The October 2015 Children’s orthopaedics Roundup360 looks at: Radiographic follow-up of DDH; When the supracondylar goes wrong; Apophyseal avulsion fractures; The ‘pulled elbow’; Surgical treatment of active or aggressive aneurysmal bone cysts in children; Improving stability in supracondylar fractures; Biological reconstruction may be preferable in children’s osteosarcoma; The paediatric hip fracture
Retrieval studies of total hip replacements with highly cross-linked ultra-high-molecular-weight polyethylene liners have shown much less surface damage than with conventional ultra-high-molecular-weight polyethylene liners. A recent revision hip replacement for recurrent dislocation undertaken after only five months revealed a highly cross-linked polyethylene liner with a large area of visible delamination. In order to determine the cause of this unusual surface damage, we analysed the bearing surfaces of the cobalt-chromium femoral head and the acetabular liner with scanning electron microscopy, energy dispersive x-ray spectroscopy and optical profilometry. We concluded that the cobalt-chromium modular femoral head had scraped against the titanium acetabular shell during the course of the dislocations and had not only roughened the surface of the femoral head but also transferred deposits of titanium onto it. The largest deposits were 1.6 μm to 4.3 μm proud of the surrounding surface and could lead to increased stresses in the acetabular liner and therefore cause accelerated wear and damage. This case illustrates that dislocations can leave titanium deposits on cobalt-chromium femoral heads and that highly cross-linked ultra-high-molecular-weight polyethylene remains susceptible to surface damage.
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
Fracture of a ceramic component in total hip
replacement is a rare but potentially catastrophic complication.
The incidence is likely to increase as the use of ceramics becomes
more widespread. We describe such a case, which illustrates how
inadequate initial management will lead to further morbidity and
require additional surgery. We present the case as a warning that
fracture of a ceramic component should be revised to another ceramic-on-ceramic
articulation in order to minimise the risk of further catastrophic
wear.
We report our early experience with the use of
a new prosthesis, the Modular Hemipelvic Prosthesis II, for reconstruction
of the hemipelvis after resection of a primary malignant peri-acetabular
tumour involving the sacroiliac joint. We retrospectively reviewed the outcome of 17 patients who had
undergone resection of a pelvic tumour and reconstruction with this
prosthesis between July 2002 and July 2010. One patient had a type I+II+III+IV resection (ilium + peri-acetabulum
+ pubis/ischium + sacrum) and 16 had a type I+II+IV resection (ilium
+ acetabulum + sacrum). The outcome was assessed at a mean follow-up
of 33 months (15 to 59). One patient was alive with disease, 11
were alive without disease and five had died of disease. The overall
five-year survival rate was 62.4%. Six patients had a local recurrence.
The mean Musculoskeletal Tumour Society score was 58% (33 to 77).
Deep infection occurred in two patients, problems with wound healing
in five and dislocation in one. For patients with a primary malignant peri-acetabular sarcoma
involving the sacroiliac joint, we believe that this new prosthesis
is a viable option for reconstruction of the bony defect left following
resection of the tumour. It results in a satisfactory functional
outcome with an acceptable rate of complications. Cite this article:
Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
The number of arthroplasties being undertaken
is expected to grow year on year, and periprosthetic joint infections will
be an increasing socioeconomic burden. The challenge to prevent
and eradicate these infections has resulted in the emergence of
several new strategies, which are discussed in this review. Cite this article:
The widespread use of MRI has revolutionised
the diagnostic process for spinal disorders. A typical protocol
for spinal MRI includes T1 and T2 weighted sequences in both axial
and sagittal planes. While such an imaging protocol is appropriate
to detect pathological processes in the vast majority of patients,
a number of additional sequences and advanced techniques are emerging.
The purpose of the article is to discuss both established techniques
that are gaining popularity in the field of spinal imaging and to
introduce some of the more novel ‘advanced’ MRI sequences with examples
to highlight their potential uses. Cite this article:
This randomised trial evaluated the outcome of
a single design of unicompartmental arthroplasty of the knee (UKA) with
either a cemented all-polyethylene or a metal-backed modular tibial
component. A total of 63 knees in 45 patients (17 male, 28 female)
were included, 27 in the all-polyethylene group and 36 in the metal-backed
group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up
of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were
revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and
two metal-backed components were revised (at one and five years).
One revision in both groups was for unexplained pain, one in the
metal-backed group was for progression of osteoarthritis. The others
in the all-polyethylene group were for aseptic loosening. The survivorship
at seven years calculated by the Kaplan–Meier method for the all-polyethylene
group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for
the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at
risk 16) This difference was statistically significant (p <
0.001).
At the most recent follow-up, significantly better mean Western
Ontario and McMaster Universities Arthritis Index Scores were found
in the all-polyethylene group (13.4 This randomised study demonstrates that all-polyethylene components
in this design of fixed bearing UKA had unsatisfactory results with
significantly higher rates of failure before ten years compared
with the metal-back components. Cite this article:
Revision knee arthroplasty presents a number
of challenges, not least of which is obtaining solid primary fixation
of implants into host bone. Three anatomical zones exist within
both femur and tibia which can be used to support revision implants.
These consist of the joint surface or epiphysis, the metaphysis
and the diaphysis. The methods by which fixation in each zone can
be obtained are discussed. The authors suggest that solid fixation
should be obtained in at least two of the three zones and emphasise
the importance of pre-operative planning and implant selection. Cite this article:
Clinical studies of patients with bone sarcomas have been challenged
by insufficient numbers at individual centres to draw valid conclusions.
Our objective was to assess the feasibility of conducting a definitive
multi-centre randomised controlled trial (RCT) to determine whether
a five-day regimen of post-operative antibiotics, in comparison
to a
24-hour regimen, decreases surgical site infections in patients
undergoing endoprosthetic reconstruction for lower extremity primary
bone tumours. We performed a pilot international multi-centre RCT. We used
central randomisation to conceal treatment allocation and sham antibiotics
to blind participants, surgeons, and data collectors. We determined
feasibility by measuring patient enrolment, completeness of follow-up,
and protocol deviations for the antibiotic regimens. Objective
Methods
We hypothesised that a minimally invasive peroneus
brevis tendon transfer would be effective for the management of
a chronic rupture of the Achilles tendon. In 17 patients (three
women, 14 men) who underwent minimally invasive transfer and tenodesis
of the peroneus brevis to the calcaneum, at a mean follow-up of
4.6 years (2 to 7) the modified Achilles tendon total rupture score
(ATRS) was recorded and the maximum circumference of the calf of
the operated and contralateral limbs was measured. The strength
of isometric plantar flexion of the gastrocsoleus complex and of
eversion of the ankle were measured bilaterally. Functional outcomes
were classified according to the four-point Boyden scale. At the latest review, the mean maximum circumference of the calf
of the operated limb was not significantly different from the pre-operative
mean value, (41.4 cm, 32 to 50 This procedure may be safely performed, is minimally invasive,
and allows most patients to return to pre-injury sport and daily
activities. Cite this article:
Lengthening osteotomies of the calcaneus in children are in general
grafted with bone from the iliac crest. Artificial bone grafts have
been introduced, however, their structural and clinical durability
has not been documented. Radiostereometric analysis (RSA) is a very
accurate and precise method for measurements of rigid body movements including
the evaluation of joint implant and fracture stability, however,
RSA has not previously been used in clinical studies of calcaneal
osteotomies. We assessed the precision of RSA as a measurement tool
in a lateral calcaneal lengthening osteotomy (LCLO). LCLO was performed in six fixed adult cadaver feet. Tantalum
markers were inserted on each side of the osteotomy and in the cuboideum.
Lengthening was done with a plexiglas wedge. A total of 24 radiological
double examinations were obtained. Two feet were excluded due to
loose and poorly dispersed markers. Precision was assessed as systematic
bias and 95% repeatability limits.Objectives
Methods
A moderator and panel of five experts led an
interactive session in discussing five challenging and interesting patient
case presentations involving surgery of the hip. The hip pathologies
reviewed included failed open reduction internal fixation of subcapital
femoral neck fracture, bilateral hip disease, evaluation of pain
after metal-on-metal hip arthroplasty, avascular necrosis, aseptic
loosening secondary to osteolysis and polyethylene wear, and management
of ceramic femoral head fracture.
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
Drug therapy forms an integral part of the management
of many orthopaedic conditions. However, many medicines can produce
serious adverse reactions if prescribed inappropriately, either
alone or in combination with other drugs. Often these hazards are
not appreciated. In response to this, the European Union recently
issued legislation regarding safety measures which member states
must adopt to minimise the risk of errors of medication. In March 2014 the Medicines and Healthcare products Regulatory
Agency and NHS England released a Patient Safety Alert initiative
focussed on errors of medication. There have been similar initiatives
in the United States under the auspices of The National Coordinating
Council for Medication Error and The Joint Commission on the Accreditation
of Healthcare Organizations. These initiatives have highlighted
the importance of informing and educating clinicians. Here, we discuss common drug interactions and contra-indications
in orthopaedic practice. This is germane to safe and effective clinical
care. Cite this article:
Large femoral heads have become popular in total
hip replacement (THR) as a method of reducing the risk of dislocation.
However, if large heads are used in ceramic-on-ceramic THR, the
liner must be thinner, which may increase the risk of fracture.
To compare the rates of ceramic fracture and dislocation between
28 mm and 32 mm ceramic heads, 120 hips in 109 patients (51 men
and 58 women, mean age 49.2 years) were randomised to THR with either
a 28 mm or a 32 mm ceramic articulation. A total of 57/60 hips assigned
to the 28 mm group and 55/60 hips assigned to the 32 mm group were
followed for at least five years. No ceramic component fractures
occured in any patient in either group. There was one dislocation
in the 32 mm group and none in the 28 mm group (p = 0.464). No hip
had detectable wear, focal osteolysis or prosthetic loosening. In
our small study the 32 mm ceramic articulation appeared to be safe
in terms of ceramic liner fracture. Cite this article:
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p <
0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with <
1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.