The ideal treatment method regarding various defect sizes after local aggressive tumor resection is unknown. We investigated the biomechanical properties of metaphyseal defect filling regarding different defect sizes and fixation methods. Ninety-one sheep tibias were divided into five groups as 21 tibias per four study groups and 7 tibias in the control group. Study groups were further divided into three subgroups according to 25%, 50% and 75% metaphyseal defect size. Control group tibias were left intact. In study group 1, a metaphyseal defect was created and no further process was applied. Metaphyseal defects were filled with cement without fixation in group 2. Cement filling and fixation with 2 screws were performed in group 3. In addition to cement filling, plate-screw fixation was performed in group 4. Axial loading test was applied to all tibias and the results were compared between study subgroups and control group. Plate-screw fixation was found to have the best biomechanical properties in all defect sizes.
Extensor mechanism and abductor reconstructions in total joint arthroplasty are problematic. Growing tendon into a metallic implant would have great reconstructive advantages. With the introduction of porous metal implants, it was hoped that tendons could be directly attached to implants. However, the effects of the porous metal structure on tissue growth and pore penetration is unknown. In this rat model, we investigated the effect of pore size on tendon repair fixation using printed titanium implants with differing pore sizes. There were four groups of six Sprague Dawley rats (n = 28) plus control (n=4). Implants had pore sizes of 400µm (n=8), 700µm (n=8), and 1000µm (n=8). An Achilles tendon defect was created, and the implant positioned and sutured between the cut ends. Harvest occurred at 12-weeks. Half the specimens underwent tensile
This study aims to compare the biomechanical properties of the “Double Lasso-Loop” suture anchor (DLSA) technique with the commonly performed interference screw (IS) technique in an ex vivo ovine model. Fourteen fresh sheep shoulder specimens were used in this study. Dissection was performed leaving only the biceps muscle attached to the humerus and proximal radius before sharply incised to simulate long head of biceps tendon (LHBT) tear. Repair of the LHBT tear was performed on all specimens using either DSLA or IS technique. Cyclical loading of 500 cycles followed by
Proximal humeral shaft fractures are commonly treated with long straight locking plates endangering the radial nerve distally. The aim of this study was to investigate the biomechanical competence in a human cadaveric bone model of 90°-helical PHILOS plates versus conventional straight PHILOS plates in proximal third comminuted humeral shaft fractures. Eight pairs of humeral cadaveric humeri were instrumented using either a long 90°-helical plate (group1) or a straight long PHILOS plate (group2). An unstable proximal humeral shaft fracture was simulated by means of an osteotomy maintaining a gap of 5cm. All specimens were tested under quasi-static loading in axial compression, internal and external rotation as well as bending in 4 directions. Subsequently, progressively increasing internal rotational
Recently, a new suture was designed to minimize laxity in order to preserve consistent tissue approximation while improving footprint compression after tendon repair. The aims of this study were: (1) to compare the biomechanical competence of two different high strength sutures in terms of slippage and
Bone turnover and microdamage are impacted by skeletal metastases which can contribute to increased fracture risk. Treatments for metastatic disease may further impact bone quality. This study aimed to establish an understanding of microdamage accumulation and
Operative fixation is the treatment of choice for a rupture of the distal tendon of biceps. A variety of techniques have been described including transosseous sutures and suture anchors. The poor quality of the bone of the radial tuberosity might affect the
We compared time-dependent changes in the biomechanical properties of single-and double-row repair of a simulated acute tear of the rotator cuff in rabbits to determine the effect of the fixation techniques on the healing process. A tear of the supraspinatus tendon was created in 80 rabbits which were separated into two equal groups. A single-row repair with two suture anchors was conducted in group 1 and a double-row repair with four suture anchors in group 2. A total of ten intact contralateral shoulder joints was used as a control group. Biomechanical testing was performed immediately post-operatively and at four and eight weeks, and histological analysis at four and eight weeks. The mean
Although 3D-printed porous dental implants may possess improved osseointegration potential, they must exhibit appropriate fatigue strength. Finite element analysis (FEA) has the potential to predict the fatigue life of implants and accelerate their development. This work aimed at developing and validating an FEA-based tool to predict the fatigue behavior of porous dental implants. Test samples mimicking dental implants were designed as 4.5 mm-diameter cylinders with a fully porous section around bone level. Three porosity levels (50%, 60% and 70%) and two unit cell types (Schwarz Primitive (SP) and Schwarz W (SW)) were combined to generate six designs that were split between calibration (60SP, 70SP, 60SW, 70SW) and validation (50SP, 50SW) sets. Twenty-eight samples per design were additively manufactured from titanium powder (Ti6Al4V). The samples were tested under bending compression loading (ISO 14801) monotonically (N=4/design) to determine ultimate load (F. ult. ) (Instron 5866) and cyclically at six load levels between 50% and 10% of F. ult. (N=4/design/
Introduction and Objective. Distal femoral fractures are commonly treated with a straight plate fixed to the lateral aspects of both proximal and distal fragments. However, the lateral approach may not always be desirable due to persisting soft-tissue or additional vascular injury necessitating a medial approach. These problems may be overcome by pre-contouring the plate in helically shaped fashion, allowing its distal part to be fixed to the medial aspect of the femoral condyle. The objective of this study was to investigate the biomechanical competence of medial femoral helical plating versus conventional straight lateral plating in an artificial distal femoral fracture model. Materials and Methods. Twelve left artificial femora were instrumented with a 15-hole Locking Compression Plate – Distal Femur (LCP-DF) plate, using either conventional lateral plating technique with the plate left non-contoured, or the medial helical plating technique by pre-contouring the plate to a 180° helical shape and fixing its distal end to the medial femoral condyle (n=6). An unstable extraarticular distal femoral fracture was subsequently simulated by means of an osteotomy gap. All specimens were tested under quasi-static and progressively increasing cyclic axial und torsional
Suture anchor have been used in surgical procedure of tendon or ligament repair. Recently, there has been developed an all suture anchor (soft anchor) which can be used even when the insertion area is narrow. But, the stability of soft anchors due to narrow zone has not been elucidated. This purpose of this study was to investigate stability of soft anchors with respect to their fixation intervals. Polyurethane foams with two different bone densities (10 pcf; 0.16g / cm³, 20 pcf; 0.32g / cm³) were used. All suture anchors and conventional suture anchors were fixed at 10mm, 5mm, and 2.5mm intervals. The
There is a growing trend towards using pre-clinical models of atrophic non-union. This study investigated different fixation devices, by comparing the mechanical stability at the fracture site of tibia bone fixed by either intramedullary nail, compression plate or external fixator. 40 tibias from adult male Wistar rats' cadavers were osteotomised at the mid-shaft and a gap of 1 mm was created and maintained at the fracture site to simulate criteria of atrophic non-union model. These were divided into five groups (n=8 in each): the first group was fixed with 20G intramedullary nail, the second group with 18G nail, the third group with 4-hole plate, the fourth group with 6-hole plate, and the fifth group with external fixator. Tibia was harvested by leg disarticulation from the knee and ankle joints, the soft tissues were carefully removed from the leg, and tibias were kept hydrated throughout the experiment. Each group was then subdivided into two subgroups for mechanical testing: one for axial loading (n=4) and one for 4-point bending (n=4). Statistical analysis was carried out by ANOVA with a fisher post-hoc comparison between groups. A p-value less than 0.05 was considered statistically significant. Axial
There is a growing trend towards using pre-clinical models of atrophic non-union. This study investigated different fixation devices, by comparing the mechanical stability at the fracture site of tibia bone fixed by either intramedullary nail, compression plate or external fixator. 40 tibias from adult male Wistar rats' cadavers were osteotomised at the mid-shaft and a gap of 1 mm was created and maintained at the fracture site to simulate criteria of atrophic non-union model. These were divided into five groups (n=8 in each): the first group was fixed with 20G intramedullary nail, the second group with 18G nail, the third group with 4-hole plate, the fourth group with 6-hole plate, and the fifth group with external fixator. Tibia was harvested by leg disarticulation from the knee and ankle joints, the soft tissues were carefully removed from the leg, and tibias were kept hydrated throughout the experiment. Each group was then subdivided into two subgroups for mechanical testing: one for axial loading (n=4) and one for 4-point bending (n=4). Statistical analysis was carried out by ANOVA with a fisher post-hoc comparison between groups. A p-value less than 0.05 was considered statistically significant. Axial
The concept of decellularised xenografts as a basis for anterior cruciate ligament (ACL) reconstruction was introduced to overcome limitations in alternative graft sources such as substantial remodelling delaying recovery and donor site morbidity. This study aimed to measure the biomechanical properties of decellularised porcine super flexor tendon (pSFT) processed to create ACL grafts of varying diameters, with a view to facilitating production of stratified ‘off the shelf’ products with specified functional properties for use in ACL reconstructive surgery. Decellularisation was carried out using a previously established procedure, including antibiotic washes, low concentration detergent (0.1% sodium dodecyl sulphate) washes and nuclease treatments. Decellularised pSFTs were prepared to create double-bundle grafts of 7, 8 and 9mm diameter (n=6 in each group). Femoral and tibial fixations were simulated utilising Arthrex suspension devices (Tightrope®) and interference screws in bovine bone respectively. Dynamic stiffness and creep were measured under cyclic loading between 50–250N for 1000 cycles at 1Hz. This was followed by ramp to failure at 200mm/min from which linear stiffness and
Abstract. Objectives. To evaluate mechanical properties of three suture-tendon constructs, the Krackow stitch (KS), the modified Prusik knot (PK) and the Locking SpeedWhip (LSW), using human cadaveric quadriceps grafts (QT). Methods. Thirty QT grafts were obtained from human cadaver specimens and an equal number of tendon-suture constructs were prepared for three stitches: KS, PK and LSW. The constructs were mounted in a materials testing machine (ElectroPuls E10000, Instron, Norwood, MA) and subject to tensile loading based on an established protocol. Load and displacement data for each tendon-suture construct were recorded. Results. Seven of 10 LSW specimens failed due to suture pullout before completing cyclic loading. Comparisons of the 3 successful LSW specimens (LSW3) were made to the KS and PK groups. All KS and PK specimens failed by suture breakage in
Introduction. Suture anchor have been used in surgical procedure of tendon or ligament repair. Recently, there has been developed an all suture anchor (soft anchor) which can be used even when the insertion area is narrow. But, the stability of soft anchors due to narrow zone has not been elucidated. This purpose of this study was to investigate stability of soft anchors with respect to their fixation intervals. Methods. Polyurethane foams with two different bone densities (10 pcf; 0.16g / cm³, 20 pcf; 0.32g / cm) were used. All suture anchors and conventional suture anchors were fixed at 10mm, 5mm, and 2.5mm intervals. The
A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in
The anterolateral ligament (ALL) has been recently recognized as a distinct stabilizer for internal rotation in the ACL-deficient knee and it has been hypothesized that ALL reconstruction may play an important role in improving anterolateral instability following ACL reconstruction. Both the gracilis tendon (GT) and a portion of the iliotibial band (ITB) have been suggested as graft materials for ALL reconstruction, however, there is an ongoing debate concerning whether GT or ITB are appropriate grafting materials. Furthermore, there is limited knowledge in how the mechanical properties of these potential grafts compare to the native ALL. Consequently, the aim of this study was to characterize the elastic (Young's modulus and failure load) and viscoelastic (dynamic and static creep) mechanical properties of the ALL and compare these results with the characteristics of the grafting materials (GT and ITB), in order to provide guidance to clinicians with respect to graft material choice. Fourteen fresh-frozen cadaveric knees (85.2±12.2 yr) were obtained. The ALL, ITB, and the distal (GTD) and proximal gracilis tendons (GTP) (bisected at mid portion) were harvested from each donor and tested with a dynamic material testing frame. Prior to testing, the cross-sectional area of each tissue was measured using a casting method and the force required to achieve a min-max stress (1.2–12 MPa) for the testing protocol was calculated (preconditioning (20 cycles, 3–6 MPa), sinusoidal cycle (200 cycles, 1.2–12 MPa), dwell at constant load (100 s, 12 MPa), and
Coronoid fractures account for 2 to 15% of the cases with elbow dislocations and usually occur as part of complex injuries. Comminuted fractures and non-unions necessitate coronoid fixation, reconstruction or replacement. The aim of this biomechanical study was to compare the axial stability achieved via an individualized 3D printed prosthesis with curved cemented intramedullary stem to both radial head grafted reconstruction and coronoid fixation with 2 screws. It was hypothesized that the prosthetic replacement will provide superior stability over the grafted reconstruction and screw fixation. Following CT scanning, 18 human cadaveric proximal ulnas were osteotomized at 40% of the coronoid height and randomized to 3 groups (n = 6). The specimens in Group 1 were treated with an individually designed 3D printed stainless steel coronoid prosthesis with curved cemented intramedullary stem, individually designed based on the contralateral coronoid scan. The ulnas in Group 2 were reconstructed with an ipsilateral radial head autograft fixed with two anteroposterior screws, whereas the osteotomized coronoids in Group 3 were fixed in situ with two anteroposterior screws. All specimens were biomechanically tested under ramped quasi-static axial
Summary. We found an increased natural expression of the growth factors bFGF, BMP-12, VEGF, and TGF-b1 during tendon healing of rat Achilles tendons. External application of these growth factors improved the tendons