Devane PA Horne JG Chu A
Full Access

We present minimum 20 year results of a randomized, prospective double blinded trial (RCT) of cross-linked versus conventional polyethylene (PE), using a computer assisted method of PE wear measurement. After Ethics Committee approval, 122 patients were enrolled into an RCT comparing Enduron (non cross-linked PE) and highly cross-linked Marathon PE (DePuy, Leeds, UK). Other than the PE liners, identical components were used, a Duraloc 300 metal shell with one screw, a 28mm CoCr femoral head and a cemented Charnley Elite femoral stem. All patients were followed with anteroposterior (A∼P) and lateral radiographs at 3 days, 6 weeks, 3 months, 6 months, 1, 2, 3, 4, 5, 10 and 20 years. PE wear was measured with PolyMig, which has a phantom validated accuracy of ± 0.09mm. At minimum 20 year follow-up, 47 patients had died, 5 of which had been revised prior to their death. Another 32 patients were revised and alive, leaving 43 patients unrevised and alive (15 Enduron, 28 Marathon). No patients were lost to follow-up, but 2 were not able to be radiographed (dementia), leaving 41 patients (15 Enduron, 26 Marathon) available for PE wear measurement. After the bedding-in period, Enduron liners had a wear rate of 0.182 mm/year, and Marathon liners had a wear rate of 0.028 mm/year. At 20 years follow-up, 37 patients had required revision. Patients with conventional PE had three times the revision rate (28/37) of those who received XLPE (9/37). This is the longest term RCT showing substantially improved clinical and radiological results when XLPE is used as the bearing surface


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 801 - 810
1 Jul 2022
Krull P Steinbrück A Grimberg AW Melsheimer O Morlock M Perka C

Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model. Results. Median follow-up was 868 days (interquartile range 418 to 1,364). The offset liner design reduced the risk of revision (hazard ratio (HR) 0.68 (95% confidence interval (CI) 0.50 to 0.92)), while the angulated/offset liner increased the risk of revision for mechanical failure (HR 1.81 (95% CI 1.38 to 2.36)). The cumulative incidence of revision was lowest for the offset liner at one and seven years (1.0% (95% CI 0.7 to 1.3) and 1.8% (95% CI 1.0 to 3.0)). No difference was found between standard, lipped, and angulated liner designs. Higher age at index primary THA and an Elixhauser Comorbidity Index greater than 0 increased the revision risk in the first year after surgery. Implantation of a higher proportion of a single design of liner in a hospital reduced revision risk slightly but significantly (p = 0.001). Conclusion. The use of standard acetabular component liners remains a good choice in primary uncemented THA, as most modified liner designs were not associated with a reduced risk of revision for mechanical failure. Offset liner designs were found to be beneficial and angulated/offset liner designs were associated with higher risks of revision. Cite this article: Bone Joint J 2022;104-B(7):801–810


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 71 - 77
1 Jul 2020
Gonzalez Della Valle A Shanaghan KA Nguyen J Liu J Memtsoudis S Sharrock NE Salvati EA

Aims. We studied the safety and efficacy of multimodal thromboprophylaxis in patients with a history of venous thromboembolism (VTE) who undergo total hip arthroplasty (THA) within the first 120 postoperative days, and the mortality during the first year. Multimodal prophylaxis includes discontinuation of procoagulant medications, VTE risk stratification, regional anaesthesia, an intravenous bolus of unfractionated heparin prior to femoral preparation, rapid mobilization, the use of pneumatic compression devices, and chemoprophylaxis tailored to the patient’s risk of VTE. Methods. Between 2004 to 2018, 257 patients with a proven history of VTE underwent 277 primary elective THA procedures by two surgeons at a single institution. The patients had a history of deep vein thrombosis (DVT) (186, 67%), pulmonary embolism (PE) (43, 15.5%), or both (48, 17.5%). Chemoprophylaxis included aspirin (38 patients), anticoagulation (215 patients), or a combination of aspirin and anticoagulation (24 patients). A total of 50 patients (18%) had a vena cava filter in situ at the time of surgery. Patients were followed for 120 days to record complications, and for one year to record mortality. Results. Postoperative VTE was diagnosed in seven patients (2.5%): DVT in five, and PE with and without DVT in one patient each. After hospitalization, three patients required readmiss-ion for evacuation of a haematoma, one for wound drainage, and one for monitoring of an elevated international normalized ratio (INR). Seven patients died (2.5%). One patient died five months postoperatively of a PE during open thrombectomy. She had discontinued anticoagulation. One patient died of a haemorrhagic stroke while receiving Coumadin. PE or bleeding was not suspected in the remaining five fatalities. Conclusion. Multimodal prophylaxis is safe and effective in patients with a history of VTE. Postoperative anticoagulation should be prudent as very few patients developed VTE (2.5%) or died of suspected or confirmed PE. Mortality during the first year was mostly unrelated to either VTE or bleeding. Cite this article: Bone Joint J 2020;102-B(7 Supple B):71–77


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_11 | Pages 23 - 23
7 Jun 2023
Coveney E Hutton C Patel N Whitehouse S Howell J Wilson M Hubble M Kassam A
Full Access

Venous thromboembolism (VTE) is a preventable cause of morbidity and mortality in patients undergoing elective hip arthroplasty surgery. The balance of post-operative VTE prophylaxis and risk of post-operative haemorrhage remains at the forefront of surgeon's mind. The National Institute for Clinical Excellence (NICE) has altered their prophylaxis guidance in the setting of total hip arthroplasty (THA). The aim of this study was to present the VTE incidence in 8,890 patients who underwent total hip arthroplasty between January 1997 and March 2018 with Aspirin as the primary agent for pharmacological thromboprophylaxis. Analysis of prospective data collection from consecutive patients undergoing THA was performed with the incidence of deep vein thrombosis (DVT) and pulmonary embolism (PE) occurring within 6 months of the index operation as the primary outcome measure. 90-day all-cause mortality of this cohort of patients was also analysed. 8890 patients were reviewed. This included 7235 primary, 224 complex primary and 1431 revision cases. The incidence of DVT was 0.64% after elective THA and the incidence of PE was 0.54%. There was no difference in the incidence between primary and revision cases. The 90-day all-cause mortality was 0.88%. Cardiovascular and respiratory disease were the main causes of death following surgery. Only 0.03% of deaths (n= 3) within 90 days of index surgery were due to VTE. Our results support the use of aspirin as an effective form of prophylaxis against VTE following THA. It is not associated with an increased incidence in symptomatic DVT, PE or death compared to other published studies. The fact that it is inexpensive, readily available, requires no monitoring and does not pose an increased risk of bleeding are other attractive advantages of using aspirin for VTE prophylaxis


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 15 - 15
2 May 2024
Williams S Smeeton M Isaac G Anderson J Wilcox R Board T Williams S
Full Access

Dual Mobility (DM) Total Hip Replacements (THRs), are becoming widely used but function in-vivo is not fully understood. The aim of this study was to compare the incidence of impingement of a modular dual mobility with that of a standard cup. A geometrical model of one subject's bony anatomy \[1\] was developed, a THR was implanted with the cup at a range of inclination and anteversion positions (Corail® stem, Pinnacle® cup (DePuy Synthes)). Two DM variants and one STD acetabular cup were modelled. Joint motions were taken from kinematic data of activities of daily living associated with dislocation \[2\] and walking. The occurrence of impingement was assessed for each component combination, orientation and activity. Implant-implant impingement can occur between the femoral neck and the metal or PE liner (DM or STD constructs respectively) or neck-PE mobile liner (DM only). The results comprise a colour coded matrix which sums the number of impingement events for each cup position and activity and for each implant variant. Neck-PE mobile liner impingement, occurred for both DM sizes, for all activities, and most cup placement positions indicating that the PE mobile liner is likely to move at the start of all activities including walking. For all constructs no placement positions avoided neck-metal (DM) or neck-PE liner (STD) impingementevents in all activities. The least number of events occurred at higher inclination and anteversion component positions. In addition to implant-implant impingement, some instances of bone-bone and implant-bone impingement were also observed. Consistent with DM philosophy, neck-PE mobile liner impingement and liner motion occurred for all activities including walking. Neck-liner impingement frequency was comparable between both DM sizes (metal liner) and a standard cup (PE liner)


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 96 - 96
19 Aug 2024
Gauthier P Garceau S Parisien A Beaulé PE
Full Access

The purpose of our study is to examine the outcome of patients undergoing outpatient total hip arthroplasty with a BMI >35.

Case-control matching on age, gender (46% female;54%male), and ASA (mean 2.8) with 51 outpatients BMI≥35 kg/m2 (mean of 40 (35–55)), mean age of 61 (38–78) matched to 51 outpatients BMI<35 kg/m2 (mean of 27 (17–34)) mean age 61 (33–78). Subsequently 47 inpatients BMI≥35 kg/m2 (mean of 40 (35–55)) mean age 62 (34–77) were matched outpatients BMI≥35 kg/m2. For each cohort, adverse events, readmission in 90 days, reoperations were recorded.

Rate of adverse events was significantly higher in BMI ≥35: 15.69% verus 1.96% (p=0.039) with 5 reoperations in the BMI≥35 cohort vs 0 in the BMI<35 kg/m2 (p= 0.063). Readmissions did not differ between groups (p=0.125). No significant difference for all studied outcomes between the outpatient and inpatients cohorts with BMI≥35 kg/m2. The most complications requiring surgery/medical intervention (3B) were in the inpatient cohort of patients >35. The prevalence of Diabetes and Obstructive Sleep apnea was 21.6% and 29.4% for BMI>35 compared to 9.8% and 11.8%, for BMI <35, respectively.

Severely obese patients have an overall higher rate of adverse events and reoperations however it should not be used a sole variable for deciding if the patient should be admitted or not.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 24 - 24
1 Oct 2020
Beaule PE
Full Access

Introduction

Cemented femoral fixation has been shown to carry a lower risk of peri-prosthetic fracture (PPF). The aim of this study was to determine whether adequate (1) stem position and (2) cement mantle (i.e. factors associated with outcome of cemented stems) can be achieved with the anterior approach as compared to the posterior approach.

Methods

This is a prospective, multi-surgeon, single center, consecutive, case-matched series. Twenty patients/hips (age: 76±SD 14) that underwent cemented fixation of their femoral component via an anterior approach (AA) were matched with twenty hips that received the same cemented femoral components via the posterior approach by the same surgeons. Outcome measures of interest included Barrack classification (A, B, C or D), Coronal alignment defined as Varus (>5º), mild varus (3–5º), neutral, mild valgus (3–5º) and valgus (>5º) of the stem and sagittal component alignment defined as anterior to posterior, neutral, posterior to anterior (±3 degrees) and cement mantle thickness in all Gruen zones.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 28 - 28
1 Nov 2021
Perka C Krull P Steinbrück A Morlock M
Full Access

Special acetabular polyethylene (PE) liners are intended to increase the stability of the artificial hip joint, yet registry studies on them are limited. The pupose of this study was to investigate differences in revision rates for mechanical complications in primary cementless total hip arthroplasty (THA) with standard and special PE acetabular liners in patients with ostheoarthritis. Data from the German Arthroplasty Registry (EPRD) between 2012 until 2020 were analysed. Patients with diagnosed ostheoarthritis of the hip without relevant prior surgeries, who received a primary cementless THA with a ceramic/PE bearing articulation were included. Cumulative incidences of revision for mechanical complications for Standard and 4 special PE liners (Lipped, Increased Offset, Angulated, Angulated|Increased Offset) were determined using the Kaplan-Meier Estimator. Confounding factors were investigated with a Cox proportional-hazards model. In total 151.104 cases were included. 7-year unadjusted revision-free survival for mechanical complications compared to Standard liners (97.7%) was lower for Angulated (97.4%), Lipped (97.2%) and Angulated|Increased Offset liners (94.7%), but higher for Increased Offset liners (98.1%). Risk of revision for mechanical complications was not significantly different between Standard, Lipped and Angulated liners. Increased Offset liners (HR=0.68; 95% CI=0.5–0.92) reduced, while Angulated|Increased Offset liners (HR= 1.81; 95% CI=1.38–2.36) increased the risk. Higher age at admission and an Elixhauser comorbidity index greater zero increased the risk, whereas a larger liner share slightly reduced the risk. Only the use of Increased Offset liners reduced the risk of revision for mechanical complications compared to Standard liners — other special liners did not


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 1 - 1
1 Nov 2021
Mallett K Trousdale R Sierra R Abdel M Berry D
Full Access

Previous work has demonstrated increased implant failure in patients with DDH compared to osteoarthritis with historic methods and implants. This study examines outcomes of modern total hip arthroplasty (THA) techniques using uncemented fixation and cross-linked PE (XLPE) bearings for DDH. A consecutive series of 879 patients with DDH who underwent primary THA with uncemented components and a XLPE bearing at a single institution between 1999 and 2016 were identified. Mean age at index arthroplasty was 51 years, with 78% females. Mean follow-up was 8 years. 5- and 10-year survivorships free of revision were 98.4% and 98%, respectively. 5- and 10-year survivorships free of reoperation for any reason were 97% and 96%, respectively. Survivorship free of acetabular revision for aseptic loosening was 99.8% at 5 and 10 years, while survivorships free of femoral revision for aseptic loosening were 99.7% at 5 years and 99.5% at 10 years. Survivorship free of reoperation for PE damage (2 liner fractures, 2 rim damage from impingement) was 99.3% at 10 years. There were no revisions for bearing surface wear or osteolysis. 13% of patients experienced complications (4% dislocation, 3.3% wound complication rate). Ten-year survivorship free of reoperation for instability was 99.2%. 8 patients developed infection requiring operation (<1%). 100 patients had a prior pelvic osteotomy (11%), which did not increase rate of complications (p=0.22) or reoperations (p=0.51). Fixation with modern uncemented implants has dramatically reduced implant loosening in patients with DDH, and the use of XLPE bearings has markedly reduced revision for PE wear/osteolysis in this young patient population, leading to dramatically improved mid/long-term survivorship compared to historic series


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_5 | Pages 27 - 27
1 Jul 2020
Wyatt M Whitehouse M Kieser D Frampton C Hooper G
Full Access

Background. Reduced dislocation rates using lipped polyethylene (PE) liners in modular uncemented acetabular components has been shown, yet there may be increased wear because of impingement, which may lead to aseptic loosening. We used New Zealand Joint Registry (NZJR) data to compare survival rates, revision rates for dislocation and aseptic loosening between lipped and neutral liners. Methods. 31,247 primary THAs using the four commonly used uncemented modular cups were identified (January 1, 1999 to December 31, 2018). The lipped liner group comprised 49% males vs 42% in the neutral group (p < 0.001); 96% had OA vs 95% in the neutral group (p < 0.001). There was no difference in other patient characteristics. Mean follow-up was 5.1 years (SD 3.9); longest follow-up 19.3 years. Kaplan Meier survival rates were compared (20,240 lipped and 11,007 neutral PE liners). Highly cross-linked PE was used in 99% of lipped liner cups vs 85% of neutral liner cups. Associated hazard ratios were calculated using a Cox regression analysis. Results. KM survival at 10 years for lipped PE liners was 96% for lipped (95%CI 95.4%-96.2%) and 95% for neutral liners (95%CI 94.7%-95.9%). Controlling for age, gender approach, head size, image guidance, the all-cause revision risk was greater for neutral liner (HR 1.17 [95% CI 1.06 to 1.36]; p = 0.032). There was a higher risk of revision for dislocation in those with neutral liners (HR 1.84 [95%CI 1.41–2.41]; p < 0.001) but no difference in the revision rate for aseptic acetabular loosening (HR 0.85 [95%CI 0.52–1.38]; p = 0.511). Conclusions. Using lipped PE liners is not associated with a higher rate of aseptic loosening in patients who undergo primary THA. Lipped PE liners are associated with lower rates of dislocation and lower all-cause revision rates without any increased association with revision rates for wear and aseptic loosening


Bone & Joint Open
Vol. 5, Issue 1 | Pages 28 - 36
18 Jan 2024
Selmene MA Moreau PE Zaraa M Upex P Jouffroy P Riouallon G

Aims

Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications.

Methods

Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.


Bone & Joint Open
Vol. 3, Issue 10 | Pages 804 - 814
13 Oct 2022
Grammatopoulos G Laboudie P Fischman D Ojaghi R Finless A Beaulé PE

Aims

The primary aim of this study was to determine the ten-year outcome following surgical treatment for femoroacetabular impingement (FAI). We assessed whether the evolution of practice from open to arthroscopic techniques influenced outcomes and tested whether any patient, radiological, or surgical factors were associated with outcome.

Methods

Prospectively collected data of a consecutive single-surgeon cohort, operated for FAI between January 2005 and January 2015, were retrospectively studied. The cohort comprised 393 hips (365 patients; 71% male (n = 278)), with a mean age of 34.5 years (SD 10.0). Over the study period, techniques evolved from open surgical dislocation (n = 94) to a combined arthroscopy-Hueter technique (HA + Hueter; n = 61) to a pure arthroscopic technique (HA; n = 238). Outcome measures of interest included modes of failures, complications, reoperation, and patient-reported outcome measures (PROMs). Demographic, radiological, and surgical factors were tested for possible association with outcome.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 59 - 59
23 Jun 2023
Hernigou P
Full Access

The variables involved in a robotic THA can exceed 52: many parameters as pelvic orientation with CT scan, templating, offset, and leg-length, acetabular reaming, femoral osteotomy, mapping the anatomy; predefining safe zones, robotic execution, femoral head size, thickness of PE etc. with several variables for each parameter, with a total number of variables exceeding 52. This familiar number is the number of cards in a standard deck. The number of possible combinations (factorial 52! = 10^67) to shuffle the cards (and may be to perform a THA) is greater than the number of atoms on earth! Thinking that artificial intelligence and robotics can solve these problems, some surgeons and implant manufacturers have turned to artificial intelligence and robotics. We asked two questions:1) can robot with artificial intelligence really process 52 variables that represent 10^67 combinations? 2) the safety of the technology was ascertained by interrogating Food and Drug Administration (FDA) database about software-related recalls in computer-assisted and robotic arthroplasty [1], between 2017 and 2022. 1). The best computers can only calculate around 100 thousand billion combinations (10^14), and with difficulty: it takes more than 100 days to arrive at this number of digits (10^14) after the decimal point for the number π (pi). We can, therefore, expect the robot to be imperfect. 2). For the FDA software-related recalls, 4634 units were involved. The FDA determined root causes were: software design (66.6%), design change (22.2%), manufacturing deployment (5.6%), design manufacturing process (5.6%). Among the manufacturers’ reasons for recalls, a specific error was declared in 88.9%. a coding error in 43.8%. 94.4% software-related recalls were classified as class 2. Return of the device was the main action taken by firms (44.4%), followed by software update (38.9%). 3). In the same period, no robot complained about its surgeon!. Hip surgeon is as intelligent as a robot and almost twice as safe


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 29 - 29
2 May 2024
Gibbs V Champaneria R Sandercock J Welton N Geneen L Brunskill S Doree C Kimber C Palmer A Estcourt L
Full Access

Preoperative anaemia and intraoperative blood loss result in ∼90% of individuals being anaemic following hip and knee arthroplasty. Reducing blood loss offers the opportunity to improve outcomes and reduce the risk of transfusion and costs. This review's aim was to determine the effectiveness of drugs for preventing blood loss, and identify optimal dose, route, and timing of administration. Cochrane network meta-analysis of randomised controlled trials was conducted. Inclusion criteria: adults undergoing primary or revision elective hip or knee arthroplasty. Drugs studied: tranexamic acid (TXA), aprotinin, epsilon-aminocaproic acid, desmopressin, factor VIIa and XIII, fibrinogen, fibrin sealants, and non-fibrin sealants. Primary outcomes: need for allogenic blood transfusion, all• cause mortality (within 30 days). Secondary outcomes: mean number of transfusion episodes, re-operation, length of hospital stay and adverse events (DVT, PE, CVA, MI). 102 studies with 8418 participants. Trials included more women (63%). 47 studies (4398 participants) were included within the blood transfusion NMA. TXA given intra-articularly and orally at a total dose of greater than 3g pre-incision, intraoperatively and postoperatively ranked the highest, with anticipated absolute effect of 147 fewer transfusions per 1000 (53% chance ranking 1st) (relative risk(RR) 0.02, 95% credible interval(CrI) 0–0.31); moderate-certainty). Aprotinin (RR 0.59, 95%:CrI 0.36–0.86; low certainty evidence), fibrin (RR 0.86, CrI 0.25–2.93; very-low certainty) and EACA (RR 0.60, 95%:CrI 0.29–1.27; very-low certainty) were not shown to be as effective as TXA. TXA was the most effective drug for preventing bleeding in lower limb arthroplasty. Aprotinin and EACA were not as effective. Currently, the optimal dose, route and timing of administration of TXA is unclear. However, TXA given at higher doses and via mixed routes ranked higher in the treatment hierarchy. Oral TXA may be as effective as intavenous. There was no evidence of harm associated with higher doses of TXA


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 60 - 60
19 Aug 2024
Lau LCM Cheung MH Ng FY Fu H Chan PK Chiu P
Full Access

In total hip arthroplasty (THA), cementless cup without screw holes has the putative benefits of maximizing host bone contact and reducing osteolysis by eliminating channels to backside wear particles. However, supplemental trans-acetabular screws cannot be used. 74 hips in 60 patients receiving same model of cementless cup without screw holes (Depuy Duraloc 100 HA cup) from 6/1999 to 3/2003 were prospectively followed up. All patients were allowed to have immediate full weight bearing. Age at THA was 53 ± 13 (range 24–74) years. Osteonecrosis was the leading hip disease (45% of hips). Survivorship was assessed using revision of the cup as the end point. Radiological parameters, including lateral opening angle, vertical and horizontal migration distances of the cups were measured. Paired t-test was used to compare between the measurements in early postoperative period and at final follow up. 51 hips were assessed at minimum 20 years follow-up. The mean follow-up was 22.6 (range 21 – 25) years. All the cups were well fixed. There were two cup revisions. Conventional polyethylene (PE) was used in both hips; osteolysis occurred 17 and 18 years later. Both cups were well fixed but were revised, one due to cup mal-positioning, one due to need in upsizing the articulation. 14 other hips were revised but these cups were well fixed and not revised; 9 loosened stems (most were cemented Elite plus stems), 5 PE wear and osteolysis (all were conventional PE). At 20 years, the survivorship of cups was 96.1%. Changes in lateral opening angle, vertical and horizontal migration distances were 0.44±1.59°, 0.01±1.52mm and -0.32±1.47mm respectively, without statistical significance. This study provided evidence of excellent long-term survivorship of cementless cup without screw holes. Immediate postoperative weight-bearing also did not lead to cup migration in the long-term


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 25 - 25
23 Jun 2023
Ricard M Pacheco L Koorosh K Poitras S Carsen S Grammatopoulos G Wilkin G Beaulé PE
Full Access

Our understanding of pre-arthritic hip disease has evolved tremendously but challenges remain in categorizing diagnosis, which ultimately impacts choice of treatments and clinical outcomes. This study aims to report patient reported outcome measures (PROMs) comparing four different condition groups within hip preservation surgery by a group of fellowship-trained surgeons.

From 2018 to 2021, 380 patients underwent hip preservation surgery at our center and were classified into five condition groups: dysplasia: 82 (21.6%), femoro-acetabular impingement (FAI): 173 (45.4%), isolated labral tear: 103 (27.1%), failed hip preservation: 20 (5.3%) and history of childhood disease/other: 2 (0.5%). International hip outcomes Tool 12 (IHOT-12), numeric pain score and patient-reported outcomes measurement information system (PROMIS) were collected pre-operatively and at 3 months and 1 year post-operatively, with 94% and 82% follow-up rate respectively.

Arthroscopy (75.5%) was the most common procedure followed by peri-acetabular osteotomy (PAO) (22.4%) and surgical dislocation (2.1%). Re-operation rate were respectively 18.3% (15), 5.8% (10), 4.9% (5), 30% (6) and 0%. There were 36 re-operations in the cohort, 14 (39%) for unintended consequences of initial surgery, 10 (28%) for mal-correction leading to a repeat operation, 8 (22%) progression of arthritis, and 4 (11%) for incorrect initial diagnosis/intervention. Most common re-operations were hardware removal 31% (7 PAO, 3 surgical hip dislocation and 1 femoral de-rotational osteotomy), arthroscopy 31% (11) and arthroplasty 28% (10). All groups had significant improvements in their IHOT-12 as well as PROMIS physical and numerical pain scales, except those with failed hip preservation. Dysplasia group showed a slower recovery.

Overall, this study demonstrated a clear relation between the condition groups, their respective intervention and the significant improvements in PROMs with isolated labral pathology being a valid diagnosis. Establishing tertiary referral centers for hip preservation and longer follow-up is needed to monitor the overall survivorship of these various procedures.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 39 - 39
1 Oct 2019
Valle AGD Shanaghan KA Salvati EA
Full Access

Introduction. We studied the safety and efficacy of multimodal thromboprophylaxis (MMP) in patients with a history of venous thromboembolism (VTE) undergoing total hip arthroplasty (THA). MMP includes discontinuation of procoagulant medications, VTE risk stratification, regional anesthesia, an intravenous bolus of unfractionated heparin before femoral work, rapid mobilization, the use of pneumatic compression devices, and chemoprophylaxis tailored to the patient's risk. Material and methods. From 2004 to 2018, 257 patients (mean age: 67 years; range: 26–95) with a history of VTE underwent 277 primary, elective THAs procedures (128 right, 100 left, 9 single-stage bilateral, 20 staged bilateral) by two orthopaedic surgeons at a single institution. The patients had a history deep vein thrombosis (DVT) 186 (67%), pulmonary embolism (PE) 43 (15.5%), or both 48 (17.5%). Chemoprophylaxis included aspirin (38 patients) and anticoagulation (239 patients; Coumadin: 182, low-molecular-weight heparin: 3, clopidogrel: 1, rivaroxaban: 3, and a combination: 50). Forty eight patients (17.3%) had a vena cava filter at the time of surgery. Patients were followed for 120 days to detect complications, and for a year to detect mortality. Results. Postoperative VTE was diagnosed in seven patients (2.5%): DVT in five, and PE with and without DVT in one patient each. Bleeding complications occurred in 2 patients, one requiring surgical evacuation of a hematoma. Seven patients died during the first year (2.5%). One patient died 5 months postoperatively of a fatal PE during open thrombectomy, and one patient died of a hemorrhagic stroke while receiving Coumadin. PE or bleeding was not suspected in any of the remaining 5 fatalities. Conclusions. The result of this study spanning over 13 years, suggests that MMP is safe and effective. Postoperative anticoagulation should be prudent as very few patients developed postoperative VTE (2.5%) or died of suspected or confirmed PE. Mortality during the first year was mostly unrelated to VTE or bleeding. For any tables or figures, please contact the authors directly


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 64 - 64
1 Oct 2018
Glassberg MB Lachiewicz PF
Full Access

Introduction. Many pharmacologic agents have been used for venous thromboembolism (VTE) prophylaxis after elective total hip arthroplasty (THA). Rivaroxaban was the first novel oral anticoagulant approved for THA patients, but its actual efficacy and safety in clinical practice, beyond randomized trials, is unknown. Materials and Methods. This is a retrospective study, using the Truven Health MarketScan database, of anticoagulation medication prescriptions after elective THA, in both commercially insured (CI) and Medicare supplement insured (MS) patients, from 2010 to 2015. After exclusions, there were 83,179 CI and 50,534 MS patients available for analysis. There were 12,876 new users of warfarin (W) and 10,892 new users of rivaroxaban (R) in CI patients, and 7,416 new users of W and 4,739 new users of R in MS patients. We asked the following questions: (1) What were the trends and predictive factors for anticoagulant use after elective THA? (2) What was the actual clinical efficacy: frequency of deep vein thrombosis (DVT) and pulmonary embolism (PE), and frequency of adverse events within 90 days with the two most commonly used oral agents, rivaroxaban and warfarin, from June 2011 to September 2015? Data was analyzed for each anticoagulant by odds ratios using logistic regression models with stabilized inverse probability treatment weighting. Results. There was a change in use of anticoagulants after R approval. Use of W decreased from approximately 50% each in 2010 in both insurance cohorts to 10% in CI patients and 30% in MS patients in 4th quarter 2015. The use of R increased from 0 to 33% in both cohorts from 2011 to 2015. In the multivariate analysis, in CI patients, females had lower odds of getting R, and patients in Western region had higher odds of getting R; in MS patients, increasing age had reduced odds of getting rivaroxiban, but Western region and surgery in 2015 had higher odds. Patients with capitated insurance plans and renal impairment had lower odds of R initiation, but a history of cardiovascular disease or hypertension had higher odds. In 90 days after THA, patients given R had significantly lower odds ratio of both DVT and PE in both CI patients (DVT: 1.54 with W, 0.54 with R; PE: 2.12 with W, 0.73 with R) and MS patients (DVT: 3.01 W, 1.73 R; PE: 4.09 W, 1.88 R). With logistic regression analysis, users of W had significantly higher odds ratio of both DVT (CI 2.63 and MS 1.78) and PE (CI 2.60 and MS 2.09) than R. There was no significant difference in rates of bleeding between W and R, but W had higher odds ratio than R of prosthetic joint infection (PJI) in both CI (1.574) and MS (1.790) cohorts. Conclusions. There has been an increase in VTE prophylaxis with R, and a decrease in both W and LMWH use after elective THA over four years. Patient factors, insurance type, and comorbidities were associated with this change. In actual clinical efficacy, R had lower odds ratio of both DVT and PE than W, and bleeding risks were similar. The association of W with an increased odds ratio of PJI compared to R requires further study


Bone & Joint Research
Vol. 10, Issue 12 | Pages 780 - 789
1 Dec 2021
Eslam Pour A Lazennec JY Patel KP Anjaria MP Beaulé PE Schwarzkopf R

Aims

In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement.

Methods

Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 11 - 11
1 Apr 2022
McGoldrick NP Cochran M Biniam B Bhullar R Beaulé PE Kim PR Gofton W Grammatopoulos G
Full Access

Short cementless femoral stems are increasingly popular as they allow for less dissection for insertion. Use of such stems with the anterior approach (AA) may be associated with considerable per-operative fracture risk. This study's primary aim was to evaluate whether patient-specific femoral- and pelvic- morphology and surgical technique, influence per-operative fracture risk. In doing so, we aimed to describe important anatomical thresholds alerting surgeons.

This is a single-centre, multi-surgeon retrospective, case-control matched study. Of 1145 primary THAs with a short, cementless stem inserted via the AA, 39 periprosthetic fractures (3.4%) were identified. These were matched for factors known to increase fracture risk (age, gender, BMI, side, Dorr classification, stem offset and indication for surgery) with 78 THAs that did not sustain a fracture. Radiographic analysis was performed using validated software to measure femoral- (canal flare index [CFI], morphological cortical index [MCI], calcar-calcar ratio [CCR]) and pelvic- (Ilium-ischial ratio [IIR], ilium overhang, and ASIS to greater trochanter distance) morphologies and surgical technique (% canal fill). Multivariate and Receiver-Operator Curve (ROC) analysis was performed to identify predictors of fracture.

Femoral factors that differed included CFI (3.7±0.6 vs 2.9±0.4, p<0.001) and CCR (0.5±0.1 vs 0.4±0.1, p=0.006). The mean IIR was higher in fracture cases (3.3±0.6 vs 3.0±0.5, p<0.001). % Canal fill was reduced in fracture cases (82.8±7.6 vs 86.7±6.8, p=0.007). Multivariate analysis and ROC analyses revealed a threshold CFI of 3.17 was predictive of fracture (sensitivity:84.6% / specificity:75.6%). Fracture risk was 29 times higher when patients had CFI>3.17 and II ratio>3 (OR:29.2 95%CI: 9.5–89.9, p<0.001).