We performed an independent survivorship analysis on 208 Kinematic Condylar knee replacements with a minimum follow-up of ten years and a mean of 12 years. Seven patients had been lost to follow-up. At ten years the estimated survival was 92% (95% confidence limits 95% and 87%) and when stratified for diagnosis and thickness of polyethylene there was no statistical difference (p >
0.05) in survivorship of knees with osteoarthritis or rheumatoid arthritis. We conclude that the original design of the Kinematic Condylar knee replacement has a good record and that adequate evaluation of new designs of implant should be undertaken before they are widely introduced.
We reviewed 32 deep-frozen irradiated allografts used for the reconstruction of bone defects in 20 knees. They were subdivided into bulk grafts, cortical strut grafts, and morsellised bone. The average follow-up was 4.2 years (2 to 7.2). Radiographs showed union of the allograft to the host in all cases. Two allografts later fractured and three knees required further surgery because of infection. The allografts effectively filled large bone defects around the knee, lessening the need for custom-made and constrained prostheses.
A review of patients with an infected resurfacing prosthesis is presented. Eight patients with a loose infected prosthesis were treated by a one-stage exchange arthroplasty; six others with a well-fixed infected prosthesis were treated by drainage and antibiotics. All eight treated by exchange arthroplasty remained free of infection as did five of those treated by drainage. In four of these last five patients, the prosthesis was inserted without cement; the possible role of polymethylmethacrylate in the persistence of infection is discussed.
Aims. To investigate the optimal thresholds and diagnostic efficacy of commonly used serological and synovial fluid detection indexes for diagnosing periprosthetic joint infection (PJI) in patients who have rheumatoid arthritis (RA). Methods. The data from 348 patients who had RA or osteoarthritis (OA) and had previously undergone a total knee (TKA) and/or a total hip arthroplasty (THA) (including RA-PJI: 60 cases, RA-non-PJI: 80 cases; OA-PJI: 104 cases, OA-non-PJI: 104 cases) were retrospectively analyzed. A receiver operating characteristic curve was used to determine the optimal thresholds of the CRP, ESR, synovial fluid white blood cell count (WBC), and polymorphonuclear neutrophil percentage (PMN%) for diagnosing RA-PJI and OA-PJI. The diagnostic efficacy was evaluated by comparing the area under the curve (AUC) of each index and applying the results of the combined index diagnostic test. Results. For PJI prediction, the results of serological and synovial fluid indexes were different between the RA-PJI and OA-PJI groups. The optimal cutoff value of CRP for diagnosing RA-PJI was 12.5 mg/l, ESR was 39 mm/hour, synovial fluid WBC was 3,654/μl, and PMN% was 65.9%; and those of OA-PJI were 8.2 mg/l, 31 mm/hour, 2,673/μl, and 62.0%, respectively. In the RA-PJI group, the specificity (94.4%), positive predictive value (97.1%), and AUC (0.916) of synovial fluid WBC were higher than those of the other indexes. The optimal cutoff values of synovial fluid WBC and PMN% for diagnosing RA-PJI after THA were significantly higher than those of
Aims. Antibiotic prophylaxis involving timely administration of appropriately dosed antibiotic is considered effective to reduce the risk of surgical site infection (SSI) after total hip and
Aims. While preoperative bloodwork is routinely ordered, its value in determining which patients are at risk of postoperative readmission following
Aims. The aims of this study were to assess mapping models to predict the three-level version of EuroQoL five-dimension utility index (EQ-5D-3L) from the Oxford Knee Score (OKS) and validate these before and after
Aims. The aim of this meta-analysis was to determine the pooled incidence of postoperative urinary retention (POUR) following total hip and knee arthroplasty (total joint replacement (TJR)) and to evaluate the risk factors and complications associated with POUR. Methods. Two authors conducted searches in PubMed, Embase, Web of Science, and Scopus on TJR and urinary retention. Eligible studies that reported the rate of POUR and associated risk factors for patients undergoing TJR were included in the analysis. Patient demographic details, medical comorbidities, and postoperative outcomes and complications were separately analyzed. The effect estimates for continuous and categorical data were reported as standardized mean differences (SMDs) and odds ratios (ORs) with 95% CIs, respectively. Results. A total of 31 studies were included in the systematic review. Of these, 29 studies entered our meta-analysis, which included 3,273 patients diagnosed with POUR and 11,583 patients without POUR following TJR. The pooled incidence of POUR was 28.06%. Demographic risk factors included male sex (OR 1.81, 95% CI 1.26 to 2.59), increasing age (SMD 0.16, 95% CI 0.04 to 0.27), and American Society of Anesthesiologists grade 3 to 4 (OR 1.39, 95% CI 1.10 to 1.77). Patients with a history of benign prostatic hyperplasia (OR 1.99, 95% CI 1.41 to 2.83) and retention (OR 3.10, 95% CI 1.58 to 6.06) were more likely to develop POUR. Surgery-related risk factors included spinal anaesthesia (OR 1.44, 95% CI 1.19 to 1.74) and postoperative epidural analgesia (OR 2.82, 95% CI 1.65 to 4.82). Total hip arthroplasty was associated with higher odds of POUR compared to
Aims. Sagittal plane imbalance (SPI), or asymmetry between extension and flexion gaps, is an important issue in
Aims. Tibial fixation in revision total knee arthroplasty (rTKA) can present surgical challenges. It has been suggested that appropriate fixation in at least two of the three anatomical zones (epiphysis, metaphysis, and diaphysis) is essential for implant survival. However, supporting clinical data are lacking. In this retrospective case-control study, we investigated the relationship between zonal fixation of hybrid rTKA tibial components and re-revision
Aims. The transepicondylar axis is a well-established reference for the determination of femoral component rotation in
We have investigated the errors in the identification of the transepicondylar axis and the anteroposterior axis between a minimally-invasive and a conventional approach in four fresh-frozen cadaver knees. The errors in aligning the femoral prosthesis were compared with the reference transepicondylar axis as established by CT. The error in the identification of the transepicondylar axis was significantly higher in the minimal approach (4.5° of internal rotation,
Aims. Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Methods. Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating
Estimates of knee joint loadings were calculated for 12 normal subjects from kinematic and kinetic measures obtained during both level and downhill walking. The maximum tibiofemoral compressive force reached an average load of 3.9 times body-weight (BW) for level walking and 8 times BW for downhill walking, in each instance during the early stance phase. Muscle forces contributed 80% of the maximum bone-on-bone force during downhill walking and 70% during level walking whereas the ground reaction forces contributed only 20% and 30% respectively. Most total knee designs provide a tibiofemoral contact area of 100 to 300 mm2. The yield point of these polyethylene inlays will therefore be exceeded with each step during downhill walking. Future evaluation of total knee designs should be based on a tibiofemoral joint load of 3.5 times BW at 20° knee flexion, 8 times BW at 40° and 6 times BW at 60°.
Maquet's line passes from the centre of the femoral head to the centre of the body of the talus. The distance of this line from the centre of the knee on a long-leg radiograph provides the most accurate measure of coronal alignment. Malalignment causes abnormal forces which may lead to loosening after knee replacement. We report a series of 115 Denham knee replacements performed between 1976 and 1981 using the earliest design of components, inserted with intramedullary guide rods. Patients were assessed clinically and long-leg standing radiographs were taken before operation, soon after surgery and up to 12 years later. In two-thirds of the knees (68%) Maquet's line passed through the middle third of the prosthesis on postoperative films and the incidence of subsequent loosening was 3%. When Maquet's line was medial or lateral to this, an error of approximately +/- 3 degrees, the incidence of loosening at a median period of eight years was 24%. This difference is highly significant (p = 0.001). Accurate coronal alignment appears to be an important factor in prevention of loosening. Means of improving the accuracy of alignment and of measuring it on long-leg radiographs are discussed.
We reviewed 40 cementless Porous Coated Anatomic knee replacements in 34 consecutive patients. The average follow-up was 12.9 months (range 6 to 36 months). At review, loose beads were identified in 23 knees, in 11 on the femoral side and in 16 on the tibial side (in five knees on both sides). One loose bead was adjacent to the patella. In four knees there were intra-articular beads. Radiolucent lines at the interface between bone and prosthesis were seen in 19 knees, in six on the femoral side and in 16 on the tibial side (in three knees on both sides). In relation to the tibial component loose beads were associated with a radiolucent line in all except one case. Progressive loosening with an increasing number of loose beads was seen in three knees. At this early stage of follow-up there was no correlation between the clinical results and the presence of loose beads.
Aims. Simultaneous bilateral total knee arthroplasty (TKA) has been used due to its financial advantages, overall resource usage, and convenience for the patient. The training model where a trainee performs the first