The purpose of this study was to evaluate whether concerns about the release of
Aims. Significant correction of an adolescent idiopathic scoliosis in the coronal plane through a posterior approach is associated with hypokyphosis. Factors such as the magnitude of the preoperative coronal curve, the use of hooks, number of levels fused, preoperative kyphosis, screw density, and rod type have all been implicated. Maintaining the normal thoracic kyphosis is important as hypokyphosis is associated with proximal junctional failure (PJF) and early onset degeneration of the spine. The aim of this study was to determine if coronal correction per se was the most relevant factor in generating hypokyphosis. Methods. A total of 95 patients (87% female) with a median age of 14 years were included in our study. Pre- and postoperative radiographs were measured and the operative data including upper instrumented vertebra (UIV), lower instrumented vertebra (LIV),
Aims. To investigate metallosis in patients with magnetically controlled growing rods (MCGRs) and characterize the
Aims. We present a case series of five patients who had revision surgery
following magnetic controlled growing rods (MGCR) for early onset
scoliosis. Metallosis was found during revision in four out of five
patients and we postulated a mechanism for rod failure based on
retrieval analysis. Patients and Methods. Retrieval analysis was performed on the seven explanted rods.
The mean duration of MCGR from implantation to revision was 35 months
(17 to 46). The mean age at revision was 12 years (7 to 15; four
boys, one girl). Results. A total of six out of seven rods had tissue metallosis and pseudo-capsule
surrounding the actuator. A total of four out of seven rods were
pistoning. There were two rods which were broken. All rods had abrasive
circumferential markings. A significant amount of
The aim of this study was to use diffusion tensor imaging (DTI) to investigate changes in diffusion metrics in patients with cervical spondylotic myelopathy (CSM) up to five years after decompressive surgery. We correlated these changes with clinical outcomes as scored by the Modified Japanese Orthopedic Association (mJOA) method, Neck Disability Index (NDI), and Visual Analogue Scale (VAS). We used multi-shot, high-resolution, diffusion tensor imaging (ms-DTI) in patients with cervical spondylotic myelopathy (CSM) to investigate the change in diffusion metrics and clinical outcomes up to five years after anterior cervical interbody discectomy and fusion (ACDF). High signal intensity was identified on T2-weighted imaging, along with DTI metrics such as fractional anisotropy (FA). MJOA, NDI, and VAS scores were also collected and compared at each follow-up point. Spearman correlations identified correspondence between FA and clinical outcome scores.Aims
Methods
Magnetically controlled growing rods (MCGR) have been gaining popularity in the management of early-onset scoliosis (EOS) over the past decade. We present our experience with the first 44 MCGR consecutive cases treated at our institution. This is a retrospective review of consecutive cases of MCGR performed in our institution between 2012 and 2018. This cohort consisted of 44 children (25 females and 19 males), with a mean age of 7.9 years (3.7 to 13.6). There were 41 primary cases and three revisions from other rod systems. The majority (38 children) had dual rods. The group represents a mixed aetiology including idiopathic (20), neuromuscular (13), syndromic (9), and congenital (2). The mean follow-up was 4.1 years, with a minimum of two years. Nine children graduated to definitive fusion. We evaluated radiological parameters of deformity correction (Cobb angle), and spinal growth (T1-T12 and T1-S1 heights), as well as complications during the course of treatment.Aims
Methods
With the identification of literature shortfalls on the techniques employed in intraoperative navigated (ION) spinal surgery, we outline a number of measures which have been synthesised into a coherent operative technique. These include positioning, dissection, management of the reference frame, the grip, the angle of attack, the drill, the template, the pedicle screw, the wire, and navigated intrathecal analgesia. Optimizing techniques to improve accuracy allow an overall reduction of the repetition of the surgical steps with its associated productivity benefits including time, cost, radiation, and safety. Cite this article:
We aimed to retrospectively assess the accuracy and safety of
CT navigated pedicle screws and to compare accuracy in the cervical
and thoracic spine (C2-T8) with (COMB) and without (POST) prior
anterior surgery (anterior cervical discectomy or corpectomy and
fusion with ventral plating: ACDF/ACCF). A total of 592 pedicle screws, which were used in 107 consecutively
operated patients (210 COMB, 382 POST), were analysed. The accuracy
of positioning was determined according to the classification of
Gertzbein and Robbins on post-operative CT scans.Aims
Patients and Methods
Magnetically controlled growing rod (MCGR) systems use non-invasive
spinal lengthening for the surgical treatment of early-onset scoliosis
(EOS). The primary aim of this study was to evaluate the performance
of these devices in the prevention of progression of the deformity.
A secondary aim was to record the rate of complications. An observational study of 31 consecutive children with EOS, of
whom 15 were male, who were treated between December 2011 and October
2017 was undertaken. Their mean age was 7.7 years (2 to 14). The
mean follow-up was 47 months (24 to 69). Distractions were completed
using the tailgating technique. The primary outcome measure was
correction of the radiographic deformity. Secondary outcomes were
growth, functional outcomes and complication rates.Aims
Patients and Methods
The primary aim of this study was to evaluate the performance
and safety of magnetically controlled growth rods in the treatment
of early onset scoliosis. Secondary aims were to evaluate the clinical
outcome, the rate of further surgery, the rate of complications,
and the durability of correction. We undertook an observational prospective cohort study of children
with early onset scoliosis, who were recruited over a one-year period
and followed up for a minimum of two years. Magnetically controlled
rods were introduced in a standardized manner with distractions
performed three-monthly thereafter. Adverse events which were both related
and unrelated to the device were recorded. Ten children, for whom
relevant key data points (such as demographic information, growth
parameters, Cobb angles, and functional outcomes) were available,
were recruited and followed up over the period of the study. There
were five boys and five girls. Their mean age was 6.2 years (2.5
to 10).Aims
Patients and Methods
The aim of this study was to determine whether the sequential
application of povidone iodine-alcohol (PVI) followed by chlorhexidine
gluconate-alcohol (CHG) would reduce surgical wound contamination
to a greater extent than PVI applied twice in patients undergoing
spinal surgery. A single-centre, interventional, two arm, parallel group randomised
controlled trial was undertaken, involving 407 patients who underwent
elective spinal surgery. For 203 patients, the skin was disinfected before surgery using
PVI (10% [w/w (1% w/w available iodine)] in 95% industrial denatured
alcohol, povidone iodine; Videne Alcoholic Tincture) twice, and
for 204 patients using PVI once followed by CHG (2% [w/v] chlorhexidine
gluconate in 70% [v/v] isopropyl alcohol; Chloraprep with tint).
The primary outcome measure was contamination of the wound determined
by aerobic and anaerobic bacterial growth from samples taken after
disinfection.Aims
Patients and Methods
In order to evaluate the effectiveness of the Mobi-C implant
in cervical disc degeneration, a randomised study was conducted,
comparing the Mobi-C prosthesis arthroplasty with anterior cervical
disc fusion (ACDF) in patients with single level cervical spondylosis. From January 2008 to July 2009, 99 patients were enrolled and
randomly divided into two groups, those having a Mobi-C implant
(n = 51; 30 men, 21 women) and those undergoing ACDF (n = 48; 28
men, 20 women).The patients were followed up for five years, with
the primary outcomes being the Japanese Orthopaedic Association
score, visual analogue scale for pain and the incidence of further
surgery. The secondary outcomes were the Neck Disability Index and
range of movement (ROM) of the treated segment.Aims
Patients and Methods
Wrong-level surgery is a unique pitfall in spinal
surgery and is part of the wider field of wrong-site surgery. Wrong-site
surgery affects both patients and surgeons and has received much
media attention. We performed this systematic review to determine
the incidence and prevalence of wrong-level procedures in spinal
surgery and to identify effective prevention strategies. We retrieved
12 studies reporting the incidence or prevalence of wrong-site surgery
and that provided information about prevention strategies. Of these,
ten studies were performed on patients undergoing lumbar spine surgery
and two on patients undergoing lumbar, thoracic or cervical spine procedures.
A higher frequency of wrong-level surgery in lumbar procedures than
in cervical procedures was found. Only one study assessed preventative
strategies for wrong-site surgery, demonstrating that current site-verification protocols
did not prevent about one-third of the cases. The current literature
does not provide a definitive estimate of the occurrence of wrong-site
spinal surgery, and there is no published evidence to support the
effectiveness of site-verification protocols. Further prevention
strategies need to be developed to reduce the risk of wrong-site surgery.
The aim of this study was to evaluate the feasibility
of using the intact S1 nerve root as a donor nerve to repair an avulsion
of the contralateral lumbosacral plexus. Two cohorts of patients
were recruited. In cohort 1, the L4–S4 nerve roots of 15 patients
with a unilateral fracture of the sacrum and sacral nerve injury
were stimulated during surgery to establish the precise functional
distribution of the S1 nerve root and its proportional contribution
to individual muscles. In cohort 2, the contralateral uninjured
S1 nerve root of six patients with a unilateral lumbosacral plexus
avulsion was transected extradurally and used with a 25 cm segment
of the common peroneal nerve from the injured leg to reconstruct
the avulsed plexus. The results from cohort 1 showed that the innervation of S1 in
each muscle can be compensated for by L4, L5, S2 and S3. Numbness
in the toes and a reduction in strength were found after surgery
in cohort 2, but these symptoms gradually disappeared and strength
recovered. The results of electrophysiological studies of the donor
limb were generally normal. Severing the S1 nerve root does not appear to damage the healthy
limb as far as clinical assessment and electrophysiological testing
can determine. Consequently, the S1 nerve can be considered to be
a suitable donor nerve for reconstruction of an avulsed contralateral
lumbosacral plexus. Cite this article:
Adolescent idiopathic scoliosis affects about
3% of children. Non-operative measures are aimed at altering the natural
history to maintain the size of the curve below 40° at skeletal
maturity. The application of braces to treat spinal deformity pre-dates
the era of evidence-based medicine, and there is a paucity of irrefutable
prospective evidence in the literature to support their use and
their effectiveness has been questioned. This review considers this evidence. The weight of the evidence
is in favour of bracing over observation. The most recent literature
has moved away from addressing this question, and instead focuses
on developments in the design of braces and ways to improve compliance. Cite this article:
Posterior lumbar interbody fusion (PLIF) is indicated
for many patients with pain and/or instability of the lumbar spine.
We performed 36 PLIF procedures using the patient’s lumbar spinous
process and laminae, which were inserted as a bone graft between
two vertebral bodies without using a cage. The mean lumbar lordosis
and mean disc height to vertebral body ratio were restored and preserved
after surgery. There were no serious complications. These results suggest that this procedure is safe and effective.
Osteoporotic vertebral compression fractures
(VCFs) are an increasing public health problem. Recently, randomised
controlled trials on the use of kyphoplasty and vertebroplasty in
the treatment of these fractures have been published, but no definitive conclusions
have been reached on the role of these interventions. The major
problem encountered when trying to perform a meta-analysis of the
available studies for the use of cementoplasty in patients with
a VCF is that conservative management has not been standardised.
Forms of conservative treatment commonly used in these patients
include bed rest, analgesic medication, physiotherapy and bracing. In this review, we report the best evidence available on the
conservative care of patients with osteoporotic VCFs and associated
back pain, focusing on the role of the most commonly used spinal
orthoses. Although orthoses are used for the management of these patients,
to date, there has been only one randomised controlled trial published
evaluating their value. Until the best conservative management for
patients with VCFs is defined and standardised, no conclusions can
be drawn on the superiority or otherwise of cementoplasty techniques
over conservative management.
We reviewed 15 consecutive patients, 11 women and four men, with a mean age of 48.7 years (37.3 to 62.6), who between July 2004 and August 2007 had undergone percutaneous sacroiliac fusion using hollow modular anchorage screws filled with demineralised bone matrix. Each patient was carefully assessed to exclude other conditions and underwent pre-operative CT and MR scans. The diagnosis of symptomatic sacroiliac disease was confirmed by an injection of local anaesthetic and steroid under image intensifier control. The short form-36 questionnaire and Majeed’s scoring system were used for pre- and post-operative functional evaluation. Post-operative radiological evaluation was performed using plain radiographs. Intra-operative blood loss was minimal and there were no post-operative clinical or radiological complications. The mean follow-up was for 17 months (9 to 39). The mean short form-36 scores improved from 37 (23 to 51) to 80 (67 to 92) for physical function and from 53 (34 to 73) to 86 (70 to 98) for general health (p = 0.037). The mean Majeed’s score improved from 37 (18 to 54) pre-operatively to 79 (63 to 96) post-operatively (p = 0.014). There were 13 good to excellent results. The remaining two patients improved in short form-36 from a mean of 29 (26 to 35) to 48 (44 to 52). Their persistent pain was probably due to concurrent lumbar pathology. We conclude that percutaneous hollow modular anchorage screws are a satisfactory method of achieving sacroiliac fusion.
Inflammatory markers such as the C-reactive protein (CRP), white blood cell count and body temperature are easy to measure and are used as indicators of infection. The way in which they change in the early post-operative period after instrumented spinal surgery has not been reported in any depth. We measured these markers pre-operatively and at one, four, seven and 14 days postoperatively in 143 patients who had undergone an instrumented posterior lumbar interbody fusion. The CRP proved to be the only sensitive marker and had returned to its normal level in 48% of patients after 14 days. The CRP on day 7 was never higher than that on day 4. Age, gender, body temperature, operating time and blood loss were not related to the CRP level. A high CRP does not in itself suggest infection, but any increase after four days may presage infection.
Between January 1990 and December 2000 we carried out 226 SB Charité III disc replacements for lumbar disc degeneration in 160 patients. They were reviewed at a mean follow-up of 79 months (31 to 161) to determine the clinical and radiological outcome. The clinical results were collected by an independent observer, who was not involved in patient selection, treatment or follow-up, using a combination of outcome measures, including the Oswestry Disability Index. Pain was recorded using a visual analogue score, and the most recent radiographs were reviewed. Survival of the device was analysed by the Kaplan-Meier method and showed a cumulative survival of 35% at 156 months when radiological failure was taken as the endpoint. The mean improvement in the Oswestry disability index scores after disc replacement was 14% (6% to 21%) and the mean improvement in the pain score was 1.6 (0.46 to 2.73), both falling below the clinically significant threshold. Removal of the implant was required in 12 patients, four because of implant failure. These poor results indicate that further use of this implant is not justified.