Adjusting an external fixator can be a daunting process for patients. Despite comprehensive training, patients often request supervision for the initial adjustments which may result in a prolonged hospital stay. Following the introduction of telemedicine during the pandemic we believed that this could be utilised to support patients with their fixator adjustments. A quality improvement project was implemented to assess and evaluate a change in practice from existing Face to Face support to a telemedicine format. The aim of the project was to reduce median length of stay (LOS). The telemedicine platform was introduced in our unit from April 2021 with the change in practice. Using the life QI platform, run charts were used to record the numbers of patients whose LOS was 4 days or less. Median LOS was assessed prior to and following introduction of the telemedicine platform. Service user experience with telemedicine as well as overall training and education by the CNS team was sought through on-line questionnaires.Introduction
Materials & Methods
The FitBone lengthening nail (Orthofix UK) is an intramedullary device licensed for the lengthening of long bones in adults in the UK. It contains a motor powered by electricity transmitted via an induction coil placed underneath the skin. It was developed in Germany two decades ago but uptake in the UK has only started more recently. The aim of this study was to review the first cohort of FitBone lengthening nails in a unit with significant experience of other lengthening nails (including PRECICE and Stryde). Demographic, clinical and radiological data was prospectively collected on all FitBone cases starting in February 2022. Accuracy of lengthening rate, patient satisfaction and implant issues were all considered. Complications and learning points were recorded and discussed by the multidisciplinary team involved in the patients care.Introduction
Materials & Methods
A greater emphasis has been placed on fracture related infection (FRI) orthopaedic practice as a separate entity in recent years. Since the publication of the FRI consensus definition and guidelines, there has been an increase in the published literature on the topic and a move towards considering FRI as separate from general orthopaedic practice and as work that requires a more specialist approach. The aim of this study was to audit current FRI practice in the UK. Orthopaedic practice related to FRI in the UK was audited using a semi-structured questionnaire. Respondents were from a range of institutions, specialties and clinical roles to reflect the multi-disciplinary nature of treating FRI. The online tool SurveyMonkey was used to share the survey at the 2022 annual meeting of the British Limb Reconstruction Society. Twenty-one questions were asked in the following domains: scope of practice, theatre and clinic capacity, availability of the multidisciplinary team, renumeration for work and scope of FRI networks.Introduction
Materials & Methods
Patients undergoing limb reconstruction surgery often face a challenging and lengthy process to complete their treatment journey. The majority of existing outcome measures do not adequately capture the patient-reported outcomes relevant to this patient group in a single measure. Following a previous systematic review, the Stanmore Limb Reconstruction Score (SLRS) was designed with the intent to address this need for an effective instrument to measure patient-reported outcomes in limb reconstruction patients. We aim to assess the face validity of this score in a pilot study. The SLRS was designed following structured interviews with several groups including patients who have undergone limb reconstruction surgery, limb reconstruction surgeons, specialist nurses, and physiotherapists. This has subsequently undergone further adjustment for language and clarity. The score was then trialled on ten patients who had undergone limb reconstruction surgery, with subsequent structured questioning to understand the perceived suitability of the score.Aims
Methods
The use of intramedullary lengthening devices is becoming increasingly popular. There are no published data regarding the incidence of venous thromboembolism following intramedullary lengthening and no reports or guidance for current practices on use of thromboprophylaxis. Following a case of post-operative deep vein thrombosis in our institution, we felt that it is important to assess best practice. We conducted this survey to collect data that would describe current practice and help guide consensus for treatment. We have identified surgeons across the UK that perform intramedullary lengthening through the British Limb Reconstruction Society membership and a Precise Users database. Surgeons were contacted and asked to respond to an online survey (SurveyMonkey - SVMK Inc.). Responses to thromboprophylaxis regimes employed in their practice and cases of venous thromboembolism were collated.Introduction
Materials and Methods
The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited with regards to both outcomes and potential issues. In this paper we report on our early experience and raise awareness for the potential of adverse effects from this device. This is a review of all patients treated in our institution using this implant. Data were prospectively recorded. We report on demographics, nail accuracy, reliability, consolidation index and cases where concerning clinical and radiological findings were encountered.Introduction
Materials and Methods
Patients undergoing limb reconstruction surgery often face a challenging and often lengthy process to complete their treatment journey. The majority of existing outcome measures do not adequately capture the patient reported outcomes relevant to this patient group in a single measure. Following a previous systematic review, the Stanmore Limb Reconstruction Score (SLRS) was designed with the intent to address this need for an effective instrument to measure patient reported outcomes in limb reconstruction patients. The SLRS was designed following the use of structured interviews with a group of patients who have undergone limb reconstruction surgery, limb reconstruction surgeons, specialist nurses and physiotherapists. This has undergone further adjustment for language and clarity. The score was then trialled on 10 patients who have been through the process of limb reconstruction surgery, with subsequent structured questioning to understand the perceived suitability.Introduction
Materials and Methods
Pixel Value Ratio (PVR) is a radiographic measure of the relative density of the regenerate to the adjacent bone. This has been reported as an objective criterion for regenerate healing and a guide for when to allow full weight bearing (FWB) in lengthening with intramedullary telescopic nails. The threshold for which magnitude of PVR is adequate to allow bearing full weight is not yet agreed. The aim of this study was to identify from our cohort of adult limb lengthening patients the time to FWB following lengthening, the PVR at this point, and how this compared with the recommended values in the literature. A retrospective database review identified 30 adult patients treated with the PRECICE femoral nail by two senior authors. Time from completion of lengthening to instruction to fully weight bear was noted. The PVR was calculated for each cortex on plain radiographs taken at each postoperative visit following completion of lengthening. Significance was set at Introduction
Materials and Methods
Ring breakage is a rare but significant complication requiring revision surgery and prolonging the course of treatment. We have encountered three cases with Taylor Spatial Frames (TSF) with breakage at the half ring junction of the distal ring. This experimental study examines the strains produced at different locations on the distal ring during loading and the effects of altering the construct in order to develop techniques to minimise the risk of breakage. We mounted different TSF constructs on tibia sawbone models. Construct 1 reproducing the configuration of cases where failure was seen, Construct 2 with different wire and half pin configuration and construct 3 with the distal ring rotated 60 degrees. Strain Gauges were attached to different locations and measurements were collected during loading. Statistical analysis was subsequently performed.Introduction
Materials and Methods
To quantify the micro-motion at the fracture gap in a tibial fracture model stabilised with an external fixator. A surrogate model of a tibia and a cadaver leg were fractured and stabilised using a two-ring hexapod external fixator. They were tested initially under static loading and then subjected to vibration.Aim
Method
The management of a significant bony defect following excision
of a diaphyseal atrophic femoral nonunion remains a challenge. We
present the outcomes using a combined technique of acute femoral
shortening, stabilized with a long retrograde intramedullary nail,
accompanied by bifocal osteotomy compression and distraction osteogenesis with
a temporary monolateral fixator. Eight men and two women underwent the ‘rail and nail’ technique
between 2008 and 2016. Proximal locking of the nail and removal
of the external fixator was undertaken once the length of the femur
had been restored and prior to full consolidation of the regenerate.Aims
Patients and Methods
Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared.Aims
Patients and Methods
Several low energy osteotomy techniques are described in the literature but there is limited evidence comparing healing indices. We present a retrospective review of two techniques to evaluate an optimum method. Two cohorts of patients underwent osteotomy of the tibia using a Gigli saw (n=15) or DeBastiani corticotomy (n=12) technique. Indications for surgery included limb lengthening and bone transport for defect reconstruction with a minimal distraction of 2 cm. The patient radiographs were anonymised and the regenerate assessed by the two senior authors who were blinded to the osteotomy type. Bone quality was recorded along the anterior, posterior, medial and lateral cortices, graded 1–5 from absent to full consolidation over time in frame. The time to 3 cortices healed/regenerate length was calculated. The time to consolidation of the anterior, posterior, medial and lateral cortices were compared. The mean 3 cortices index in the Gigli group was 2.0 months/cm and in the DeBastiani group 1.8 months/cm, This was not a significant difference. In both groups anterior bone formation was slower, and in 50% and 33% of the Gigli and DeBastiani groups respectively the anterior cortex did not fully heal by the time of frame removal.Method:
Results:
The Exogen Ultrasound Device is licensed for fracture non-union under NICE guideline MTG-12. Reimbursement is offered if there is no fracture union after 250 days of treatment as per manufacturer guidelines. Quoted healing rates vary between 62% and 100%: we compare our results to these. A retrospective audit of all patients who received Exogen for long bone non-union was undertaken. The indication, duration of treatment, fracture outcome and compliance with the recommended treatment pathway were recorded.Background:
Method:
Little is known about the forces carried by the Taylor Spatial Frame (TSF) hexapod fixator. Our aim was to measure the TSF resultant force and how this changed during the consolidation phase. Five patients undergoing correction of tibial deformities were recruited. Measurements were taken at 2, 4, 8 and 12 weeks post-correction during various activities. Instrumented struts incorporating strain gauges measuring axial force were temporarily used each time. Strut forces and lengths were used to determine frame kinetics. The resultant axial fixator forces and moments were calculated relative to sitting. Ground reaction forces (GRF) were measured using the treadmill force plates.Background:
Method:
External fixators are not as well tolerated around the femur when compared to the tibia. Lengthening with an intramedullary device is therefore attractive. We reviewed all cases of femoral lengthening performed at our unit from 2007 to 2014. Cases of non-unions, concurrent deformities, congenital limb deficiencies and lengthening with an unstable hip were excluded. This left 33 cases for review. Healing index, implant tolerance and complications were compared.Background:
Method:
Leg length discrepancy related to absence of the femoral head with proximal migration of the femur presents a treatment dilemma. Late sequelae of neonatal hip sepsis and chronic hip dislocation due to developmental dysplasia are the most common causes. Traditional teaching dictates that the hip is stabilised prior to limb lengthening. Reconstructive options alter the shape of the proximal femur which complicates future surgery. We retrospectively reviewed 3 cases of femoral lengthening with an ‘unstable’ hip without prior stabilization. The aetiology was neonatal hip sepsis in 2 cases and chronic hip dislocation due to developmental dysplasia in 1 case. Lengthening was performed with the use of a retrograde Precice lengthening nail.Background:
Methods:
Pin site care in external fixation remains a major challenge. The ideal dressing regime still remains controversial as does the type of dressing. We present an audit following the use of a sponge compression dressing in comparison to previously used sterile gauze. All pins and wires were inserted using a standardised technique. Dressings were applied during surgery to prevent haematoma formation. The new sponge dressing was applied to the wire or half-pin sites and compressed to the skin by either pierced tubing over the wire or clips at the end of the procedure. Dressings were changed weekly unless the pin sites were discharging serous fluid or frankly infected, when they were changed as required. The pin sites were evaluated for evidence of infection or irritation using the good, bad or ugly grading system. Each evaluation was performed by two of the authors on two separate occasions to assess the inter- and intra-observer error. The results were compared to those previously reported using sterile gauze as the pin site dressing. 1035 pin sites were assessed. 985 pin sites were graded as “good” (95.2%), 49 “bad” (4.7%) and 1 “ugly”. The kappa values demonstrated excellent inter- and intra-observer correlation, 0.892 and 0.905 respectively. The previous study using gauze reported 600 pin sites with 514 graded “good” (85.7%), 80 “bad” (13.3%) and 6 “ugly” (1%). The use of compression sponge dressing appears to reduce the rate of pin site irritation and infection. Further anecdotal evidence by patients who have undergone surgical treatment using both techniques demonstrated a preference for the sponge dressings due to ease of dressing change and general comfort once applied. We therefore would recommend this type of dressing in the use of external fixation.
The aim of our study was to assess the use of the Clavicular Hook Plate in treating acromio–clavicular joint dislocations and fractures of the distal clavicle. The prospective study was carried out at two hospitals- a teaching hospital and a district general hospital. Between 2001 and 2004 a total of 37 patients with AC joint injuries and distal clavicle fractures were treated surgically with this device. Four of the patients had sustained a Neers Type 2 fracture of the distal clavicle, while 33 patients had acromio-clavicular joint dislocation (Rockwood Type 3 or higher). Mean age of the study group was 35.2 years. Post operatively, shoulder pendulum exercises were commenced on the second day and all patients discharged within 48 hours. During the first few weeks, we restricted shoulder abduction to 90 degrees. At the first postoperative follow up appointment at 2 weeks, average shoulder abduction was 30 degrees and forward elevation −40 degrees. This improved at 6 weeks to 85 degrees and 105 degrees respectively. The plates were removed at an average time interval of 11 weeks for the ACJ dislocations (range 8–12 weeks) and 15 weeks for the clavicle fractures (range 12–16 weeks). At three months after plate removal, we evaluated patients to measure the Visual Analogue Score(VAS) and Constant Score. The mean VAS was 1.4 (range 0–6) and the mean Constant score was 92 (range 72 to 98). Wound healing problems occurred in two patients, while two had a stress riser clavicle fracture. These had to be subsequently fixed with a Dynamic Compression Plate. One patient developed a superficial wound infection. Seven patients had problems due to impingement between the hook and the under surface of the acromion. A 45 year old female patient developed ACJ instability after plate removal. Radiographs revealed widening of the AC joint and some osteophyte formation. She went on to develop frozen shoulder which was treated with intensive physiotherapy. The AO hook plate represents an improvement over previous implants in treating injuries around the AC Joint. However, the need for a second operation to remove the plate remains a significant problem. Complications resulting from impingement were common in our patients and represent a major drawback of this implant.
Intramedullary Nailing is now accepted as the standard treatment for most femoral diaphyseal fractures. Most intramedullary nails are designed for proximal and distal locking with screws. We describe our experience with the Brooker Wills femoral nail. This nail is unique as distal fixation is achieved with as transverse fixator deployed through slots in the nail, a concept first enunciated by deCamargo in 1952. The fins of the fixator pierce the distal cortex when deployed thereby conferring rotational stability. The entire nail (including the proximal and distal fixation devices) can be inserted through a single proximal incision in the skin. We treated 17 patients with femoral shaft fractures using this system. 93% of the patients were males. Motor vehicle accidents accounted for 80 % of the fractures. Most fractures involved the middle third of the femur (54%), followed by distal third (33%) and proximal third (13%). 67 % of the fractures showed Winquist and Hansen Grade 3 or 4 comminution. All the nailings were performed in the supine position. Static locking was done in 16 cases. Post operative weight bearing was individualized with 86% of the patients bearing full weight before the end of 16 weeks. The average time to full weight bearing was 14 weeks. The mean time to union was 17.1 weeks, with proximal, middle and distal third fractures showing average healing times of 19, 15.6 and 18.8 weeks respectively. All the fractures united. There was one case of delayed union. Intra-operative complications included–inability to deploy the distal fixator in one case and unwinding of the reamer in another case. There was one case of superficial infection where a patient developed a sinus over the metallic fin of the distal fixation device, which had penetrated the cortex. This healed after the distal fixation device was removed. There was one case of proximal nail migration, but the fracture went onto union with some limb shortening. Results were evaluated using the scoring system devised by Sanders etal (1991). The functional criteria in their scoring system include- knee flexion, presence of pain, femur deformity, walking distance, stair climbing and pre injury functional status. We achieved excellent results in 53% of patients, good- 27%, fair–13 % and poor-7%. We feel that the main advantage of the Brooker Wills Nail is the ease of distal locking. This saves time and also leads to lower level of radiation exposure, as the image intensifier is sparingly used for distal locking. In fact, this nail is an attractive option in centres which lack an image intensifier. We achieved good results with this nailing system and feel that it is an acceptable alternative to other nailing systems with conventional locking systems with screws.