Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis. At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p <
0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate. There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.
Aims. The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel coating that is capable of a controlled antibiotic release for
Gram-negative periprosthetic joint infection (PJI) has been poorly studied despite its rapidly increasing incidence. Treatment with one-stage revision using intra-articular (IA) infusion of antibiotics may offer a reasonable alternative with a distinct advantage of providing a means of delivering the drug in high concentrations. Carbapenems are regarded as the last line of defense against severe Gram-negative or polymicrobial infection. This study presents the results of one-stage revision using intra-articular carbapenem infusion for treating Gram-negative PJI, and analyzes the characteristics of bacteria distribution and drug sensitivity. We retrospectively reviewed 32 patients (22 hips and 11 knees) who underwent single-stage revision combined with IA carbapenem infusion between November 2013 and March 2020. The IA and intravenous (IV) carbapenem infusions were administered for a single Gram-negative infection, and IV vancomycin combined with IA carbapenems and vancomycin was applied for polymicrobial infection including Gram-negative bacteria. The bacterial community distribution, drug sensitivity, infection control rate, functional recovery, and complications were evaluated. Reinfection or death caused by PJI was regarded as a treatment failure.Aims
Methods
Survival analysis is an important tool for assessing the outcome of total joint replacement. The Kaplan-Meier method is used to estimate the incidence of revision of a prosthesis over time, but does not account appropriately for competing events which preclude revision. In the presence of competing death, this method will lead to statistical bias and the curve will lose its interpretability. A valid comparison of survival results between studies using the method is impossible without accounting for different rates of competing events. An alternative and easily applicable approach, the cumulative incidence of competing risk, is proposed. Using three simulated data sets and realistic data from a cohort of 406 consecutive
Prophylaxis against venous thromboembolism after elective total hip replacement is routinely recommended. Our preference has been to use mechanical prophylaxis without anticoagulant drugs. A randomised controlled trial was performed to evaluate whether the incidence of post-operative venous thromboembolism was reduced by using pharmacological anticoagulation with either fondaparinux or enoxaparin in addition to our prophylactic mechanical regimen. A total of 255 Japanese patients who underwent primary unilateral
To investigate whether chronic kidney disease (CKD) is associated with the risk of all-cause revision or revision due to a periprosthetic joint infection (PJI) after primary hip or knee arthroplasty. This retrospective cohort study comprised 18,979 consecutive hip and knee arthroplasties from a single high-volume academic hospital. At a median of 5.6 years (interquartile range (IQR) 3.5 to 8.1), all deaths and revisions were counted. To overcome the competing risk of death, competing risk analysis using the cumulative incidence function (CIF) was applied to analyze the association between different stages of CKD and revisions. Confounding factors such as diabetes and BMI were considered using either a stratified CIF or the Fine and Gray model.Aims
Methods
The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
In the absence of an identified organism, single-stage revision is contraindicated in prosthetic joint infection (PJI). However, no studies have examined the use of intra-articular antibiotics in combination with single-stage revision in these cases. In this study, we present the results of single-stage revision using intra-articular antibiotic infusion for treating culture-negative (CN) PJI. A retrospective analysis between 2009 and 2016 included 51 patients with CN PJI who underwent single-stage revision using intra-articular antibiotic infusion; these were compared with 192 culture-positive (CP) patients. CN patients were treated according to a protocol including intravenous vancomycin and a direct intra-articular infusion of imipenem and vancomycin alternately used in the morning and afternoon. In the CP patients, pathogen-sensitive intravenous (IV) antibiotics were administered for a mean of 16 days (12 to 21), and for resistant cases, additional intra-articular antibiotics were used. The infection healing rate, Harris Hip Score (HHS), and Hospital for Special Surgery (HSS) knee score were compared between CN and CP groups.Aims
Methods
Antifibrinolytic agents, including tranexamic acid (TXA) and epsilon-aminocaproic acid (EACA), have been shown to be safe and effective for decreasing perioperative blood loss and transfusion following total hip arthroplasty (THA) and total knee arthroplasty (TKA). However, there are few prospective studies that directly compare these agents. The purpose of this study was to compare the benefits of intraoperative intravenous TXA with EACA. A total of 235 patients (90 THA and 145 TKA) were enrolled in this prospective, randomized controlled trial at a single tertiary-care referral centre. In the THA cohort, 53.3% of the patients were female with a median age of 59.8 years (interquartile range (IQR) 53.3 to 68.1). In the TKA cohort, 63.4% of the patients were female with a median age of 65.1 years (IQR 59.4 to 69.5). Patients received either TXA (n = 119) or EACA (n = 116) in two doses intraoperatively. The primary outcome measures included change in haemoglobin level and blood volume, postoperative drainage, and rate of transfusion. Secondary outcome measures included postoperative complications, cost, and length of stay (LOS).Aims
Patients and Methods
We report our results using three different threaded acetabular components (Mecring A, Mecring B and Weill) in 715 hips with a follow-up of between one and ten years (median: 99.1, 56.5, 38.3 months, respectively). All cups were implanted with one type of
We report a prospective randomly controlled trial to examine the effectiveness of continuous passive motion (CPM) in improving postoperative function and range of movement after total knee arthroplasty (TKA). We allocated 53 patients (57 knees) to one of three postoperative regimes: no CPM (n = 19); CPM at 0 to 40° (0 to 40 CPM; n = 18); and CPM at 0 to 70° (0 to 70 CPM; n = 20). Those in the CPM groups had CPM for 48 hours and all patients had an identical regime of physiotherapy. There was an even distribution of various cemented and
The aims of this study were to determine the indications and
frequency of ordering a CT pulmonary angiography (CTPA) following
primary arthroplasty of the hip and knee, and to determine the number
of positive scans in these patients, the location of emboli and
the outcome for patients undergoing CTPA. We analyzed the use of CTPA, as an inpatient and up to 90 days
as an outpatient, in a cohort of patients and reviewed the medical
records and imaging for each patient undergoing CTPA.Aims
Patients and Methods
Periprosthetic fracture is a significant complication of total
hip and knee arthroplasty. This study aimed to describe the survival
of patients sustaining periprosthetic femoral fractures and compare
this with that of the general population, as well as to identify
the factors that influence survival. A total of 151 patients (women: men 116:35, mean age 74.6 years,
standard deviation 11.5) that sustained a periprosthetic fracture
between January 2005 and October 2012 were retrospectively analysed.
Epidemiological data, comorbidities, type of surgical management,
type of implant, and mortality data were studied.Aims
Patients and Methods
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article:
Increasing demand for total hip and knee arthroplasty (THA/TKA)
and associated follow-up has placed huge demands on orthopaedic
services. Feasible follow-up mechanisms are therefore essential. We conducted an audit of clinical follow-up decision-making for
THA/TKA based on questionnaire/radiograph review compared with local
practice of Arthroplasty Care Practitioner (ACP)-led outpatient
follow-up. In all 599 patients attending an ACP-led THA/TKA follow-up
clinic had a pelvic/knee radiograph, completed a pain/function questionnaire
and were reviewed by an ACP. An experienced orthopaedic surgeon
reviewed the same radiographs and questionnaires, without patient
contact or knowledge of the ACP’s decision. Each pathway classified
patients into: urgent review, annual monitoring, routine follow-up
or discharge. Aims
Methods
Periprosthetic joint infection (PJI) is a devastating
complication for patients and results in greatly increased costs
of care for both healthcare providers and patients. More than 15
500 revision hip and knee procedures were recorded in England, Wales
and Northern Ireland in 2013, with infection accounting for 13%
of revision hip and 23% of revision knee procedures. We report our experience of using antibiotic eluting absorbable
calcium sulphate beads in 15 patients (eight men and seven women
with a mean age of 64.8 years; 41 to 83) as part of a treatment
protocol for PJI in revision arthroplasty. The mean follow-up was 16 months (12 to 22). We report the outcomes
and complications, highlighting the risk of hypercalcaemia which
occurred in three patients. We recommend that serum levels of calcium be routinely sought
following the implantation of absorbable calcium sulphate beads
in orthopaedic surgery. Cite this article:
We compared the length of hospitalisation, rate
of infection, dislocation of the hip and revision, and mortality following
primary hip and knee arthroplasty for osteoarthritis in patients
with Alzheimer’s disease (n = 1064) and a matched control group
(n = 3192). The data were collected from nationwide Finnish health
registers. Patients with Alzheimer’s disease had a longer peri-operative
hospitalisation (median 13 days Cite this article:
We have investigated the accuracy of the templating of digital radiographs in planning total hip replacement using two common object-based calibration methods with the ball placed laterally (method 1) or medially (method 2) and compared them with two non-object-based methods. The latter comprised the application of a fixed magnification of 121% (method 3) and calculation of magnification based on the object-film-distance (method 4). We studied the post-operative radiographs of 57 patients (19 men, 38 women, mean age 73 years (53 to 89)) using the measured diameter of the prosthetic femoral head and comparing it with the true value. Both object-based methods (1 and 2) produced large errors (mean/maximum: 2.55%/17.4% and 2.04%/6.46%, respectively). Method 3 applying a fixed magnification and method 4 (object-film-distance) produced smaller errors (mean/maximum 1.42%/5.22% and 1.57%/4.24%, respectively; p <
0.01). The latter results were clinically relevant and acceptable when planning was allowed to within one implant size. Object-based calibration (methods 1 and 2) has fundamental problems with the correct placement of the calibration ball. The accuracy of the fixed magnification (method 3) matched that of object-film-distance (method 4) and was the most reliable and efficient calibration method in digital templating.
Balancing service provision and surgical training is a challenging issue that affects all healthcare systems. A multicentre prospective study of 1501 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and the grade of the operating surgeon, and whether there is any difference in outcome if surgeons’ assistants assist with the operation, rather than orthopaedic trainees. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of revision and dislocation, operating time, and length of hospital stay. There was no significant difference in ΔOHS or complication rates between operations undertaken by trainers and trainees, or those at which surgeons’ assistants and trainees were the assistant. However, there was a significant difference in the duration of surgery, with a mean reduction of 28 minutes in those in which a surgeons’ assistant was the assistant. This study provides evidence that total hip replacements can be performed safely and effectively by appropriately trained surgeons in training, and that there are potential benefits of using surgeons’ assistants in orthopaedic surgery.
Implantation of allograft bone is an integral part of revision surgery of the hip. One major concern with its use is the risk of transmission of infective agents. There are a number of methods of processing allograft bone in order to reduce this risk. One method requires washing the tissue using pulsed irrigation immediately before implantation. We report the incidence of deep bacterial infection in 138 patients (144 revision hip arthroplasties) who had undergone implantation of allograft bone. The bone used was fresh-frozen, non-irradiated and pulse-washed with normal saline before implantation. The deep infection rate at a minimum follow-up of one year was 0.7%. This method of processing appears to be associated with a very low risk of allograft-related bacterial infection.