Objectives. Excessive acetabular coverage is the most common cause of pincer-type
femoroacetabular
Third-body wear is believed to be one trigger for adverse results
with metal-on-metal (MOM) bearings. Impingement and subluxation
may release metal particles from MOM replacements. We therefore
challenged MOM bearings with relevant debris types of cobalt–chrome
alloy (CoCr), titanium alloy (Ti6Al4V) and polymethylmethacrylate
bone cement (PMMA). Cement flakes (PMMA), CoCr and Ti6Al4V particles (size range
5 µm to 400 µm) were run in a MOM wear simulation. Debris allotments
(5 mg) were inserted at ten intervals during the five million cycle
(5 Mc) test. Objectives
Methods
The effects of disease progression and common tendinopathy treatments
on the tissue characteristics of human rotator cuff tendons have
not previously been evaluated in detail owing to a lack of suitable
sampling techniques. This study evaluated the structural characteristics
of torn human supraspinatus tendons across the full disease spectrum,
and the short-term effects of subacromial corticosteroid injections
(SCIs) and subacromial decompression (SAD) surgery on these structural
characteristics. Samples were collected inter-operatively from supraspinatus tendons
containing small, medium, large and massive full thickness tears
(n = 33). Using a novel minimally invasive biopsy technique, paired
samples were also collected from supraspinatus tendons containing
partial thickness tears either before and seven weeks after subacromial
SCI (n = 11), or before and seven weeks after SAD surgery (n = 14).
Macroscopically normal subscapularis tendons of older patients (n
= 5, mean age = 74.6 years) and supraspinatus tendons of younger
patients (n = 16, mean age = 23.3) served as controls. Ultra- and
micro-structural characteristics were assessed using atomic force
microscopy and polarised light microscopy respectively. Objectives
Methods
Acetabular component orientation in total hip arthroplasty (THA)
influences results. Intra-operatively, the natural arthritic acetabulum
is often used as a reference to position the acetabular component.
Detailed information regarding its orientation is therefore essential. The
aim of this study was to identify the acetabular inclination and
anteversion in arthritic hips. Acetabular inclination and anteversion in 65 symptomatic arthritic
hips requiring THA were measured using a computer navigation system.
All patients were Caucasian with primary osteoarthritis (29 men,
36 women). The mean age was 68 years (SD 8). Mean inclination was
50.5° (SD 7.8) in men and 52.1° (SD 6.7) in women. Mean anteversion
was 8.3° (SD 8.7) in men and 14.4° (SD 11.6) in women. Objectives
Methods