The August 2015 Hip &
Pelvis Roundup360 looks at: The well-fixed acetabular revision; Predicting complications in revision arthroplasty; Is infection associated with fixation?; Front or back? An enduring question in hip surgery; Muscle-sparing approaches?; Gabapentin as a post-operative analgesic adjunct; An Indian take on AVN of the hip; Weight loss and arthroplasty
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
Although atlantoaxial rotatory fixation (AARF) is a common cause
of torticollis in children, the diagnosis may be delayed. The condition
is characterised by a lack of rotation at the atlantoaxial joint
which becomes fixed in a rotated and subluxed position. The management of
children with a delayed presentation of this condition is controversial.
This is a retrospective study of a group of such children. Children who were admitted to two institutions between 1988 and
2014 with a diagnosis of AARF were included. We identified 12 children
(four boys, eight girls), with a mean age of 7.3 years (1.5 to 13.4),
in whom the duration of symptoms on presentation was at least four weeks
(four to 39). All were treated with halo traction followed by a
period of cervical immobilisation in a halo vest or a Minerva jacket.
We describe a simple modification to the halo traction that allows
the child to move their head whilst maintaining traction. The mean follow-up
was 59.6 weeks (24 to 156).Aims
Patients and Methods
An atypical femoral fracture (AFF), with a transverse
fracture radiologically through the lateral cortex is a rare but serious
condition. In order to improve our ability to identify patients
with this condition, we retrospectively surveyed all subtrochanteric,
peri-implant and diaphyseal femoral fractures in patients aged ≥ 65
years who underwent surgical treatment at our hospital between 2004
and 2011. We describe the incidence of atypical fractures and their characteristics,
with observational data including a review of the hospital and general
practitioner records. Clinical outcomes were evaluated using the
Harris hip score (HHS) and the timed up-and-go (TUG) test. Atypical fractures only occurred in women with an incidence of
9.8 per 100 000 person-years. The incidence in those who were treated
with bisphosphonates was 79.0 per 100 000 person-years; eight of
17 fractures occurred around metal implants. There was a high incidence
of delayed union and revision surgery. A total of nine patients (ten
AFFs) were available for review at a mean follow-up of 36.5 months
(10 to 104). The clinical outcome was poor with a mean HHS of 58.9
(95% CI 47.4 to 70.4) and a mean TUG test of 25.7 s (95% CI 12.7
to 38.8). The delay in diagnosis and treatment of AFF may result from a
lack of knowledge of this condition. Cite this article:
To investigate the longevity of uncemented fixation of a femoral
component in total hip arthroplasty (THA) in patients with Dorr
type C proximal femoral morphology. A total of 350 consecutive uncemented THA in 320 patients were
performed between 1983 and 1987, by a single surgeon using the Taperloc
femoral component. The 63 patients (68 hips) with Dorr type C proximal
femoral morphology were the focus of this review. The mean age of
the patients was 69 years (24 to 88) and mean follow-up was 16.6
years (ten to 29). Survival analysis included eight patients (eight
hips) who died without undergoing revision surgery prior to obtaining
ten years follow-up. All 55 surviving patients (60 hips) were available
for clinical assessment and radiographic review. As a comparator
group, the survival and implant fixation in the remaining 282 THAs
(257 patients) with Dorr type A and B morphology were evaluated.
The mean age of these patients was 52 years (20 to 82).Aims
Patients and Methods
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of failure and modes of failure. We subsequently investigated
the factors that may affect the potential for the restoration of bone
stock and the durability of the re-revision reconstruction using
multivariate analysis. At the time of re-revision, there were ten host acetabula with
no significant defects, 14 with contained defects, nine with minor-column,
seven with major-column defects and ten with pelvic discontinuity.
When bone defects at re-revision were compared with those at the
previous revision, there was restoration of bone stock in 31 hips, deterioration
of bone stock in nine and remained unchanged in ten. This was a
significant improvement (p <
0.001). Morselised allografting
at the index revision was not associated with the restoration of
bone stock. In 17 hips (34%), re-revision was possible using a simple acetabular
component without allograft, augments, rings or cages. There were
47 patients with a mean follow-up of 70 months (6 to 146) available
for survival analysis. Within this group, the successful cases had
a minimum follow-up of two years after re-revision. There were 22 clinical
or radiological failures (46.7%), 18 of which were due to aseptic
loosening. The five and ten year Kaplan–Meier survival rate was
75% (95% CI, 60 to 86) and 56% (95% CI, 40 to 70) respectively with
aseptic loosening as the endpoint. The rate of aseptic loosening
was higher for hips with pelvic discontinuity (p = 0.049) and less
when the allograft had been in place for longer periods (p = 0.040). The use of a cage or ring over structural allograft bone for
massive uncontained defects in acetabular revision can restore host
bone stock and facilitate subsequent re-revision surgery to a certain
extent. Cite this article:
This review examines the future of total hip arthroplasty, aiming to avoid past mistakes
Oxidised zirconium (OxZi) has been developed
as an alternative bearing surface for femoral heads in total hip arthroplasty
(THA). This study has investigated polyethylene wear, functional
outcomes and complications, comparing OxZi and cobalt–chrome (CoCr)
as part of a three-arm, multicentre randomised controlled trial.
Patients undergoing THA from four institutions were prospectively
randomised into three groups. Group A received a CoCr femoral head
and highly cross-linked polyethylene (XLPE) liner; Group B received
an OxZi femoral head and XLPE liner; Group C received an OxZi femoral
head and ultra-high molecular weight polyethylene (UHMWPE) liner.
At five years, 368 patients had no statistically significant differences
in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western
Ontario and McMaster Universities Osteoarthritis Index (p = 0.847),
pain scores
(p = 0.458) or complications. The mean rate of linear wear was 0.028
mm/year (standard deviation ( Cite this article:
A pelvic discontinuity occurs when the superior
and inferior parts of the hemi-pelvis are no longer connected, which
is difficult to manage when associated with a failed total hip replacement.
Chronic pelvic discontinuity is found in 0.9% to 2.1% of hip revision
cases with risk factors including severe pelvic bone loss, female
gender, prior pelvic radiation and rheumatoid arthritis. Common
treatment options include: pelvic plating with allograft, cage reconstruction,
custom triflange implants, and porous tantalum implants with modular augments.
The optimal technique is dependent upon the degree of the discontinuity,
the amount of available bone stock and the likelihood of achieving
stable healing between the two segments. A method of treating pelvic
discontinuity using porous tantalum components with a distraction
technique that achieves both initial stability and subsequent long-term
biological fixation is described. Cite this article:
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
We investigated the incidence of soft-tissue
lesions after small head metal-on-metal total hip replacement (MoM THR).
Between December 1993 and May 1999, 149 patients (195 hips) underwent
primary cementless MoM THR. During the follow-up period, three patients (five THRs) died
and eight patients (14 THRs) were lost to follow-up. We requested
that all patients undergo CT evaluation. After exclusion of five
patients (six THRs) who had undergone a revision procedure, and
22 (28 THRs) who were unwilling to take part in this study, 111
patients (142 THRs) were evaluated. There were 63 men (88 THRs)
and 48 women (54 THRs) with a mean age of 45.7 years (37 to 56)
at the time of surgery. The mean follow-up was 15.4 years (13 to
19). A soft-tissue lesion was defined as an abnormal peri-prosthetic
collection of fluid, solid lesion or asymmetrical soft-tissue mass. At final follow-up, soft-tissue lesions were found in relation
to 28 THRs (19.7%), including 25 solid and three cystic lesions.
They were found in 20 men and eight women; 26 lesions were asymptomatic
and two were symptomatic. The mean maximal diameter of the soft-tissue
lesion was 42.3 mm (17 to 135). The relatively high rate of soft-tissue lesions
observed with small head MoM THR remains a concern. Cite this article:
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
To assess the extent of osteointegration in two designs of shoulder
resurfacing implants. Bony integration to the Copeland cylindrical
central stem design and the Epoca RH conical-crown design were compared. Implants retrieved from six patients in each group were pair-matched.
Mean time to revision surgery of Copeland implants was 37 months
(standard deviation (Aims
Patients and Methods
Induced membrane technique is a relatively new technique in the reconstruction of large bone defects. It involves the implantation of polymethylmethacrylate (PMMA) cement in the bone defects to induce the formation of membranes after radical debridement and reconstruction of bone defects using an autologous cancellous bone graft in a span of four to eight weeks. The purpose of this study was to explore the clinical outcomes of the induced membrane technique for the treatment of post-traumatic osteomyelitis in 32 patients. A total of 32 cases of post-traumatic osteomyelitis were admitted to our department between August 2011 and October 2012. This retrospective study included 22 men and ten women, with a mean age of 40 years (19 to 70). Within this group there were 20 tibias and 12 femurs with a mean defect of 5 cm (1.5 to 12.5). Antibiotic-loaded PMMA cement was inserted into the defects after radical debridement. After approximately eight weeks, the defects were implanted with bone graft.Objectives
Methods
There is no single standardised method of measuring
the orientation of the acetabular component on plain radiographs
after total hip arthroplasty. We assessed the reliability and accuracy
of three methods of assessing anteversion of the acetabular component
for 551 THAs using the PolyWare software and the methods of Liaw
et al, and of Woo and Morrey. All measurements of the three methods
had excellent intra- and inter-observer reliability. The values
of the PolyWare software, which determines version of the acetabular
component by edge detection were regarded as the reference standard.
Although the PolyWare software and the method of Liaw et al were similarly
precise, the method of Woo and Morrey was significantly less accurate
(p <
0.001). The method of Liaw et al seemed to be more accurate
than that of Woo and Morrey when compared with the measurements
using the PolyWare software. If the qualified lateral radiograph
was selected, anteversion measured using the method of Woo and Morrey
was considered to be relatively reliable. Cite this article:
This review summarises the evidence for the treatment
of displaced fractures of the femoral neck in elderly patients.
Results from randomised clinical trials and national register studies are
presented when available. The advantages of arthroplasty compared with internal fixation
are supported by several studies. A number of studies contribute
to the discussions of total hip arthroplasty (THA) The direct lateral approach reduces the risk of dislocation compared
with the posterior approach. Cemented implants lower the risk of
periprosthetic fracture and its subsequent morbidity and mortality.
As the risk of peri-operative death related to bone cement can be reduced
by adequate measures, cemented implants are recommended in fracture
cases. Take home message: There remains a great variation in the surgical
management of patients with a hip fracture, and an evidence-based
approach should improve the outcomes for this vulnerable patient
group. Cite this article:
We report the kinematic and early clinical results
of a patient- and observer-blinded randomised controlled trial in which
CT scans were used to compare potential impingement-free range of
movement (ROM) and acetabular component cover between patients treated
with either the navigated ‘femur-first’ total hip arthroplasty (THA) method
(n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional
THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75).
The Hip Osteoarthritis Outcome Score, the Harris hip score, the
Euro-Qol-5D and the Mancuso THA patient expectations score were
assessed at six weeks, six months and one year after surgery. A
total of 48 of the patients (84%) in the navigated ‘femur-first’
group and 43 (65%) in the conventional group reached all the desirable
potential ROM boundaries without prosthetic impingement for activities
of daily living (ADL) in flexion, extension, abduction, adduction
and rotation (p = 0.016). Acetabular component cover and surface
contact with the host bone were >
87% in both groups. There was
a significant difference between the navigated and the conventional
groups’ Harris hip scores six weeks after surgery (p = 0.010). There
were no significant differences with respect to any clinical outcome
at six months and one year of follow-up. The navigated ‘femur-first’
technique improves the potential ROM for ADL without prosthetic
impingement, although there was no observed clinical difference
between the two treatment groups. Cite this article:
The cause of dissatisfaction following total
knee arthroplasty (TKA) remains elusive. Much attention has been
focused on static mechanical alignment as a basis for surgical success and
optimising outcomes. More recently, research on both normal and
osteoarthritic knees, as well as kinematically aligned TKAs, has
suggested that other specific and dynamic factors may be more important
than a generic target of 0 ± 3º of a neutral axis. Consideration
of these other variables is necessary to understand ideal targets
and move beyond generic results. Cite this article:
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods