Prosthetic Hip dislocations remain one of the most common major complications after total hip arthroplasty procedures, which has led to much debate and refinement geared to the optimization of implant and bearing options, surgical approaches, and technique. The implementation of larger femoral heads has afforded patients a larger excursion distance and primary arc range motion before impingement, leading to lowered risk of hip dislocation. However, studies suggest that while the above remains true, the use of larger heads may contribute to increased volumetric wear, trunnion related corrosion, and an overall higher prevalence of loosening, pain, and patient dissatisfaction, which may require revision hip arthroplasty. More novel designs such as the dual mobility hip have been introduced into the United States to optimize stability and range of motion, while possibly lowering the frictional torque and modes of failure associated with larger fixed bearing articulations. Therefore, the aim of this study is to compare the effect of bearing design and anatomic angles on frictional torque using a clinically relevant model8. Two bearing designs at various anatomical angles were used; a fixed and a mobile acetabular component at anatomical angles of 0°,20°,35°,50°, and 65°. The fixed design consisted of a 28/56mm inner diameter/outer diameter acetabular hip insert that articulated against a 28mm CoCr femoral head (n=6). The mobile design consisted of a 28mm CoCr femoral head into a 28/56mm inner diameter/outer diameter polyethylene insert that articulates against a 48mm metal shell (n=6). The study was conducted dynamically following a physiologically relevant frictional model8. A statistical difference was found only between the anatomical angles comparison of 0vs65 degrees in the mobile bearing design. In the fixed bearing design, a statistical difference was found between the anatomical angles comparison of 20vs35 degrees, 20vs50 degrees, and 35vs65 degrees. No anatomical angle effect on frictional torque between each respective angle or bearing design was identified.
Introduction.
The frictional torque of ceramic-on-ceramic bearings tended to increase with increasing the bearings size (32, 48, 56mm). However, the frictional torque was significantly lower than that measured on metal-on-metal bearings under well positioned and well lubricated conditions. Larger head size in total hip replacement theoretically provides increased range of motion and enhanced stability. However, there are potential clinical concerns regarding increased frictional torques with large diameter metal-on-metal bearings causing loosening of the acetabular cups and corrosion at the taper. The aim of this study was to determine the frictional torques of large diameter BIOLOX® delta ceramic-on-ceramic bearings.Summary Statement
Introduction
With the demand for total hip arthroplasty continuing it is essential that the longest follow-up cases remain under continuous review. Fifty one of the primary Charnley low-frictional torques arthroplasties have passed 30 years follow-up: mean 31 years (30–36): 42 are women and 9 men. Their mean age at surgery was 47 years (24–64). At the latest follow-up two have had a late deep infection, three a dislocation, nine a radiologically loose cup, two a loose stem and one had had a fracture of the shaft of the femur, and one a fractured stem, with only the fractured stem coming to revision. Clinical results remain satisfactory. Wear and loosening of the cup remains the long term problem. Further improvement and even longer follow-up and successful results of the Charnley LFA will come from materials which offer the greatest resistance to wear.
High failure rates with large diameter, metal on metal hip replacements have highlighted a potential issue with the head/stem taper junction as one of the significant sources of metal ion release. Postulated reasons as to why this may be such a problem with large head metal on metal hip replacements is due to the increased torque achieved by the larger head size. This may be responsible for applying greater micromotion between the head and stem taper and consequently greater amounts of fretting corrosion. The aim of this study was to perform short term in vitro electrochemical tests to assess the effect of increasing head diameter and torque on the fretting corrosion susceptibility of the head/stem taper interface and to investigate its effect on different material combinations. 36 mm Cobalt Chrome (CoCr) femoral heads were coupled with either a CoCr or Titanium (Ti) stem with 12/14 tapers, all with a smooth surface finish. Increasing perpendicular horizontal offsets in the sagittal plane created incremental increases in torque. Offset increments of 0 mm, 5.4 mm and 7.5 mm were selected (Figure 1) to simulate the torque force equivalent to 9 Nm, 12 Nm and 17 Nm. An inverted hip replacement setup was used (ASTM F1875-98) (Figure 2). Components were statically loaded at 0 kN and 2.3 kN prior to sinusoidal cyclic loading and electrochemical testing. Mean & fretting currents were calculated every 50 cycles up to a maximum of 1000 cycles of sinusoidal cyclic loading at 3 Hz along with the Overall Mean Current (OMC), Overall Mean Fretting Current (OMFC) and Overall Current change (OCC).Introduction:
Methods:
A prospective study of Charnley low friction arthroplasty in patients under the age of 51 at the time of surgery. 1092 patients, 668 women and 424 men, mean age 41 years (12–51) at the time of surgery had 1434 Charnley low-friction arthroplasties carried out between November 1962 and December 1990. At mean follow-up 14.2 years (1 – 32), 742 patients (973 hips) are still attending and patients’ satisfaction with the outcome is 96.2%. Survivorship was 95.24% at 10 years and 58.67% at 27 years. Patients who had had previous hip surgery had revision rate 24.8% compared with 14.1%. Patients with rheumatoid arthritis had fewer revisions than those with developmental hip dysplasia or primary osteoarthritis. Large 43 mm diameter cup gave lower revisions for aseptic cup loosening as compared with 40mm cup. Presence and preservation of subchondral plate, rim support compared to medialization of the cup, use of acetabular cement pressurizer and the reduced diameter neck (10mm) all made a contribution to reducing the incidence of revisions for aseptic cup loosening. Closing the medullary canal with bone block reduced the incidence of aseptic stem loosening. Use of the brace reamers did not affect the outcome and there appears to have been no advantage with the flanged stem. The long-term problem was the increasing incidence of revisions for aseptic cup loosening. This was exponentially related to the depth of cup penetration by the head of the femoral component. The long term clinical results of the Charnley LFA remain excellent even in young patients. Rim support of the cup, preservation of the subchondral bone of the acetabulum, cup flange and pressurising of the acetabular cement all make a significant contribution. Distal closure of the medullary canal and central position of the stem are of benefit, but reaming the medullary canal to cortex must be avoided. The long-term problem has been highlighted again as: wear and cup loosening.
Burroughs et al showed that frictional torque increases with increasing head size in a simple in vitro model and showed differences in frictional torque with different polyethylene materials [1]. Therefore, the purpose of this study was to evaluate the influence of bearing material and bearing size on the frictional torque of hip bearings utilizing a more physiologically relevant hip simulator model. A total of four hip bearing combinations (Crosslinked PE/CoCr, Conventional PE/CoCr, Crosslinked PE/Delta and Alumina /Alumina) with various bearing sizes were evaluated. The sizes tested in this study range from 22 mm to 44 mm; it is important to note that the study only evaluated bearing combinations (size and material combination) currently commercially available. A total of three samples per bearing combination were tested, with the exception of conventional PE, which included a total of 4 samples. A MTS hip joint simulator was used. All components were oriented anatomically with the femoral head mounted below on a rotating angled block which imparts a 23° biaxial rocking motion onto the head. Loading was held constant at each load level (500N, 1000N, 1500N, 2000N, 2450N) for at least two rotational cycles while all 3 axes of load and all 3 axes of moments were measured at 10 khz. Fresh Alpha Calf Fraction serum was utilized as a lubricant. Results show that frictional torque increases with the increase of head size regardless of head material for all polyethylene combinations (p > 0.05), as shown in Figure 1 and 2. However, results showed no change in frictional behavior for the Alumina/Alumina combination regardless of the bearing size. The results of this test did not show any significant difference between crosslinked PE and conventional PE materials for sizes 28 mm and 32 mm when paired against a CoCr head (p > 0.05) (Figure 3). The Alumina/Alumina bearing combination had the lowest frictional torque among all the bearing material combinations evaluated in this study. This data suggests that there is a strong correlation between increased head size and increased frictional torque (R2 = 0.6906, 0.8847) for the polyethylenes evaluated here regardless of head material. No correlation can be concluded for the Alumina /Alumina bearing combination (R2 = 0.0217). The combination of Alumina /Alumina seems to have the most favorable frictional properties. This data also suggests no effect on frictional properties regardless of the polyethylene material (crosslinked and conventional) for sizes 28 mm and 32 mm. The frictional torque values recorded in this study are different than those published by Burroughs et al [1]. This difference may be attributed to the testing methodology. The current study utilizes a hip simulator, which closely mimics the natural joint providing a more physiologically relevant model whereas the Burroughs et al study utilizes a single axis machine. It is important to understand that frictional behavior in hip bearings may be highly sensitive to bearing clearance, cup thickness, and stiffness, which may outweight the effect of head diameter. Further evaluation is necessary to isolate and investigate those parameters.
Recent studies on large diameter femoral head hip replacements have implicated the modular taper junction as one of the significant sources of wear and corrosion products and this has been attributed to increased torque and bending on the taper interface. The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper junction and to investigate whether different material combinations also had an effect. We examined 1) Cobalt Chromium (CoCr) heads on CoCr stems 2) CoCr heads on Titanium alloy (Ti) stems and 3) Ceramic heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the femoral stem in the anterior posterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm where the force generated was equivalent to 0Nm, 9Nm, 14Nm and 18Nm. In Test 2 we investigated the effect of increasing bending moment by offsetting the application of axial load from the midline in the medial-lateral (ML). Offset increments equivalent to +0, +7 and +14 heads were used. For each test we used n=3 for each different material combination.Introduction
Patients/Materials & Methods
Cross linked polyethylene (XLPE) has gained popularity as a bearing surface of choice for younger patients despite only medium term results being available for wear rates. Concern remains regarding the long-term stability and durability of these materials. In order to address these issues we present the longest radiological and clinical follow-up of XLPE. Since 1986, we have prospectively studied a group of 17 patients (19 hips) that underwent a cemented Charnley low friction arthroplasty using a combination of 22.225mm alumina ceramic femoral head, a modified Charnley flanged stem and a chemically cross-linked polyethylene cup. We now report the 28 year clinical and radiological results.Introduction
Patients/Materials & Methods
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
Recent issues related to trunionosis have created a new paradigm in choosing femoral head material in total hip arthroplasty. While many consider highly-crosslinked polyethylene (XLPE) to be the gold standard currently in acetabular liner bearing surface, the debate remains whether metal or ceramic heads are best paired with XLPE. Wear characteristics are similar within an order of magnitude when comparing cobalt chrome femoral heads with ceramic when used in combination with XLPE. Therefore, discernable differences between the two femoral head materials with respect to outcomes would be the result of other biomechanical factors. Notably the fretting and corrosion of metal heads at the modular taper of femoral components is a serious concern and represents a significant deterrent when considering this material. The fretting corrosion that occurs with metal femoral heads has recently been well documented in multiple reports, and can be associated with adverse local tissue reactions necessitating revision hip arthroplasty.
INTRODUCTION. Modular metal-on-metal hip implants show increased revision rates due to fretting and corrosion at the interface. High frictional torque potentially causes such effects at the head-taper interface, especially for large hip bearings. The aim of this study was to investigate fretting and corrosion of sleeved ceramic heads for large ceramic-on-ceramic (CoC) bearings. METHODS. The investigated system consists of a ceramic head (ISO 6474-2; BIOLOX® Option), a metal sleeve (Ti-6Al-4V, ISO 5832-3) and different metal stem tapers (Ti-6Al-4V, ISO 5832-3; stainless steel, ISO 5832-1; CoCrMo, ISO 5832-12). Three different test methods were used to assess corrosion behaviour and connection strength of head-sleeve-taper interfaces:
. –. Fretting corrosion acc. to ASTM F1- Corrosion under in-vivo relevant loads. –.
An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the taper junctions. The relationship
between taper wear rates and a number of variables, including bearing
diameter and orientation of the acetabular component, was investigated.Objectives
Methods