Revision total hip arthroplasty after removal of a fractured well-fixed extensively porous-coated femoral component using a trephine



Download PDF

Abstract

When fracture of an extensively porous-coated femoral component occurs, its removal at revision total hip arthroplasty (THA) may require a femoral osteotomy and the use of a trephine. The remaining cortical bone after using the trephine may develop thermally induced necrosis. A retrospective review identified 11 fractured, well-fixed, uncemented, extensively porous-coated femoral components requiring removal using a trephine with a minimum of two years of follow-up.

The mean time to failure was 4.6 years (1.7 to 9.1, standard deviation (sd) 2.3). These were revised using a larger extensively porous coated component, fluted tapered modular component, a proximally coated modular component, or a proximal femoral replacement. The mean clinical follow-up after revision THA was 4.9 years (2 to 22, sd 3.1). The mean diameter of the femoral component increased from 12.7 mm (sd 1.9) to 16.2 mm (sd 3.4; p >  0.001). Two revision components had radiographic evidence of subsidence that remained radiographically stable at final follow-up. The most common post-operative complication was instability affecting six patients (54.5%) on at least one occasion.

A total of four patients (36.4%) required further revision: three for instability and one for fracture of the revision component. There was no statistically significant difference in the mean Harris hip score before implant fracture (82.4; sd 18.3) and after trephine removal and revision THA (81.2; sd 14.8, p = 0.918).

These findings suggest that removal of a fractured, well-fixed, uncemented, extensively porous-coated femoral component using a trephine does not compromise subsequent fixation at revision THA and the patient’s pre-operative level of function can be restored. However, the loss of proximal bone stock before revision may be associated with a high rate of dislocation post-operatively.

Cite this article: Bone Joint J 2015;97-B:1192–6.