While Oxidized Zirconium (OxZr) femoral heads matched with highly cross-linked polyethylene (XLPE) have demonstrated the lowest rate of revision compared to other bearing couples in the Australian National Joint Registry, it has been postulated that these results may, in part, be due to the fact that a single company offers this bearing option with a limited combination of femoral and acetabular prostheses. The purpose of this study was to assess clinical and radiographic outcomes in a matched cohort of total hip replacements (THR) utilizing an identical cementless femoral stem and acetabular component with either an Oxidized Zirconium (OxZr) or Cobalt-Chrome (CoCr) femoral heads at a minimum of 10 years follow-up. We reviewed our institutional database to identify all patients whom underwent a THR with a single cementless femoral stem, acetabular component, XLPE liner and OxZr femoral head with a minimum of 10 years of follow-up. These were then matched to patients who underwent a THR with identical prosthesis combinations with CoCr femoral head by gender, age and BMI. All patients were prospectively evaluated with WOMAC, SF-12 and Harris Hip Score (HHS) preoperatively and postoperatively at 6 weeks, 3 months, 1 and 2 years and every 2 years thereafter. Charts and radiographs were reviewed to determine the revision rates and survivorship (both all cause and aseptic) at 10 years for both cohorts. Paired analysis was performed to determine if differences exist in patient reported outcomes. There were 208 OxZr THRs identified which were matched with 208 CoCr THRs. There was no difference in average age (OxZr, 54.58 years, CoCr, 54.75 years), gender (OxZr 47.6% female, CoCr 47.6% female), and average body max index (OxZr, 31.36 kg/m2, CoCr, 31.12 kg/m2) between the two cohorts. There were no significant differences preoperatively in any of the outcome scores between the two groups (WOMAC (p=0.449), SF-12 (p=0.379), HHS(p=0.3718)). Both the SF12 (p=0.446) and the WOMAC (p=0.278) were similar between the two groups, however the OxZr THR cohort had slightly better HHS compared to the CoCr THR cohort (92.6 vs. 89.7, p=0.039). With revision for any reason as the end point, there was no significant difference in 10 years survivorship between groups (OxZr 98.5%, CoCr 96.6%, p=0.08). Similarly, aseptic revisions demonstrated comparable survivorship rates at 10 year between the OxZr (99.5%) and CoCr groups (97.6%)(p=0.15). Both THR cohorts demonstrated outstanding survivorship and improvement in patient reported outcomes. The only difference was a slightly better HHS score for the OxZr cohort which may represent selection bias, where OxZr implants were perhaps implanted in more active patients. Implant survivorship was excellent and not dissimilar for both the OxZr and CoCr groups at 10 years. Therefore, with respect to implant longevity at the end of the first decade, there appears to be no clear advantage of OxZr heads compared to CoCr heads when paired with XLPE for patients with similar demographics. Further follow-up into the second and third decade may be required to demonstrate if a difference does exist.
Surgeons performing a total knee replacement (TKR) have two available techniques available to help them achieve the proper bone resections and ligament tension – gap balancing (GB) and measured resection (MR). GB relies on balancing ligaments prior to bony resections whereas bony resections are made based on anatomical landmarks in MR. Many studies have been done to compare the joint kinematics between the two techniques, however the results have been varied. These studies were not done with anatomically designed prostheses. The Journey II (Smith & Nephew, Memphis, TN) is one such design which attempts to mimic the normal knee joint structure to return more natural kinematics to the joint, with emphasis on eliminating both paradoxical anterior motion and reduced posterior femoral rollback. Given the design differences between anatomical and non-anatomical prostheses, it is important to investigate whether one technique provides superior kinematics when an anatomical design is used. We hypothesize that there will be no difference between the two techniques. A total of 56 individuals were recruited to receive a Journey II prosthesis and randomized evenly to groups where the GB technique or MR technique is used. For all patients in the study, a series of radiostereometric analysis (RSA) images were acquired at 3-months post-operatively at different knee flexion angles, ranging in 20° increments from 0° to 120°. Model-based RSA software (RSACore, Leiden, Netherlands) was used to obtain the 3D positions and orientations of the femoral and tibial implant components, which were in turn used to obtain kinematic measures (contact locations and magnitude of excursion) for each condyle.Introduction
Methods
Despite improvements in the survivorship of total knee replacements (TKR) over the years, patient satisfaction following TKR has not improved, with approximately 20% of patients recording dissatisfaction with their new knee joint. It is unclear why many patients feel this way, but it may relate in part to implant designs that do not provide a “natural” feeling knee. Implant manufacturers continue to introduce new concepts for implant design, which are essential for reaching the goal of a “normal” knee after TKR surgery. The Journey II TKR (Smith & Nephew) was developed with this goal in mind. Its anatomical design attempts to mimic the normal knee joint structure to return more natural kinematics to the joint, with emphasis on eliminating both paradoxical anterior motion and reduced posterior femoral rollback. Our objective is to examine patients receiving the Journey II TKR to measure the knee joint contact kinematics of the Journey II TKR compared to a non-anatomically designed implant by the same manufacturer. We hypothesize that the Journey II TKR will have more natural contact kinematics that differ from the non-anatomically designed implant. A total of 28 individuals were recruited to receive a Journey II TKR, matching an existing prior cohort with a non-anatomical design from the same manufacturer (Legion TKR, Smith & Nephew). For both groups, a series of radiostereometric analysis (RSA) images were acquired at different knee flexion angles, ranging in 20° increments from 0° to 120°. Model-based RSA software (RSACore, Leiden, Netherlands) was used to obtain the 3D positions and orientations of the femoral and tibial implant components, which were in turn used to obtain kinematic measures (contact locations and magnitude of excursion) for each condyle. Results from the Journey II TKR group at 3 months post-operation were compared to the 2-year post-operative measurements from the Legion TKR group.Introduction
Methods
Patient specific instrumentation (PSI) for total knee replacement (TKR) has demonstrated mixed success in simplifying the operation, reducing its costs, and improving limb alignment. Evaluation of PSI with tools such as radiostereometric analysis (RSA) has been limited, especially for cut-through style guides providing mechanical alignment. The primary goal of the present study was to compare implant migration following TKR using conventional and PSI surgical techniques, with secondary goals to examine whether the use of PSI reduces operative time, instrumentation, and surgical waste. The study was designed as a prospective, randomized controlled trial of 50 patients, with 25 patients each in the PSI and conventional groups, powered for the RSA analysis. Patients in the PSI group received an MRI and standing 3-foot x-rays to construct patient-specific cut-through surgical guides for the femur and tibia with a mechanical alignment. All patients received the same posterior-stabilized implant, with marker beads inserted in the bone around the implants to enable RSA imaging. Intraoperative variables such as time, number of instrumentation trays used, and mass of surgical waste were recorded. Patients underwent supine RSA exams at multiple time points (2&6 weeks, 3&6 months and yearly) with 6 months data currently available. Migration of the tibial and femoral components was calculated using model-based RSA software. WOMAC, SF-12, EQ5D, and UCLA outcome measures were recorded pre-operatively and post-operatively.Background
Methods
The advent of highly cross-linked polyethylene has resulted in improved wear rates and reduced osteolysis with at least intermediate follow-up when compared to conventional polyethylene. However, the role of alternative femoral head bearing materials in decreasing wear is less clear. The purpose of this study was to determine in-vivo polyethylene wear rates across ceramic, Oxinium, and cobalt chrome femoral head articulations. A review of our institutional database was performed to identify patients who underwent a total hip arthroplasty using either ceramic or oxidized zirconium (Oxinium) femoral head components on highly cross-linked polyethylene between 2008 and 2011. These patients were then matched on implant type, age, sex and BMI with patients who had a cobalt chrome bearing implant during the same time period. RSA analysis was performed using the center index method to measure femoral head penetration (polyethylene wear). Secondary quality of life outcomes were collected using WOMAC and HHS Scores. Paired analyses were performed to detect differences in wear rate (mm/year) between the cobalt chrome cohorts and their matched ceramic and Oxinium cohorts. Additional independent group comparisons were performed by analysis of variance with the control groups collapsed to determine wear rate differences between all three cohorts.Background
Methods
Adequate fixation of implant components is an important goal for all arthroplasty procedures. Aseptic loosening is one of the leading causes of revision surgery in total knee arthroplasty. Radiostereometric analysis (RSA) is an imaging technique to measure implant migration, with established migration thresholds for well-fixed, at risk, and unacceptably migrating components. The purpose of the present study was to examine the long-term fixation of a cemented titanium fixed bearing polished tibial baseplate. Patients enrolled in a previous two-year prospective trial were recalled at ten years. All patients received a cemented, posterior-stabilised total knee replacement of the same design implanted by one of three surgeons. Of the original 35 patients, 16 were available for long-term follow-up, with one patient lost to follow-up, nine patients deceased, and a further nine patients unwilling to return to the clinic. Each patient underwent RSA imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes as well as maximum total point motion (MTPM) was compared between all time points (baseline, six weeks, three months, six months, one year, two years) up to the ten year follow-up visits. Outcome scores including the Knee Society Score (KSS), WOMAC, SF-12, and UCLA Activity Score were recorded. At ten years, the mean migrations of the tibial component were less than 0.1 mm and 0.1 degree in all planes relative to the post-operative RSA exam. There was no significant difference in tibial component migration between time points. However, MTPM increased significantly over time (p = 0.002), from 0.23 ± 0.18 mm at six weeks to 0.42 ± 0.20 mm at ten years. At one year, 13 patients had an acceptable MTPM level, three patients had an ‘at risk’ level, and no patient had an ‘unacceptable’ level. No patients were revised at ten years. WOMAC and KSS were significantly improved (p < 0.0001) at the latest follow-up compared to pre-operatively, but there was no difference in SF-12. The median UCLA Activity Score at latest follow-up was six (range, two to eight). The tibial baseplate demonstrated solid fixation at ten years. No patients had an unacceptable MTPM level at one year and no patients were revised at ten years, supporting the use of RSA to predict long-term loosening risk. The low level of tibial baseplate migration found in the present study correlates to the low rate of revision for this implant as reported in individual studies and in joint replacement registries.
Infection following total hip arthroplasty (THA) represents a devastating complication and is one of the main causes for revision surgery. This complication may be treated by irrigation and debridement with head and polyethylene exchange (IDHPE) or a two-stage revision (2SR). Previous studies have reported on the eradication success rates but few have reported patient outcome scores. The purpose of this study was to report patient outcome scores for both IDHPE and 2SR and compare these to a non-infected matched cohort. We hypothesised that both cohorts would have worse outcomes than the control group, and that those who failed an initial IDHPE and required a 2SR would have a worse outcome than those treated initially with a 2SR. A retrospective review identified 137 patients from our institutional arthroplasty database who had an infected primary THA between 1986–2013. We excluded patients with less than one-year follow-up. Mean follow-up was 60 months (12–187 months). A control cohort was identified and matched according to age and Charlton Comorbidity Index (CCI). Harris Hip Scores, Short Form 12 and WOMAC scores were compared between our control group and our infected cohort. Sixty-eight patients were treated with a 2SR and 69 patients were treated with an IDHPE. There was a 59% success rate in eradicating the infection with an IDHPE. All of the 28 patients who failed an IDHPE later went on to a 2SR. Outcome scores for the 2SR cohort were significantly worse than the non-infected controls (p0.05). There was no difference in outcome scores when comparing our 2SR cohort to our failed IDHPE (p>0.05). Previous studies have focused on eradication rates. However, it is important to consider patient outcome scores when deciding the best treatment. Infected patients treated with a successful IDHPE had similar outcomes to non-infected patients. Patients that failed IDHPE and went onto 2SR had similar outcomes to those that had a 2SR alone. IDHPE should still be considered in the treatment algorithm of infected THA.
Previous retrieval studies demonstrate increased tibial baseplate roughness leads to higher polyethylene backside wear in total knee arthroplasty (TKA). Micromotion between the polyethylene backside and tibial baseplate is affected by the locking mechanism design and can further increase backside wear. The purpose of this study was to examine modern locking mechanisms, in the setting of both roughened and polished tibial baseplates, on backside tibial polyethylene wear. Five TKA models were selected, all with different tibial baseplate and/or locking mechanism designs. Six retrieval tibial polyethylenes from each TKA model were matched based on time There were no differences among the groups based on TIV (p=0.962), age (p=0.651), BMI (p=0.951), gender, revision number, or reason for revision. There was a significant difference across groups for mean total damage score (p=0.029). The polished tibial design with a partial peripheral capture locking mechanism and anterior constraint demonstrated a significantly lower score compared to one of the roughened tibial designs with a complete peripheral-rim locking mechanism (13.0 vs. 22.1, p=0.018). Otherwise, mean total damage scores were not significant between groups. As far as modes of wear, there were identifiable differences among the groups based on abrasions (p=0.005). The polished design with a tongue-in-groove locking mechanism demonstrated a significantly higher score compared to both groups with roughened tibial baseplates (5.83 vs. 0.83, p=0.024 and 5.83 vs. 0.92, p=0.033). Only the two designs with roughened tibial baseplates demonstrated dimpling (5.67 and 8.67) which was significant when compared against all other groups (p0.99). No other significant differences were identified when examining burnishing, cold flow, scratching, or pitting. No polyethylene components exhibited embedded debris or delamination. Total damage scores were similar between all groups except when comparing one of the polished TKA design to one of the roughened designs. The other TKA model with a roughened tibial baseplate had similar damage scores to the polished designs, likely due to its updated locking mechanism. Dimpling wear patterns were specific for roughened tibial baseplates while abrasive wear patterns were identified in the design with a tongue-in-groove locking mechanism. Our study showed even in the setting of a roughened tibial baseplate, modern locking mechanisms decrease backside wear similar to that of other current generation TKA designs.
Studies that have previously examined the relationship between inclination angle and polyethylene wear have shown increased wear of conventional polyethylene with high inclination angles. To date, there have been no long term An institutional arthroplasty database was used to identify patients who had metal-on-highly crosslinked polyethylene primary total hip arthroplasty (THA) using the same component design with a minimum follow up of 10 years ago. A modified RSA examination setup was utilised, recreating standard anteroposterior (AP) and cross-table lateral exams in a single biplane RSA acquisition. Three dimensional head penetration was measured using the centre index method. The same radiographs were used to measure inclination angle and anteversion. Spearman correlation was used to show an association between the parameters of acetabular position and wear rate. A total of 43 hips were included for analysis in this study. Average follow-up was 12.3 ± 1.2 years. The average linear wear rate was calculated to be 0.066 ± 0.066 mm/year. Inclination angle was not correlated with polyethylene wear rate (p=0.82). Anteversion was also not correlated with polyethylene wear rate (p=0.11). There was no statistical difference between wear rates of hips within Lewinnek's “safe zone” and those outside this “safe zone” (p=0.11). Males had a higher wear rate of 0.094 ± 0.089 mm/year compared to females with a wear rate of 0.046 ± 0.032 mm/year (p=0.045). At long term follow up of greater than 10 years, highly cross linked polyethylene has very low wear rates. This excellent tribology is independent of acetabular position, but gender did impact wear rates. Due to the low wear rates, follow-up of even longer term is suggested to examine variables affecting wear.
Complications involving the patellofemoral joint are a source of anterior knee pain, instability, and dysfunction following total knee arthroplasty. “Overstuffing” the patello-femoral joint refers to an increase in the thickness of the patellofemoral joint after a total knee replacement compared to the preoperative thickness. While biomechanical studies have indicated that overstuffing the patellofemoral joint may lead to adverse clinical outcomes, limited clinical evidence exists to support this notion. The purpose of this study is to evaluate the effect of changing the thickness of the patellafemoral joint on functional outcomes following total knee arthroplasty. Our institutional arthroplasty database was used to identify 1347 patients who underwent a primary total knee arthroplasty between 2006 and 2012 with the same component design. Standard preoperative and postoperative anteroposterior, lateral, and skyline radiographs were collected and measured for patello-femoral overstuffing. These measurements included anterior patellar displacement, anterior femoral offset, and anteroposterior femoral size. These measurements were correlated with patient outcome data using WOMAC, KSS scores, and postoperative range of motion. Multiple linear regression analysis was used to assess the association between stuffing and functional outcomes. A total of 1031 patients who underwent total knee arthroplasty were included. Increased anterior patellar displacement, a measure of patellofemoral joint thickness, was associated with decreased WOMAC scores (p=0.02). Anterior femoral offset (p=0.210) and anteroposterior femoral size (p=0.091) were not significantly associated with patient functional outcomes. Postoperative range of motion (ROM) was not associated with patellofemoral stuffing (p=0.190). The current study demonstrated that functional outcomes are adversely affected by patellofemoral overstuffing. Based on these results, caution is encouraged against increasing the thickness of the patellofemoral joint, particularly on the patellar side of the joint.
Tranexamic acid (TEA), an antifibrinolytic agent, is routinely used for reduction of blood loss in total hip arthroplasty (THA). However, use of intravenous (IV) TEA has been questioned due to safety concerns and a lack of biochemical data in the arthroplasty literature. Tranexamic acid given topically as a periarticular solution is a promising alternative route of administration. The purpose of this study is to identify differences in systemic absorption for intravenous and topical TEA administered during primary THA. In a blinded randomised controlled trial of patients undergoing primary cementless total hip arthroplasty, 29 participants received a weight-based bolus infusion of intravenous TEA (20 mg/kg) 10 minutes prior to skin incision. Conversely, 15 participants received a 1.5 g bolus dose of TEA administered topically into the periarticular region of the operative hip at the time of arthrotomy closure. A blood sample was drawn one hour post-administration for measurement of serum TEA concentration (µg/mL) by tandem mass spectrometry. In addition to comparing mean concentration levels for both treatment arms, each sample concentration was referenced to a pre-determined TEA concentration threshold of 10 µg/mL, a value known to represent 80% tissue plasminogen activator (tPA) inhibition Those participants receiving topical TEA had four-fold lower TEA levels at one hour postoperatively (mean 12.44 ± 17.59 versus 52.54 ± 23.94 µg/mL, p<0.05). These results demonstrate significantly lower circulating TEA at one hour after topical administration. Intravenous TEA must travel through the intravascular compartment in order to reach the operative hip. Topical administration of TEA targets bleeding tissues within the surgical field without necessitating parenteral administration. This results in less inhibition of tPA away from the operative site, potentially decreasing the risk of developing a pro-thrombotic state postoperatively. Correlating these results with outcomes from clinical efficacy trials comparing intravenous and topical TEA use in THA will further clarify optimal dosing strategies.
The philosophy of cemented total hip arthroplasty (THA) femoral components has become polarised. At one extreme are polished, collarless, tapered devices that are expected to subside; at the other extreme are roughened, non-tapered implants with a collar designed not to subside. Radiostereometric analysis (RSA) allows the accurate measurement of implant movement and has been extensively used for measurement of the Data from two previous prospective RSA trials with two-year follow-up were pooled. The first group included 36 patients who received a Spectron (Smith & Nephew, Memphis, USA) cemented stem. The second group included 13 patients who received an Exeter (Stryker, Mahwah, USA) cemented stem, and 15 patients who received a CPCS (Smith & Nephew, Memphis, USA) cemented stem. All patients underwent RSA examinations shortly post-operation, at 6 weeks, 3 months, 6 months, 1 years, and 2 years. Migration and rotation of the femoral stems was measured at each time point relative to the post-operative exam, and compared between the two groups. There was no difference in age at surgery (Spectron 78 ± 6 years, Exeter/CPCS 77 ± 5 years, p = 0.43), BMI (Spectron and Exeter/CPCS 28 ± 5 kg/m2, p = 0.92), or percentage of male patients (Spectron 23% male, Exeter/CPCS 21% male) between the implant groups. Subsidence was significantly greater (p < 0.0001) at all time points from three months to two years for the Exeter and CPCS stems (0.94 ± 0.39 mm at two years) compared to the Spectron stem (0.05 ± 0.16 mm at two years). There was no significant difference between the stem types for medial-lateral translation (p = 0.07) or anterior-posterior translation (p = 0.49), or for anterior-posterior tilt (p = 0.15), internal-external rotation (p = 0.89), or varus-valgus rotation (p = 0.05). Implant material, design, and surface finish are all factors in the long-term performance of cemented femoral hip implants. In this study, both femoral stem designs had a magnitude of sub-sidence that was within the limits of what is considered to be safe with respect to long-term performance. The continuous subsidence of the Exeter and CPCS stems is consistent with previous reports in the literature.
Trunnionosis in modular hip arthroplasty has recently been recognised to be clinically important. Gaining an understanding of how the material interface at the head-trunnion affects the tribology at the modular junctions has current clinical implications as well as an implication on future implant selection and material choice. This matched-cohort study aims to compare tribocorrosion between ceramic and cobalt-chromium trunnions and to investigate other factors that contribute to the difference in tribocorrosion if present. All hip prostheses retrieved between 1999 and 2015 at one centre were reviewed. Fifty two ceramic heads were retrieved, and these were matched to a cobalt-chromium cohort according to taper design, head size, neck length and implantation time in that order. The two cohorts were similar in male:female ratio (p=0.32) and body mass index (p=0.15) though the ceramic group was younger than the cobalt-chromium group (56.6 (+/−)13.5 years for ceramic group vs 66.3 (+/−14.4) years for cobalt-chromium group; p=0.001). There was no significant difference in the reasons for revision between the two groups (p=0.42). The femoral head trunnions were examined by two independent observers using a previously published 4-point scoring technique. The trunnions were divided into three zones: apex, middle and base. The observers were blinded to clinical and manufacturing data where possible. Ceramic head trunnions demonstrated a lower median fretting and corrosion score at the base zone (p<0.001), middle zone (p<0.001) and in the combined score (p<0.001). In a subgroup analysis by head size, ceramic heads had a lower fretting and corrosion score at 28mm head diameter (p<0.001). Within the ceramic group, taper design had a significant effect on fretting and corrosion in the apex zone (p=0.04). Taper design also had a similar effect in the cobalt-chromium group in the apex zone (p=0.03). For the ceramic trunnions, the largest effect was contributed by the difference between the 11/13 taper and the 12/14 taper. For the cobalt-chromium trunnions, the largest effect was contributed by the difference between the 5 degree 38′ 37″ taper and type 1 taper. Ceramic head trunnions showed a significantly lower fretting and corrosion score as compared to cobalt-chromium trunnions. Ceramic heads had a lower score than cobalt-chromium heads at 28mm head diameter. Taper design had an effect on fretting and corrosion within each cohort.
The advent of highly cross-linked polyethylene has resulted in improved wear rates and reduced osteolysis with at least intermediate follow-up when compared to conventional polyethylene. However, the role of alternative femoral head bearing materials in decreasing wear is less clear. The purpose of this study was to determine in-vivo polyethylene wear rates across ceramic, Oxinium, and cobalt chrome femoral head articulations. A review of our institutional database was performed to identify patients who underwent a total hip arthroplasty using either ceramic or oxidised zirconium (Oxinium) femoral head components on highly cross-linked polyethylene between 2008 and 2011. These patients were then matched on implant type, age, sex and BMI with patients who had a cobalt chrome bearing implant during the same time period. RSA analysis was performed using the centre index method to measure femoral head penetration (polyethylene wear). Secondary quality of life outcomes were collected using WOMAC and HHS Scores. Paired analyses were performed to detect differences in wear rate (mm/year) between the cobalt chrome cohorts and their matched ceramic and Oxinium cohorts. Additional independent group comparisons were performed by analysis of variance with the control groups collapsed to determine wear rate differences between all three cohorts. A total of 68 patients underwent RSA analysis. Fifteen patients with a ceramic femoral head component and 14 patients with an Oxinium femoral head component along with the same number of matched patients with cobalt chrome femoral head component were included in the analysis. The time This study presents evidence of a greater wear rate (mm/year) of the Oxinium femoral head component compared to a cobalt chrome femoral head component. This difference was not demonstrated in the ceramic femoral head component. Despite this difference, there were no clinical differences as measured by the HHS and WOMAC. Future research should focus on factors that may contribute to the higher wear rate seen in the Oxinium cohort.
Total joint arthroplasty is commonly associated with post-operative anemia. Blood conservation programs have been developed to optimise patients prior to surgery. Epoetin Alfa (Eprex) or intravenous (IV) iron transfusions are two modalities that can be used pre-operatively to optimise hemoglobin and ferritin levels. There are, however, potential complications and increased costs associated with their use. Oral iron is a less costly option for those undergoing surgery but requires more time to take effect. There are no studies to date that examine the effects of an early screening program utilising oral iron supplementation prior to total joint arthroplasty. The purpose of this study is to evaluate the effect of implementing early pre-operative oral iron supplementation on patients prior total joint arthroplasty. A retrospective review of patients undergoing total joint arthroplasty was performed using our institution clinical informatics database. We identified all patients seen in pre-admission clinic (PAC) between Jan 1, 2009 and March 31, 2010 representing our control group. We then identified all patients seen in PAC between October 1, 2012 and December 31, 2013. Patients in this cohort received screening blood work when booked for surgery, and oral iron supplementation was given to patients with hemoglobin of less than 135g/L or ferritin less than 100ug/L, thus representing our treatment group. Patients undergoing revision, uni-compartment knee arthroplasty and bilateral arthroplasties were excluded from the study. Pearson Chi-Square tests were used to calculate significance between groups with main outcomes including pre-admission hemoglobin, and pre-operative requirements for Eprex or IV iron. In our control group, we identified 354 patients (25.6%) with hemoglobin less than 130 g/L at time of pre-admission clinic. In our treatment group, this number dropped significantly to only 16.4% of patients (p<0.005). Implementation of an early screening program using oral iron supplementation resulted in a decrease in the number of patients with hemoglobin lower than 130 g/L at the time of pre-admission clinic. There was also a significant decrease in the use of Eprex and IV iron pre-operatively in the patients in the early screening program. These results encourage the use of early oral iron supplementation for patients with hemoglobin less than 135 g/L or ferritin less than 100ug/L in order to optimise patients prior to total joint arthroplasty.
The development of new bearing surfaces for total joint replacement is constantly evolving. Oxidized zirconium (Oxinium) has been introduced for use in both total hip arthroplasty (THA) and total knee arthroplasty (TKA). The aetiology of wear is multifactorial and includes adhesive, abrasive, third-body and fatigue wear mechanisms. Oxinium femoral components have demonstrated clear improvements in wear characteristics in-vitro. The purpose of this prospective study was to evaluate the mid-term (minimum 5 year) clinical and radiographic results and survivorship of the Genesis II™ knee implant system using an Oxinium femoral component. Between January 2001 and December 2008, 382 Genesis II Oxinium (Smith & Nephew) primary total knee arthroplasties (TKA) (313 patients) were implanted at our institution. A comparison with a cohort of 317 patients (382 knees) who received a Genesis II knee implant using a ‘conventional’ cobalt-chrome (Co-Cr) femoral component was performed during the same time period. Prospective data was collected on all patients including demographics (age, BMI, diagnosis) as well as pre and postoperative clinical outcome scores (SF-12, WOMAC, and knee society clinical rating scores (KSCRS). Radiological analysis for evidence of osteolysis and loosening was performed in all patients. Comparisons were performed to determine differences between the Oxinium and Cobalt Chrome cohorts. Kaplan-Meier survival analysis was performed to show cumulative survival over time. Failure was defined as femoral component revision due to any cause.INTRODUCTION
METHODS
Total knee arthroplasty (TKA) has proven clinical success with reported longterm survivorship of 92% in the elderly population. Concerns regarding increased loosening rates and potential need for multiple revision surgeries in patients younger than 60 years have traditionally discouraged TKA in younger patients. The purpose of this study was to review the longterm clinical and radiographic results of patients under the age of 45 yrs who underwent a total knee replacement. A retrospective review of our institutional database was performed between January 1996 and December 2004 The criteria for inclusion in the study were as follows: age 45 years or younger at index arthroplasty, cemented condylar prosthesis, and a minimum follow-up of 9 years. A total of 39 consecutive primary cemented condylar knee arthroplasties in patients with end stage arthritis were performed in 36 patients who were 45 years or younger. The median length of followup was 13.6 years. There were 18 men and 21 women. The mean age was 40.6 years (range, 28–44). There were 23 right, 16 left, and 3 bilateral procedures. The mean body mass index was 31.2 kg/m2 Clinical and radiological evaluations were performed before surgery and at 6 weeks, 3 months, 1 year and every 2 years subsequently. Knee scores were calculated using the Knee Society Clinical Rating Scores (KSCRS), SF-12 and WOMAC scores to assess pre, and post-operative function. Postoperative anteroposterior and lateral views were assessed for femoral and tibial component position, alignment, and presence or progression of radiolucent lines at the bone-cement and prosthesis-cement interfaces Statistical analysis was performed using the 2-sample test or Wilcoxon rank sum test for comparison of continuous variables. Kaplan-Meier analysis of implant survival was performed with failure defined as femoral component revision due to any cause.Introduction
Materials and Methods
The effect of polyethylene wear and lift-off between the tibial and femoral components on the mechanical axis was assessed in primary TKA (Total Knee Arthroplasty) based on retrieval data and full leg radiographs. Controversy exist regarding performing a TKA with component placement in physiologic versus neutral alignment. Some literature indicates good survivorship and superior clinical outcome in undercorrected TKA's for varus osteoarthritic knees. However, other literature indicates decreased survivorship and coronal plane alignment is still one of the contributing factors to wear in total knee arthroplasty (TKA). The two determinants of the intra-articular deformity in TKA arepolyethylene wear and lift-off between the tibial and femoral compartment. The goal of this study was to evaluate the impact of wear and lift-off on the mechanical axis in neutral and varus aligned TKA's.Summary
Introduction
Increased modularity of total hip arthroplasty components has occurred, with theoretical advantages and disadvantages. Recent literature indicates the potential for elevated revision rates of modular neck systems and the potential for metallosis and ALVAL (Aseptic Lymphocyte dominated Vasculitis Associated Lesion) formation at the modular neck/stem site. Retrieval analysis of one modular neck implant design including SEM (Scanning Electron Microscopy) assessment was done and correlated to FEA (Finite Element Analysis) as well as clinical features of patient demographics, implant and laboratory analysis. Correlation of the consistent corrosion locations to FEA indicates that the material and design features of this system may result in a biomechanical reason for failure. The stem aspect of the modular neck/stem junction may be at particular risk.
Coronal plane alignment is one of the contributing factors to polyethylene wear in total knee arthroplasty (TKA). The goal of this study was to evaluate the wear and damage patterns of retrieved tibial polyethylene inserts in relationship to the overall mechanical alignment and to the position of the tibial component. Based on full-length radiographs, ninety-five polyethylene inserts retrieved from primary TKA's with a minimum time in-vivo of five years were analysed for wear and damage. Four alignment groups were compared: valgus, neutral, mild varus and moderate varus. Varus and valgus positioning of the tibial component was analysed for damage score for the neutral and varus aligned groups.Introduction
Materials and methods