Advertisement for orthosearch.org.uk
Results 21 - 40 of 43
Results per page:
Bone & Joint Research
Vol. 8, Issue 11 | Pages 563 - 569
1 Nov 2019
Koh Y Lee J Lee H Kim H Kang K

Objectives

Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty with isolated medial or lateral compartment osteoarthritis. However, polyethylene wear can significantly reduce the lifespan of UKA. Different bearing designs and materials for UKA have been developed to change the rate of polyethylene wear. Therefore, the objective of this study is to investigate the effect of insert conformity and material on the predicted wear in mobile-bearing UKA using a previously developed computational wear method.

Methods

Two different designs were tested with the same femoral component under identical kinematic input: anatomy mimetic design (AMD) and conforming design inserts with different conformity levels. The insert materials were standard or crosslinked ultra-high-molecular-weight polyethylene (UHMWPE). We evaluated the contact pressure, contact area, wear rate, wear depth, and volumetric wear under gait cycle loading conditions.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims

The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer.

Materials and Methods

Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1310 - 1319
1 Oct 2018
Langton DJ Wells SR Joyce TJ Bowsher JG Deehan D Green S Nargol AVF Holland JP

Aims

There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs.

Patients and Methods

We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 68 - 76
1 Jun 2019
Jones CW Choi DS Sun P Chiu Y Lipman JD Lyman S Bostrom MPG Sculco PK

Aims

Custom flange acetabular components (CFACs) are a patient-specific option for addressing large acetabular defects at revision total hip arthroplasty (THA), but patient and implant characteristics that affect survivorship remain unknown. This study aimed to identify patient and design factors related to survivorship.

Patients and Methods

A retrospective review of 91 patients who underwent revision THA using 96 CFACs was undertaken, comparing features between radiologically failed and successful cases. Patient characteristics (demographic, clinical, and radiological) and implant features (design characteristics and intraoperative features) were collected. There were 74 women and 22 men; their mean age was 62 years (31 to 85). The mean follow-up was 24.9 months (sd 27.6; 0 to 116). Two sets of statistical analyses were performed: 1) univariate analyses (Pearson’s chi-squared and independent-samples Student’s t-tests) for each feature; and 2) bivariable logistic regressions using features identified from a random forest analysis.


Aims

This study aimed to compare the change in health-related quality of life of patients receiving a traditional cemented monoblock Thompson hemiarthroplasty compared with a modern cemented modular polished-taper stemmed hemiarthroplasty for displaced intracapsular hip fractures.

Patients and Methods

This was a pragmatic, multicentre, multisurgeon, two-arm, parallel group, randomized standard-of-care controlled trial. It was embedded within the WHiTE Comprehensive Cohort Study. The sample size was 964 patients. The setting was five National Health Service Trauma Hospitals in England. A total of 964 patients over 60 years of age who required hemiarthroplasty of the hip between February 2015 and March 2016 were included. A standardized measure of health outcome, the EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and at four months following the operation.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 807 - 816
1 Jun 2014
Rajaee SS Kanim LEA Bae HW

Using the United States Nationwide Inpatient Sample, we identified national trends in revision spinal fusion along with a comprehensive comparison of comorbidities, inpatient complications and surgical factors of revision spinal fusion compared to primary spinal fusion.

In 2009, there were 410 158 primary spinal fusion discharges and 22 128 revision spinal fusion discharges. Between 2002 and 2009, primary fusion increased at a higher rate compared with revision fusion (56.4% vs 51.0%; p < 0.001). In 2009, the mean length of stay and hospital charges were higher for revision fusion discharges than for primary fusion discharges (4.2 days vs 3.8 days, p < 0.001; USD $91 909 vs. $87 161, p < 0.001). In 2009, recombinant human bone morphogenetic protein (BMP) was used more in revision fusion than in primary fusion (39.6% vs 27.6%, p < 0.001), whereas interbody devices were used less in revision fusion (41.8% vs 56.6%, p < 0.001).

In the multivariable logistic regression model for all spinal fusions, depression (odds ratio (OR) 1.53, p < 0.001), psychotic disorders (OR 1.49, p < 0.001), deficiency anaemias (OR 1.35, p < 0.001) and smoking (OR 1.10, p = 0.006) had a greater chance of occurrence in revision spinal fusion discharges than in primary fusion discharges, adjusting for other variables. In terms of complications, after adjusting for all significant comorbidities, this study found that dural tears (OR 1.41; p < 0.001) and surgical site infections (OR 3.40; p < 0.001) had a greater chance of occurrence in revision spinal fusion discharges than in primary fusion discharges (p < 0.001). A p-value < 0.01 was considered significant in all final analyses.

Cite this article: Bone Joint J 2014;96-B:807–16.


Bone & Joint Research
Vol. 5, Issue 6 | Pages 206 - 214
1 Jun 2016
Malak TT Broomfield JAJ Palmer AJR Hopewell S Carr A Brown C Prieto-Alhambra D Glyn-Jones S

Objectives

High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA).

Methods

We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 766 - 773
1 Jun 2017
Graves SE de Steiger R Davidson D Donnelly W Rainbird S Lorimer MF Cashman KS Vial RJ

Aims

Femoral stems with exchangeable (modular) necks were introduced to offer surgeons an increased choice when determining the version, offset and length of the femoral neck during total hip arthroplasty (THA). It was hoped that this would improve outcomes and reduce complications, particularly dislocation. In 2010, the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) first reported an increased rate of revision after primary THA using femoral stems with an exchangeable neck. The aim of this study was to provide a more comprehensive up-to-date analysis of primary THA using femoral stems with exchangeable and fixed necks.

Materials and Methods

The data included all primary THA procedures performed for osteoarthritis (OA), reported to the AOANJRR between 01 September 1999 and 31 December 2014. There were 9289 femoral stems with an exchangeable neck and 253 165 femoral stems with a fixed neck. The characteristics of the patients and prostheses including the bearing surface and stem/neck metal combinations were examined using Cox proportional hazard ratios (HRs) and Kaplan-Meier estimates of survivorship.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 892 - 900
1 Jul 2016
Atrey A Heylen S Gosling O Porteous MJL Haddad FS

Joint replacement of the hip and knee remain very satisfactory operations. They are, however, expensive. The actual manufacturing of the implant represents only 30% of the final cost, while sales and marketing represent 40%. Recently, the patents on many well established and successful implants have expired. Companies have started producing and distributing implants that purport to replicate existing implants with good long-term results.

The aims of this paper are to assess the legality, the monitoring and cost saving implications of such generic implants. We also assess how this might affect the traditional orthopaedic implant companies.

Cite this article: Bone Joint J 2016;98-B:892–900.


Bone & Joint Research
Vol. 5, Issue 8 | Pages 338 - 346
1 Aug 2016
MacLeod AR Sullivan NPT Whitehouse MR Gill HS

Objectives

Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection.

Materials and Methods

Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.


Bone & Joint 360
Vol. 4, Issue 6 | Pages 2 - 5
1 Dec 2015
Dodd L Sharpe I Mandalia VI Toms AD Phillips JRA

The global economy has been facing a financial crisis. Healthcare costs are spiraling, and there is a projected £30 billion health funding gap by 2020 in the UK.1 This has prompted a drive for efficiency in healthcare provision in the UK, and in 2012, the Health and Social Care Act was introduced, heralding a fundamental change to the structure of the National Health Service, especially in the way that healthcare is funded in England.2

What is happening in the UK is a reflection of a global problem. Rationing of healthcare is a topic of much discussion; as unless spending is capped, providing healthcare will become unsustainable. Who decides how money is spent, and which services should be rationed?

In this article we aim to discuss the impact that rationing may have on orthopaedic surgery, and we will discuss our own experiences of attempts to ration local services.3 We also seek to inform and educate the general orthopaedic community on this topic.


Bone & Joint 360
Vol. 4, Issue 4 | Pages 8 - 11
1 Aug 2015
McBride A Nicol S Monsell F


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 997 - 1003
1 Jul 2015
Papanagiotou M Dailiana ZH Karachalios T Varitimidis S Vlychou M Hantes M Malizos KN

We report the outcome of 84 nonunions involving long bones which were treated with rhBMP-7, in 84 patients (60 men: 24 women) with a mean age 46 years (18 to 81) between 2003 and 2011. The patients had undergone a mean of three previous operations (one to 11) for nonunion which had been present for a mean of 17 months (4 months to 20 years). The nonunions involved the lower limb in 71 patients and the remainder involved the upper limb. A total of 30 nonunions were septic. Treatment was considered successful when the nonunion healed without additional procedures. The relationship between successful union and the time to union was investigated and various factors including age and gender, the nature of the nonunion (location, size, type, chronicity, previous procedures, infection, the condition of the soft tissues) and type of index procedure (revision of fixation, type of graft, amount of rhBMP-7) were analysed. The improvement of the patients’ quality of life was estimated using the Short Form (SF) 12 score.

A total of 68 nonunions (80.9%) healed with no need for further procedures at a mean of 5.4 months (3 to 10) post-operatively. Multivariate logistic regression analysis of the factors affecting union suggested that only infection significantly affected the rate of union (p = 0.004).Time to union was only affected by the number of previous failed procedures (p = 0.006). An improvement of 79% and 32.2% in SF-12 physical and mental score, respectively, was noted within the first post-operative year.

Rh-BMP-7 combined with bone grafts, enabled healing of the nonunion and improved quality of life in about 80% of patients. Aseptic nonunions were much more likely to unite than septic ones. The number of previous failed operations significantly delayed the time to union.

Cite this article: Bone Joint J 2015;97-B:997–1003.


Bone & Joint 360
Vol. 3, Issue 5 | Pages 36 - 37
1 Oct 2014
Di Martino A


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 101 - 104
1 Nov 2014
Lombardi Jr AV Berend KR Adams JB

Previous studies of failure mechanisms leading to revision total knee replacement (TKR) performed between 1986 and 2000 determined that many failed early, with a disproportionate amount accounted for by infection and implant-associated factors including wear, loosening and instability. Since then, efforts have been made to improve implant performance and instruct surgeons in best practice. Recently our centre participated in a multi-centre evaluation of 844 revision TKRs from 2010 to 2011. The purpose was to report a detailed analysis of failure mechanisms over time and to see if failure modes have changed over the past 10 to 15 years. Aseptic loosening was the predominant mechanism of failure (31.2%), followed by instability (18.7%), infection (16.2%), polyethylene wear (10.0%), arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to failure was 5.9 years (ten days to 31 years), 35.3% of all revisions occurred at less than two years, and 60.2% in the first five years. With improvements in implant and polyethylene manufacture, polyethylene wear is no longer a leading cause of failure. Early mechanisms of failure are primarily technical errors. In addition to improving implant longevity, industry and surgeons must work together to decrease these technical errors. All reports on failure of TKR contain patients with unexplained pain who not infrequently have unmet expectations. Surgeons must work to achieve realistic patient expectations pre-operatively, and therefore, improve patient satisfaction post-operatively.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):101–4.


Bone & Joint 360
Vol. 4, Issue 1 | Pages 6 - 11
1 Feb 2015
Manktelow A Bloch B

This review examines the future of total hip arthroplasty, aiming to avoid past mistakes


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 10 - 18
1 Jan 2015
Sabah SA Henckel J Cook E Whittaker R Hothi H Pappas Y Blunn G Skinner JA Hart AJ

Arthroplasty registries are important for the surveillance of joint replacements and the evaluation of outcome. Independent validation of registry data ensures high quality. The ability for orthopaedic implant retrieval centres to validate registry data is not known. We analysed data from the National Joint Registry for England, Wales and Northern Ireland (NJR) for primary metal-on-metal hip arthroplasties performed between 2003 and 2013. Records were linked to the London Implant Retrieval Centre (RC) for validation. A total of 67 045 procedures on the NJR and 782 revised pairs of components from the RC were included. We were able to link 476 procedures (60.9%) recorded with the RC to the NJR successfully. However, 306 procedures (39.1%) could not be linked. The outcome recorded by the NJR (as either revised, unrevised or death) for a primary procedure was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval linkage and correct assignment of outcome code improved over time. The rates of error for component reference numbers on the NJR were as follows: femoral head category number 14/229 (5.0%); femoral head batch number 13/232 (5.3%); acetabular component category number 2/293 (0.7%) and acetabular component batch number 24/347 (6.5%).

Registry-retrieval linkage provided a novel means for the validation of data, particularly for component fields. This study suggests that NJR reports may underestimate rates of revision for many types of metal-on-metal hip replacement. This is topical given the increasing scope for NJR data. We recommend a system for continuous independent evaluation of the quality and validity of NJR data.

Cite this article: Bone Joint J 2015;97-B:10–18.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1578 - 1585
1 Dec 2014
Rankin KS Sprowson AP McNamara I Akiyama T Buchbinder R Costa ML Rasmussen S Nathan SS Kumta S Rangan A

Trauma and orthopaedics is the largest of the surgical specialties and yet attracts a disproportionately small fraction of available national and international funding for health research. With the burden of musculoskeletal disease increasing, high-quality research is required to improve the evidence base for orthopaedic practice. Using the current research landscape in the United Kingdom as an example, but also addressing the international perspective, we highlight the issues surrounding poor levels of research funding in trauma and orthopaedics and indicate avenues for improving the impact and success of surgical musculoskeletal research.

Cite this article: Bone Joint J 2014; 96-B:1578–85.


Bone & Joint 360
Vol. 2, Issue 1 | Pages 16 - 19
1 Feb 2013

The February 2013 Knee Roundup360 looks at: mobile-bearing TKRs; arthroscopic ACL reconstruction; the use of chondrocytes for osteochondral defects; ACL reconstruction and the return to pivoting sports; ACLs and the MOON study; the benefit of knee navigation; and trabecular metal.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1183 - 1191
1 Sep 2015
Briggs TWR Hanna SA Kayani B Tai S Pollock RC Cannon SR Blunn GW Carrington RWJ

The long term biological effects of wear products following total hip arthroplasty (THA) are unclear. However, the indications for THA are expanding, with increasingly younger patients undergoing the procedure.

This prospective, randomised study compared two groups of patients undergoing THA after being randomised to receive one of two different bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal (MoM) n = 23. We investigated the relationship between three variables: bearing surface (MoP vs MoM), whole blood levels of chromium (Cr) and cobalt (Co) and chromosomal aberrations in peripheral lymphocyte pre-operatively and at one, two and five years post-surgery.

Our results demonstrated significantly higher mean cobalt and chromium (Co and Cr) blood levels in the MoM group at all follow-up points following surgery (p < 0.01), but there were no significant differences in the chromosomal aberration indices between MoM and MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p = 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34) for translocations, breaks, loss and gain of chromosomes respectively. Regression analysis showed a strong linear relationship between Cr levels and the total chromosomal aberration indices in the MoM group (R2 = 0.90016), but this was not as strong for Co (R2 = 0.68991). In the MoP group, the analysis revealed a poor relationship between Cr levels and the total chromosomal aberration indices (R2 = 0.23908) but a slightly stronger relationship for Co (R2 = 0.64292). Across both groups, Spearman’s correlation detected no overall association between Co and Cr levels and each of the studied chromosomal aberrations. There remains no clear indication which THA bearing couple is the most biocompatible, especially in young active patients. While THA continues to be very successful at alleviating pain and restoring function, the long-term biological implications of the procedure still require further scrutiny.

Cite this article: Bone Joint J 2015;97-B:1183–91.