Advertisement for orthosearch.org.uk
Results 61 - 80 of 460
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 315 - 320
1 Mar 2011
Hart AJ Ilo K Underwood R Cann P Henckel J Lewis A Cobb J Skinner J

We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine.

We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The acetabular inclination angle was strongly positively correlated with the rate of wear. Femoral version was weakly negatively correlated with the rate of wear. Application of a threshold of > 5 μm/year for the rate of wear in order to separate the revisions into low or high wearing groups showed that more high wearing components were implanted outside Lewinnek’s safe zone, but that this was mainly due to the inclination of the acetabular component, which was the only parameter that significantly differed between the groups.

We were unable to show that excess version of the acetabular component alone or combined with femoral version was associated with an increase in the rate of wear based on our assessment of version using CT.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 12 - 19
1 Jan 2010
Vendittoli P Roy A Mottard S Girard J Lusignan D Lavigne M

We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component.

Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001).

The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 307 - 314
1 Mar 2011
Matthies A Underwood R Cann P Ilo K Nawaz Z Skinner J Hart AJ

This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR).

There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared.

Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.

Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur.

These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 20 - 27
1 Jan 2018
Sabah SA Moon JC Jenkins-Jones S Morgan CL Currie CJ Wilkinson JM Porter M Captur G Henckel J Chaturvedi N Kay P Skinner JA Hart AH Manisty C

Aims

The aim of this study was to determine whether patients with metal-on-metal (MoM) arthroplasties of the hip have an increased risk of cardiac failure compared with those with alternative types of arthroplasties (non-MoM).

Patients and Methods

A linkage study between the National Joint Registry, Hospital Episodes Statistics and records of the Office for National Statistics on deaths was undertaken. Patients who underwent elective total hip arthroplasty between January 2003 and December 2014 with no past history of cardiac failure were included and stratified as having either a MoM (n = 53 529) or a non-MoM (n = 482 247) arthroplasty. The primary outcome measure was the time to an admission to hospital for cardiac failure or death. Analysis was carried out using data from all patients and from those matched by propensity score.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1527 - 1534
1 Nov 2010
Grammatopoulos G Pandit H Murray DW Gill HS

Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients).

Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 (sd 0.10) vs mean 1.30 (sd 0.08) p = 0.001). At operation the patients with pseudotumours had a greater reduction in the size of their femoral heads (p = 0.035) and subsequently had greater neck narrowing (mean 10.1% (sd 7.2) vs mean 3.8% (sd 3.2) p < 0.001). No female patient with a pre-operative HNR ≤ 1.3 developed a pseudotumour.

We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and edge loading, and may contribute to high wear and pseudotumour formation. As the incidence of pseudotumour is low in men, it appears safe to perform resurfacing in men. However, this study suggests that it is also reasonable to resurface in women with a pre-operative HNR ≤ 1.3.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 57 - 61
1 Jun 2019
Chalmers BP Mangold DG Hanssen AD Pagnano MW Trousdale RT Abdel MP

Aims

Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR.

Patients and Methods

Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 592 - 600
1 May 2017
Matharu GS Nandra RS Berryman F Judge A Pynsent PB Dunlop DJ

Aims

To determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure.

Patients and Methods

We retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression.


The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 33 - 40
1 Apr 2017
Galea VP Laaksonen I Matuszak SJ Connelly JW Muratoglu O Malchau H

Aims

Our first aim was to determine whether there are significant changes in the level of metal ions in the blood at mid-term follow-up, in patients with an Articular Surface Replacement (ASR) arthroplasty. Secondly, we sought to identify risk factors for any increases.

Patients and Methods

The study involved 435 patients who underwent unilateral, metal-on-metal (MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These patients all had one measurement of the level of metal ions in the blood before seven years had passed post-operatively (early evaluation) and one after seven years had passed post-operatively (mid-term evaluation). Changes in ion levels were tested using a Wilcoxon signed-rank test. We identified subgroups at the highest risk of increase using a multivariable linear logistic regression model.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1035 - 1041
1 Sep 2019
Markel DC Bou-Akl T Rossi MD Pizzimenti N Wu B Ren W

Aims

The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components.

Patients and Methods

A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 2 | Pages 177 - 184
1 Mar 2004
Daniel J Pynsent PB McMinn DJW

The results of conventional hip replacement in young patients with osteoarthritis have not been encouraging even with improvements in the techniques of fixation and in the bearing surfaces. Modern metal-on-metal hip resurfacing was introduced as a less invasive method of joint reconstruction for this particular group.

This is a series of 446 hip resurfacings (384 patients) performed by one of the authors (DJWM) using cemented femoral components and hydroxyapatite-coated uncemented acetabular components with a maximum follow-up of 8.2 years (mean 3.3). Their survival rate, Oxford hip scores and activity levels are reviewed.

Six patients died due to unrelated causes. There was one revision (0.02%) out of 440 hips. The mean Oxford score of the surviving 439 hips is 13.5. None of the patients were told to change their activities at work or leisure; 31% of the men with unilateral resurfacings and 28% with bilateral resurfacings were involved in jobs that they considered heavy or moderately heavy; 92% of men with unilateral hip resurfacings and 87% of the whole group participate in leisure-time sporting activity.

The extremely low rate of failure in spite of the resumption of high level occupational and leisure activities provides early evidence of the suitability of this procedure for young and active patients with arthritis.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1304 - 1312
1 Oct 2017
Langton DJ Sidaginamale RP Joyce TJ Meek RD Bowsher JG Deehan D Nargol AVF Holland JP

Aims

We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system.

Patients and Methods

We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 7 | Pages 1055 - 1056
1 Sep 2003
Cumming D Fordyce MJF

We describe a patient who sustained a peri-prosthetic fracture of the femoral neck two weeks after undergoing a resurfacing procedure for osteoarthritis of the hip. The fracture was minimally displaced and a satisfactory outcome was obtained following conservative treatment.


The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1455 - 1462
1 Nov 2016
Matharu GS Berryman F Brash L Pynsent PB Dunlop DJ Treacy RBC

Aims

We investigated whether blood metal ion levels could effectively identify patients with bilateral Birmingham Hip Resurfacing (BHR) implants who have adverse reactions to metal debris (ARMD).

Patients and Methods

Metal ion levels in whole blood were measured in 185 patients with bilateral BHRs. Patients were divided into those with ARMD who either had undergone a revision for ARMD or had ARMD on imaging (n = 30), and those without ARMD (n = 155). Receiver operating characteristic analysis was used to determine the optimal thresholds of blood metal ion levels for identifying patients with ARMD.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1045 - 1051
1 Aug 2013
Arthroplasty Society C

The purpose of this study was twofold: first, to determine whether the five-year results of hip resurfacing arthroplasty (HRA) in Canada justified the continued use of HRA; and second, to identify whether greater refinement of patient selection was warranted.

This was a retrospective cohort study that involved a review of 2773 HRAs performed between January 2001 and December 2008 at 11 Canadian centres. Cox’s proportional hazards models were used to analyse the predictors of failure of HRA. Kaplan–Meier survival analysis was performed to predict the cumulative survival rate at five years. The factors analysed included age, gender, body mass index, pre-operative hip pathology, surgeon’s experience, surgical approach, implant sizes and implant types. The most common modes of failure were also analysed.

The 2773 HRAs were undertaken in 2450 patients: 2127 in men and 646 in women. The mean age at operation was 50.5 years (sd 8.72; 18 to 82) and mean follow-up was 3.4 years (sd 2.1; 2.0 to 10.1). At the last follow-up a total of 101 HRAs (3.6%) required revision. Using revision for all causes of failure as the endpoint, Kaplan–Meier survival analysis showed a cumulative survival of 96.4% (95% confidence interval (CI) 96.1 to 96.9) at five years. With regard to gender, the five-year overall survival was 97.4% in men (95% CI 97.1 to 97.7) and 93.6% in women (95% CI 92.6 to 94.6). Female gender, smaller femoral components, specific implant types and a diagnosis of childhood hip problems were associated with higher rates of failure. The most common cause of failure was fracture of the femoral neck, followed by loosening of the femoral component.

The failure rates of HRA at five years justify the ongoing use of this technique in men. Female gender is an independent predictor of failure, and a higher failure rate at five years in women leads the authors to recommend this technique only in exceptional circumstances for women.

Cite this article: Bone Joint J 2013;95-B:1044–50.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 42 - 49
1 Apr 2012
Kwon Y Mellon SJ Monk P Murray DW Gill HS

Objectives

Pseudotumours (abnormal peri-prosthetic soft-tissue reactions) following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have been associated with elevated metal ion levels, suggesting that excessive wear may occur due to edge-loading of these MoM implants. This study aimed to quantify in vivo edge-loading in MoMHRA patients with and without pseudotumours during functional activities.

Methods

The duration and magnitude of edge-loading in vivo was quantified during functional activities by combining the dynamic hip joint segment contact force calculated from the three-dimensional (3D) motion analysis system with the 3D reconstruction of orientation of the acetabular component and each patient’s specific hip joint centre, based on CT scans.


Bone & Joint Research
Vol. 1, Issue 4 | Pages 56 - 63
1 Apr 2012
Langton DJ Sidaginamale R Lord JK Nargol AVF Joyce TJ

Objectives

An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer.

Methods

Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular bearing surfaces and also the surfaces of the taper junctions. The relationship between taper wear rates and a number of variables, including bearing diameter and orientation of the acetabular component, was investigated.


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1172 - 1177
1 Sep 2013
Matharu GS McBryde CW Pynsent WB Pynsent PB Treacy RBC

We report the long-term survival and functional outcome of the Birmingham Hip Resurfacing (BHR) in patients aged < 50 years at operation, and explore the factors affecting survival. Between 1997 and 2006, a total of 447 BHRs were implanted in 393 patients (mean age 41.5 years (14.9 to 49.9)) by one designing surgeon. The mean follow-up was 10.1 years (5.2 to 14.7), with no loss to follow-up. In all, 16 hips (3.6%) in 15 patients were revised, giving an overall cumulative survival of 96.3% (95% confidence interval (CI) 93.7 to 98.3) at ten years and 94.1% (95% CI 84.9 to 97.3) at 14 years. Using aseptic revision as the endpoint, the survival for men with primary osteoarthritis (n = 195) was 100% (95% CI 100 to 100) at both ten years and 14 years, and in women with primary osteoarthritis (n = 109) it was 96.1% (95% CI 90.1 to 99.9) at ten years and 91.2% (95% CI 68.6 to 98.7) at 14 years. Female gender (p = 0.047) and decreasing femoral head size (p = 0.044) were significantly associated with an increased risk of revision. The median Oxford hip score (OHS, modified as a percentage with 100% indicating worst outcome) at last follow-up was 4.2% (46 of 48; interquartile range (IQR) 0% to 24%) and the median University of California, Los Angeles (UCLA) score was 6.0 (IQR 5 to 8). Men had significantly better OHS (p = 0.02) and UCLA scores (p = 0.01) than women. The BHR provides excellent survival and functional results in men into the second decade, with good results achieved in appropriately selected women.

Cite this article: Bone Joint J 2013;95-B:1172–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 471 - 476
1 Apr 2012
Holland JP Langton DJ Hashmi M

We present the clinical results and survivorship of consecutive 100 Birmingham Hip Resurfacings in 90 patients at a minimum follow-up of ten years. All procedures were carried out by an independent surgeon who commenced a prospective study in 1998. Patients were assessed clinically using the Western Ontario and McMaster Universities osteoarthritis index, Short-Form 36, Harris hip score and University of California, Los Angeles activity score. Radiological analysis was performed by an independent observer and blood metal ion levels concentrations were measured at ten years post-operatively in 62 patients. The median acetabular component inclination was 46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median chromium concentration in the unilateral group was 1.74 (0.41 to 15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38 to 19.32).

In total there were eight failures giving an overall survival at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3). BHR in male patients had an improved survivorship of 94.6% (95% CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5), but this did not reach statistical significance (p = 0.119). Four of the nine BHRs with a bearing diameter of 42 mm failed.

The overall results were consistent with data produced from other centres in that the clinical outcome of large male patients was extremely encouraging, whereas the survival of the smaller joints was less satisfactory.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 746 - 754
1 Jun 2012
Jameson SS Baker PN Mason J Porter ML Deehan DJ Reed MR

Modern metal-on-metal hip resurfacing has been widely performed in the United Kingdom for over a decade. However, the literature reports conflicting views of the benefits: excellent medium- to long-term results with some brands in specific subgroups, but high failure rates and local soft-tissue reactions in others. The National Joint Registry for England and Wales (NJR) has collected data on all hip resurfacings performed since 2003. This retrospective cohort study recorded survival time to revision from a resurfacing procedure, exploring risk factors independently associated with failure. All patients with a primary diagnosis of osteoarthritis who underwent resurfacing between 2003 and 2010 were included in the analyses. Cox’s proportional hazard models were used to analyse the extent to which the risk of revision was related to patient, surgeon and implant covariates.

A total of 27 971 hip resurfacings were performed during the study period, of which 1003 (3.59%) underwent revision surgery. In the final adjusted model, we found that women were at greater risk of revision than men (hazard ratio (HR) = 1.30, p = 0.007), but the risk of revision was independent of age. Of the implant-specific predictors, five brands had a significantly greater risk of revision than the Birmingham Hip Resurfacing (BHR) (ASR: HR = 2.82, p < 0.001, Conserve: HR = 2.03, p < 0.001, Cormet: HR = 1.43, p = 0.001, Durom: HR = 1.67, p < 0.001, Recap: HR = 1.58, p = 0.007). Smaller femoral head components were also significantly more likely to require revision (≤ 44 mm: HR = 2.14, p < 0.001, 45 to 47 mm: HR = 1.48, p = 0.001) than medium or large heads, as were operations performed by low-volume surgeons (HR = 1.36, p <  0.001). Once these influences had been removed, in 4873 male patients < 60 years old undergoing resurfacing with a BHR, the five-year estimated risk of revision was 1.59%.

In summary, after adjustment for a range of covariates we found that there were significant differences in the rate of failure between brands and component sizes. Younger male patients had good five-year implant survival when the BHR was used.