Advertisement for orthosearch.org.uk
Results 1 - 15 of 15
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 360 - 360
1 Mar 2013
Wang T Pelletier M Bertollo N Crosky A Walsh W
Full Access

Introduction. Implant contamination prior to cement application has the potential to affect the cement-implant bond. the consequences of implant contamination were investigated in vitro using static shear loading with bone cement and titanium dowels of differing surface roughness both with, and without contamination by substances that are likely to be present during surgery. Namely; saline, fat, blood and oil, as a negative control. Methods. Fifty Titanium alloy (Ti-6Al-4V) dowels were prepared with two surface finishes comparable to existing stems. The roughness (Ra and Rq) of the dowel surface was measured before and after the pushout test. Four contaminants (Phosphate Buffered Saline (PBS), ovine marrow, ovine blood, olive oil) were prepared and heated to 37°C. Each contaminant was smeared on the dowel surface completely and uniformly approximately 4 minutes prior to implantation. Samples were separated into ten groups (n=5 per group) based on surface roughness and contaminant. Titanium alloy dowels was placed in the center of Polyvinyl chloride (PVC) tubes with bone cement, and equilibrated at 37°C in PBS for 7 days prior to mechanical testing. The push out test was performed at 1 mm per minute. The dowel surface and cement mantel were analyzed using a Scanning Electron Microscopy (SEM) to determine the distribution and composition of any debris and contaminates on the surface. Results. All contaminants decreased stem-bone cement interfacial shear strength. Saline produced the greatest decrease, followed by blood. The effect of fat was less pronounced and similar to that of oil likely due to the strong lipid solvent properties of the methacrylate monomer. For rough dowels, there were differences in ultimate shear strength between control and contaminated groups (p<0.001). Blood and saline groups had lower ultimate shear strength compared to fat and oil (p<0.05) (fig. 1). The ultimate shear strength for smooth samples was not significantly affected by contamination. Increasing surface roughness increased the interfacial bonding strength, even in the presence of contaminants. In control, fat and oil groups, the effect of roughness are significant (p<0.001, p<0.05 and p<0.001 respectively) (fig. 1). Scanning Electron Microscopy (SEM) showed that contaminants influence the interfacial bond by different mechanisms. Although rough surfaces were associated with higher bond strength, they also generated more debris, which could negatively affect the longevity of the implant bond (fig. 2 and fig. 3). Conclusion. The results of this study underscores the importance of keeping an implant free from contamination, and that if contamination does occur, a saline rinse may further decrease the stability of an implant. Contaminants did not significantly affect the bond strength between bone cement and smooth Ti stem, although a trend of improved properties was seen in the presence of lipid based contaminants. Therefore, the influence of contaminants is more important to the shape-closed type stem. Increasing surface roughness dramatically improved the load carrying capability of the implant-cement interface even with contaminants


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_12 | Pages 19 - 19
1 Dec 2022
Eltit F Wang Q Xu S Satra M Liu D Wang R Charest-Morin R Cox M
Full Access

One out of nine Canadian males would suffer prostate cancer (PC) during his lifetime. Life expectancy of males with PC has increased with modern therapy and 90% live >10 years. However, 20% of PC-affected males would develop incurable metastatic diseases. Bone metastases (BM) are present in ~80% of metastatic PC patients, and are the most severe complication of PC, generating severe pain, fractures, spinal cord compression, and death. Interestingly, PC-BMs are mostly osteoblastic. However, the structure of this newly formed bone and how it relates to pain and fracture are unknown. Due to androgen antagonist treatment, different PC phenotypes develop with differential dependency on androgen receptor (AR) signaling: androgen-dependent (AR+), double negative (AR-) and neuroendocrine. How these phenotypes are related to changes in bone structure has not been studied. Here we show a state-of-the-art structural characterization of PCBM and how PC phenotypes are associated to abnormal bone formation in PCBM. Cadaveric samples (n=14) obtained from metastases of PC in thoracic or lumbar vertebrae (mean age 74yo) were used to analyze bone structure. We used micro-computed tomography (mCT) to analyze the three-dimensional structure of the bone samples. After imaging, the samples were sectioned and one 3mm thick section was embedded in epoxy-resin, ground and polished. Scanning electron microscopy (SEM)/energy-dispersive X-ray spectroscopy (EDS) and quantitative backscattering electron (qBSE) imaging were used to determine mineral morphology and composition. Another section was used for histological analysis of the PC-affected bone. Collagen structure, fibril orientation and extracellular matrix composition were characterized using histochemistry. Additionally, we obtained biopsies of 3 PCBM patients undergoing emergency decompression surgery following vertebral fracture and used them for immunohistological characterization. By using mCT, we observed three dysmorphic bone patterns: osteolytic pattern with thinned trabecula of otherwise well-organized structures, osteoblastic pattern defined as accumulation of disorganized matrix deposited on pre-existing trabecula, and osteoblastic pattern with minimum residual trabecula and bone space dominated by accumulation of disorganized mineralized matrix. Comparing mCT data with patho/clinical parameters revealed a trend for higher bone density in males with larger PSA increase. Through histological sections, we observed that PC-affected bone, lacks collagen alignment structure, have a higher number of lacunae and increased amount of proteoglycans as decorin. Immunohistochemistry of biopsies revealed that PC-cells inside bone organize into two manners: i) glandular-like structures where cells maintain their polarization in the expression of prostate markers, ii) diffuse infiltrate that spreads along bone surfaces, with loss of cell polarity. These cells take direct contact with osteoblasts in the surface of trabecula. We define that PCBM are mostly composed by AR+ with some double negative cells. We did not observe neuroendocrine phenotype cells. PCBMs generate predominantly osteoblastic lesions that are characterized by high lacunar density, lack of collagen organization and elevated proteoglycan content. These structural changes are associated with the infiltration of PC cells that are mostly androgen-dependent but have lost their polarization and contact directly with osteoblasts, perhaps altering their function. These changes could be associated with lower mechanical properties that led to fracture and weakness of the PCBM affected bone


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 9 - 9
1 Feb 2017
Aldinger P Pawar V
Full Access

INTRODUCTION. Due to increasing interest into taper corrosion observed primarily in hip arthroplasty devices with modular tapers, efforts towards characterizing the corrosion byproducts are prevalent in the literature [1–4]. As a result of this motivation, several studies postulate cellular induced corrosion due to the presence of remarkable features in the regions near taper junction regions and articulating surfaces [3–5]. Observations made on explanted devices from a retrieval database as well as laboratory tests have led to the alternative proposal of electrocautery-electrosurgery damage as the cause of these features. These surgical instruments are commonly used for hemostasis or different degrees of tissue dissection. METHODS. Scanning electron microscopy (SEM) and energy dispersive spectroscopy (EDS) were used to evaluate the features observed on retrieved devices. Retrieved devices consisted of OXINIUM and cobalt-chromium-molybdenum (CoCrMo) femoral implants, a Titanium-alloy hip stem, and a CoCrMo metal-on-metal femoral head. Electrocautery-electrosurgery damage was created using a SurgiStat II (Valleylab, Colorado) onto various components (CoCrMo, OXINIUM femoral heads as well as Ti-6Al-4V and CoCrMo alloy test stem constructs). Test components were evaluated using the same methods as the retrieved devices. RESULTS. Remarkable features were present on retrieved devices (Figure 1) which were similar to previous studies (3–5). The appearance of these features could be described as crater-like, pitted, scratched, molten or splattered material, and ruffled. These features were present on articulating and non-articulating regions as well as near taper junctions. Testing performed on samples using the SurgiStat II, created features that were similar in appearance (Figure 1). Additionally, material transfer that included an iron peak based on EDS in addition to the cobalt and chromium (present due to native material) was detected in the regions of contact (Figure 2). CONCLUSIONS. It was possible to re-create damage features similar to those previously characterized as remarkable features created by cellular-induced corrosion [3–5]. It is theorized that the high-voltage based electrocautery (commonly Bovie) or high-frequency based electrosurgical devices can result in localized degradation/alteration of oxides and passive regions of commonly used orthopaedic alloys. These surgical instruments, specifically the cutting electrodes, are frequently made of stainless steels which can result in iron transfer during contact with the device. During the surgical use of the electrocautery-electrosurgery instrument, it may be necessary to remove tissue, bone, or cauterize near the implant or explant which may have led to the damage features noted in this study and the previous literature [3–5]. If this damage occurs during the initial implantation of the devices, it may further exacerbate corrosion in the damaged region and/or alter the mechanical integrity of the constructs (i.e. fatigue performance)


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 3 - 3
1 Sep 2012
Langton D Ghandi J Sidaginamale R Mereddy P Joyce T Lord J Natu S Nargol A
Full Access

Introduction. We conducted independent wear analysis of retrieved metal on metal (MoM) hip components from around the world. All patients with resurfaced hips who developed adverse reactions to metal debris (ARMD) were found to have increased wear of the bearing surfaces. This was untrue in patients with large diameter (?36mm) MoM total hip replacements. This led us to search for other factors leading to ARMD. Methods. MoM THR explants retrieved from 78 patients suffering ARMD underwent full volumetric wear analysis of bearing surface and taper-junctions using coordinate measuring machine. Scanning electron microscopy (SEM) used to characterise material composition of specific areas. Results. 34 MoM THRs were found to have relatively low bearing surface wear (< 3mm. 3. /year). In each of these cases, material loss up to 60 microns wear depth was identified on the internal taper-junctions of femoral components. However, volumetric loss was rarely >5mm. 3. Similarly only 65% of metal ion levels of these patients were found to be greater than the MHRA guidance figure (7µg/L). Patterns of material loss at the tapers were consistent with antero-posterior force splaying open the taper-junction. This characteristic pattern was identified in number of commercially available devices (titanium and cobalt chromium stems). Soft tissue lesions were severe in patients found to have taper damage. Histology confirmed severe ALVAL with lymphoid neogenesis in majority cases, suggesting that wear debris from taper junctions may have greater potential to stimulate adverse immune response. Discussion. The results suggest that forces transmitted from large diameter hard-on-hard bearing surfaces are sufficient to cause mechanical damage to modular junctions with secondary localised corrosion. We urge caution in the use of these designs and recommend a re-evaluation of the stem head interface


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 136 - 136
1 May 2016
Lapaj L Mroz A Wendland J Markuszewski J Kruczynski J
Full Access

Introduction. Dislocation of an uncemented total hip replacement (THR) can cause damage to the femoral hear, when it passes through the rim of metal acetabular shell. This can lead to metal transfer on the surface of the head or chipping of bulk head material. Although dislocation is one of most common complications in total hip arthroplasty (THA), little is known if causes any further damage to the articulating surface of ceramic heads in long term observations. Aim of the study. To evaluate, if dislocations of THR with ceramic on polyethylene bearing causes structural damage to the articulating surface of the femoral head in a follow-up of minium 10 years. Materials and methods. MATERIAL. This study included four groups of third generation (CeramTec Biolox forte) ceramic femoral heads with a diameter of 28 mm:. 8 heads dislocated in the first year(managed nonsurgically) where implants functioned for at least 10 years. 9 heads removed within 12 months after implantation because of recurrent dislocations. 11 heads retrieved after a minimum of 10 years from hips where no dislocation occurred. 8 unused heads. METHODS. Surface topography of retrieved heads was evaluated using Scanning Electron Microscopy with Energy – Dispersive X-ray Spectroscopy; roughness measurements were performed in three directions (0, 45,90 degrees) over a distance of 1,5 mm using a contact profilometer. Results. On the dislocated heads dark stained scratches were present in the lower, non weight-bearing part; SEM studies demonstrated presence of multidirectional smear-like metal deposits (Fig 1) with a small number of chipped ceramic grains. EDS analysis of scratches indicated presence of Ti or Ti,Al,V - dependant on construction alloy of the acetabular component. There were no differences in morphology of scratched areas between heads removed within a few months after dislocation and components retrieved after at least 10 years. Interestingly SEM findings within the weight-bearing part of dislocated heads were identical within corresponding areas of femoral heads articulating for 10 or more years as well as unused components (Fig 2). In these areas we observed a small number of pinholes and scratches, which we believe resulted from machining and subsequent polishing. Roughness measurements were consistent with SEM findings (Table 1). Areas covered by metal deposits had a significantly higher roughness compared to undamaged area. We did not observe statistical differences between roughness of weight-bearing area in all examined groups. Conclusions. Our study indicates, that dislocation of femoral head in a uncemented total hip replacement with ceramic on polyethylene bearing results in contained damage in the area which contacted the metal acetabular shell. Dislocation does not affect the weight-bearing surface of the head, even in a long-term follow-up. This study was financed by a National Science Centre grant No. 2012/05/D/NZ5/01840


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 11 - 11
1 May 2016
Russo A Lopomo N Bianchi M Boi M Ortolani A Gambardella A Marchiori G Maltarello M Visani A Marcacci M
Full Access

Introduction. Protective hard coatings are appealing for several technological applications and even for orthopaedic implants and prosthetic devices. For what concerns the application to prosthetic components, coating of the surface of the metallic part with low-friction and low-wear materials has been proposed [1, 2]; at the same time, concerning use of ceramic materials in joint arthroplasty, zirconia-toughned-alumina (ZTA) ceramic material has shown high strength, fracture toughness, elasticity, hardness, and wear resistance [3, 4]. The purpose of this study was to directly deposit ZTA coatings by using a novel sputter-based electron deposition technique, namely Pulsed Plasma Deposition (PPD) [5]. Preliminary characterization of realized coatings from the point of view of morphology, wettability, adhesion and friction coefficients was performed. Materials and methods. PPD technique was used to deposit ZTA coatings; this technique is able to maintain the stoichiometry of the starting target. In this study we started from a cylindrical ZTA target (30 mm diameter × 5 mm thickness, 75% alumina / 25% zirconia) and followed the procedure described by Bianchi et al [5]. Characterization of morphology, micro-structure and chemistry of deposited coatings was performed by Scanning Electron Microscopy (SEM) equipped with Energy Dispersive X-ray Spectroscopy (EDS) and Atomic Force Microscope (AFM). Coating-substrate interface quality were investigated by micro-scratch tests. Measurement of the contact angle between a drop of 1 ml of ultrapure water and the surface of the sample was performed to estimate the degree of wetting. A ZTA-coated stainless steel ball (AISI 420, 3 mm radius) was coupled against medical grade UHMWPE to evaluate the friction of the proposed coupling in preliminary ball-on-disk tribological tests. Results. Coatings up to several microns thick have been deposited by PPD. Deposited ZTA films exhibited a smooth nanostructured surface. Mechanical tests showed a well-adherent films were deposited. In particular, scratch tests assessed a good interface adhesion: first formation of cracking in the coating during testing was reported at about 0.8 N. Hydrophobic behavior of the coating was revealed by contact angles (average contact angle 116° ± 2°), probably due to the nano-roughness of the coating itself. Good average friction coefficient ranging from 0.12 to 0.15 was showed by preliminary tribological tests carried out in deionized water after up to 10000 m tracks. Conclusions. The preliminary results of a novel approach aiming to drastically improve the performance of prosthetic couplings by introducing hard ceramic coating was presented. Showed results suggested the feasibility of pursuing this approach of realizing ZTA coatings by means of PPD technique. Further analyses on mechanical properties, nanoroughness and tribology are ongoing. Well-adherent ZTA films deposited directly on the prosthetic components surface would then allow a drastic improvement of the actual prosthetic behaviour


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 10 - 10
1 May 2016
Russo A Bianchi M Lopomo N Boi M Ortolani A Marchiori G Gambardella A Maltarello M Visani A Marcacci M
Full Access

Introduction. Total joint arthroplasty is frequently necessary when a traumatic or degenerative disease leads to develop osteoarthritis (OA). Nowadays, the main reason for long term prosthesis failure is due to osteolysys and aseptic loosening of the implant itself, that are related to UHMWPE wear debris [1–3]. Different solutions to overcome this issue have been proposed, including different couplings like metal-on-metal and ceramic-on-ceramic. Our hypothesis was that a hard ceramic thin film realized on the plastic component (i.e. UHMWPE) could improve the friction and wear performance in a prosthetic coupling. The purpose of the presented study was therefore to characterize from the point of view of structure and mechanical performance of this ceramic-coated plastic component. The thin films were specifically realized by means of the novel Pulsed Plasma Deposition (PPD) technique [4]. Materials and methods. PPD technique was used to deposit Yttria-stabilized zirconia (YSZ at 3%) films on medical-grade UHMWPE substrates [4]. The morphology and micro-structure were characterized by Scanning Electron Microscopy (SEM) equipped with Energy Dispersive X-ray Spectroscopy (EDS), X-ray diffraction (XRD) and X-ray Photoelectron Spectroscopy (XPS). By means of nanoindentation and scratch tests mechanical properties were investigated. Ball-on-disk tribological tests were carried out in air, deionized water and physiological solution against alumina balls (6 mm diameter, grade 200) used as counterpart; friction evaluation of the proposed approach and the corresponding worn track were analyzed by SEM-EDS. Results. Deposited zirconia films up to several micron thickness exhibited a fully cubic structure and a smooth nanostructured surface. Good high hardness and Young's modulus values (17 GPa and 154 GPa respectively) were revealed by nanoindentation tests, while no radial cracks, spalling or pile-up phenomena were observable at critical fracture tests. Assessment of the very strong interface adhesion was observed by scratch tests (with initial delamination at about 2 N load). When the UHMWPE substrate was covered by a ceramic film as thin as 1.5 micron, an indentation depth reduction of about 330% was registered. Further, the material yielding under an applied constant load (creep) was larger for UHMWPE compared to coated UHMWPE, whose total creep being only the 19% of the total creep of UHMWPE, respectively. Finally, preliminary tribological tests carried out in air against an alumina ball counterpart showed wear rate as low as 3.2*10-6mm3N-1m-1 after 500.000 cycles, showing an average friction coefficient evaluated on unpolished materials ranging from 0.15 to 0.3 in air. Conclusions. The proposed approach was able to directly coat the plastic insert of a commercial implant joint with hard ceramic materials, thus providing specific additional mechanical and superficial properties, while preserving the well established mechanical properties of UHMWPE. The results of this study showed an alternative and promising approach to improve UHMWPE mechanical properties in arthroplasty


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 20 - 20
1 Mar 2013
Zywiel M Syed K Overgaard CB Cheung AC Turgeon T Brandt J
Full Access

Revision of fractured ceramic-on-ceramic total hip replacements with a cobalt-chromium (CoCr) alloy-on-polyethylene articulation can facilitate metallosis and require further expensive revision surgery [1–3]. In the present study, a fifty-two year old male patient suffered from fatal cardiomyopathy after undergoing revision total hip arthroplasty. The patient had received a polyethylene-ceramic acetabular liner and a ceramic femoral head as his primary total hip replacement. The polyethylene-ceramic sandwich acetabular liner fractured in vivo after 58 months and the patient underwent his first revision surgery where he received a Vitamin E stabilized acetabular Polyethylene (PE) liner and a CoCr alloy femoral head with documented synovectomy at that time. After 15 months, the patient was admitted to hospital in cardiogenic shock, with retrieval of the bearing components. Before the second revision surgery, peak serum cobalt levels measured 6,521 μg/L, 78-times greater than serum cobalt levels of 83μg/L associated with cobalt poisoning [4]. Serum titanium levels found in the patient measured 17.5 μg/L) normal, healthy range 0–1.4 μg/L). The retrieved CoCr alloy femoral head had lost a total of 28.3g (24% or an estimated amount of 102 × 10. −9. wear particles (∼2 μm diameter) [1]) within 16 months of in vivo service. Despite initiating a cobalt chelating therapy, the patients' cardiac left ventricular ejection fraction remained reduced at 6%. This was followed by multi-organ failure, and ultimately the patient passed away shortly after being taken off life support. Embedded ceramic particles were found on the backside and articular surfaces of the Vitamin E-stabilized PE acetabular liner. Evidence of fretting wear on the titanium (Ti) alloy acetabular shell was present, possibly explaining the increased serum Ti levels. Scanning electron microscopy and energy dispersive X-ray analyses confirmed Ti alloy transfer on the embedded ceramic particles on the backside PE liner surface and CoCr alloy transfer on the embedded ceramic particles on the articular PE liner surface. A fractured ceramic-on-ceramic total hip replacement should not be revised to a CoCr alloy-on-polyethylene articulation irrespective of concurrent synovectomy [5] as it can cause severe, third-body wear to the CoCr alloy femoral head that can lead to metallosis with fatal, systemic consequences


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_12 | Pages 4 - 4
1 Mar 2013
Hexter A Panagiotidou A Singh J Skinner J Hart A
Full Access

In metal-on-metal (MoM) total hip arthroplasty, the taper interface is where the femoral head (female taper surface) attaches to the trunnion (male taper) of the femoral stem. Corrosion is well reported in metal-on-polyethylene hips but little is known about taper corrosion in MoM devices. The aim of this study was to quantify corrosion in modern-generation stemmed MoM hip systems and gain insight into the nature of the underlying corrosive attack. Taper corrosion was quantified in 161 failed MoM components (head components n=128; femoral stem n=33) from nine hip types with the use of a qualitative subjective scoring system. An unanticipated finding on preliminary inspection of the hips was a region on the female taper surface that contained ridges that directly corresponded with the ridged microthread on the trunnion. The ridges were not present on unimplanted (control) female taper surfaces and therefore a novel four-scale subjective scoring system was devised to quantify the prevalence and severity of this ‘imprinting’ phenomenon. Evidence of corrosion was observed in 81% (131/161) of components, with at least moderate corrosion observed in 58% (94/161). Corrosion was greater on the female taper surface than on the male taper (p=0.034) and the two scores were associated (r=0.784, p=0.001). Imprinting affected all manufacturers and was observed in 64% (82/128) of head components. The corrosion and imprinting scores were strongly correlated (r=0.694, p=0.001). Corrosion was largely confined to the area of the female taper interface where imprinting had occurred i.e. the region that had been in contact with the trunnion microthread. Scanning electron microscopy showed evidence of fretting corrosion and substantial mechanical wear within the ridged region on the female taper surface. Our study indicates that MoM hips are susceptible to taper corrosion. We believe it occurs by a process of “mechanically-assisted crevice corrosion,” involving the following sequence of events: joint fluid enters the taper junction as a result of pumping of fluid along the machined microthread of the trunnion. This results in galvanic corrosion of the anodic surface (the cobalt-chromium femoral head or taper sleeve). The pattern of corrosion of the head taper is determined by the surface profile of the screw thread of the trunnion, thus leaving an imprinted appearance. Historically the ridged microthread was introduced to trunnions to minimise the risk of burst fracture of ceramic heads. However this study indicates that the ridges are detrimental in MoM hips by causing extensive mechanical wear. Thus the possibility that cobalt-chrome and ceramic femoral head components require different trunnion designs needs urgent investigation


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_8 | Pages 134 - 134
1 May 2016
Lapaj L Mroz A Wendland J Markuszewski J
Full Access

Introduction. Titanium nitride (TiN) coatings are used in total hip arthroplasty to reduce friction of bearing couples or to decrease the allergic potential of orthopaedic alloys. Little is known about performance of currently manufactured implants, since only few retrieval studies were performed, furthermore they included a small number of implants manufactured over 15 years ago. Aim of study. To examine wear and degradation of retrieved TiN coated femoral heads articulating with ultra-high molecular weight polyethylene (UHMWPE). Methods. We included eight femoral heads with a made od TiAl6V4 alloy and coated with TiN using Physical Vapour Deposition (PVD). All heads (28 and 32 mm) were retrieved after at least 12 months of use (range 12–56). The reason for revision was aseptic loosening in 6 cases, septic loosening in one case and recurrent dislocations (five episodes) in one uncemented prosthesis. One unused head was included as reference sample. All implants were evaluated with light microscopy, Scanning Electron Microscopy (SEM) with Energy-Dispersive X-ray Spectroscopy (EDS). 30 SEM images from each implant were digitally analysed using ImageJ software to compare damage in loaded and non weight-bearing parts of the heads. Results. Studies with light microscopy revealed severe damage to the dislocated femoral head, with multiple metallic scratches. SEM studies indicated presence of multiple scratches and pinholes with a diameter of 1–10 µm (Fig1a,b,). Residue from the manufacturing process was present in all implants in form of pure Ti droplets found in round voids. In all implants we found irregular areas (diam. 20–50 µm, Fig 1c,d) where the coating was delaminated from the substrate metal with cracks arising from coating defects (Fig1e-h). Some of these debonded fragments were embedded into the PVD layer in weight-bearing parts of all heads. In one head, which was subjected to dislocations we observed deposits of titanium alloy from the acetabular shell (Fig 2a,b). The deposits were accompanied by large patches of delaminated coating as well as multiple cracks (Fig 2c,d). Small fragments of the acetabular titanium alloy damaged the coating in third body mechanism. Surprisingly in three implants we EDS analysis revealed similar spheres (diam. 1–10 µm) containing Niobium (Nb), although this element is not a part of any of the components used in the implants(Fig 2e,f). Interestingly presence of Nb droplets were associated with a higher number of other defect in these heads both in weight-baring and non weight-bearing parts of the heads suggesting inferior coating quality in these cases (Fig 2 g,h; Fig3, cases H2,H5,H8). Conclusions. Compared to previous studies we did not observe severe wear or the coating, however we observed some degradation of the film in vivo. Our results indicate that dislocation can lead to severe failure of the coating in vivo. Moreover presence of Nb residues and coexisting defects in some implants suggests inferior coating quality in these implants and indicates the need for strict monitoring of the production process. This study was funded by a grant from the National Science Centre nr 2012/05/D/NZ5/01840


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 133 - 133
1 Feb 2017
Gascoyne T Turgeon T
Full Access

Introduction. Detailed analysis of retrieved total hip replacements (THRs) is valuable for assessing implant and material successes and failures. Reduction of bearing wear and corrosion and fretting of the head-neck trunnion is essential to implant durability and patient health. This research quantifies and characterizes taper and bearing surface damage on retrieved oxidized zirconium THRs. Methods. Initially, 11 retrieved oxidized zirconium femoral heads were examined along with their associated femoral stems. Relevant patient and retrieval data was collected from clinical charts and radiographs. Taper corrosion (Figure 1) and fretting damage (Figure 2) scoring was performed following the Dyrkacz [1] method. A coordinate measuring machine was used to obtain a detailed surface map of the male and female taper surfaces. Taper surface maps were best-fit with an idealized cone followed by volume subtraction to quantify the amount of material removed as a result of fretting and corrosion processes. Scanning electron microscopy was performed on select samples to identify specific damage modes. Unique surface bumps were noted on the articular surface of select femoral heads (Figure 3). Seventeen femoral heads were added to the analysis specifically for identification of these bumps. Articular surfaces were searched under SEM magnification and bumps were identified and counted. Parametric statistical correlations were performed with SAS v9.3. Results. Mean patient age was 61 years (Range: 35–95) with mean implantation period being 2.0 years (Range: 0.1–11.4) and mean body mass index of 29 kg/m. 2. (Range: 22–46). Revision for infection (n=11), peri-prosthetic fracture (n=5) and dislocation (n=5) were the main reasons for revision. Mean corrosion damage scores were 2.0 and 3.6 (head, neck) while mean fretting damage scores were 8.5 and 5.8 (head, neck). Fretting damage score was weakly correlated with implantation period (p=0.07) while corrosion damage score was not. Mean corrosion and fretting volume measured 0.40 mm. 3. and 0.87 mm. 3. (head, neck). Volume of corrosion and fretting damage did not correlate with implantation period; however neck volume correlated with inclination angle of the acetabular cup (p<0.01). Bearing diameter was not found to correlate with corrosion and fretting damage score or volume. The unique surface bumps were identified in 12 of 28 samples, with 3 samples having <10 bumps. Presence of these bumps did not appear to be related to bearing diameter, implantation period, or any damage metrics. Conclusion. Fretting damage was found to correlate with implantation period, suggesting that is a continuous in vivo process; however, this was not found for corrosion damage. Fretting damage volume correlated with acetabular cup angle; however, this may be coincidence as only 8 samples were included in the analysis. Overall, our corrosion damage scores (2.0–3.6) were lower than previously published values for 28mm & 36mm cobalt-chrome heads (4.5–13.1) [1]. However, our fretting damage scores (5.8–8.5) were higher than previously published (2.8–4.4) [1]. Greater fretting damage on the oxidized zirconium heads may be explained by the softer zirconium alloy compared to that of cobalt-chromium. Further subsurface investigation of the surface bumps is underway using a focused ion beam mill


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 155 - 155
1 Jan 2016
Lopomo N Bianchi M Boi M Maltarello MC Liscio F Visani A Ortolani A Marcacci M Russo A
Full Access

Introduction. Protective hard coatings are appealing for several technological applications like solar cells, organic electronics, fuel cells, cutting tools and even for orthopaedic implants and prosthetic devices. At present for what concerns the application to prosthetic components, the coating of the surface of the metallic part with low-friction and low-wear materials has been proposed [1]. Concerning the use of ceramic materials in joint arthroplasty, zirconia-toughned-alumina (ZTA) reported high strength, fracture toughness, elasticity, hardness, and wear resistance [2]. The main goal of this study was to directly deposit ZTA coating by using a novel sputter-based electron deposition technique, namely Pulsed Plasma Deposition (PPD) [3]. The realized coatings have been preliminary characterized from the point of view of morphology, wettability, adhesion and friction coefficients. Materials and methods. ZTA coatings were deposited by PPD technique, which is able to maintain the stoichiometry of the starting target. In this case we started from a cylindrical ZTA target (30 mm diameter × 5 mm thickness, 75% alumina / 25% zirconia). The morphology, micro-structure and chemistry of deposited coatings were characterized by Scanning Electron Microscopy (SEM) equipped with Energy Dispersive X-ray Spectrosopy (EDS) and Atomic Force Microcscope (AFM). Coating-substrate interface quality were investigated by microscratch tests. The degree of wetting was estimated by measuring the contact angle between a drop of 1 ml of ultrapure water and the surface of the sample. Preliminary ball-on-disk tribological tests were carried out in air and deionized water coupling ZTA-coated stainless steel ball (AISI 420, 3 mm radius, grade 200) against medical grade UHMWPE to evaluate the friction of the proposed coupling. Results. Deposited ZTA films exhibited a smooth nanostructured surface. Coatings up to several microns thick have been deposited by PPD [Fig. 1 – SEM image (left) and cross section (right)]. Mechanical tests showed a well-adherent films were deposited. In particular, the good interface adhesion was assessed by scratch tests, reporting at about 0.8 N the first formation of cracking in the coating during testing. The contact angles revealed an hydrophobic behavior of the coating (average contact angle 116° ± 2°), probably due to the nano-roughness of the coating itself [Fig. 2 – Contact angle]. Preliminary tribological tests carried out in deionized water after up to 10000 m tracks showed good average friction coefficient ranging from 0.12 to 0.15 [Fig. 3 – Friction coefficient]. Conclusions. We have presented the preliminary results of a novel approach aiming to the drastically improve the performance of prosthetic couplings by introducing hard ceramic coating. The results showed suggested the feasibility of pursuing this approach of realizing ZTA coatings by means of PPD technique. Further analyese on mechanical properties, nano-roughness and tribology should be performed. Well-adherent ZTA films deposited directly on the surface of prosthetic components of a joint implant would then allow a drastic improvement of the actual prosthetic behaviour


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 251 - 251
1 Jun 2012
Unwin P Pickford M Shawcross J Blunn G Cannon S Grimer R
Full Access

Infected mega-endoprostheses are difficult to treat with systemic antibiotics due to encapsulation of the implant by fibrous tissue, formation of biofilms and antibiotic resistant bacteria. Modifying the implant surface by incorporating a bactericidal agent may reduce infection. Infection rates are typically in the range of 8% to 30%. This study describes a novel process method of “stitching-in” ionic silver into the implant surface, in vitro testing and its early clinical usage. A novel process has been developed to “stitch in” ionic silver into the upper surface of titanium alloy (Ti6Al4V). The process produces a modification by anodisation of the titanium alloy in dilute phosphoric acid, followed by absorption of silver from an aqueous solution. The engineered surface modification is therefore integral with the substrate and loaded with silver by an ion exchange reaction. Using this technique the maximum inventory of silver for typical a mega-prosthesis is 6mg and this is greater than 300 times lower than the No Observable Adverse Affects Level (NOAEL). Scanning electron microscopy revealed that the silver was concentrated in pits and forming reservoirs of ionic silver exposed to the body tissues. Laboratory-based studies focusing on the safety and efficacy of silver as a bactericidal agent have included investigation into cytotoxicity using fibroblast and osteoblast cell lines, the impact of silver in reducing corrosion and laboratory testing to establish if the modified surface has an effect on the wear and mechanical characteristics. A range of fatigue, static, tensile pull off tests were performed. The silver elution profiles for both silver loaded and HA coated over a silver loaded surface have been examined. Histological studies were also performed to examine the impact of the silver on osseointegration. The in vitro results confirm that silver is an effective antimicrobial agent. The mechanical characterization studies have identified that the surface treatment has no or minimal impact on the implant surface. Early results of the elution studies are encouraging showing that the HA coating of a silver loaded surface does not “seal” in the silver. To date (May 10) 147 silver treated mega-prostheses have been implanted since March 2006. The majority of implants were distal femoral (29%), proximal tibial (23%) or hemiplevic (10%). The most common indication was revision of a failed limb salvage reconstruction (58%), with the dominant cause of failure being infection. The next most common indication was bone tumour (31%) and the large majority were used in the high risk skeletal locations of the tibia (44%) and the pelvis (27%). Early clinical results are encouraging indicating a significant reduction in the incidence of infection. Three implants have been retrieved. An analysis of a proximal humeral replacement that had been in situ for 6 mths identified that there was 10-20% remaining on the implant surface. This novel process of “stitching-in” silver appears to be a safe and effective surface treatment in helping to control infections of mega-prostheses. This technology has the potential to be transferred to other arthroplasty joints


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_10 | Pages 35 - 35
1 May 2016
Shin T Kang K Park S Kwon S Lim Y Lim D
Full Access

Introduction. Cementless arthroplasty has been widely used for younger patients with osteoarthritis and other joint pathology. Cementless arthroplasty will be required to porous surface which is to similar to the trabecular bone for bone ingrowth. Titanium Plasma Spray (TPS) has been worldwide used for the porous coating method on arthroplasty. However, TPS coating is limited that would not to establish optimal porosity for bone ingrowth due to arbitary position of melted powder by plasma gas on substrate. Therefore, it is reported coating detached from its substrate (i.e. arthroplasty) is induced implant loosening. Thus, a novel Laser-aided Direct Metal Tooling (DMT) based on Additive Manufacturing (AM) was developed to overcome these limitations. In this study, we were done to assess stereological analysis, static tensile, shear, abrasion test, and physical analysis for evaluation of the efficacy of DMT which was newly-developed coating technology. Then, mechanical characteristics of DMT coating were compared to commercial TPS coating's. Materials and Methods. First, porosity of the DMT coating was evaluated using Microphotography and Scanning Electron Microscopy (SEM), as described in Figure 1. Static tensile and shear test for assessment of mechanical characteristic in relation to the DMT and TPS coating specimens were conducted on the basis of ASTM F1147 and F1044 using universal testing machine (Endolab®, Servohydraulic Test Frame, DE). Maximum tensile strength and maximum shear strength were evaluated for each specimen (n=5). Abrasion test was performed based on ASTM F1978 using Taber® Rotary Platform Abraser Model 5135 (TABER®Industries, USA). Abrasion losses for each specimen (n=6) were measured at 2, 5, 10, and 100 cycles, respectively. Results. Porosity of the DMT coating was found to be 64 ± 11%. Maximum tensile strength (mean ± SD) of the DMT coating (48.6 ± 4.3 MPa) was lower than TPS's (51.5 ± 11.6 MPa) about 5.6%, but detached position of one of the TPS specimens was observed at coating layer not in adhesive layer. Maximum shear strength of the DMT coating (46.3 ± 1.9 MPa) was 10.2% higher, compared to TPS's (42.0 ± 0.6 MPa). Abrasion losses of the DMT (2 cycles, 1.0 ± 0.5 mg; 5 cycles, 2.5 ± 0.9 mg; 10 cycles, 4.2 ± 0.7 mg; 100 cycles, 20 ± 1.4 mg) were significantly higher (71.9% – 77.8% higher) than that of TPS's (2 cycles, 4.5 ± 1.4 mg; 5 cycles, 9.8 ± 2.8 mg; 10 cycles, 17.0 ± 3.4 mg; 100 cycles, 71.1 ± 4.2 mg), as shown in Figure 2. Roughness of the DMT coating was Ra 62.5 ± 2 μm, Rz 316.1 ± 8.1 μm which were 33.5%, 40.6% (Ra, Rz sequence) higher than that of TPS coating (Ra 46.8 ± 8.9 μm, Rz 224.9 ± 28.8 μm), as shown in Figure 3. Conclusions. Our results suggested that a novel DMT coating technology was feasible to apply coating method on the surface of the arthroplasty in terms of outstanding mechanical characteristics which were compared to commercial TPS coating


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 115 - 115
1 Apr 2019
Bock R Cullen D McEntire B More K Leonard D Bal BS
Full Access

Introduction. The osteogenic capability of any biomaterial is governed by a number of critical surface properties such as surface energy, surface potential, and topography. Prior work suggested that the Si-Y-O-N phase(s) present in the form of a thin (<150 nm), interrupted film at the surface of an annealed silicon nitride bioceramic may be responsible for an observed upregulation of osteoblastic activity due to passive surface properties and dissolution of chemical species. In this study high- resolution analytical electron microscopy was utilized to identify the Si-Y-O-N phase present on the annealed silicon nitride surface, and dissolution studies were employed to elucidate mechanisms of the material's favorable cell interactions. Materials and Methods. Si. 3. N. 4. discs (12.7 mm diameter × 1 mm thick) containing Y. 2. O. 3. and Al. 2. O. 3. sintering aids were processed using conventional techniques and subsequently subjected to annealing in a nitrogen atmosphere. Pre-cultured SaOS-2 osteosarcoma cells at a concentration of 5 × 10. 5. cells/ml were seeded onto sterile polished nitrogen-annealed Si. 3. N. 4. discs in an osteogenic medium consisting of DMEM supplemented with about 50 µg/mL ascorbic acid, 10 mM β-glycerol phosphate, 100 mM hydrocortisone, and 10% fetal bovine calf serum. The samples were incubated for up to 7 days at 37°C with two medium replenishments. Transmission electron microscopy (TEM) images were acquired from focused ion beam (FIB)-prepared samples using a Hitachi HF-3300 TEM (300 kV). Scanning transmission electron microscopy (STEM) images were recorded using a Nion UltraSTEM 100 (60 kV). STEM high-angle annular dark-field (HAADF) imaging and energy dispersive X-ray spectroscopy (EDS) analyses were performed on a JEOL JEM2200FS (200 kV) equipped with a third-order CEOS aberration corrector and a Bruker XFlash silicon drift detector. Results. A cross-section of the of the Si. 3. N. 4. /extracellular polymer (ECP) interface is illustrated in Fig. 1(a)∼(b) as a high- angle annular dark field (HAADF) STEM image (a) with and EDS map overlay (b) highlighting locations of Ca, Y, and Si. The underlying Si. 3. N. 4. microstructure is covered by a yttrium-rich intergranular phase (IGP) film. Deposition of cell-derived hydroxyapatite (HAp) occurred directly onto this IGP film. In Fig. 2, a bright field TEM image (electron diffraction pattern inset) shows the interface between the partially-crystalline HAp and the Y-Si-O-N phase, identified as monoclinic yttrium disilicate (i.e., m-Y. 2. Si. 2. O. 7. ) with a 2 atomic% N impurity, at teh atomic scale. Although rapid electron damage of the mineralized ECP was observed, EDS analyses suggested a Ca/P ratio of ∼1.43, along with the incorporation of Si. Conclusions. The osteogenic Si-Y-O-N phase was successfully identified as a minority concentration of Si. 3. N. 4. dissolved into a m-Y. 2. Si. 2. O. 7. matrix. Evidence of the release of (SiO. 4. ). 4−. tetrahedra from this phase into the local biological microenvironment and their incorporation into the cell-derived HAp layer was also observed. Identification of this phase paves the way for ongoing work to understand and optimize this novel biomaterial. For any figures or tables, please contact the authors directly