We made a prospective study of 58 consecutive Neer II total shoulder replacements in 49 rheumatoid patients. Cemented glenoid and press-fit
Fifteen patients with rheumatoid arthritis had 19 elbow arthroplasties with the Wadsworth type of surface replacement prosthesis. After a follow-up period averaging 30 months, 11 patients with 15 elbow arthroplasties were entirely satisfied with their freedom from pain and range of movement. Radiologically, however, the
Aims. We evaluated clinical and radiographic outcomes of total shoulder
arthroplasty (TSA) using the second-generation Trabecular Metal
(TM) Glenoid component. The first generation component was withdrawn
in 2005 after a series of failures were reported. Between 2009 and
2012, 40 consecutive patients with unilateral TSA using the second-generation
component were enrolled in this clinical study. The mean age of
the patients was 63.8 years (40 to 75) and the mean follow-up was
38 months (24 to 42). Methods. Patients were evaluated using the Constant score (CS), the American
Shoulder and Elbow Surgeons (ASES) score and routine radiographs. Results. Significant differences were found between the pre- and post-operative
CS (p = 0.003), ASES (p = 0.009) scores and CS subscores of pain
(p <
0.001), strength (p <
0.001) and mobility items (p <
0.05). No glenoid or
Failed internal fixation of a fracture of the
proximal humerus produces many challenges with limited surgical options.
The aim of this study was to evaluate the clinical outcomes after
the use of a reverse shoulder arthroplasty under these circumstances.
Between 2007 and 2012, 19 patients (15 women and four men, mean
age 66 years; 52 to 82) with failed internal fixation after a proximal
humeral fracture, underwent implant removal and reverse shoulder arthroplasty
(RSA). The mean follow-up was 36 months (25 to 60). The mean American
Shoulder and Elbow Score improved from 27.8 to 50.1 (p = 0.019).
The mean Simple Shoulder Test score improved from 0.7 to 3.2 (p
= 0.020), and the mean visual analogue scale for pain improved from
6.8 to 4.3 (p = 0.012). Mean forward flexion improved from 58.7°
to 101.1° (p <
0.001), mean abduction from 58.7° to 89.1° (p
= 0.012), mean external rotation from 10.7° to 23.1° (p = 0.043)
and mean internal rotation from buttocks to L4 (p = 0.034). A major
complication was recorded in five patients (26%) (one intra-operative
fracture, loosening of the
Determining and accurately restoring the flexion-extension
axis of the elbow is essential for functional recovery after total
elbow arthroplasty (TEA). We evaluated the effect of morphological
features of the elbow on variations of alignment of the components
at TEA. Morphological and positioning variables were compared by
systematic CT scans of 22 elbows in 21 patients after TEA. There were five men and 16 women, and the mean age was 63 years
(38 to 80). The mean follow-up was 22 months (11 to 44). The anterior offset and version of the
To prevent insufficiency of the triceps after
total elbow arthroplasty, we have, since 2008, used a triceps-sparing ulnar
approach. This study evaluates the clinical results and post-operative
alignment of the prosthesis using this approach. We reviewed 25 elbows in 23 patients. There were five men and
18 women with a mean age of 69 years (54 to 83). There were 18 elbows
with rheumatoid arthritis, six with a fracture or pseudoarthrosis
and one elbow with osteoarthritis. Post-operative complications included one intra-operative fracture,
one elbow with heterotopic ossification, one transient ulnar nerve
palsy, and one elbow with skin necrosis, but no elbow was affected
by insufficiency of the triceps. Patients were followed for a mean of 42 months (24 to 77). The
mean post-operative Japanese Orthopaedic Association Elbow Score
was 90.8 (51 to 100) and the mean Mayo Elbow Performance score 93.8
(65 to 100). The mean post-operative flexion/extension of the elbow
was 135°/-8°. The Manual Muscle Testing score of the triceps was
5 in 23 elbows and 2 in two elbows (one patient). The mean alignment
of the implants examined by 3D-CT was 2.8° pronation (standard deviation
(. sd). 5.5), 0.3° valgus (. sd. 2.7), and 0.7° extension
(. sd. 3.2) for the
Medium-term results of the Discovery elbow replacement
are presented. We reviewed 51 consecutive primary Discovery total
elbow replacements (TERs) implanted in 48 patients. The mean age
of the patients was 69.2 years (49 to 92), there were 19 males and
32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69).
A total of six patients were lost to follow-up. Statistically significant
improvements in range movement and Oxford Elbow Score were found
(p <
0.001). Radiolucent lines were much more common in, and
aseptic loosening was exclusive to, the
The February 2014 Shoulder &
Elbow Roundup. 360 . looks at: whether arthroscopic acromioplasty is a cost-effective intervention; shockwave therapy in cuff tear; whether microfracture relieves short-term pain in cuff repair; the promising early results from L-PRF augmented cuff repairs; rehabilitation following cuff repair; supination strength following biceps tendon rupture; whether longer is better in
Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary
Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey. NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.Aims
Methods
Metallosis after shoulder replacement has not previously been described in the literature. We report a patient who developed extensive metallosis after implantation of an uncemented Nottingham shoulder replacement. He underwent a revision procedure. Examination of the retrieved prosthesis showed that the titanium porous coating was separating from the humeral stem and becoming embedded in the ultra-high-molecular-weight polyethylene glenoid component, resulting in abrasive wear of the
The liner design is a key determinant of the constraint of a reverse total shoulder arthroplasty (rTSA). The aim of this study was to compare the degree of constraint of rTSA liners between different implant systems. An implant company’s independent 3D shoulder arthroplasty planning software (mediCAD 3D shoulder v. 7.0, module v. 2.1.84.173.43) was used to determine the jump height of standard and constrained liners of different sizes (radius of curvature) of all available companies. The obtained parameters were used to calculate the stability ratio (degree of constraint) and angle of coverage (degree of glenosphere coverage by liner) of the different systems. Measurements were independently performed by two raters, and intraclass correlation coefficients were calculated to perform a reliability analysis. Additionally, measurements were compared with parameters provided by the companies themselves, when available, to ensure validity of the software-derived measurements.Aims
Methods
The aim of this study is to report the implant survival and factors associated with revision of total elbow arthroplasty (TEA) using data from the Dutch national registry. All TEAs recorded in the Dutch national registry between 2014 and 2020 were included. The Kaplan-Meier method was used for survival analysis, and a logistic regression model was used to assess the factors associated with revision.Aims
Methods
The aims of this study were to identify and evaluate the current literature examining the prognostic factors which are associated with failure of total elbow arthroplasty (TEA). Electronic literature searches were conducted using MEDLINE, Embase, PubMed, and Cochrane. All studies reporting prognostic estimates for factors associated with the revision of a primary TEA were included. The risk of bias was assessed using the Quality In Prognosis Studies (QUIPS) tool, and the quality of evidence was assessed using the modified Grading of Recommendations, Assessment, Development, and Evaluations (GRADE) framework. Due to low quality of the evidence and the heterogeneous nature of the studies, a narrative synthesis was used.Aims
Methods
We undertook a radiological analysis of 186 standard Souter implants to determine survivorship and to analyse the pattern of failure in those needing revision. The implants had been inserted as a primary procedure in patients with rheumatoid arthritis of the elbow at our hospital over the last 12 years. Taking revision as an endpoint, the survivorship after 12 years was 87%. If, however, revision and loosening, defined as the Hindex value equivalent to demarcation of 1 mm around the whole implant, are also included, the survivorship falls to 80%. Of the 24 implants revised, 18 (75%) were for problems with the
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
The current evidence comparing the two most common approaches for reverse total shoulder arthroplasty (rTSA), the deltopectoral and anterosuperior approach, is limited. This study aims to compare the rate of loosening, instability, and implant survival between the two approaches for rTSA using data from the Dutch National Arthroplasty Registry with a minimum follow-up of five years. All patients in the registry who underwent a primary rTSA between January 2014 and December 2016 using an anterosuperior or deltopectoral approach were included, with a minimum follow-up of five years. Cox and logistic regression models were used to assess the association between the approach and the implant survival, instability, and glenoid loosening, independent of confounders.Aims
Methods
We have studied the three-dimensional geometry of the proximal humerus on human cadaver specimens using a digitised measuring device linked to a computer. Our findings demonstrated the variable shape of the proximal humerus as well as its variable dimensions. The articular surface, which is part of a sphere varies individually in its orientation as regards inclination and retroversion, and it has variable medial and posterior offsets. These variations cannot be accommodated by the designs of most contemporary
We report the results of a series of 59 unconstrained total arthroplasties of the elbow after a mean follow-up of 6.5 years (4 to 10). All the patients had rheumatoid arthritis. The indication for surgery was pain in all but one. Outcome was assessed by the Mayo Elbow Performance Index (MEPI). Of the 59 arthroplasties, two were lost to follow-up. Ten patients died, but as two of their arthroplasties were failures we included them in the results. The outcome in the remaining 49 was excellent in 26 (53%), good in 15 (31%), fair in one (2%) and poor or a failure in seven (14%). There was an improvement in the pain score (p <
0.001), movement (p <
0.001) and function (p <
0.001). Two patients developed instability, but neither required further surgery. There was a mean increase of 21° in flexion and of 7° in extension. The overall rate of complications was 33.9%. Lesions of the ulnar nerve, one of which did not recover, occurred in four patients, deep infection in two and stiffness in five. The rates of complications were similar to those in recent reports of other elbow arthroplasties. We carried out a radiological analysis of 39 arthroplasties which showed radiolucent lines around the
This study aimed to assess the impact of using the metal-augmented glenoid baseplate (AGB) on improving clinical and radiological outcomes, as well as reducing complications, in patients with superior glenoid wear undergoing reverse shoulder arthroplasty (RSA). From January 2016 to June 2021, out of 235 patients who underwent primary RSA, 24 received a superior-AGB after off-axis reaming (Group A). Subsequently, we conducted propensity score matching in a 1:3 ratio, considering sex, age, follow-up duration, and glenoid wear (superior-inclination and retroversion), and selected 72 well-balanced matched patients who received a standard glenoid baseplate (STB) after eccentric reaming (Group B). Superior-inclination, retroversion, and lateral humeral offset (LHO) were measured to assess preoperative glenoid wear and postoperative correction, as well as to identify any complications. Clinical outcomes were measured at each outpatient visit before and after surgery.Aims
Methods