Abstract

The stem design in total hip arthroplasty (THA) is constantly evolving. The impact of the collar on the risk of periprosthetic fracture remains controversial. This study aimed to determine whether adding a collar to the femoral stem impacts the early periprosthetic fracture risk within 90 days of surgery.

This retrospective study included 1,623 patients who underwent primary total hip arthroplasty in a single orthopedic department between January 2010 and December 2020. The inclusion criteria were uncemented stem with or without a collar, in a primary intention, without previous hip surgery with a similar “corail like” design. The assessed data were demographic characteristics (age, gender, number of obese (BMI > 30)), single or dual mobility, the surgical approach, the early complication, particularly the periprosthetic femoral fractures. Of the 1,623 patients, 1,380 received a collared stem (85%), and 243 received a collarless stem (15%). A multivariate analysis was performed to determine the collar's effect on the risk of early periprosthetic fracture (<90 days).

Nine (0.55%) early periprosthetic fractures were identified in the whole cohort. There were four fractures (1.65%) in the collarless stem group and five fractures (0.36%) in the collared stem group (p=0,005). One patient required revision surgery in the collared stem group, while two patients required revision surgery in the collarless stem group. The multinomial logistic regression model indicated a statistically significant lower (p<0.05) risk of early periprosthetic fracture within 90 days of surgery in the collared stems group. No other risk factor for early periprosthetic fractures has been identified.

Using collared stems in cementless THA protects early periprosthetic femoral fractures within 90 days of surgery.