Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated We found that long-term alumina wear in association with a loose
acetabular component could be categorised into three groups. Of
20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight
‘severe wear’, which was characterised by a change in the physical
shape of the bearing and a loss of volume. This suggests that the
wear in alumina–alumina bearings in association with a loose acetabular
component may be variable in pattern, and may explain, in part,
why the wear of a ceramic head
Polyethylene wear debris in TKA arises from several sources, including the tibiofemoral articulation and the interface between the backside surface of the tibial insert and the metal tibial tray. In this study we identify a new source of abrasive damage to the polyethylene bearing surface: impingement of resected bony surfaces, osteophytes and overhanging acrylic cement on the tibial bearing surface during joint motion. One hundred forty-eight tibial components of 24 different designs in a retrieval collection were examined. A digital image of the articular surface of each insert was recorded. The presence, location and projected area of abrasive wear to the non-articulating edges of the insert were assessed using image analysis software. Significant abrasive wear was observed in 24% of the retrievals with cemented femoral components and 9% from non-cemented components. Of the retrievals exhibiting this abrasive wear mode, 46% experienced multiple site damage. The average damage area for each individual abrasive scar was 78±11mm^2. Within the group of worn inserts, the abrasive scars were seen with a frequency of 69% on the extreme medial edge, 19% on the extreme lateral edge, 38% on the posterior-medial edge and 23% on the posterior-lateral edge. In posterior stabilized components with an open femoral box design, scarring of the superior surface of the tibial post was also observed. This proposed mode of damage was confirmed with several retrieved femoral components containing either fixed cement protruding from the posterior condyles, from the medial and lateral edges or osteophytes embedded in the posterior capsule. The corresponding inserts exhibited significant abrasive scarring at those locations. We have observed a previously unrecognized source of polyethylene damage resulting in gouging, abrasion and severe localized damage in cemented and uncemented total knee replacement. Clearly, acrylic cement, in bulk or particulate form, often contributes to severe damage of the tibial surface and improvements to instruments and techniques for cementing are needed to prevent this wear mechanism.
Artificial hip joints have been in use for a number of years; various combinations of metals and polymers have been tested both in vitro and in vivo. Modern ceramics have found application as bearings in hip replacement due to the enhanced wear and friction that they offer. It has been hypothesised that during the swing phase of gait it is possible for the Femoral head and the Acetabular cup to dislocate, before relocating during heel contact. Severe loading such as this could cause greater levels of wear to occur in artificial hip joints. This study provides comparative analysis between ceramic-on-ceramic hip joint pairings under both severe and standard loading profiles. Five zirconia-toughened alumina (ZTA) 28mm diameter bearing pairs were tested on a ProSim Hip Simulator for 5.3 million cycles (MC), two under severe loading and three under standard loading conditions. Additionally a Loaded Soak Control, Soak Control and Environmental Control were used. Wear was recorded every 0.5 MC by gravimetric measurement. Surface microscopy images from a Zygo New View 100 and an Atomic Force Microscope (AFM) were taken before testing and then at, 0.5 MC. 2.5 MC, and 5.3 MC. The standard loading profile followed . ISO14242. -1 standard with 2650±50N maximum force, ±10° internal-external rotation and −15–30° flexion-extension. To simulate aggressive wear condition, microseparation inferiorly and micro-lateralisation laterally were applied during the swing phase. Dual acting cylinders were used to apply a constant force of 350±50N in opposition to the standard loading profile to enable separation between the Femoral Head and the Acetabular Cup during the swing phase. This microseparation was measured by means of a Linear Variable Differential Transformer (LVDT) and the setting gave a reading of 1.2mm ± 0.1mm at the start of each 0.5 million run. The value for microlateralisation was 0.9mm whilst the inferior separation was 1.2mm. Wear rates for the ceramic cups under
INTRODUCTION. Implant wear testing is traditionally undertaken using standardized inputs set out by ISO or ASTM. These inputs are based on a single individual performing a single activity with a specific implant. Standardization helps ensure that implants are tested to a known set of parameters from which comparisons may be drawn but it has limitations as patients perform varied activities, with different implant sizes and designs that produce different kinematics/kinetics. In this study, wear performance has been evaluated using gait implant specific loading/kinematics and comparing to a combination deep knee bend (DKB), step down (SD) and gait implant specific loading on cruciate retaining (CR) rotating platform (RP) total knee replacements (TKR). This combination activity profile better replicates patient activities of daily living (ADL). METHODS. Two sets of three ATTUNE. ®. size 5 right leg CR RP TKRs (DePuy Synthes, Warsaw, IN) were used in a study to evaluate ADL implant wear. Implant specific loading profiles were produced via a validated finite element lower limb model [1] that uses activity data such as gait (K1L_110108_1_86p), SD (K1L_240309_2_144p), and DKB (K9P_2239_0_9_I1) from the Orthoload database [2] to produce external boundary conditions. Each set of components were tested using a VIVO joint simulator (AMTI, Watertown, MA, Figure 1) for a total of 4.5 million cycles (Mcyc). All cycles were conducted at 0.8Hz in force-control with flexion driven in displacement control. Bovine calf serum lubricant was prepared to a total protein concentration of 18g/L and maintained at 37°±2°C. Wear of the tibial inserts was quantified via gravimetric methods per ISO14243–2:2009(E). Polyethylene tibial insert weights were taken prior to testing and every 0.5Mcyc there after which corresponded to serum exchange intervals. The multi-activity test intervals were split into10 loops of 1,250 DKB, 3,000 SD, and 45,750 gait cycles in series. Based on activity data presented by Wimmer et al. the number of cycles per activity and activities used is sufficient for a person that is considered active [3]. A loaded soak control was used to compensate for fluid absorption in wear rate calculations. Wear rates were calculated using linear regression. RESULTS SECTION. The wear rate for the gait-only activity test was calculated to be 0.20±0.04mg/Mcyc conversely the wear rate for the multi-activity test was 2.65±0.67mg/Mcyc (Figure 2). Wear scars can be found in Figure 3. Using a two-sided t-test of unequal variance, it was found that there was a significant difference between the two wear rates (p=0.004). DISCUSSION. Adding activities to the wear simulation test significantly increased the average wear rate of the test samples, confirming that changes in cross shear from different activities will tend to increase the wear of an implant. The results of this study prove that single activity wear testing may not be the most clinically
Posterior stabilized (PS) total knee arthroplasty (TKA), wherein mechanical engagement of the femoral cam and tibial post prevents abnormal anterior sliding of the knee, is a proven surgical technique. However, many patients complain about abnormal clicking sensation, and several reports of
Background. Distal femoral replacements (DFR) are used in children for limb-salvage procedures after bone tumor surgery. These are typically modular devices involving a hinged knee axle that has peripheral metal-on-polyethylene (MoP) and central metal-on-metal (M-M) articulations. While modular connections and M-M surfaces in hip devices have been extensively studied, little is known about long-term wear or corrosion mechanisms of DFRs. Retrieved axles were examined to identify common features and patterns of surface damage, wear and corrosion. Methods. The cobalt chromium alloy axle components from 13 retrieved DFRs were cleaned and examined by eye and with a stereo microscope up to 1000× magnification. Each axle was marked into 6 zones for visual inspection: the proximal and distal views, and the middle (M-M) and 2 peripheral (MoP) zones. The approximate percentage of the following features were recorded per zone: polishing, abrasion or scratching, gouges or detectable wear, impingement wear (i.e. from non- intentional articulation), discoloration and pitting. Results. In each case, the middle M-M zones showed more damage features compared with peripheral MoP zones. Brown discoloration, presumably due to tribofilm deposits, was the predominant M-M area feature, particularly at the junction between the MoP and M-M zones. Higher magnification showed areas of polishing underlying the discoloration, suggesting repetitive removal of the surface metal and re-deposition of tribofilms (Fig 2B). 9 cases demonstrated reflective patches resembling “thumbprint” or “fish scale” markings, which, under higher magnification, showed signs of scratching and grooving in a radial pattern (Figs 2D, 3A). Pits were occasionally present but appeared to be from third-body damage as signs of corrosion were absent. Features that resembled carbides, sometimes with associated “comet” patterns of scratching were apparent under higher magnification in some areas. The MoP zones showed variable scratching, abrasion and wear polishing. The MoP to M-M junctional areas were demarcated by a distinct band corresponding, in some cases, to a narrow wear groove or gouge. 3 axles showed evidence of
Aims: The goal was to determine if young patients respect the instructions for medical supervision control instructions after a total hip arthroplasty (THA). Methods:We performed 50 THA (Harris cups and ABG I stems) in 39 patients, mean age 38 years ± 11 (15–58). There were thirty-eight 28mm femoral heads (30 zirconia, 8 chromium-cobalt) and 12 chromium-cobalt heads in 22mm. Each patient was told how necessary a regular medical supervision was with the following schedule: controls at 2 months, 1 year then every two years. In 2000, all the patients were evaluated (no lost for follow-up). Wear was measured according to Livermore with a digitizer. Results: At 72 ± 14 months of follow-up (5–9 years), 51% of the patients (20/39) did not come back for the intermediate controls. No factor could be accounted for that, except being a male (p=0.04). Wear was rather severe (average 0.23 mm/year). 37% of the patients with a
The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time.Aims
Methods
The aim of this study was to longitudinally compare the clinical and radiological outcomes of anatomical total shoulder arthroplasty (aTSA) up to long-term follow-up, when using cemented keel, cemented peg, and hybrid cage peg glenoid components and the same humeral system. We retrospectively analyzed a multicentre, international clinical database of a single platform shoulder system to compare the short-, mid-, and long-term clinical outcomes associated with three designs of aTSA glenoid components: 294 cemented keel, 527 cemented peg, and 981 hybrid cage glenoids. Outcomes were evaluated at 4,746 postoperative timepoints for 1,802 primary aTSA, with a mean follow-up of 65 months (24 to 217).Aims
Methods
Retrieved alumina-on-alumina hip joints frequently exhibit a localised region of high wear, commonly called ‘stripe wear’. This ‘stripe wear’ can be replicated in vitro by the introduction of micro-separation, where the joint contact shifts laterally reproducing edge loading during the simulated walking cycle. While the origin of stripe wear is clearly associated with the micro-scale impact resulting from micro-separation, the wear processes leading to its formation and the wear mechanisms elsewhere on the joint are not so well understood. The purpose of this study was to compare the surface microstructure of in vivo and in vitro alumina hip prostheses, and investigate the origins of the damage accumulation mechanisms that lead to prosthetic failure. The in vivo alumina hip prosthesis was Biolox (Ceram-Tec, AG, Plochingen, Gemany) implanted for 11 years [. 1. ]. The in vitro alumina hip prosthesis was Biolox-forte (CeramTec, AG, Plochingen, Gemany), which had been tested in a hip joint simulator under micro-separation at Leeds University using the procedures given in [2]. The worn surfaces of the alumina hip prostheses were investigated using a Scanning Electron Microscopy (SEM). Similar worn surfaces were seen for both in vivo and in vitro samples. Focused ion beam (FIB) microscopy was used to determine the sub-surface damage across the stripe wear. Samples were subsequently removed for Transmission Electron Microscopy (TEM). Sub-surface damage was found to be limited to a few μm beneath the surface; ~ 2μm for in vivo samples and ~1μm for in vitro samples. The transition from mild wear to more
Wear of the underside of modular tibial inserts (“backside wear”) has been reported by several authors. However, the actual volume of material lost through wear of the backside surface has not been quantified. This study reports the results of computerized measurements of tibial inserts of one design known to have a high incidence of backside wear in situ. A series of retrieved TKA components of one design (AMK, Depuy) with evidence of
Two big problems exist with the all polyethylene cemented tibial component; the polyethylene and the cement. The polyethylene is too weak and flexible to bear high tibial load, so it deforms and loosens. The interface stresses are too high when two flexible structures are poorly bonded and heavily loaded. Modularity between the polyethylene tibial component and the metal-backed tray was introduced in the mid-80's for versatility and to facilitate screw fixation for cementless implants. These designs allow exchange of various polyethylene thicknesses, and aids the addition of stems and wedges. Other advantages include the reduction of inventory, and the potential for isolated tibial polyethylene exchanges as a simpler revision procedure. Several studies have documented the high failure rate of isolated polyethylene exchange procedures, because technical problems related to the original components are left uncorrected. However, revision for wear is the simplest revision ever!. Since the late 1980's the phenomena of polyethylene wear and osteolysis have been observed much more frequently when compared with earlier eras. The reasons for this increased prevalence of synovitis, progressive osteolysis, and
Introduction. Titanium nitride (TiN) coatings are used in total hip arthroplasty to reduce friction of bearing couples or to decrease the allergic potential of orthopaedic alloys. Little is known about performance of currently manufactured implants, since only few retrieval studies were performed, furthermore they included a small number of implants manufactured over 15 years ago. Aim of study. To examine wear and degradation of retrieved TiN coated femoral heads articulating with ultra-high molecular weight polyethylene (UHMWPE). Methods. We included eight femoral heads with a made od TiAl6V4 alloy and coated with TiN using Physical Vapour Deposition (PVD). All heads (28 and 32 mm) were retrieved after at least 12 months of use (range 12–56). The reason for revision was aseptic loosening in 6 cases, septic loosening in one case and recurrent dislocations (five episodes) in one uncemented prosthesis. One unused head was included as reference sample. All implants were evaluated with light microscopy, Scanning Electron Microscopy (SEM) with Energy-Dispersive X-ray Spectroscopy (EDS). 30 SEM images from each implant were digitally analysed using ImageJ software to compare damage in loaded and non weight-bearing parts of the heads. Results. Studies with light microscopy revealed severe damage to the dislocated femoral head, with multiple metallic scratches. SEM studies indicated presence of multiple scratches and pinholes with a diameter of 1–10 µm (Fig1a,b,). Residue from the manufacturing process was present in all implants in form of pure Ti droplets found in round voids. In all implants we found irregular areas (diam. 20–50 µm, Fig 1c,d) where the coating was delaminated from the substrate metal with cracks arising from coating defects (Fig1e-h). Some of these debonded fragments were embedded into the PVD layer in weight-bearing parts of all heads. In one head, which was subjected to dislocations we observed deposits of titanium alloy from the acetabular shell (Fig 2a,b). The deposits were accompanied by large patches of delaminated coating as well as multiple cracks (Fig 2c,d). Small fragments of the acetabular titanium alloy damaged the coating in third body mechanism. Surprisingly in three implants we EDS analysis revealed similar spheres (diam. 1–10 µm) containing Niobium (Nb), although this element is not a part of any of the components used in the implants(Fig 2e,f). Interestingly presence of Nb droplets were associated with a higher number of other defect in these heads both in weight-baring and non weight-bearing parts of the heads suggesting inferior coating quality in these cases (Fig 2 g,h; Fig3, cases H2,H5,H8). Conclusions. Compared to previous studies we did not observe
Introduction. Due to the predictability of outcomes achieved with reverse shoulder arthroplasty (rTSA), rTSA is increasingly being used in patients where glenoid fixation is compromised due to presence of glenoid wear. There are various methods to achieve glenoid fixation in patients with glenoid wear, including the use of bone grafting behind the glenoid baseplate or the use of augmented glenoid baseplates. This clinical study quantifies clinical outcomes achieved using both techniques in patients with
One of serious issues in total hip arthroplasty (THA) is the osteolysis which results in aseptic loosening caused by the wear particles from a polyethylene (PE) acetabular cup. In addition, oxidation degradation of PE cup resulting in the fracture or the
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with
Multiaxial rotation of femoral component is generated in a wide range against UHMWPE tibial insert during ambulation or deep bending activities. Simultaneously, microscopic oscillation and twisting might accompany with such a wide-range motion. Such a combined in-vivo kinetics is expected to bring more
We examined 86 polyethylene inserts, retrieved from total and unicompartmental knee prostheses after an average of 39.5 months in situ, grading them from 0 to 3 for seven modes of polyethylene degradation.
We removed 23 Charité artificial discs in 19 patients due to severe back-and legpain and performed an anterior as well posterior fusion. Mean interval between insertion and retrieval is 8,4 years (3,0–16,0 years). The mean age at retrieval is 49,4 years (40–72 years). The cause of persistent or recurrent pain was one or more of several problems thought to be related to the pain: subsidence, migration, wear with or without breakage of the metal marker ring, facetjoint or adjacent degeneration. In all cases wear of the polyethylene core was seen in major or minor degree. The wear was related to posterior impingement possibly due to hyperlordosis and extension instability and was subsequently more pronounced in the peripheral rim than in the central dome. Also a relationship was noted to in vivo oxidation of the polyethylene, especially in the rim. In the periprosthetic tissue an inflammatory reaction was found, more in the
Ceramic-on-ceramic bearing is an attractive alternative to metal-on-polyethylene bearing due to the unique tri-bological advantages of alumina. However, despite the long-term satisfactory results obtained so far in the vast majority of patients, failure may occur in a few cases. Clinical, radiographic, laboratory and microbiological data of 30 consecutive subjects with failed alumina-on-alumina total hip arthroplasties (THA) were analyzed to define if foreign body reaction to wear debris may be responsible for periprosthetic bone resorption, as in conventional metal-to-polyethylene bearings. In all cases, clinical and radiographical material was reviewed, retrieved implants were examined, and histology of periprosthetic tissues was analyzed. Massive osteolysis was never observed. Apart from 5 five patients for which revision surgery was necessary due to the occurrence of late infection, in all other cases failure had occurred due to secondary implant instability (as in the case of screwed sockets, 19 cases) or to malpositioning of the implant (5 cases). One patient suffered from chronic dislocation. In the vast majority of cases, ceramic wear debris was absent or scarce, and did not induce any tissue reaction. In a few cases with