Introduction. Aseptic loosening is a major cause of revision of total joint arthroplasty (TJA). Although crosslinked
Introduction.
INTRODUCTION.
Since artificial joints are expected to operate for more than decades in human body, animal and clinical studies are not suitable for evaluation of their durability. Instead, in-vitro mechanical tests have been employed, but they cannot fully reproduce complex in-vivo mechanical and biochemical environment. For instance, lipids in synovial fluid have been known to be absorbed in
About 2% of primary total joint replacement arthroplasty (TJA) procedures become infected. Periprosthetic joint infection (PJI) is currently one of the main reasons requiring costly TJA revisions, posing a burden on patients, physicians and insurance companies.1 Currently used drug-eluting polymers such as bone cements offer limited drug release profiles, sometimes unable to completely clear out bacterial microorganisms within the joint space. For this study we determined the safety and efficacy of an antibiotic-eluting UHMWPE articular surface that delivered local antibiotics at optimal concentrations to treat PJI in a rabbit model. Skeletally mature adult male New Zealand White rabbits received either two non-antibiotic eluting UHMWPE (CONTROL, n=5) or vancomycin-eluting UHMWPE (TEST, n=5) (3 mm in diameter and 6 mm length) in the patellofemoral groove (Introduction
Materials and Methods
Alpha Lipoic Acid (L.A.) is an effective natural antioxidant discovered in the human body in 1951 from L.J. Reed and I.C. Gunslaus from liver. It is inside broccoli, spinach and red meats, especially liver and spleen. Actually it is largely used as antioxidant in antiaging products according to the low toxicity level of the product. The present study take into consideration the possibility to reduce oxidation of medical irradiated UHMWPE GUR 1050, mixing together polyethylene powder and Alpha Lipoic Acid powder. The study is composed of two parts. Part 1 Thermostability of alpha lipoic acid during polyethylene fusion Part 2 detection of oxygen level in artificially aged irradiated polyethylene Solid pieces were made with Gur 1050 powder (Ticona Inc., Bayport, Tex, USA) and mixed with Alpha Lipoic Acid (Talamonti, Italy, Stock 1050919074) 0,1%-‐0,3%-‐0,5%-‐1%-‐2% and gamma ray irradiated with 30 kGy (Isomedix, Northborough, MA). An owen (80° Celsius) was used to produce an aging effect for 35 days in the doped and control samples (Conventional not doped polyethylene). This process simulate an aging effect of 10 years into the human body. Part 1 : THERMAL STABILITY : a Fourier Transfer Infra Red (FTIR) test was made in pieces molded in a cell at 150° and 200°Celsius and pressure of 200 MPa comparing to the UHMWPE powder mixed with alpha lipoic acid. The presence of Alpha Lipoic Acid in the polyethylene was found at any depth in the manufacts. Part 2 : OXIDATION OF THE SURFACE : After 5 weeks at 80° Celsius in a owen (ASTM standard F-‐2003-‐02)A FOURIER TRANSFER INFRA RED TEST (FTIR) was made in the superficial layer and deeper on the undersurface of doped 0,1% and conventional UHMWPE. The antioxidation limit is defined as the ratio of the area under 1740cm/−‐1 carbonyl and 1370 cm/−‐1 Methylene absorbance peaks. In conventional UHMWPE oxidation is detected on the surface and decreases in the deeper layers down to zero under 1500 Micron. In the doped UHMWPE, FTIR demonstrate a very low oxidation limit on the surface and at any depth, comparing to conventional UHMWPE. The examples show that Lipoic Acid is effective as antioxidant in irradiated UHMWPE and it is stable with respect to thermal treatment. For any figures or tables, please contact authors directly.
Osteolysis caused by wear of the ultrahigh molecular weight polyethylene (UHMWPE) often leads to failure. Cross-linking improves wear, but also produces residual free radicals that decrease oxidative stability. 58 patients (64 observed hips), all with osteoarthritis, gave informed consent to participate in a 5 year RSA study. Each patient received a VEPE liner, a Regenerex™ acetabular shell, and an uncemented stem with either a 32mm or 36 mm cobalt chrome femoral head. Tantalum beads were inserted into the VEPE, the pelvic and the femoral bone to measure head penetration into the polyethylene, and shell and stem stability over time, using RSA. RSA radiographs were scheduled immediately postoperatively (up to 6 weeks) and 6 months, 1, 2, 3, and 5 years after surgery. The Wilcoxon signed-ranks nonparametric test was used to determine if changes in penetration or migration were significant over time at p≤0.05.Introduction
Methods
This prospective randomised controlled trial aims to compare the clinical and radiological outcomes of ceramic on ceramic, cobalt chrome on
Purpose. This prospective randomised controlled trial aims to compare the clinical and radiological outcomes of ceramic on ceramic, cobalt chrome on
Aim. This prospective randomised controlled trial aims to compare the clinical and radiological outcomes of ceramic on ceramic, cobalt chrome on
Wear of the
The purpose of this investigation is to assess the rate of wear the effect once the “bedding in period”/ poly creep had been eliminated. Creep is the visco-elastic deformation that polyethylene exhibits in the first 6–12 weeks. We also assessed the wear pattern of four different bearing couples in total hip arthroplasty (THA): cobalt-chrome (CoCr) versus oxidized zirconium (OxZir) femoral heads with
Polyethylene wear-debris induced inflammatory osteolysis is known as the main cause of aseptic loosening and long term revision total hip arthroplasty. Although recent reports suggest that antioxidant impregnated
Background. The anatomy of the human knee is very different than the tibiofemoral surface geometry of most modern total knee replacements (TKRs). Many TKRs are designed with simplified articulating surfaces that are mediolaterally symmetrical, resulting in non-natural patterns of motion of the knee joint [1]. Recent orthopaedic trends portray a shift away from basic tibiofemoral geometry towards designs which better replicate natural knee kinematics by adding constraint to the medial condyle and decreasing constraint on the lateral condyle [2]. A recent design concept has paired this theory with the concept of guided kinematic motion throughout the flexion range [3]. The purpose of this study was to validate the kinematic pattern of motion of the surface-guided knee concept through in vitro, mechanical testing. Methods. Prototypes of the surface-guided knee implant were manufactured using cobalt chromium alloy (femoral component) and
Introduction. Highly crosslinked,
Recent advancements in biomaterial technology have created novel options for acetabular fixation in primary total hip arthroplasty (THA). For example, cementless acetabular fixation has become the preferred option, however, there is continued debate concerning whether long-term survivorship is comparable to that of cemented component fixation. Many doubts previously associated with early cementless designs have been addressed with newer features such as improved locking mechanisms, enhanced congruity between the acetabular liner and the shell, and the inclusion of highly cross-linked
Introduction. Despite the positive outcomes in shoulder joint replacements in the last two decades, polyethylene wear debris in metal-on-polyethylene artificial shoulder joints is well-known as a limitation in the long-term survival of shoulder arthroplasties systems. Consequently, there is an interest in the use of novel materials as an alternative to hard bearing surfaces such as pyrolytic carbon layer (PyroCarbon). Materials and Methods. In the present study, the unique Newcastle Shoulder Wear Simulator was used (Smith et al., 2015; Smith et al., 2016) to evaluate the wear behavior of four commercially available PyroCarbon humeral heads 43 mm diameter, articulating against conventional
Introduction. Orthopedic implants are subject to wear and release
INTRODUCTION. Shoulder joint prostheses have become the most commonly replaced after knee and hip artificial implants. Reverse shoulder arthroplasty (RSA) is the treatment option for patients with severe osteoarthritis, rotator cuff arthropathy or a massive rotator cuff tear with pseudoparalysis. Though successful, the long-term survival of such implants are limited by wear of the materials in contact [1, 2]. The aim of this study was to investigate RSA wear in vitro using a clinically relevant activities of daily living (ADLs). MATERIALS AND METHODS. Four new JRI Orthopaedics Reverse Shoulder 42 mm diameter VAIOS with cobalt-chromium (CoCr) glenospheres and
Introduction. Total hip replacement with metal-on-polymer (MoP) hip prostheses is a successful treatment for late-stage osteoarthritis. However, the wear debris generated from the polymer acetabular liners remains a problem as it can be associated with osteolysis and aseptic loosening of the implant. This has led to the investigation of more wear resistant polymers in orthopaedics. Cross-linked polyethylene (XLPE) is now the gold-standard acetabular liner material. However, we asked if carbon fibre reinforced polyether ether ketone (CFR-PEEK) might be a lower wear material. In addition, we sought to understand the influence of contact stress on the wear of both XLPE and CFR-PEEK as this has not previously been reported. Materials and Methods. A 50-station circularly translating pin-on-disc (SuperCTPOD) machine was used to wear test both XLPE and CFR-PEEK pins against cobalt chromium (CoCr) discs to investigate the influence of contact stress on their wear rates. Fifty XLPE and 50 CFR-PEEK pins were articulated against CoCr discs. The pins, 9 mm in outer diameter and 12 mm in height, were drilled with different diameter holes to generate different sized annuli and thus, different contact areas. The pins were tested at 1.10, 1.38, 1.61, 2.00 and 5.30 MPa, which are typical contact stresses observed in the natural hip joint. An additional pin for every test group was used as a control to track the lubricant uptake. The discs were polished to 0.015 μm Sa prior to testing. The test stations contained 16 ml of diluted newborn calf serum (protein concentration: 22 g/L). Wear was measured gravimetrically with a balance (resolution: 10 μm) every 500,000 cycles. A standardised cleaning and weighing protocol was followed. Results and Discussion. The wear rates for the XLPE pins were calculated as 1.05, 0.90, 0.77, 0.48 and 0.28 mg/million cycles for the different pin stress groups respectively. The wear rates decreased with increasing contact stress, which was similar to what was observed for