Structural allografts may be used to manage uncontained
bone defects in revision total knee replacement (TKR). However,
the availability of cadaver grafts is limited in some areas of Asia.
The aim of this study was to evaluate the mid-term outcome of the
use of femoral head allografts for the reconstruction of uncontained
defects in revision TKR, focusing on complications related to the
graft. We retrospectively reviewed 28 patients (30 TKRs) with Anderson
Orthopaedic Research Institute (AORI) type 3 bone defects, who underwent
revision using femoral head allografts and stemmed components. The
mean number of femoral heads used was 1.7 (1 to 3). The allograft–host
junctions were packed with cancellous autograft. At a mean follow-up of 76 months (38 to 136) the mean American
Knee Society knee score improved from 37.2 (17 to 60) pre-operatively
to 90 (83 to 100) (p <
0.001). The mean function score improved
from 26.5 (0 to 50) pre-operatively to 81 (60 to 100) (p <
0.001).
All the grafts healed to the host bone. The mean time to healing
of the graft was 6.6 months
(4 to 16). There have been no complications of collapse of the graft,
nonunion, infection or implant loosening. No revision surgery was
required. The use of femoral head allografts in conjunction with a stemmed
component and autogenous bone graft in revision TKR in patients
with uncontained bone defects resulted in a high rate of healing
of the graft with minimal complications and a satisfactory outcome.
Longer follow-up is needed to observe the evolution of the graft. Cite this article:
We investigated whether an asymmetric extension
gap seen on routine post-operative radiographs after primary total
knee replacement (TKR) is associated with pain at three, six, 12
and 24 months’ follow-up. On radiographs of 277 patients after primary
TKR we measured the distance between the tibial tray and the femoral
condyle on both the medial and lateral sides. A difference was defined
as an asymmetric extension gap. We considered three groups (no asymmetric
gap, medial-opening and lateral-opening gap) and calculated the
associations with the Western Ontario and McMaster Universities
osteoarthritis index pain scores over time. Those with an asymmetric extension gap of ≥ 1.5 mm had a significant
association with pain scores at three months’ follow-up; patients
with a medial-opening extension gap reported more pain and patients
with a lateral-opening extension gap reported less pain (p = 0.036).
This effect was still significant at six months (p = 0.044), but had
lost significance by 12 months (p = 0.924). When adjusting for multiple
cofounders the improvement in pain was more pronounced in patients
with a lateral-opening extension gap than in those with a medial-opening extension
gap at three (p = 0.037) and six months’ (p = 0.027) follow-up. Cite this article:
Ancient Egypt was a highly developed agrarian society with a massive civil engineering capability. Trauma and skeletal disease were common and vestiges of the evidence for that survive, largely in the form of hieratic images and papyri dedicated to the practice of medicine. The earliest treatise on trauma is the
Ensuring correct rotation of the femoral component
is a challenging aspect of patellofemoral replacement surgery. Rotation
equal to the epicondylar axis or marginally more external rotation
is acceptable. Internal rotation is associated with poor outcomes.
This paper comprises two studies evaluating the use of the medial
malleolus as a landmark to guide rotation. We used 100 lower-leg anteroposterior radiographs to evaluate
the reliability of the medial malleolus as a landmark. Assessment
was made of the angle between the tibial shaft and a line from the
intramedullary rod entry site to the medial malleolus. The femoral
cut was made in ten cadaver knees using the inferior tip of the
medial malleolus as a landmark for rotation. Rotation of the cut
relative to the anatomical epicondylar axis was assessed using CT.
The study of radiographs found the position of the medial malleolus
relative to the tibial axis is consistent. Using the inferior tip
of the medial malleolus in the cadaver study produced a mean external
rotation of 1.6° (0.1° to 3.7°) from the anatomical epicondylar
axis. Using the inferior tip of the medial malleolus to guide the
femoral cutting jig avoids internal rotation and introduces an acceptable
amount of external rotation of the femoral component.
The October 2012 Wrist &
Hand Roundup360 looks at: osteoarticular flaps to the PIPJ; prognosis after wrist arthroscopy; adipofascial flaps and post-traumatic adhesions; the torn TFCC alone; ulna-shortening osteotomy for ulnar impaction syndrome; Dupuytren’s disease; when a wrist sprain is not a sprain; and shrinking the torn intercarpal ligament.
The management of bone loss in revision replacement of the knee remains a challenge despite an array of options available to the surgeon. Bone loss may occur as a result of the original disease, the design of the prosthesis, the mechanism of failure or technical error at initial surgery. The aim of revision surgery is to relieve pain and improve function while addressing the mechanism of failure in order to reconstruct a stable platform with transfer of load to the host bone. Methods of reconstruction include the use of cement, modular metal augmentation of prostheses, custom-made, tumour-type or hinged implants and bone grafting. The published results of the surgical techniques are summarised and a guide for the management of bone defects in revision surgery of the knee is presented.
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite coating, has improved the outcomes of uncemented
TKA over time.
We investigated the clinical and radiological outcome of trochleaplasty for recurrent patellar dislocation in association with trochlear dysplasia in 38 consecutive patients (45 knees) with a mean follow-up of 8.3 years (4 to 14). None had recurrence of dislocation after trochleaplasty. Post-operatively, patellofemoral pain, present pre-operatively in only 35 knees, became worse in 15 (33.4%), remained unchanged in four (8.8%) and improved in 22 (49%). Four knees which had no pain pre-operatively (8.8%) continued to have no pain. A total of 33 knees were available for radiological assessment. Post-operatively, all but two knees (93.9%) had correction of trochlear dysplasia radiologically but degenerative changes of the patellofemoral joint developed in 30% (10) of the knees. We conclude that recurrent patellar dislocation associated with trochlear dysplasia can be treated successfully by trochleaplasty, but the impact on patellofemoral pain and the development of patellofemoral osteoarthritis is less predictable. Overall, subjective patient satisfaction with restored patellofemoral stability after trochleaplasty appeared to outweigh its possible sequelae.
We reviewed nine patients at a mean period of 11 years (6 to 16) after curettage and cementing of a giant-cell tumour around the knee to determine if there were any long-term adverse effects on the cartilage. Plain radiography, MRI, delayed gadolinium-enhanced MRI of the cartilage and measurement of the serum level of cartilage oligomeric matrix protein were carried out. The functional outcome was evaluated using the Lysholm knee score. Each patient was physically active and had returned to their previous occupation. Most participated in recreational sports or exercise. The mean Lysholm knee score was 92 (83 to 100). Only one patient was found to have cartilage damage adjacent to the cement. This patient had a history of intra-articular fracture and local recurrence, leading to degenerative changes. Interpretation of the data obtained from delayed gadolinium-enhanced MRI of the cartilage was difficult, with variation in the T1 values which did not correlate with the clinical or radiological findings. We did not find it helpful in the early diagnosis of degeneration of cartilage. We also found no obvious correlation between the serum cartilage oligomeric matrix protein level and the radiological and MR findings, function, time after surgery and the age of the patient. In summary, we found no evidence that the long-term presence of cement close to the knee joint was associated with the development of degenerative osteoarthritis.
A variety of materials were used for early hip prostheses. The introduction of Vitallium by Smith-Petersen represented a further advance in this surgical field. We present the longest known follow-up of a Smith-Petersen Vitallium mould arthroplasty.
Matrix-assisted autologous chondrocyte transplantation (MACT)
has been developed and applied in the clinical practice in the last
decade to overcome most of the disadvantages of the first generation
procedures. The purpose of this systematic review is to document
and analyse the available literature on the results of MACT in the
treatment of chondral and osteochondral lesions of the knee. All studies published in English addressing MACT procedures were
identified, including those that fulfilled the following criteria:
1) level I-IV evidence, 2) measures of functional or clinical outcome,
3) outcome related to cartilage lesions of the knee cartilage.Objectives
Methods
We studied 11 patients (14 elbows) with gross rheumatoid deformity of the elbow, treated by total arthroplasty using the Kudo type-5 unlinked prosthesis, and who were evaluated between five and 11 years after operation. Massive bone defects were augmented by autogenous bone grafts. There were no major complications such as infection, subluxation or loosening. In most elbows relief from pain and stability were achieved. The results, according to the Mayo Elbow Performance Score, were excellent in eight, good in five and fair in one. In most elbows there was minimal or no resorption of the grafted bone. There were no radiolucent lines around the stems of the cementless components. This study shows that even highly unstable rheumatoid elbows can be replaced successfully using an unlinked prosthesis, with augmentation by grafting for major defects of bone.
Seven stiff total knee arthroplasties are presented
to illustrate the roles of: 1) manipulation under general anesthesia;
2) multiple concurrent diagnoses in addition to stiffness; 3) extra-articular
pathology; 4) pain as part of the stiffness triad (pain and limits
to flexion or extension); 5) component internal rotation; 6) multifactorial
etiology; and 7) surgical exposure in this challenging clinical
problem.
Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score.Objectives
Methods
Numerous complications following total knee replacement (TKR)
relate to the patellofemoral (PF) joint, including pain and patellar
maltracking, yet the options for A total of three knees with end-stage osteoarthritis and three
knees that had undergone TKR at more than one year’s follow-up were
investigated. In each knee, sequential biplane radiological images
were acquired from the sagittal direction (i.e. horizontal X-ray
source and 10° below horizontal) for a sequence of eight flexion
angles. Three-dimensional implant or bone models were matched to
the biplane images to compute the six degrees of freedom of PF tracking
and TF kinematics, and other clinical measures.Objectives
Methods
Elastic stable intramedullary nailing (ESIN)
is generally acknowledged to be the treatment of choice for displaced diaphyseal
femoral fractures in children over the age of three years, although
complication rates of up to 50% are described. Pre-bending the nails
is recommended, but there are no published data to support this.
Using synthetic bones and a standardised simulated fracture, we
performed biomechanical testing to determine the influence on the
stability of the fracture of pre-bending the nails before implantation.
Standard ESIN was performed on 24 synthetic femoral models with
a spiral fracture. In eight cases the nails were inserted without
any pre-bending, in a further eight cases they were pre-bent to
30° and in the last group of eight cases they were pre-bent to 60°. Mechanical
testing revealed that pre-bending to 60° produced a significant
increase in the stiffness or stability of the fracture. Pre-bending
to 60° showed a significant positive influence on the stiffness
compared with unbent nails. Pre-bending to 30° improved stiffness
only slightly. These findings validate the recommendations for pre-bending,
but the degree of pre-bend should exceed 30°. Adopting higher degrees
of pre-bending should improve stability in spiral fractures and
reduce the complications of varus deformity and shortening.
It has been suggested that an increased posterior
tibial slope (PTS) and a narrow notch width index (NWI) increase
the risk of anterior cruciate ligament (ACL) injury. The aim of
this study was to establish why there are conflicting reports on
their significance. A total of fifty patients with a ruptured ACL
and 50 patients with an intact ACL were included in the study. The
group with ACL rupture had a statistically significantly increased
PTS (p <
0.001) and a smaller NWI (p <
0.001) than the control
group. When a high PTS and/or a narrow NWI were defined as risk
factors for an ACL rupture, 80% of patients had at least one risk
factor present; only 24% had both factors present. In both groups
the PTS was negatively correlated to the NWI (correlation coefficient
= -0.28, p = 0.0052). Using a univariate model, PTS and NWI appear
to be correlated to rupture of the ACL. Using a logistic regression
model, the PTS (p = 0.006) and the NWI (p <
0.0001) remain significant
risk factors. From these results, either a steep PTS or a narrow
NWI predisposes an individual to ACL injury. Future studies should
consider these factors in combination rather than in isolation.
Despite many claims of good wear properties following
total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral
component, there are conflicting clinical results. We hypothesised
that there would be no difference in either the mid-term clinical
and radiological outcomes or the characteristics of the polyethylene
wear particles (weight, size and shape) in patients using an OxZr
or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent
bilateral TKR, receiving an OxZr femoral component in one knee and
a CoCr femoral component in the other. The mean follow-up was 7.5
years (6 to 8). Following aspiration, polyethylene wear particles
were analysed using thermogravimetric methods and scanning electron
microscopy. At the most recent follow-up, the mean Knee Society
score, Western Ontario and McMaster Universities Osteoarthritis
Index score, range of movement and satisfaction score were not significantly
different in the two groups. The mean weight, size, aspect ratio
and roundness of the aspirated wear particles were similar for each
femoral component. Survivorship of the femoral, tibial and patellar
components was 100% in both groups. In the absence of evidence of an advantage in the medium term
we cannot justify the additional expense of an OxZr femoral component.