Hip resurfacing arthroplasty (HRA) became a popular procedure in the early 90s because of the improved wear characteristic, preserving nature of the procedure and the optimal stability and range of motion. Concerns raised since 2004 when metal ions were seen in blood and urine of patients with a MoM implant. Design of the prosthesis, acetabular component malpositioning, contact-patch-to-rim distance (CPR) and a reduced joint size all seem to play a role in elevated metal ion concentrations. Little is known about the influence of physical activity (PA) on metal ion concentrations. Implant wear is thought to be a function of use and thus of patient activity levels. Wear of polyethylene acetabular bearings was positively correlated with patient's activity in previous studies. It is hypothesized that daily habitual physical activity of patients with a unilateral resurfacing prosthesis, measured by an activity monitor, is associated with habitual physical activity. A prospective, explorative study was conducted. Only patients with a unilateral hip resurfacing prosthesis and a follow-up of 10 ± 1 years were included. Metal ion concentrations were determined using ICP-MS. Habitual physical activity of subjects was measured in daily living using an acceleration-based activity monitor. Outcome consisted of quantitative and qualitative activity parameters. In total, 16 patients were included. 12 males (75%) and 4 females (25%) with a median age at surgery of 55.5 ± 9.7 years [43.0 – 67.9] and median follow-up of 9.9 ± 1.0 years [9.1 – 10.9]. The median cobalt and chromium ion concentrations were 25 ± 13 and 38 ± 28 nmol/L. A significant relationship, when adjusting for age at surgery, BMI, cup size and cup inclination, between sit-stand transfers (p = .034) and high intensity peaks (p = .001) with cobalt ion concentrations were found (linear regression analysis). This study showed that a high number of sit-stand transfers and a high number of high intensity peaks is significantly correlated with high metal ion concentrations, but results should be interpreted with care. For patients it seems save to engage in activities with low intensity peaks like walking or cycling without triggering critical wear or metal ions being able to achieve important general health benefits and quality of life, although the quality (high intensity peaks) of physical activity and behaviour of patients (sit-stand-transfers) seem to influence metal ion concentrations.
An increase in metal ion levels is seen after implantation of all MoM hip prosthesis due to release from the surface directly, more so during articulation and corrosion of the bearing surfaces. The bearing surfaces in MoM prosthesis consist of cobalt, chromium and molybdenum. Several case-reports of cobalt toxicity due to a MoM prosthesis have been published in the last decade. Cobalt intoxication may lead to a variety of symptoms: neuro-ocular toxicity (tinnitus, vertigo, deafness, blindness, convulsions, headaches and peripheral neuropathy), cardiotoxicity and thyroid toxicity. Nausea, anorexia and unexplained weight loss have been described. Systemic effects from metal ions even with well functioning implants or with ion concentrations lower than those associated with known adverse effects may exist and warrant investigation. The aim of this study is to investigate self-reported systemic complaints in association with cobalt ion concentrations in patients with any type of MoM hip prosthesis. A cohort study was conducted. Patients with both unilateral and bilateral, resurfacing and large head metal on metal total hip arthroplasties were included for the current study. Blood metal ion concentrations (cobalt and chromium) were measured by inductively coupled plasma mass spectrometry (ICP-MS). Based on the known cobalt toxicity symptoms of case-reports and toxicology reports a new non-validated questionnaire was developed. questions were subdivided in general questions/symptoms, vestibular symptoms, neurological symptoms, emotional health and cardio- and thyroid toxicity symptoms. Independent samples T test, Fishers Exact Test and Pearsons (R) correlation were used. Analysis was performed on two groups; a low cobalt ion concentration group and a high cobalt ion concentration group A total of 62 patients, 36 (58%) men and 26 (42%) women, were included with a mean age at surgery of 60.8 ± 9.3 years (41.6 – 78.1) and a mean follow up of 6.3 ± 1.4years (3.7 – 9.6). In these patients a total of 71 prosthesis were implanted: 53 unilateral and 9 bilateral. Of these, 44 were resurfacing and 27 large head metal on metal (LHMoM) total hip arthroplasties. Mean cobalt and chromium ion concentrations were 104 ± 141 nmol/L (9 – 833) and 95 ± 130nmol/L (6 – 592), respectively. Based on the different thresholds (120 – 170 or 220 nmol/L) the low cobalt ion concentration group consisted of 44 (71%), 51 (82%) or 55 (89%) subjects respectively. No differences were found in general characteristics, independently of the threshold. The composite score of vestibular symptoms (vision, hearing, tinnitus, dizziness) was significantly higher (p < .050) in all high cobalt ion concentrations groups, independent of the threshold value This study aimed to detect a trend in self-reported systemic complaints in patients with metal-on-metal hip arthroplasty due to raised cobalt ion concentrations. Vestibular symptoms were more common in high cobalt ion concentration groups independent of the three threshold levels tested. The upper limit of acceptable cobalt ion concentrations remains uncertain. With regards to proactively inquired, self-reported symptoms the threshold where effects may be present could be lower than values currently applied in clinical follow-up. It is unknown what exposure to elevated metal ion concentrations for a longer period of time causes with aging subjects. Further research with a larger cohort and a more standardized questionnaire is necessary to detect previously undiscovered or under-reported effects.
Acetabular component orientation is an important determinant of outcome following total hip arthroplasty (THA). Although surgeons aim to achieve optimal cup orientation, many studies demonstrate their inability to consistently achieve this. Factors that contribute are pelvic orientation and the surgeon's ability to correctly orient the cup at implantation. The goal of this study was to determine the accuracy with which surgeons can achieve cup orientation angles. In this Introduction
Methods
Metal-on-metal hip implants can produce adverse tissue reactions to wear debris. Increased metal ion concentrations in the blood are measured as a proxy to wear and the complications it can trigger. Many studies have examined various factors influencing the metal ion concentrations. This is the first study to investigate the effect of physical activity level, as objectively measured in daily life, on blood ion levels, expecting higher concentrations for higher patient activity. Thirty-three patients (13F/20M, 55.8 ± 6.2 years at surgery) with a unilateral resurfacing hip prosthesis were included. At last follow-up (6.8 ± 1.5 years) cobalt and chromium concentrations in the blood were determined by inductively coupled plasma mass spectrometry. Physical activity was measured during 4 successive days using a 3D-acceleration-based activity monitor. Data was analysed using validated algorithms, producing quantitative and qualitative parameters. Acetabular cup position was measured radiographically. Correlations were tested with Pearson's r'.Background
Methods
The second generation metal-on-metal (MoM) prosthesis of the hip became a worldwide success in the 90s. However, after the placement of a MoM prosthesis the cobalt ion concentrations raise significantly. This may lead to systemic complaints and even cobalt toxicity. Sixty-one patients (26F/35M) with both an unilateral and bilateral resurfacing or large-head MoM (LHMoM) hip prosthesis were included. At last follow-up (5.77 ± 1.57 yrs) cobalt concentrations in the blood were determined by ICP-MS. Based on the known cobalt toxicity symptoms we developed a non-validated questionnaire. Analysis was done on two groups; a low cobalt concentration group and a high cobalt concentration group. We used 170 nmol/L as the upper limit of well functioning prosthesis as defined by the Dutch Orthopaedic Society (NOV). Independent samples T test and Pearson correlation coefficient were done.Background
Methods
A comparative study on CT- and MRI-based patient specific matched guides (PSG) from the same manufacturer for the implantation of total knee arthroplasty (TKA) has not been undertaken. A total of 64 knees operated with CT based PSG was divided into two groups, with (n=32, CTHK) or without (n=32, CTNA) a history of a knee operation, and matched with a control group operated with MRI based PSG(n=64). Alignment of the biomechanical axis of the leg (HKA angle) and accuracy of individual implant alignment were measured on digital long-standing AP and sagittal radiographs. HKA and implant angles <3° deviation of the preoperative planned alignment were defined as correct. Peroperative implant size, OR time (min) and blood loss (ml) were compared. The average HKA angle in the CTHK group (177.0, 170.5 to 181.5, p=0.016) and mean varus-valgus alignment of the tibia component in the MRI group (90.6, 85.6 to 94.1, p=0.003) were statistically significant different. None of the outcome on the frontal femoral and lateral tibial component were statistically significant different. Percentage <3° deviation of the preoperative planned femoral flexion-extension alignment was better in the MRI group (84%, p=0.002), compared to the CTHK and CTNA group (respectively 30% and 42%). Average operation time was statistically significant shorter in favour of the MRI group (53.1, 34 to 80, p≤0.00), compared to the CTHK(70.8, 44 to 114) and CTNA group (59.2, 41 to 78). There is discrepancy between CT and MRI based PSG from the same manufacture because of patients who were not suitable for MRI due to history of a knee operation in the past. Whether these differences are clinically relevant is questionable. Future research needs to emphasise whether one of these two modalities, MRI or CT is superior compared to the other.
This is the first report of a new technique for unicompartmental to total knee arthroplasty revision surgery in which patient specific guides are formed based on preoperative CT imaging. This technique can help to make revision surgery less technically demanding. Unicompartmental to total knee arthroplasty revision surgery can be a technically demanding procedure. Joint line restoration, rotation and augmentations can cause difficulties. This study describes a new technique in which single way fitting guides serve to position knee system cutting blocks.Summary Statement
Introduction
This paper is the first to compare the results of unicompartmental to total knee arthroplasty revision surgery between cases with explained pain and cases with unexplained pain. Revision surgery for unexplained pain usually results in a less favourable outcome. Although it is suggested in literature that results of UKA to TKA revision surgery improve when the mechanism of failure is understood, a comparative study regarding this topic is lacking.Summary Statement
Introduction
Alignment results did not differ between PSG and conventional instrumentation. A small reduction in operation time and blood loss was found with the PSG system, but is unlikely of clinical significance. Length of hospital stay was identical for both groups. Several techniques for aligning a TKA exist nowadays. Patient-specific guiding (PSG) has relatively recently been introduced to try to resolve the shortcomings of existing techniques while optimising the operative procedure. Still few reports have been published on the clinical outcome and on the peroperative results of this new technique. This prospective, double-blind, randomised controlled trial was designed to address the following research questions: 1. Is there a significant difference in outliers in alignment in the frontal and sagittal plane between PSG TKA and conventional TKA. 2. Is there a significant difference in operation time, blood loss and length of hospital stay between the 2 techniques.Summary
Introduction
Patient-specific guiding (PSG) is a relatively new technique for aligning a total knee arthroplasty (TKA). Limited data exist on the precise accuracy of the technique. The purpose of this study is to investigate whether there was significant difference between the alignment of the individual femoral and tibial components (in all three anatomical planes) as calculated pre-operatively and the actually achieved alignment Twenty-six patients were included. Software permitted matching of the pre-operative MRI-scan (and therefore calculated prosthesis position) to a pre-operative full-leg CT-scan. After surgery a post-operative full-leg CT-scan could be superimposed onto the pre-operative CT-scan to accurately determine deviations from planning (see figure 1 and 2). This 3D-technique has an accuracy of 0.7–1.0 degrees.Background:
Methods:
The full leg x-ray is a widely used imaging modality for post-operative assessment of total knee replacement (TKR). However, these assessments require controlled conditions and precise measuring in order to be accurate. inter-observer reliability remains a matter of concern as well. This study examines whether intersurgeon differences are significant. Post-operative lateral and full-leg frontal x-rays of 26 patients were assessed by 6 surgeons according to a strict measuring protocol. Four measurements (Figure 1 and 2) were taken of which two were on the femur (Femoral Varus Angle FVA and Femoral Flexion Angle FFA) and two, on the tibia (Tibial Varus Angle TVA; Tibial Slope Angle TSA). A random effects, two-way ANOVA was performed on the data using Minitab (v 16.0, Minitab Inc., Pennsylvania, USA) to determine whether a surgeon has influence on the results (α = 0.05). Intra-class correlation coefficients (ICC) and standard error of measurements (SEM) resulting in smallest detectable changes (SDC) were also calculated [1].Background
Method