Aims. Our primary aim was to establish the proportion of female orthopaedic consultants who perform arthroplasty via cases submitted to the
The National Joint Registry for England, Wales and Northern Ireland
(NJR) has extended its scope to report on hospital, surgeon and
implant performance. Data linkage of the NJR to the London Implant
Retrieval Centre (LIRC) has previously evaluated data quality for
hip primary procedures, but did not assess revision records. We analysed metal-on-metal hip revision procedures performed
between 2003 and 2013. A total of 69 929 revision procedures from
the NJR and 929 revised pairs of components from the LIRC were included.Aims
Methods
Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost. In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.Aims
Methods
Recent publications have drawn attention to the fact that some brands of joint replacement may contain variants which perform significantly worse (or better) than their ‘siblings’. As a result, the
The June 2024 Shoulder & Elbow Roundup. 360. looks at: Reverse versus anatomical total shoulder replacement for osteoarthritis? A UK national picture; Acute rehabilitation following traumatic anterior shoulder dislocation (ARTISAN): pragmatic, multicentre, randomized controlled trial; acid for rotator cuff repair: a systematic review and meta-analysis of randomized controlled trials; Metal or ceramic humeral head total shoulder arthroplasty: an analysis of data from the
Data of high quality are critical for the meaningful interpretation of registry information. The
The December 2023 Knee Roundup. 360. looks at: Obesity is associated with greater improvement in patient-reported outcomes following primary total knee arthroplasty; Does mild flexion of the femoral prosthesis in total knee arthroplasty result in better early postoperative outcomes?; Robotic or manual total knee arthroplasty: a randomized controlled trial; Patient-relevant outcomes following first revision total knee arthroplasty, by diagnosis: an analysis of implant survivorship, mortality, serious medical complications, and patient-reported outcome measures using the
Aims. Despite the increasing numbers of ankle arthroplasties, there are limited studies on their survival and comparisons between different implants. The primary aim of this study was to determine the failure rates of primary ankle arthroplasties commonly used in the UK. Methods. A data linkage study combined
Aims. The Exeter V40 femoral stem is the most implanted stem in the
Aims. We compared the risks of re-revision and mortality between two-stage revision surgery and single-stage revision surgery among patients with infected primary knee arthroplasty. Methods. Patients with a periprosthetic joint infection (PJI) of their primary knee arthroplasty, initially revised with a single-stage or a two-stage procedure in England and Wales between 2003 and 2014, were identified from the
Aims. The Oxford Shoulder Score (OSS) is a 12-item measure commonly used for the assessment of shoulder surgeries. This study explores whether computerized adaptive testing (CAT) provides a shortened, individually tailored questionnaire while maintaining test accuracy. Methods. A total of 16,238 preoperative OSS were available in the
Aims. The aim of this study was to review the provision of total elbow arthroplasties (TEAs) in England, including the incidence, the characteristics of the patients and the service providers, the types of implant, and the outcomes. Methods. We analyzed the primary TEAs recorded in the
Aims. The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes. Methods. A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the
Aims. This study evaluates the association between consultant and hospital volume and the risk of re-revision and 90-day mortality following first-time revision of primary hip arthroplasty for aseptic loosening. Methods. We conducted a cohort study of first-time, single-stage revision hip arthroplasties (RHAs) performed for aseptic loosening and recorded in the
Aims. Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA. Methods. A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the
Aims. We compared the risks of re-revision and mortality between two-stage and single-stage revision surgeries among patients with infected primary hip arthroplasty. Methods. Patients with a periprosthetic joint infection (PJI) of their primary arthroplasty revised with single-stage or two-stage procedure in England and Wales between 2003 and 2014 were identified from the
Aims. Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the
Aims. When a total ankle arthroplasty (TAA) fails, it can be converted to a fusion or a revision arthroplasty. Despite the increasing numbers of TAAs being undertaken, there is little information in the literature about the management of patients undergoing fusion following a failed TAA. The primary aim of this study was to analyze the survival of fusions following a failed TAA using a large dataset from the
Aims. This study describes the variation in the annual volumes of revision hip arthroplasty (RHA) undertaken by consultant surgeons nationally, and the rate of accrual of RHA and corresponding primary hip arthroplasty (PHA) volume for new consultants entering practice. Methods.
Aims. Our main aim was to describe the trend in the comorbidities of patients undergoing elective total hip arthroplasties (THAs) and knee arthroplasties (KAs) between 1 January 2005 and 31 December 2018 in England. Methods. We combined data from the
Aims. The number of revision total ankle arthroplasties (TAAs) which are undertaken is increasing. Few studies have reported the survival after this procedure. The primary aim of this study was to analyze the survival of revision ankle arthroplasties using large datasets. Secondary aims were to summarize the demographics of the patients, the indications for revision TAA, further operations, and predictors of survival. Methods. The study combined data from the
Aims. The aim of this study was to identify variables associated with time to revision, demographic details associated with revision indication, and type of prosthesis employed, and to describe the survival of hinge knee arthroplasty (HKA) when used for first-time knee revision surgery and factors that were associated with re-revision. Methods. Patient demographic details, BMI, American Society of Anesthesiologists (ASA) grade, indication for revision, surgical approach, surgeon grade, implant type (fixed and rotating), time of revision from primary implantation, and re-revision if undertaken were obtained from the
Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods. Data from the
Aims.
Aims. The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit. Methods. A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the
Aims. Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA. Methods. We performed an observational study using
Aims. The aim of this study was to capture 12-month outcomes from a representative multicentre cohort of patients undergoing total ankle arthroplasty (TAA), describe the pattern of patient-reported outcome measures (PROMs) at 12 months, and identify predictors of these outcome measures. Methods. Patients listed for a primary TAA at 19 NHS hospitals between February 2016 and October 2017 were eligible. PROMs data were collected preoperatively and at six and 12 months including: Manchester-Oxford Foot and Ankle Questionnaire (MOXFQ (foot and ankle)) and the EuroQol five-dimension five-level questionnaire (EQ-5D-5L). Radiological pre- and postoperative data included Kellgren-Lawrence score and implant position measurement. This was supplemented by data from the
Aims. The main objective of this study was to examine whether the Oxford Shoulder Score (OSS) demonstrated floor or ceiling effects when used to measure outcomes following shoulder arthroplasty in a large national cohort. Secondary objectives were to assess its pain and function subscales, and to identify independent predictors for patients achieving a postoperative ceiling score following shoulder arthroplasty. Methods. Secondary database analysis of the
Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the
Aims. Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. Methods. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK
Aims. To determine if primary cemented acetabular component geometry (long posterior wall (LPW), hooded, or offset reorientating) influences the risk of revision total hip arthroplasty (THA) for instability or loosening. Methods. The
Aims. The primary aim is to estimate the current and potential number of patients on NHS England orthopaedic elective waiting lists by November 2020. The secondary aims are to model recovery strategies; review the deficit of hip and knee arthroplasty from
Aims. The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). Methods. We undertook a registry study combining the
Aims. To determine the trajectories of patient reported pain and functional disability over five years following total hip arthroplasty (THA) or total knee arthroplasty (TKA). Methods. A prospective, longitudinal cohort sub-study within the
Aims. “Get It Right First Time” (GIRFT) and NHS England’s Best Practice Tariff (BPT) have published directives advising that patients over the ages of 65 (GIRFT) and 69 years (BPT) receiving total hip arthroplasty (THA) should receive cemented implants and have brought in financial penalties if this policy is not observed. Despite this, worldwide, uncemented component use has increased, a situation described as a ‘paradox’. GIRFT and BPT do, however, acknowledge more data are required to support this edict with current policies based on the
Aims. Debate remains whether the patella should be resurfaced during total knee replacement (TKR). For non-resurfaced TKRs, we estimated what the revision rate would have been if the patella had been resurfaced, and examined the risk of re-revision following secondary patellar resurfacing. Methods. A retrospective observational study of the
Aims. We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre. Methods. Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year. Results. Eight patients (ten ankles) died during follow-up, but none required revision. Of the surviving 106 patients (108 ankles: rheumatoid arthritis (RA), n = 15; osteoarthritis (OA), n = 93), 38 were women and 68 were men, with a mean age of 68.2 years (48 to 86) at the time of surgery. Mean follow-up was 5.1 years (2.1 to 9.0). A total of ten implants failed (8.5%), thus requiring revision. The implant survival at seven years, using revision as an endpoint, was 88.2% (95% CI 100% to 72.9%). There was a mean improvement in Manchester-Oxford Foot and Ankle Questionnaire (MOXFQ) from 85.0 to 32.8 and visual analogue scale (VAS) scores from 7.0 to 3.2, and overall satisfaction was 89%. The three commonest complications were malleolar fracture (14.4%, n = 17), wound healing (13.6%, n = 16), and superficial infection (12.7%, n = 15). The commonest reason for revision was aseptic loosening. No patients in our study were revised for deep infection. Conclusion. Our results show that Zenith survival rates are comparable with those in the literature for other implants and in the
Aims. To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). Methods. In this registry study, data from the
Aims. The rate of dislocation when traditional single bearing implants are used in revision total hip arthroplasty (THA) has been reported to be between 8% and 10%. The use of dual mobility bearings can reduce this risk to between 0.5% and 2%. Dual mobility bearings are more expensive, and it is not clear if the additional clinical benefits constitute value for money for the payers. We aimed to estimate the cost-effectiveness of dual mobility compared with single bearings for patients undergoing revision THA. Methods. We developed a Markov model to estimate the expected cost and benefits of dual mobility compared with single bearing implants in patients undergoing revision THA. The rates of revision and further revision were calculated from the
Aims. To determine mortality risk after first revision total hip arthroplasty (THA) for periprosthetic femoral fracture (PFF), and to compare this to mortality risk after primary and first revision THA for other common indications. Methods. The study cohort consisted of THAs recorded in the
Aims. The aim of this study was to investigate whether the use of antibiotic-loaded bone cement influenced the risk of revision surgery after primary total hip arthroplasty (THA) for osteoarthritis. Methods. The study involved data collected by the
Aims. We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital. Methods. All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with
Aims. The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model. Materials and Methods. A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the
Aims. We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. Patients and Methods. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a
Aims. The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. Patients and Methods. A descriptive observational study was undertaken using data from the
Aims. The aim of this study was to determine whether patients with
metal-on-metal (MoM) arthroplasties of the hip have an increased
risk of cardiac failure compared with those with alternative types
of arthroplasties (non-MoM). Patients and Methods. A linkage study between the
Aim. There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several
Aims. Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients. Materials and Methods. Data from the
Objectives. We used the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man (NJR) to investigate the risk of revision due to prosthetic joint infection (PJI) for patients undergoing primary and revision hip arthroplasty, the changes in risk over time, and the overall burden created by PJI. Methods. We analysed revision total hip arthroplasties (THAs) performed due to a diagnosis of PJI and the linked index procedures recorded in the NJR between 2003 and 2014. The cohort analysed consisted of 623 253 index primary hip arthroplasties, 63 222 index revision hip arthroplasties and 7585 revision THAs performed due to a diagnosis of PJI. The prevalence, cumulative incidence functions and the burden of PJI (total procedures) were calculated. Overall linear trends were investigated with log-linear regression. Results. We demonstrated a prevalence of revision THA due to prosthetic joint infection of 0.4/100 procedures following primary and 1.6/100 procedures following revision hip arthroplasty. The prevalence of revision due to PJI in the three months following primary hip arthroplasty has risen 2.3-fold (95% confidence interval (CI) 1.3 to 4.1) between 2005 and 2013, and 3.0-fold (95% CI 1.1 to 8.5) following revision hip arthroplasty. Over 1000 procedures are performed annually as a consequence of hip PJI, an increase of 2.6-fold between 2005 and 2013. Conclusions. Although the risk of revision due to PJI following hip arthroplasty is low, it is rising and, coupled with the established and further predicted increased incidence of both primary and revision hip arthroplasty, this represents a growing and substantial treatment burden. Cite this article: E. Lenguerrand, M. R. Whitehouse, A. D. Beswick, S. A. Jones, M. L. Porter, A. W. Blom. Revision for prosthetic joint infection following hip arthroplasty: Evidence from the
Aims. The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Methods. Prospective cohort study using linked
Aims. We performed a systematic review of the current literature regarding
the outcomes of unconstrained metacarpophalangeal joint (MCPJ) arthroplasty. Materials and Methods. We initially identified 1305 studies, and 406 were found to be
duplicates. After exclusion criteria were applied, seven studies
were included. Outcomes extracted included pre- and post-operative
pain visual analogue scores, range of movement (ROM), strength of
pinch and grip, satisfaction and patient reported outcome measures
(PROMs). Clinical and radiological complications were recorded.
The results are presented in three groups based on the design of
the arthroplasty and the aetiology (pyrocarbon-osteoarthritis (pyro-OA),
pyrocarbon-inflammatory arthritis (pyro-IA), metal-on-polyethylene
(MoP)). Results. Results show that pyrocarbon implants provide an 85% reduction
in pain, 144% increase of pinch grip and 13° improvements in ROM
for both OA and IA combined. Patients receiving MoP arthroplasties
had a reduction in pinch strength. Satisfaction rates were 91% and
92% for pyrocarbon-OA and pyrocarbon-IA groups, respectively. There were
nine failures in 87 joints (10.3%) over a mean follow-up of 5.5
years (1.0 to 14.3) for pyro-OA. There were 18 failures in 149 joints
(12.1%) over a mean period of 6.6 years (1.0 to 16.0) for pyro-IA.
Meta-analysis was not possible due to the heterogeneity of the studies
and the limited presentation of data. Conclusion. We would recommend prospective data collection for small joint
arthroplasties of the hand consisting of PROMs that would allow
clinicians to come to stronger conclusions about the impact on function
of replacing the MCPJs. A
The National Institute for Clinical Excellence (NICE) produces recommendations on appropriate treatment within the National Health Service (NHS) in England and Wales. The NICE guidelines on prophylaxis for venous thromboembolism in orthopaedic surgery recommend that all patients be offered a low molecular weight heparin (LMWH). The linked hospital episode statistics of 219 602 patients were examined to determine the rates of complications following lower limb arthroplasty for the 12-month periods prior to and following the publication of these guidelines. These were compared with data from the
The effective capture of outcome measures in
the healthcare setting can be traced back to Florence Nightingale’s
investigation of the in-patient mortality of soldiers wounded in
the Crimean war in the 1850s. Only relatively recently has the formalised collection of outcomes
data into Registries been recognised as valuable in itself. With the advent of surgeon league tables and a move towards value
based health care, individuals are being driven to collect, store
and interpret data. Following the success of the
Aims. Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding ceramic fracture.
This study aims to report the risk of revision for fracture of modern
CoC bearings and identify factors that might influence this risk,
using data from the
Unicompartmental knee arthroplasty (UKA) has
advantages over total knee arthroplasty but
Recent guidance recommends the use of a well-proven
cemented femoral stem for hemiarthroplasty in the management of
fractures of the femoral neck, and the Exeter Trauma Stem (ETS)
has been suggested as an example of such an implant. The design
of this stem was based on the well-proven Exeter Total Hip Replacement
stem (ETHRS). This study assessed the surface finish of the ETS
in comparison with the ETHRS. Two ETSs and two ETHRSs were examined
using a profilometer with a precision of 1 nm and compared with
an explanted Exeter Matt stem. The mean roughness average (RA) of
the ETSs was approximately ten times higher than that of the ETHRSs (0.235 μm
(0.095 to 0.452) versus 0.025 μm (0.011 to 0.059);
p <
0.001). The historical Exeter Matt stem roughness measured
a mean RA of 0.973 μm (0.658 to 1.159). The change of the polished
Exeter stem to a matt surface finish in 1976 resulted in a high
stem failure rate. We do not yet know whether the surface differences
between ETS and ETHRS will be clinically significant. We propose
the inclusion of hemiarthroplasty stems in
Aims . The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. Patients and Methods. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison. Results. The mean serum cobalt (Co) and chromium (Cr) ion levels remained
above baseline in both groups (CoM: Co 1.16 μg/l (0.41 to 14.67),
Cr 1.05 μg/l (0.16 to 12.58); MoM: Co 2.93 μg/l (0.35 to 30.29),
Cr 1.85 μg/l (0.36 to 17.00)) but the increase was significantly
less in the CoM cohort (Co difference p = 0.001, Cr difference p
= 0.002). These medium-term results, coupled with lower revision
rates from
Objectives. High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The
There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent
Whether to use total or unicompartmental knee
replacement (TKA/UKA) for end-stage knee osteoarthritis remains controversial.
Although UKA results in a faster recovery, lower rates of morbidity
and mortality and fewer complications, the long-term revision rate
is substantially higher than that for TKA. The effect of each intervention on
patient-reported outcome remains unclear. The aim of this study
was to determine whether six-month patient-reported outcome measures
(PROMs) are better in patients after TKA or UKA, using data from
a large
Objective. Mortality rates reported by the
New brands of joint prosthesis are released for general implantation with limited evidence of their long-term performance in patients. The CUSUM continuous monitoring method is a statistical testing procedure which could be used to provide prospective evaluation of brands as soon as implantation in patients begins and give early warning of poor performance. We describe the CUSUM and illustrate the potential value of this monitoring tool by applying it retrospectively to the 3M Capital Hip experience. The results show that if the clinical data and methodology had been available, the CUSUM would have given an alert to the underperformance of this prosthesis almost four years before the issue of a Hazard Notice by the Medical Devices Agency. This indicates that the CUSUM can be a valuable tool in monitoring joint prostheses, subject to timely and complete collection of data. Regional or
Aims. The aim of this study was to present data on 11 459 patients
who underwent total hip (THA), total knee (TKA) or unicompartmental
knee arthroplasty (UKA) between November 2002 and April 2014 with
aspirin as the primary agent for pharmacological thromboprophylaxis. . Patients and Methods. We analysed the incidence of deep vein thrombosis (DVT) and pulmonary
embolism (PE) then compared the 90-day all-cause mortality with
the corresponding data in the
Total knee arthroplasty (TKA) and total hip arthroplasty
(THA) are recognised and proven interventions for patients with
advanced arthritis. Studies to date have demonstrated a steady increase
in the requirement for primary and revision procedures. Projected
estimates made for the United States show that by 2030 the demand
for primary TKA will grow by 673% and for revision TKA by 601% from
the level in 2005. For THA the projected estimates are 174% and
137% for primary and revision surgery, respectively. The purpose
of this study was to see if those predictions were similar for England
and Wales using data from the
United Kingdom National Institute for Health
and Clinical Excellence guidelines recommend the use of total hip replacement
(THR) for displaced intracapsular fractures of the femoral neck
in cognitively intact patients, who were independently mobile prior
to the injury. This study aimed to analyse the risk factors associated
with revision of the implant and mortality following THR, and to
quantify risk.
The survivorship of contemporary resurfacing arthroplasty of the hip using metal-on-metal bearings is better than that of first generation designs, but short-term failures still occur. The most common reasons for failure are fracture of the femoral neck, loosening of the component, osteonecrosis of the femoral head, reaction to metal debris and malpositioning of the component. In 2008 the Australian
Patient-reported outcome measures (PROMs) are
increasingly being used to assess functional outcome and patient satisfaction.
They provide a framework for comparisons between surgical units,
and individual surgeons for benchmarking and financial remuneration.
Better performance may bring the reward of more customers as patients and
commissioners seek out high performers for their elective procedures.
Using
We assessed the outcome of patients who were
lost to follow-up after arthroplasty by a single surgeon. The aim was
to validate the surgeon’s data set with the Australian Orthopaedic
Association National Joint Replacement Registry and determine the
outcome of those patients lost to follow-up. Prospective data on patient demographics, operative details and
outcomes of the surgeon’s 1192 primary unicompartmental knee arthroplasty
(UKA) procedures were analysed. There were 69 knees in patients
who were lost to follow-up, among whom the Registry identified 31
deaths and eight revisions. The cumulative percentage revision (CPR) at seven years using
the additional Registry data was 8.8% (95% confidence interval (CI)
7 to 11). Using the surgeon’s data, the CPR at seven years was 8%
(95% CI 6.3 to 10.1) for the best-case scenario where loss to follow-up
was excluded, and 16% (95% CI 13.8 to 19.4) for the worst-case scenario, where
all patients lost to follow-up were deemed to have been revised.
There was a significantly higher mortality rate in those patients
lost to follow-up. This study demonstrates that a
The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery. A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series.Aims
Methods
Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening. A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.Aims
Methods
The October 2024 Arthroplasty Roundup360 looks at: Breaking the mould: female representation in arthroplasty surgery remains low, with elbow leading the way; Post COVID-19: where are we with the 'catch up' in England and Wales?; Prevalence and clinical impact of sarcopenia in patients undergoing total joint replacement: a systematic review and a meta-analysis; Total joint replacement and sleep: the state of the evidence.
The February 2023 Foot & Ankle Roundup360 looks at: Joint inflammatory response in ankle and pilon fractures; Tibiotalocalcaneal fusion with a custom cage; Topical application of tranexamic acid can reduce blood loss in calcaneal fractures; Risk factors for failure of total ankle arthroplasty; Pain catastrophizing: the same as pain forecasting?.
The February 2024 Foot & Ankle Roundup360 looks at: Survival of revision ankle arthroplasty; Tibiotalocalcaneal nail for the management of open ankle fractures in the elderly patient; Accuracy of a patient-specific total ankle arthroplasty instrumentation; Fusion after failed primary ankle arthroplasty: can it work?; Treatment options for osteochondral lesions of the talus; Managing hair tourniquet syndrome of toe: a rare emergency; Ultrasound-guided collagenase therapy for recurrent plantar fibromatosis: a promising line of therapy?.
The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component. We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.Aims
Methods
In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component. Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).Aims
Methods
The aims of this study were to identify and evaluate the current literature examining the prognostic factors which are associated with failure of total elbow arthroplasty (TEA). Electronic literature searches were conducted using MEDLINE, Embase, PubMed, and Cochrane. All studies reporting prognostic estimates for factors associated with the revision of a primary TEA were included. The risk of bias was assessed using the Quality In Prognosis Studies (QUIPS) tool, and the quality of evidence was assessed using the modified Grading of Recommendations, Assessment, Development, and Evaluations (GRADE) framework. Due to low quality of the evidence and the heterogeneous nature of the studies, a narrative synthesis was used.Aims
Methods
Cite this article:
Cite this article:
The December 2022 Knee Roundup360 looks at: Effect of physical therapy versus arthroscopic partial meniscectomy: the ESCAPE trial at five years; Patellofemoral arthroplasty or total knee arthroplasty: a randomized controlled trial; Rehabilitation versus surgical reconstruction for anterior cruciate ligament injury; End-stage knee osteoarthritis in Australia: the effect of obesity; Do poor patient-reported outcome measures at six months relate to knee revision?; What is the cost of nonoperative interventions for knee osteoarthritis?
The April 2024 Knee Roundup360 looks at: Challenging the status quo: re-evaluating the impact of obesity on unicompartmental knee arthroplasty outcomes; Timing matters: the link between ACL reconstruction delays and cartilage damage; Custom fit or off the shelf: evaluating patient outcomes in tailored versus standard knee replacements; Revolutionizing knee replacement: a comparative study on robotic-assisted and computer-navigated techniques; Pre-existing knee osteoarthritis and severe joint depression are associated with the need for total knee arthroplasty after tibial plateau fracture in patients aged over 60 years; Modern digital therapies?; A matched study on fracture rates following knee replacement surgeries;
The October 2024 Foot & Ankle Roundup360 looks at: Hemiarthroplasty for hallux rigidus; Fixed or mobile-bearing ankle arthroplasty? A meta-analysis; Bone grafting for periprosthetic bone cysts following total ankle arthroplasty; Diabetic foot ulcer after first-ray amputation; Early motion after ankle surgery: the path to faster recovery?; Are podiatrists and orthopaedic surgeons approaching zone 2 fifth metatarsal fractures in the same way?.
The June 2023 Hip & Pelvis Roundup360 looks at: Machine learning to identify surgical candidates for hip and knee arthroplasty: a viable option?; Poor outcome after debridement and implant retention; Can you cement polyethylene liners into well-fixed acetabular shells in hip revision?; Revision stem in primary arthroplasties: the Exeter 44/0 125 mm stem; Depression and anxiety: could they be linked to infection?; Does where you live affect your outcomes after hip and knee arthroplasties?; Racial disparities in outcomes after total hip arthroplasty and total knee arthroplasty are substantially mediated by socioeconomic disadvantage both in black and white patients.
Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF. Cite this article:
The April 2023 Knee Roundup360 looks at: Does bariatric surgery reduce complications after total knee arthroplasty?; Mid-flexion stability in total knee arthroplasties implanted with kinematic alignment: posterior-stabilized versus medial-stabilized implants; Inflammatory response in robotic-arm-assisted versus conventional jig-based total knee arthroplasty; Journey II bicruciate stabilized (JII-BCS) and GENESIS II total knee arthroplasty: the CAPAbility, blinded, randomized controlled trial; Lifetime risk of revision and patient factors; Platelet-rich plasma use for hip and knee osteoarthritis in the USA; Where have the knee revisions gone?; Tibial component rotation in total knee arthroplasty: CT-based study of 1,351 tibiae.
The October 2023 Foot & Ankle Roundup360 looks at: Risk factors for failure of total ankle arthroplasties; Effects of synovial fluid fracture haematoma to tissue-engineered cartilage; Coronal plane deformity in CMT-cavovarus feet using automated 3D measurements; Immediate weightbearing after ankle fracture fixation – is it safe?; Unlocking the mystery of Mueller-Weiss disease; Diabetic foot management: predictors of failure.