Advertisement for orthosearch.org.uk
Results 1 - 5 of 5
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 112 - 112
1 Mar 2017
Jang Y Yoo O Lee Y Lee M Elazab A Choi D
Full Access

Background

Open-wedge high tibial osteotomy (OWHTO) is an operation involving proper load re-distribution in the treatment for medial uni-compartmental arthritis of the knee joint. Therefore, stable fixation is mandatory for safe healing of this additive type of osteotomy to minimize the risk of non-union and loss of correction. For stability, screws provide optimal support and anchorage of the fixator in the condylar area without risking penetration of either the articulating surface. The purpose of the study was to evaluate the screw insertion angle and orientation with an anatomical plate that is post-contoured to the surface geometry of the proximal tibia after OWHTO.

Methods

From March 2012 to June 2014, 31 uni-planar and 38 bi-planar osteotomies were evaluated. Postoperative computed tomography data obtained after open wedge high tibial osteotomy using a locking plate were used for reconstruction of the 3 dimensional model with Mimics v.16.0 of the proximal tibia and locking plate. Measurement data were compared between 2 groups (gap lesser than or equal to 10 mm (Group 1) and gap greater than 10 mm(Group 2)). These data were also compared between the uniplanar (Group 3) and bi-planar (Group 4) osteotomy groups.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_4 | Pages 48 - 48
1 Feb 2017
Kang K Trinh T Yoo O Jang Y Lee M Lim D
Full Access

Introduction

The Rotational alignment is an important factor for survival total knee Arthroplasty. Rotational malalignment causes knee pain, global instability, and wear of the polyethylene inlay. Also, the anterior cortex line was reported that more reliable and more easily identifiable landmark for correct tibial component alignment. The aims of the current study is to identify effect of inserting the tibial baseplate of using anterior cortex line landmark of TKA on stress/strain distributions within cortical bone and bone cement. Through the current study, final aim is to suggest an alternative position of tibia baseplate for reduction of TKA failures with surgical convenience.

Materials and Method

A three-dimensional tibia FE model with TKA was generated based on a traditional TKA surgical guideline. Here, a commercialized TKA (LOSPA, Corentc, Korea) was considered corresponded to a patient specific tibia morphology. Tibia baseplate was positioned at anterior cortex line. Alternative two positions were also considered based on tibia tuberosity 1/3 line and tibia tuberosity end line known as a gold standard (Fig. 1-A). Loading and boundary conditions for the FE analysis were determined based on five activities of daily life of persons with TKA (Fig. 1-B). FE model was additionally validated comparing with an actual mechanical test.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 81 - 81
1 May 2016
Kang K Trinh T Jang Y Yoo O Lee M Lim D
Full Access

Introduction

Revision total knee arthroplasy (TKA) has been often used with a metal block augmentation for patients with poor bone quality. However, bone resorption beneath metal block augmentation has been still reported and little information about the reasons of the occurrence of bone resorption is available. The aim of the current study is to identify a possibility of the potential occurrence of bone resorption beneath metal block augmentation, through evaluation of strain distribution beneath metal block augmentation in revision TKA with metal block augmentation, during high deep flexion.

Materials and Method

LOSPA, revision TKA with a metal block augmentation (Baseplate size #5, Spacer size #5, Stem size Φ9, L30, Augment #5 T5) was considered in this study. For the test, the tibia component of LOSPA was implanted to the tibia sawbone (left, #3401, Sawbones EuropeAB, Malmö, Sweden), which was corresponded to a traditional TKR surgical guideline. The femoral component of LOSPA was mounted to a customized jig attached to the Instron 8872 (Instron, Norwood, MA, USA), which was designed specially to represent the angles ranged from 0° to 140° with consideration of a rollback of knee joint (Figure. 1). Here, a compressive load of 1,600N (10N/s) was applied for each angle. Strain distribution was then measured from rossete strain gauge (Half Bridge type, CAS, Seoul, Korea) together (Figure 1).


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 99 - 99
1 May 2016
Kim J Yoo O Seo J Jang Y Kim J Sun D Kim Y
Full Access

Introduction

A stem extension improves fixation stability of a tibial component. We need caution not to contact the tibial cortex with an offset adaptor. A symmetric tibial stem design often requires the component's re-positioning with negative effects. Therefore, the objective of this study was to validate clinical efficacy of a tibial baseplate with asymmetric stemmed position (TB-ASP) using aligning outlier rate. We hypothesized that TB-ASP design will be better aligned without unessential offset adaptor than a tibial baseplate with symmetric stemmed position (TB-SSP).

Methods

TB-ASP was designed based on the anthropometric standard model (58 female cadavers, 54.7±11.4 years)(Figure 1.). To validate the stem position, 3D bone models of 20 OA patients (71.8±7.2 years) was reconstructed. All virtual surgery has done by one surgeon with consistent surgical procedure for the analysis criteria.

An analysis of TB-ASP's aligning outlier was proceeded by following steps; 1) aligning tibial baseplate to the line from medial 1/3 tuberosity to the center of PCL, 2) selecting tibial baseplate's size for maximal bone coverage without problematic overhang, 3) trying to displace tibial baseplate and stem extension(120mm long) not to contact tibial cortex. A case invading tibial cortex was considered to be an outlier. The ratio using offset adaptor was compared to those of TB-SSP. Statistical analysis was performed using paired t-test.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 84 - 84
1 May 2016
Trinh T Kang K Lim D Yoo O Lee M Jang Y
Full Access

Introduction

Revision total knee arthroplasty (TKA) has been often used with a metal block augmentation for patients with poor bone quality. However, bone defects are frequently detected in revision TKA used with metal block augmentation. This study focused on identification of a potential possibility of the bone defect occurrence through the evaluation of the strain distribution on the cortical bone of the tibia implanted revision TKA with metal block augmentation, during high deep flexion.

Materials and Methods

Composite tibia finite element (FE) model was developed and revision TKA FE model with a metal block augmentation (Baseplate size #5 44AP/67ML, Spacer size #5 44AP/67ML, Stem size Φ9, L30, Augment #5 44AP/67ML thickness 5mm) was integrated with the composite tibia FE model. 0°, 30° 60°, 90°, 120° and 140° flexion positions were then considered with femoral rollback phenomenon [Fig 1.A]. A compressive load of 1,600N through the femoral component was applied to the composite tibia FE model integrated with the tibia component, sharing by the medial and lateral condyles, simulating a stance phase before toe-off [Fig 1.B].