Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated We found that long-term alumina wear in association with a loose
acetabular component could be categorised into three groups. Of
20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight
‘severe wear’, which was characterised by a change in the physical
shape of the bearing and a loss of volume. This suggests that the
wear in alumina–alumina bearings in association with a loose acetabular
component may be variable in pattern, and may explain, in part,
why the wear of a ceramic head
The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time.Aims
Methods
The aim of this study was to longitudinally compare the clinical and radiological outcomes of anatomical total shoulder arthroplasty (aTSA) up to long-term follow-up, when using cemented keel, cemented peg, and hybrid cage peg glenoid components and the same humeral system. We retrospectively analyzed a multicentre, international clinical database of a single platform shoulder system to compare the short-, mid-, and long-term clinical outcomes associated with three designs of aTSA glenoid components: 294 cemented keel, 527 cemented peg, and 981 hybrid cage glenoids. Outcomes were evaluated at 4,746 postoperative timepoints for 1,802 primary aTSA, with a mean follow-up of 65 months (24 to 217).Aims
Methods
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with
We examined 86 polyethylene inserts, retrieved from total and unicompartmental knee prostheses after an average of 39.5 months in situ, grading them from 0 to 3 for seven modes of polyethylene degradation.
This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA). From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant malposition, infection, and extensor mechanism problems were excluded.Aims
Methods
The aim of this study was to establish the results of isolated exchange of the tibial polyethylene insert in revision total knee arthroplasty (RTKA) in patients with well-fixed femoral or tibial components. We report on a series of RTKAs where only the polyethylene was replaced, and the patients were followed for a mean of 13.2 years (10.0 to 19.1). Our study group consisted of 64 non-infected, grossly stable TKA patients revised over an eight-year period (1998 to 2006). The mean age of the patients at time of revision was 72.2 years (48 to 88). There were 36 females (56%) and 28 males (44%) in the cohort. All patients had received the same cemented, cruciate-retaining patella resurfaced primary TKA. All subsequently underwent an isolated polyethylene insert exchange. The mean time from the primary TKA to RTKA was 9.1 years (2.2 to 16.1).Aims
Patients and Methods
We compared and quantified the modes of failure and patterns of wear of 11 Mittelmeier and 11 Ceraver-Ostal retrieved alumina-alumina hip prostheses with reference to the corresponding clinical and radiological histories. Macroscopic wear was assessed using a three-dimensional co-ordinate measuring machine. Talysurf contacting profilometry was used to measure surface roughness on a microscopic scale and SEM to determine mechanisms of wear at the submicron level. The components were classified into one of three categories of wear: low (no visible/measurable wear), stripe (elliptical wear stripe on the heads and larger worn areas on the cups) and
Osteoarthritis of the medial compartment of the knee often shows a specific pattern of anterior wear. Review of our revisions from a series of medial metal-backed Brigham unicondylar knee replacements performed between 1983 and 1989 showed that this wear pattern was common on the tibial polyethylene surface. We reviewed these cases retrospectively to compare the pattern of preoperative erosion with the wear of the prosthesis. In all 14 knees with
Aims
Patients and Methods
A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
We wished to investigate the influence of metal debris exposure
on the subsequent immune response and resulting soft-tissue injury
following metal-on-metal (MoM) hip arthroplasty. Some reports have
suggested that debris generated from the head-neck taper junction
is more destructive than equivalent doses from metal bearing surfaces. We investigated the influence of the source and volume of metal
debris on chromium (Cr) and cobalt (Co) concentrations in corresponding
blood and hip synovial fluid samples and the observed agglomerated
particle sizes in excised tissues using multiple regression analysis
of prospectively collected data. A total of 199 explanted MoM hips
(177 patients; 132 hips female) were analysed to determine rates
of volumetric wear at the bearing surfaces and taper junctions. Aims
Patients and Methods
Third-body wear is believed to be one trigger for adverse results
with metal-on-metal (MOM) bearings. Impingement and subluxation
may release metal particles from MOM replacements. We therefore
challenged MOM bearings with relevant debris types of cobalt–chrome
alloy (CoCr), titanium alloy (Ti6Al4V) and polymethylmethacrylate
bone cement (PMMA). Cement flakes (PMMA), CoCr and Ti6Al4V particles (size range
5 µm to 400 µm) were run in a MOM wear simulation. Debris allotments
(5 mg) were inserted at ten intervals during the five million cycle
(5 Mc) test. Objectives
Methods
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
We measured the contact areas and contact stresses at the post-cam mechanism of a posterior-stabilised total knee arthroplasty when a posterior force of 500 N was applied to the Kirschner Performance, Scorpio Superflex, NexGen LPS Flex Fixed, and NexGen LPS Flex Mobile knee systems. Measurements were made at 90°, 120°, and 150° of flexion both in neutral rotation and 10° of internal rotation of the tibial component. Peak contact stresses at 90°, 120°, and 150° were 24.0, 33.9, and 28.8 MPa, respectively, for the Kirschner; 26.0, 32.4, and 22.1 MPa, respectively, for the Scorpio; and 34.1, 31.5, and 32.5 MPa, respectively, for the NexGen LPS Flex Fixed. With an internally rotated tibia, the contact stress increased significantly with all the fixed-bearing arthroplasties but not with the NexGen LPS Flex Mobile arthroplasty. The post-cam design should be modified in order to provide a larger contact area whilst avoiding any impingement and edge loading.
We investigated the characteristics of patients
who achieved Japanese-style deep flexion (seiza-sitting) after total knee
replacement (TKR) and measured three-dimensional positioning and
the contact positions of the femoral and tibial components. Seiza-sitting
was achieved after surgery by 23 patients (29 knees) of a series
of 463 TKRs in 341 patients. Pre-operatively most of these patients
were capable of seiza-sitting, had a lower body mass index and a favourable
attitude towards the Japanese lifestyle (27 of 29 knees). According
to two-/three-dimensional image registration analysis in the seiza-sitting
position, flexion, varus and internal rotation angles of the tibial
component relative to the femoral component had means of 148° ( Cite this article:
We report on the long-term results of 163 bicruciate-retaining
Hermes 2C total knee replacements in 130 patients at a mean follow-up
of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament
had a partially degenerative appearance it was preserved as long
as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The
description and surgical technique of this minimally constrained
prosthesis were published in 1983 and the ten-year clinical results
in 1999. A total of 12% of the knees (20 of 163) in this study were revised
because of wear of the polyethylene tibial insert. Excellent stability
was achieved and the incidence of aseptic component loosening was
4.3% (seven of 163). The survival rate using revision for any reason
as the endpoint was 82% (95% confidence interval 76.2 to 88.0). Although this series included a relatively small number of replacements,
it demonstrated that the anterior cruciate ligament, even when partially
degenerated at the time of TKR, remained functional and provided
adequate stability at a long-term follow-up. Cite this article:
The practice of removing a well-fixed cementless
femoral component is associated with high morbidity. Ceramic bearing
couples are low wearing and their use minimises the risk of subsequent
further revision due to the production of wear debris. A total of
165 revision hip replacements were performed, in which a polyethylene-lined acetabular
component was revised to a new acetabular component with a ceramic
liner, while retaining the well-fixed femoral component. A titanium
sleeve was placed over the used femoral trunnion, to which a ceramic
head was added. There were 100 alumina and 65 Delta bearing couples
inserted. The mean Harris hip score improved significantly from 71.3 (9.0
to 100.0) pre-operatively to 91.0 (41.0 to 100.0) at a mean follow
up of 4.8 years (2.1 to 12.5) (p <
0.001). No patients reported
squeaking of the hip. There were two fractures of the ceramic head, both in alumina
bearings. No liners were seen to fracture. No fractures were observed
in components made of Delta ceramic. At 8.3 years post-operatively
the survival with any cause of failure as the endpoint was 96.6%
(95% confidence interval (CI) 85.7 to 99.3) for the acetabular component and
94.0% (95% CI 82.1 to 98.4) for the femoral component. The technique of revising the acetabular component in the presence
of a well-fixed femoral component with a ceramic head placed on
a titanium sleeve over the used trunnion is a useful adjunct in
revision hip practice. The use of Delta ceramic is recommended. Cite this article: