at three and 12 months.Aims
Patients and Methods
The aim of this study was to determine whether the rates of revision
for metal-on-metal (MoM) total hip arthroplasties (THAs) with Pinnacle
components varied according to the year of the initial operation,
and compare these with the rates of revision for other designs of
MoM THA. Data from the National Joint Registry for England and Wales included
36 mm MoM THAs with Pinnacle acetabular components which were undertaken
between 2003 and 2012 with follow-up for at least five years (n
= 10 776) and a control group of other MoM THAs (n = 13 817). The
effect of the year of the primary operation on all-cause rates of revision
was assessed using Cox regression and interrupted time-series analysis.Aims
Patients and Methods
The aim of this study was to investigate the differences in 30-day outcomes between patients undergoing revision for an infected total hip arthroplasty (THA) compared with an aseptic revision THA. This was a retrospective review of prospectively collected data from the American College of Surgeons National Surgical Quality Improvement Program (ACS NSQIP) database, between 2012 and 2017, using Current Procedural Terminology (CPT) codes for patients undergoing a revision THA (27134, 27137, 27138). International Classification of Diseases Ninth Revision/Tenth Revision (ICD-9-CM, ICD-10-CM) diagnosis codes for infection of an implant or device were used to identify patients undergoing an infected revision THA. CPT-27132 coupled with ICD-9-CM/ICD-10-CM codes for infection were used to identify patients undergoing a two-stage revision. A total of 13 556 patients were included; 1606 (11.8%) underwent a revision THA due to infection and there were 11 951 (88.2%) aseptic revisions.Aims
Patients and Methods
Osteolysis, secondary to local and systemic physiological effects, is a major challenge in total hip arthroplasty (THA). While osteolytic defects are commonly observed in long-term follow-up, how such lesions alter the distribution of stress is unclear. The aim of this study was to quantitatively describe the biomechanical implication of such lesions by performing subject-specific finite-element (FE) analysis on patients with osteolysis after THA. A total of 22 hemipelvis FE models were constructed in order to assess the transfer of load in 11 patients with osteolysis around the acetabular component of a THA during slow walking and a fall onto the side. There were nine men and two women. Their mean age was 69 years (55 to 81) at final follow-up. Changes in peak stress values and loads to fracture in the presence of the osteolytic defects were measured.Aims
Patients and Methods
Cementless primary total hip arthroplasty (THA) is associated with risks of bleeding and thromboembolism. Anticoagulants are effective as venous thromboprophylaxis, but with an increased risk of bleeding. Tranexamic acid (TXA) is an efficient antifibrinolytic agent, but the mode and timing of its administration remain controversial. This study aimed to determine whether two intravenous (IV) TXA regimens (a three-hour two-dose (short-TXA) and 11-hour four-dose (long-TXA)) were more effective than placebo in reducing perioperative real blood loss (RBL, between baseline and day 3 postoperatively) in patients undergoing THA who receive rivaroxaban as thromboprophylaxis. The secondary aim was to assess the non-inferiority of the reduction of blood loss of the short protocol A multicentre, prospective, randomized, double-blind, placebo-controlled trial was undertaken involving 229 patients undergoing primary cementless THA using a posterior approach, whose extended rivaroxaban thromboprophylaxis started on the day of surgery. There were 98 male and 131 female patients, with a mean age of 65.5 years (32 to 91). The primary outcome, perioperative RBL, was evaluated at 72 hours postoperatively. The efficacy of short- and long-TXA protocols in the reduction of perioperative RBL was compared with a placebo group.Aims
Patients and Methods
We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr. For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood.Objectives
Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
The aim of this study was to report the implant survival and patient-reported outcome measures (PROMs) in a consecutive series of patients aged less than 50 years at the time of arthroplasty using the Birmingham Hip Resurfacing system (BHR), with a minimum follow-up of ten years. A total of 226 patients with osteoarthritis of the hip, who underwent BHR and presented to a single surgeon, were included in the study. Survival of the implant was confirmed by cross-checking with the Australian Orthopaedic Association National Joint Replacement Registry. Kaplan–Meier survival curves with 95% confidence intervals (CIs) were constructed. Pre- and postoperative PROMs were compared with Aims
Patients and Methods
Severe acetabular bone loss and pelvic discontinuity (PD) present particular challenges in revision total hip arthroplasty. To deal with such complex situations, cup-cage reconstruction has emerged as an option for treating this situation. We aimed to examine our success in using this technique for these anatomical problems. We undertook a retrospective, single-centre series of 35 hips in 34 patients (seven male, 27 female) treated with a cup-cage construct using a trabecular metal shell in conjunction with a titanium cage, for severe acetabular bone loss between 2011 and 2015. The mean age at the time of surgery was 70 years (42 to 85) and all patients had an acetabular defect graded as Paprosky Type 2C through to 3B, with 24 hips (69%) having PD. The mean follow-up was 47 months (25 to 84).Aims
Patients and Methods
The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. A descriptive observational study was undertaken using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man between April 2003 and March 2017. All patients aged ≤ 20 years at the time of THA were included and the primary outcome was revision surgery. Descriptive statistics were used to summarize the data and Kaplan–Meier estimates calculated for the cumulative implant survival.Aims
Patients and Methods
The aim of this study was to report the initial results of the
Exeter V40 stem, which became available in 2000. A total of 540 total hip arthroplasties (THAs) were performed
in our unit using this stem between December 2000 and May 2002.
Our routine protocol is to review patients postoperatively and at
one, five, and ten years following surgery.Aims
Patients and Methods
The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison.Aims
Patients and Methods
Alarm over the reported high failure rates for metal-on-metal (MoM) hip implants as well as their potential for locally aggressive Adverse Reactions to Metal Debris (ARMDs) has prompted government agencies, internationally, to recommend the monitoring of patients with MoM hip implants. Some have advised that a blood ion level >7 µg/L indicates potential for ARMDs. We report a systematic review and meta-analysis of the performance of metal ion testing for ARMDs. We searched MEDLINE and EMBASE to identify articles from which it was possible to reconstruct a 2 × 2 table. Two readers independently reviewed all articles and extracted data using explicit criteria. We computed a summary receiver operating curve using a Bayesian random-effects hierarchical model.Objectives
Methods
The aim of this study was to determine whether patients with
metal-on-metal (MoM) arthroplasties of the hip have an increased
risk of cardiac failure compared with those with alternative types
of arthroplasties (non-MoM). A linkage study between the National Joint Registry, Hospital
Episodes Statistics and records of the Office for National Statistics
on deaths was undertaken. Patients who underwent elective total
hip arthroplasty between January 2003 and December 2014 with no
past history of cardiac failure were included and stratified as
having either a MoM (n = 53 529) or a non-MoM (n = 482 247) arthroplasty.
The primary outcome measure was the time to an admission to hospital
for cardiac failure or death. Analysis was carried out using data
from all patients and from those matched by propensity score.Aims
Patients and Methods
We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip Resurfacing (BHR)
implants who have adverse reactions to metal debris (ARMD). Metal ion levels in whole blood were measured in 185 patients
with bilateral BHRs. Patients were divided into those with ARMD
who either had undergone a revision for ARMD or had ARMD on imaging
(n = 30), and those without ARMD (n = 155). Receiver operating characteristic
analysis was used to determine the optimal thresholds of blood metal
ion levels for identifying patients with ARMD.Aims
Patients and Methods
To determine the outcomes following revision surgery of metal-on-metal
hip arthroplasties (MoMHA) performed for adverse reactions to metal
debris (ARMD), and to identify factors predictive of re-revision. We performed a retrospective observational study using National
Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery
for ARMD between 2008 and 2014. The outcomes studied following revision were
intra-operative complications, mortality and re-revision surgery.
Predictors of re-revision were identified using competing-risk regression
modelling.Aims
Patients and Methods
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
The aims of this study were to compare the diagnostic test characteristics
of ultrasound alone, metal artefact reduction sequence MRI (MARS-MRI)
alone, and ultrasound combined with MARS-MRI for identifying intra-operative
pseudotumours in metal-on-metal hip resurfacing (MoMHR) patients
undergoing revision surgery. This retrospective diagnostic accuracy study involved 39 patients
(40 MoMHRs). The time between imaging modalities was a mean of 14.6
days (0 to 90), with imaging performed at a mean of 5.3 months (0.06
to 12) before revision. The prevalence of intra-operative pseudotumours
was 82.5% (n = 33).Aims
Methods
The acetabular labrum is a soft-tissue structure
which lines the acetabular rim of the hip joint. Its role in hip
joint biomechanics and joint health has been of particular interest
over the past decade. In normal hip joint biomechanics, the labrum
is crucial in retaining a layer of pressurised intra-articular fluid
for joint lubrication and load support/distribution. Its seal around
the femoral head is further regarded as a contributing to hip stability through
its suction effect. The labrum itself is also important in increasing
contact area thereby reducing contact stress. Given the labrum’s
role in normal hip joint biomechanics, surgical techniques for managing
labral damage are continuously evolving as our understanding of
its anatomy and function continue to progress. The current paper
aims to review the anatomy and biomechanical function of the labrum
and how they are affected by differing surgical techniques. Take home message: The acetabular labrum plays a critical role
in hip function and maintaining and restoring its function during
surgical intervention remain an essential goal. Cite this article:
We assessed the difference in hospital based and early clinical
outcomes between the direct anterior approach and the posterior
approach in patients who undergo total hip arthroplasty (THA). The outcome was assessed in 448 (203 males, 245 females) consecutive
patients undergoing unilateral primary THA after the implementation
of an ‘Enhanced Recovery’ pathway. In all, 265 patients (mean age:
71 years (49 to 89); 117 males and 148 females) had surgery using
the direct anterior approach (DAA) and 183 patients (mean age: 70
years (26 to 100); 86 males and 97 females) using a posterior approach.
The groups were compared for age, gender, American Society of Anesthesiologists
grade, body mass index, the side of the operation, pre-operative
Oxford Hip Score (OHS) and attendance at ‘Joint school’. Mean follow-up
was 18.1 months (one to 50).Aims
Patients and Methods
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
The inherent challenges of total hip replacement
(THR) in children include the choice of implant for the often atypical
anatomical morphology, its fixation to an immature growing skeleton
and the bearing surface employed to achieve a successful long-term
result. We report the medium-term results of 52 consecutive uncemented
THRs undertaken in
35 paediatric patients with juvenile idiopathic arthritis. The mean
age at the time of surgery was 14.4 years (10 to 16). The median
follow-up was 10.5 years (6 to 15). During the study period 13 THRs
underwent revision surgery. With revision as an endpoint, subgroup
analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs
and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This
resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral
component and 62% (95% CI 41.0 to 78.0) of the acetabular component.
Revision of the acetabular component for wear and osteolysis were
the most common reasons for failure accounting for 11 of the 13
revisions. The success seen in patients with a ceramic-on-ceramic articulation
seems to indicate that this implant strategy has the potential to
make a major difference to the long-term outcome in this difficult
group of patients.
We hypothesised that the use of tantalum (Ta)
acetabular components in revision total hip arthroplasty (THA) was protective
against subsequent failure due to infection. We identified 966 patients
(421 men, 545 women and 990 hips) who had undergone revision THA
between 2000 and 2013. The mean follow up was 40.2 months (3 months
to 13.1 years). The mean age of the men and women was 62.3 years
(31 to 90) and 65.1 years (25 to 92), respectively. Titanium (Ti) acetabular components were used in 536 hips while
Ta components were used in 454 hips. In total, 73 (7.3%) hips experienced
subsequent acetabular failure. The incidence of failure was lower
in the Ta group at 4.4% (20/454) compared with 9.9% (53/536) in
the Ti group (p <
0.001, odds ratio 2.38; 95% CI 1.37 to 4.27).
Among the 144 hips (64 Ta, 80 Ti) for which revision had been performed
because of infection, failure due to a subsequent infection was
lower in the Ta group at 3.1% (2/64) compared with 17.5% (14/80)
for the Ti group (p = 0.006). Thus, the use of Ta acetabular components during revision THA
was associated with a lower incidence of failure from all causes
and Ta components were associated with a lower incidence of subsequent
infection when used in patients with periprosthetic joint infection. Cite this article:
We report a systematic review and meta-analysis of published randomised controlled trials evaluating the efficacy of tranexamic acid (TXA) in reducing blood loss and transfusion in total hip replacement (THR). The data were evaluated using the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group. We identified 11 clinical trials which were suitable for detailed extraction of data. There were no trials that used TXA in revision THR. A total of seven studies (comprising 350 patients) were eligible for the blood loss outcome data. The use of TXA reduced intra-operative blood loss by a mean of 104 ml (95% confidence interval (CI) −164 to −44, p = 0.0006, heterogeneity I2 0%), postoperative blood loss by a mean of 172 ml (95% CI −263 to −81, p = 0.0002, heterogeneity I2 63%) and total blood loss by a mean of 289 ml (95% CI −440 to −138, p <
0.0002, heterogeneity I2 54%). TXA led to a significant reduction in the proportion of patients requiring allogeneic blood transfusion (risk difference −0.20, 95% CI −0.29 to −0.11, p <
0.00001, I2 15%). There were no significant differences in deep-vein thrombosis, pulmonary embolism, infection rates or other complications among the study groups.
We are currently facing an epidemic of periprosthetic
fractures around the hip. They may occur either during surgery or
post-operatively. Although the acetabulum may be involved, the femur
is most commonly affected. We are being presented with new, difficult
fracture patterns around cemented and cementless implants, and we
face the challenge of an elderly population who may have grossly
deficient bone and may struggle to rehabilitate after such injuries.
The correct surgical management of these fractures is challenging.
This article will review the current choices of implants and techniques
available to deal with periprosthetic fractures of the femur. Cite this article:
In the majority of patients with slipped upper
femoral epiphysis only one hip is involved at primary diagnosis. However,
the contralateral hip often becomes involved over time. There are
no reliable factors predicting a contralateral slip. Whether or
not the contralateral hip should undergo prophylactic fixation is
a matter of controversy. We present a number of essential points
that have to be considered both when choosing to fix the contralateral
hip prophylactically as well as when refraining from surgery and
instead following the patients with repeat radiographs.
We undertook a retrospective cohort study to
determine clinical outcomes following the revision of metal-on-metal (MoM)
hip replacements for adverse reaction to metal debris (ARMD), and
to identify predictors of time to revision and outcomes following
revision. Between 1998 and 2012 a total of 64 MoM hips (mean age
at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings
and 18 (28%) total hip replacements) were revised for ARMD at one specialist
centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision
for ARMD there were 13 hips (20.3%) with post-operative complications
and eight (12.5%) requiring re-revision. The Kaplan–Meier five-year survival rate for ARMD revision was
87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding
re-revisions, the median absolute Oxford hip score (OHS) following
ARMD revision using the percentage method (0% best outcome and 100%
worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%),
which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using
the modified OHS. Histopathological response did not affect time
to revision for ARMD (p = 0.334) or the subsequent risk of re-revision
(p = 0.879). Similarly, the presence or absence of a contralateral
MoM hip bearing did not affect time to revision for ARMD (p = 0.066)
or the subsequent risk of re-revision (p = 0.178). Patients revised to MoM bearings had higher rates of re-revision
(five of 16 MoM hips re-revised; p = 0.046), but those not requiring
re-revision had good functional results (median absolute OHS 14.6%
or 41.0/48). Short-term morbidity following revision for ARMD was
comparable with previous reports. Caution should be exercised when choosing
bearing surfaces for ARMD revisions. Cite this article:
Metal artefact reduction (MAR) MRI is now widely
considered to be the standard for imaging metal-on-metal (MoM) hip
implants. The Medicines and Healthcare Products Regulatory Agency
(MHRA) has recommended cross-sectional imaging for all patients
with symptomatic MoM bearings. This paper describes the natural
history of MoM disease in a 28 mm MoM total hip replacement (THR)
using MAR MRI. Inclusion criteria were patients with MoM THRs who had
not been revised and had at least two serial MAR MRI scans. All
examinations were reported by an experienced observer and classified
as A (normal), B (infection) or C1–C3 (mild, moderate, severe MoM-related
abnormalities). Between 2002 and 2011 a total of 239 MRIs were performed
on 80 patients (two to four scans per THR); 63 initial MRIs (61%)
were normal. On subsequent MRIs, six initially normal scans (9.5%)
showed progression to a disease state; 15 (15%) of 103 THRs with
sequential scans demonstrated worsening disease on subsequent imaging. Most patients with a MoM THR who do not undergo early revision
have normal MRI scans. Late progression (from normal to abnormal,
or from mild to more severe MoM disease) is not common and takes
place over several years. Cite this article:
Despite the worldwide usage of the cemented Contemporary
acetabular component (Stryker), no published data are available
regarding its use in patients aged <
50 years. We undertook a
mid- to long-term follow-up study, including all consecutive patients
aged
<
50 years who underwent a primary total hip replacement using
the Contemporary acetabular component with the Exeter cemented stem
between January 1999 and January 2006. There were 152 hips in 126
patients, 61 men and 65 women, mean age at surgery 37.6 years (16
to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to
12.0). All clinical questionnaire scores, including Harris hip score,
Oxford hip score and several visual analogue scales, were found
to have improved. The eight year survivorship of all acetabular
components for the endpoints revision for any reason or revision
for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2
to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up
was complete for 146 implants. The eight year survival for the endpoint
radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving
implants were considered radiologically loose but were asymptomatic.
The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent
lines (n = 20, 13.9%) in the 144 surviving cups indicates a need
for continued observation in the second decade of follow-up in order
to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival
rate >
90% in young patients support the use of the Contemporary
acetabular component in this specific patient group. Cite this article:
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
New brands of joint prosthesis are released for general implantation with limited evidence of their long-term performance in patients. The CUSUM continuous monitoring method is a statistical testing procedure which could be used to provide prospective evaluation of brands as soon as implantation in patients begins and give early warning of poor performance. We describe the CUSUM and illustrate the potential value of this monitoring tool by applying it retrospectively to the 3M Capital Hip experience. The results show that if the clinical data and methodology had been available, the CUSUM would have given an alert to the underperformance of this prosthesis almost four years before the issue of a Hazard Notice by the Medical Devices Agency. This indicates that the CUSUM can be a valuable tool in monitoring joint prostheses, subject to timely and complete collection of data. Regional or national joint registries provide an opportunity for future centralised, continuous monitoring of all hip and knee prostheses using these techniques.
Pseudotumours (abnormal peri-prosthetic soft-tissue reactions)
following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have
been associated with elevated metal ion levels, suggesting that
excessive wear may occur due to edge-loading of these MoM implants.
This study aimed to quantify The duration and magnitude of edge-loading Objectives
Methods
Juvenile idiopathic arthritis (JIA) is a chronic disease of childhood; it causes joint damage which may require surgical intervention, often in the young adult. The aim of this study was to describe the long-term outcome and survival of hip replacement in a group of adult patients with JIA and to determine predictors of survival for the prosthesis. In this retrospective comparative study patients were identified from the database of a regional specialist adult JIA clinic. This documented a series of 47 hip replacements performed in 25 adult patients with JIA. Surgery was performed at a mean age of 27 years (11 to 47), with a mean follow-up of 19 years (2 to 36). The mean Western Ontario and McMaster Universities osteoarthritis index questionnaire (WOMAC) score at the last follow-up was 53 (19 to 96) and the mean Health Assessment Questionnaire score was 2.25 (0 to 3). The mean pain component of the WOMAC score (60 (20 to 100)) was significantly higher than the mean functional component score (46 (0 to 97)) (p = 0.02). Kaplan-Meier survival analysis revealed a survival probability of 46.6% (95% confidence interval 37.5 to 55.7) at 19 years, with a trend towards enhanced survival with the use of a cemented acetabular component and a cementless femoral component. This was not, however, statistically significant (acetabular component, p = 0.76, femoral component, p = 0.45). Cox’s proportional hazards regression analysis showed an implant survival rate of 54.9% at 19 years at the mean of covariates. Survival of the prosthesis was significantly poorer (p = 0.001) in patients who had been taking long-term corticosteroids and significantly better (p = 0.02) in patients on methotrexate.
We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham hip resurfacing (BHR) in two groups of 54 young patients matched for age, gender, body mass index and pre-operative levels of activity. The clinical outcome was assessed by the University of California, Los Angeles (UCLA) activity score, the Oxford Hip Score (OHS) and the EuroQol scores. Radiologically, all hips were assessed for migration and osteolysis, the hybrid THRs for polyethylene wear and the BHRs for a pedestal sign. The mean follow-up of the patients with a hybrid THR was ten years and for those with a BHR, nine years. Four patients with a hybrid THR and one with a BHR had died. In each group five were lost to follow-up. The revision rate of the hybrid THRs was 16.7% (9 of 54) and of the BHRs 9.3% (5 of 54) (p = 0.195). Radiographs of a further eight hybrid THRs demonstrated wear and osteolysis, and they await revision (p = 0.008). Of the unrevised BHRs 90% had radiological changes, of which approximately 50% had progressed over the previous four years. All hybrid THRs demonstrated linear polyethylene wear with a mean of 1.24 mm (0.06 to 3.03). The BHRs recorded superior OHS (p = 0.013), UCLA (p = 0.008), and EuroQol visual analogue scores (p = 0.009). After nine years, patients with BHRs remained more active and had a lower rate of revision than those with hybrid THRs. Both groups demonstrated progressive radiological changes at medium-term follow-up.
Lately, concerns have arisen following the use of large metal-on-metal bearings in hip replacements owing to reports of catastrophic soft-tissue reactions resulting in implant failure and associated complications. This review examines the literature and contemporary presentations on current clinical dilemmas in metal-on-metal hip replacement.
We report the outcome of total hip replacement in 29 failed metal-on-metal resurfacing hip replacements in which the primary surgery was performed between August 1995 and February 2005. The mean length of follow-up was five years (1.7 to 11.7). Of the 29 hip resurfacings, 19 acetabular components and all the femoral components were revised (28 uncemented stems and one cemented stem). There were no deaths and none of the patients was lost to follow-up. None of the hips underwent any further revision. The results of the revision resurfacing group were compared with those of a control group of age-matched patients. In the latter group there were 236 primary total hip replacements and 523 resurfacings performed during the same period by the same surgeons. The outcome of the revision resurfacing group was comparable with that of the stemmed primary hip replacement group but was less good than that of the primary hip resurfacing group. Long-term follow-up is advocated to monitor the outcome of these cases.
We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an uncemented Allofit acetabular component with a Metasul liner. A total of 15 patients (16 hips) died before their follow-up at ten years and three were lost to follow-up. The study group therefore comprised 82 patients (85 hips). The mean Oxford score at ten years was 20.7 (12 to 42). Six of 85 hips required revision surgery. One was performed because of infection, one for aseptic loosening of the acetabular component and four because of unexplained pain. Histological examination showed an aseptic lymphocytic vasculitis associated lesion-type tissue response in two of these. Continued follow-up is advocated in order to monitor the long-term performance of the Metasul bearing and tissue responses to metal debris.
The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis. We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years. After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial. We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.
We have analysed the management and clinical outcome of a series of consecutive patients who had a total hip replacement and developed post-operative surgical site infection (SSI) with methicillin-resistant Methicillin-resistant
We have analysed from the Finnish Arthroplasty Register the long-term survivorship of the 12 most commonly-used cemented implants between 1980 and 2005 in patients aged 55 years or older with osteoarthritis. Only two designs of femoral component, the Exeter Universal and the Müller Straight femoral component had a survivorship of over 95% at ten years with revision for aseptic loosening as the endpoint. At 15 years of the femoral and acetabular component combinations, only the Exeter Universal/Exeter All-poly implant had a survival rate of over 90% with revision for aseptic loosening as the endpoint. In the subgroup of patients aged between 55 and 64 years, survivorship overall was less than 90% at ten years. The variation in the long-term rates of survival of different cemented hip implants was considerable in patients aged 55 years or older. In those aged between 55 and 64 years, none of the cemented prostheses studied yielded excellent long-term survival rates (≥ 90% at 15 years).
We assessed the outcome of patients with Vancouver type B2 and B3 periprosthetic fractures treated with femoral revision using an uncemented extensively porous-coated implant. A retrospective clinical and radiographic assessment of 22 patients with a mean follow-up of 33.7 months was performed. The mean time from the index procedure to fracture was 10.8 years. There were 17 patients with a satisfactory result. Complications in four patients included subsidence in two, deep sepsis in one, and delayed union in one. Concomitant acetabular revision was required in 19 patients. Uncemented extensively porous-coated femoral stems incorporate distally allowing stable fixation. We found good early survival rates and a low incidence of nonunion using this implant.