Advertisement for orthosearch.org.uk
Results 1 - 47 of 47
Results per page:
Bone & Joint Open
Vol. 3, Issue 10 | Pages 795 - 803
12 Oct 2022
Liechti EF Attinger MC Hecker A Kuonen K Michel A Klenke FM

Aims. Traditionally, total hip arthroplasty (THA) templating has been performed on anteroposterior (AP) pelvis radiographs. Recently, additional AP hip radiographs have been recommended for accurate measurement of the femoral offset (FO). To verify this claim, this study aimed to establish quantitative data of the measurement error of the FO in relation to leg position and X-ray source position using a newly developed geometric model and clinical data. Methods. We analyzed the FOs measured on AP hip and pelvis radiographs in a prospective consecutive series of 55 patients undergoing unilateral primary THA for hip osteoarthritis. To determine sample size, a power analysis was performed. Patients’ position and X-ray beam setting followed a standardized protocol to achieve reproducible projections. All images were calibrated with the KingMark calibration system. In addition, a geometric model was created to evaluate both the effects of leg position (rotation and abduction/adduction) and the effects of X-ray source position on FO measurement. Results. The mean FOs measured on AP hip and pelvis radiographs were 38.0 mm (SD 6.4) and 36.6 mm (SD 6.3) (p < 0.001), respectively. Radiological view had a smaller effect on FO measurement than inaccurate leg positioning. The model showed a non-linear relationship between projected FO and femoral neck orientation; at 30° external neck rotation (with reference to the detector plane), a true FO of 40 mm was underestimated by up to 20% (7.8 mm). With a neutral to mild external neck rotation (≤ 15°), the underestimation was less than 7% (2.7 mm). The effect of abduction and adduction was negligible. Conclusion. For routine THA templating, an AP pelvis radiograph remains the gold standard. Only patients with femoral neck malrotation > 15° on the AP pelvis view, e.g. due to external rotation contracture, should receive further imaging. Options include an additional AP hip view with elevation of the entire affected hip to align the femoral neck more parallel to the detector, or a CT scan in more severe cases. Cite this article: Bone Jt Open 2022;3(10):795–803


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 852 - 858
1 Jul 2022
Grothe T Günther K Hartmann A Blum S Haselhoff R Goronzy J

Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Methods. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti. 12. Mo. 6. Zr. 2. Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients. Results. A total of 20 ALTRs were identified on MRI, with an incidence of 26%. Patients with ALTR had significantly higher median Co levels compared with those without ALTR (2.96 μg/l (interquartile range (IQR) 1.35 to 4.98) vs 1.44 μg/l (IQR 0.79 to 2.5); p = 0.019). Radiological evidence of osteolysis was also significantly associated with ALTR (p = 0.014). Median Cr levels were not significantly higher in those with ALTR compared with those without one (0.97 μg/l (IQR 0.72 to 1.9) v 0.67 μg/l (IQR 0.5 to 1.19; p = 0.080). BMI, sex, age, type of femoral component, head length, the inclination of the acetabular component, and heterotopic ossification formation showed no significant relationship with ALTR. Conclusion. Due to the high incidence of local ALTR in our cohort after more than ten years postoperatively, we recommend regular follow-up investigation even in asymptomatic patients with V40 taper and metal heads. As cobalt levels correlate with ALTR occurrence, routine metal ion screening and consecutive MRI investigation upon elevation could be discussed. Cite this article: Bone Joint J 2022;104-B(7):852–858


Bone & Joint Research
Vol. 7, Issue 6 | Pages 388 - 396
1 Jun 2018
Langton DJ Sidaginamale RP Joyce TJ Bowsher JG Holland JP Deehan D Nargol AVF Natu S

Objectives. We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). Patients and Methods. We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties. Results. In the resurfacing cohort (n = 76), the statistical modelling indicated that the presence of severe ALVAL and a large fluid collection were associated with greater joint fluid Co concentrations after adjustment for volumetric wear rates (p = 0.005). These findings were replicated in the mixed implant group (n = 178), where the presence of severe ALVAL and a large fluid collection were significantly associated with greater fluid Co concentrations (p < 0.001). Conclusion. The development of severe ALVAL is associated with elevations in metal ion concentrations far beyond those expected from the volumetric loss from the prosthetic surfaces. This finding may aid the understanding of the sequence of events leading to soft-tissue reactions following MoM hip arthroplasties. Cite this article: D. J. Langton, R. P. Sidaginamale, T. J. Joyce, J. G. Bowsher, J. P. Holland, D. Deehan, A. V. F. Nargol, S. Natu. Aseptic lymphocyte-dominated vasculitis-associated lesions are related to changes in metal ion handling in the joint capsules of metal-on-metal hip arthroplasties. Bone Joint Res 2018;7:388–396. DOI: 10.1302/2046-3758.76.BJR-2018-0037


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 365 - 371
1 Apr 2019
Nam D Salih R Nahhas CR Barrack RL Nunley RM

Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (. sd. 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis. Results. Mean Harris Hip Scores improved from 54.1 (. sd. 20.5) to 91.2 (. sd. 10.8) at two years postoperatively (p < 0.001). All patients had radiologically well-fixed components, no patients experienced any instability, and no patients required any further intervention. Mean cobalt levels increased from 0.065 ug/l (. sd. 0.03) preoperatively to 0.30 ug/l (. sd. 0.51) at one year postoperatively (p = 0.01) but decreased at two years postoperatively to 0.16 ug/l (. sd. 0.23; p = 0.2). Four patients (9.3%) had a cobalt level outside the reference range (0.03 ug/l to 0.29 ug/l) at two years postoperatively, with values from 0.32 ug/l to 0.94 ug/l. The mean femoral BMD ratio was maintained in Gruen zones 2 to 7 at both one and two years postoperatively using this stem design. At two years postoperatively, mean BMD in the medial calcar was 101.5% of the baseline value. Conclusion. Use of a modular DM prosthesis and cementless, tapered femoral stem has shown encouraging results in young, active patients undergoing primary THA. Elevation in mean cobalt levels and the presence of four patients outside the reference range at two years postoperatively demonstrates the necessity of continued surveillance in this cohort. Cite this article: Bone Joint J 2019;101-B:365–371


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 775 - 782
1 Jul 2023
Koper MC Spek RWA Reijman M van Es EM Baart SJ Verhaar JAN Bos PK

Aims

The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate.

Methods

A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1298 - 1303
1 Oct 2017
Schouten R Malone AA Frampton CM Tiffen C Hooper G

Aims . The primary aim of this independent prospective randomised trial was to compare serum metal ion levels for ceramic-on-metal (CoM) and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty (THA). Our one-year results demonstrated elevation in metal ion levels above baseline with no significant difference between the CoM and MoM groups. This paper reviews the five-year data. Patients and Methods. The implants used in each patient differed only in respect to the type of femoral head (ceramic or metal). At five-year follow-up of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available for comparison. Results. The mean serum cobalt (Co) and chromium (Cr) ion levels remained above baseline in both groups (CoM: Co 1.16 μg/l (0.41 to 14.67), Cr 1.05 μg/l (0.16 to 12.58); MoM: Co 2.93 μg/l (0.35 to 30.29), Cr 1.85 μg/l (0.36 to 17.00)) but the increase was significantly less in the CoM cohort (Co difference p = 0.001, Cr difference p = 0.002). These medium-term results, coupled with lower revision rates from national joint registries, suggest that the performance of CoM THA may be superior to that of MoM. . Conclusion. While both bearing combinations have since been withdrawn these results provide useful information for planning clinical surveillance of CoM THAs and warrants continued monitoring. Cite this article: Bone Joint J 2017;99-B:1298–1303


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 193 - 199
1 Feb 2022
Wang Q Wang H A G Xiao T Kang P

Aims

This study aimed to use intraoperative free electromyography to examine how the placement of a retractor at different positions along the anterior acetabular wall may affect the femoral nerve during total hip arthroplasty (THA) when undertaken using the direct anterior approach (THA-DAA).

Methods

Intraoperative free electromyography was performed during primary THA-DAA in 82 patients (94 hips). The highest position of the anterior acetabular wall was defined as the “12 o’clock” position (middle position) when the patient was in supine position. After exposure of the acetabulum, a retractor was sequentially placed at the ten, 11, 12, one, and two o’clock positions (right hip; from superior to inferior positions). Action potentials in the femoral nerve were monitored with each placement, and the incidence of positive reactions (defined as explosive, frequent, or continuous action potentials, indicating that the nerve was being compressed) were recorded as the primary outcome. Secondary outcomes included the incidence of positive reactions caused by removing the femoral head, and by placing a retractor during femoral exposure; and the incidence of femoral nerve palsy, as detected using manual testing of the strength of the quadriceps muscle.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 42 - 49
1 Apr 2012
Kwon Y Mellon SJ Monk P Murray DW Gill HS

Objectives. Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities. Methods. The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient’s specific hip joint centre, based on CT scans. Results. Edge-loading in the hips with pseudotumours occurred with a four-fold increase in duration and magnitude of force compared with the hips without pseudotumours (p = 0.02). Conclusions. The study provides the first in vivo evidence to support that edge-loading is an important mechanism that leads to localised excessive wear (edge-wear), with subsequent elevation of metal ion levels in MoMHRA patients with pseudotumours


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 301 - 305
1 Mar 2007
Ziaee H Daniel J Datta AK Blunt S McMinn DJW

Metal-on-metal bearings are being increasingly used in young patients. The potential adverse effects of systemic metal ion elevation are the subject of ongoing investigation. The purpose of this study was to investigate whether cobalt and chromium ions cross the placenta of pregnant women with a metal-on-metal hip resurfacing and reach the developing fetus. Whole blood levels were estimated using high-resolution inductively-coupled plasma mass spectrometry. Our findings showed that cobalt and chromium are able to cross the placenta in the study patients with metal-on-metal hip resurfacings and in control subjects without any metal implants. In the study group the mean concentrations of cobalt and chromium in the maternal blood were 1.39 μg/l (0.55 to 2.55) and 1.28 μg/l (0.52 to 2.39), respectively. The mean umbilical cord blood concentrations of cobalt and chromium were comparatively lower, at 0.839 μg/l (0.42 to 1.75) and 0.378 μg/l (0.14 to 1.03), respectively, and this difference was significant with respect to chromium (p < 0.05). In the control group, the mean concentrations of cobalt and chromium in the maternal blood were 0.341 μg/l (0.18 to 0.54) and 0.199 μg/l (0.12 to 0.33), and in the umbilical cord blood they were 0.336 μg/l (0.17 to 0.5) and 0.194 μg/l (0.11 to 0.56), respectively. The differences between the maternal and umbilical cord blood levels in the controls were marginal, and not statistically significant (p > 0.05). The mean cord blood level of cobalt in the study patients was significantly greater than that in the control group (p < 0.01). Although the mean umbilical cord blood chromium level was nearly twice as high in the study patients (0.378 μg/l) as in the controls (0.1934 μg/l), this difference was not statistically significant. (p > 0.05). The transplacental transfer rate was in excess of 95% in the controls for both metals, but only 29% for chromium and 60% for cobalt in study patients, suggesting that the placenta exerts a modulatory effect on the rate of metal ion transfer


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims

Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA.

Methods

This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m2, and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 14 - 20
1 Jan 2016
Zywiel MG Cherian JJ Banerjee S Cheung AC Wong F Butany J Gilbert C Overgaard C Syed K Jacobs JJ Mont MA

As adverse events related to metal on metal hip arthroplasty have been better understood, there has been increased interest in toxicity related to the high circulating levels of cobalt ions. However, distinguishing true toxicity from benign elevations in cobalt levels can be challenging. The purpose of this review is to examine the use of cobalt alloys in total hip arthroplasty, to review the methods of measuring circulating cobalt levels, to define a level of cobalt which is considered pathological and to review the pathophysiology, risk factors and treatment of cobalt toxicity. To the best of our knowledge, there are 18 published cases where cobalt metal ion toxicity has been attributed to the use of cobalt-chromium alloys in hip arthroplasty. Of these cases, the great majority reported systemic toxic reactions at serum cobalt levels more than 100 μg/L. This review highlights some of the clinical features of cobalt toxicity, with the goal that early awareness may decrease the risk factors for the development of cobalt toxicity and/or reduce its severity. Take home message: Severe adverse events can arise from the release of cobalt from metal-on-metal arthroplasties, and as such, orthopaedic surgeons should not only be aware of the presenting problems, but also have the knowledge to treat appropriately. Cite this article: Bone Joint J 2016;98-B:14–20


Bone & Joint Open
Vol. 2, Issue 8 | Pages 611 - 617
10 Aug 2021
Kubik JF Bornes TD Klinger CE Dyke JP Helfet DL

Aims

Surgical treatment of young femoral neck fractures often requires an open approach to achieve an anatomical reduction. The application of a calcar plate has recently been described to aid in femoral neck fracture reduction and to augment fixation. However, application of a plate may potentially compromise the regional vascularity of the femoral head and neck. The purpose of this study was to investigate the effect of calcar femoral neck plating on the vascularity of the femoral head and neck.

Methods

A Hueter approach and capsulotomy were performed bilaterally in six cadaveric hips. In the experimental group, a one-third tubular plate was secured to the inferomedial femoral neck at 6:00 on the clockface. The contralateral hip served as a control with surgical approach and capsulotomy without fixation. Pre- and post-contrast MRI was then performed to quantify signal intensity in the femoral head and neck. Qualitative assessment of the terminal arterial branches to the femoral head, specifically the inferior retinacular artery (IRA), was also performed.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 586 - 590
1 May 2007
Sauvé P Mountney J Khan T De Beer J Higgins B Grover M

Metal-on-metal bearings for total hip replacement (THR) are becoming increasingly popular. Improved wear characteristics mean that these articulations are being inserted into younger patients in the form of THR and resurfacing procedures. This has led to concerns regarding potential carcinogenicity because of the increased exposure to metal ions that the procedure brings. We have studied the serum cobalt and chromium concentrations in patients who had primary, well-fixed Ring metal-on-metal THRs for more than 30 years. The levels of cobalt and chromium were elevated by five and three times, respectively compared with those in our reference groups. Metal-on-metal articulations appear to be the source of metal ions throughout the life of the prosthesis. In three patients who had undergone revision of a previous metal-on-metal THR to a metal-on-polyethylene replacement the levels of metal ions were within the normal range. The elevations of cobalt and chromium ions seen in our study were comparable with those in patients with modern metal-on-metal THRs


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 53 - 58
1 Jul 2021
Lawrie CM Bechtold D Schwabe M Clohisy JC

Aims

The direct anterior approach (DAA) for total hip arthroplasty (THA) has potential advantages over other approaches and is most commonly performed with the patient in the supine position. We describe a technique for DAA THA with the patient in the lateral decubitus position and report the early clinical and radiological outcomes, the characteristics of the learning curve, and perioperative complications.

Methods

All primary DAA THAs performed in the lateral position by a single surgeon over a four-year period from the surgeon’s first case using the technique were identified from a prospectively collected database. Modified Harris Hip Scores (mHHS) were collected to assess clinical outcome, and routine radiological analysis was performed. Retrospective review of the medical records identified perioperative complications, the characteristics of the learning curve, and revisions.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 359 - 365
1 Apr 2004
Sakano S Hasegawa Y Torii Y Kawasaki M Ishiguro N

We reviewed the outcome of curved intertrochanteric varus osteotomy in the treatment of osteonecrosis of the femoral head in 20 hips. A mean varus angulation of 31° was obtained by the osteotomy. The ratio of intact area on the weight-bearing portion increased from 19% to 61%. The mean elevation and lateral displacement of the greater trochanter were 1.2 cm and 0.5 cm, respectively. These changes in the position of the greater trochanter were very small when compared with those after conventional varus wedge osteotomy. Nonunion or delayed union was not observed. Quantitative analyses showed aggressive bone remodelling in the medial intertrochanteric region. Eighteen hips survived without collapse after a mean follow-up of 48 months. We conclude that curved varus osteotomy can be used to preserve the hip joint in patients with osteonecrosis of the femoral head


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 57 - 61
1 Jul 2020
Woolson ST

Aims

To establish whether there was a consensus among the members of the Hip Society (HS) on the role of direct anterior approach (DAA) contemporary primary total hip arthroplasty (THA).

Methods

An online survey was sent to all 112 active and senior members of the HS, to which 71 members responded. The survey was constructed to determine whether they believed that evidence-based medicine proves, in modern clinical practice, that the DAA has significant benefits compared to risks when contrasted with other approaches. In addition, they were asked if they currently used the DAA.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 57 - 61
1 Jun 2019
Chalmers BP Mangold DG Hanssen AD Pagnano MW Trousdale RT Abdel MP

Aims

Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR.

Patients and Methods

Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 378 - 385
1 Apr 2019
García-Rey E Carbonell-Escobar R Cordero-Ampuero J García-Cimbrelo E

Aims

We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up.

Patients and Methods

Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).


Bone & Joint Research
Vol. 8, Issue 3 | Pages 146 - 155
1 Mar 2019
Langton DJ Natu S Harrington CF Bowsher JG Nargol AVF

Objectives

We investigated the reliability of the cobalt-chromium (CoCr) synovial joint fluid ratio (JFR) in identifying the presence of a severe aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response and/or suboptimal taper performance (SOTP) following metal-on-metal (MoM) hip arthroplasty. We then examined the possibility that the CoCr JFR may influence the serum partitioning of Co and Cr.

Methods

For part A, we included all revision surgeries carried out at our unit with the relevant data, including volumetric wear analysis, joint fluid (JF) Co and Cr concentrations, and ALVAL grade (n = 315). Receiver operating characteristic curves were constructed to assess the reliability of the CoCr JFR in identifying severe ALVAL and/or SOTP. For part B, we included only patients with unilateral prostheses who had given matched serum and whole blood samples for Co and Cr analysis (n = 155). Multiple regression was used to examine the influence of JF concentrations on the serum partitioning of Co and Cr in the blood.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 898 - 902
1 Jul 2018
Lachiewicz PF O’Dell JA

Aims

To report our experience with trunnion corrosion following metal-on-polyethylene total hip arthroplasty, in particular to report the spectrum of presentation and determine the mean time to presentation.

Patients and Methods

We report the presenting symptoms and signs, intraoperative findings, and early results and complications of operative treatment in nine patients with a mean age of 74 years (60 to 86). The onset of symptoms was at a mean of seven years (3 to 18) after index surgery.


Bone & Joint Research
Vol. 7, Issue 1 | Pages 85 - 93
1 Jan 2018
Saleh A George J Faour M Klika AK Higuera CA

Objectives

The diagnosis of periprosthetic joint infection (PJI) is difficult and requires a battery of tests and clinical findings. The purpose of this review is to summarize all current evidence for common and new serum biomarkers utilized in the diagnosis of PJI.

Methods

We searched two literature databases, using terms that encompass all hip and knee arthroplasty procedures, as well as PJI and statistical terms reflecting diagnostic parameters. The findings are summarized as a narrative review.


The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used.

The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm3 (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r² = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient’s mass, age and body mass index) did not correlate with the amount of material loss observed (p >  0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis.

Cite this article: Bone Joint J 2015;97-B:1350–7.


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1024 - 1030
1 Aug 2015
Whitehouse MR Endo M Zachara S Nielsen TO Greidanus NV Masri BA Garbuz DS Duncan CP

Adverse reaction to wear and corrosion debris is a cause for concern in total hip arthroplasty (THA). Modular junctions are a potential source of such wear products and are associated with secondary pseudotumour formation.

We present a consecutive series of 17 patients treated at our unit for this complication following metal-on-highly cross-linked polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as infection, and present the aggregate laboratory results and pathological findings in this series.

The clinical presentation was pain, swelling or instability. Solid, cystic and mixed soft-tissue lesions were noted on imaging and confirmed intra-operatively. Corrosion at the head–neck junction was noted in all cases. No bacteria were isolated on multiple pre- and intra-operative samples yet the mean erythrocyte sedimentation rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and stromal polymorphonuclear cell counts were noted in nine cases.

Adverse soft–tissue reactions can occur in MoP THA owing to corrosion products released from the head–neck junction. The diagnosis should be carefully considered when investigating pain after THA. This may avoid the misdiagnosis of periprosthetic infection with an unidentified organism and mitigate the unnecessary management of these cases with complete single- or two-stage exchange.

Cite this article: Bone Joint J 2015;97-B:1024–1030.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 68 - 72
1 Jan 2016
Goodman GP Engh Jr CA

The custom triflange is a patient-specific implant for the treatment of severe bone loss in revision total hip arthroplasty (THA). Through a process of three-dimensional modelling and prototyping, a hydroxyapatite-coated component is created for acetabular reconstruction. There are seven level IV studies describing the clinical results of triflange components. The most common complications include dislocation and infection, although the rates of implant removal are low. Clinical results are promising given the challenging problem. We describe the design, manufacture and implantation process and review the clinical results, contrasting them to other methods of acetabular reconstruction in revision THA.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):68–72.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 326 - 333
1 Mar 2016
Morvan A Moreau S Combourieu B Pansard E Marmorat JL Carlier R Judet T Lonjon G

Aims

The primary aim of this study was to analyse the position of the acetabular and femoral components in total hip arthroplasty undertaken using an anterior surgical approach.

Patients and Methods

In a prospective, single centre study, we used the EOS imaging system to analyse the position of components following THA performed via the anterior approach in 102 patients (103 hips) with a mean age of 64.7 years (sd 12.6). Images were taken with patients in the standing position, allowing measurement of both anatomical and functional anteversion of the acetabular component.


The Bone & Joint Journal
Vol. 97-B, Issue 10 | Pages 1345 - 1349
1 Oct 2015
Regev GJ Drexler M Sever R Dwyer T Khashan M Lidar Z Salame K Rochkind S

Sciatic nerve palsy following total hip arthroplasty (THA) is a relatively rare yet potentially devastating complication. The purpose of this case series was to report the results of patients with a sciatic nerve palsy who presented between 2000 and 2010, following primary and revision THA and were treated with neurolysis. A retrospective review was made of 12 patients (eight women and four men), with sciatic nerve palsy following THA. The mean age of the patients was 62.7 years (50 to 72; standard deviation 6.9). They underwent interfascicular neurolysis for sciatic nerve palsy, after failing a trial of non-operative treatment for a minimum of six months. Following surgery, a statistically and clinically significant improvement in motor function was seen in all patients. The mean peroneal nerve score function improved from 0.42 (0 to 3) to 3 (1 to 5) (p < 0.001). The mean tibial nerve motor function score improved from 1.75 (1 to 4) to 3.92 (3 to 5) (p = 0.02).The mean improvement in sensory function was a clinically negligible 1 out of 5 in all patients. In total, 11 patients reported improvement in their pain following surgery.

We conclude that neurolysis of the sciatic nerve has a favourable prognosis in patients with a sciatic nerve palsy following THA. Our findings suggest that surgery should not be delayed for > 12 months following injury.

Cite this article: Bone Joint J 2015;97-B:1345–9


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1618 - 1622
1 Dec 2014
von Roth P Abdel MP Wauer F Winkler T Wassilew G Diederichs G Perka C

Intact abductors of the hip play a crucial role in preventing limping and are known to be damaged through the direct lateral approach. The extent of trauma to the abductors after revision total hip replacement (THR) is unknown. The aim of this prospective study was to compare the pre- and post-operative status of the gluteus medius muscle after revision THR. We prospectively compared changes in the muscle and limping in 30 patients who were awaiting aseptic revision THR and 15 patients undergoing primary THR. The direct lateral approach as described by Hardinge was used for all patients. MRI scans of the gluteus medius and functional analyses were recorded pre-operatively and six months post-operatively. The overall mean fatty degeneration of the gluteus medius increased from 35.8% (1.1 to 98.8) pre-operatively to 41% (1.5 to 99.8) after multiple revision THRs (p = 0.03). There was a similar pattern after primary THR, but with considerably less muscle damage (p = 0.001), indicating progressive muscle damage. Despite an increased incidence of a positive Trendelenburg sign following revision surgery (p = 0.03) there was no relationship between the cumulative fatty degeneration in the gluteus medius and a positive Trendelenburg sign (p = 0.26). The changes associated with other surgical approaches to the hip warrant investigation.

Cite this article: Bone Joint J 2014;96-B:1618–22.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 88 - 91
1 Nov 2013
Su EP Su SL

Surface hip replacement (SHR) is generally used in younger, active patients as an alternative conventional total hip replacement in part because of the ability to preserve femoral bone. This major benefit of surface replacement will only hold true if revision procedures of SHRs are found to provide good clinical results.

A retrospective review of SHR revisions between 2007 and 2012 was presented, and the type of revision and aetiologies were recorded. There were 55 SHR revisions, of which 27 were in women. At a mean follow-up of 2.3 years (0.72 to 6.4), the mean post-operative Harris hip score (HHS) was 94.8 (66 to 100). Overall 23 were revised for mechanical reasons, nine for impingement, 13 for metallosis, nine for unexplained pain and one for sepsis. Of the type of revision surgery performed, 14 were femoral-only revisions; four were acetabular-only revisions, and 37 were complete revisions.

We did not find that clinical scores were significantly different between gender or different types of revisions. However, the mean post-operative HHS was significantly lower in patients revised for unexplained pain compared with patients revised for mechanical reasons (86.9 (66 to 100) versus 99 (96 to 100); p = 0.029). There were two re-revisions for infection in the entire cohort.

Based on the overall clinical results, we believe that revision of SHR can have good or excellent results and warrants a continued use of the procedure in selected patients. Close monitoring of these patients facilitates early intervention, as we believe that tissue damage may be related to the duration of an ongoing problem. There should be a low threshold to revise a surface replacement if there is component malposition, rising metal ion levels, or evidence of soft-tissue abnormalities.

Cite this article: Bone Joint J 2013;95-B, Supple A:88–91.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1617 - 1625
1 Dec 2013
Schmitz MWJL Timmer C Rijnen WHC Gardeniers JWM Schreurs BW

Despite the worldwide usage of the cemented Contemporary acetabular component (Stryker), no published data are available regarding its use in patients aged < 50 years. We undertook a mid- to long-term follow-up study, including all consecutive patients aged < 50 years who underwent a primary total hip replacement using the Contemporary acetabular component with the Exeter cemented stem between January 1999 and January 2006. There were 152 hips in 126 patients, 61 men and 65 women, mean age at surgery 37.6 years (16 to 49 yrs). One patient was lost to follow-up.

Mean clinical follow-up of all implants was 7.6 years (0.9 to 12.0). All clinical questionnaire scores, including Harris hip score, Oxford hip score and several visual analogue scales, were found to have improved. The eight year survivorship of all acetabular components for the endpoints revision for any reason or revision for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2 to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up was complete for 146 implants. The eight year survival for the endpoint radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving implants were considered radiologically loose but were asymptomatic. The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent lines (n = 20, 13.9%) in the 144 surviving cups indicates a need for continued observation in the second decade of follow-up in order to observe their influence on long-term survival.

The clinical and radiological data resulting in a ten-year survival rate > 90% in young patients support the use of the Contemporary acetabular component in this specific patient group.

Cite this article: Bone Joint J 2013;95-B:1617–25.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1464 - 1473
1 Nov 2013
Vendittoli P Rivière C Roy AG Barry J Lusignan D Lavigne M

A total of 219 hips in 192 patients aged between 18 and 65 years were randomised to 28-mm metal-on-metal uncemented total hip replacements (THRs, 107 hips) or hybrid hip resurfacing (HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3) there was no significant difference between the THR and HR groups regarding rate of revision (4.0% (4 of 99) vs 5.8% (6 of 104), p = 0.569) or re-operation rates without revision (5.1% (5 of 99) vs 2.9% (3 of 104), p = 0.428). In the THR group one recurrent dislocation, two late deep infections and one peri-prosthetic fracture required revision, whereas in the HR group five patients underwent revision for femoral head loosening and one for adverse reaction to metal debris. The mean University of California, Los Angeles activity scores were significantly higher in HR (7.5 (sd 1.7) vs 6.9 (sd 1.7), p = 0.035), but similar mean Western Ontario and McMaster Universities Osteoarthritis Index scores were obtained (5.8 (sd 9.5) in HR vs 5.1 (sd 8.9) in THR, p = 0.615) at the last follow-up. Osteolysis was found in 30 of 81 THR patients (37.4%), mostly in the proximal femur, compared with two of 83 HR patients (2.4%) (p < 0.001). At five years the mean metal ion levels were < 2.5 μg/l for cobalt and chromium in both groups; only titanium was significantly higher in the HR group (p = 0.001). Although revision rates and functional scores were similar in both groups at mid-term, long-term survival analysis is necessary to determine whether one procedure is more advantageous than the other.

Cite this article: Bone Joint J 2013;95-B:1464–73.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 570 - 573
1 Apr 2012
Whittingham-Jones P Mann B Coward P Hart AJ Skinner JA

Fracture of a ceramic component in total hip replacement is a rare but potentially catastrophic complication. The incidence is likely to increase as the use of ceramics becomes more widespread. We describe such a case, which illustrates how inadequate initial management will lead to further morbidity and require additional surgery. We present the case as a warning that fracture of a ceramic component should be revised to another ceramic-on-ceramic articulation in order to minimise the risk of further catastrophic wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1611 - 1617
1 Dec 2012
Jameson SS Baker PN Mason J Gregg PJ Brewster N Deehan DJ Reed MR

Despite excellent results, the use of cemented total hip replacement (THR) is declining. This retrospective cohort study records survival time to revision following primary cemented THR using the most common combination of components that accounted for almost a quarter of all cemented THRs, exploring risk factors independently associated with failure. All patients with osteoarthritis who had an Exeter V40/Contemporary THR (Stryker) implanted before 31 December 2010 and recorded in the National Joint Registry for England and Wales were included in the analysis. Cox’s proportional hazard models were used to analyse the extent to which risk of revision was related to patient, surgeon and implant covariates, with a significance threshold of p < 0.01. A total of 34 721 THRs were included in the study. The overall seven-year rate of revision for any reason was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final adjusted model the risk of revision was significantly higher in THRs with the Contemporary hooded component (hazard ratio (HR) 1.88, p < 0.001) than with the flanged version, and in smaller head sizes (< 28 mm) compared with 28 mm diameter heads (HR 1.50, p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69 to 1.63) with a 28 mm diameter head and flanged component. The overall risk of revision was independent of age, gender, American Society of Anesthesiologists grade, body mass index, surgeon volume, surgical approach, brand of cement/presence of antibiotic, femoral head material (stainless steel/alumina) and stem taper size/offset. However, the risk of revision for dislocation was significantly higher with a ‘plus’ offset head (HR 2.05, p = 0.003) and a hooded acetabular component (HR 2.34, p < 0.001).

In summary, we found that there were significant differences in failure between different designs of acetabular component and sizes of femoral head after adjustment for a range of covariates.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 725 - 729
1 Jun 2007
Ikemura S Yamamoto T Jingushi S Nakashima Y Mawatari T Iwamoto Y

Transtrochanteric curved varus osteotomy was designed to avoid some of the disadvantages of varus wedge osteotomy, such as post-operative leg-length discrepancy. In this retrospective study we investigated the leg-length discrepancy and clinical outcome after transtrochanteric curved varus osteotomy undertaken in patients with osteonecrosis of the femoral head. Between January 1993 and March 2004, this osteotomy was performed in 42 hips of 36 patients with osteonecrosis of the femoral head. There were 15 males and 21 females with a mean age at surgery of 34 years (15 to 68). The mean follow-up was 5.9 years (2.0 to 12.5). The mean pre-operative Harris hip score was 64.0 (43 to 85) points, which improved to a mean of 88.7 (58 to 100) points at final follow-up. The mean varus angulation post-operatively was 25° (12° to 38°) and the post-operative mean leg-length discrepancy was 13 mm (4 to 25). The post-operative leg-length discrepancy showed a strong correlation with varus angulation (Pearson’s correlation coefficient; r = 0.9530, p < 0.0001), which may be useful for predicting the leg-length discrepancy which can occur even after transtrochanteric curved varus osteotomy.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 471 - 476
1 Apr 2012
Holland JP Langton DJ Hashmi M

We present the clinical results and survivorship of consecutive 100 Birmingham Hip Resurfacings in 90 patients at a minimum follow-up of ten years. All procedures were carried out by an independent surgeon who commenced a prospective study in 1998. Patients were assessed clinically using the Western Ontario and McMaster Universities osteoarthritis index, Short-Form 36, Harris hip score and University of California, Los Angeles activity score. Radiological analysis was performed by an independent observer and blood metal ion levels concentrations were measured at ten years post-operatively in 62 patients. The median acetabular component inclination was 46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median chromium concentration in the unilateral group was 1.74 (0.41 to 15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38 to 19.32).

In total there were eight failures giving an overall survival at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3). BHR in male patients had an improved survivorship of 94.6% (95% CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5), but this did not reach statistical significance (p = 0.119). Four of the nine BHRs with a bearing diameter of 42 mm failed.

The overall results were consistent with data produced from other centres in that the clinical outcome of large male patients was extremely encouraging, whereas the survival of the smaller joints was less satisfactory.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1001 - 1004
1 Aug 2011
Fisher J

Bioengineering reasons for increased wear and failure of metal-on-metal (MoM) bearings in hip prostheses have been described. Low wear occurs in MoM hips when the centre of the femoral head is concentric with the centre of the acetabular component and the implants are correctly positioned. Translational or rotational malpositioning of the components can lead to the contact-patch of the femoral component being displaced to the rim of the acetabular component, resulting in a ten- to 100-fold increase in wear and metal ion levels. This may cause adverse tissue reactions, loosening of components and failure of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1351 - 1355
1 Oct 2010
Veitch SW Whitehouse SL Howell JR Hubble MJW Gie GA Timperley AJ

We report the outcome of the flangeless, cemented all-polyethylene Exeter acetabular component at a mean of 14.6 years (10 to 17) after operation. Of the 263 hips in 243 patients, 122 prostheses are still in situ; 112 patients (119 hips) have died, 18 hips have been revised, and three patients (four hips) were lost to follow-up. Radiographs at the final review were available for 110 of the 122 surviving hips. There were acetabular radiolucent lines in 54 hips (49%). Two acetabular components had migrated but neither patient required revision. The Kaplan-Meier survivorship at 15 years with 61 hips at risk with revision for any cause as the endpoint was 89.9% (95% confidence interval (CI) 84.6 to 95.2) and for aseptic loosening of the acetabular component or lysis 91.7% (95% CI 86.6 to 96.8). In 210 hips with a diagnosis of primary osteoarthritis, survivorship with revision for any cause as the endpoint was 93.2% (95% CI 88.1 to 98.3), and for aseptic loosening of the acetabular component 95.0% (95% CI 90.3 to 99.7).

The cemented all-polyethylene Exeter acetabular component has an excellent long-term survivorship.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 781 - 786
1 Jun 2010
Zhao G Yamamoto T Ikemura S Motomura G Mawatari T Nakashima Y Iwamoto Y

We investigated the factors related to the radiological outcome of a transtrochanteric curved varus osteotomy in patients with osteonecrosis of the hip. We reviewed 73 hips in 62 patients with a mean follow-up of 12.4 years (5 to 31.1). There were 28 men and 34 women, with a mean age of 33.3 years (15 to 68) at the time of surgery. The 73 hips were divided into two groups according to their radiological findings: group 1 showed progression of collapse and/or joint-space narrowing; group 2 had neither progressive collapse nor joint-space narrowing. Both of these factors and the radiological outcomes were analysed by a stepwise discriminant analysis. A total of 12 hips were categorised as group 1 and 61 as group 2. Both the post-operative intact ratio and the localisation of the necrotic lesion correlated with the radiological outcome. The cut-off point of the postoperative intact ratio to prevent the progression of collapse was 33.6%, and the cut-off point to prevent both the progression of collapse and joint-space narrowing was 41.9%.

The results of this study indicate that a post-operative intact ratio of 33.0% is necessary if a satisfactory outcome is to be achieved after this varus osteotomy.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 356 - 361
1 Mar 2010
Kwon Y- Glyn-Jones S Simpson DJ Kamali A McLardy-Smith P Gill HS Murray DW

The presence of pseudotumours, which are soft-tissue masses relating to the hip, after metal-on-metal hip resurfacing arthroplasty has been associated with elevated levels of metal ions in serum, suggesting that pseudotumours occur when there is increased wear. We aimed to quantify the wear in vivo of implants revised for pseudotumours (eight) and of a control group of implants (22) revised for other reasons of failure.

We found that the implant group with pseudotumours had a significantly higher rate of median linear wear of the femoral component at 8.1 μm/year (2.75 to 25.4) than the 1.79 μm/year (0.82 to 4.15; p = 0.002) of the non-pseudotumour group. For the acetabular component a significantly higher rate of median linear wear of 7.36 μm/year (1.61 to 24.9) was observed in the pseudotumour group compared with 1.28 μm/year (0.81 to 3.33, p = 0.001) in the other group. Wear of the acetabular component in the pseudotumour group always involved the edge of the implant, indicating that edge-loading had occurred.

Our findings are the first direct evidence that pseudotumour is associated with increased wear at the metal-on-metal articulation. Furthermore, edge-loading with the loss of fluid-film lubrication may be an important mechanism of generation of wear in patients with a pseudotumour.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 12 - 19
1 Jan 2010
Vendittoli P Roy A Mottard S Girard J Lusignan D Lavigne M

We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component.

Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001).

The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 140 - 144
1 Feb 2008
Nagoya S Kaya M Sasaki M Tateda K Yamashita T

We evaluated triple-phase bone scintigraphy in the differential diagnosis of peri-prosthetic infection in 46 patients with a total hip replacement or bipolar hemiarthroplasty who were due for revision surgery. There were 18 men and 28 women, with a mean age at operation of 64.6 years (28 to 81). We defined peri-prosthetic infection as an increased uptake of radioisotope in all the phases of triple-phase bone scintigraphy and validated these results against the histological and/or microbiology results in every case.

The positive and negative predictive values for the presence of infection were 83% and 93%, respectively. The diagnostic sensitivity was 88% and the specificity was 90%.

This study indicates that triple-phase bone scintigraphy is a useful tool in the detection of peri-prosthetic infection and offers a cost-effective method of screening.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1580 - 1583
1 Dec 2006
Ling ZX Kumar VP

We investigated the relationship of the inferior gluteal nerve to gluteus maximus by dissecting the muscle in 12 fresh-frozen and formalin-treated cadavers. The anatomy was recorded using still digital photography. The course of the inferior gluteal nerve was carefully traced and was noted to enter the deep surface of gluteus maximus approximately 5 cm from the tip of the greater trochanter of the femur. The susceptibility of the nerve to injury during a posterior approach to the hip may be explained by its close relationship to the deep surface of gluteus maximus. It is easily damaged before it has been seen if the muscle is split and parted more than 5 cm from the tip of the greater trochanter of the femur. We suggest that a modified posterior approach be used to expose the hip to avoid damage to this nerve.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 715 - 723
1 Jun 2008
McMinn DJW Daniel J Ziaee H Pradhan C

The dysplasia cup, which was devised as an adjunct to the Birmingham Hip Resurfacing system, has a hydroxyapatite-coated porous surface and two supplementary neutralisation screws to provide stable primary fixation, permit early weight-bearing, and allow incorporation of morcellised autograft without the need for structural bone grafting.

A total of 110 consecutive dysplasia resurfacing arthroplasties in 103 patients (55 men and 48 women) performed between 1997 and 2000 was reviewed with a minimum follow-up of six years. The mean age at operation was 47.2 years (21 to 62) and 104 hips (94%) were Crowe grade II or III.

During the mean follow-up of 7.8 years (6 to 9.6), three hips (2.7%) were converted to a total hip replacement at a mean of 3.9 years (2 months to 8.1 years), giving a cumulative survival of 95.2% at nine years (95% confidence interval 89 to 100). The revisions were due to a fracture of the femoral neck, a collapse of the femoral head and a deep infection. There was no aseptic loosening or osteolysis of the acetabular component associated with either of the revisions performed for failure of the femoral component. No patient is awaiting a revision.

The median Oxford hip score in 98 patients with surviving hips at the final review was 13 and the 10th and the 90th percentiles were 12 and 23, respectively.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 16 - 20
1 Jan 2007
Gill HS Campbell PA Murray DW De Smet KA

Resurfacing arthroplasty of the hip is being used increasingly as an alternative to total hip replacement, especially for young active patients. There is concern about necrosis of the femoral head after resurfacing which can result in fracture and loosening. Most systems use a cemented femoral component, with the potential for thermal necrosis of the cancellous bone of the reamed femoral head. We used thermal probes to record temperatures close to the cement-bone interface during resurfacing arthroplasty.

The maximum temperature recorded at the cement-bone interface in four cases was approximately 68°C which was higher than that reported to kill osteocytes. A modified surgical technique using insertion of a suction cannula into the lesser trochanter, generous pulsed lavage and early reduction of the joint significantly reduced the maximum recorded cancellous bone temperature to approximately 36°C in five cases (p = 0.014).

We recommend the modified technique since it significantly reduces temperatures at the cement-bone interface.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 562 - 569
1 May 2008
Daniel J Pradhan A Pradhan C Ziaee H Moss M Freeman J McMinn DJW

We report a retrospective review of the incidence of venous thromboembolism in 463 consecutive patients who underwent primary total hip arthroplasty (487 procedures). Treatment included both total hip replacement and hip resurfacing, and the patients were managed without anticoagulants. The thromboprophylaxis regimen included an antiplatelet agent, generally aspirin, hypotensive epidural anaesthesia, elastic compression stockings and early mobilisation. In 258 of these procedures (244 patients) performed in 2005 (cohort A) mechanical compression devices were not used, whereas in 229 (219 patients) performed during 2006 (cohort B) bilateral intermittent pneumatic calf compression was used.

All operations were performed through a posterior mini-incision approach. Patients who required anticoagulation for pre-existing medical problems and those undergoing revision arthroplasty were excluded. Doppler ultrasonographic screening for deep-vein thrombosis was performed in all patients between the fourth and sixth post-operative days. All patients were reviewed at a follow-up clinic six to ten weeks after the operation. In addition, reponse to a questionnaire was obtained at the end of 12 weeks post-operatively.

No symptomatic calf or above-knee deep-vein thrombosis or pulmonary embolism occurred. In 25 patients in cohort A (10.2%) and in ten patients in cohort B (4.6%) asymptomatic calf deep-vein thromboses were detected ultrasonographically. This difference was statistically significant (p = 0.03). The regimen followed by cohort B offers the prospect of a low incidence of venous thromboembolism without subjecting patients to the higher risk of bleeding associated with anticoagulant use.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 770 - 775
1 Jun 2005
Hsieh P Shih C Chang Y Lee MS Yang W Shih H

We have carried out in 24 patients, a two-stage revision arthroplasty of the hip for infection with massive bone loss. We used a custom-made, antibiotic-loaded cement prosthesis as an interim spacer. Fifteen patients had acetabular deficiencies, eight had segmental femoral bone loss and one had a combined defect.

There was no recurrence of infection at a mean follow-up of 4.2 years (2 to 7). A total of 21 patients remained mobile in the interim period. The mean Merle D’Aubigné and Postel hip score improved from 7.3 points before operation to 13.2 between stages and to 15.8 at the final follow-up. The allograft appeared to have incorporated into the host bone in all patients. Complications included two fractures and one dislocation of the cement prosthesis.

The use of a temporary spacer maintains the function of the joint between stages even when there is extensive loss of bone. Allograft used in revision surgery after septic conditions restores bone stock without the risk of recurrent infection.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 870 - 876
1 Jul 2006
Khan RJK Fick D Alakeson R Haebich S de Cruz M Nivbrant B Wood D

We treated 34 patients with recurrent dislocation of the hip with a constrained acetabular component. Roentgen stereophotogrammetric analysis was performed to assess migration of the prosthesis.

The mean clinical follow-up was 3.0 years (2.2 to 4.8) and the radiological follow-up was 2.7 years (2.0 to 4.8). At the latest review six patients had died and none was lost to follow-up. There were four acetabular revisions, three for aseptic loosening and one for deep infection. Another acetabular component was radiologically loose with progressive radiolucent lines in all Gruen zones and was awaiting revision. The overall rate of aseptic loosening was 11.8% (4 of 34). Roentgen stereophotogrammetric analysis in the non-revised components confirmed migration of up to 1.06 mm of translation and 2.32° of rotation at 24 months. There was one case of dislocation and dissociation of the component in the same patient. Of the 34 patients, 33 (97.1%) had no further episodes of dislocation.

The constrained acetabular component reported in our study was effective in all but one patient with instability of the hip, but the rate of aseptic loosening was higher than has been reported previously and requires further investigation.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 628 - 631
1 May 2005
Dunstan E Sanghrajka AP Tilley S Unwin P Blunn G Cannon SR Briggs TWR

Metal-on-metal hip bearings are being implanted into younger patients. The consequence of elevated levels of potentially carcinogenic metal ions is therefore a cause for concern. We have determined the levels of cobalt (Co), chromium (Cr), titanium (Ti) and vanadium (Va) in the urine and whole blood of patients who had had metal-on-metal and metal-on-polyethylene articulations in situ for more than 30 years. We compared these with each other and with the levels for a control group of subjects.

We found significantly elevated levels of whole blood Ti, Va and urinary Cr in all arthroplasty groups. The whole blood and urine levels of Co were grossly elevated, by a factor of 50 and 300 times respectively in patients with loose metal-on-metal articulations when compared with the control group. Stable metal-on-metal articulations showed much lower levels. Elevated levels of whole blood or urinary Co may be useful in identifying metal-on-metal articulations which are loose.