Aims. Knee osteonecrosis in advanced stages may lead to joint degeneration. Total knee arthroplasty (TKA) for osteonecrosis has traditionally been associated with suboptimal results. We analyzed outcomes of contemporary TKAs for osteonecrosis, with particular emphasis on: survivorship free from aseptic loosening, any revision, and any reoperation plus the clinical outcomes, complications, and radiological results. Patients and Methods. In total, 156 patients undergoing 167 primary TKAs performed for osteonecrosis between 2004 and 2014 at a single institution were reviewed. The mean age at index TKA was 61 years (14 to 93) and the mean body mass index (BMI) was 30 kg/m. 2. (18 to 51) The mean follow-up was six years (2 to 12). A total of 110 TKAs (66%) were performed for primary osteonecrosis and 57 TKAs (34%) for secondary osteonecrosis. Overall, 15 TKAs (9%) had tibial stems, while 12 TKAs (7%) had
Aims. There is little literature about total knee arthroplasty (TKA) after distal femoral osteotomy (DFO). Consequently, the purpose of this study was to analyze the outcomes of TKA after DFO, with particular emphasis on: survivorship free from aseptic loosening, revision, or any re-operation; complications; radiological results; and clinical outcome. Patients and Methods. We retrospectively reviewed 29 patients (17 women, 12 men) from our total joint registry who had undergone 31 cemented TKAs after a DFO between 2000 and 2012. Their mean age at TKA was 51 years (22 to 76) and their mean body mass index 32 kg/m. 2. (20 to 45). The mean time between DFO and TKA was ten years (2 to 20). The mean follow-up from TKA was ten years (2 to 16). The prostheses were posterior-stabilized in 77%, varus-valgus constraint (VVC) in 13%, and cruciate-retaining in 10%. While no patient had metaphyseal fixation (e.g. cones or sleeves), 16% needed a
Introduction. The number of medial unicompartmental knee replacements (UKR) performed for arthritis has increased and as such, revisions to total knee replacement (TKR) is increasing. Previous studies have investigated survivorship of UKR to TKR revision and functional outcomes compared to TKR to TKR revision, but have failed to detail the surgical considerations involved in these revisions. Our objectives are to investigate the detailed surgical considerations involved in UKR to TKR revisions. Methods. This study is a retrospective comparative analysis of a prospectively collected database. From 2005 to 2017, 61 revisions of UKR to TKR were completed at a single center. Our inclusion criteria included: revision of UKR to TKR or TKR to TKR with minimum 1 year follow-up. Our exclusion criteria include: single component and liner revisions and revision for infection. The 61 UKR to TKR revisions were matched 2:1 with respect to age, ASA and BMI to a group of 122 TKR to TKR revisions. The following data was collected: indication for and time to revision, operative skin to skin surgical time, the use of specialized equipment (augment size/location, stem use), intraoperative and postoperative complications, re-operations and outcome scores (WOMAC, Oxford 12, SF 12, satisfaction score). Results. There were no statistical differences between the demographic data from either group (age, BMI, ASA, sex and follow-up range). Progression of arthritis was the most common reason for revision in the UKR to TKR group (30/61, 49%, p < 0.001). Aseptic loosening was the most common reason for revision in the TKR to TKR group (73/122, 60%,) and was encountered more often than aseptic loosening in the UKR to TKR group (21/61, 35%, p=0.002). The operative time was longer in the TKR to TKR group (77 vs 112 min, p< 0.001). Femoral augmentation was required for one 1/61 (1.64%) UNI and 92/122 (75%) TKR revisions, respectively (p <0.001). Medial tibial augments were required in 9/61 (14.8%) of the UKR to TKR group while 12/122 (10%) and 10/122 (8%) of the TKR to TKR group required medial and full tibial augments, respectively (p=0.7). UKR to TKR revisions never required
Introduction. In the setting of periprosthetic joint infection, the complete removal of implants and cement can be challenging with well-fixed, cemented implants about the knee. This can get especially complex in the setting of long cemented
Aims. Varus-valgus constrained (VVC) implants are often used during revision total knee arthroplasty (TKA) to gain coronal plane stability. However, the increased mechanical torque applied to the bone-cement interface theoretically increases the risk of aseptic loosening. We assessed mid-term survivorship, complications, and clinical outcomes of a fixed-bearing VVC device in revision TKAs. Methods. A total of 416 consecutive revision TKAs (398 patients) were performed at our institution using a single fixed-bearing VVC TKA from 2007 to 2015. Mean age was 64 years (33 to 88) with 50% male (199). Index revision TKA diagnoses were: instability (n = 122, 29%), aseptic loosening (n = 105, 25%), and prosthetic joint infection (PJI) (n = 97, 23%). All devices were cemented on the epiphyseal surfaces.
Aseptic loosening is the most common cause of failure following cemented total knee arthroplasty (TKA), and has been linked to poor cementation technique. We aimed to develop a consensus on the optimal technique for component cementation in TKA. A UK-based, three-round, online modified Delphi Expert Consensus Study was completed focusing on cementation technique in TKA. Experts were identified as having a minimum of five years’ consultant experience in the NHS and fulfilling any one of the following criteria: a ‘high volume’ knee arthroplasty practice (> 150 TKAs per annum) as identified from the National joint Registry of England, Wales, Northern Ireland and the Isle of Man; a senior author of at least five peer reviewed articles related to TKA in the previous five years; a surgeon who is named trainer for a post-certificate of comletion of training fellowship in TKA.Aims
Methods
Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.Aims
Methods
Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones. A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m2 (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their knee alignment based on the type of stem used. Overall, 22 knees had cementless stems on both sides, 52 had cemented stems on both sides, and 15 had mixed stems. Postoperative long-standing radiographs were evaluated for coronal and sagittal plane alignment. Adjusted logistic regression models were run to assess malalignment hip-knee-ankle (HKA) alignment beyond ± 3° and sagittal alignment of the tibial and femoral components ± 3° by stem type.Aims
Methods
To describe the risk of periprosthetic joint infection (PJI) and reoperation in patients who have an acute, traumatic wound dehiscence following total knee arthroplasty (TKA). From January 2002 to December 2018, 16,134 primary TKAs were performed at a single institution. A total of 26 patients (0.1%) had a traumatic wound dehiscence within the first 30 days. Mean age was 68 years (44 to 87), 38% (n = 10) were female, and mean BMI was 34 kg/m2 (23 to 48). Median time to dehiscence was 13 days (interquartile range (IQR) 4 to 15). The dehiscence resulted from a fall in 22 patients and sudden flexion after staple removal in four. The arthrotomy was also disrupted in 58% (n = 15), including a complete extensor mechanism disruption in four knees. An irrigation and debridement with component retention (IDCR) was performed within 48 hours in 19 of 26 knees and two-thirds were discharged on antibiotic therapy. The mean follow-up was six years (2 to 15). The association of wound dehiscence and the risk of developing a PJI was analyzed.Aims
Methods
Objectives. Orthopaedic surgeons use stems in revision knee surgery to obtain
stability when metaphyseal bone is missing. No consensus exists
regarding stem size or method of fixation. This in vitro study
investigated the influence of stem length and method of fixation
on the pattern and level of relative motion at the bone–implant
interface at a range of functional flexion angles. Methods. A custom test rig using differential variable reluctance transducers
(DVRTs) was developed to record all translational and rotational
motions at the bone–implant interface. Composite femurs were used.
These were secured to permit variation in flexion angle from 0°
to 90°. Cyclic loads were applied through a tibial component based
on three peaks corresponding to 0°, 10° and 20° flexion from a normal
walking cycle. Three different femoral components were investigated
in this study for cementless and cemented interface conditions. Results. Relative motions were found to increase with flexion angle. Stemmed
implants reduced relative motions in comparison to stemless implants
for uncemented constructs. Relative motions for cemented implants
were reduced to one-third of their equivalent uncemented constructs. Conclusions. Stems are not necessary for cemented implants when the metaphyseal
bone is intact. Short cemented
Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system. We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).Aims
Methods
Preservation of posterior condylar offset (PCO) has been shown to correlate with improved functional results after primary total knee arthroplasty (TKA). Whether this is also the case for revision TKA, remains unknown. The aim of this study was to assess the independent effect of PCO on early functional outcome after revision TKA. A total of 107 consecutive aseptic revision TKAs were performed by a single surgeon during an eight-year period. The mean age was 69.4 years (39 to 85) and there were 59 female patients and 48 male patients. The Oxford Knee Score (OKS) and Short-form (SF)-12 score were assessed pre-operatively and one year post-operatively. Patient satisfaction was also assessed at one year. Joint line and PCO were assessed radiographically at one year.Objectives
Methods
Risk of revision following total knee arthroplasty (TKA) is higher
in patients under 55 years, but little data are reported regarding
non-revision outcomes. This study aims to identify predictors of
dissatisfaction in these patients. We prospectively assessed 177 TKAs (157 consecutive patients,
99 women, mean age 50 years; 17 to 54) from 2008 to 2013. Age, gender,
implant, indication, body mass index (BMI), social deprivation,
range of movement, Kellgren-Lawrence (KL) grade of osteoarthritis
(OA) and prior knee surgery were recorded. Pre- and post-operative
Oxford Knee Score (OKS) as well as Short Form-12 physical (PCS)
and mental component scores were obtained. Post-operative range
of movement, complications and satisfaction were measured at one
year.Aims
Patients and Methods
The purpose of this study was to develop an accurate, reliable and easily applicable method for determining the anatomical location of the joint line during revision knee arthroplasty. The transepicondylar width (TEW), the perpendicular distance between the medial and lateral epicondyles and the distal articular surfaces (DMAD, DLAD) and the distance between the medial and lateral epicondyles and the posterior articular surfaces (PMAD, DLAD) were measured in 40 knees from 20 formalin-fixed adult cadavers (11 male and nine female; mean age at death 56.9 years, Objectives
Methods
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total knee prosthesis. The osteolytic defects were types II and
III (A or B) according to the Anderson Orthopaedic Research Institute
(AORI) classification. The mean age of the patients was 72 years
(62 to 84) and there were 15 men and 11 women. In this series patients had
undergone a mean of 2.34 previous total knee arthroplasties. The
main objective was to restore anatomy along with stability and function
of the knee joint to allow immediate full weight-bearing and active
knee movement. Outcomes were measured using Knee Society scores,
Oxford knee scores, range of movement of the knee and serial radiographs.
Patients were followed for a mean of 36 months (24 to 49). The mean
Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20
(32 to 38) (p <
0.001) and mean American Knee Society scores
improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p <
0.001).
No radiolucent lines suggestive of loosening were seen around the trabecular
metal cones, and by one year all the radiographs showed good osteo-integration.
There was no evidence of any collapse or implant migration. Our
early results confirm the findings of others that trabecular metal
cones offer a useful way of managing severe bone loss in revision
TKR. Cite this article:
This was a retrospective analysis of the medium-
to long-term results of 46 TC3 Sigma revision total knee replacements
using long uncemented stems in press-fit mode. Clinical and radiological analysis took place pre-operatively,
at two years post-operatively, and at a mean follow-up of 8.5 years
(4 to 12). The mean pre-operative International Knee Society (IKS)
clinical score was 42 points (0 to 74), improving to 83.7 (52 to
100) by the final follow-up. The mean IKS score for function improved
from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up.
At the final follow-up 30 knees (65.2%) had an excellent result, seven
(15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor
result. There were two failures, one with anteroposterior instability
and one with aseptic loosening. The TC3 revision knee system, when used with press-fit for long
intramedullary stems and cemented femoral and tibial components,
in both septic and aseptic revisions, results in a satisfactory
clinical and radiological outcome, and has a good medium- to long-term
survival rate.
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.
Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR. A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by >
3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47). We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.
It is unclear whether there is a limit to the amount of distal bone required to support fixation of supracondylar periprosthetic femoral fractures. This retrospective multicentre study evaluated lateral locked plating of periprosthetic supracondylar femoral fractures and compared the results according to extension of the fracture distal with the proximal border of the femoral prosthetic component. Between 1999 and 2008, 89 patients underwent lateral locked plating of a supracondylar periprosthetic femoral fracture, of whom 61 patients with a mean age of 72 years (42 to 96) comprising 53 women, were available after a minimum follow-up of six months or until fracture healing. Patients were grouped into those with fractures located proximally (28) and those with fractures that extended distal to the proximal border of the femoral component (33). Delayed healing and nonunion occurred respectively in five (18%) and three (11%) of more proximal fractures, and in two (6%) and five (15%) of the fractures with distal extension (p = 0.23 for delayed healing; p = 0.72 for nonunion, Fisher’s exact test). Four construct failures (14%) occurred in more proximal fractures, and three (9%) in fractures with distal extension (p = 0.51). Of the two deep infections that occurred in each group, one resolved after surgical debridement and antibiotics, and one progressed to a nonunion. Extreme distal periprosthetic supracondylar fractures of the femur are not a contra-indication to lateral locked plating. These fractures can be managed with internal fixation, with predictable results, similar to those seen in more proximal fractures.