The benefits of HXLPE in total knee arthroplasty (TKA) have not been as evident as total hip arthroplasty (THA). A systematic review and meta-analysis to assess the impact of highly-crosslinked polyethylene (HXLPE) on TKA outcomes compared to
Introduction. Highly cross-linked (HXL) polyethylene has demonstrated clinical advantages as a wear resistant acetabular bearing material in total hip arthroplasty (THA) [1]. In vitro wear testing has predicted a tenfold reduction in the wear rate of HXL polyethylene, as compared to its conventional, non-HXL counterpart [2]. To date, radiographic studies of head penetration represent the state-of-the-art in determining clinical wear of polyethylene hip liners [3]. However, as the amount of wear drops to very low levels, it becomes important to develop a precise and reliable method for measuring wear, facilitating a comparison of clinical results to expectations. This study focuses on locating and quantifying the maximum linear wear of retrieved acetabular poly liners using a coordinate measuring machine (CMM). Specifically, HXL liners are compared to a baseline of conventional, non-HXL bearings. Methods. An IRB-approved retrieval laboratory received 63 HXL acetabular bearing retrievals from 5 manufacturers with in vivo durations of 1.01–14.85 years. These were compared with 32 conventional, non-HXL controls (including gas plasma, gamma-barrier and EtO) from 3 manufacturers with in vivo durations of 1.03–20.89 years. Liners were mounted in a tripod of axial contacts with the liner face positioned in a vertical plane. Each bearing was scanned with a CMM dual-probe head, with one horizontal probe scanning the articular surface and the other scanning the non-articular, sequentially. Surface-normal wall thickness values along each latitude were calculated using a custom developed algorithm (Figure 1). Because the liners are axially symmetric as manufactured, deviation in wall thickness at a given latitude represents linear wear [4]. Results. Total wear penetration for the HXL liners ranged from 0.02 to 1.03 mm, and for the conventional, non-HXL controls ranged from 0.07 to 6.85 mm. The HXL liners had an average linear wear rate of 0.02 mm/year, compared to 0.20 mm/year for the conventional, non-HXL controls (Figure 2). The direction of maximum wear, as measured in degrees from the cup pole, ranged from 8.32 to 73.86 degrees. Differences in wear rates as a function of crosslinking dose, as well as presence/absence of a lip can be identified. Discussion. This wear measurement study of retrievals is the first application of a novel CMM technique to locate and quantify wear in HXL liners compared to
Introduction:. Although commonly used, the clinical performance of highly crosslinked polyethylene in total knee arthroplasty (TKA) remains unknown and concerns exist regarding fatigue resistance and oxidation, particularly in posterior-stabilized (PS) designs. The purpose of this study is to compare highly crosslinked and
Purpose. Cross-linking of polyethylene greatly reduces its wear rate in hip simulator studies. We conducted a systematic review and meta-analysis of randomized controlled trials comparing cross-linked to
Introduction. Ceramic femoral heads have proven to be more scratch resistant with better wettability and improved wear characteristics compared to metal heads in the laboratory setting. The objective of this study was to compare long-term survivorship and in vivo wear rates of ceramic and metal femoral heads against
Purpose. The objective of this study was to compare the wear characteristics and damage scores in highly crosslinked (XLPE) and
Introduction. Total knee replacement (TKR) implant designs and materials have been shown to have a significant impact on tibial insert wear. A medial-pivot (MP) design theoretically should generate less wear due to a large contact area in the medial compartment and lower contact stresses. Synovial fluid aspiration studies have confirmed that a first generation MP TKR system (ADVANCE®, MicroPort Orthopedics Inc., Arlington, TN, USA) generates less wear debris than is seen with other implant designs articulating against
We report the outcome of 320 primary Total Hip Arthroplasties (THA) with minimum 10-year follow-up (range 10–17 years, mean 12.6 years), performed by a single surgeon in Tauranga New Zealand, with the Exeter Contemporary Flanged all-polyethylene cup and Exeter femoral stem via a posterior approach. The aim of the study is to compare the results with the published results from the design centre and create a baseline cohort for further outcomes research in this centre. All patients were prospectively followed at 6 weeks, 1 year, 5 years, 10 years, (and 15 years when available). Of 333 cases that matched the inclusion criteria, 13 procedures in 12 patents were excluded because of concomitant bone grafting and/or supplementary cage fixation, leaving 320 primary THA procedures in 280 patients, including 26 bilateral procedures in 13 patients. Mean follow-up of the surviving cases was 12.6 (range 5.0-17.1) years. There were 12 revisions – 2 for fracture, 5 for instability, 1 for impingement pain and 4 for infection. There were no revisions for aseptic cup loosening. Kaplan-Meier survivorship with revision for aseptic loosening as the endpoint was 100% at 15.0 years (with minimum 40 cases remaining at risk). All-cause acetabular revision in 12 cases result in a Kaplan-Meier survival of 95.9% (95% CI: 93.5 to 98.3%). Cemented THA with the Exeter Contemporary Flanged cup and the Exeter stem is a durable combination with results that can be replicated outside of the design centre. The Exeter Contemporary Flanged cup has excellent survivorship at 15 years when used with the Exeter stem. Cemented THA with well-proven components should be considered the benchmark against which newer designs and materials should be compared.
The 2021 Australian Orthopaedic Association National Joint Replacement Registry report indicated that total shoulder replacement using both mid head (TMH) length humeral components and reverse arthroplasty (RTSA) had a lower revision rate than stemmed humeral components in anatomical total shoulder arthroplasty (aTSA) - for all prosthesis types and diagnoses. The aim of this study was to assess the impact of component variables in the various primary total arthroplasty alternatives for osteoarthritis in the shoulder. Data from a large national arthroplasty registry were analysed for the period April 2004 to December 2020. The study population included all primary aTSA, RTSA, and TMH shoulder arthroplasty procedures undertaken for osteoarthritis (OA) using either cross-linked polyethylene (XLPE) or non-cross-linked polyethylene (non XLPE). Due to the previously documented and reported higher revision rate compared to other anatomical total shoulder replacement options, those using a cementless metal backed glenoid components were excluded. The rate of revision was determined by Kaplan-Meir estimates, with comparisons by Cox proportional hazard models. Reasons for revision were also assessed. For a primary diagnosis of OA, aTSA with a cemented XLPE glenoid component had the lowest revision rate with a 12-year cumulative revision rate of 4.7%, compared to aTSA with cemented non-XLPE glenoid component of 8.7%, and RTSA of 6.8%. The revision rate for TMH was lower than aTSA with cemented non-XLPE, but was similar to the other implants at the same length of follow-up. The reason for revision for cemented aTSR was most commonly component loosening, not rotator cuff deficiency. Long stem humeral components matched with XLPE in aTSA achieve a lower revision rate compared to shorter stems, long stems with
Contemporary polyethylene liners for total hip replacements were introduced in the late 1990's to address osteolysis associated with wear of
Isolated liner and head exchange procedure has been an established treatment method for polyethylene wear and osteolysis when the acetabular component remains well-fixed. In this study, its mid-term results were evaluated retrospectively in 34 hips. Among the consecutive patients operated upon from September 1995, 2 patients (3 hips) were excluded because of inadequate follow-up and the results of remaining 34 hips of 34 patients were evaluated. They were 20 men and 14 women with a mean age of 49 years at the time of index surgery.
INTRODUCTION. Since the early 2000s, highly cross-linked (HXL) UHMWPE's have become a popular option with multiple experimental and clinical studies showing that gamma or electron radiation doses between 50–100kGY reduce wear and potentially extend the bearing life of UHMWPE. However, the increased wear resistance came at a compromise to mechanical properties due to the cross-linking process. Vitamin E has been added to some HXL UHMWPE materials to offer a solution to the compromise by increasing oxidation resistance and maintaining sufficient fatigue strength. However, limited data is available on the effect of the fabrication process, especially the method of irradiation, on the properties of the Vitamin E blended HXL UHMWPE. The purpose of this study was to evaluate the effects of adding the antioxidant vitamin E to highly crosslinked UHMWPE on wear rates. METHODS. Wear testing was performed on six highly crosslinked UHMWPE acetabular liners containing vitamin E (0.1% wt. alpha tocopherol) fabricated using the Cold Irradiation Mechanically Annealed (CIMA) process, initially cross-linked with approximately 100 kGy gamma irradiation, and terminally gamma sterilized. The liners were paired with three 40mm CoCr femoral heads and 40mm three ceramic femoral heads. Testing was completed per ASTM F1714 and ISO 14242 on an orbital hip joint wear simulator (Shore Western, California) and lubricated with 90% bovine calf serum, 20mM EDTA, 0.2% wt. NaN. 3. and DI water. A 1.1Hz Paul-type loading waveform with a peak of 2kN was used for a total of 5E6 wear cycles. Three loaded soak controls were used in parallel to adjust for fluid absorption. Samples were weighed every 5E5 wear cycles. RESULTS. The wear rates for the HXL blended vitamin-E liners were calculated using the slope of the linear regression over the steady state and resulted in a wear rate of 0.49mg/Mc. This is a decrease of approximately 95% compared to the 9.54 mg/Mc 28mm ID conventional UHMWPE wear rates as well as a notable difference for the other HXL UHMWPE liner wear rates discussed in the review. DISCUSSION. HXL blended vitamin-E 40mm liners demonstrated an approximate 95% reduction in wear rates compared to a 28mm ID conventional UHMWPE. The reduced wear rate confirmed the design expectation that a higher irradiation dose in the fabrication process resulted in an increased amount of polymer crosslinking. Additionally, the wear rate of the HXL blended vitamin-E liners studied was well below 20mg/Mc, which was shown by Dowd et al. to be the threshold of osteolysis in THA. SIGNIFICANCE. The HXL vitamin E blended UHMWPE liner tested in this study demonstrated reduced wear rates by approximately 95% compared to
Introduction. Ceramic femoral heads have superior scratch resistant with better wettability and improved wear characteristics compared to metal heads in the laboratory setting. The objective of this study was to compare long-term in vivo wear rates of ceramic and metal femoral heads against
INTRODUCTION. Aseptic loosening is the most common failure mode for Total Elbow Arthroplasty (TEA) and is considered to be associated with accelerated polyethylene bearing wear [1, 2]. This study aimed to evaluate three commercially available implant designs under loads associated with daily living. The hypothesis was that more recent designs (Discovery and Nexel) provide greater articular contact areas resulting in lower polyethylene stresses compared to the Coonrad/Morrey (CM). METHODS. Motion tracking was performed on a healthy volunteer during elbow flexion at 0, 45, and 90° shoulder abduction because most daily activities occur with some shoulder abduction [3] resulting in varus stress about the elbow. This kinematic data was used in an OpenSim upper extremity musculoskeletal model [4] to estimate muscle and joint reaction loads with 5lb in hand, consistent with the common clinical restrictions following TEA. Computer aided assemblies of the smallest size implants for each system were imported to ANSYS for finite element analysis. Metallic components were treated as rigid and polyethylene components were modeled using a nonlinear elastoplastic constitutive model calibrated to material data. Articular contacts were frictional. Physiologic joint reaction forces and moments quantified in OpenSim were applied and the resulting peak articular contact area and peak bearing von Mises stresses were assessed. RESULTS. Simulated deformation patterns of CM bearings corresponded well to those reported in retrievals studies [1, 2] supporting the clinical relevance of the modeling approach. Peak stresses for CM and Nexel were consistently found in the central and side bearings respectively. The central bearing stresses remained 2–2.6 times lower in Nexel compared to CM. Peak stress for all three TEA systems increased with shoulder abduction (Fig.1, 2). Highest peak stresses (Fig.2) were obtained in CM and consistently exceeded the polyethylene yield limit; CM showed the lowest contact area (Fig.3). Nexel and Discovery experienced peak polyethylene stresses 26–34% and 17–39% lower than CM respectively (Fig.2). DISCUSSION. Our results support the hypothesis that newer TEA systems provide increased articular contact area and reduced bearing stresses during physiological loading. The cylindrical CM central bearing carries both the joint reaction force and moment leading to edge loading and high stresses (Fig.1). The design of the Nexel central bearing provides limited resistance to varus-valgus moment, thus transferring the moment to the side bearings and reducing central bearing stresses. The hemispherical Discovery bearing design was confirmed to offer a large articular contact area. However, non-concentricity of the contact spheres can lead to edge loading and high polyethylene stresses under off-axis forces. CM and Discovery utilize
Introduction. Technological advances in the processing of polyethylene have led to improved survivorship of total hip arthroplasty. The purpose of this study was to determine if a second generation highly cross-linked polyethylene could improve upon wear rates compared to conventional and first generation cross-linked polyethylene in patients undergoing primary total hip arthroplasty. Methods. Linear and volumetric wear rates of a second generation highly cross-linked polyethylene were evaluated following primary total hip arthroplasty. There were 44 patients with an average age of 68.6 years and mean follow-up of 5.3 years. Patients were evaluated at six weeks, one, two and five years. Wear rates were determined from digitized AP Pelvis radiograph by an independent observer using Martell's software. Acetabular inclination and femoral head size were also evaluated to determine variability in wear rates. Results. The mean linear wear rate for the entire group was 0.015mm/year (±0.055). There was a 64% decrease in linear wear rate when compared to a first generation highly cross-linked polyethylene from the same institution. There was a 90% decrease in wear rate compared to
Total hip arthroplasty (THA) is one of the most successful and commonly performed surgical interventions worldwide. Based on registry data, at one-year post THA, implant survivorship is nearly 100% and patient satisfaction is 90%. A novel, porous coated acetabular implant was introduced in Europe and Australia in 2007. Several years after its introduction, warnings were issued for the system when used with metal-on-metal bearings due to adverse local tissue reaction, with one study reporting a 24% failure rate (Dramis et al. 2014). A subsequent 2018 study by Teoh et al. showed that the acetabular system had a survival rate of 98.9% at five years when used with
Introduction. Dislocation is one of the leading causes of revision after primary total hip arthroplasty (THA). Polyethylene wear is one of the risk factors for late dislocations (>2 years). It can induce an inflammatory response resulting in distension and thinning of the pseudocapsule, predisposing the hip to dislocation. Alternatively, eccentric seating of the femoral head in a worn out socket may result in an asymmetric excursion arc predisposing the hip to impingement, levering out and dislocation. Highly cross linked polyethylene has a significantly lower wear rate as compared to
Dual mobility components for total hip arthroplasty provide for an additional articular surface, with the goals of improving range of motion, jump distance, and overall stability of the prosthetic hip joint. A large polyethylene head articulates with a polished metal acetabular component, and an additional smaller metal or ceramic head is snap-fit into the large polyethylene. In some European centers, these components are routinely used for primary total hip arthroplasty. However, their greatest utility will be to prevent and manage recurrent dislocation in the setting of revision total hip arthroplasty. Several retrospective series have shown satisfactory results for this indication at medium-term follow-up times. The author has used dual mobility components on two occasions to salvage a failed constrained liner. At least one center reports that dual mobility outperforms 40mm femoral heads in revision arthroplasty. Modular dual mobility components, with screw fixation, are the author's first choice for the treatment of recurrent dislocation, revision of failed metal-on-metal resurfacing or total hips, unipolar arthroplasties, and salvage of failed constrained liners. There are concerns of elevated metal levels with one design, and acute early intra-prosthetic dissociation following attempted closed reduction. Total hip surgeons no longer use
INTRODUCTION. Many studies have looked at the effects of titanium tibial baseplates compared to cobalt chrome baseplates on backside wear. However, the surface finish of the materials is usually different (polished/unpolished) [1,2]. Backside wear may be a function not only of tray material but also of the locking mechanism. The purpose of this study was to evaluate the wear performance of
Dual mobility components for total hip arthroplasty provide for an additional articular surface, with the goals of improving range of motion, jump distance, and overall stability of the prosthetic hip joint. A large polyethylene head articulates with a polished metal acetabular component, and an additional smaller metal or ceramic head is snap-fit into the large polyethylene. In some European centers, these components are routinely used for primary total hip arthroplasty. However, their greatest utility will be to prevent and manage recurrent dislocation in the setting of revision total hip arthroplasty. Several retrospective series have shown satisfactory results for this indication at medium-term follow-up times. The author has used dual mobility components on two occasions to salvage a failed constrained liner. At least one center reports that dual mobility outperforms 40mm femoral heads in revision arthroplasty. Modular dual mobility components, with screw fixation, are the author's first choice for the treatment of recurrent dislocation, revision of failed metal-metal resurfacing, total hips, unipolar arthroplasties, and salvage of failed constrained liners. There are concerns of elevated metal levels with one design, and acute early intra-prosthetic dissociation following attempted closed reduction. Total hip surgeons no longer cement Charnley acetabular components, use