Advertisement for orthosearch.org.uk
Results 1 - 20 of 99
Results per page:
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 110 - 110
1 May 2016
Wada H Mishima H Yoshizawa T Sugaya H Nishino T Yamazaki M
Full Access

Introduction. Hydroxyapatite and poly-L-lactide (HA/PLLA) composites are osteoconductive and biodegradable. They have already been used clinically to treat fractured bones by inducing osteosynthesis and serving as the bone filling material. During revision of total hip arthroplasty, we have grafted bone onto the bone defect and covered it with an HA/PLLA mesh instead of using a metal mesh on the non-load bearing portion of the cup (Figure 1). However, whether the interface between the HA/PLLA and the titanium alloy cup was stable remains unclear. Objectives. The purpose of this study was to determine and compare the histological osteoconductivity and osteoinductivity of HA/PLLA and titanium alloy. Methods. Semicylindrical samples (diameter: 3 mm; height: 5 mm) were fabricated from a composite unsintered-HA/PLLA containing 40% fine HA particles and from titanium alloy (Ti-6AL-4V). These two samples were combined to form one cylinder (Figure 2). Defects 3 mm in diameter were drilled into both femoral condyles of nine Japanese white rabbits, and the samples were implanted by press fitting. The rabbits were euthanized at 2, 4, 8, and 25 weeks after implantation, and undecalcified ground samples were prepared. New bone formation was examined histologically using Toluidine blue and Villanueva Goldner stains. Results. New bone formation was observed around the sample at 4 weeks, and the amount increased by 8 weeks. In addition, partial remodeling of the trabecular bones and absorption of the HA/PLLA were found at 25 weeks. Small amounts of new bone formation were found at 4 weeks between the HA/PLLA and titanium alloy materials (Figure 3: Toluidine blue stain), and the amount increased at 8 and 25 weeks. The HA/PLLA had been slightly absorbed and new bone was formed in the gap, which was close to the border between the materials, at 25 weeks. However, the amount of absorption was limited, and no new bone was found in samples where the materials were firmly in contact. Conclusions. HA/PLLA was only slight absorbed at 25 weeks, suggesting that it was stable in vivo and has good osteoconductive and osteoinductive properties. No new bone was found in the regions where the sample was stable and had no gaps between the HA/PLLA and titanium alloy, probably because there was no space for new bone to form in those regions. In contrast, new bone formation was found in gaps of more than 20 μm. Clinically, many gaps likely exist, allowing new bone formation to occur even in a stable implant. This may stabilize the HA/PLLA and titanium alloy materials for longer times. As expected, the HA/PLLA and titanium alloy were mostly stablein vivo


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 4 - 4
1 Apr 2019
Tamura J Asada Y Oota M Matsuda Y
Full Access

Introduction. We have investigated the long-term (minimum follow-up period; 10 years) clinical results of the total hip arthroplasty (THA) using K-MAX HS-3 tapered stem. Materials and Methods. In K-MAX HS-3 THA (Kyocera Medical, Kyoto, Japan), cemented titanium alloy stem and all polyethylene cemented socket are used. This stem has the double tapered symmetrical stem design, allowing the rotational stability and uniform stress distribution. The features of this stem are; 1. Vanadium-free high-strength titanium alloy (Ti-15Mo-5Zr-3Al), 2. Double-tapered design, 3. Smooth surface (Ra 0.4µm), 4. Broad proximal profile, 5. Small collar. Previous type stem, which was made of the same smooth-surface titanium alloy, has the design with cylindrical stem tip, allowing the maximum filling of the femoral canal. Osteolysis at the distal end of the stem had been reported in a few cases in previous type with cylindrical stem tip, probably due to the local stress concentration. Therefore the tapered stem was designed, expecting better clinical results. 157 THAs using HS-3 taper type stem were performed at Kitano Hospital between March 2004 and March 2008. And 101 THAs, followed for more than 10 years, were investigated (follow-up rate; 64.3%). The average age of the patients followed at the operation was 61.7 years and the average follow-up period was 10.9 years. The all-polyethylene socket was fixed by bone cement, and the femoral head material was CoCr (22mm; 5 hips, 26 mm; 96 hips). Results. Two hips were revised, one was due to late infection, and the other due to breakage of the implant in trauma. Japanese orthopaedic association (JOA) score improved from 40 to 86 points. Postoperative complication was three periprosthetic fractures (one femoral shaft fracture and two greater trochanteric fractures) and femoral shaft fracture case was operated. Dislocation was not observed. Socket loosening (Hodgkinson, Type 3, 4) and stem loosening (Harris, definite and probable) were not observed radiographically. Cortical hypertrophy was observed in 7.9%. The survival rate of HS-3 tapered stem was 98% for revision due to any reason and 100% for revision due to aseptic loosening. Discussion. The long-term clinical results of K-MAX HS-3 tapered stem were excellent. The osteolysis at the stem tip was not observed in this type, which was observed in a few cases in previous type. From the X-ray finding, it was suggested that this taperd stem had more uniform stress distribution to the femoral bone than previous type. Moreover, the problems associated with titanium alloy usage were not observed. From the present investigation, good farther long-term results of the tapered titanium stem were expected


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 43 - 43
1 Mar 2017
Tamura J Asada Y Ota M Matsuda Y
Full Access

Introduction. We have compared the middle-term (average follow-up period; 10 years) clinical results of the K-MAX HS-3 tapered stem with those of the previous type having cylindrical tip. Materials and Methods. In K-MAX HS-3 THA (Kyocera Medical, Kyoto, Japan), cemented titanium alloy stem and all polyethylene cemented socket are used. This stem has the double tapered symmetrical stem design, allowing the rotational stability and uniform stress distribution (Type T) (Fig. 1). The features of this stem are; 1. Vanadium-free high-strength titanium alloy (Ti-15Mo-5Zr-3Al), 2. Double-tapered design, 3. Smooth surface (Ra 0.4μm), 4. Broad proximal profile, 5. Small collar. In contrast, previous type stem, which was made of the same smooth-surface titanium alloy, has the design with cylindrical stem tip, allowing the maximum filling of the femoral canal (Type C) (Fig. 2). Osteolysis at the distal end of the stem had been reported in a few cases in Type C, probably due to the local stress concentration. Therefore the tapered stem was designed, expecting better clinical results. All surgery was performed at Kitano Hospital between September 2003 and June 2006. 72 THA were performed (Type T; 52 hips, Type C; 20 hips). The average age of the patients at the operation was 61 and 69 years and the average follow-up period was 10.1 and 10.4 years for the Type T and C, respectively. The all-polyethylene socket was fixed by bone cement, and the femoral head material was alumina or CoCr (22 or 26 mm). Results. One hip was revised in Type C. Japanese orthopaedic association (JOA) score improved from 39/37 to 84/77 points (Type T/C). Postoperative complication was dislocation in one case (Type C; 1). Socket loosening was not observed radiographically. Stem loosening was observed in one hip in Type C, demonstrating osteolysis at the distal end of the stem. In this case, revision THA was performed 3.5 years postoperatively. Bone resorption was more frequently observed in Type C than Type T. Cortical hypertrophy was observed in 7.7% in Type T and 25% in Type C. Discussion. The middle-term clinical results of K-MAX HS-3 taper stem (Type T) was excellent. The osteolysis at the stem tip was not observed in this type, which was observed in a few cases in previous Type C. From the X-ray finding, it was suggested that Type T had more uniform stress distribution to the femoral bone than Type C. Moreover, the problems associated with titanium alloy usage were not observed. From the present investigation, good long-term results of the tapered titanium stem (Type T) was expected. For any figures or tables, please contact authors directly (see Info & Metrics tab above).


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 82 - 82
1 Feb 2020
Zobel S Huber G King M Pfeiffer D Morlock M
Full Access

Introduction. During revision surgery, the active electrode of an electrocautery device may get close to the implant, potentially provoking a flashover. Incidents have been reported, where in situ retained hip stems failed after isolated cup revision. Different sizes of discoloured areas, probably induced by electrocautery contact, were found at the starting point of the fracture. The effect of the flashover on the implant material is yet not fully understood. The aim of this study was to investigate the fatigue strength reduction of Ti-6Al-4V titanium alloy after electrocautery contact. Material and Methods. 16 titanium rods (Ti-6Al-4V, extra low interstitial elements, according to DIN 17851, ⊘ 5 mm, 120 mm length) were stress-relief annealed (normal atmosphere, holding temperature 622 °C, holding time 2 h) and cooled in air. An implant specific surface roughness was achieved by chemical and electrolytic polishing (Ra = 0.307, Rz = 1.910). Dry (n = 6) and wet (n = 6, 5 µl phosphate buffered saline) flashovers were applied with a hand-held electrode of a high-frequency generator (Aesculap AG, GN 640, monopolar cut mode, output power 300 W, modelled patient resistance 500 Ω). The size of the generated discoloured area on the rod's surface - representative for the heat affected zone (HAZ) - was determined using laser microscopy (VK-150x, Keyence, Japan). Rods without flashover (n = 4) served as control. The fatigue strength of the rods was determined under dynamic (10 Hz, load ratio R = 0.1), force-controlled four-point bending (FGB Steinbach GmbH, Germany) with swelling load (numerical bending stress 852 MPa with a bending moment of 17.8 Nm) until failure of the rods. The applied bending stress was estimated using a finite-element-model of a hip stem during stumbling. Metallurgical cuts were made to analyse the microstructure. Results. The control rods failed at the pushers of the setup (median: 94,550, range: 194,000 cycles). The rods with flashover failed directly at the HAZ significantly earlier than the control rods (p = 0.018). The analysis of the microstructure showed a transformation of the equiaxed α+β microstructure to a bimodal state. The size of the HAZs were equal for the dry (median: 1.51 mm. 2. , range: 5.68 mm. 2. ) and wet flashovers (median: 0.92 mm. 2. , range: 2.50 mm. 2. , p = 0.792). The cycles to failure were smaller for the dry flashover (median: 22,650 cycles, range: 5,700) than the wet flashover but not reaching statistical significance (median: 32,200, range: 57,900; p = 0.052). No correlation between the dimension of the HAZs and the cycles to failure was found (dry: r. 2. = 0.019, p = 0.8; wet: r. 2. = 0.015, p = 0.721). Discussion. Flashovers induced by an electrocautery device reduce the fatigue strength of Ti-6Al-4V. Since no correlation between the size of the HAZs and the cycles to failure was found, every contact between electrocautery devices and metal implants should be avoided. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 144 - 144
1 Sep 2012
Perez-Jorge C Perez-Tanoira R Arenas M Matykina E Conde A Gomez-Barrena E
Full Access

INTRODUCTION. Biomaterial-related infections are an important complication in orthopaedic surgery [1], and Staphylococcus sp. accounts for more than half of the prosthetic joint infection cases [2]. Adhesion of bacteria to biomaterial surfaces is a key step in pathogenesis of such infections [3]. Titanium alloys are widely used in orthopaedic implants because their biocompatibility [4]. Surface incorporation of ions with antimicrobial properties, like fluorine, is one strategy previously studied with good results [5]. MATERIAL AND METHODS. A 18mm diameter rod of Ti–6Al–4V alloy ELI grade according to the standard ASTMF136-02 supplied by SURGIVAL was cut into 2 mm thick disk specimens, ground through successive grades of SiC paper to 1200 grade, degreased with a conventional detergent and rinsed in tap water followed by deionised water. The specimens were then chemically polished (CP). The disks were anodized only on one side by using a two electrode cell in a suitable electrolyte. TiO. 2. barrier layers, without fluoride (BL), were produced by anodizing in 1 M H. 2. SO. 4. at 15 mA cm-2 to 90 V, reaching 200 nm of thickness. Fluoride barrier layers (FBL) were produced in an electrolyte containing 1 M NH. 4. H. 2. PO. 4. and 0.15 M NH. 4. F, at constant voltage controlled at 20 V for 120 min at 20°C; the thickness of the layer is 140 nm. Laboratory biofilm-forming strains of Staphylococcus aureus 15981 [6] and Staphylococcus epidermidis ATCC 35984 were used in adherence studies, which were performed using the protocol by Kinnari et al [7]. Photographs obtained were studied by ImageJ software. Statistical analysis was performed by EPI-INFO software. The experiments were performed in triplicates. RESULTS. Lower adherence was detected when compared FBL with unmodified controls (CP and BL). A statistical significant difference (p<0.01) was detected in the adhesion to modified material between both species, being the adherence of S. aureus lower than that of S. epidermidis (Figure 1). DISCUSSION & CONCLUSIONS. There is currently a discussion about the actual antibacterial properties of fluorine when incorporated in biomaterial surfaces. In this study we have demonstrated that both S. aureus and S. epidermidis strains showed a decrease of bacterial adhesion to modified surfaces with fluorine, a decrease that cannot be due to other surface modifications. Further studies, including adhesion studies with clinical strains [8], must be performed to confirm these results, which can lead to the development of new materials with a potential use in orthopaedic surgery


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 4 - 4
1 Jan 2016
MacDonald D Clarkin-Wright G Parvizi J Lee G Klein G Rimnac C Gilbert J Kurtz S
Full Access

Introduction. In THA, fretting corrosion at the head-stem taper junction has emerged as a clinical concern that may result in adverse local tissue reactions, even in patients with a metal-on-polyethylene bearing [1]. Taper junctions that employ a ceramic head have demonstrated reduced corrosion at the interface [2]. However, during revision surgery with a well-fixed stem, a titanium sleeve is used in conjunction with a ceramic head to ensure proper fit of the head onto the stem and better stress distribution. In vitro testing has suggested that corrosion is not a concern in sleeved ceramic heads [3]; however, little is known about the in vivo fretting corrosion of the sleeves. The purpose of this study was to investigate fretting corrosion in sleeved ceramic heads. Materials and Methods. Between 2001 and 2014, 35 sleeved ceramic heads were collected during revision surgery as part of a multi-center retrieval program. The sleeves were all fabricated from titanium alloy and manufactured by 4 companies (CeramTec (n=14), Smith & Nephew (Richards, n=11), Stryker (n=5), and Zimmer (n=5)). The femoral heads were made from 3 ceramics (Alumina (n=7), Zirconia (n=11), and Zirconia-toughened Alumina (n=17)). Sleeve dimensions (length and thickness) were measured using calibrated calipers. Fretting corrosion of the sleeves and available associated stems was scored using a 4-point, semi-quantitative scoring system [4], with 1 being little-to-no damage, and 4 corresponded to severe fretting corrosion. Five sleeves could not be extracted; thus the external surface was not scored. Results. Moderate-to-severe fretting corrosion scores (Score ≥ 2) were observed in 97% (34/35) of internal tapers (sleeve-femoral stem contact), 57% (17/30) of external tapers (sleeve-femoral head contact), and 65% (11/17) of the stems. The internal sleeve had higher fretting corrosion scores than the external taper (Mean Score Difference [MSD] = 1.1; p = 0.001) and stem (MSD = 0.7; p = 0.016). Fretting corrosion scores were correlated with implantation time at all surfaces (Rho ≥ 0.53; p ≤ 0.015). Scores were not correlated with sleeve dimensions (p > 0.05). Fretting corrosion scores of the external sleeve correlated directly with activity level (p = 0.005) and inversely with patient age (p = 0.03). Discussion. The retrieval data shows that fretting corrosion can occur in these components, particularly on the internal surface of the sleeve. The corrosion scores were similar to levels observed in prior studies of tapers in CoCr heads [2]. Implantation time was the main predictor of increased fretting corrosion. The impact of ceramic material and sleeve design currently remain unclear as the analyses were confounded with implantation time. Thus, more detailed and quantitative analyses are required to fully determine the factors that influence fretting corrosion of sleeved ceramic heads in THA


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 429 - 429
1 Dec 2013
Mitchell B
Full Access

Additive manufacturing (AM) techniques have gained attraction in orthopedic implant design with their ability to create unique shapes and structures. Depending on the application, there are different mechanical properties required. This study evaluated the mechanical properties of direct metal laser sintered (DMLS) Titanium alloy (Ti6Al4V) with and without hot isostatic pressure (HIP) treatment. Three dimensional computer modeling and the DMLS manufacturing assisted in building net or near-net samples for testing. The material testing consisted of uniaxial tension, Charpy impact, rotating beam fatigue (RBF), density, and hardness. Two sets of Ti6Al4V samples were created for testing using a DMLS process and stress relieved in a vacuum furnace prior to removal from the build platform. One set of samples were HIP treated. The two sets of samples were tested and the material properties of the non-HIP treated samples were compared to those with HIP treatment. Tension testing was conducted on fifteen (15) samples per treatment according to ASTM E8/E8M on as-built samples designed to a round specimen 3 per the standard. Fifteen (15) Charpy impact samples per treatment were built to near-net shapes. A low stress grind was performed on all surfaces and a notch was placed in the sample to comply with ASTM E23 and testing was performed in accordance with the standard. Fifteen (15) samples were built per treatment and machined for RBF per ISO 1143. RBF was performed on all samples at a frequency of 100 Hz with run out conditions of 10M cycles or failure. Density and hardness was measured on three (3) samples from each set using Archimedes' Principle and Rockwell hardness techniques respectively. The average (standard deviation) tensile strengths between the two groups were statistically different (p < 0.05). The non-HIP treated samples had an average ultimate strength of 956(10) MPa, yield strength of 896(13) MPa, and modulus of 118(2) GPa (Table 1). The HIP treated samples had an average ultimate strength of 909(4) MPa, yield strength of 832(9) MPa, and modulus of 112(3) GPa (Table 1). There was also statistical differences in the impact strength with the HIP treatment samples having a higher required force of 23.4(1.6) J compared to the non-HIP treated group of 19.8(1.8) J (Table 1). The fatigue strength of the samples HIP treated compared to the non-HIP treated group was 650 MPa and 396 MPa respectively (Table 1). This study shows that the HIP treatment of DMLS Ti6Al4V diminishes some mechanical strengths while greatly improving the fatigue life of the material. As we continue to evaluate these “new” materials for orthopedic devices, these mechanical and physical properties will help us understand the capabilities of this process and material


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 109 - 109
1 May 2016
Papapietro N Di Martino A Marinozzi A Cancilleri F Denaro V
Full Access

Introduction. One of the major concerns in the use of modular hip prostheses is the structural failure of one or more components of the prosthesis, with total mechanical failure. In literature there are sporadic cases of breakage of the prosthetic neck in patients with high functional demand. Material and methods. In our experience we have implanted a model of modular hip prostheses in 38 patients. In 17 of these patients it was implanted a neck made of titanium alloy, while in the remaining 21 was utilised a CrCo neck. In 5 of the 17 patients with neck titanium it has been a spontaneous rupture of the neck in a time ranging from 12 to 24 months. We proceeded with the replacement of the titanium neck with similar neck made of CrCo alloy in 5 patients with failed neck, and we decided to call the remaining 12 patients with titanium neck, offering them the revision of the system. All patients accepted the revision surgery. Results. implants explanted and subject to breakage of the neck were analyzed by an independent metallurgical laboratory which has produced a document in which were found the main signs of wear which led to the deterioration of metal involved. Conclusions. The rupture of the modular hip prostheses is a possible event both because of the high functional loads required by some types of patients, that of the building materials and design of the modular components. In this study we are presented the results of the laboratory tests that allow us to pose hypotheses about the causes and failure mechanisms of modular hip prostheses


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 393 - 393
1 Dec 2013
Murphy S Le D
Full Access

Introduction

Adverse Local Tissue Reactions (ALTR) have been reported in association with both wear and corrosion. Tissue reactions have been reported in association with corrosion at CoCr head-CoCr neck, CoCr head-TiAl6V4 neck, and CoCr modular neck on beta-titanium (TMZF) stem junctions. The current abstract reports on 3 cases of ALTR in association with CoCr modular necks on convention titanium (TiAl6V4) stem junctions.

Case 1. A 67 year old male (87 kg, 1.73 m, BMI 29.1) presented with new onset hip irritation 11 months after surgery. Radiographs show no abnormalities. Further investigation revealed the following: ESR = 95, CRP = 5, Cr level = 1.0, Co level = 4.1, leukocyte transformation testing = highly reactive to nickel. Hip aspiration was culture negative with 11,250 wbc. Metal artifact reduction MR showed cystic local reaction in the region of the greater trochanter.

Case 2. A 52 year old male (89 kg, 1.83 m, BMI 26.5) presented with new onset hip irritation 30 months after surgery. Radiographs show no abnormalities. Further investigation revealed the following: ESR = 7, CRP = 5.4, Cr level = 2.1, Co level = 4.8, leukocyte transformation testing = reactive to nickel. Hip aspiration was culture negative with 3995 wbc. Metal artifact reduction MR showed cystic local reaction in the region of the iliopsoas.

Case 3. A 52 year old male (104 kg, 1.85 m, BMI 30.1) presented with new onset hip irritation 26 months after surgery. Radiographs show no abnormalities. Further investigation revealed the following: ESR = 33, CRP = 34.9, Cr level = 1.0, Co level = 3.7, leukocyte transformation testing = no reactivity to any of the biomaterials. Hip aspiration was culture negative with 3,780 wbc. Metal artifact reduction MR showed cystic local reaction in the region of the iliopsoas.

Discussion

All three of these patients are scheduled for revision surgery. All three had ceramic-ceramic bearings. We have experience with 1029 ceramic-ceramic THA with fixed neck conventional titanium and modular titanium neck implants with minimum 2 yr f/u and have never diagnosed an adverse reaction in any of these patients. It is possible that corrosion at the CoCr neck on TiAl6V4 stem junction is the root cause of these reactions. Although the incidence of diagnosed reactions is roughly 1%, it appears that the use of CoCr at any junction under significant mechanical stress can result in adverse local tissue reaction and therefore should either be avoided or used with great caution and compelling indications.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 137 - 137
1 Apr 2019
Oladokun A Vangolu Y Aslam Z Harrington J Brown A Hall R Neville A Bryant M
Full Access

Introduction. Titanium and its alloys are attractive biomaterials attributable to their desirable corrosion, mechanical, biocompatibility and osseointegration properties. In particular, β – titanium alloys like the TMZF possess other advantages such as its lower modulus compared to Ti6Al4V alloy. This reduces stress shielding effect in Total Hip Arthroplasty (THA) and the replacement of V in the Ti6Al4V alloy, eliminates in-vivo V-induced toxicity. Unfortunately, implants made of TMZF were later recalled by the FDA due to higher than acceptable revision rates. The purpose of this study was to compare the fretting corrosion characteristics of Ti6Al4V and TMZF titanium alloys. It is hoped the findings will inform better design of β – titanium alloys for future applications in THA. Method. A ball-on-flat configuration was utilised in this study to achieve a Hertzian point contact for CoCrMo – Ti6Al4V and CoCrMo – TMZF material combinations. These were assessed at a fretting displacement of ±50 µm at an initial contact pressure of 1 GPa. Each fretting test lasted 6000 cycles at a frequency of 1 Hz. A two-electrode cell set-up was used to monitor in-situ open circuit potential (OCP). The simulated physiological solution consisted of Foetal Bovine Serum (FBS) diluted to 25% with Phosphate Buffered Saline (PBS) and 0.03% Sodium Azide (SA) balance. The temperature was kept at ∼37°C. Corrosion products on the worn surfaces and subsurface transformations in both alloys were characterised using the Scanning and Transmission Electron Microscopy (SEM/TEM) to obtain high resolution micrographs. The samples were prepared using a FIB-SEM. Bright-field, dark-field and selected area electron diffraction (SAED) patterns were all captured using a scanning TEM (STEM) and Energy Dispersed X-Ray spectroscopy (EDX) mapping was carried out. Results. The results showed that fretting regime transition from partial-slip to gross slip was delayed a few hundred cycles for TMZF relative to the Ti6Al4V (Figure 1). This indicates that the lower modulus of TMZF influences the degree of elastic deformation accommodated prior to the initiation of plastic shear at the fretting interface. The OCP directly corresponded to the fretting regime for both material combinations (Figure 2). Surface and subsurface characterisation of both alloys show differences in the structure of their mechanically mixed corrosion products and metallurgical transformations. Interestingly, an amorphous Co-rich layer was seen across the TMZF surface (Figure 3) whereas, pitting corrosion products from the CoCrMo alloy was seen on the Ti6Al4V alloy. Conclusion. In summary, the difference in the fretting behavior of Ti6Al4V and TMZF directly corresponds to the combined differences in their elastic modulus and surface chemistry. This corresponds to the differences observed in their electrochemical behavior. However, the main differences observed were the properties of their corrosion products and subsurface metallurgical transformations. These observed characteristic differences are to be considered in further examination of the cause of higher failure rates in TMZF alloys. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_5 | Pages 1 - 1
1 Apr 2018
Jo S Lee S Lim W Kim D Lee J
Full Access

Introduction. Cobalt chrome femoral head has been used widely in total hip arthroplasty and has shown favorable outcome. However, there is still of concern of potential metal toxicity from the wear debris. In the other hand, titanium is well known for its biocompatibility but it is not used in bearing surface of arthroplasty due to its brittleness. Recently, coating of the prosthesis using plasma electrolytic oxidation (PEO) has shown favorable surface protection. Thus, in this study, we tried to find out whether the PEO coating on the titanium surface would provide surface protection. Materials and methods. Five Titanium alloy (Ti-6Al-4V) ball mimicking femoral head was manufactured and was coted using plasma electrolytic oxidation. Wear rate was tested using validated wear tester with 10N compression force at 80rpm. The amount of wear was detected by measuring change of weight after wear test was completed. This was compared with femoral head manufactured with titanium alloy without PEO coating. Toxicity of the debris was also tested using MTT assay with human osteoblast cell line. Results. Compare to the base titanium metal, PEO coated metal head has shown to provide surface protection. The wear rate has significantly decreased with PEO coating (median value : 0.00015g/mm. 2. vs 0.00006 g/mm. 2. ). MTT assay revealed no cytotoxicity with the amount of debris generated from the wear test. Conclusion. The result of the current study indicate that the PEO coating on the titanium femoral head can significantly decrease the wear rate and is non cytotoxic. This indicates that the femoral head manufactured with titanium alloy and PEO coating maybe a potential alternative to be used in total hip arthroplasty


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 242 - 242
1 Mar 2013
Lim YW Kim Y Kwon SY Chang YJ Kim KS
Full Access

Introduction. Inspired by mussel-adhesion phenomena in nature can integrate inorganic hydroxyapatite crystals within versatile materials. This is a simple, aqueous, two-step functionalization approach, called polydopamine-assisted hydroxyapatite formation (pHAF), that consists of i) the chemical activation of material surfaces via polydopamine coating and ii) the growth of hydroxyapatite in a simulated body fluid (SBF). We presumed polydopamine coating on the surface of titanium alloy would improve the ability of cementless stems to osseointegrate. We therefore compared the in vitro ability of cells to adhere to polydopamine coated Ti alloy and machined Ti alloy. Method. We performed energy-dispersive x-ray spectroscopy and scanned electron microscopy investigations to assess the structure and morphology of the surfaces. Biologic and morphologic responses to osteoblast cell lines (MC3T-E1) were then examined by measuring cell proliferation, cell differentiation (alkaline phosphatase activity), and avb3 integrin. Results. Cell proliferation, alkaline phosphatase activity, migration, and adhesion were not increased in the polydopamine coated Ti alloy compared to other group. And, the polydopamine coated Ti alloy shows better apatite forming ability than the untreated one, as evidenced by apatite formation after SBF immersion for 14 days. Discussion. The surface modifications we used (polydopamine coating) enhanced the apatite formation, but did not change the biocompatibility (proliferation and migration of osteoblastlike cells) of Ti alloy


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 99 - 99
1 May 2016
Oladokun A Pettersson M Bryant M Hall R Neville A
Full Access

Introduction. Cobalt-Chromium-Molybdenum (CoCr) and Titanium-Aluminium-Vanadium (Ti) alloys are the most commonly used alloys used for Total Hip Replacement due to their excellent biocompatibility and mechanical properties. However, both are susceptible to fretting corrosion In-vivo. The objective of this study was to understand the damage mechanism of both combinations through a sub-surface damage assessment of the alloys at various fretting amplitudes using the Transmission Electron Microscopy (TEM – CM200 FEGTEM). The TEM was used to attain a cross sectional view of the alloys in orderto see the effect of high shear stress on the grain structure. Methods. The two combinations were fretted at a maximum contact pressure of 1 GPa in a Ball – on – Plate configuration for displacement amplitudes of 10μm, 25μm, 50μm and 150μm. The contact was lubricated with 25% v/v Foetal Bovine Serum (FBS), diluted with Phosphate Buffered Saline (PBS). The material loss through wear and corrosion from the fretting contact were quantified using the Visual Scanning Interferometry (VSI). The TEM samples were obtained using the Focused Ion Beam (FIB – FEA Nova 200 Nanolab). Samples were obtained from regions of high stress (shaded in red) [Fig. 1] for both CoCr and Ti flat of the CoCr–CoCr and CoCr–Ti couples respectively. Result. Total volume loss result vs. Dissipated Energy was plotted from displacement amplitudes of 10μm, 25μm and 50μm for both couples consecutively [Fig. 2]. The TEM images [Fig. 3] of CoCr alloy (denoted as CC) reveal a progressive damage to the topmost surface of the alloy and loss of nano-crystalline layer. Evidence of severe grain damage from the topmost surface can also be seen at 50μm. On the other hand, the Ti alloy (denoted as CT) at [Fig. 3 (CT–25μm)] reveal some recrystallization at the topmost surface and a progressive recrystallization of the bulk alloy was observed at 150μm. Damage to the surface was also visible at this displacement amplitude which initiated a crack as circled in red in the image [Fig. 3 (CT–150μm)]. Discussion. Fouvry et al1 discussed the effect of the interfacial shear work done (dissipated energy) on a fretted material; this energy is mainly expended on material structure transformation (as observed in Ti alloy) and/or wear generation (as observed in CoCr alloy) [Fig. 2]. This intermediate damage mechanism helps to identify that CoCr–CoCr follows a wear dominated mechanism while CoCr–Ti preferably exhibits fatigue behaviour until large displacement amplitudes are applied leading to accelerated wear of the top surface [Fig. 3 (CT–150μm)]. The recrystallization was observed over 2μm below the surface at displacement amplitude of 150μm. Consequentially, this could modify the metallurgy of the Ti alloy and may contribute to the clinically observed phenomena whereby, the softer Ti wears the harder CoCr component2. Conclusion. TEM micrographs reveal large granular damage on the CoCr alloy and deep bulk recrystallization of the Ti alloy as a result of interfacial shear stress. This suggests that the Ti alloy may experience a change in its mechanical behaviour. On the other hand, it is identified that a CoCr–CoCr couple experiences a wear dominated mechanism


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 25 - 25
1 Apr 2019
Cazzola M Ferraris S Stella B Orlygsson G Ng CH Cempura G Scolaro C Prenesti E Yamaguchi S Pezzotti G Cochis A Rimondini L Spriano S
Full Access

In order to improve fast osseointegration, to modulate inflammatory response and to avoid biofilm formation, several attempts of surface modifications of titanium alloy in term of surface topography and chemistry have been performed over years, but this is still an open issue. In our research work, a patented chemical treatment was developed and tailored to improve fast osseointegration and to allow further surface functionalization in order to get a multifunctional surface. After the chemical treatment, Ti6Al4V shows a micro and nano-textured surface oxide layer with high density of hydroxyls groups, as summarized Figure 1: it is able to induce apatite precipitation (during soaking in Simulated Body Fluid), high wettability by blood, specific protein adsorption, positive osteoblast response and surface mechanical resistance to implantation friction. Hydroxyl groups exposed by the treated surface also allow binding natural biomolecules such as polyphenols, which can further improve the rate and quality of osseointegration by adding anti-inflammatory, antibacterial and antitumoral effects suitable for implants in critical situations. Polyphenols have the further added value of being a low cost and eco-sustainable product, extractable from byproducts of wine and food industry. On the chemically treated and functionalized samples, the surface characterization was performed using Folin&Ciocalteu test, fluorescence microscopy and XPS analysis in order to check the presence and activity of the grafted biomolecules (polyphenols from red grape pomace and green tea leaves). Cell tests were performed with Kusa A-1 cells highlighting the ability of polyphenols to improve osteoblasts differentiation and deposition of mineralized extracellular matrix. Surface functionalization can also be performed with chitin derived biomolecules to reduce inflammation. With the purpose of obtaining the antibacterial effect, during the chemical treatment a silver precursor can also be added to obtain in situ reduced silver nanoparticles embedded in the nano-structured oxide layer. The samples containing nanoparticles on the surface were characterized by means of TEM and FESEM observation highlighting the presence of well distributed and small-sized nanoparticles on the surface and through the thickness of the oxide layer. A long-lasting release in water was observed up to 14 days and antibacterial tests on Staphylococcus aureus showed the ability of the surface to reduce bacteria viability avoiding biofilm formation. The results showed that the patented chemical treatment can improve the response of osteoblasts to titanium alloy implants, but is also a promising way to obtain multifunctional surfaces with antibacterial, antioxidant, anti-inflammatory and antitumoral properties that can be the future of orthopedic implants


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 135 - 135
1 May 2016
Lapaj L Mroz A Markuszewski J Kruczynski J Wendland J
Full Access

Introduction. Backside wear of polyethylene (PE) inlays in fixed-bearing total knee replacement (TKR) generates high number of wear debris, but is poorly studied in modern plants with improved locking mechanisms. Aim of study. Retrieval analysis of PE inlays from contemporary fixed bearing TKRs - to evaluate the relationship between backside wear and liner locking mechanism and material type and roughness of the tibial tray. Methods. MATERIAL. We included five types of implants, revised after min. 12 months (14–71): three models with a peripheral locking rim and two models with a dove-tail locking mechanism. Altogether this study included 15 inlays were removed from TKRs with CoCr alloy tray with a roughened surface and a peripheral locking lip liner (Stryker Triathlon, Ra 5,61 µm), 9 from CoCr trays with peripheral locking lip and untreated surface (Aesculap Search, Ra 0,81 µm), 13 from Ti alloy trays with peripheral locking lip and untreated surface (DePuy PFC Sigma 0,61 µm), 11 from Ti alloy trays with untreated surface and dovetail locking mechanism (Zimmer NexGen, 0,34 µm), and 9 from iplants with a Ti alloy tibial tray with mirror polished surface and dovetail locking mechanism (Smitn&Nephew Genesis II, 0,11 µm). METHODS. Wear of bearing surface and back side of retrieved inlays was examined in 10 sectors under a light microscope. Seven modes of wear were analysed and quantified according to the Hood scale: surface deformation, pitting, embedded third bodies, pitting, scratching, burnishing (polishing), abrasion and delamination. Damage of inlays caused by backside wear was also evaluated using scanning electron microscopy (SEM). Roughness of tibial tray was evaluated using a contact profilometer. Results. We found no differences between wear scores on the articulating surface in all group, they did not correlate with backside wear scores in all groups as well. Compared to all other groups, backside wear scores were significantly higher in implants with untreated Ti alloy tibial tray (P<0,001 Wilcoxon test). Lowest wear rates were found in implants from both Ti and CoCr alloys and peripheral locking rim. Interestingly there was no difference between wear of implants with polished and untreated surface (Fig. 1). SEM analysis demonstrated different wear modes in implants with dovetail mechanism and peripheral rim. The first group demonstrated signs of gross rotational instability, with severe abrasion with an arch-shaped pattern and delaminated PE (Fig 2). In one design we observed severe extrusion of PE into screw holes of the tibial tray. Inlays from trays with peripheral rim presented two types of wear: flattening of machining marks or protrusion of the material caused by the rough surface (Fig 3). Conclusions. This study demonstrates that backside wear is still a problem in modern TKR. Our findings suggest that it is predominantly affected by type of locking mechanism (with peripheral rim performing better), to a lesser extent by surface roughness of the tibial component, while material type does not seem to play an important role. This study was funded by a grant from the National Science Centre nr 2012/05/D/NZ5/01840. To view tables/figures, please contact authors directly


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 76 - 76
1 May 2016
Nakao M Uchida K Sakai R Naruse K Takaso M Mabuchi K
Full Access

Background. Residual stress remains in bone tissues after press-fit-fixation of a joint prosthesis, recently employed for joint arthroplasty. The response of bone tissues to the residual stress is, however, unknown because it is not physiological. This unnatural stimulus may have adverse effects on bone tissues, including causing thigh pain or bone resorption. In the present study, we designed an experimental method to apply a stationary load from inside an animal femur using a loop spring of titanium alloy with super elasticity. The femoral response was assessed based on the migration of the wire into bone twelve weeks after implantation. As the results, wire migration was noted in 10 of 11 cases. Methods. We developed a method using a loop spring made of super elastic titanium alloy, which can maintain sufficient stress in a rat femur for a prolonged period. This titanium alloy, which contains 43.94% titanium and 56.06% nickel, was supplied as a wire (WDL1, Actment Co., Ltd., Kasukabe, Japan). In the present study, an experimental method was designed to apply a stationary load from inside a rat femur by inserting a loop spring made of super elastic wire. Results. Ten weeks after implantation, migration of the spring wire into the cortical bone was noted in 10 of 11 cases. To assess spring migration in cortical bone, we measured the distance from the endocortical surface to the tip of springs on micro-CT images. The line of the endocortical surface was extrapolated from the adjacent to the wire contact area. The estimated load was distributed from 1.19 to 3.28 N. The migration depth on anterior and posterior sides was not associated with the estimated load. Discussion. In the present study, we developed a method of generating a stationary stress field in a rat femur using a loop spring made from Ni-Ti alloy with super ermore, implantation of the pin was presumed to be able to interpretation by the thelasticity. The load that originated from elastic deformation was large enough to apply mechanical stimulation to bone tissue. The estimated load was distributed from 1.19 to 3.28 N. Migration of the implanted loop spring in the femur was observed in ten of the eleven. The migration depth apparently did not increase with the increase in estimated load. Therefore, regardless of the load, that is stationary load is applied has been suggested to be involved in the migration of the pin. Furtheory of bone remodeling. n a state where the pin load of is applied, the bone implanted pin to which was destroyed by osteoclasts, was presumed to occur is bone formation in the implanted state. Interpretation. The present findings suggest that an excessive stationary load at the implant surface induces endosteal bone resorption together with the migration or protrusion of a prosthesis


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_3 | Pages 117 - 117
1 Jan 2016
Park Y Moon Y Lim S Kim D Jang S
Full Access

Introduction. Micro-arc oxidation (MAO) is an electrochemical method used to treat metal surfaces. It provides nanoporous pits, and thick oxide layers, and incorporates calcium and phosphorus into the coating layer of titanium alloy. This modification on the surface of titanium alloy by MAO coating would improve the ability of cementless stems to osseointegrate. In spite of these structural and chemical advantages, clinical study of total hip arthroplasty (THA) using MAO coated stem has not yet been reported. In this study, we evaluated the clinical and radiographic results associated with cementless grit-blasted tapered-wedge stems that were identical in geometry but differed with regard to surface treatment with or without MAO coating. Materials & Methods. We performed a retrospective review of 141 THAs using MAO coated stem for a minimum of 5 years and compared them to 219 THAs using the same geometry stem without MAO coating. A cementless Bencox femoral component (Corentec, Seoul, Korea) was used in all hips. It is made of titanium alloy with a straight, double wedged, tapered stem with a rectangular cross-section. Surface treatment was performed using grit blasting with a roughness of 5.5ųm that was available either with or without MAO coating. Clinical and radiographic evaluations were performed preoperatively and at 6 weeks, 3 months, 6 months, and 1 year postoperatively, and then annually. Results. In the MAO coating group, the mean Harris hip score improved from 43.7 points preoperatively to 93.9 points postoperatively. The mean WOMAC score and UCLA activity score at the final follow-up was 17 points (range, 6–34 points) and 6.9 points (range, 5–10 points), respectively. Thigh pain at final follow-up was reported in 2 hips (1.4%), but neither of these hips showed signs of implant loosening or limited daily activities. Complications included one squeaking, one iliopsoas tendonitis, and one deep vein thrombosis. Postoperative Harris hip scores, WOMAC scores, UCLA activity scores, and complication rates did not differ between the two groups. In both groups, no femoral or acetabular component showed radiographic evidence of mechanical loosening, and no components had been revised at the final follow-up. Conclusions. Primary THA using a cementless grit-blasted tapered-wedge stem with MAO coating showed encouraging medium-term outcomes. Further prospective controlled study is required to investigate the long-term implant survival, possible complications, and cost-effectiveness of utilizing MAO coating in THA


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_8 | Pages 4 - 4
1 May 2014
Blaha J
Full Access

Having previously been a proponent of the advantages of the modular neck in total hip arthroplasty, I now take the opposite argument because of corrosion that happens with all taper junctions. The advantage of the modular neck is the “uncoupling” of femoral stem position from the final position of the femoral head. Surgical priorities frequently compete, whether positioning the stem for the best press-fit (for cementless fixation) or the best cement mantle (for cemented fixation), and positioning of the stem for preventing dislocation and improving function. My personal use of the modular neck spanned approximately 4 years from 2003–2008 and encompassed a total of 390 primary and revision cases. Excellent functional results were obtained, but some problems occurred that were associated with the modular neck and with large diameter head metal-metal articulations. The modular neck was designed and studied at the Rizzoli Institute in Bologna, Italy with the conclusion that the strength of construct (titanium alloy neck in the titanium alloy stem) was sufficient and the potential for fretting at the modular junction was small and acceptable. Pre-market testing of the device met and exceeded all FDA suggested benchmarks. The first modular neck fracture in my personal series occurred more than 3 years after implantation, in a large man with a long, varus modular neck. Within a year another fracture of a long, varus modular neck occurred in a heavy man. I now know of 6 modular neck fractures among the 390 cases. We have found evidence of corrosion, some very severe, in modular necks that we have revised (both fractured and intact modular necks). This corrosion is caused by Mechanically Assisted Crevice Corrosion associated with fretting at the modular junction which leads to removal of the titanium oxide “passivation” layer that generally forms on a titanium implant. This exposes more of the substrate metal to oxidation and can create pits that, in the notch-sensitive titanium alloy, can lead to the initiation of fracture. The hydrogen that is created from the corrosion reaction and diffused into the metal can cause “embrittlement” which predisposes it to fracture. We also have seen “hydrogen pneumarthrosis” associated with corrosion of the titanium modular neck in which the corrosion concentrated the hydrogen gas in the femoral stem below the modular neck and suddenly was released into the joint with significant pain. The hydrogen gas is irritating to the joint capsule and the patient presents with intense pain and gas in the joint, a clinical picture that can be confused with infection in the joint with a gas-forming organism. We now know that the condition is self-limiting, but suggests that revision of the modular neck construct would be a reasonable course of action. Recently cobalt chromium modular necks have replaced those made of titanium alloy. Since cobalt-chromium is harder and stiffer, the milieu of the taper junction will be different than that of the titanium-titanium junction, and it has been suggested that this will allow safe and long-term use of the modular neck. The first titanium alloy necks were introduced in the early 1990s and it took until the mid-2000s to recognise problems. Last year the Stryker modular neck used with the Rejuvenate stem was recalled because of significant reaction associated with corrosion at the neck-to-stem junction. Corrosion is inevitable at modular junctions exposed to cyclic loading, especially in the milieu of body fluids. We now know that ALTR occurs in response to taper junction corrosion as well, and the more modular junctions there are in a total hip construct, the more debris and potential reaction likely. Fixed neck stems provide satisfactory long-term fixation and function for patients, so despite a functional advantage to the modular neck, it is “a bridge too far”


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_6 | Pages 106 - 106
1 Jul 2020
Dion C Lanting B Howard J Teeter M Willing R
Full Access

During revision total knee arthroplasty (rTKA), proximal tibial bone loss is frequently encountered and can result in a less-stable bone-implant fixation. A 3D printed titanium alloy (Ti6Al4V) revision augment that conforms to the irregular shape of the proximal tibia was recently developed. The purpose of this study was to evaluate the fixation stability of rTKA with this augment in comparison to conventional cemented rTKA. Eleven pairs of thawed fresh-frozen cadaveric tibias (22 tibias) were potted in custom fixtures. Primary total knee arthroplasty (pTKA) surgery was performed on all tibias. Fixation stability testing was conducted using a three-stage eccentric loading protocol. Static eccentric (70% medial/ 30% lateral) loading of 2100 N was applied to the implants before and after subjecting them to 5×103 loading cycles of 700 N at 2 Hz using a joint motion simulator. Bone-implant micromotion was measured using a high-resolution optical system. The pTKA were removed. The proximal tibial bone defect was measured. One tibia from each pair was randomly allocated to the experimental group, and rTKA was performed with a titanium augment printed using selective laser melting. The contralateral side was assigned to the control group (revision with fully cemented stems). The three-stage eccentric loading protocol was used to test the revision TKAs. Independent t-tests were used to compare the micromotion between the two groups. After revision TKA, the mean micromotion was 23.1μm ± 26.2μm in the control group and 12.9μm ± 22.2μm in the experimental group. There was significantly less micromotion in the experimental group (p= 0.04). Prior to revision surgery, the control and experimental group had no significant difference in primary TKA micromotion (p= 0.19) and tibial bone loss (p= 0.37). This study suggests that early fixation stability of revision TKA with the novel 3D printed titanium augment is significantly better then the conventional fully cemented rTKA. The early press-fit fixation of the augment is likely sufficient for promoting bony ingrowth of the augment in vivo. Further studies are needed to investigate the long-term in-vivo fixation of the novel 3D printed augment


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_7 | Pages 11 - 11
1 Apr 2017
Murphy S
Full Access

Distal neck modularity places a modular connection at a mechanically critical location, which is also the location that confers perhaps the greatest clinical utility. The benefits of increased clinical options at that location must be weighed against the potential risks of adding an additional junction to the construct. Those risks include prosthetic neck fracture, taper corrosion, metal hypersensitivity, and adverse local tissue reaction. Further, in-vitro testing of ultimate or fatigue strength of femoral component designs has repeatedly failed to predict behavior in-vivo, raising questions about the utility of in-vitro testing that does not incorporate the effect of mechanically assisted crevice corrosion into the test design. The material properties of Ti alloy and CoCr alloy place limits on design considerations in the proximal femur. The smaller taper junctions that are necessary for primary reconstruction are particularly vulnerable to failure whereas larger taper junctions commonly used in revision modular femoral component designs have greater opportunity for success. Modular junctions of CoCr alloy on conventional Ti alloy have been shown to have a greater incidence of clinically significant mechanically assisted crevice corrosion and adverse reaction. Designs that have proven clinical strength and utility universally have larger, more robust junctions, that extend into the metaphysis of the femur. While these designs are primarily designed for revision total hip replacement (THR), they are occasionally indicated for primary THR. Overall, however, while design options at the neck-stem junction have unmatched clinical utility, no design that does not extend into the metaphysis has proven to be universally reliable. While routine use of modular neck components for primary THR does not appear to be clinically indicated based on current evidence, modular designs with proven successful proximal junctions appear to be indicated for revision THR and rare primary THR with extreme version or other anatomical circumstances