The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance. A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.Aims
Methods
There are concerns regarding complications and longevity of total elbow arthroplasty (TEA) in young patients, and the few previous publications are mainly limited to reports on linked elbow devices. We investigated the clinical outcome of unlinked TEA for patients aged less than 50 years with rheumatoid arthritis (RA). We retrospectively reviewed the records of 26 elbows of 21 patients with RA who were aged less than 50 years who underwent primary TEA with an unlinked elbow prosthesis. The mean patient age was 46 years (35 to 49), and the mean follow-up period was 13.6 years (6 to 27). Outcome measures included pain, range of motion, Mayo Elbow Performance Score (MEPS), radiological evaluation for radiolucent line and loosening, complications, and revision surgery with or without implant removal.Aims
Methods
Aims. Reverse total shoulder arthroplasty (RTSA) using trabecular metal (TM)-backed glenoid implants has been introduced with the aim to increase implant survival. Only short-term reports on the outcomes of TM-RTSA have been published to date. We aim to present the seven-year survival of TM-backed glenoid implants along with minimum five-year clinical and radiological outcomes. Methods. All consecutive elective RTSAs performed at a single centre between November 2008 and October 2014 were reviewed. Patients who had primary TM-RTSA for rotator cuff arthropathy and osteoarthritis with deficient cuff were included. A total of 190 shoulders in 168 patients (41 male, 127 female) were identified for inclusion at a mean of 7.27 years (SD 1.4) from surgery. The primary outcome was survival of the implant with all-cause revision and aseptic glenoid loosening as endpoints. Secondary outcomes were clinical, radiological, and patient-related outcomes with a five-year minimum follow-up. Results. The implant was revised in ten shoulders (5.2%) with a median time to revision of 21.2 months (interquartile range (IQR) 9.9 to 41.8). The Kaplan-Meier survivorship estimate at seven years was 95.9% (95% confidence interval (CI) 91.7 to 98; 35 RTSAs at risk) for aseptic
Aims. Shoulder arthroplasty using short humeral components is becoming increasingly popular. Some such components have been associated with relatively high rates of adverse radiological findings. The aim of this retrospective review was to evaluate the radiological humeral bone changes and
The aim of this study was to report the outcomes of different treatment options for glenoid loosening following reverse shoulder arthroplasty (RSA) at a minimum follow-up of two years. We retrospectively studied the records of 79 patients (19 men, 60 women; 84 shoulders) aged 70.4 years (21 to 87) treated for aseptic loosening of the glenosphere following RSA. Clinical evaluation included pre- and post-treatment active anterior elevation (AAE), external rotation, and Constant score.Aims
Patients and Methods
Aim. We present the medium-term clinical results of a reverse total
shoulder arthroplasty with a trabecular metal glenoid base plate. Patients and Methods. We reviewed 125 consecutive primary reverse total shoulder arthroplasties
(RTSA) implanted in 124 patients for rotator cuff arthropathy. There
were 100 women and 24 men in the study group with a mean age of
76 years (58 to 89). The mean follow-up was 32 months (24 to 60).
No patient was lost to follow-up. Results. There were statistically significant improvements in the mean
range of movement and Oxford Shoulder Score (p <
0.001). Kaplan-Meier
survivorship at five years was 96.7% (95% confidence interval 91.5
to 98.7) with aseptic glenoid failure as the end point. . Radiologically, 63 shoulders (50.4%) showed no evidence of notching,
51 (40.8%) had grade 1 notching, ten (8.0%) had grade 2 notching
and one (0.8%) had grade 4 notching. Radiolucency around the glenoid
base plate was found in one patient (0.8%) and around the humeral
stem in five (4.0%). In all, three RTSA (2.4%) underwent revision surgery
for aseptic
The LockDown device (previously called Surgilig)
is a braided polyester mesh which is mostly used to reconstruct the
dislocated acromioclavicular joint. More than 11 000 have been implanted
worldwide. Little is known about the tissue reaction to the device
nor to its wear products when implanted in an extra-articular site
in humans. This is of importance as an adverse immunological reaction
could result in osteolysis or damage to the local tissues, thereby affecting
the longevity of the implant. We analysed the histology of five LockDown implants retrieved
from five patients over the last seven years by one of the senior
authors. Routine analysis was carried out in all five cases and
immunohistochemistry in one. The LockDown device acts as a scaffold for connective tissue
which forms an investing fibrous pseudoligament. The immunological
response at the histological level seems favourable with a limited
histiocytic and giant cell response to micron-sized wear particles.
The connective tissue envelope around the implant is less organised
than a native ligament. Cite this article:
Medium-term results of the Discovery elbow replacement
are presented. We reviewed 51 consecutive primary Discovery total
elbow replacements (TERs) implanted in 48 patients. The mean age
of the patients was 69.2 years (49 to 92), there were 19 males and
32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69).
A total of six patients were lost to follow-up. Statistically significant
improvements in range movement and Oxford Elbow Score were found
(p <
0.001). Radiolucent lines were much more common in, and
aseptic loosening was exclusive to, the humeral component. Kaplan–Meier
survivorship at five years was 92.2% (95% CI 74.5% to 96.4%) for
aseptic loosening. In four TERs, periprosthetic infection occurred
resulting in failure. A statistically significant association between
infection and increased BMI was found (p = 0.0268). Triceps failure
was more frequent after the Mayo surgical approach and TER performed
after previous trauma surgery. No failures of the implant were noted. Our comparison shows that the Discovery has early clinical results
that are similar to other semi-constrained TERs. We found continued
radiological surveillance with particular focus on humeral lucency
is warranted and has not previously been reported. Despite advances
in the design of total elbow replacement prostheses, rates of complication
remain high. Cite this article: