Advertisement for orthosearch.org.uk
Results 1 - 20 of 31
Results per page:
Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 43 - 43
19 Aug 2024
Lustig S Batailler C
Full Access

The stem design in total hip arthroplasty (THA) is constantly evolving. The impact of the collar on the risk of periprosthetic fracture remains controversial. This study aimed to determine whether adding a collar to the femoral stem impacts the early periprosthetic fracture risk within 90 days of surgery. This retrospective study included 1,623 patients who underwent primary total hip arthroplasty in a single orthopedic department between January 2010 and December 2020. The inclusion criteria were uncemented stem with or without a collar, in a primary intention, without previous hip surgery with a similar “corail like” design. The assessed data were demographic characteristics (age, gender, number of obese (BMI > 30)), single or dual mobility, the surgical approach, the early complication, particularly the periprosthetic femoral fractures. Of the 1,623 patients, 1,380 received a collared stem (85%), and 243 received a collarless stem (15%). A multivariate analysis was performed to determine the collar's effect on the risk of early periprosthetic fracture (<90 days). Nine (0.55%) early periprosthetic fractures were identified in the whole cohort. There were four fractures (1.65%) in the collarless stem group and five fractures (0.36%) in the collared stem group (p=0,005). One patient required revision surgery in the collared stem group, while two patients required revision surgery in the collarless stem group. The multinomial logistic regression model indicated a statistically significant lower (p<0.05) risk of early periprosthetic fracture within 90 days of surgery in the collared stems group. No other risk factor for early periprosthetic fractures has been identified. Using collared stems in cementless THA protects early periprosthetic femoral fractures within 90 days of surgery


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 24 - 24
1 Apr 2022
Partridge T Osborne S Marsh M Holland J
Full Access

We present a consecutive case cohort of the first 100 Birmingham Hip Resurfacing (BHR)'s in 90 patients with a minimum follow up of 20 years. All procedures were performed by a single surgeon having commenced the study in 1998. The original cohort included 68 males with 75 hips (7 bilateral) and 22 females with 25 hips (3 bilateral). The mean age at index procedure was 52. Patients were recalled to review in clinic as per Medicines and Healthcare products Regulatory Agency guidelines with x-rays, metal ions and Harris Hip Scores recorded. After a minimum of 20 year follow-up review the known overall revision rate is 11%. 11 have died and 7 have been lost to review. In males the known revision rate is 4/75 (5%), 3 of which were due to early fractures and 1 revision for infection at 5 years. The known revision rate in females is 7/25 (28%) of which 6 were due to adverse reactions to metal debris; 3 at 7–8 yrs, 1 at 15 yrs, 2 at 18 years, and 1 at 2 months for avascular necrosis. The mean 20 year metal ions results were cobalt 36.6 nmol/l (range 7.4–232.1) and chromium 32.1 nmol/l (range 6–120.8). The mean Harris hip score was 88.5 (range 38.6–99.8). This case series with 20 years follow-up demonstrates excellent long term survival of the BHRs further to registry BHR data and similar to THR for same age recipients. Applying today's strict criteria for patient selection would likely improve survival rate further. BHR has not demonstrated the catastrophic revision rates associated with some metal-on-metal resurfacing and remains a safe option for those who meet the criteria


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 33 - 37
1 Jul 2021
Ennin KA Elsharkawy KA Dasgupta S Emerson RH

Aims. To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications?. Methods. A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group. Results. The cementless group had a higher rate of femoral complications (8 vs 0; p = 0.011). There were two loose stems and six fractures, all requiring revision. Fractures occurred a mean 14.5 days (2 to 31) postoperatively and loosening at 189 days and 422 days postoperatively. Femoral cementing can be done using the DA approach safely and reduces the number of complications compared with a contemporary cementless series. Conclusion. A higher rate of early fractures and loosening occurred with cementless stems. This was not observed in our cemented stem cohort and cementing was safely accomplished through the DA approach. The modern femoral cementing process with the DA approach does not add to surgical complexity or time, has fewer early complications, and is a safer option for older patients compared to cementless femoral arthroplasties. Cite this article: Bone Joint J 2021;103-B(7 Supple B):33–37


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_4 | Pages 40 - 40
1 Apr 2022
Holleyman R Khan S Charlett A Inman D Johansen A Brown C Barnard S Fox S Baker P Deehan D Burton P Gregson C
Full Access

Hip fracture principally affects the frailest in society, many of whom are care dependent, and are disproportionately at risk of contracting COVID-19. We examined the impact of COVID-19 infection on hip fracture mortality in England. We conducted a cohort study of patients with hip fracture recorded in the National Hip Fracture Database between 1st February 2019 and 31st October 2020, in England. Data were linked to Hospital Episode Statistics to quantify patient characteristics and comorbidities, Office for National Statistics mortality data, and Public Health England's SARS-CoV-2 testing results. Multivariable Cox regression examined determinants of 90-day mortality. Excess mortality attributable to COVID-19 was quantified using Quasi-Poisson models. Analysis of 102,900 hip fractures (42,630 occurring during the pandemic) revealed that amongst those with COVID-19 infection at presentation (n=1,120) there was a doubling of 90-day mortality; hazard ratio (HR) 2.05 (95%CI 1.86–2.26), while for infections arising between 8–30 days after presentation (n=1,644) the figure was even higher at 2.52 (2.32–2.73). Malnutrition [1.44 (1.19–1.75)] and non-operative treatment [2.89 (2.16–3.86)] were the only modifiable risk factors for death in COVID-19 positive patients. Patients with previous COVID-19 initially had better survival compared to those who contracted COVID-19 around the time of their hip fracture; however, survival rapidly declined and by 365 days the combination of hip fracture and COVID-19 infection was associated with a 50% mortality rate. Between 1st January and 30th June 2020, 1,273 (99.7%CI 1,077–1,465) excess deaths occurred within 90 days of hip fracture, representing an excess mortality of 23% (20%–26%), with most deaths occurring within 30 days. COVID-19 infection more than doubled early hip fracture mortality; the first 30-days after injury were most critical, suggesting that targeted interventions in this period may have most benefit in improving survival


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 779 - 786
1 Jul 2019
Lamb JN Baetz J Messer-Hannemann P Adekanmbi I van Duren BH Redmond A West RM Morlock MM Pandit HG

Aims. The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model. Materials and Methods. A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an in vitro model using paired fresh frozen cadaveric femora. Results. The prevalence of early revision for periprosthetic fracture was 0.34% (1180/337 647) and 44.0% (520/1180) occurred within 90 days of surgery. Implant risk factors included: collarless stem, non-grit-blasted finish, and triple-tapered design. In the in vitro model, a medial calcar collar consistently improved the stability and resistance to fracture. Conclusion. Analysis of features of stem design in registry data is a useful method of identifying implant characteristics that affect the risk of early periprosthetic fracture around a cementless femoral stem. A collar on the calcar reduced the risk of an early periprosthetic fracture and this was confirmed by biomechanical testing. This approach may be useful in the analysis of other uncommon modes of failure after THA. Cite this article: Bone Joint J 2019;101-B:779–786


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 254 - 260
1 Mar 2023
Bukowski BR Sandhu KP Bernatz JT Pickhardt PJ Binkley N Anderson PA Illgen R

Aims

Osteoporosis can determine surgical strategy for total hip arthroplasty (THA), and perioperative fracture risk. The aims of this study were to use hip CT to measure femoral bone mineral density (BMD) using CT X-ray absorptiometry (CTXA), determine if systematic evaluation of preoperative femoral BMD with CTXA would improve identification of osteopenia and osteoporosis compared with available preoperative dual-energy X-ray absorptiometry (DXA) analysis, and determine if improved recognition of low BMD would affect the use of cemented stem fixation.

Methods

Retrospective chart review of a single-surgeon database identified 78 patients with CTXA performed prior to robotic-assisted THA (raTHA) (Group 1). Group 1 was age- and sex-matched to 78 raTHAs that had a preoperative hip CT but did not have CTXA analysis (Group 2). Clinical demographics, femoral fixation method, CTXA, and DXA data were recorded. Demographic data were similar for both groups.


The Bone & Joint Journal
Vol. 104-B, Issue 9 | Pages 1039 - 1046
1 Sep 2022
Özdemir E Kuijpers MFL Visser J Schreurs BW Rijnen WHC

Aims

The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique.

Methods

We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 451 - 457
1 Apr 2017
Gromov K Bersang A Nielsen CS Kallemose T Husted H Troelsen A

Aims. The aim of this study was to identify patient- and surgery-related risk factors for sustaining an early periprosthetic fracture following primary total hip arthroplasty (THA) performed using a double-tapered cementless femoral component (Bi-Metric femoral stem; Biomet Inc., Warsaw, Indiana). Patients and Methods. A total of 1598 consecutive hips, in 1441 patients receiving primary THA between January 2010 and June 2015, were retrospectively identified. Level of pre-operative osteoarthritis, femoral Dorr type and cortical index were recorded. Varus/valgus placement of the stem and canal fill ratio were recorded post-operatively. Periprosthetic fractures were identified and classified according to the Vancouver classification. Regression analysis was performed to identify risk factors for early periprosthetic fracture. Results. The mean follow-up was 713 days (1 to 2058). A total of 48 periprosthetic fractures (3.0%) were identified during the follow-up and median time until fracture was 16 days, (interquartile range 10 to 31.5). Patients with femoral Dorr type C had a 5.2 times increased risk of post-operative periprosthetic fracture compared with type B, while female patients had a near significant two times increased risk over time for post-operative fracture. Conclusion. Dorr type C is an independent risk factor for early periprosthetic fracture, following THA using a double tapered cementless stem such as the Bi-Metric. Surgeons should take bone morphology into consideration when planning for primary THA and consider using cemented femoral components in female patients with poor bone quality. Cite this article: Bone Joint J 2017;99-B:451–7


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 46 - 52
1 Jul 2021
McGoldrick NP Fischman D Nicol GM Kreviazuk C Grammatopoulos G Beaulé PE

Aims

The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach.

Methods

A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.


Bone & Joint Open
Vol. 2, Issue 5 | Pages 293 - 300
3 May 2021
Lewis PM Khan FJ Feathers JR Lewis MH Morris KH Waddell JP

Aims

“Get It Right First Time” (GIRFT) and NHS England’s Best Practice Tariff (BPT) have published directives advising that patients over the ages of 65 (GIRFT) and 69 years (BPT) receiving total hip arthroplasty (THA) should receive cemented implants and have brought in financial penalties if this policy is not observed. Despite this, worldwide, uncemented component use has increased, a situation described as a ‘paradox’. GIRFT and BPT do, however, acknowledge more data are required to support this edict with current policies based on the National Joint Registry survivorship and implant costs.

Methods

This study compares THA outcomes for over 1,000 uncemented Corail/Pinnacle constructs used in all age groups/patient frailty, under one surgeon, with identical pre- and postoperative pathways over a nine-year period with mean follow-up of five years and two months (range: nine months to nine years and nine months). Implant information, survivorship, and regular postoperative Oxford Hip Scores (OHS) were collected and two comparisons undertaken: a comparison of those aged over 65 years with those 65 and under and a second comparison of those aged 70 years and over with those aged under 70.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 13 - 13
1 Oct 2018
Greco NJ Lombardi AV Morris MJ Berend KR
Full Access

Background. Despite growing interest in direct anterior approach total hip arthroplasty, perioperative femoral fracture and early aseptic loosening are increasingly recognized complications. Previous research has documented the role of surgeon experience in association with these femoral complications. The purpose of this study was to explore the relationship between femoral component design and early periprosthetic femoral complications. Methods. This was an extension of previous work with an updated patient cohort of 5090 consecutive direct anterior primary total hip arthroplasties at a single institution with a single-taper, wedge femoral stem with 4 variants involving length and geometry: Group 1) full length, standard profile; Group 2) full length, reduced distal profile; Group 3) short length, standard profile; and Group 4) short length, reduced distal profile. Records were reviewed retrospectively for the incidence of early periprosthetic fracture or aseptic loosening and analyzed with regard to patient demographics and femoral stem type. Results. There were 42 (0.83%) periprosthetic femur complications observed in the early postoperative period. Increased age (p<0.001) and female gender (p=0.023) were significantly associated with incidence of femoral complications in univariate analysis, while age maintained this significant relationship in multivariate analysis (P<0.001). There was a trend toward increased complication rate in patients receiving a short stem with full profile taper (1.27%, p=0.0539). Conclusion. Despite an overall low rate of femoral complications after direct anterior total hip arthroplasty, the risk is increased in elderly patients and females. Furthermore, femoral stem design may portend an elevated risk of these complications


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 479 - 485
1 Mar 2021
Nugent M Young SW Frampton CM Hooper GJ

Aims

Joint registries typically use revision of an implant as an endpoint and report survival rates after a defined number of years. However, reporting lifetime risk of revision may be more meaningful, especially in younger patients. We aimed to assess lifetime risk of revision for patients in defined age groups at the time of primary surgery.

Methods

The New Zealand Joint Registry (NZJR) was used to obtain rates and causes of revision for all primary total hip arthroplasties (THAs) performed between January 1999 and December 2016. The NZJR is linked to the New Zealand Registry of Births, Deaths and Marriages to obtain complete and accurate data. Patients were stratified by age at primary surgery, and lifetime risk of revision calculated according to age, sex, and American Society of Anesthesiologists (ASA) classification. The most common causes for revision were also analyzed for each age group.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 28 - 37
1 Jan 2010
Jameson SS Langton DJ Nargol AVF

We present the early clinical and radiological results of Articular Surface Replacement (ASR) resurfacings in 214 hips (192 patients) with a mean follow-up of 43 months (30 to 57). The mean age of the patients was 56 years (28 to 74) and 85 hips (40%) were in 78 women. The mean Harris hip score improved from 52 (11 to 81) to 95 (27 to 100) at two years and the mean University of California, Los Angeles activity score from 3.9 (1 to 10) to 7.4 (2 to 10) in the same period. Narrowing of the neck (to a maximum of 9%) was noted in 124 of 209 hips (60%). There were 12 revisions (5.6%) involving four (1.9%) early fractures of the femoral neck and two (0.9%) episodes of collapse of the femoral head secondary to avascular necrosis. Six patients (2.8%) had failure related to metal wear debris. The overall survival for our series was 93% (95% confidence interval 80 to 98) and 89% (95% confidence interval 82 to 96) for hips with acetabular components smaller than 56 mm in diameter. The ASR implant has a lower diametrical clearance and a subhemispherical acetabular component when compared with other more frequently implanted metal-on-metal hip resurfacings. These changes may contribute to the higher failure rate than in other series, compared with other designs. Given our poor results with the small components we are no longer implanting the smaller size


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_9 | Pages 17 - 17
1 May 2018
Woodnutt D Mullins M Mohammed A Dodd M
Full Access

Introduction. Proximal short stems have gained in popularity for perceived bone preservation but more recently, physiological loading. We report the medium term success of a calcar loading, short stem in a large cohort from a single unit with multiple surgeons. Patients/Materials & Methods. Prospectively collected sequential data, with no loss to follow-up, was retrospectively analysed from our own local database. Demographic data was assimilated to collect age, sex, BMI smoking history. Revision cases were analysed for cause and a PTIR and Kaplan Meier Curve constructed to quantify survival. Radiographs reviewed in 100 of the oldest (2009 to 2011) cases to look specifically for calcar resorption and stem sink. Imatri software was used for calculations: sink was categorised into greater or less than 0.5cm; calcar resorption was estimated using maximal osteotomy surface change of slope. Patient pre-operative and post-operative outcome scores (using Harris Hip, Charnley and Oxford scales) were extracted from the database. Results. 752 cases (in 685 patients) were identified: 313 male, 372 females with a combined average age of 61 (19 to 84); and, an average BMI of 29 kg/m. 2. ; (18–49); and, 4% were noted to be smoking on their last assessment with a maximum implantation of 9 (average 4.2) years. 74% (556) of cases utilised a 36mm Solid Ceramic head but 28, 32, 40 and 44mm heads in CrCo and TiN were also used. The majority (84%) of cases were performed for Osteoarthritis (Inflammatory arthropathy was a relative contraindication); but, 22 cases were performed for revision from a resurfacing arthroplasty. 17 patients were noted to have died (all more than 90 days post-operatively). 14 cases were revised for any cause (1.8%): 5 for cup related reasons (including dislocation); 4 deep infection, 2 for failure of fixation; 3 for periprosthetic fracture. All failures occurred within the first 4.5 years and all within the first 100 cases for each surgeon. The KM survival curve gave a 5 year survival (185 at risk) of 96.1% all causes and 99.1% for femur only. The PTIR (non-competing) was 6.5 (per 1000 years exposed) and 6.3 (competing). No reported cases of thigh pain were observed; one ceramic on ceramic bearing reported transient but recurrent squeak within the first 2 years before settling. Average sink was less than 1mm but greater than 0.5cm in 5 cases (2 of which required revision for non-fixation). All sink occurred within the first 6 months. Calcar resorption was only noted in 3 cases and was greater than 45° from the prosthesis bone interface. Discussion. Our results are similar to unpublished, smaller outcomes of the same stem in other centres. Two relatively new modes of failure were noted with sink and early periprosthetic fracture accounting for 56% of stem related failures. In this novel designed stem, there appears to be a narrow window between under and over-sizing with the methods of failure of sink or fracture probable


The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 49 - 55
1 Apr 2017
Hossain F Konan S Volpin A Haddad FS

Aims

The aim of this study was to compare early functional and health related quality of life outcomes (HRQoL) in patients who have undergone total hip arthroplasty (THA) using a bone conserving short stem femoral component and those in whom a conventional length uncemented component was used. Outcome was assessed using a validated performance based outcome instrument as well as patient reported outcome measures (PROMs).

Patients and Methods

We prospectively analysed 33 patients whose THA involved a contemporary proximally porous coated tapered short stem femoral component and 53 patients with a standard conventional femoral component, at a minimum follow-up of two years. The mean follow-up was 31.4 months (24 to 39). Patients with poor proximal femoral bone quality were excluded. The mean age of the patients was 66.6 years (59 to 77) and the mean body mass index was 30.2 kg/m2 (24.1 to 41.0). Outcome was assessed using the Oxford Hip Score (OHS) and the University College Hospital (UCH) hip score which is a validated performance based instrument. HRQoL was assessed using the EuroQol 5D (EQ-5D).


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 468 - 474
1 Apr 2016
Abdel MP Houdek MT Watts CD Lewallen DG Berry DJ

Aims

The goals of this study were to define the risk factors, characteristics, and chronology of fractures in 5417 revision total hip arthroplasties (THAs).

Patients and Methods

From our hospital’s prospectively collected database we identified all patients who had undergone a revision THA between 1969 and 2011 which involved the femoral stem. The patients’ medical records and radiographs were examined and the relevant data extracted. Post-operative periprosthetic fractures were classified using the Vancouver system. A total of 5417 revision THAs were identified.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 461 - 467
1 Apr 2016
Abdel MP Watts CD Houdek MT Lewallen DG Berry DJ

Aim and Methods

The goals of this study were to define the risk factors, nature, chronology, and treatment strategies adopted for periprosthetic femoral fractures in 32 644 primary total hip arthroplasties (THAs).

Results

There were 564 intra-operative fractures (1.7%); 529 during uncemented stem placement (3.0%) and 35 during cemented stem placement (0.23%). Intra-operative fractures were more common in females and patients over 65 years (p < 0.001). The majority occurred during placement of the femoral component (60%), and involved the calcar (69%). There were 557 post-operative fractures (20-year probability: 3.5%; 95% confidence interval (CI) 3.2 to 3.9); 335 fractures after placement of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to 9.1) and 222 after placement of a cemented stem (20-year probability: 2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture within 30 days after an uncemented stem was ten times higher than a cemented stem. The most common post-operative fracture type was a Vancouver AG (32%; n = 135), with 67% occurring after a fall. In all, 36% (n = 152) were treated with revision arthroplasty.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 442 - 448
1 Apr 2014
Feyen H Shimmin AJ

Many different lengths of stem are available for use in primary total hip replacement, and the morphology of the proximal femur varies greatly. The more recently developed shortened stems provide a distribution of stress which closely mimics that of the native femur. Shortening the femoral component potentially comes at the cost of decreased initial stability. Clinical studies on the performance of shortened cemented and cementless stems are promising, although long-term follow-up studies are lacking. We provide an overview of the current literature on the anatomical features of the proximal femur and the biomechanical aspects and clinical outcomes associated with the length of the femoral component in primary hip replacement, and suggest a classification system for the length of femoral stems.

Cite this article: Bone Joint J 2014;96-B:442-8.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1344 - 1350
1 Oct 2012
Penny JO Ding M Varmarken JE Ovesen O Overgaard S

Radiostereometric analysis (RSA) can detect early micromovement in unstable implant designs which are likely subsequently to have a high failure rate. In 2010, the Articular Surface Replacement (ASR) was withdrawn because of a high failure rate. In 19 ASR femoral components, the mean micromovement over the first two years after implantation was 0.107 mm (sd 0.513) laterally, 0.055 mm (sd 0.204) distally and 0.150 mm (sd 0.413) anteriorly. The mean backward tilt around the x-axis was -0.08° (sd 1.088), mean internal rotation was 0.165° (sd 0.924) and mean varus tilt 0.238° (sd 0.420). The baseline to two-year varus tilt was statistically significant from zero movement, but there was no significant movement from one year onwards.

We conclude that the ASR femoral component achieves initial stability and that early migration is not the mode of failure for this resurfacing arthroplasty.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1045 - 1051
1 Aug 2013
Arthroplasty Society C

The purpose of this study was twofold: first, to determine whether the five-year results of hip resurfacing arthroplasty (HRA) in Canada justified the continued use of HRA; and second, to identify whether greater refinement of patient selection was warranted.

This was a retrospective cohort study that involved a review of 2773 HRAs performed between January 2001 and December 2008 at 11 Canadian centres. Cox’s proportional hazards models were used to analyse the predictors of failure of HRA. Kaplan–Meier survival analysis was performed to predict the cumulative survival rate at five years. The factors analysed included age, gender, body mass index, pre-operative hip pathology, surgeon’s experience, surgical approach, implant sizes and implant types. The most common modes of failure were also analysed.

The 2773 HRAs were undertaken in 2450 patients: 2127 in men and 646 in women. The mean age at operation was 50.5 years (sd 8.72; 18 to 82) and mean follow-up was 3.4 years (sd 2.1; 2.0 to 10.1). At the last follow-up a total of 101 HRAs (3.6%) required revision. Using revision for all causes of failure as the endpoint, Kaplan–Meier survival analysis showed a cumulative survival of 96.4% (95% confidence interval (CI) 96.1 to 96.9) at five years. With regard to gender, the five-year overall survival was 97.4% in men (95% CI 97.1 to 97.7) and 93.6% in women (95% CI 92.6 to 94.6). Female gender, smaller femoral components, specific implant types and a diagnosis of childhood hip problems were associated with higher rates of failure. The most common cause of failure was fracture of the femoral neck, followed by loosening of the femoral component.

The failure rates of HRA at five years justify the ongoing use of this technique in men. Female gender is an independent predictor of failure, and a higher failure rate at five years in women leads the authors to recommend this technique only in exceptional circumstances for women.

Cite this article: Bone Joint J 2013;95-B:1044–50.