Full Access

Hip

TOTAL HIP ARTHROPLASTY IN PATIENTS WITH DEVELOPMENTAL DYSPLASIA OF THE HIP: MODERN IMPLANTS IMPROVE RATES OF LOOSENING, WEAR, AND REVISION

International Hip Society (IHS) Closed Meeting, Berlin, Germany, 3–5 November 2021.



Abstract

Previous work has demonstrated increased implant failure in patients with DDH compared to osteoarthritis with historic methods and implants. This study examines outcomes of modern total hip arthroplasty (THA) techniques using uncemented fixation and cross-linked PE (XLPE) bearings for DDH.

A consecutive series of 879 patients with DDH who underwent primary THA with uncemented components and a XLPE bearing at a single institution between 1999 and 2016 were identified. Mean age at index arthroplasty was 51 years, with 78% females. Mean follow-up was 8 years.

5- and 10-year survivorships free of revision were 98.4% and 98%, respectively. 5- and 10-year survivorships free of reoperation for any reason were 97% and 96%, respectively. Survivorship free of acetabular revision for aseptic loosening was 99.8% at 5 and 10 years, while survivorships free of femoral revision for aseptic loosening were 99.7% at 5 years and 99.5% at 10 years. Survivorship free of reoperation for PE damage (2 liner fractures, 2 rim damage from impingement) was 99.3% at 10 years. There were no revisions for bearing surface wear or osteolysis. 13% of patients experienced complications (4% dislocation, 3.3% wound complication rate). Ten-year survivorship free of reoperation for instability was 99.2%. 8 patients developed infection requiring operation (<1%). 100 patients had a prior pelvic osteotomy (11%), which did not increase rate of complications (p=0.22) or reoperations (p=0.51).

Fixation with modern uncemented implants has dramatically reduced implant loosening in patients with DDH, and the use of XLPE bearings has markedly reduced revision for PE wear/osteolysis in this young patient population, leading to dramatically improved mid/long-term survivorship compared to historic series.