Advertisement for orthosearch.org.uk
Results 1 - 100 of 396
Results per page:
The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1304 - 1312
1 Oct 2017
Langton DJ Sidaginamale RP Joyce TJ Meek RD Bowsher JG Deehan D Nargol AVF Holland JP

Aims. We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system. Patients and Methods. We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy. Results. After adjustment for other factors, CoCr stem tapers were found to have significantly greater volumetric material loss than the equivalent Ti stem tapers. Conclusion. When taper junction damage is identified during revision of a LD MoM hip, it should be suspected that a male taper composed of a standard CoCr alloy has sustained significant changes to the taper cone geometry which are likely to be more extensive than those affecting a Ti alloy stem. Cite this article: Bone Joint J 2017;99-B:1304–12


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 925 - 933
1 Jul 2016
Sidaginamale RP Joyce TJ Bowsher JG Lord JK Avery PJ Natu S Nargol AVF Langton DJ

Aims. We wished to investigate the influence of metal debris exposure on the subsequent immune response and resulting soft-tissue injury following metal-on-metal (MoM) hip arthroplasty. Some reports have suggested that debris generated from the head-neck taper junction is more destructive than equivalent doses from metal bearing surfaces. . Patients and Methods. We investigated the influence of the source and volume of metal debris on chromium (Cr) and cobalt (Co) concentrations in corresponding blood and hip synovial fluid samples and the observed agglomerated particle sizes in excised tissues using multiple regression analysis of prospectively collected data. A total of 199 explanted MoM hips (177 patients; 132 hips female) were analysed to determine rates of volumetric wear at the bearing surfaces and taper junctions. . Results. The statistical modelling suggested that a greater source contribution of metal debris from the taper junction was associated with smaller aggregated particle sizes in the local tissues and a relative reduction of Cr ion concentrations in the corresponding synovial fluid and blood samples. Metal debris generated from taper junctions appears to be of a different morphology, composition and therefore, potentially, immunogenicity to that generated from bearing surfaces. Conclusion. The differences in debris arising from the taper and the articulating surfaces may provide some understanding of the increased incidence of soft-tissue reactions reported in patients implanted with MoM total hip arthroplasties compared with patients with hip resurfacings. Cite this article: Bone Joint J 2016;98-B:925–33


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 463 - 472
1 Apr 2015
Panagiotidou A Meswania J Osman K Bolland B Latham J Skinner J Haddad FS Hart A Blunn G

The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper interface of a modular femoral component and to investigate whether different combinations of material also had an effect. The combinations we examined were 1) cobalt–chromium (CoCr) heads on CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14 Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending moment by offsetting the application of axial load from the midline in the mediolateral plane. Increments of offset equivalent to head + 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing torque for all combinations of material. However, Ti stems showed the highest corrosion currents. Increased bending moments associated with using larger offset heads produced more corrosion: Ti stems generally performed worse than CoCr stems. Using ceramic heads did not prevent corrosion, but reduced it significantly in all loading configurations. Cite this article: Bone Joint J 2015;97-B:463–72


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 852 - 858
1 Jul 2022
Grothe T Günther K Hartmann A Blum S Haselhoff R Goronzy J

Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Methods. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti. 12. Mo. 6. Zr. 2. Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients. Results. A total of 20 ALTRs were identified on MRI, with an incidence of 26%. Patients with ALTR had significantly higher median Co levels compared with those without ALTR (2.96 μg/l (interquartile range (IQR) 1.35 to 4.98) vs 1.44 μg/l (IQR 0.79 to 2.5); p = 0.019). Radiological evidence of osteolysis was also significantly associated with ALTR (p = 0.014). Median Cr levels were not significantly higher in those with ALTR compared with those without one (0.97 μg/l (IQR 0.72 to 1.9) v 0.67 μg/l (IQR 0.5 to 1.19; p = 0.080). BMI, sex, age, type of femoral component, head length, the inclination of the acetabular component, and heterotopic ossification formation showed no significant relationship with ALTR. Conclusion. Due to the high incidence of local ALTR in our cohort after more than ten years postoperatively, we recommend regular follow-up investigation even in asymptomatic patients with V40 taper and metal heads. As cobalt levels correlate with ALTR occurrence, routine metal ion screening and consecutive MRI investigation upon elevation could be discussed. Cite this article: Bone Joint J 2022;104-B(7):852–858


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1310 - 1319
1 Oct 2018
Langton DJ Wells SR Joyce TJ Bowsher JG Deehan D Green S Nargol AVF Holland JP

Aims. There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. Patients and Methods. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling. Results. A total of 95 Exeter MoP and 249 MoM THAs were examined. The median volumetric loss from the MoM cohort was over four times larger than that from the MoP cohort (1.01 mm. 3. vs 0.23 mm. 3. , p < 0.001), despite a significantly shorter median period in vivo for the MoM group (48 months vs 90 months, p < 0.001). Multiple regression modelling indicated that the dominant variables leading to greater female taper material loss were bearing diameter (p < 0.001), larger female taper angles (p < 0.001), and male titanium stem tapers (p < 0.001). Conclusion. Consistent with the long-term clinical success of the device, the volumetric material loss from Exeter femoral head tapers was, in general, small compared with that from larger-diameter MoM head tapers. Cite this article: Bone Joint J 2018;100-B:1310–9


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 125 - 132
1 May 2024
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP

Aims. Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components. Methods. We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m. 2. (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used. Results. The 15-year cumulative incidence of any revision was 24%. There were 48 revisions, with the most common reasons being dislocation (n = 25) and infection (n = 16). The 15-year cumulative incidence of any reoperation was 28%. Only 13 revisions involved the fluted tapered component (FTC), for a 15-year cumulative incidence of any FTC revision of 8%. Only two FTCs were revised for aseptic loosening, resulting in a 15-year cumulative incidence of FTC revision for aseptic loosening of 1%. Stem subsidence ≥ 5 mm occurred in 2% of unrevised cases. All stems were radiologically stable at most recent follow-up. Mean Harris Hip Score was 69 (SD 20) at most recent follow-up. Conclusion. This series demonstrated that MFT components were durable and reliable in the setting of two-stage reimplantation THA for infection. While the incidence of aseptic loosening was very low, the incidence of any revision was 24% at 15 years, primarily due to dislocation and recurrent PJI. Cite this article: Bone Joint J 2024;106-B(5 Supple B):125–132


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 33 - 40
1 Jul 2020
Gustafson JA Pourzal R Levine BR Jacobs JJ Lundberg HJ

Aims. The aim of this study was to develop a novel computational model for estimating head/stem taper mechanics during different simulated assembly conditions. Methods. Finite element models of generic cobalt-chromium (CoCr) heads on a titanium stem taper were developed and driven using dynamic assembly loads collected from clinicians. To verify contact mechanics at the taper interface, comparisons of deformed microgroove characteristics (height and width of microgrooves) were made between model estimates with those measured from five retrieved implants. Additionally, these models were used to assess the role of assembly technique—one-hit versus three-hits—on the taper interlock mechanical behaviour. Results. The model compared well to deformed microgrooves from the retrieved implants, predicting changes in microgroove height (mean 1.1 μm (0.2 to 1.3)) and width (mean 7.5 μm (1.0 to 18.5)) within the range of measured changes in height (mean 1.4 μm (0.4 to 2.3); p = 0.109) and width (mean 12.0 μm (1.5 to 25.4); p = 0.470). Consistent with benchtop studies, our model found that increasing assembly load magnitude led to increased taper engagement, contact pressure, and permanent deformation of the stem taper microgrooves. Interestingly, our model found assemblies using three hits at low loads (4 kN) led to decreased taper engagement, contact pressures and microgroove deformations throughout the stem taper compared with tapers assembled with one hit at the same magnitude. Conclusion. These findings suggest additional assembly hits at low loads lead to inferior taper interlock strength compared with one firm hit, which may be influenced by loading rate or material strain hardening. These unique models can estimate microgroove deformations representative of real contact mechanics seen on retrievals, which will enable us to better understand how both surgeon assembly techniques and implant design affect taper interlock strength. Cite this article: Bone Joint J 2020;102-B(7 Supple B):33–40


Aims. Revision total hip arthroplasty in patients with Vancouver type B3 fractures with Paprosky type IIIA, IIIB, and IV femoral defects are difficult to treat. One option for Paprovsky type IIIB and IV defects involves modular cementless, tapered, revision femoral components in conjunction with distal interlocking screws. The aim of this study was to analyze the rate of reoperations and complications and union of the fracture, subsidence of the stem, mortality, and the clinical outcomes in these patients. Methods. A total of 46 femoral components in patients with Vancouver B3 fractures (23 with Paprosky type IIIA, 19 with type IIIB, and four with type IV defects) in 46 patients were revised with a transfemoral approach using a modular, tapered, cementless revision Revitan curved femoral component with distal cone-in-cone fixation and prospectively followed for a mean of 48.8 months (SD 23.9; 24 to 112). The mean age of the patients was 80.4 years (66 to 100). Additional distal interlocking was also used in 23 fractures in which distal cone-in-cone fixation in the isthmus was < 3 cm. Results. One patient (2.2%) died during the first postoperative year. After six months, 43 patients (93.5%) had osseous, and three had fibrous consolidation of the fracture and the bony flap, 42 (91.3%) had bony ingrowth and four had stable fibrous fixation of the stem. No patient had radiolucency around the interlocking screws and no screw broke. One patient had non-progressive subsidence and two had a dislocation. The mean Harris Hip Score increased from of 57.8 points (SD 7.9) three months postoperatively to 76.1 points (SD 10.7) 24 months postoperatively. Conclusion. The 2° tapered, fluted revision femoral component with distal cone-in-cone-fixation, combined with additional distal interlocking in patients with bony deficiency at the isthmus, led to reproducibly good results in patients with a Vancouver B3 fracture and Paprosky type IIIA, IIIB, and IV defects with regard to union of the fracture, subsidence or loosening of the stem, and clinical outcomes. Cite this article: Bone Joint J 2024;106-B(4):344–351


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 481 - 486
1 May 2023
Scott CEH Jain S Moran M Haddad FS

The Unified Classification System (UCS), or Vancouver system, is a validated and widely used classification system to guide the management of periprosthetic femoral fractures. It suggests that well-fixed stems (type B1) can be treated with fixation but that loose stems (types B2 and B3) should be revised. Determining whether a stem is loose can be difficult and some authors have questioned how to apply this classification system to polished taper slip stems which are, by definition, loose within their cement mantle. Recent evidence has challenged the common perception that revision surgery is preferable to fixation surgery for UCS-B periprosthetic fractures around cemented polished taper slip stems. Indications for fixation include an anatomically reducible fracture and cement mantle, a well-fixed femoral bone-cement interface, and a well-functioning acetabular component. However, not all type B fractures can or should be managed with fixation due to the risk of early failure. This annotation details specific fracture patterns that should not be managed with fixation alone. Cite this article: Bone Joint J 2023;105-B(5):481–486


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 46 - 52
1 Jul 2021
McGoldrick NP Fischman D Nicol GM Kreviazuk C Grammatopoulos G Beaulé PE

Aims. The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach. Methods. A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups. Results. No significant differences were identified between groups in Barrack grade on the anteroposterior (AP) (p = 0.640) or lateral views (p = 0.306), or for alignment on the AP (p = 0.603) or lateral views (p = 0.254). An adequate cement mantle (Barrack A or B) was achieved in 77.6% (anterior group, n = 45) and in 86% (posterior group, n = 49), respectively. Multivariate analysis revealed factors associated with unsatisfactory cement mantle (Barrack C or D) included higher BMI, left side, and Dorr Type C morphology. A mean cement mantle thickness of ≥ 2 mm was achieved in all Gruen zones for both approaches. The mean cement mantle was thicker in zone 7 (p < 0.001) and thinner in zone 9 for the anterior approach (p = 0.032). Incidence of cement mantle defects between groups was similar (6.9% (n = 4) vs 8.8% (n = 5), respectively; p = 0.489). Conclusion. An adequate cement mantle and good alignment can be achieved using a collarless polished tapered femoral component inserted through the anterior approach. Cite this article: Bone Joint J 2021;103-B(7 Supple B):46–52


The Bone & Joint Journal
Vol. 103-B, Issue 4 | Pages 644 - 649
1 Apr 2021
Alsousou J Oragu E Martin A Strickland L Newman S Kendrick B Taylor A Glyn-Jones S

Aims. The aim of this prospective cohort study was to evaluate the early migration of the TriFit cementless proximally coated tapered femoral stem using radiostereometric analysis (RSA). Methods. A total of 21 patients (eight men and 13 women) undergoing primary total hip arthroplasty (THA) for osteoarthritis of the hip were recruited in this study and followed up for two years. Two patients were lost to follow-up. All patients received a TriFit stem and Trinity Cup with a vitamin E-infused highly cross-linked ultra-high molecular weight polyethylene liner. Radiographs for RSA were taken postoperatively and then at three, 12, and 24 months. Oxford Hip Score (OHS), EuroQol five-dimension questionnaire (EQ-5D), and adverse events were reported. Results. At two years, the mean subsidence of the head and tip for the TriFit stem was 0.38 mm (SD 0.32) and 0.52 mm (SD 0.36), respectively. The total migration of the head and tip was 0.55 mm (SD 0.32) and 0.71 mm (SD 0.38), respectively. There were no statistically significant differences between the three to 12 months' migration (p = 0.105) and 12 to 24 months' migration (p = 0.694). The OHS and EQ-5D showed significant improvements at two years. Conclusion. The results of this study suggest that the TriFit femoral stem achieves initial stability and is likely to be stable in the mid and long term. A long-term outcome study is required to assess late mechanisms of failure and the effects of bone mineral density (BMD) related changes. Cite this article: Bone Joint J 2021;103-B(4):644–649


The Bone & Joint Journal
Vol. 106-B, Issue 3 | Pages 220 - 223
1 Mar 2024
Kayani B Luo TD Haddad FS


The Bone & Joint Journal
Vol. 97-B, Issue 3 | Pages 312 - 317
1 Mar 2015
Amanatullah DF Howard JL Siman H Trousdale RT Mabry TM Berry DJ

Revision total hip arthroplasty (THA) is challenging when there is severe loss of bone in the proximal femur. The purpose of this study was to evaluate the clinical and radiographic outcomes of revision THA in patients with severe proximal femoral bone loss treated with a fluted, tapered, modular femoral component. Between January 1998 and December 2004, 92 revision THAs were performed in 92 patients using a single fluted, tapered, modular femoral stem design. Pre-operative diagnoses included aseptic loosening, infection and peri-prosthetic fracture. Bone loss was categorised pre-operatively as Paprosky types III-IV, or Vancouver B3 in patients with a peri-prosthetic fracture. The mean clinical follow-up was 6.4 years (2 to 12). A total of 47 patients had peri-operative complications, 27 of whom required further surgery. However, most of these further operations involved retention of a well-fixed femoral stem, and 88/92 femoral components (97%) remained in situ. Of the four components requiring revision, three were revised for infection and were well fixed at the time of revision; only one (1%) was revised for aseptic loosening. The most common complications were post-operative instability (17 hips, 19%) and intra-operative femoral fracture during insertion of the stem (11 hips, 12%). Diaphyseal stress shielding was noted in 20 hips (22%). There were no fractures of the femoral component. At the final follow-up 78% of patients had minimal or no pain. . Revision THA in patients with extensive proximal femoral bone loss using the Link MP fluted, tapered, modular stem led to a high rate of osseointegration of the stem at mid-term follow-up. Cite this article: Bone Joint J 2015; 97-B:312–17


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1031 - 1037
1 Aug 2015
da Assunção RE Pollard TCB Hrycaiczuk A Curry J Glyn-Jones S Taylor A

Periprosthetic femoral fracture (PFF) is a potentially devastating complication after total hip arthroplasty, with historically high rates of complication and failure because of the technical challenges of surgery, as well as the prevalence of advanced age and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty using a modular, titanium, tapered, conical stem for PFF in a series of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of 27 patients had an American Society of Anesthesiologists grade of at least 3. At a mean follow-up of 35 months (4 to 66) the mean Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly associated with a poorer OHS. All fractures united and no stem needed to be revised. Three hips in three patients required further surgery for infection, recurrent PFF and recurrent dislocation and three other patients required closed manipulation for a single dislocation. One stem subsided more than 5 mm but then stabilised and required no further intervention. . In this series, a modular, tapered, conical stem provided a versatile reconstruction solution with a low rate of complications. Cite this article: Bone Joint J 2015;97-B:1031–7


The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 49 - 55
1 Apr 2017
Hossain F Konan S Volpin A Haddad FS

Aims. The aim of this study was to compare early functional and health related quality of life outcomes (HRQoL) in patients who have undergone total hip arthroplasty (THA) using a bone conserving short stem femoral component and those in whom a conventional length uncemented component was used. Outcome was assessed using a validated performance based outcome instrument as well as patient reported outcome measures (PROMs). Patients and Methods. We prospectively analysed 33 patients whose THA involved a contemporary proximally porous coated tapered short stem femoral component and 53 patients with a standard conventional femoral component, at a minimum follow-up of two years. The mean follow-up was 31.4 months (24 to 39). Patients with poor proximal femoral bone quality were excluded. The mean age of the patients was 66.6 years (59 to 77) and the mean body mass index was 30.2 kg/m. 2. (24.1 to 41.0). Outcome was assessed using the Oxford Hip Score (OHS) and the University College Hospital (UCH) hip score which is a validated performance based instrument. HRQoL was assessed using the EuroQol 5D (EQ-5D). Results. There were no major peri-operative complications. There was no difference in the mean post-operative OHS, EQ-5D or function subscale of the UCH hip scores between the two groups. The mean pre-operative OHS and EQ-5D scores improved significantly (all p <  0.001). The mean functional component of the UCH hip score at final follow-up was 42.5 and 40.6 in the short stem and conventional stem groups, respectively. There was no statistically significant difference between the groups (p = 0.42). A total of seven patients (21.2%) in the short stem group and nine (16.98%) in the conventional group achieved a ceiling effect using the OHS; none did using the function subscale of the UCH hip score. Conclusion. The proximally porous coated tapered short stem femoral component achieves comparable short-term functional outcomes when compared with a conventional longer stem uncemented femoral component when THA is undertaken in patients with good bone quality. Cite this article: Bone Joint J 2017;99-B(4 Supple B):49–55


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 191 - 197
1 Feb 2020
Gabor JA Padilla JA Feng JE Schnaser E Lutes WB Park KJ Incavo S Vigdorchik J Schwarzkopf R

Aims. Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS), early results showed high rates of subsidence. Advances in stem design may mitigate these concerns. This study reports on the use of a current monolithic TFTS for a variety of indications. Methods. A multi-institutional retrospective study of all consecutive total hip arthroplasty (THA) and revision total hip arthroplasty (rTHA) patients who received the monolithic TFTS was conducted. Surgery was performed by eight fellowship-trained arthroplasty surgeons at four institutions. A total of 157 hips in 153 patients at a mean follow-up of 11.6 months (SD7.8) were included. Mean patient age at the time of surgery was 67.4 years (SD 13.3) and mean body mass index (BMI) was 28.9 kg/m. 2. (SD 6.5). Outcomes included intraoperative complications, one-year all-cause re-revisions, and subsidence at postoperative time intervals (two weeks, six weeks, six months, nine months, and one year). Results. There were eight intraoperative complications (4.9%), six of which were intraoperative fractures; none occurred during stem insertion. Six hips (3.7%) underwent re-revision within one year; only one procedure involved removal of the prosthesis due to infection. Mean total subsidence at latest follow-up was 1.64 mm (SD 2.47). Overall, 17 of 144 stems (11.8%) on which measurements could be performed had >5 mm of subsidence, and 3/144 (2.1%) had >10 mm of subsidence within one year. A univariate regression analysis found that additional subsidence after three months was minimal. A multivariate regression analysis found that subsidence was not significantly associated with periprosthetic fracture as an indication for surgery, the presence of an extended trochanteric osteotomy (ETO), Paprosky classification of femoral bone loss, stem length, or type of procedure performed (i.e. full revision vs conversion/primary). Conclusion. Advances in implant design, improved trials, a range of stem lengths and diameters, and high offset options mitigate concerns of early subsidence and dislocation with monolithic TFTS, making them a valuable option for femoral revision. Cite this article: Bone Joint J 2020;102-B(2):191–197


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 50 - 53
1 Jan 2016
Konan S Garbuz DS Masri BA Duncan CP

Tapered fluted titanium stems are increasingly used for femoral revision arthroplasty. They are available in modular and non-modular forms. Modularity has advantages when the bone loss is severe, the proximal femur is mis shapen or the surgeon is unfamiliar with the implant, but it introduces the risk of fracture of the stem at the junction between it and the proximal body segment. For that reason, and while awaiting intermediate-term results of more recently introduced designs of this junction, non-modularity has attracted attention, at least for straightforward revision cases. . We review the risks and causes of fracture of tapered titanium modular revision stems and present an argument in favour of the more selective use of modular designs. Cite this article: Bone Joint J 2016;98-B(1 Suppl A):50–3


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 56 - 59
1 Nov 2014
Konan S Garbuz DS Masri BA Duncan CP

Non-modular tapered fluted, titanium stems are available for use in femoral revision. The combination of taper and flutes on the stem provides axial and rotational stability, respectively. The material and surface properties of the stem promotes bone on-growth. If the surgeon is confident and reasonably experienced in the surgical use of this sort of design and the case is relatively straightforward, a non-modular design is effective. It also potentially reduces implant inventory, and circumvents the potential problems of taper junction corrosion and fatigue fracture. There are reports of excellent survival, good clinical and functional results and evidence of subsequent increase in proximal bone stock. Cite this article: Bone Joint J 2014;96-B(11 Suppl A):56–9


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 889 - 894
1 Jul 2012
Burston BJ Barnett AJ Amirfeyz R Yates PJ Bannister GC

We prospectively followed 191 consecutive collarless polished tapered (CPT) femoral stems, implanted in 175 patients who had a mean age at operation of 64.5 years (21 to 85). At a mean follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were still alive. The fate of all original stems is known. The 16-year survivorship with re-operation for any reason was 80.7% (95% confidence interval 72 to 89.4). There was no loss to follow-up, with clinical data available on all 95 hips and radiological assessment performed on 90 hips (95%). At latest follow-up, the mean Harris hip score was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48). Stems subsided within the cement mantle, with a mean subsidence of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem (0.5%) has been revised due to aseptic loosening. In total seven stems were revised for any cause, of which four revisions were required for infection following revision of the acetabular component. A total of 21 patients (11%) required some sort of revision procedure; all except three of these resulted from failure of the acetabular component. Cemented acetabular components had a significantly lower revision burden (three hips, 2.7%) than Harris Galante uncemented components (17 hips, 21.8%) (p < 0.001). The CPT stem continues to provide excellent radiological and clinical outcomes at 15 years following implantation. Its results are consistent with other polished tapered stem designs


The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm. 3. (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r² = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient’s mass, age and body mass index) did not correlate with the amount of material loss observed (p >  0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis. Cite this article: Bone Joint J 2015;97-B:1350–7


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1210 - 1215
1 Sep 2005
Pospischill M Knahr K

We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1). The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant were found. One patient needed revision surgery because of a late deep infection. In 11 hips (10.7%), the reason for revision was progressive wear of the polyethylene liner. Exchange of the acetabular component because of aseptic loosening without detectable liner wear was carried out in three hips (2.9%). After 15 years the survivorship with aseptic loosening as the definition for failure was 95.6% for the acetabular component and 100% for the femoral component


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 921 - 927
1 Jul 2005
Glyn-Jones S Gill HS Beard DJ McLardy-Smith P Murray DW

Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability. Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°, . sd. 0.42°) and internally rotate (mean 1.28°, . sd. 0.99°). The CPS-Plus showed no significant valgus rotation (mean 0.2°, . sd. 0.42°) or internal rotation (mean −0.03°, . sd. 0.75°). A wider, more rectangular cross-section improves rotational stability and may have a better long-term outcome


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 58 - 60
1 Nov 2012
Munro JT Garbuz DS Masri BA Duncan CP

Tapered, fluted, modular, titanium stems have a long history in Europe and are increasing in popularity in North America. We have reviewed the results at our institution looking at stem survival and clinical outcomes. Radiological outcomes and quality of life assessments have been performed and compared to cylindrical non-modular cobalt chromium stems. Survival at five years was 94%. This fell to 85% at ten years due to stem breakage with older designs. Review of radiology showed maintenance or improvement of bone stock in 87% of cases. Outcome scores were superior in tapered stems despite worse pre-operative femoral deficiency. Tapered stems have proved to be a useful alternative in revision total hip arthroplasty across the spectrum of femoral bone deficiency


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1000 - 1004
1 Aug 2008
Wraighte PJ Howard PW

Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998. The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification. The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively. Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p < 0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97)


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 917 - 924
1 Jul 2016
Whittaker RK Hothi HS Meswania JM Berber R Blunn GW Skinner JA Hart AJ

Aims

Surgeons have commonly used modular femoral heads and stems from different manufacturers, although this is not recommended by orthopaedic companies due to the different manufacturing processes.

We compared the rate of corrosion and rate of wear at the trunnion/head taper junction in two groups of retrieved hips; those with mixed manufacturers (MM) and those from the same manufacturer (SM).

Materials and Methods

We identified 151 retrieved hips with large-diameter cobalt-chromium heads; 51 of two designs that had been paired with stems from different manufacturers (MM) and 100 of seven designs paired with stems from the same manufacturer (SM). We determined the severity of corrosion with the Goldberg corrosion score and the volume of material loss at the head/stem junction. We used multivariable statistical analysis to determine if there was a significant difference between the two groups.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 755 - 761
1 Jun 2015
Flatøy B Röhrl SM Rydinge J Dahl J Diep LM Nordsletten L

Cemented femoral stems with force closed fixation designs have shown good clinical results despite high early subsidence. A new triple-tapered stem in this category (C-stem AMT) was introduced in 2005. This study compares this new stem with an established stem of similar design (Exeter) in terms of migration (as measured using radiostereometric analysis), peri-prosthetic bone remodelling (measured using dual energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs.

A total of 70 patients (70 hips) with a mean age of 66 years (53 to 78) were followed for two years. Owing to missing data of miscellaneous reasons, the final analysis represents data from 51 (RSA) and 65 (DXA) patients. Both stems showed a typical pattern of migration: Subsidence and retroversion that primarily occurred during the first three months. C-stem AMT subsided less during the first three months (p = 0.01), before stabilising at a subsidence rate similar to the Exeter stem from years one to two. The rate of migration into retroversion was slightly higher for C-stem AMT during the second year (p = 0.03). Whilst there were slight differences in movement patterns between the stems, the C-stem AMT exhibits good early clinical outcomes and displays a pattern of migration and bone remodelling that predicts good clinical performance.

Cite this article: Bone Joint J 2015;97-B:755–61.


Aims

To evaluate the hypothesis that failed osteosynthesis of periprosthetic Vancouver type B1 fractures can be treated successfully with stem revision using a transfemoral approach and a cementless, modular, tapered revision stem with reproducible rates of fracture healing, stability of the revision stem, and clinically good results.

Patients and Methods

A total of 14 patients (11 women, three men) with a mean age of 72.4 years (65 to 90) undergoing revision hip arthroplasty after failed osteosynthesis of periprosthetic fractures of Vancouver type B1 were treated using a transfemoral approach to remove the well-fixed stem before insertion of a modular, fluted titanium stem which obtained distal fixation. These patients were clinically and radiologically followed up for a mean 52.2 months (24 to 144).


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 889 - 895
1 Jul 2014
Fink B Urbansky K Schuster P

We report our experience of revision total hip replacement (THR) using the Revitan curved modular titanium fluted revision stem in patients with a full spectrum of proximal femoral defects. A total of 112 patients (116 revisions) with a mean age of 73.4 years (39 to 90) were included in the study. The mean follow-up was 7.5 years (5.3 to 9.1). A total of 12 patients (12 hips) died but their data were included in the survival analysis, and four patients (4 hips) were lost to follow-up. The clinical outcome, proximal bone regeneration and subsidence were assessed for 101 hips.

The mean Harris Hip Score was 88.2 (45.8 to 100) after five years and there was an increase of the mean Barnett and Nordin-Score, a measure of the proximal bone regeneration, of 20.8 (-3.1 to 52.7). Five stems had to be revised (4.3%), three (2.9%) showed subsidence, five (4.3%) a dislocation and two of 85 aseptic revisions (2.3%) a periprosthetic infection.

At the latest follow-up, the survival with revision of the stem as the endpoint was 95.7% (95% confidence interval 91.9% to 99.4%) and with aseptic loosening as the endpoint, was 100%. Peri-prosthetic fractures were not observed.

We report excellent results with respect to subsidence, the risk of fracture, and loosening after femoral revision using a modular curved revision stem with distal cone-in-cone fixation. A successful outcome depends on careful pre-operative planning and the use of a transfemoral approach when the anatomy is distorted or a fracture is imminent, or residual cement or a partially-secured existing stem cannot be removed. The shortest appropriate stem should, in our opinion, be used and secured with > 3 cm fixation at the femoral isthmus, and distal interlocking screws should be used for additional stability when this goal cannot be realised.

Cite this article: Bone Joint J 2014;96-B:889–95.


Aims

This study aimed to compare the change in health-related quality of life of patients receiving a traditional cemented monoblock Thompson hemiarthroplasty compared with a modern cemented modular polished-taper stemmed hemiarthroplasty for displaced intracapsular hip fractures.

Patients and Methods

This was a pragmatic, multicentre, multisurgeon, two-arm, parallel group, randomized standard-of-care controlled trial. It was embedded within the WHiTE Comprehensive Cohort Study. The sample size was 964 patients. The setting was five National Health Service Trauma Hospitals in England. A total of 964 patients over 60 years of age who required hemiarthroplasty of the hip between February 2015 and March 2016 were included. A standardized measure of health outcome, the EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and at four months following the operation.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 43 - 47
1 Jan 2014
Craig P Bancroft G Burton A Collier S Shaylor P Sinha A

The issues surrounding raised levels of metal ions in the blood following large head metal-on-metal total hip replacement (THR), such as cobalt and chromium, have been well documented. Despite the national popularity of uncemented metal-on-polyethylene (MoP) THR using a large-diameter femoral head, few papers have reported the levels of metal ions in the blood following this combination. Following an isolated failure of a 44 mm Trident–Accolade uncemented THR associated with severe wear between the femoral head and the trunnion in the presence of markedly elevated levels of cobalt ions in the blood, we investigated the relationship between modular femoral head diameter and the levels of cobalt and chromium ions in the blood following this THR.

A total of 69 patients received an uncemented Trident–Accolade MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean age 67.0 years) were recruited and had levels of cobalt and chromium ions in the blood measured between May and June 2012. The patients were then divided into three groups according to the diameter of the femoral head used: 12 patients in the 28 mm group (controls), 18 patients in the 36 mm group and 13 patients in the 40 mm group. A total of four patients had identical bilateral prostheses in situ at phlebotomy: one each in the 28 mm and 36 mm groups and two in the 40 mm group.

There was a significant increase in the mean levels of cobalt ions in the blood in those with a 36 mm diameter femoral head compared with those with a 28 mm diameter head (p = 0.013). The levels of cobalt ions in the blood were raised in those with a 40 mm diameter head but there was no statistically significant difference between this group and the control group (p = 0.152). The levels of chromium ions in the blood were normal in all patients.

The clinical significance of this finding is unclear, but we have stopped using femoral heads with a diameter of ≤ 36 mm, and await further larger studies to clarify whether, for instance, this issue particularly affects this combination of components.

Cite this article: Bone Joint J 2014;96-B:43–7.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 121 - 129
1 Mar 2024
Orce Rodríguez A Smith PN Johnson P O'Sullivan M Holder C Shimmin A

Aims. In recent years, the use of a collared cementless femoral prosthesis has risen in popularity. The design intention of collared components is to transfer some load to the resected femoral calcar and prevent implant subsidence within the cancellous bone of the metaphysis. Conversely, the load transfer for a cemented femoral prosthesis depends on the cement-component and cement-bone interface interaction. The aim of our study was to compare the three most commonly used collared cementless components and the three most commonly used tapered polished cemented components in patients aged ≥ 75 years who have undergone a primary total hip arthroplasty (THA) for osteoarthritis (OA). Methods. Data from the Australian Orthopaedic Association National Joint Replacement Registry from 1 September 1999 to 31 December 2022 were analyzed. Collared cementless femoral components and cemented components were identified, and the three most commonly used components in each group were analyzed. We identified a total of 11,278 collared cementless components and 47,835 cemented components. Hazard ratios (HRs) from Cox proportional hazards models, adjusting for age and sex, were obtained to compare the revision rates between the groups. Results. From six months postoperatively onwards, patients aged ≥ 75 years undergoing primary THA with primary diagnosis of OA have a lower risk of all-cause revision with collared cementless components than with a polished tapered cemented component (HR 0.78 (95% confidence interval 0.64 to 0.96); p = 0.018). There is no difference in revision rate prior to six months. Conclusion. Patients aged ≥ 75 years with a primary diagnosis of OA have a significantly lower rate of revision with the most common collared cementless femoral component, compared with the most common polished tapered cemented components from six months postoperatively onwards. The lower revision rate is largely due to a reduction in revisions for fracture and infection. Cite this article: Bone Joint J 2024;106-B(3 Supple A):121–129


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 31 - 37
1 Mar 2024
Bunyoz KI Tsikandylakis G Mortensen K Gromov K Mohaddes M Malchau H Troelsen A

Aims. In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients. Methods. A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively. Results. At five years, the median levels of chromium, cobalt, and titanium were 0.5 μg/l (interquartile range (IQR) 0.50 to 0.62), 0.24 μg/l (IQR 0.18 to 0.30), and 1.16 μg/l (IQR 1.0 to 1.68) for the 32 mm group, and 0.5 μg/l (IQR 0.5 to 0.54), 0.23 μg/l (IQR 0.17 to 0.39), and 1.30 μg/l (IQR 1 to 2.05) for the 36 mm to 44 mm group, with no significant difference between the groups (p = 0.825, p = 1.000, p = 0.558). There were increased levels of metal ions at two years postoperatively in seven patients in the 32 mm group, compared with four in the 36 mm to 44 mm group, and at five years postoperatively in six patients in the 32 mm group, compared with seven in the 36 mm to 44 mm group. There was no significant difference in either the OHS (p = 0.665) or UCLA (p = 0.831) scores between patients with or without an increased level of metal ions. Conclusion. In patients who underwent MoP THA, we found no differences in the levels of metal ions five years postoperatively between those with a femoral head of 32 mm and those with a femoral head of between 36 mm and 44 mm, and no corrosion-related revisions. As taper corrosion can start after five years, there remains a need for longer-term studies investigating the relationship between the size of the femoral head size and corrosion in patients undergoing MoP THA. Cite this article: Bone Joint J 2024;106-B(3 Supple A):31–37


The Bone & Joint Journal
Vol. 105-B, Issue 11 | Pages 1196 - 1200
1 Nov 2023
Parker MJ Chatterjee R Onsa M Cawley S Gurusamy K

Aims. The aim of this study was to report the three-year follow-up for a series of 400 patients with a displaced intracapsular fracture of the hip, who were randomized to be treated with either a cemented polished tapered hemiarthroplasty or an uncemented hydroxyapatite-coated hemiarthroplasty. Methods. The mean age of the patients was 85 years (58 to 102) and 273 (68%) were female. Follow-up was undertaken by a nurse who was blinded to the hemiarthroplasty that was used, at intervals for up to three years from surgery. The short-term follow-up of these patients at a mean of one year has previously been reported. Results. A total of 210 patients (52.5%) died within three years of surgery. One patient was lost to follow-up. Recovery of mobility was initially significantly better in those treated with a cemented hemiarthroplasty, although by three years after surgery this difference became statistically insignificant. The mortality was significantly lower in those treated with a cemented hemiarthroplasty (p = 0.029). There was no significant difference in pain scores, or in the incidence of implant-related complications or revision surgery, between the two groups. Conclusion. These results further support the use of a cemented hemiarthroplasty for the routine management of elderly patients with a displaced intracapsular fracture of the hip. Cite this article: Bone Joint J 2023;105-B(11):1196–1200


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 911 - 916
1 Jul 2015
Del Balso C Teeter MG Tan SC Lanting BA Howard JL

Tribocorrosion at the head–neck taper interface – so-called ‘taperosis’ – may be a source of metal ions and particulate debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and corrosion in retrieved head–neck tapers in vivo for a minimum of two years (mean 8.7 years; 2.6 to 15.9). A total of 56 femoral heads ranging from 28 mm to 3 mm to 28 mm + 8 mm, and 17 femoral stems featuring a single taper design were included in the study. Fretting and corrosion were scored in three horizontally oriented concentric zones of each taper by stereomicroscopy. Head length was observed to affect fretting (p = 0.03), with 28 mm + 8 mm femoral heads showing greater total fretting scores than all other head lengths. The central zone of the femoral head bore taper was subject to increased fretting damage (p = 0.01), regardless of head length or stem offset. High-offset femoral stems were associated with greater total fretting of the bore taper (p = 0.04). Increased fretting damage is seen with longer head lengths and high-offset femoral stems, and occurs within a central concentric zone of the femoral head bore taper. Further investigation is required to determine the effect of increased head size, and variations in head–neck taper design. Cite this article: Bone Joint J 2015; 97-B:911–16


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1238 - 1246
1 Jul 2021
Hemmerling KJ Weitzler L Bauer TW Padgett DE Wright TM

Aims. Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. Methods. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy. Results. Fretting was present on 88% (53/60) of the retrieved liners, and corrosion was present on 97% (58/60). Fretting was most often found on the lip of the taper at the transition between the lip and the dome regions. Macrophages and particles reflecting an innate inflammatory reaction to corrosion debris were noted in six of the 48 cases for which periprosthetic tissues were examined, and all were associated with retrieved components that had high corrosion scores. Conclusion. Our results show that corrosion occurs at the interface between MDM liners and shells and that it can be associated with reactions in the local tissues, suggesting continued concern that this problem may become clinically important with longer-term use of these implants. Cite this article: Bone Joint J 2021;103-B(7):1238–1246


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy JW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims. Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. Methods. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment. Results. A total of 97 patients matched the inclusion criteria (50 Exeter and 47 C-Stem AMT components). There were no significant differences between the patient demographic data in either group. Mean follow-up was 9.7 years. A significant improvement in Oxford Hip Score (OHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 12-item Short-Form Survey (SF-12) scores was observed in both cohorts. Leg lengths were significantly shorter in the Exeter group, with a mean of -4 mm in this cohort compared with 0 mm in the C-Stem AMT group. One patient in the Exeter group had early evidence of radiological loosening. In total, 16 patients (15%) underwent further revision of the femoral component (seven in the C-Stem AMT group and nine in the Exeter group). No femoral components were revised for aseptic loosening. There were two cases of femoral component fracture in the Exeter group. Conclusion. Our series shows promising mid-term outcomes for the cement-in-cement revision technique using either the Exeter or C-Stem AMT components. These results demonstrate that cement-in-cement revision using a double or triple taper-slip design is a safe and reliable technique when used for the correct indications. Cite this article: Bone Joint J 2021;103-B(7):1215–1221


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 386 - 389
1 Apr 2019
Hampton C Weitzler L Baral E Wright TM Bostrom MPG

Aims. The aim of this study was to evaluate fretting and corrosion in retrieved oxidized zirconium (OxZr; OXINIUM, Smith & Nephew, Memphis, Tennessee) femoral heads and compare the results with those from a matched cohort of cobalt-chromium (CoCr) femoral heads. Patients and Methods. A total of 28 OxZr femoral heads were retrieved during revision total hip arthroplasty (THA) and matched to 28 retrieved CoCr heads according to patient demographics. The mean age at index was 56 years (46 to 83) in the OxZr group and 70 years (46 to 92) in the CoCr group. Fretting and corrosion scores of the female taper of the heads were measured according to the modified Goldberg scoring method. Results. The OxZr-retrieved femoral heads showed significantly lower mean corrosion scores than the CoCr heads (1.3 (1 to 2.75) vs 2.1 (1 to 4); p < 0.01). Mean fretting scores were also significantly lower in the OxZr cohort when compared with the CoCr cohort (1.3 (1 to 2) vs 1.5 (1 to 2.25); p = 0.02). OxZr heads had more damage in the proximal region compared with the distal region of the head. Location had no impact on damage of CoCr heads. A trend towards increased corrosion in large heads was seen only in the CoCr heads, although this was not statistically significant. Conclusion. Retrieval analysis of OxZr femoral heads showed a decreased amount of fretting and corrosion compared with CoCr femoral heads. OxZr seems to be effective at reducing taper damage. Cite this article: Bone Joint J 2019;101-B:386–389


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 3 - 6
1 Nov 2013
Wassef AJ Schmalzried TP

A modular femoral head–neck junction has practical advantages in total hip replacement. Taper fretting and corrosion have so far been an infrequent cause of revision. The role of design and manufacturing variables continues to be debated. Over the past decade several changes in technology and clinical practice might result in an increase in clinically significant taper fretting and corrosion. Those factors include an increased usage of large diameter (36 mm) heads, reduced femoral neck and taper dimensions, greater variability in taper assembly with smaller incision surgery, and higher taper stresses due to increased patient weight and/or physical activity. Additional studies are needed to determine the role of taper assembly compared with design, manufacturing and other implant variables. . Cite this article: Bone Joint J 2013;95-B, Supple A:3–6


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 309 - 320
1 Feb 2021
Powell-Bowns MFR Oag E Ng N Pandit H Moran M Patton JT Clement ND Scott CEH

Aims. The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). Methods. This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality. Results. Fractures (B1 n = 74 (49%); B2 n = 50 (33%); and B3 n = 28 (18%)) occurred at median of 4.2 years (interquartile range (IQR) 1.2 to 9.2) after primary total hip arthroplasty (THA) (n = 138) or hemiarthroplasty (n = 14). Rates of revision and reoperation were significantly higher following revision arthroplasty compared to ORIF for B2 (p = 0.001) and B3 fractures (p = 0.050). Five-year survival was significantly better following ORIF: 92% (95% confidence interval (CI) 86.4% to 97.4%) versus 63% (95% CI 41.7% to 83.3%), p < 0.001. ORIF was associated with reduced blood transfusion requirement and reoperations, but there were no differences in medical complications, hospital stay, or mortality between surgical groups. No independent predictors of revision following ORIF were identified: where the bone-cement interface was intact, fixation of B2 or B3 fractures was not associated with an increased risk of revision. Conclusion. When the bone-cement interface was intact and the fracture was anatomically reducible, all Vancouver B fractures around Exeter stems could be managed with fixation as opposed to revision arthroplasty. Fixation was associated with reduced need for blood transfusion and lower risk of revision surgery compared with revision arthroplasty. Cite this article: Bone Joint J 2021;103-B(2):309–320


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims. Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA. Methods. This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m. 2. , and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported. Results. A total of 43 patients were enrolled. At minimum follow-up of five years, 23 (53.4%) returned for clinical and radiological review, 25 (58.1%) had metal ion analysis performed, 19 (44.2%) underwent dual energy x-ray absorptiometry scans, and 25 (58%) completed a pain-drawing questionnaire. The mean modified Harris Hip Scores improved significantly from 54.8 (SD 19) preoperatively to 93.08 (SD 10.5) five years postoperatively (p < 0.001). One patient was revised for aseptic acetabular loosening. The mean cobalt levels increased from 0.065 μg/l (SD 0.03) to 0.08 (SD 0.05) and the mean titanium levels increased from 0.35 (SD 0.13) to 0.78 (SD 0.29). The femoral BMD ratio decreased in Gruen Zone 1 (91.9%) at five years postoperatively compared with the baseline scores at six weeks potoperatively. The femoral BMD ratio was maintained in Gruen zones 2 to 7. Conclusion. The use of a modular DM component and a cementless, tapered femoral stem shows excellent mid-term survivorship with minimal concerns for corrosion and metal ion release in a cohort of young, active patients undergoing primary THA. Cite this article: Bone Joint J 2021;103-B(7 Supple B):73–77


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 365 - 371
1 Apr 2019
Nam D Salih R Nahhas CR Barrack RL Nunley RM

Aims. Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell (vs a monoblock acetabular component) have the advantage of allowing supplementary screw fixation, but the potential for corrosion between the liner and acetabulum has raised concerns. While DM prostheses have shown improved stability in patients deemed ‘high-risk’ for dislocation undergoing total hip arthroplasty (THA), their performance in young, active patients has not been reported. This study’s purpose was to assess clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) in young, active patients receiving a modular DM acetabulum and recently introduced titanium, proximally coated, tapered femoral stem design. Patients and Methods. This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m. 2. and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (. sd. 6.5)) were enrolled. All patients had a minimum of two years’ clinical follow-up. Patient-reported outcome measures, whole blood metal ion levels (ug/l), and periprosthetic femoral BMD were measured at baseline, as well as at one and two years postoperatively. Power analysis indicated 40 patients necessary to demonstrate a five-fold increase in cobalt levels from baseline (alpha = 0.05, beta = 0.80). A mixed model with repeated measures was used for statistical analysis. Results. Mean Harris Hip Scores improved from 54.1 (. sd. 20.5) to 91.2 (. sd. 10.8) at two years postoperatively (p < 0.001). All patients had radiologically well-fixed components, no patients experienced any instability, and no patients required any further intervention. Mean cobalt levels increased from 0.065 ug/l (. sd. 0.03) preoperatively to 0.30 ug/l (. sd. 0.51) at one year postoperatively (p = 0.01) but decreased at two years postoperatively to 0.16 ug/l (. sd. 0.23; p = 0.2). Four patients (9.3%) had a cobalt level outside the reference range (0.03 ug/l to 0.29 ug/l) at two years postoperatively, with values from 0.32 ug/l to 0.94 ug/l. The mean femoral BMD ratio was maintained in Gruen zones 2 to 7 at both one and two years postoperatively using this stem design. At two years postoperatively, mean BMD in the medial calcar was 101.5% of the baseline value. Conclusion. Use of a modular DM prosthesis and cementless, tapered femoral stem has shown encouraging results in young, active patients undergoing primary THA. Elevation in mean cobalt levels and the presence of four patients outside the reference range at two years postoperatively demonstrates the necessity of continued surveillance in this cohort. Cite this article: Bone Joint J 2019;101-B:365–371


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 579 - 584
1 May 2016
Osman K Panagiotidou AP Khan M Blunn G Haddad FS

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface. We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice. Cite this article: Bone Joint J 2016;98-B:579–84


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1565 - 1571
1 Dec 2018
Kristensen TB Dybvik E Furnes O Engesæter LB Gjertsen J

Aims. The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck. Patients and Methods. From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan–Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis. Results. The one-year survival was 96.0% (95% confidence interval (CI) 95.6 to 96.4) for the Exeter stem, 97.0% (95% CI 96.4 to 97.6) for the Lubinus SP2 stem, 97.6% (95% CI 97.0 to 98.2) for the Charnley stem, 98.1% (95% CI 97.3 to 98.9) for the Spectron EF stem, and 96.4% (95% CI 95.6 to 97.2) for the Charnley Modular stem, respectively. The hazard ratio for reoperation after one year was lower for Lubinus SP2 (HR 0.77, 95% CI 0.60 to 0.97), Charnley (HR 0.64, 95% CI 0.48 to 0.86), and Spectron EF stems (HR 0.44, 95% CI 0.29 to 0.67) compared with the Exeter stem. Reoperation for periprosthetic fracture occurred almost exclusively after the use of polished tapered stems. Conclusion. We were able to confirm that implant survival after cemented hemiarthroplasty for a hip fracture is high. Differences in rates of reoperation seem to favour anatomical and straight stems compared with polished tapered stems, which had a higher risk of periprosthetic fracture


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 95 - 97
1 Nov 2013
Cross MB Paprosky WG

If a surgeon is faced with altered lesser trochanter anatomy when revising the femoral component in revision total hip replacement, a peri-prosthetic fracture, or Paprosky type IIIb or type IV femoral bone loss, a modular tapered stem offers the advantages of accurately controlling femoral version and length. The splines of the taper allow rotational control, and improve the fit in femoral canals with diaphyseal bone loss. In general, two centimetres of diaphyseal contact is all that is needed to gain stability with modular tapered stems. By allowing the proximal body trial to rotate on a well-fixed distal segment during trial reduction, appropriate anteversion can be obtained in order to improve intra-operative stability, and decrease the dislocation risk. However, modular stems should not be used for all femoral revisions, as implant fracture and corrosion at modular junctions can still occur. Cite this article: Bone Joint J 2013;95-B, Supple A:95–7


The Bone & Joint Journal
Vol. 100-B, Issue 1_Supple_A | Pages 44 - 49
1 Jan 2018
Berstock JR Whitehouse MR Duncan CP

Aims. To present a surgically relevant update of trunnionosis. . Materials and Methods. Systematic review performed April 2017. Results. Trunnionosis accounts for approximately 2% of the revision total hip arthroplasty (THA) burden. Thinner (reduced flexural rigidity) and shorter trunnions (reduced contact area at the taper junction) may contribute to mechanically assisted corrosion, exacerbated by high offset implants. The contribution of large heads and mixed metallurgy is discussed. Conclusion. Identifying causative risk factors is challenging due to the multifactorial nature of this problem. Cite this article: Bone Joint J 2018;100-B(1 Supple A):44–9


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 11 - 16
1 Jan 2020
Parker MJ Cawley S

Aims. Debate continues about whether it is better to use a cemented or uncemented hemiarthroplasty to treat a displaced intracapsular fracture of the hip. The aim of this study was to attempt to resolve this issue for contemporary prostheses. Methods. A total of 400 patients with a displaced intracapsular fracture of the hip were randomized to receive either a cemented polished tapered stem hemiarthroplasty or an uncemented Furlong hydroxyapatite-coated hemiarthroplasty. Follow-up was conducted by a nurse blinded to the implant at set intervals for up to one year from surgery. Results. A total of 115 patients died in the year after surgery. There was a tendency towards a slightly higher mortality in those treated with the uncemented prosthesis after one year (64 vs 51; p = 0.18). For the survivors, there was no significant difference in pain score at any of the time intervals. Patients treated using the cemented hemiarthroplasty recovered mobility better than those treated with the uncemented hemiarthroplasty (mean decrease in mobility score at one year: 1.7 vs 1.1, SD 1.9; p = 0.008). There was a tendency to more periprosthetic fractures in the uncemented group (five vs two cases; p = 0.45), but overall the need for further surgery was similar in both groups (nine vs seven cases). There were four perioperative deaths in the cemented group. Conclusion. These results indicate that a contemporary cemented hemiarthroplasty gives better results than an uncemented hemiarthroplasty for patients with a displaced intracapsular fracture of the hip. When the condition of the patient permits, a cemented hemiarthroplasty should be used. Cite this article: Bone Joint J. 2020;102-B(1):11–16


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 112 - 117
1 May 2024
Hickie KL Neufeld ME Howard LC Greidanus NV Masri BA Garbuz DS

Aims

There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes.

Methods

We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR system from January 2000 to December 2007. We included 106 patients (108 hips) with a mean follow-up of 14.5 years (2.3 to 22.3). Mean patient age was 69.2 years (37.0 to 89.4), and 51.9% were female (n = 55). Indications for index revision included aseptic loosening (73.1%), infection (16.7%), fracture (9.3%), and stem fracture (0.9%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. At most recent follow-up, Oxford Hip Scores (OHS) were collected, and radiological stem stability was determined using the Engh classification.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 89 - 96
1 Mar 2024
Heckmann ND Chung BC Liu KC Chen XT Lovro LR Kistler NM White E Christ AB Longjohn DB Oakes DA Lieberman JR

Aims

Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations.

Methods

A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 895 - 900
1 Jul 2012
Gill IPS Webb J Sloan K Beaver RJ

We present a series of 35 patients (19 men and 16 women) with a mean age of 64 years (36.7 to 75.9), who underwent total hip replacement using the ESKA dual-modular short stem with metal on-polyethylene bearing surfaces. This implant has a modular neck section in addition to the modular head. Of these patients, three presented with increasing post-operative pain due to pseudotumour formation that resulted from corrosion at the modular neck-stem junction. These patients underwent further surgery and aseptic lymphocytic vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the neck-stem taper. Blood cobalt and chromium levels were measured at a mean of nine months (3 to 28) following surgery. These were compared with the levels in seven control patients (three men and four women) with a mean age of 53.4 years (32.1 to 64.1), who had an identical prosthesis and articulation but with a prosthesis that had no modularity at neck-stem junction. The mean blood levels of cobalt in the study group were raised at 50.75 nmol/l (5 to 145) compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important source of metal ion release and pseudotumour formation requiring revision surgery. Finite element modelling of the dual modular stem demonstrated high stresses at the modular stem-neck junction. Dual modular cobalt-chrome hip prostheses should be used with caution due to these concerns


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 515 - 515
1 May 2024
Kayani B D. Luo T S. Haddad F


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 935 - 940
1 Jul 2010
McCalden RW Charron KD Yuan X Bourne RB Naudie DD MacDonald SJ

This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless, tapered, polished cemented stems, the only difference being a slight lateral to medial taper with the CPCS stem. A total of 34 patients were enrolled in a single-blinded randomised controlled trial in which 17 patients received a dedicated radiostereometric CPCS stem and 17 a radiostereometric Exeter stem. No difference was found in any of the outcome measures pre-operatively or post-operatively between groups. At two years, the mean subsidence for the CPCS stem was nearly half that seen for the Exeter stem (0.77 mm (−0.943 to 1.77) and 1.25 mm (0.719 to 1.625), respectively; p = 0.032). In contrast, the mean internal rotation of the CPCS stem was approximately twice that of the Exeter (1.61° (−1.07° to 4.33°) and 0.59° (0.97° to 1.64°), respectively; p = 0.048). Other migration patterns were not significantly different between the stems. The subtle differences in designs may explain the different patterns of migration. Comparable migration with the Exeter stem suggests that the CPCS design will perform well in the long term


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 879 - 883
1 Sep 2024
Kayani B Staats K Haddad FS


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 115 - 120
1 Mar 2024
Ricotti RG Flevas DA Sokrab R Vigdorchik JM Mayman DJ Jerabek SA Sculco TP Sculco PK

Aims

Periprosthetic femoral fracture (PPF) is a major complication following total hip arthroplasty (THA). Uncemented femoral components are widely preferred in primary THA, but are associated with higher PPF risk than cemented components. Collared components have reduced PPF rates following uncemented primary THA compared to collarless components, while maintaining similar prosthetic designs. The purpose of this study was to analyze PPF rate between collarless and collared component designs in a consecutive cohort of posterior approach THAs performed by two high-volume surgeons.

Methods

This retrospective series included 1,888 uncemented primary THAs using the posterior approach performed by two surgeons (PKS, JMV) from January 2016 to December 2022. Both surgeons switched from collarless to collared components in mid-2020, which was the only change in surgical practice. Data related to component design, PPF rate, and requirement for revision surgery were collected. A total of 1,123 patients (59.5%) received a collarless femoral component and 765 (40.5%) received a collared component. PPFs were identified using medical records and radiological imaging. Fracture rates between collared and collarless components were analyzed. Power analysis confirmed 80% power of the sample to detect a significant difference in PPF rates, and a Fisher’s exact test was performed to determine an association between collared and collarless component use on PPF rates.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 548 - 554
1 Jun 2024
Ohyama Y Minoda Y Masuda S Sugama R Ohta Y Nakamura H

Aims

The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group).

Methods

The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 1 - 2
1 May 2024
Berry DJ Haddad FS


The Bone & Joint Journal
Vol. 106-B, Issue 1 | Pages 11 - 15
1 Jan 2024
Jain S Lamb JN Pandit H

Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF.

Cite this article: Bone Joint J 2024;106-B(1):11–15.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 593 - 601
1 Jun 2023
Scott CEH Yapp LZ Howard T Patton JT Moran M

Periprosthetic femoral fractures are increasing in incidence, and typically occur in frail elderly patients. They are similar to pathological fractures in many ways. The aims of treatment are the same, including 'getting it right first time' with a single operation, which allows immediate unrestricted weightbearing, with a low risk of complications, and one that avoids the creation of stress risers locally that may predispose to further peri-implant fracture. The surgical approach to these fractures, the associated soft-tissue handling, and exposure of the fracture are key elements in minimizing the high rate of complications. This annotation describes the approaches to the femur that can be used to facilitate the surgical management of peri- and interprosthetic fractures of the femur at all levels using either modern methods of fixation or revision arthroplasty.

Cite this article: Bone Joint J 2023;105-B(6):593–601.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 468 - 474
1 May 2024
d'Amato M Flevas DA Salari P Bornes TD Brenneis M Boettner F Sculco PK Baldini A

Aims

Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis).

Methods

Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 736 - 740
1 Jun 2016
Sassoon AA Barrack RL

The use of large-diameter metal-on-metal (MoM) components in total hip arthroplasty (THA) is associated with an increased risk of early failure due to adverse local tissue reaction to metal debris (ARMD) in response to the release of metal ions from the bearing couple and/or head-neck taper corrosion. The aim of this paper was to present a review of the incidence and natural history of ARMD and the forms of treatment, with a focus on the need for and extent of resection or debulking of the pseudotumour. An illustrative case report is presented of a patient with an intra-pelvic pseudotumour associated with a large diameter MoM THA, which was treated successfully with revision of the bearing surface to a dual mobility couple and retention of the well-fixed acetabular and femoral components. The pseudotumour was left in situ. Resolution of the intra-pelvic mass and normalisation of metal ion levels was observed seven months post-operatively. Cite this article: Bone Joint J 2016;98-B:736–40


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1020 - 1027
1 Aug 2017
Matharu GS Judge A Pandit HG Murray DW

Aims. To determine the outcomes following revision surgery of metal-on-metal hip arthroplasties (MoMHA) performed for adverse reactions to metal debris (ARMD), and to identify factors predictive of re-revision. Patients and Methods. We performed a retrospective observational study using National Joint Registry (NJR) data on 2535 MoMHAs undergoing revision surgery for ARMD between 2008 and 2014. The outcomes studied following revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using competing-risk regression modelling. Results. Intra-operative complications occurred in 40 revisions (1.6%). The cumulative five-year patient survival rate was 95.9% (95% confidence intervals (CI) 92.3 to 97.8). Re-revision surgery was performed in 192 hips (7.6%). The cumulative five-year implant survival rate was 89.5% (95% CI 87.3 to 91.3). Predictors of re-revision were high body mass index at revision (subhazard ratio (SHR) 1.06 per kg/m. 2 . increase, 95% CI 1.02 to 1.09), modular component only revisions (head and liner with or without taper adapter; SHR 2.01, 95% CI 1.19 to 3.38), ceramic-on-ceramic revision bearings (SHR 1.86, 95% CI 1.23 to 2.80), and acetabular bone grafting (SHR 2.10, 95% CI 1.43 to 3.07). These four factors remained predictive of re-revision when the missing data were imputed. Conclusion. The short-term risk of re-revision following MoMHA revision surgery performed for ARMD was comparable with that reported in the NJR following all-cause non-MoMHA revision surgery. However, the factors predictive of re-revision included those which could be modified by the surgeon, suggesting that rates of failure following ARMD revision may be reduced further. Cite this article: Bone Joint J 2017;99-B:1020–7


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 50 - 59
1 Jan 2017
Carli AV Negus JJ Haddad FS

Aims. Periprosthetic femoral fractures (PFF) following total hip arthroplasty (THA) are devastating complications that are associated with functional limitations and increased overall mortality. Although cementless implants have been associated with an increased risk of PFF, the precise contribution of implant geometry and design on the risk of both intra-operative and post-operative PFF remains poorly investigated. A systematic review was performed to aggregate all of the PFF literature with specific attention to the femoral implant used. Patients and Methods. A systematic search strategy of several journal databases and recent proceedings from the American Academy of Orthopaedic Surgeons was performed. Clinical articles were included for analysis if sufficient implant description was provided. All articles were reviewed by two reviewers. A review of fundamental investigations of implant load-to-failure was performed, with the intent of identifying similar conclusions from the clinical and fundamental literature. Results. In total 596 articles were initially identified, with 34 being eligible for analysis. Aggregate analysis of 1691 PFFs in 342 719 primary THAs revealed a significantly higher number of PFFs with cementless femoral implants (p < 0.001). Single-wedge and double-wedge (fit-and-fill) femoral implants were associated with a threefold increase in PFF rates (p < 0.001) compared with anatomical, fully coated and tapered/rounded stems. Within cemented stems, loaded-taper (Exeter) stems were associated with more PFFs than composite-beam (Charnley) stems (p = 0.004). Review of the fundamental literature revealed very few studies comparing cementless component designs. Conclusion. Very few studies within the PFF literature provide detailed implant information. Cementless implants, specifically those of single-wedge and double-wedge, have the highest PFF rates in the literature, with most investigations recommending against their use in older patients with osteoporotic bone. This review illustrates the need for registries and future PFF studies to record implant name and information for future analysis. Furthermore, future biomechanical investigations comparing modern implants are needed to clarify the precise contribution of implant design to PFF risk. Cite this article: Bone Joint J 2017;99-B(1 Supple A):50–9


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 864 - 871
1 Aug 2023
Tyas B Marsh M de Steiger R Lorimer M Petheram TG Inman DS Reed MR Jameson SS

Aims

Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty.

Methods

Patients recorded by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) between 1 September 1999 and 31 December 2020 who underwent hemiarthroplasty for the treatment of a hip fracture with the following implants were included: a cemented monoblock Exeter Trauma Stem (ETS), cemented Exeter V40 with a bipolar head, a monoblock Thompsons prosthesis (Cobalt/Chromium or Titanium), and an Exeter V40 with a Unitrax head. Overall and age-defined cumulative revision rates were compared over the ten years following surgery.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 352 - 358
1 Apr 2024
Wilson JM Trousdale RT Bedard NA Lewallen DG Berry DJ Abdel MP

Aims

Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct.

Methods

We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 104 - 109
1 Mar 2024
Sugano N Maeda Y Fuji H Tamura K Nakamura N Takashima K Uemura K Hamada H

Aims

Femoral component anteversion is an important factor in the success of total hip arthroplasty (THA). This retrospective study aimed to investigate the accuracy of femoral component anteversion with the Mako THA system and software using the Exeter cemented femoral component, compared to the Accolade II cementless femoral component.

Methods

We reviewed the data of 30 hips from 24 patients who underwent THA using the posterior approach with Exeter femoral components, and 30 hips from 24 patients with Accolade II components. Both groups did not differ significantly in age, sex, BMI, bone quality, or disease. Two weeks postoperatively, CT images were obtained to measure acetabular and femoral component anteversion.


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1032 - 1038
1 Sep 2022
García-Rey E Cruz-Pardos A Saldaña L

Aims

A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade.

Methods

Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 1000 - 1007
1 Sep 2024
Gong T Lu M Sheng H Li Z Zhou Y Luo Y Min L Tu C

Aims

Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes.

Methods

This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 1000 - 1006
1 Sep 2023
Macken AA Haagmans-Suman A Spekenbrink-Spooren A van Noort A van den Bekerom MPJ Eygendaal D Buijze GA

Aims

The current evidence comparing the two most common approaches for reverse total shoulder arthroplasty (rTSA), the deltopectoral and anterosuperior approach, is limited. This study aims to compare the rate of loosening, instability, and implant survival between the two approaches for rTSA using data from the Dutch National Arthroplasty Registry with a minimum follow-up of five years.

Methods

All patients in the registry who underwent a primary rTSA between January 2014 and December 2016 using an anterosuperior or deltopectoral approach were included, with a minimum follow-up of five years. Cox and logistic regression models were used to assess the association between the approach and the implant survival, instability, and glenoid loosening, independent of confounders.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 144 - 150
1 Feb 2024
Lynch Wong M Robinson M Bryce L Cassidy R Lamb JN Diamond O Beverland D

Aims

The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component.

Methods

We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 17 - 23
1 Mar 2024
LaValva SM LeBrun DG Canoles HG Ren R Padgett DE Su EP

Aims

Professional dancers represent a unique patient population in the setting of hip arthroplasty, given the high degree of hip strength and mobility required by their profession. We sought to determine the clinical outcomes and ability to return to professional dance after total hip arthroplasty (THA) or hip resurfacing arthroplasty (HRA).

Methods

Active professional dancers who underwent primary THA or HRA at a single institution with minimum one-year follow-up were included in the study. Primary outcomes included the rate of return to professional dance, three patient-reported outcome measures (PROMs) (modified Harris Hip Score (mHHS), Hip disability and Osteoarthritis Outcome Score for Joint Replacement (HOOS-JR), and Lower Extremity Activity Scale (LEAS)), and postoperative complications.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 540 - 547
1 Jun 2024
Nandra RS Elnahal WA Mayne A Brash L McBryde CW Treacy RBC

Aims

The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years.

Methods

The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.


The Bone & Joint Journal
Vol. 104-B, Issue 12 | Pages 1292 - 1303
1 Dec 2022
Polisetty TS Jain S Pang M Karnuta JM Vigdorchik JM Nawabi DH Wyles CC Ramkumar PN

Literature surrounding artificial intelligence (AI)-related applications for hip and knee arthroplasty has proliferated. However, meaningful advances that fundamentally transform the practice and delivery of joint arthroplasty are yet to be realized, despite the broad range of applications as we continue to search for meaningful and appropriate use of AI. AI literature in hip and knee arthroplasty between 2018 and 2021 regarding image-based analyses, value-based care, remote patient monitoring, and augmented reality was reviewed. Concerns surrounding meaningful use and appropriate methodological approaches of AI in joint arthroplasty research are summarized. Of the 233 AI-related orthopaedics articles published, 178 (76%) constituted original research, while the rest consisted of editorials or reviews. A total of 52% of original AI-related research concerns hip and knee arthroplasty (n = 92), and a narrative review is described. Three studies were externally validated. Pitfalls surrounding present-day research include conflating vernacular (“AI/machine learning”), repackaging limited registry data, prematurely releasing internally validated prediction models, appraising model architecture instead of inputted data, withholding code, and evaluating studies using antiquated regression-based guidelines. While AI has been applied to a variety of hip and knee arthroplasty applications with limited clinical impact, the future remains promising if the question is meaningful, the methodology is rigorous and transparent, the data are rich, and the model is externally validated. Simple checkpoints for meaningful AI adoption include ensuring applications focus on: administrative support over clinical evaluation and management; necessity of the advanced model; and the novelty of the question being answered.

Cite this article: Bone Joint J 2022;104-B(12):1292–1303.


The Bone & Joint Journal
Vol. 104-B, Issue 8 | Pages 987 - 996
1 Aug 2022

Aims

The aim of this study was to describe the demographic details of patients who sustain a femoral periprosthetic fracture (PPF), the epidemiology of PPFs, PPF characteristics, and the predictors of PPF types in the UK population.

Methods

This is a multicentre retrospective cohort study including adult patients presenting to hospital with a new PPF between 1 January 2018 and 31 December 2018. Data collected included: patient characteristics, comorbidities, anticoagulant use, social circumstances, level of mobility, fracture characteristics, Unified Classification System (UCS) type, and details of the original implant. Descriptive analysis by fracture location was performed, and predictors of PPF type were assessed using mixed-effects logistic regression models.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 451 - 457
1 Apr 2017
Gromov K Bersang A Nielsen CS Kallemose T Husted H Troelsen A

Aims. The aim of this study was to identify patient- and surgery-related risk factors for sustaining an early periprosthetic fracture following primary total hip arthroplasty (THA) performed using a double-tapered cementless femoral component (Bi-Metric femoral stem; Biomet Inc., Warsaw, Indiana). Patients and Methods. A total of 1598 consecutive hips, in 1441 patients receiving primary THA between January 2010 and June 2015, were retrospectively identified. Level of pre-operative osteoarthritis, femoral Dorr type and cortical index were recorded. Varus/valgus placement of the stem and canal fill ratio were recorded post-operatively. Periprosthetic fractures were identified and classified according to the Vancouver classification. Regression analysis was performed to identify risk factors for early periprosthetic fracture. Results. The mean follow-up was 713 days (1 to 2058). A total of 48 periprosthetic fractures (3.0%) were identified during the follow-up and median time until fracture was 16 days, (interquartile range 10 to 31.5). Patients with femoral Dorr type C had a 5.2 times increased risk of post-operative periprosthetic fracture compared with type B, while female patients had a near significant two times increased risk over time for post-operative fracture. Conclusion. Dorr type C is an independent risk factor for early periprosthetic fracture, following THA using a double tapered cementless stem such as the Bi-Metric. Surgeons should take bone morphology into consideration when planning for primary THA and consider using cemented femoral components in female patients with poor bone quality. Cite this article: Bone Joint J 2017;99-B:451–7


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 61 - 64
1 Nov 2012
Mayle Jr RE Paprosky WG

Revision total hip arthroplasty (THA) is projected to increase by 137% from the years 2005 to 2030. Reconstruction of the femur with massive bone loss can be a formidable undertaking. The goals of revision surgery are to create a stable construct, preserve bone and soft tissues, augment deficient host bone, improve function, provide a foundation for future surgery, and create a biomechanically restored hip. Options for treatment of the compromised femur include: resection arthroplasty, allograft prosthetic composite (APC), proximal femoral replacement, cementless fixation with a modular tapered fluted stem, and impaction grafting. The purpose of this article is to review the treatment options along with their associated outcomes in the more severe femoral defects (Paprosky types IIIb and IV) in revision THA.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1192 - 1196
1 Sep 2015
Amanatullah DF Siman H Pallante GD Haber DB Sierra RJ Trousdale RT

When fracture of an extensively porous-coated femoral component occurs, its removal at revision total hip arthroplasty (THA) may require a femoral osteotomy and the use of a trephine. The remaining cortical bone after using the trephine may develop thermally induced necrosis. A retrospective review identified 11 fractured, well-fixed, uncemented, extensively porous-coated femoral components requiring removal using a trephine with a minimum of two years of follow-up. . The mean time to failure was 4.6 years (1.7 to 9.1, standard deviation (. sd. ) 2.3). These were revised using a larger extensively porous coated component, fluted tapered modular component, a proximally coated modular component, or a proximal femoral replacement. The mean clinical follow-up after revision THA was 4.9 years (2 to 22, . sd. 3.1). The mean diameter of the femoral component increased from 12.7 mm (. sd. 1.9) to 16.2 mm (. sd. 3.4; p >  0.001). Two revision components had radiographic evidence of subsidence that remained radiographically stable at final follow-up. The most common post-operative complication was instability affecting six patients (54.5%) on at least one occasion. . A total of four patients (36.4%) required further revision: three for instability and one for fracture of the revision component. There was no statistically significant difference in the mean Harris hip score before implant fracture (82.4; . sd. 18.3) and after trephine removal and revision THA (81.2; . sd. 14.8, p = 0.918). . These findings suggest that removal of a fractured, well-fixed, uncemented, extensively porous-coated femoral component using a trephine does not compromise subsequent fixation at revision THA and the patient’s pre-operative level of function can be restored. However, the loss of proximal bone stock before revision may be associated with a high rate of dislocation post-operatively. Cite this article: Bone Joint J 2015;97-B:1192–6


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 820 - 825
1 Jul 2022
Dhawan R Baré JV Shimmin A

Aims

Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR) of patients with adverse SPM who received a DMA.

Methods

A multicentre study was performed using data from 227 patients undergoing primary total hip arthroplasty (THA), enrolled consecutively. All the patients who had one or more adverse spine or pelvic mobility parameter had a DMA inserted at the time of their surgery. The mean age was 76 years (22 to 93) and 63% were female (n = 145). At a mean of 14 months (5 to 31) postoperatively, the AOANJRR was analyzed for follow-up information. Reasons for revision and types of revision were identified.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 66 - 69
1 Nov 2014
Lombardi Jr AV Berend KR Adams JB

The common recommended treatment for infected total hip replacement is two-staged exchange including removal of all components. However, removal of well-fixed femoral stems can result in structural bone damage. We recently reported on an alternative treatment of partial two-stage exchange used in selected cases, in which a well-fixed femoral stem was left and only the acetabular component removed, the joint space was debrided thoroughly, an antibiotic-laden polymethylmethacrylate spacer was moulded using a bulb-type syringe and placed in the acetabulum, intravenous antibiotics were administered during the interval, and delayed re-implantation was performed. In 19 patients treated with this technique from January 2000 to January 2011, 89% were free of infection at a mean follow-up of four years (2 to 11). Since then, disposable silicone moulds have become available to fabricate spacers in separate femoral and head units. The head spacer mould, which incorporates various neck taper adapter options, greatly facilitates the technique of partial two-stage exchange. We report our early experience using disposable silicone head spacer moulds for partial two-stage exchange in seven patients with infected primary hip replacements. Cite this article: Bone Joint J 2014;96-B(11 Suppl A):66–9


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 55 - 57
1 Nov 2012
Schmalzried TP

The differential diagnosis of the painful total hip arthroplasty (resurfacing or total hip) includes infection, failure of fixation (loosening), tendinitis, bursitis, synovitis, adverse local tissue reaction (ALTR) to cobalt-chromium alloys, and non-hip issues, such as spinal disorders, hernia, gynecologic, and other pelvic pain. Assuming that the hip is the source of pain, the first level question is prosthetic or non-prosthetic pain generator? The second level prosthetic question is septic or aseptic? The third level question (aseptic hips) is well-fixed or loose? ALTR is best diagnosed by cross-sectional imaging. Successful treatment is dependent on correct identification and elimination of the pain generator. Treatment recommendations for ALTR and taper corrosion are evolving


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1611 - 1619
1 Dec 2016
Wilson MJ Hook S Whitehouse SL Timperley AJ Gie GA

Aims. Femoral impaction bone grafting was first developed in 1987 using morselised cancellous bone graft impacted into the femoral canal in combination with a cemented, tapered, polished stem. We describe the evolution of this technique and instrumentation since that time. . Patients and Methods. Between 1987 and 2005, 705 revision total hip arthroplasties (56 bilateral) were performed with femoral impaction grafting using a cemented femoral stem. All surviving patients were prospectively followed for a mean of 14.7 years (9.8 to 28.3) with no loss to follow-up. By the time of the final review, 404 patients had died. Results. There were 76 further revisions (10.8%) involving the stem; seven for aseptic loosening, 23 for periprosthetic fracture, 24 for infection, one for malposition, one for fracture of the stem and 19 cement-in-cement exchanges of the stem during acetabular revision. The 20-year survival rate for the entire series was 98.8% (95% confidence interval (CI) 97.8 to 99.8) with aseptic loosening as the endpoint, and 87.7% (95% CI 82.8 to 92.6) for revision for any reason. Survival improved with the evolution of the technique, although this was not statistically significant due to the overall low rate of further revision. Conclusion. This is the largest series of revision total hip arthroplasties with femoral impaction grafting, and the results support the continued use of this technique. Cite this article: Bone Joint J 2016;98-B:1611–19


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (. sd.  1.71) vs 0.32 mm (. sd. 0.45), p = 0.66; rotation 0.96° (. sd. 1.49) vs 1.41° (. sd. 2.95), p = 0.88; and total migration 1.09 mm (. sd. 1.74) vs 0.73 mm (. sd. 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem. Cite this article: Bone Joint J 2015; 97-B:595–602


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 124 - 130
1 Jan 2009
Deuel CR Jamali AA Stover SM Hazelwood SJ

Bone surface strains were measured in cadaver femora during loading prior to and after resurfacing of the hip and total hip replacement using an uncemented, tapered femoral component. In vitro loading simulated the single-leg stance phase during walking. Strains were measured on the medial and the lateral sides of the proximal aspect and the mid-diaphysis of the femur. Bone surface strains following femoral resurfacing were similar to those in the native femur, except for proximal shear strains, which were significantly less than those in the native femur. Proximomedial strains following total hip replacement were significantly less than those in the native and the resurfaced femur. These results are consistent with previous clinical evidence of bone loss after total hip replacement, and provide support for claims of bone preservation after resurfacing arthroplasty of the hip


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 92 - 94
1 Nov 2013
Gehrke T Gebauer M Kendoff D

Femoral revision after cemented total hip replacement (THR) might include technical difficulties, following essential cement removal, which might lead to further loss of bone and consequently inadequate fixation of the subsequent revision stem. . Femoral impaction allografting has been widely used in revision surgery for the acetabulum, and subsequently for the femur. In combination with a primary cemented stem, impaction grafting allows for femoral bone restoration through incorporation and remodelling of the impacted morsellized bone graft by the host skeleton. Cavitary bone defects affecting meta-physis and diaphysis leading to a wide femoral shaft, are ideal indications for this technique. Cancellous allograft bone chips of 1 mm to 2 mm size are used, and tapered into the canal with rods of increasing diameters. To impact the bone chips into the femoral canal a prosthesis dummy of the same dimensions of the definitive cemented stem is driven into the femur to ensure that the chips are very firmly impacted. Finally, a standard stem is cemented into the neo-medullary canal using bone cement. . To date several studies have shown favourable results with this technique, with some excellent long-term results reported in independent clinical centres worldwide. Cite this article: Bone Joint J 2013;95-B, Supple A:92–4


The Bone & Joint Journal
Vol. 104-B, Issue 6 | Pages 747 - 757
1 Jun 2022
Liang H Yang Y Guo W Yan L Tang X Li D Qu H Zang J Du Z

Aims

The aim of this study was to investigate the feasibility of application of a 3D-printed megaprosthesis with hemiarthroplasty design for defects of the distal humerus or proximal ulna following tumour resection.

Methods

From June 2018 to January 2020, 13 patients with aggressive or malignant tumours involving the distal humerus (n = 8) or proximal ulna (n = 5) were treated by en bloc resection and reconstruction with a 3D-printed megaprosthesis with hemiarthroplasty, designed in our centre. In this paper, we summarize the baseline and operative data, oncological outcome, complication profiles, and functional status of these patients.


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 598 - 603
1 May 2022
Siljander MP Gausden EB Wooster BM Karczewski D Sierra RJ Trousdale RT Abdel MP

Aims

The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes.

Methods

We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 608 - 615
1 May 2011
Bolland BJRF Culliford DJ Langton DJ Millington JPS Arden NK Latham JM

This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1202 - 1206
1 Sep 2014
Kumar V Sharma S James J Hodgkinson JP Hemmady MV

Despite a lack of long-term follow-up, there is an increasing trend towards using femoral heads of large diameter in total hip replacement (THR), partly because of the perceived advantage of lower rates of dislocation. However, increasing the size of the femoral head is not the only way to reduce the rate of dislocation; optimal alignment of the components and repair of the posterior capsule could achieve a similar effect. . In this prospective study of 512 cemented unilateral THRs (Male:Female 230:282) performed between 2004 and 2011, we aimed to determine the rate of dislocation in patients who received a 22 mm head on a 9/10 Morse taper through a posterior approach with capsular repair and using the transverse acetabular ligament (TAL) as a guide for the alignment of the acetabular component. The mean age of the patients at operation was 67 years (35 to 89). The mean follow-up was 2.8 years (0.5 to 6.6). Pre- and post-operative assessment included Oxford hip, Short Form-12 and modified University of California Los Angeles and Merle D’Aubigne scores. The angles of inclination and anteversion of the acetabular components were measured using radiological software. There were four dislocations (0.78%), all of which were anterior. In conclusion, THR with a 22 mm diameter head performed through a posterior approach with capsular repair and using the TAL as a guide for the alignment of the acetabular component was associated with a low rate of dislocation. Cite this article: Bone Joint J 2014;96-B:1202–6


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1320 - 1326
1 Oct 2011
Regis D Sandri A Bonetti I Braggion M Bartolozzi P

Revision after failed femoral components may be technically demanding due to loss of peri-prosthetic bone. This retrospective study evaluated the long-term results of femoral revision using the cementless Wagner Self-Locking stem. Between 1992 and 1998, 68 consecutive hips in 66 patients underwent femoral revision using this implant. A total of 25 patients died from unrelated causes without further revision; the remaining 41 hips in 41 patients (12 men and 29 women) with a mean age of 61 years (29 to 80) were reviewed at a mean follow-up of 13.9 years (10.4 to 15.8). A transfemoral approach was used in 32 hips. A total of five stems required further revision because of infection in two, progressive subsidence in two and recurrent dislocation in one. Four hips had dislocated and eight stems had subsided ≥ 10 mm. The mean Harris hip score improved from 33 points pre-operatively to 75 points at final follow-up (p < 0.001). In all, 33 stems (91.7%) showed radiological signs of stable bone fixation. The cumulative survival rates at 15.8 years with femoral revision for any reason and for stem failure as the endpoints were 92.0% (95% confidence interval (CI) 86.0% to 98.4%) and 96.6% (95% CI 92.2% to 100%), respectively. The survivorship with revision and ≥ 10 mm migration of the stem as the endpoint was 83.6% (95% CI 76.6% to 91.4%). This study shows quite good survival and moderate clinical outcome when using a monoblock tapered titanium stem for supporting the regeneration of bone in complex revision hip surgery.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 63 - 66
1 Nov 2013
Rodriguez JA Cooper HJ

Large ceramic femoral heads offer several advantages that are potentially advantageous to patients undergoing both primary and revision total hip replacement. Many high-quality studies have demonstrated the benefit of large femoral heads in reducing post-operative instability. Ceramic femoral heads may also offer an advantage in reducing polyethylene wear that has been reported in vitro and is starting to become clinically apparent in mid-term clinical outcome studies. Additionally, the risk of taper corrosion at a ceramic femoral head–neck junction is clearly lower than when using a metal femoral head. With improvements in the material properties of both modern ceramic femoral heads and polyethylene acetabular liners that have reduced the risk of mechanical complications, large ceramic heads have gained popularity in recent years. Cite this article: Bone Joint J 2013;95-B, Supple A:63–6


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 212 - 220
1 Feb 2022
Fishley WG Selvaratnam V Whitehouse SL Kassam AM Petheram TG

Aims

Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique.

Methods

We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 315 - 320
1 Mar 2006
Field RE Singh PJ Latif AMH Cronin MD Matthews DJ

We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims

Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study.

Methods

In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 83 - 90
1 Jan 2022
Batten TJ Gallacher S Evans JP Harding RJ Kitson J Smith CD Thomas WJ

Aims

The use and variety of stemless humeral components in anatomical total shoulder arthroplasty (TSA) have proliferated since their advent in 2004. Early outcomes are reassuring but independent mid-term results are scarce. This independent study reports a consecutive series of 143 Eclipse stemless shoulder prostheses with a minimum five-year (5 to 10) follow-up.

Methods

Outcomes of 143 procedures undertaken for all indications in 131 patients were reviewed, with subset analysis of those for osteoarthritis (OA) (n = 99). The primary outcome was the Oxford Shoulder Score (OSS) at a minimum of five years. Secondary outcomes were ranges of motion and radiological analysis of humeral radiolucency, rotator cuff failure, and glenoid loosening.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1611 - 1617
1 Dec 2012
Jameson SS Baker PN Mason J Gregg PJ Brewster N Deehan DJ Reed MR

Despite excellent results, the use of cemented total hip replacement (THR) is declining. This retrospective cohort study records survival time to revision following primary cemented THR using the most common combination of components that accounted for almost a quarter of all cemented THRs, exploring risk factors independently associated with failure. All patients with osteoarthritis who had an Exeter V40/Contemporary THR (Stryker) implanted before 31 December 2010 and recorded in the National Joint Registry for England and Wales were included in the analysis. Cox’s proportional hazard models were used to analyse the extent to which risk of revision was related to patient, surgeon and implant covariates, with a significance threshold of p < 0.01. A total of 34 721 THRs were included in the study. The overall seven-year rate of revision for any reason was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final adjusted model the risk of revision was significantly higher in THRs with the Contemporary hooded component (hazard ratio (HR) 1.88, p < 0.001) than with the flanged version, and in smaller head sizes (< 28 mm) compared with 28 mm diameter heads (HR 1.50, p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69 to 1.63) with a 28 mm diameter head and flanged component. The overall risk of revision was independent of age, gender, American Society of Anesthesiologists grade, body mass index, surgeon volume, surgical approach, brand of cement/presence of antibiotic, femoral head material (stainless steel/alumina) and stem taper size/offset. However, the risk of revision for dislocation was significantly higher with a ‘plus’ offset head (HR 2.05, p = 0.003) and a hooded acetabular component (HR 2.34, p < 0.001). In summary, we found that there were significant differences in failure between different designs of acetabular component and sizes of femoral head after adjustment for a range of covariates


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 66 - 72
1 Jul 2021
Hernandez NM Hinton ZW Wu CJ Lachiewicz PF Ryan SP Wellman SS

Aims

Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years.

Methods

Using the database of a single academic centre, 126 revision THAs in 117 patients using a single design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 665 - 677
1 May 2011
Sköldenberg OG Salemyr MO Bodén HS Lundberg A Ahl TE Adolphson PY

Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck. We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years. Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D. Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years. The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1409 - 1418
1 Nov 2006
Scheerlinck T Casteleyn P

We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from in vitro studies and finite element modelling were examined with reference to the survivorship results for each of the concepts of fixation. It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed