As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
Despite advances in the prevention and treatment of osteoporotic fractures, their prevalence continues to increase. Their operative treatment remains a challenge for the surgeon, often with unpredictable outcomes. This review highlights the current aspects of management of these fractures and focuses on advances in implant design and surgical technique.
The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented
Post-mortem retrieval of canine, cemented femoral components was analysed to assess the performance of these implants in the dog as a model for human total hip replacement (THR). Mechanical testing and radiological analysis were performed to determine the stability of the implant and the quality of the cement. Thirty-eight implants from 29 dogs were retrieved after time intervals ranging from 0.67 to 11.67 years. The incidence of aseptic loosening was 63.2%, much higher than in human patients (6% in post-mortem studies). Failure of the femoral implants began with debonding at the cement-metal interface, similar to that in implants in man. The incidence of aseptic loosening was much lower in bilateral than in unilateral implants. Significant differences were observed for three different designs of implant. While the dog remains the animal model of choice for THR, results from this study provide insight into interspecies differences in the performance of implants. For example, the performance of THR in dogs should be compared with that in young rather than in elderly human patients.
The reduced stability of hydroxyapatite (HA)-coated implants in osteopenic conditions is considered to be a major problem. We therefore developed a model of a boosted cementless implantation in osteopenic rats. Twelve-week-old rats were either ovariectomised (OVX) or sham-operated (SO), and after 24 weeks plain or HA-coated implants were inserted. They were treated with either a prostaglandin EP4 receptor agonist (ONO-4819) or saline for one month. The EP4 agonist considerably improved the osteoporosis in the OVX group. Ultrastructural analysis and mechanical testing showed an improvement in the implant-bone attachment in the HA-coated implants, which was further enhanced by the EP4 agonist. Although the stability of the HA-coated implants in the saline-treated OVX rats was less than in the SO normal rats, the administration of the EP4 agonist significantly compensated for this shortage. Our results showed that the osteogenic effect of the EP4 agonist augmented the osteoconductivity of HA and significantly improved the stability of the implant-bone attachment in the osteoporotic rat model.
This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite coating. Over time the hydroxyapatite coating appears to encourage filling of interface gaps remaining after surgery. Our results compare favourably with those of series describing cemented or porous-coated knee replacements, and suggest that fixation with hydroxyapatite is a reliable option in primary total knee replacement.
A pelvic discontinuity occurs when the superior
and inferior parts of the hemi-pelvis are no longer connected, which
is difficult to manage when associated with a failed total hip replacement.
Chronic pelvic discontinuity is found in 0.9% to 2.1% of hip revision
cases with risk factors including severe pelvic bone loss, female
gender, prior pelvic radiation and rheumatoid arthritis. Common
treatment options include: pelvic plating with allograft, cage reconstruction,
custom triflange implants, and porous tantalum implants with modular augments.
The optimal technique is dependent upon the degree of the discontinuity,
the amount of available bone stock and the likelihood of achieving
stable healing between the two segments. A method of treating pelvic
discontinuity using porous tantalum components with a distraction
technique that achieves both initial stability and subsequent long-term
biological fixation is described. Cite this article:
We report the five year outcomes of a two-stage
approach for infected total hip replacement. This is a single-surgeon
experience at a tertiary centre where the more straightforward cases
are treated using single-stage exchange. This study highlights the
vital role of the multidisciplinary team in managing these cases. A total of 125 patients (51 male, 74 female) with a mean age
of 68 years (42 to 78) were reviewed prospectively. Functional status
was assessed using the Harris hip score (HHS). The mean HHS improved
from 38 (6 to 78.5) pre-operatively to 81.2 (33 to 98) post-operatively.
Staphylococcus species were isolated in 85 patients (68%). The rate of control of infection was 96% at five years. In all,
19 patients died during the period of the study. This represented
a one year mortality of 0.8% and an overall mortality of 15.2% at
five years. No patients were lost to follow-up. We report excellent control of infection in a series of complex
patients and infections using a two-stage revision protocol supported
by a multidisciplinary approach. The reason for the high rate of
mortality in these patients is not known. Cite this article:
Aseptic loosening of the femoral component is
an important indication for revision surgery in unicompartmental knee
replacement (UKR). A new design of femoral component with an additional
peg was introduced for the cemented Oxford UKR to increase its stability.
The purpose of this study was to compare the primary stability of
the two designs of component. Medial Oxford UKR was performed in 12 pairs of human cadaver
knees. In each pair, one knee received the single peg and one received
the twin peg design. Three dimensional micromotion and subsidence
of the component in relation to the bone was measured under cyclical
loading at flexion of 40° and 70° using an optical measuring system.
Wilcoxon matched pairs signed-rank test was performed to detect
differences between the two groups. There was no significant difference in the relative micromotion
(p = 0.791 and 0.380, respectively) and subsidence (p = 0.301 and
0.176, respectively) of the component between the two groups at
both angles of flexion. Both designs of component offered good strength
of fixation in this cadaver study. Cite this article:
In developmental dysplasia of the hip (DDH),
a bone defect is often observed superior to the acetabulum after
the reconstruction at the level of the true acetabulum during total
hip replacement (THR). However, the essential amount of uncemented
acetabular component coverage required for a satisfactory outcome
remains controversial. The purpose of this study was to assess the
stability and function of acetabular components with a lack of coverage >
30% (31% to 50%). A total of 760 DDH patients underwent THR with
acetabular reconstruction at the level of the true floor. Lack of
coverage above the acetabular component of >
30% occurred in 56
patients. Intra-operatively, autogenous morcellised bone grafts
were used to fill the uncovered portion. Other than two screws inserted through
the acetabular shell, no additional structural supports were used
in these hips. In all, four patients were lost to follow-up. Therefore,
52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1
years (42 to 78) were available for this study at a mean of 4.8
years (3 to 7). There were no instances of prosthesis revision or
marked loosening during the follow-up. The Harris hip score improved
from a mean of 40.7 points ( Cite this article:
Because of the contradictory body of evidence related to the
potential benefits of helical blades in trochanteric fracture fixation,
we studied the effect of bone compaction resulting from the insertion
of a proximal femoral nail anti-rotation (PFNA). We developed a subject-specific computational model of a trochanteric
fracture (31-A2 in the AO classification) with lack of medial support
and varied the bone density to account for variability in bone properties
among hip fracture patients.Objectives
Methods
We analysed the long-term clinical and radiological results of 63 uncemented Low Contact Stress total knee replacements in 47 patients with rheumatoid arthritis. At a mean follow-up of 12.9 years (10 to 16), 36 patients (49 knees) were still alive; three patients (five knees) were lost to follow-up. Revision was necessary in three knees (4.8%) and the rate of infection was 3.2%. The mean clinical and functional Knee Society scores were 90 (30 to 98) and 59 (25 to 90), respectively, at final follow-up and the mean active range of movement was 104° (55° to 120°). The survival rate was 94% at 16 years but 85.5% of patients lost to follow-up were considered as failures. Radiological evidence of impending failure was noted in one knee.
We have developed a CT-based navigation system using infrared light-emitting diode markers and an optical camera. We used this system to perform cementless total hip replacement using a ceramic-on-ceramic bearing couple in 53 patients (60 hips) between 1998 and 2001. We reviewed 52 patients (59 hips) at a mean of six years (5 to 8) postoperatively. The mid-term results of total hip replacement using navigation were compared with those of 91 patients (111 hips) who underwent this procedure using the same implants, during the same period, without navigation. There were no significant differences in age, gender, diagnosis, height, weight, body mass index, or pre-operative clinical score between the two groups. The operation time was significantly longer where navigation was used, but there was no significant difference in blood loss or navigation-related complications. With navigation, the acetabular components were placed within the safe zone defined by Lewinnek, while without, 31 of the 111 components were placed outside this zone. There was no significant difference in the Merle d’Aubigne and Postel hip score at the final follow-up. However, hips treated without navigation had a higher rate of dislocation. Revision was performed in two cases undertaken without navigation, one for aseptic acetabular loosening and one for fracture of a ceramic liner, both of which showed evidence of neck impingement on the liner. A further five cases undertaken without navigation showed erosion of the posterior aspect of the neck of the femoral component on the lateral radiographs. These seven impingement-related mechanical problems correlated with malorientation of the acetabular component. There were no such mechanical problems in the navigated group. We conclude that CT-based navigation increased the precision of orientation of the acetabular component and control of limb length in total hip replacement, without navigation-related complications. It also reduced the rate of dislocation and mechanical problems related to impingement.
The Oxford mobile-bearing unicompartmental knee
replacement (UKR) is an effective and safe treatment for osteoarthritis
of the medial compartment. The results in the lateral compartment
have been disappointing due to a high early rate of dislocation
of the bearing. A series using a newly designed domed tibial component
is reported. The first 50 consecutive domed lateral Oxford UKRs in 50 patients
with a mean follow-up of three years (2.0 to 4.3) were included.
Clinical scores were obtained prospectively and Kaplan-Meier survival
analysis was performed for different endpoints. Radiological variables
related to the position and alignment of the components were measured. One patient died and none was lost to follow-up. The cumulative
incidence of dislocation was 6.2% (95% confidence interval (CI)
2.0 to 17.9) at three years. Survival using revision for any reason
and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85
to 99) at three years, respectively. Outcome scores, visual analogue
scale for pain and maximum knee flexion showed a significant improvement
(p <
0.001). The mean Oxford knee score was 43 ( Clinical results are excellent and short-term survival has improved
when compared with earlier series. The risk of dislocation remains
higher using a mobile-bearing UKR in the lateral compartment when
compared with the medial compartment. Patients should be informed
about this complication. To avoid dislocations, care must be taken
not to elevate the lateral joint line.
None
Massive endoprostheses using a cemented intramedullary stem are widely used to allow early resumption of activity after surgery for tumours. The survival of the prosthesis varies with the anatomical site, the type of prosthesis and the mode of fixation. Revision surgery is required in many cases because of aseptic loosening. Insertion of a second cemented endoprosthesis may be difficult because of the poor quality of the remaining bone, and loosening recurs quickly. We describe a series of 14 patients with triplate fixation in difficult revision or joint-sparing tumour surgery with a minimum follow-up of four years. The triplate design incorporated well within a remodelled cortex to achieve osseomechanical integration with all patients regaining their original level of function within five months. Our preliminary results suggest that this technique may provide an easy, biomechanically friendly alternative to insertion of a further device with an intramedullary stem, which has a shorter lifespan in revision or joint-sparing tumour surgery. A short segment of bone remaining after resection of a tumour will not accept an intramedullary stem, but may be soundly fixed using this method.
Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
Orthopaedic surgeons use stems in revision knee surgery to obtain
stability when metaphyseal bone is missing. No consensus exists
regarding stem size or method of fixation. This A custom test rig using differential variable reluctance transducers
(DVRTs) was developed to record all translational and rotational
motions at the bone–implant interface. Composite femurs were used.
These were secured to permit variation in flexion angle from 0°
to 90°. Cyclic loads were applied through a tibial component based
on three peaks corresponding to 0°, 10° and 20° flexion from a normal
walking cycle. Three different femoral components were investigated
in this study for cementless and cemented interface conditions.Objectives
Methods