When fracture of an extensively porous-coated
femoral component occurs, its removal at revision total hip arthroplasty
(THA) may require a femoral osteotomy and the use of a trephine.
The remaining cortical bone after using the trephine may develop
thermally induced necrosis. A retrospective review identified 11
fractured, well-fixed, uncemented, extensively porous-coated femoral
components requiring removal using a trephine with a minimum of
two years of follow-up. The mean time to failure was 4.6 years (1.7 to 9.1, standard
deviation ( A total of four patients (36.4%) required further revision: three
for instability and one for fracture of the revision component.
There was no statistically significant difference in the mean Harris
hip score before implant fracture (82.4; These findings suggest that removal of a fractured, well-fixed,
uncemented, extensively porous-coated femoral component using a
trephine does not compromise subsequent fixation at revision THA
and the patient’s pre-operative level of function can be restored.
However, the loss of proximal bone stock before revision may be associated
with a high rate of dislocation post-operatively. Cite this article:
The April 2015 Children’s orthopaedics Roundup360 looks at: Reducing the incidence of DDH – is ‘back carrying’ the answer?; Surgical approach and AVN may not be linked in DDH; First year routine radiographic follow up for scoliosis not necessary; Diagnosis of osteochondritis dessicans; Telemedicine in paediatrics; Regional anesthesia in supracondylar fractures?
Previous classification systems of failure of
limb salvage focused primarily on endoprosthetic failures and lacked sufficient
depth for the effective study of the causes of failure. In order
to address these inadequacies, the International Society of Limb
Salvage (ISOLS) formed a committee to recommend revisions of the
previous systems. The purpose of this study was to report on their
recommendations. The modifications were prepared using an earlier,
evidence-based model with subclassification based on the existing
medical literature. Subclassification for all five primary types
of failure of limb salvage following endoprosthetic reconstruction
were formulated and a complementary system was derived for the failure
of biological reconstruction. An additional classification of failure
in paediatric patients was also described. Limb salvage surgery presents a complex array of potential mechanisms
of failure, and a complete and precise classification of types of
failure is required. Earlier classification systems lacked specificity,
and the evidence-based system outlined here is designed to correct
these weaknesses and to provide a means of reporting failures of
limb salvage in order to allow the interpretation of outcome following
reconstructive surgery. Cite this article:
Rarely, the extent of a malignant bone tumour
may necessitate resection of the complete humerus to achieve adequate
oncological clearance. We present our experience with reconstruction
in such cases using a total humeral endoprosthesis (THER) in 20
patients (12 male and eight female) with a mean age of 22 years
(6 to 59). We assessed the complications, the oncological and functional
outcomes and implant survival. Surgery was performed between June
2001 and October 2009. The diagnosis included osteosarcoma in nine,
Ewing’s sarcoma in eight and chondrosarcoma in three. One patient
was lost to follow-up. The mean follow-up was 41 months (10 to 120)
for all patients and 56 months (25 to 120) in survivors. There were
five local recurrences (26.3%) and 11 patients were alive at time
of last follow-up, with overall survival for all patients being
52% (95% confidence interval (CI) 23.8 to 74) at five years. The
mean Musculoskeletal Tumor Society score for the survivors was 22
(73%; 16 to 23). The implant survival was 95% (95% CI 69.5 to 99.3)
at five years. The use of a THER in the treatment of malignant tumours of bone
is oncologically safe; it gives consistent and predictable results
with low rates of complication.
The June 2012 Oncology Roundup360 looks at: avoiding pelvic hemipelvectomy; proximal femoral metastasis; extendible prostheses; rotationplasty; soft-tissue sarcomas; osteosarcoma of the pelvis; recurrent chondrosarcoma ; MRI and the differentiation between benign and malignant lesions; and malignant fibrous histiocytoma.
There are no recent studies comparing cable with
wire for the fixation of osteotomies or fractures in total hip replacement
(THR). Our objective was to evaluate the five-year clinical and
radiological outcomes and complication rates of the two techniques.
We undertook a review including all primary and revision THRs performed
in one hospital between 1996 and 2005 using cable or wire fixation.
Clinical and radiological evaluation was performed five years post-operatively.
Cables were used in 51 THRs and wires in 126, and of these, 36 THRs
with cable (71%) and 101 with wire (80%) were evaluated at follow-up.
The five-year radiographs available for 33 cable and 91 wire THRs
revealed rates of breakage of fixation of 12 of 33 (36%) and 42
of 91 (46%), respectively. With cable there was a significantly
higher risk of metal debris (68% In conclusion, we found a higher incidence of complications and
a trend towards increased infection and foreign-body reaction with
the use of cables.
We reviewed our initial seven-year experience
with a non-invasive extendible prosthesis in 34 children with primary bone
tumours. The distal femur was replaced in 25 cases, total femur
in five, proximal femur in one and proximal tibia in three. The
mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain
alive. The prostheses were lengthened by an electromagnetic induction
mechanism in an outpatient setting and a mean extension of 32 mm
(4 to 80) was achieved without anaesthesia. There were lengthening
complications in two children: failed lengthening in one and the
formation of scar tissue in the other. Deep infection developed
in six patients (18%) and local recurrence in three. A total of
11 patients required further surgery to the leg. Amputation was
necessary in five patients (20%) and a two-stage revision in another.
There were no cases of loosening, but two patients had implant breakage
and required revision. The mean Musculoskeletal Tumor Society functional
score was 85% (60% to 100%) at last known follow-up. These early
results demonstrate that the non-invasive extendible prosthesis
allows successful lengthening without surgical intervention, but
the high incidence of infection is a cause for concern.
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
Guiding growth by harnessing the ability of growing bone to undergo plastic deformation is one of the oldest orthopaedic principles. Correction of deformity remains a major part of the workload for paediatric orthopaedic surgeons and recently, along with developments in limb reconstruction and computer-directed frame correction, there has been renewed interest in surgical methods of physeal manipulation or ‘guided growth’. Manipulating natural bone growth to correct a deformity is appealing, as it allows gradual correction by non- or minimally invasive methods. This paper reviews the techniques employed for guided growth in current orthopaedic practice, including the basic science and recent advances underlying mechanical physeal manipulation of both healthy and pathological physes.
There have been recent reports linking alendronate and a specific pattern of subtrochanteric insufficiency fracture. We performed a retrospective review of all subtrochanteric fractures admitted to our institution between 2001 and 2007. There were 20 patients who met the inclusion criteria, 12 of whom were on long-term alendronate. Alendronate-associated fractures tend to be bilateral (Fisher’s exact test, p = 0.018), have unique radiological features (p <
0.0005), be associated radiologically with a pre-existing ellipsoid thickening of the lateral femoral cortex and are likely to be preceded by prodromal pain. Biomechanical investigations did not suggest overt metabolic bone disease. Only one patient on alendronate had osteoporosis prior to the start of therapy. We used these findings to develop a management protocol to optimise fracture healing. We also advocate careful surveillance in individuals at-risk, and present our experience with screening and prophylactic fixation in selected patients.
Despite advances in the prevention and treatment of osteoporotic fractures, their prevalence continues to increase. Their operative treatment remains a challenge for the surgeon, often with unpredictable outcomes. This review highlights the current aspects of management of these fractures and focuses on advances in implant design and surgical technique.
We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation.
In developing countries locally-made low-cost prostheses are mainly used in limb-salvage surgery to alleviate the economic burden. We retrospectively collected data on 104 patients treated by limb-salvage surgery between July 1997 and July 2005. We used a locally-designed and fabricated stainless-steel endoprosthesis in each case. Oncological and functional outcomes were evaluated at a mean follow-up of 47 months (12 to 118). A total of 73 patients (70.2%) were free from disease, nine (8.7%) were alive with disease, 19 (18.2%) had died from their disease and three (2.9%) from unrelated causes. According to the Musculoskeletal Tumor Society scoring system, the mean functional score was 76.3% (SD 17.8). The five-year survival for the implant was 70.5%. There were nine cases (8.7%) of infection, seven early and two late, seven (6.7%) of breakage of the prosthesis, three (2.9%) of aseptic loosening and two (1.9%) of failure of the polyethylene bushing. Multivariate analysis showed that a proximal tibial prosthesis and a resection length of 14 cm or more were significant negative prognostic factors. Our survival rates and Musculoskeletal Tumor Society functional scores are similar to those reported in the literature. Although longer follow-up is needed to confirm our results, we believe that a low-cost custom-made endoprosthesis is a cost-effective and reliable reconstructive option for limb salvage in developing countries.
We investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the distal femur after resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database, review of the design of the implant and case records. In total, 335 patients underwent a distal femoral replacement, 162 having a fixed-hinge design and 173 a rotating-hinge. The median age of the patients was 24 years (interquartile range 17 to 48). A total of 192 patients remained alive with a mean follow-up of 12 years (5 to 30). The risk of revision for any reason was 17% at five years, 33% at ten years and 58% at 20 years. Aseptic loosening was the main reason for revision of the fixed-hinge knees while infection and fracture of the stem were the most common for the rotating-hinge implant. The risk of revision for aseptic loosening was 35% at ten years with the fixed-hinge knee, which has, however, been replaced by the rotating-hinge knee with a hydroxyapatite collar. The overall risk of revision for any reason fell by 52% when the rotating-hinge implant was used. Improvements in the design of distal femoral endoprostheses have significantly decreased the need for revision operations, but infection remains a serious problem. We believe that a cemented, rotating-hinge prosthesis with a hydroxyapatite collar offers the best chance of long-term survival of the prosthesis.