Aims. Navigation
Instability is a common cause of failure after total hip arthroplasty. A novel reverse total hip has been developed, with a femoral cup and acetabular ball, creating enhanced mechanical stability. The purpose of this study was to assess the implant fixation using radiostereometric analysis (RSA), and the clinical safety and efficacy of this novel design. Patients with end-stage osteoarthritis were enrolled in a prospective cohort at a single centre. The cohort consisted of 11 females and 11 males with mean age of 70.6 years (SD 3.5) and BMI of 31.0 kg/m2 (SD 5.7). Implant fixation was evaluated using RSA as well as Western Ontario and McMaster Universities Osteoarthritis Index, Harris Hip Score, Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, 38-item Short Form survey, and EuroQol five-dimension health questionnaire scores at two-year follow-up. At least one acetabular screw was used in all cases. RSA markers were inserted into the innominate bone and proximal femur with imaging at six weeks (baseline) and six, 12, and 24 months. Independent-samples Aims
Methods
To develop prediction models using machine-learning (ML) algorithms for 90-day and one-year mortality prediction in femoral neck fracture (FNF) patients aged 50 years or older based on the Hip fracture Evaluation with Alternatives of Total Hip arthroplasty versus Hemiarthroplasty (HEALTH) and Fixation using Alternative Implants for the Treatment of Hip fractures (FAITH) trials. This study included 2,388 patients from the HEALTH and FAITH trials, with 90-day and one-year mortality proportions of 3.0% (71/2,388) and 6.4% (153/2,388), respectively. The mean age was 75.9 years (SD 10.8) and 65.9% of patients (1,574/2,388) were female. The algorithms included patient and injury characteristics. Six algorithms were developed, internally validated and evaluated across discrimination (c-statistic; discriminative ability between those with risk of mortality and those without), calibration (observed outcome compared to the predicted probability), and the Brier score (composite of discrimination and calibration).Aims
Methods
Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years.Aims
Methods
Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon. A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.Aims
Methods
The purpose of this study is to report our updated results at a minimum follow-up of 30 years using a first generation uncemented tapered femoral component in primary total hip arthroplasty (THA). The original cohort consisted of 145 consecutive THAs performed by a single surgeon in 138 patients. A total of 37 patients (40 hips) survived a minimum of 30 years, and are the focus of this review. The femoral component used in all cases was a first-generation Taperloc with a non-modular 28 mm femoral head. Clinical follow-up at a minimum of 30 years was obtained on every living patient. Radiological follow-up at 30 years was obtained on all but four.Aims
Methods
Large acetabular bone defects encountered in revision total hip arthroplasty (THA) are challenging to restore. Metal constructs for structural support are combined with bone graft materials for restoration. Autograft is restricted due to limited volume, and allogenic grafts have downsides including cost, availability, and operative processing. Bone graft substitutes (BGS) are an attractive alternative if they can demonstrate positive remodelling. One potential product is a biphasic injectable mixture (Cerament) that combines a fast-resorbing material (calcium sulphate) with the highly osteoconductive material hydroxyapatite. This study reviews the application of this biomaterial in large acetabular defects. We performed a retrospective review at a single institution of patients undergoing revision THA by a single surgeon. We identified 49 consecutive patients with large acetabular defects where the biphasic BGS was applied, with no other products added to the BGS. After placement of metallic acetabular implants, the BGS was injected into the remaining bone defects surrounding the new implants. Patients were followed and monitored for functional outcome scores, implant fixation, radiological graft site remodelling, and revision failures.Aims
Methods
Cementing in arthroplasty for hip fracture is associated with improved postoperative function, but may have an increased risk of early mortality compared to uncemented fixation. Quantifying this mortality risk is important in providing safe patient care. This study investigated the association between cement use in arthroplasty and mortality at 30 days and one year in patients aged 50 years and over with hip fracture. This retrospective cohort study used linked data from the Australian Hip Fracture Registry and the National Death Index. Descriptive analysis and Kaplan-Meier survival curves tested the unadjusted association of mortality between cemented and uncemented procedures. Multilevel logistic regression, adjusted for covariates, tested the association between cement use and 30-day mortality following arthroplasty. Given the known institutional variation in preference for cemented fixation, an instrumental variable analysis was also performed to minimize the effect of unknown confounders. Adjusted Cox modelling analyzed the association between cement use and mortality at 30 days and one year following surgery.Aims
Methods
The primary aim of this study was to compare surgical methods (sliding hip screw (SHS) vs intramedullary nailing (IMN)) for trochanteric hip fracture in relation to death within 120 days after surgery and return to independent living. The secondary aim was to assess whether the associations between surgical method and death or ability to return to independent living varied depending on fracture subtype or other patient characteristics. A total of 27,530 individuals from the Swedish Hip Fracture Register RIKSHÖFT (SHR) aged ≥ 70 years, admitted to hospital between 1 January 2014 and 31 December 2019 with trochanteric hip fracture, were included. Within this cohort, 12,041 individuals lived independently at baseline, had follow-up information in the SHR, and were thus investigated for return to independent living. Death within 120 days after surgery was analyzed using Cox regression with SHS as reference and adjusted for age and fracture type. Return to independent living was analyzed using logistic regression adjusted for age and fracture type. Analyses were repeated after stratification by fracture type, age, and sex.Aims
Methods
This study aimed to assess the carbon footprint associated with total hip arthroplasty (THA) in a UK hospital setting, considering various components within the operating theatre. The primary objective was to identify actionable areas for reducing carbon emissions and promoting sustainable orthopaedic practices. Using a life-cycle assessment approach, we conducted a prospective study on ten cemented and ten hybrid THA cases, evaluating carbon emissions from anaesthetic room to recovery. Scope 1 and scope 2 emissions were considered, focusing on direct emissions and energy consumption. Data included detailed assessments of consumables, waste generation, and energy use during surgeries.Aims
Methods
Although the Fitmore Hip Stem has been on the market for almost 15 years, it is still not well documented in randomized controlled trials. This study compares the Fitmore stem with the CementLeSs (CLS) in several different clinical and radiological aspects. The hypothesis is that there will be no difference in outcome between stems. In total, 44 patients with bilateral hip osteoarthritis were recruited from the outpatient clinic at a single tertiary orthopaedic centre. The patients were operated with bilateral one-stage total hip arthroplasty. The most painful hip was randomized to either Fitmore or CLS femoral component; the second hip was operated with the femoral component not used on the first side. Patients were evaluated at three and six months and at one, two, and five years postoperatively with patient-reported outcome measures, radiostereometric analysis, dual-energy X-ray absorptiometry, and conventional radiography. A total of 39 patients attended the follow-up visit at two years (primary outcome) and 35 patients at five years. The primary outcome was which hip the patient considered to have the best function at two years.Aims
Methods
The aim of this study was to explore the functional results in a fitter subgroup of participants in the Hip Fracture Evaluation with Alternatives of Total Hip Arthroplasty versus Hemiarthroplasty (HEALTH) trial to determine whether there was an advantage of total hip arthroplasty (THA) versus hemiarthroplasty (HA) in this population. We performed a post hoc exploratory analysis of a fitter cohort of patients from the HEALTH trial. Participants were aged over 50 years and had sustained a low-energy displaced femoral neck fracture (FNF). The fittest participant cohort was defined as participants aged 70 years or younger, classified as American Society of Anesthesiologists grade I or II, independent walkers prior to fracture, and living at home prior to fracture. Multilevel models were used to estimate the effect of THA versus HA on functional outcomes. In addition, a sensitivity analysis of the definition of the fittest participant cohort was performed.Aims
Methods
Pelvic discontinuity is a rare but increasingly common complication of total hip arthroplasty (THA). This single-centre study evaluated the performance of custom-made triflange acetabular components in acetabular reconstruction with pelvic discontinuity by determining: 1) revision and overall implant survival rates; 2) discontinuity healing rate; and 3) Harris Hip Score (HHS). Retrospectively collected data of 38 patients (39 hips) with pelvic discontinuity treated with revision THA using a custom-made triflange acetabular component were analyzed. Minimum follow-up was two years (mean 5.1 years (2 to 11)).Aims
Methods
Our objective was describing an algorithm to identify and prevent vascular injury in patients with intrapelvic components. Patients were defined as at risk to vascular injuries when components or cement migrated 5 mm or more beyond the ilioischial line in any of the pelvic incidences (anteroposterior and Judet view). In those patients, a serial investigation was initiated by a CT angiography, followed by a vascular surgeon evaluation. The investigation proceeded if necessary. The main goal was to assure a safe tissue plane between the hardware and the vessels.Aims
Methods
The primary aim of our study was to assess the influence of age on hip-specific outcome following total hip arthroplasty (THA). Secondary aims were to assess health-related quality of life (HRQoL) and level of activity according to age. A prospective cohort study was conducted. All patients were fitted with an Exeter stem with a 32 mm head on highly cross-linked polyethylene (X3RimFit) cemented acetabulum. Patients were recruited into three age groups: < 65 years, 65 to 74 years, and ≥ 75 years, and assessed preoperatively and at three, 12, 24, and 60 months postoperatively. Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Harris Hip Score (HHS), and Hip disability and Osteoarthritis Outcome Score (HOOS), were used to assess hip-specific outcome. EuroQol five-dimension five-level questionnaire (EQ-5D-5L) and 36-Item Short Form Survey (SF-36) scores were used to assess HRQoL. The Lower Extremity Activity Scale (LEAS) and Timed Up and Go (TUG) were used to assess level of activity.Aims
Methods
Perthes’ disease is an uncommon hip disorder with limited data on the long-term outcomes in adulthood. We partnered with community-based foundations and utilized web-based survey methodology to develop the Adult Perthes Survey, which includes demographics, childhood and adult Perthes’ disease history, the University of California Los Angeles (UCLA) Activity Scale item, Short Form-36, the Hip disability and Osteoarthritis Outcome Score, and a body pain diagram. Here we investigate the following questions: 1) what is the feasibility of obtaining > 1,000 survey responses from adults who had Perthes’ disease using a web-based platform?; and 2) what are the baseline characteristics and demographic composition of our sample? The survey link was available publicly for 15 months and advertised among support groups. Of 1,505 participants who attempted the Adult Perthes survey, 1,182 completed it with a median timeframe of 11 minutes (IQR 8.633 to 14.72). Participants who dropped out were similar to those who completed the survey on several fixed variables. Participants represented 45 countries including the USA (n = 570; 48%), UK (n = 295; 25%), Australia (n = 133; 11%), and Canada (n = 46; 4%). Of the 1,182 respondents, 58% were female and the mean age was 39 years (SD 12.6).Aims
Methods
The aim of the HIPGEN consortium is to develop the first cell therapy product for hip fracture patients using PLacental-eXpanded (PLX-PAD) stromal cells. HIPGEN is a multicentre, multinational, randomized, double-blind, placebo-controlled trial. A total of 240 patients aged 60 to 90 years with low-energy femoral neck fractures (FNF) will be allocated to two arms and receive an intramuscular injection of either 150 × 106 PLX-PAD cells or placebo into the medial gluteal muscle after direct lateral implantation of total or hemi hip arthroplasty. Patients will be followed for two years. The primary endpoint is the Short Physical Performance Battery (SPPB) at week 26. Secondary and exploratory endpoints include morphological parameters (lean body mass), functional parameters (abduction and handgrip strength, symmetry in gait, weightbearing), all-cause mortality rate and patient-reported outcome measures (Lower Limb Measure, EuroQol five-dimension questionnaire). Immunological biomarker and in vitro studies will be performed to analyze the PLX-PAD mechanism of action. A sample size of 240 subjects was calculated providing 88% power for the detection of a 1 SPPB point treatment effect for a two-sided test with an α level of 5%.Aims
Methods
With increasing burden of revision hip arthroplasty (THA), one of the major challenges is the management of proximal femoral bone loss associated with previous multiple surgeries. Proximal femoral arthroplasty (PFA) has already been popularized for tumour surgeries. Our aim was to describe the outcome of using PFA in these demanding non-neoplastic cases. A retrospective review of 25 patients who underwent PFA for non-neoplastic indications between January 2009 and December 2015 was undertaken. Their clinical and radiological outcome, complication rates, and survival were recorded. All patients had the Stanmore Implant – Modular Endo-prosthetic Tumour System (METS).Aims
Methods
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
Large-diameter metal-on-metal (MoM) total hip arthroplasty (THA) has demonstrated unexpected high failure rates and pseudotumour formation. The purpose of this prospective cohort study is to report ten-year results in order to establish revision rate, prevalence of pseudotumour formation, and relation with whole blood cobalt levels. All patients were recalled according to the guidelines of the Dutch Orthopaedic Association. They underwent clinical and radiographical assessments (radiograph and CT scan) of the hip prosthesis and whole blood cobalt ion measurements. Overall, 94 patients (95 hips) fulfilled our requirements for a minimum ten-year follow-up.Aims
Methods