Aims. Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. Methods. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate. Results. Five patient-specific variables and seven implant-specific variables were evaluated for significant association with lower insert wear rate. Six were significant when controlling for other factors: greater patient age, female sex, shorter duration in vivo, polished tray, highly cross-linked PE (HXLPE), and constrained knee design. Conclusion. This study confirmed that
Aim of this study was the development of a dynamic FE-framework to identify worst-case size combinations and kinematics in a virtual wear simulator setup covering five daily activities and high, dynamic loads. Two cruciate sacrificing knee designs (D1 & D2) were tested physically on a wear-testing machine prior the model development using a high demanding, daily activity protocol (HDA) [1]. A simplified FE-setup was generated, reduced to the 3D geometries of the assembly whereas the representation of the mechanical wear simulator conditions and the load transmission was achieved by joint elements. Inertial and other time-related effects of the physical situation were compensated by a system of spring- and damper elements. Using a time-series signal optimization approach on the anterior-posterior translation and the internal-external rotation results for each activity, 38 variable parameters were varied in between pre-defined limits in a semiautomatic workflow. For each design, two consecutive cycles of a single activity were analysed and the results of the second cycle were used for the optimization. Based on the determined values, a single set of averaged parameter settings was identified that covers all activity cycles sufficiently. A total of 1010 dynamic analyses were carried out in order to find a sharable set of parameter values. In this study, an efficient simulation workflow for design evaluation was developed. Therefore, a HDA wear-testing machine was simplified to boundary conditions and stabilizing elements, using a single set of parameters for all activities. The calculated kinematics were in a comparable range to the machine output. Further applications of the method were found in systematic analyses of entire implant systems to achieve consistent kinematics over the size compatibility range in the design process of new implant systems.
Total knee arthroplasty is a well established treatment for degenerative joint disease with good clinical results. However, complications may occur due to a biological response to polyethylene wear particles, leading to osteolysis and aseptic loosening, as well as local and systemic hypersensitivity reactions triggered by metal ions and particles such as chromium, cobalt and molybdenum. Moreover, there is an increasing demand on the performance of these implants, as this treatment is also performed in heavier, younger and middle-aged adults who have a significant physical activity and higher life expectancy. The purpose of the following study was to compare the wear characteristics and performance of a zirconium nitride (ZrN) coated knee implant, designed for patients with metal ion hypersensitivity, against the clinically established cobalt-chromium (CoCr) version under a high demanding activities wear simulation. Medium size AS Columbus® DD (Aesculap AG, Tuttlingen, Germany) femoral and tibial components with a ZrN surface were tested in comparison with the cobalt-chromium version Columbus® DD. For both groups, ultra-high-molecular weight polyethylene (UHMWPE) gliding surfaces (size T3, high 10 mm) were used. Wear simulation was performed on a load controlled 4 station
Improving the wear resistance of polyethylene is considered paramount to improving knee implant longevity. Consequently, a range of polymer fabrication techniques have evolved in the quest for a highly wear resistant material. The objective of this study was to explore the wear performance of polyethylene as fabricated in a variety of ways. The following materials were prepared, sterilised, artificially aged, and machined into wear specimens (n = 4 for each material): Compression molded GUR1050 with three levels of cross-linking (120 kGy, 65 kGy, and 0 kGy irradiation – control); ram extruded GUR4150 high modulus material; compression molded GUR4150 low modulus material; and HSS/PolySolidur/Hoechst reference polyethylene. Using a custom designed joint articular wear simulator, samples were loaded for 2 million cycles at a frequency of 0.5 Hz under loads of 2.1 kN. Tests were stopped every 250 000 cycles; and wear surfaces were examined microscopically for surface damage (pitting, cracking, delamination). After 2 million loading cycles the following specimens were pitted and delaminated: 2 GUR1050 control samples, 3 GUR4150 high modulus specimens, and all 4 reference polyethylene specimens. Burnishing, but no pitting, was seen in all GUR1050 elevated cross-linked polyethylene specimens, and in all GUR4150 low modulus specimens. The materials tested in this study represent a broad range of fabrication techniques. Differences in starting resin cannot fully account for the differences in wear behaviour seen between the groups; as damage was not limited to one resin group. The cross-linked specimens were melt-annealed, prior to cross-linking. It is possible that this processing step, and not the actual cross-linking, contributed to the improved wear performance of this group. However, of most interest is the comparable wear performance of GUR1050 cross-linked polyethylene and GUR4150 low modulus polyethylene suggesting that cross-linking polyethylene is not the only route towards obtaining a polyethylene with superior wear characteristics.
Polyethylene (PE) wear is known as a limiting factor for total knee replacements (TKR). Thus, preclinical wear testing is an important tool to assess the suitability of new designs and new materials. However, standardized testing (e.g. according to ISO 14243) does not cover the individual situation in the patient. Consequentially, this study investigates the following two parameters: Testing-Frequency: Patients with TKR's show a humiliated walking frequency (down to 0,5Hz) compared to standardized testing (1Hz±0.1). In the first part of this study, the influence of a decreased test frequency on the PE wear behavior is investigated Interval of lubricant replacement: For in-vitro testing bovine serum is used as a substitute for the synovial fluid. Physiologically a continuous regeneration and removement of destructed components is taking place. In contrast, for simulator testing the bovine serum is typically changed completely every 500.000 cycles/steps. Therefore the goal of the second part of this study was to test if the serum replacing interval affects the PE wear behavior. Wear tests were conducted on an AMTI force controlled knee simulator. A cruciate substituting (ultracongruent) implant design (TC Plus, Smith & Nephew, Rotkreuz, Switzerland) was used. First, a reference wear study with a test frequency of 1Hz and a lubricant replacement interval (RI) of 500.000 cycles according to ISO 14243-1:2009 was carried out. Tests were run to a total of 5 million cycles. A second wear test was run with a reduced frequency of 0.5 Hz. The reduced frequency resulted in an extended testing period for the same number of cycles. To exclude an influence of the extended time period, the lubricant was changed, in the first half of testing every 500.000 cycles corresponding to 12 days (cycle depending (CD)), and in the second part every 250.000 cycles corresponding to 6 days (time depending (TD)). Tests were run to a total of 3 million cycles. A third test was run with a frequency of 1 Hz. For this test a reduced serum RI of 150.000 cycles was choosen. This test was run to a total of 1.500.000 cycles.Background
Material and Methods
Purpose: Posters of study: To produce a virtual reality model of prosthetic knee joint motion, giving detail regarding contact area and contact pressures during the process of walking. Method: Using serial CT scans and data regarding the material properties of bone, cartilage, ligaments and other soft tissue structures a computational “virtual reality model” of the knee has been constructed. This computational model which is a 3D dynamic representation of a human knee joint, may be programmed to replicate the standard gait pattern of the human knee. This study details the development of this model and its validation against the accepted Stenmore Test Rig for modelling of knee joint movement and
Wear of ultra-high molecular weight polyethylene (UHMWPE) is a major factor that affects longevity of the total joint replacement. In total hips, cross-linking of polyethylene acetabular cup has been shown highly effective in reducing wear both clinically and experimentally. In TKR, Schmidig 2000 showed 90% reduction of wear rate in 10 Mrad irradiated tibial inserts compared to 3 Mrad irradiated tibial inserts. Thus crosslinking should provide substantial improvement also in the wear resistance of UHMWPE tibial inserts. Our objective was to compare 3 Mrad UHMWPE with 10 Mrad HCLPE in same design but comparing standard kinematics to more severe mal-rotation kinematics. The latter offsets the tibial tray with 15 degrees internal rotation such the central tibial eminencies became involved in the wear process. Our hypothesis was that HCLPE would be more resistant than the standard UHMPE even in the mal-rotation model. The control material was Duracon 3 Mrad UHMWPE. HCLPE was radiation crosslinked to 10 Mrad mater
Polyethylene wear in total knee arthroplasty (TKA) is a complex and mutifactorial process. It is generally recognized that wear is directly related to a material wear factor, contact stress, and sliding distance. Conventional methods of predicting polyethylene wear in TKA mainly focus on peak contact stress or subsurface shear stress using finite element method analysis. By incorporating kinematics and contact stress, a new predictor for polyethylene wear in TKA (“Wear Index”) has been developed. The Wear Index was defined by multiplying deformation by femoro-tibial sliding velocity. The purpose of this study was to determine the predictive value of the Wear Index for polyethylene wear in TKA using both a numeric and an in vitro model. Four commercially available total knee prostheses were modeled for this study. Deformation and sliding velocity were calculated based on the three-dimensional geometry of the components and the gait kinematic inputs using Hertz’s formula. One specimen of each of the four types of total knee prostheses was mounted on a custom-designed knee simulator. Vertical loads and flexion-extension uni-axial motion were simulated using computer controlled servohydraulic actuators. The same gait kinematic inputs used in the theoretical study were used in the simulation test. After the simulations, the surface of the tibial insert was examined microscopically and macroscopically and compared with the theoretically generated Wear Index. This study showed a high correlation between the numeric model and the simulation. The depth of wear on the tibial insert correlated significantly with the Wear Index. Microscopic findings also demonstrated a good correlation between the Wear Index and observed wear patterns. Sliding velocity is an important factor for understanding wear in TKA. In conclusion, this study suggests that the Wear Index is a reliable predictor of polyethylene wear in TKA, as it incorporates both contact stress and kinematics in its calculation.
The introduction of mobile bearings for unicompartimental knee implants resulted in heightened interest in this implant design in the field of orthopaedics. This study aims to determine the effect of the mobile and fixed design concepts on the wear progression in unicompartmental knee implants using a knee simulator. An unicompartmental knee implant design, which is available in a fixed and mobile version, was tested using a knee simulator. For the wear test, the medial and lateral compartments were implemented in the simulator. To account for the physiologically higher medial load compared to the lateral compartment, a medially-biased load distribution was implemented. The wear test was performed force controlled according to ISO 14243. Wear was measured gravimetrically separately for the medial and lateral compartments. To evaluate implant kinematics, AP-translation and IE-rotation were measured during the simulation. Gravimetric wear was higher medially than laterally for both designs. The mean wear rate of the medial mobile compartment was found to be 10.70 mg/10E6 cycles, whereas a mean wear rate of 6.05 mg/10E6cycles was found for the medial compartment of the fixed design. Lateral wear rates, which were about 50% lower than medial wear rates, were found to be 5.38 mg/10E6 cycles in the mobile design and 3.23 mg/10E6 cycles in the lateral design. Wear of the mobile design was higher compared to the fixed design, both medially and laterally. Surprisingly the kinematics of both designs were very similar. A low AP-translation of 2.7 mm in the mobile and 2.4 mm in the fixed designs was documented. High IE-rotations of 6.5° and 6.7° for the mobile and the fixed design, respectively, were observed. In bicondylar bearing knee designs, reduced wear has been reported for mobile polyethylene inlays. This study showed that the wear behaviour of unicompartmental knee implants differs from bicondylar implants and that the introduction of the mobile concept may lead to increased wear.
The purpose of the study was to compare the mechanical properties, oxidation and wear resistance of a vitamin E blended and moderately crosslinked polyethylene for total knee arthroplasty (MXE) in comparison with clinically established polyethylene materials. The following polyethylene materials were tested: CPE (30 kGy e-beam sterilized), XLPE (75 kGy gamma crosslinked @ 100°C), ViXLPE (0.1 % vitamin E blended, 80 kGy e-beam crosslinked @ 100°C), and MXE (0.1 % vitamin E blended polyethylene, 30 kGy gamma sterilized). For the different tests, the polyethylene materials were either unaged or artificially aged for two or six weeks according to ASTM F2003-02. The oxidation index was measured based on ASTM F2102 at a 1 mm depth. Small punch testing was performed based on ASTM F2977. Mechanical properties were measured on unaged materials according to ASTM D638. Wear simulation was performed on a load controlled 3 + 1 station
Aims. The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components. Methods. Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a
Majority of ultra-high molecular weight polyethylene (UHMWPE) medical devices used in total joint arthroplasty are crosslinked using gamma radiation to improve wear resistance. Alternative methods of crosslinking are urgently needed to replace gamma radiation due to rapid decline in its supply. Peroxide crosslinking is a candidate method with widespread industrial applications. Oxidative stability and biocompatibility, which are critical requirements for medical device applications, can be achieved using vitamin-E as an additive and by removing peroxide by-products through high temperature melting, respectively. We investigated compression molded UHMWPE/vitamin-E/di-cumyl peroxide blends followed by high-temperature melting in inert gas as a material candidate for tibial
Background. Wear and fatigue damage to polyethylene components remain major factors leading to complications after total knee and unicompartmental arthroplasty. A number of wear simulations have been reported using mechanical test equipment as well as computer models. Computational models of
Introduction. Wear of polyethylene tibial inserts has been cited as being responsible for up to 25% of revision surgeries, imposing a very significant cost burden on the health care system and increasing patient risk. Accurate measurement of material loss from retrieved knee bearings presents difficult challenges because gravimetric methods are not useful with retrievals and unworn reference dimensions are often unavailable. Geometry and the local anatomy restrict in vivo radiographic wear analysis, and no large-scale analyses have illuminated long-term comparative wear rates and their dependence on design and patient factors. Our study of a large retrieval archive of knee inserts indicates that abrasive/adhesive wear of polyethylene inserts, both on the articular surface and on the backside of modular knees is an important contributor to wear, generation of debris and integrity of locking geometry. The objective of the current study is to quantify wear performance of tibial inserts in a large archive of retrieved knees of different designs. By assessing wear in a large and diverse series, the goal is to discern the effect on wear performance of a number of different factors: patient factors that might help guide treatment, knee design factors and bearing material factors that may inform a surgeon's choice from among the array of arthroplasty device options. Methods. An IRB approved retrieval database was queried for TKA designs implanted between 1997 and 2017. 1385 devices from 5 TKA designs were evaluated. Damage was ranked according to Hood's method, oxidation was determined through FTIR, and wear was determined through direct measurement of retrieved inserts using a previously established protocol. Design features (e.g. materials, conformity, locking mechanisms, stabilization, etc.) and patient demographics (e.g. age, weight, BMI, etc.) were cataloged. Multivariate analysis was performed to isolate factors contributing to wear, oxidation, and damage. Results. Wear and oxidation were both found to scale with time in vivo in conventional and crosslinked polyethylene. Wear rate was also found to scale with time in vivo, but was not found to be a function of oxidation. Regression shows patient age and female sex to correlate negatively with wear rate. Polished trays, crosslinked polyethylene, and constrained knee designs are all correlated with decreased wear rates. Discussion. While this study indicates that loosening and infection are predominant causes for TKA revision, wear related failure remains common. We believe this to be the largest existing comparative study of modern TKA wear rates. Insert wear is shown to correlate with several patient factors. Wear performance also varies significantly between knee designs, polyethylene material choice and tray surface finish. When compared to a historical standard for
Introduction. Orthopedic implants are subject to wear and release ultra-high molecular weight polyethylene (UHMWPE) debris. Analysis of UHMWPE wear particles is critical in determining the safety and effectiveness of novel orthopedic implants. Complete digestion of periprosthetic tissue and wear fluid is necessary to ensure accurate morphological and quantitative particle analysis. Acid digestion methods are more effective than enzymatic and base digestion approaches [Baxter+ 2009]. However, optimal digestion times, quantity, and type of acid are unclear for particle isolation. In addition, imaging and analysis techniques are critical to ensure accurate reporting of particle characteristics. Here, we 1) compared the efficacy of three acid-based digestion methods in isolating particles from a) bovine serum and b) animal/human tissue, and 2) analyzed the effects of imaging location on particle quantity/morphology results. Methods. 1a) UHMWPE (GUR 150) particles were generated by Mode I
Background. Wear simulation in total knee arthroplasty (TKA) is currently based on the most frequent activity – level walking. A decade ago multi-station
Background. The anatomy of the human knee is very different than the tibiofemoral surface geometry of most modern total knee replacements (TKRs). Many TKRs are designed with simplified articulating surfaces that are mediolaterally symmetrical, resulting in non-natural patterns of motion of the knee joint [1]. Recent orthopaedic trends portray a shift away from basic tibiofemoral geometry towards designs which better replicate natural knee kinematics by adding constraint to the medial condyle and decreasing constraint on the lateral condyle [2]. A recent design concept has paired this theory with the concept of guided kinematic motion throughout the flexion range [3]. The purpose of this study was to validate the kinematic pattern of motion of the surface-guided knee concept through in vitro, mechanical testing. Methods. Prototypes of the surface-guided knee implant were manufactured using cobalt chromium alloy (femoral component) and ultra-high molecular weight polyethylene (tibial component). The prototypes were installed in a force-controlled
Background. Additive manufacturing (AM) has created many new avenues for material and manufacturing innovation. In orthopaedics, metal additive manufacturing is now widely used for production of joint replacements, spinal fusion devices, and cranial maxillofacial reconstruction. Plastic additive manufacturing on the other hand, has mostly been utilized for pre-surgical planning models and surgical cutting guides. The addition of pharmaceuticals to additively manufactured plastics is novel, particularly when done at the raw material level. The purpose of this study was to prove the concept of antibiotic elution from additively manufactured polymeric articles and demonstrate feasibility of application in orthopaedics. Methods. Using patented processes, three heat-stable antibiotics commonly used in orthopaedics were combined with six biocompatible polymers (2 bioresorbable) into filament and powder base materials for fused deposition modeling (FDM) and selective laser sintering (SLS) AM processes. Raw materials of 1%, 2%, and 5% antibiotic concentrations (by mass) were produced as well as a blend of all three antibiotics each at 1% concentration. Thin disks of 25 mm diameter were manufactured of each polymer with each antibiotic at all concentrations. Disks were applied to the center of circular petri dishes inoculated with a bacterium as per a standard zone of inhibition, or Kirby-Bauer disk diffusion tests. After 72 hours incubation, the zone of inhibited bacterial growth was measured. Periprosthetic joint infection (PJI) of the knee was selected as the proof-of-concept application in orthopaedics. A series of tibial inserts mimicking those of a common TKR system were manufactured via SLS using a bioresorbable base material (Figure 1). Three prototype inserts were tested on a
Introduction. Ideally, a patient receiving a unicondylar knee replacement will have fully functional anterior and posterior cruciate ligaments. When at least one of the cruciate ligaments is not fully functional, femoral and tibial implant contact position can potentially increase along the anterior-posterior (AP) axis. Where unicondylar implant wear testing typically uses AP resistance assuming fully functional cruciate ligaments, the authors used reduced AP resistance intended to simulate deficient cruciate ligaments. Methods. Optetrak Logic® Uni (Exactech Inc, Gainesville, FL USA) unicondylar test specimens featuring an all-UHMWPE tibial component and a cobalt chromium femoral component were used in this study. The system has a semi-constrained articular geometry. Testing was conducted at an independent testing facility (EndoLab GMBH, Thansau, Rosenheim, Germany). A four-station knee simulator was used (EndoLab knee simulator) with two unicondylar knee implants per station, giving a total of eight test specimens. Two different tibial fixation designs (keeled and peg) with identical articulating surfaces were tested. Tibial test specimens were 6 mm in thickness. Unloaded soak controls were stored in distilled water at 37°C. The test was conducted according to ISO 14243–1: 2009 [1]. Test specimens were immersed in calf serum (PAA GmBH, Cölbe, LOT B00111-5126) with a protein content of 20 g/l. Custom polyurethane molds allowed for individual component measurement. Per the ISO 14243-1, a 7% medial offset was incorporated into the set-up. The unicondylar knee implants were set at neutral position in extension. Tibial rotational restraint was 0.36 Nm/° and zero when the test specimen was within ± 6° of the reference position. This test was conducted with an AP resistance of 9.3N/mm to maximize AP displacement and simulate deficient cruciate ligaments. Typical unicondylar
INTRODUCTION. Highly cross-linked polyethylene (XLPE) inserts have shown significant improvements in decreasing wear and osteolysis in total hip arthroplasty [1]. In contrast to that, XLPE has not shown to reduce wear or aseptic loosening in total knee arthroplasty [2,3,4]. One major limitation is that current wear testing in vitro is mainly focused on abrasive-adhesive wear due to level walking test conditions and does not reflect “delamination” as an essential clinical failure mode [5,6]. The objective of our study was to use a highly demanding daily activities wear simulation to evaluate the delamination risk of polyethylene materials with and without vitamin E stabilisation. MATERIALS & METHODS. A cruciate retaining fixed bearing TKA design (Columbus® CR) with artificially aged polyethylene knee bearings (irradiation 30 & 50 kGy) blended with and without 0.1% vitamin E was used under medio-lateral load distribution and soft tissue restrain simulation. Daily patient activities measured by Bergmann et al. [7] in vivo, were applied for 5 million