Advertisement for orthosearch.org.uk
Results 861 - 880 of 953
Results per page:
Bone & Joint 360
Vol. 1, Issue 1 | Pages 10 - 12
1 Feb 2012


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 342 - 348
1 Mar 2010
El Masri F Kerboull L Kerboull M Courpied JP Hamadouche M

We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component.

At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1431 - 1437
1 Nov 2009
Biring GS Kostamo T Garbuz DS Masri BA Duncan CP

We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system.

All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores.

The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 (sd 18.3). The mean Oxford-12 score was 74.0 (sd 22.3), and the mean Short form-12 score was 53.1 (sd 9.4) (mental) and 33.5 (sd 13.5) (physical). The mean satisfaction score was 90.5 (sd 15.3).

Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 757 - 761
1 Jun 2009
Nuttall D Haines JF Trail IA

In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with offset humeral head components. Over a period of 24 months the relative movement of the glenoid component with respect to the scapula was measured using radiostereometric analysis.

Nine glenoids needed reaming for erosion. There was a significant increase in rotation about all three axes with time (p < 0.001), the largest occurring about the longitudinal axis (anteversion-retroversion), with mean values of 3.8° and 1.9° for the non-offset and offset humeral head eroded subgroups, respectively. There was also a significant difference in rotation about the anteversion-retroversion axis (p = 0.01) and the varus-valgus (p < 0.001) z-axis between the two groups. The offset humeral head group reached a plateau at early follow-up with rotation about the z-axis, whereas the mean of the non-offset humeral head group at 24 months was three times greater than that of the offset group accounting for the highly significant difference between them.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 310 - 314
1 Mar 2006
Sexton SA Stossel CA Haddad FS

The Kent hip is a distally-locked femoral stem which was developed to address severe proximal bone loss, severe bony deformity and peri-prosthetic fracture.

We reviewed the results of 145 consecutive Kent hips implanted into 141 patients between 1987 and 2000. The indications for implantation were aseptic loosening (75 hips), septic loosening (two), peri-prosthetic and prosthetic fracture (37), severe bony deformity (24), and fracture through a proximal femoral metastasis (seven).

The median time to full weight-bearing after surgery was two days and the mean length of follow-up was 5.1 years (2 to 15). Further revisions were required for 13 femoral stems. With removal of the stem for any reason as an end-point, the cumulative survival at five, ten and 15 years was 93%, 89% and 77%, respectively. In patients aged ≥ 70 years, the cumulative survival at 15 years was 92%, compared with 68% in those aged < 70 years. Because of these findings, we recommend the use of interlocking stems in patients aged ≥ 70 years, particularly in those with a peri-prosthetic fracture, for whom alternative methods are limited. Outcome scores and survival data, compared with other systems, indicate that the Kent hip should be used with caution in younger patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 982 - 988
1 Jul 2012
Puri A Gulia A Jambhekar N Laskar S

We analysed the outcome of patients with primary non-metastatic diaphyseal sarcomas who had en bloc resection with preservation of the adjoining joints and reconstruction with re-implantation of sterilised tumour bone after extracorporeal radiation (50 Gy). Between March 2005 and September 2009, 32 patients (16 Ewing’s sarcoma and 16 osteogenic sarcoma) with a mean age of 15 years (2 to 35) underwent this procedure. The femur was the most common site in 17 patients, followed by the tibia in 11, humerus in three and ulna in one. The mean resected length of bone was 19 cm (10 to 26). A total of 31 patients were available at a mean follow-up of 34 months (12 to 74). The mean time to union for all osteotomy sites was 7.3 months (3 to 28): metaphyseal osteotomy sites united quicker than diaphyseal osteotomy sites (5.8 months (3 to 10) versus 9.5 months (4 to 28)). There were three local recurrences, all in soft-tissue away from irradiated graft. At the time of final follow-up, 19 patients were free of disease, one was alive with disease and 11 had died of disease. The mean Musculoskeletal Tumor Society Score for 29 patients evaluated at the last follow-up was 26 (9 to 30).

Extracorporeal irradiation is an oncologically safe and inexpensive technique for limb salvage in diaphyseal sarcomas and has good functional results.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1286 - 1292
1 Oct 2006
McLaughlin JR Lee KR

We studied a consecutive series of 285 uncemented total hip replacements in 260 patients using the Taperloc femoral component and the T-Tap acetabular component. The outcome of every hip was determined in both living and deceased patients. A complete clinical and radiological follow-up was obtained for 209 hips in 188 living patients, followed for a mean of 14.5 years (10 to 18.9). They were divided into two groups, obese and non-obese, as determined by their body mass index. There were 100 total hip replacements in 89 patients in the obese cohort (body mass index ≥ 30 kg/m2), and 109 in 99 non-obese (body mass index < 30 kg/m2) patients. A subgroup analysis of 31 patients of normal weight (body mass index 20 kg/m2 to 25 kg/m2) (33 hips) and 26 morbidly obese patients (body mass index ≥ 35 kg/m2) (30 hips) was also carried out.

In the obese group five femoral components (5%) were revised and one (1%) was loose by radiological criteria. Femoral cortical osteolysis was seen in eight hips (8%). The acetabular component was revised in 57 hips (57%) and a further 17 (17%) were loose. The mean Harris hip score improved from 52 (30 to 66) pre-operatively to 89 (49 to 100) at final follow-up. Peri-operative complications occurred in seven patients (7%).

In the non-obese group six (6%) femoral components were revised and one (1%) was loose. Femoral cortical osteolysis occurred in six hips (6%). The acetabular component was revised in 72 hips (66%) and a further 18 (17%) were loose. The mean Harris hip score increased from 53 (25 to 73) prior to surgery to 89 (53 to 100) at the time of each patient’s final follow-up radiograph.

No statistically significant difference was identified between the obese and non-obese patients with regards to clinical and radiological outcome or complications. The subgroup analysis of patients of normal weight and those who were morbidly obese showed no statistically significant difference in the rate of revision of either component.

Our findings suggest there is no evidence to support withholding total hip replacement from obese patients with arthritic hips on the grounds that their outcome will be less satisfactory than those who are not obese.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 341 - 344
1 Mar 2006
Tanaka N Sakahashi H Ishii S Kudo H

The purpose of this study was to assess the long-term results (more than ten years) of two types of cemented ulnar component with type-5 Kudo total elbow arthroplasty in a consecutive series of 56 patients (60 elbows) with rheumatoid arthritis, and to compare the results in elbows above and below a Larsen grade IV.

There was no radiolucency around the humeral component. Patients in whom a metal-backed ulnar component and a porous-coated stem were used had better clinical results and significantly less progression of radiolucent line formation around the ulnar component. They also had a significantly better long-term survival than patients with an all-polyethylene ulnar component. The clinical results of arthroplasty using all-polyethylene ulnar components were inferior, regardless of the degree of joint destruction.

We conclude that the type-5 Kudo total elbow arthroplasty with cementless fixation of the porous-coated humeral component and cemented fixation of a metal-backed ulnar component is acceptable and well-tolerated by rheumatoid patients.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 42 - 49
1 Apr 2012
Kwon Y Mellon SJ Monk P Murray DW Gill HS

Objectives

Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities.

Methods

The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient’s specific hip joint centre, based on CT scans.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1130 - 1131
1 Aug 2007
BEAVER RJ


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1592 - 1596
1 Dec 2011
Babis GC Sakellariou VI Chatziantoniou AN Soucacos PN Megas P

We report the results of 62 hips in 62 patients (17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent revision of the acetabular component of a total hip replacement due to aseptic loosening between May 2003 and November 2007. All hips had a Paprosky type IIIa acetabular defect. Acetabular revision was undertaken using a Procotyl E cementless oblong implant with modular side plates and a hook combined with impaction allografting.

At a mean follow-up of 60.5 months (36 to 94) with no patients lost to follow-up and one died due to unrelated illness, the complication rate was 38.7%. Complications included aseptic loosening (19 hips), deep infection (3 hips), broken hook and side plate (one hip) and a femoral nerve palsy (one hip). Further revision of the acetabular component was required in 18 hips (29.0%) and a further four hips (6.4%) are currently loose and awaiting revision.

We observed unacceptably high rates of complication and failure in our group of patients and cannot recommend this implant or technique.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1045 - 1048
1 Aug 2011
Avery PP Baker RP Walton MJ Rooker JC Squires B Gargan MF Bannister GC

We reviewed the seven- to ten-year results of our previously reported prospective randomised controlled trial comparing total hip replacement and hemiarthroplasty for the treatment of displaced intracapsular fracture of the femoral neck. Of our original study group of 81 patients, 47 were still alive.

After a mean follow up of nine years (7 to 10) overall mortality was 32.5% and 51.2% after total hip replacement and hemiarthroplasty, respectively (p = 0.09). At 100 months postoperatively a significantly greater proportion of hemiarthroplasty patients had died (p = 0.026). Three hips dislocated following total hip replacement and none after hemiarthroplasty. In both the total hip replacement and hemiarthroplasty groups a deterioration had occurred in walking distance (p = 0.02 and p < 0.001, respectively). One total hip replacement required revision compared with four hemiarthroplasties which were revised to total hip replacements. All surviving patients with a total hip replacement demonstrated wear of the cemented polyethylene component and all hemiarthroplasties had produced acetabular erosion.

There was lower mortality (p = 0.013) and a trend towards superior function in patients with a total hip replacement in the medium term.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1015 - 1018
1 Aug 2007
Wroblewski BM Siney PD Fleming PA

We studied survival to 38 years after Charnley low-friction arthroplasty of the hip. We used revision as an end-point, while adopting a policy of regular follow-up and early revision for radiological changes alone if indicated.

Between November 1962 and June 2005, 22 066 primary low-friction arthroplasties (17 409 patients) had been performed at Wrightington Hospital by more than 330 surgeons. By June 2006, 1001 (4.5%) hips had been revised and 1490 patients (2662 hips, 12%) had died. At 31 years, where a minimum of 40 hips were still attending follow-up, survival with revision for infection as an endpoint was 95%, for dislocation 98%, for a fractured stem 88.6%, for a loose stem 72.5% and for a loose acetabular component 53.7%. Wear and loosening of the ultra-high-molecular-weight polyethylene acetabular component were the main long-term problems.

We conclude that regular follow-up after hip replacement is essential and that all operative findings should be recorded at revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1479 - 1486
1 Nov 2011
Park J Kim Y

The purpose of this prospective, randomised study was to evaluate the clinical and radiological results comparing the identical cemented or cementless NexGen total knee prostheses implanted bilaterally in the same patient. Sequential simultaneous bilateral total knee replacements were performed in 50 patients (100 knees). There were 39 women and 11 men with a mean age of 58.4 years (51 to 67) who received a cemented prosthesis in one knee and a cementless prosthesis in the other. The mean follow-up was 13.6 years (13 to 14). At final review, the mean Knee Society scores (96.2 (82 to 100) versus 97.7 (90 to 100)), the mean Western Ontario and McMaster Universities osteoarthritis index (34.5 (4 to 59) versus 35.6 (5 to 51)), the mean ranges of knee movement (124° (100° to 140°) versus 128° (110° to 140°)), mean patient satisfaction (8.1 (sd 1.9) versus 8.3 (sd 1.7)), and radiological results were similar in both groups. The rate of survival of the femoral components was 100% in both groups at 14 years. The rate of survival of the cemented tibial component was 100% and 98% in the cementless tibial component. No osteolysis was identified in either group. Our data have shown no advantage of cementless over cemented components in total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1093 - 1097
1 Aug 2011
Weiss KR Bhumbra R Biau DJ Griffin AM Deheshi B Wunder JS Ferguson PC

Pathological fractures of the humerus are associated with pain, morbidity, loss of function and a diminished quality of life. We report our experience of stabilising these fractures using polymethylmethacrylate and non-locking plates. We undertook a retrospective review over 20 years of patients treated at a tertiary musculoskeletal oncology centre. Those who had undergone surgery for an impending or completed pathological humeral fracture with a diagnosis of metastatic disease or myeloma were identified from our database. There were 63 patients (43 men, 20 women) in the series with a mean age of 63 years (39 to 87).

All had undergone intralesional curettage of the tumour followed by fixation with intramedullary polymethylmethacrylate and plating. Complications occurred in 14 patients (22.2%) and seven (11.1%) required re-operation. At the latest follow-up, 47 patients (74.6%) were deceased and 16 (25.4%) were living with a mean follow-up of 75 months (1 to 184). A total of 54 (86%) patients had no or mild pain and 50 (80%) required no or minimal assistance with activities of daily living. Of the 16 living patients none had pain and all could perform activities of daily living without assistance.

Intralesional resection of the tumour, filling of the cavity with cement, and plate stabilisation of the pathological fracture gives immediate rigidity and allows an early return of function without the need for bony union. The patient’s local disease burden is reduced, which may alleviate tumour-related pain and slow the progression of the disease. The cemented-plate technique provides a reliable option for the treatment of pathological fractures of the humerus.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 746 - 750
1 Jun 2011
Zampelis V Ornstein E Franzén H Atroshi I

Previously, radiostereometric analysis following hip revision performed using impacted morsellised allograft bone and a cemented Exeter stem has shown continuous subsidence of the stem for up to five years. It is not known whether the subsidence continues thereafter. In our study, 17 of 25 consecutive osteo-arthritic patients with aseptically loose stems who underwent first-time revision using impacted morsellised allograft bone and a cemented Exeter stem were followed by yearly radiostereometric examinations for nine years. The mean subsidence at six weeks was 1.1 mm (0.1 to 2.3), from six weeks to one year 1.3 mm (0 to 2.6), from one to five years 0.7 mm (0 to 2.0), and from five to nine years 0.7 mm (0.1 to 3.1). That from six weeks to nine years was 2.7 mm (0 to 6.4) (95% confidence interval 2.0 to 3.5). The Charnley pain score significantly improved after revision, and was maintained at nine years, but walking ability deteriorated slightly as follow-up extended. Of the eight patients who were not followed for nine years, two had early subsidence exceeding 11 mm.

Our findings show that in osteo-arthritic patients who undergo revision for aseptic loosening of the stem using impacted morsellised allograft bone and a cemented Exeter stem, migration of the stem continues over nine years at a slower rate after the first year, but without clinical deterioration or radiological loosening.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1449 - 1456
1 Nov 2011
Fujii M Nakashima Y Yamamoto T Mawatari T Motomura G Iwamoto Y Noguchi Y

In order to clarify how intra-articular lesions influence the survival of a periacetabular osteotomy in patients with dysplasia of the hip, we performed an observational study of 121 patients (121 hips) who underwent a transposition osteotomy of the acetabulum combined with an arthroscopy. Their mean age was 40.2 years (13 to 64) and the mean follow-up was 9.9 years (2 to 18). Labral and cartilage degeneration tended to originate from the anterosuperior part of the acetabulum, followed by the femoral side. In all, eight hips (6.6%) had post-operative progression to Kellgren–Lawrence grade 4 changes, and these hips were associated with the following factors: moderate osteoarthritis, decreased width of the joint space, joint incongruity, and advanced intra-articular lesions (subchondral bone exposure in the cartilage and a full-thickness labral tear). Multivariate analysis indicated subchondral bone exposure on the femoral head as an independent risk factor for progression of osteoarthritis (p = 0.003). In hips with early stage osteoarthritis, femoral subchondral bone exposure was a risk factor for progression of the grade of osteoarthritis.

Although the outcome of transposition osteotomy of the acetabulum was satisfactory, post-operative progression of osteoarthritis occurred at a high rate in hips with advanced intra-articular lesions, particularly in those where the degenerative process had reached the point of femoral subchondral bone exposure.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1632 - 1637
1 Dec 2007
Myers GJC Abudu AT Carter SR Tillman RM Grimer RJ

We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database.

A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic loosening was the most common reason for revision in the fixed-hinge knees, fracture of the implant in the early design of rotating hinges and infection in the current version. The risk of revision for aseptic loosening in the fixed-hinge knees was 46% at ten years. This was reduced to 3% in the rotating-hinge knee with a hydroxyapatite collar. The cemented, rotating hinge design currently offers the best chance of long-term survival of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design.

At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%).

In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 409 - 413
1 Mar 2011
McCalden RW Charron KD Davidson RD Teeter MG Holdsworth DW

We present a case of early retrieval of an Oxinium femoral head and corresponding polyethylene liner where there was significant surface damage to the head and polyethylene. The implants were retrieved at the time of revision surgery to correct leg-length discrepancy just 48 hours after the primary hip replacement. Appropriate analysis of the retrieved femoral head demonstrated loss of the Oxinium layer with exposure of the underlying substrate and transfer of titanium from the acetabular shell at the time of a reduction of the index total hip replacement. In addition, the level of damage to the polyethylene was extensive despite only 48 hours in situ.

The purpose of this report is to highlight the care that is required at the time of reduction, especially with these hard femoral counter-faces such as Oxinium. To our knowledge, the damage occurring at the time of reduction has not been previously reported following the retrieval of an otherwise well-functioning hip replacement.