Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
The purpose of this study was to clarify the appearance of the reparative tissue on the articular surface and to analyse the properties of the reparative tissue after hemicallotasis osteotomy (HCO) using MRI T1ρ and T2 mapping. Coronal T1ρ and T2 mapping and three-dimensional gradient-echo images were obtained from 20 subjects with medial knee osteoarthritis. We set the regions of interest (ROIs) on the full-thickness cartilage of the medial femoral condyle (MFC) and medial tibial plateau (MTP) of the knee and measured the cartilage thickness (mm) and T1ρ and T2 relaxation times (ms). Statistical analysis of time-dependent changes in the cartilage thickness and the T1ρ and T2 relaxation times was performed using one-way analysis of variance, and Scheffe’s test was employed for Objectives
Methods
This study aimed to characterise and qualitatively grade the severity of the corrosion particles released into the hip joint following taper corrosion. The 26 cases examined were CoC/ABG Modular (n = 13) and ASR/SROM (n = 13). Blood serum metal ion levels were collected before and after revision surgery. The haematoxylin and eosin tissue sections were graded on the presence of fibrin exudates, necrosis, inflammatory cells and corrosion products. The corrosion products were identified based on visible observation and graded on abundance. Two independent observers blinded to the clinical patient findings scored all cases. Elemental analysis was performed on corrosion products within tissue sections. X-Ray diffraction was used to identify crystalline structures present in taper debris.Objectives
Methods
We have reviewed 70 patients with bilateral simultaneous total hip arthroplasties to determine the rate of failure and to compare polyethylene wear and osteolysis between an implant with a cobalt-chrome head and Hylamer liner with that of a zirconia head and Hylamer liner. The mean thickness of the polyethylene liner was 11.0 mm (8.8 to 12.2) in the hip with a zirconia head and 10.7 mm (8.8 to 12.2) in that with a cobalt-chrome head. At follow-up at 6.4 years no acetabular or femoral component had been revised for aseptic loosening and no acetabular or femoral component was loose according to radiological criteria in both the cemented and cementless groups. The mean rate of linear wear and annual wear rate were highest in the 22 mm zirconia femoral head (1.25 mm (SD 1.05) and 0.21 mm (SD 0.18), respectively) and lowest in the 22 mm cobalt-chrome femoral head (0.70 mm (SD 0.39) and 0.12 mm (SD 0.07), respectively). The mean volumetric wear was highest in the 28 mm zirconia femoral head (730.79 mm. 3. ) and lowest in the 22 mm cobalt-chrome femoral head (264.67 mm. 3. ), but if the results were compared by size of the femoral head and type of material there was no statistical difference (p >
0.05). Sequential measurements of annual wear showed that the zirconia femoral head had a relatively higher rate of penetration than the cobalt-chrome head over the first three years; thereafter the rate of wear was reduced and compared favourably with that of cobalt-chrome heads. There was a statistically significant relationship between the wear of the polyethylene liner and the age of the patient, male gender and the degree of abduction
Nerve palsy is a well-described complication
following total hip arthroplasty, but is highly distressing and
disabling. A nerve palsy may cause difficulty with the post-operative
rehabilitation, and overall mobility of the patient. Nerve palsy
may result from compression and tension to the affected nerve(s)
during the course of the operation via surgical manipulation and
retractor placement, tension from limb lengthening or compression
from post-operative hematoma. In the literature, hip dysplasia,
lengthening of the leg, the use of an uncemented femoral component, and
female gender are associated with a greater risk of nerve palsy.
We examined our experience at a high-volume, tertiary care referral
centre, and found an overall incidence of 0.3% out of 39 056 primary
hip arthroplasties. Risk factors found to be associated with the
incidence of nerve palsy at our institution included the presence
of spinal stenosis or lumbar disc disease, age younger than 50,
and smoking. If a nerve palsy is diagnosed, imaging is mandatory
and surgical evacuation or compressive haematomas may be beneficial.
As palsies are slow to recover, supportive care such as bracing,
therapy, and reassurance are the mainstays of treatment. Cite this article:
An important measure for the diagnosis and monitoring of knee osteoarthritis is the minimum joint space width (mJSW). This requires accurate alignment of the x-ray beam with the tibial plateau, which may not be accomplished in practice. We investigate the feasibility of a new mJSW measurement method from stereo radiographs using 3D statistical shape models (SSM) and evaluate its sensitivity to changes in the mJSW and its robustness to variations in patient positioning and bone geometry. A validation study was performed using five cadaver specimens. The actual mJSW was varied and images were acquired with variation in the cadaver positioning. For comparison purposes, the mJSW was also assessed from plain radiographs. To study the influence of SSM model accuracy, the 3D mJSW measurement was repeated with models from the actual bones, obtained from CT scans.Objectives
Materials and Methods
The aims of our study were to provide long-term information on
the behaviour of the thoracolumbar/lumbar (TL/L) curve after thoracic
anterior correction and fusion (ASF) and to determine the impact
of ASF on pulmonary function. A total of 41 patients (four males, 37 females) with main thoracic
(MT) adolescent idiopathic scoliosis (AIS) treated with ASF were
included. Mean age at surgery was 15.2 years (11 to 27). Mean follow-up
period was 13.5 years (10 to 18).Aims
Patients and Methods
1. The results of repair of the sciatic nerve and of its main divisions have been analysed in a series of 118 cases, the patients having been under observation for three to eighteen years (average 11·7 years). 2. A result was satisfactory if there was some return of sensibility throughout the autonomous zone (the area of skin supplied exclusively by the damaged nerve) and if the more important muscles of the leg were capable of contraction against gravity and resistance. 3. When the whole of the sciatic nerve is damaged it is necessary to present the results separately for the lateral and medial popliteal divisions. 4. Of forty-seven cases of repair of the medial popliteal nerve 79 per cent showed useful motor and 62 per cent useful sensory recovery. In three out of four cases the correspondence between the degree of motor and of sensory recovery was fairly close. 5. Of seventy-two cases of repair of the lateral popliteal nerve 36 per cent showed useful motor and 74 per cent useful sensory recovery. The latter figure must be regarded with some reserve because sensory "recovery" in the lateral popliteal zone may be due to the ingrowth of nerve fibres from contiguous normally innervated skin. Thus it is not possible to correlate motor and sensory recovery. 6. In eighteen cases of repair of the posterior tibial nerve, there was useful sensory recovery in the sole in twelve. But although there was evidence of recovery in the plantar muscles in eleven cases it was functionally valueless. 7. In repair of the medial popliteal nerve the result was better if suture had been carried out early. In repair of the lateral popliteal nerve there was no evidence that delay was harmful; but the proportion of good results was so low (as judged by motor function alone, sensory recovery being often extraneous) that this exception to a general rule cannot be taken very seriously. 8. Gaps of up to twelve centimetres–estimated after resection of the damaged nerve ends–could be closed without difficulty by the usual technique, and the extent of the gap up to that limit had no influence on the prognosis. The closure of larger gaps, when the knee must be flexed beyond a right
We assessed the long-term (more than ten-year) outcomes of the
Kudo type-5 elbow prosthesis in patients with rheumatoid arthritis
(RA). We reviewed 41 elbows (Larsen Grade IV, n = 21; Grade V, n =
20) in 31 patients with RA who had undergone a Kudo type-5 total
elbow arthroplasty (TEA) between 1994 and 2003, and had been followed
up for more than ten years. The humeral component was cementless
and the all-polyethylene ulnar component cemented in every patient. Clinical
outcome was assessed using the Mayo elbow performance score. We
calculated the revision rate and evaluated potential risk factors
for revision. The duration of follow-up was a mean 141 months (120
to 203).Aims
Materials and Methods
In arthritis of the varus knee, a high tibial
osteotomy (HTO) redistributes load from the diseased medial compartment
to the unaffected lateral compartment. We report the outcome of 36 patients (33 men and three women)
with 42 varus, arthritic knees who underwent HTO and dynamic correction
using a Garches external fixator until they felt that normal alignment
had been restored. The mean age of the patients was 54.11 years
(34 to 68). Normal alignment was achieved at a mean 5.5 weeks (3
to 10) post-operatively. Radiographs, gait analysis and visual analogue
scores for pain were measured pre- and post-operatively, at one
year and at medium-term follow-up (mean six years; 2 to 10). Failure
was defined as conversion to knee arthroplasty. Pre-operative gait analysis divided the 42 knees into two equal
groups with high (17 patients) or low (19 patients) adductor moments.
After correction, a statistically significant (p <
0.001, At final follow-up, after a mean of 15.9 years (12 to 20), there
was a survivorship of 59% (95% CI 59.6 to 68.9) irrespective of
adductor moment group, with a mean time to conversion to knee arthroplasty
of 9.5 years (3 to 18; 95% confidence interval ± 2.5). HTO remains a useful option in the medium-term for the treatment
of medial compartment osteoarthritis of the knee but does not last
in the long-term. Cite this article:
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
Cement augmentation of pedicle screws could be used to improve screw stability, especially in osteoporotic vertebrae. However, little is known concerning the influence of different screw types and amount of cement applied. Therefore, the aim of this biomechanical A total of 54 osteoporotic human cadaver thoracic and lumbar vertebrae were instrumented with pedicle screws (uncemented, solid cemented or fenestrated cemented) and augmented with high-viscosity PMMA cement (0 mL, 1 mL or 3 mL). The insertion torque and bone mineral density were determined. Radiographs and CT scans were undertaken to evaluate cement distribution and cement leakage. Pull-out testing was performed with a material testing machine to measure failure load and stiffness. The paired Objectives
Materials and Methods
The aim of this systematic literature review was to assess the clinical level of evidence of commercially available demineralised bone matrix (DBM) products for their use in trauma and orthopaedic related surgery. A total of 17 DBM products were used as search terms in two available databases: Embase and PubMed according to the Preferred Reporting Items for Systematic Reviews and Meta Analyses statement. All articles that reported the clinical use of a DBM-product in trauma and orthopaedic related surgery were included.Objectives
Methods
A total of 30 patients with thoracolumbar/lumbar adolescent idiopathic
scoliosis (AIS) treated between 1989 and 2000 with anterior correction
and fusion surgery using dual-rod instrumentation were reviewed. Radiographic parameters and clinical outcomes were compared among
patients with lowest instrumented vertebra (LIV) at the lower end
vertebra (LEV; EV group) (n = 13) and those treated by short fusion
(S group), with LIV one level proximal to EV (n = 17 patients). Aims
Patients and Methods
This pilot study aimed to evaluate prospectively the use of inlet
radiographs of the hip as an alternative method of the assessment
of reduction after the surgical treatment of developmental dysplasia
of the hip (DDH). The children in this study underwent surgery between January
2013 and January 2015. All had inlet radiographs and CT scans post-operatively.
Data were analysed by determining inter-observer reliability and
intra-observer reproducibility, using the kappa value (K). Differences
were settled by discussion between the two observers until a consensus
was reached. The sensitivity and specificity of the radiographic
and CT results were compared. A total of 26 radiographs were obtained
from 23 children, with a mean age of 2.38 years (one to five).Aims
Patients and Methods
1. The results of treatment of 134 patients with congenital dislocation of 167 hips are reviewed. 2. Late diagnosis is still a major problem. 3. Subluxations rarely give rise to poor results, but in dislocations first treated over the age of five years there is a one-in-three risk of failure. 4. Manipulative reduction is successful less often than reduction on a frame and carries a higher risk of avascular necrosis. 5. Closed reduction on a frame was satisfactory in 58 per cent of patients under the age of three years, and can succeed up to the age of five. 6. Open reduction was required in 20 per cent of cases under the age of three, and can be used successfully up to the age of six. 7. Seven anatomical barriers to closed reduction have been recognised and two or more are commonly found in one hip when open reduction is performed. 8. The acetabular roof may fail to develop after reduction, especially if this is delayed. A C.E.
The purpose of this study was to analyse the biomechanics of
walking, through the ground reaction forces (GRF) measured, after
first metatarsal osteotomy or metatarsophalangeal joint (MTP) arthrodesis. A total of 19 patients underwent a Scarf osteotomy (50.3 years,
standard deviation (Aims
Patients and Methods
It may not be possible to undertake revision total hip arthroplasty
(THA) in the presence of massive loss of acetabular bone stock using
standard cementless hemispherical acetabular components and metal
augments, as satisfactory stability cannot always be achieved. We
aimed to study the outcome using a reconstruction cage and a porous
metal augment in these patients. A total of 22 acetabular revisions in 19 patients were performed
using a combination of a reconstruction cage and porous metal augments.
The augments were used in place of structural allografts. The mean
age of the patients at the time of surgery was 70 years (27 to 85)
and the mean follow-up was 39 months (27 to 58). The mean number
of previous THAs was 1.9 (1 to 3). All patients had segmental defects
involving more than 50% of the acetabulum and seven hips had an
associated pelvic discontinuity. Aims
Patients and Methods
Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem.Objectives
Methods