CHARNLEY LOW FRICTIONAL TORQUE ARTHROPLASTY IN THE YOUNG PATIENT. FACTORS AFFECTING OUTCOME.



Abstract

A prospective study of Charnley low friction arthroplasty in patients under the age of 51 at the time of surgery.

1092 patients, 668 women and 424 men, mean age 41 years (12–51) at the time of surgery had 1434 Charnley low-friction arthroplasties carried out between November 1962 and December 1990. At mean follow-up 14.2 years (1 – 32), 742 patients (973 hips) are still attending and patients’ satisfaction with the outcome is 96.2%. Survivorship was 95.24% at 10 years and 58.67% at 27 years.

Patients who had had previous hip surgery had revision rate 24.8% compared with 14.1%. Patients with rheumatoid arthritis had fewer revisions than those with developmental hip dysplasia or primary osteoarthritis. Large 43 mm diameter cup gave lower revisions for aseptic cup loosening as compared with 40mm cup. Presence and preservation of subchondral plate, rim support compared to medialization of the cup, use of acetabular cement pressurizer and the reduced diameter neck (10mm) all made a contribution to reducing the incidence of revisions for aseptic cup loosening. Closing the medullary canal with bone block reduced the incidence of aseptic stem loosening. Use of the brace reamers did not affect the outcome and there appears to have been no advantage with the flanged stem.

The long-term problem was the increasing incidence of revisions for aseptic cup loosening. This was exponentially related to the depth of cup penetration by the head of the femoral component.

The long term clinical results of the Charnley LFA remain excellent even in young patients. Rim support of the cup, preservation of the subchondral bone of the acetabulum, cup flange and pressurising of the acetabular cement all make a significant contribution. Distal closure of the medullary canal and central position of the stem are of benefit, but reaming the medullary canal to cortex must be avoided. The long-term problem has been highlighted again as: wear and cup loosening.

The abstracts were prepared by Mr Simon Donell. Correspondence should be addressed to him at the Department of Orthopaedics, Norfolk & Norwich Hospital, Level 4, Centre Block, Colney Lane, Norwich NR4 7UY, United Kingdom.