The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
Massive endoprostheses rely on extra-cortical bone bridging (ECBB)
to enhance fixation. The aim of this study was to investigate the
role of selective laser sintered (SLS) porous collars in augmenting
the osseointegration of these prostheses. The two novel designs of porous SLS collars, one with small pores
(Ø700 μm, SP) and one with large pores (Ø1500 μm, LP), were compared
in an ovine tibial diaphyseal model. Osseointegration of these collars
was compared with that of a clinically used solid, grooved design
(G). At six months post-operatively, the ovine tibias were retrieved and
underwent radiological and histological analysis.Aims
Materials and Methods
Using a systematic review, we investigated whether there is an
increased risk of post-operative infection in patients who have
received an intra-articular corticosteroid injection to the hip
for osteoarthritis prior to total hip arthroplasty (THA). Studies dealing with an intra-articular corticosteroid injection
to the hip and infection following subsequent THA were identified
from databases for the period between 1990 to 2013. Retrieved articles
were independently assessed for their methodological quality.Aims
Methods
We explored the literature surrounding whether
allergy and hypersensitivity has a clinical basis for implant selection
in total knee arthroplasty (TKA). In error, the terms hypersensitivity
and allergy are often used synonymously. Although a relationship
is present, we could not find any evidence of implant failure due
to allergy. There is however increasing basic science that suggests
a link between loosening and metal ion production. This is not an
allergic response but is a potential problem. With a lack of evidence
logically there can be no justification to use ‘hypoallergenic’
implants in patients who have pre-existing skin sensitivity to the
metals used in TKA. Cite this article:
Knee arthrodesis is a potential salvage procedure
for limb preservation after failure of total knee arthroplasty (TKA) due
to infection. In this study, we evaluated the outcome of single-stage
knee arthrodesis using an intramedullary cemented coupled nail without
bone-on-bone fusion after failed and infected TKA with extensor
mechanism deficiency. Between 2002 and 2012, 27 patients (ten female,
17 male; mean age 68.8 years; 52 to 87) were treated with septic
single-stage exchange. Mean follow-up duration was 67.1months (24
to 143, n = 27) (minimum follow-up 24 months) and for patients with
a minimum follow-up of five years 104.9 (65 to 143,; n = 13). A
subjective patient evaluation (Short Form (SF)-36) was obtained,
in addition to the Visual Analogue Scale (VAS). The mean VAS score was
1.44 (SD 1.48). At final follow-up, four patients had recurrent
infections after arthrodesis (14.8%). Of these, three patients were
treated with a one-stage arthrodesis nail exchange; one of the three
patients had an aseptic loosening with a third single-stage exchange,
and one patient underwent knee amputation for uncontrolled sepsis at
108 months. All patients, including the amputee, indicated that
they would choose arthrodesis again. Data indicate that a single-stage
knee arthrodesis offers an acceptable salvage procedure after failed
and infected TKA. Cite this article:
Metaphyseal bone loss is common with revision
total knee replacement (RTKR). Using the Anderson Orthopaedic Research
Institute (AORI) classification, type 2-B and type 3 defects usually
require large metal blocks, bulk structural allograft or highly
porous metal cones. Tibial and femoral trabecular metal metaphyseal
cones are a unique solution for large bone defects. These cones
substitute for bone loss, improve metaphyseal fixation, help correct
malalignment, restore the joint line and may permit use of a shorter
stem. The technique for insertion involves sculpturing of the remaining
bone with a high speed burr and rasp, followed by press-fit of the
cone into the metaphysis. The fixation and osteoconductive properties
of the porous cone outer surface allow ingrowth and encourage long-term
biological fixation. The revision knee component is then cemented
into the porous cone inner surface, which provides superior fixation
compared with cementing into native but deficient metaphyseal bone.
The advantages of the cone compared with allograft include: technical
ease, biological fixation, no resorption, and possibly a lower risk
of infection. The disadvantages include: difficult extraction and
relatively short-term follow-up. Several studies using cones report
promising short-term results for the reconstruction of large bone
defects in RTKR. Cite this article:
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
Mortality rates reported by the National Joint Registry for England
and Wales (NJR) were higher following cemented total knee replacement
(TKR) compared with uncemented procedures. The aim of this study
is to examine and compare the effects of cemented and uncemented
TKR on the activation of selected markers of inflammation, endothelium,
and coagulation, and on the activation of selected cytokines involved
in the various aspects of the systemic response following surgery. This was a single centre, prospective, case-control study. Following
enrolment, blood samples were taken pre-operatively, and further
samples were collected at day one and day seven post-operatively.
One patient in the cemented group developed a deep-vein thrombosis
confirmed on ultrasonography and was excluded, leaving 19 patients
in this cohort (mean age 67.4, (Objective
Methods
Approved by the Food and Drug Administration in 2004, the Phase
III Oxford Medial Partial Knee is used to treat anteromedial osteoarthritis
(AMOA) in patients with an intact anterior cruciate ligament. This
unicompartmental knee arthroplasty (UKA) is relatively new in the
United States, and therefore long-term American results are lacking. This is a single surgeon, retrospective study based on prospectively
collected data, analysing a consecutive series of primary UKAs using
the Phase III mobile-bearing Oxford Knee and Phase III instrumentation. Between July 2004 and December 2006, the senior author (RHE)
carried out a medial UKA in 173 patients (213 knees) for anteromedial
osteoarthritis or avascular necrosis (AVN). A total of 95 patients were men and 78 were women. Their mean
age at surgery was 67 years (38 to 89) and mean body mass index
29.87 kg/m2 (17 to 62). The mean follow-up was ten years (4 to 11).Aims
Patients and Methods
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern uncemented acetabular components have
few screw holes, or no holes, polished inner surfaces, improved
locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented
sockets can be combined with highly cross-linked polyethylene liners,
which have demonstrated very low wear and osteolysis rates after
ten to 15 years of implantation. The results of cement fixation
with a smooth or polished surface finished stem have been excellent,
virtually eliminating complications seen with cementless fixation
like peri-operative femoral fractures and thigh pain. Although mid-term
results of modern cementless stems are encouraging, the long-term
data do not show reduced revision rates for cementless stems compared
with cemented smooth stems. In this paper we review the conduct
of a hybrid THA, with emphasis on pre-operative planning, surgical
technique, hypotensive epidural anaesthesia, and intra-operative
physiology. Cite this article:
The purpose of this study was to evaluate whether
the serum level of interleukin 6 (IL-6) could be used to identify the
persistence of infection after the first stage of a two-stage revision
for periprosthetic joint infection. Between 2010 and 2011, we prospectively studied 55 patients (23
men, 32 women; mean age 69.5 years; 36 to 86) with a periprosthetic
joint infection. Bacteria were identified in two intra-operative
tissue samples during re-implantation in 16 patients. These cases
were classified as representing persistent infection. To calculate a precise cut-off value which could be used in everyday
clinical practice, a 3 x 2 contingency table was constructed and
manually defined. We found that a serum IL-6 ≥ 13 pg/mL can be regarded as indicating
infection: its positive-predictive value is 90.9%. A serum IL-6 ≤ 8
pg/mL can be regarded as indicating an absence of infection: its
negative predictive value is 92.1%. The serum IL-6 level seems to be a reasonable marker for identifying
persistent infection after the first stage of a revision joint arthroplasty
and before attempting re-implantation. Cite this article:
The aims of this study were to identify the early
in-hospital mortality rate after hip fracture, identify factors associated
with this mortality, and identify the cause of death in these patients.
A retrospective cohort study was performed on 4426 patients admitted
to our institution between the 1 January 2006 and 31 December 2013
with a hip fracture (1128 male (26%), mean age 82.0 years (60 to
105)). Admissions increased annually, but despite this 30-day mortality
decreased from 12.1% to 6.5%; 77% of these were in-hospital deaths.
Male gender (odds ratio (OR) 2.0, 95% confidence interval (CI) 1.3
to 3.0), increasing age (age ≥ 91; OR 4.1, 95% CI 1.4 to 12.2) and
comorbidity (American Society of Anesthesiologists grades 3 to 5;
OR 4.2, 95% CI 2.0 to 8.7) were independently and significantly
associated with increased odds of in-hospital mortality. From 220
post-mortem reports, the most common causes of death were respiratory
infections (35%), ischaemic heart disease (21%), and cardiac failure
(13%). A sub-group of hip fracture patients at highest risk of early
death can be identified with these risk factors, and the knowledge
of the causes of death can be used to inform service improvements
and the development of a more didactic care pathway, so that multidisciplinary
intervention can be focused for this sub-group in order to improve
their outcome. Cite this article:
Total knee arthroplasty (TKA) is a cost effective
and extremely successful operation. As longevity increases, the demand
for primary TKA will continue to rise. The success and survivorship
of TKAs are dependent on the demographics of the patient, surgical
technique and implant-related factors. Currently the risk of failure of a TKA requiring revision surgery
ten years post-operatively is 5%. The most common indications for revision include aseptic loosening
(29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses
considerable clinical burdens on patients and financial burdens
on healthcare systems. We present a current concepts review on the epidemiology of failed
TKAs using data from worldwide National Joint Registries. Cite this article:
Most published randomised controlled trials which
compare the rates of wear of conventional and cross-linked (XL) polyethylene
(PE) in total hip arthroplasty (THA) have described their use with
a cementless acetabular component. We conducted a prospective randomised study to assess the rates
of penetration of two distinct types of PE in otherwise identical
cemented all-PE acetabular components. A total of 100 consecutive patients for THA were randomised to
receive an acetabular component which had been either highly XL
then remelted or moderately XL then annealed. After a minimum of eight years follow-up, 38 hips in the XL group
and 30 hips in the annealed group had complete data (mean follow-up
of 9.1 years (7.6 to 10.7) and 8.7 years (7.2 to 10.2), respectively).
In the XL group, the steady state rate of penetration from one year
onwards was -0.0002 mm/year ( These results show that the yearly linear rate of femoral head
penetration can be significantly reduced by using a highly XLPE
cemented acetabular component. Cite this article:
We report our experience of staged revision surgery
for the treatment of infected total elbow arthroplasty (TEA). Between
1998 and 2010 a consecutive series of 33 patients (34 TEAs) underwent
a first-stage procedure with the intention to proceed to second-stage
procedure when the infection had been controlled. A single first-stage procedure
with removal of the components and cement was undertaken for 29
TEAs (85%), followed by the insertion of antibiotic-impregnated
cement beads, and five (15%) required two or more first-stage procedures.
The most common organism isolated was coagulase-negative A second-stage procedure was performed for 26 TEAs (76%); seven
patients (seven TEAs, 21%) had a functional resection arthroplasty
with antibiotic beads There were three recurrent infections (11.5%) in those patients
who underwent a second-stage procedure. The infection presented
at a mean of eight months (5 to 10) post-operatively. The mean Mayo
Elbow Performance Score (MEPS) in those who underwent a second stage
revision without recurrent infection was 81.1 (65 to 95). Staged revision surgery is successful in the treatment of patients
with an infected TEA and is associated with a low rate of recurrent
infection. However, when infection does occur, this study would
suggest that it becomes apparent within ten months of the second
stage procedure. Cite this article:
We present the results of 62 consecutive acetabular
revisions using impaction bone grafting and a cemented polyethylene
acetabular component in 58 patients (13 men and 45 women) after
a mean follow-up of 27 years (25 to 30). All patients were prospectively
followed. The mean age at revision was 59.2 years (23 to 82). We performed Kaplan–Meier (KM) analysis and also a Competing
Risk (CR) analysis because with long-term follow-up, the presence
of a competing event (i.e. death) prevents the occurrence of the
endpoint of re-revision. A total of 48 patients (52 hips) had died or had been re-revised
at final review in March 2011. None of the deaths were related to
the surgery. The mean Harris hip score of the ten surviving hips
in ten patients was 76 points (45 to 99). The KM survivorship at 25 years for the endpoint ‘re-revision
for any reason’ was 58.0% (95% confidence interval (CI) 38 to 73)
and for ‘re-revision for aseptic loosening’ 72.1% (95% CI 51 to
85). With the CR analysis we calculated the KM analysis overestimates
the failure rate with respectively 74% and 93% for these endpoints.
The current study shows that acetabular impaction bone grafting
revisions provide good clinical results at over 25 years. Cite this article:
The treatment of infected nonunions is difficult.
Antibiotic cement-coated (ACC) rods provide stability as well as delivering
antibiotics. We conducted a review of 110 infected nonunions treated
with ACC rods. Patients were divided into two groups: group A (67
patients) with an infected arthrodesis, and group B (43 patients)
with an infected nonunion in a long bone. In group A, infected arthrodesis,
the success rate after the first procedure was 38/67 (57%), 29/67
(43%) required further surgery for either control of infection or
non-union. At last follow-up, five patients required amputation,
representing a limb salvage rate of 62/67 (93%) overall. In all,
29/67 (43%) presented with a bone defect with a mean size of 6.78
cm (2 to 25). Of those with a bone defect, 13/29 (45%) required
further surgery and had a mean size of defect of 7.2 cm (3.5 to
25). The cultures were negative in 17/67 (26%) and the most common
organism cultured was methicillin-resistant Cite this article:
The relationship between post-operative bone
density and subsequent failure of total knee replacement (TKR) is
not known. This retrospective study aimed to determine the relationship
between bone density and failure, both overall and according to
failure mechanism. All 54 aseptic failures occurring in 50 patients
from 7760 consecutive primary cemented TKRs between 1983 and 2004
were matched with non-failing TKRs, and 47 failures in 44 patients
involved tibial failures with the matching characteristics of age
(65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis
(93.6% OA), date of operation, bilaterality, pre-operative alignment
(0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively).
In each case, the density of bone beneath the tibial component was assessed
at each follow-up interval using standardised, calibrated radiographs.
Failing knees were compared with controls both overall and, as a
subgroup analysis, by failure mechanism. Knees were compared with
controls using univariable linear regression. Significant and continuous elevation in tibial density was found
in knees that eventually failed by medial collapse (p <
0.001)
and progressive radiolucency (p <
0.001) compared with controls,
particularly in the medial region of the tibia. Knees failing due
to ligamentous instability demonstrated an initial decline in density
(p = 0.0152) followed by a non-decreasing density over time (p =
0.034 for equivalence). Non-failing knees reported a decline in
density similar to that reported previously using dual-energy x-ray
absorptiometry (DEXA). Differences between failing and non-failing
knees were observable as early as two months following surgery.
This tool may be used to identify patients at risk of failure following