The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery. A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series.Aims
Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
Aims.
Aims. The aim of this study was to compare the actual cost of a cemented and
Aim. To examine the effects of total knee arthroplasty on markers of inflammation and endothelial dysfunction, as surrogate markers for enhanced risk of vascular disease or precipitation of acute vascular events post-operatively. Methods. All patients undergoing an elective
Introduction. Cemented total knee arthroplasty (TKA) remains the gold standard with survivorship above 90% at greater than 10 years postoperatively. However, with younger, heavier, more active patients undergoing TKA at an increasing rate, cementless implants have the appeal of potential for improved implant fixation longevity and decreased rates of aseptic loosening. The cementless implants are more expensive than their cemented counterparts such that implant costs may create a barrier to utilization. However, such comparisons fail to consider the unavoidable additional costs of cementing, including the cost of operating room time, cement and cementing accessories. The purpose of this study is to compare the actual cost of cemented and cementless TKA. Methods. The TKA cost calculation included the cost of operative time, implants, cement and cementing accessories. The difference in operative time between cemented and cementless TKA was determined from a previously published study of 100 TKAs performed using a cemented (55) or press fit (45) implant of the same design performed at a single institution by four fellowship trained arthroplasty surgeons. The decision to use cemented or cementless design in these patients was made based on patient bone quality intraoperatively. Operative time was compared between groups using a Student's two-tailed T-test. The cost of operating room time was based on estimates in the recent literature. The cost of cement and cementing accessories was estimated based on publically available market data. The cost of implants was estimated from institutional data for multiple companies. Results. The cost comparison between cemented and
Introduction. The addition of hydroxyapatite in
Intra-articular administration of antibiotics during primary total knee arthroplasty (TKA) may represent a safe, cost-effective strategy to reduce the risk of acute periprosthetic joint infection (PJI). Vancomycin with an aminoglycoside provides antimicrobial cover for most organisms isolated from acute PJI after TKA. However, the intra-articular doses required to achieve sustained therapeutic intra-articular levels while remaining below toxic serum levels is unknown. The purpose of this study is to determine the intra-articular and serum levels of vancomycin and tobramycin over the first 24 hours postoperatively after intra-articular administration in primary cementless TKA. A prospective cohort study was performed. Patients were excluded if they had poor renal function, known allergic reaction to vancomycin or tobramycin, received intravenous vancomycin, or were scheduled for same-day discharge. All patients received 600 mg tobramycin and 1 g of vancomycin powder suspended in 25 cc of normal saline and injected into the joint after closure of the arthrotomy. Serum from peripheral venous blood and drain fluid samples were collected at one, four, and 24 hours postoperatively. All concentrations are reported in µg per ml.Aims
Methods
The ideal type of total knee arthroplasty (TKA) prosthesis remains a debatable topic with many different options available. Uncemented TKA has been a viable option due to its decreased operating room (OR) time but also because of its proposed improved long term fixation. Unfortunately, in the past uncemented TKA was associated with increased blood loss. Surgical technique and perioperative treatments have changed since these original studies and tranexamic acid (TXA) has become the gold standard for TKA blood loss management. The objective of this study was to evaluate if there was a difference in hemoglobin and hematocrit change, along with blood loss volume during surgery between cemented and cementless TKA when modern blood loss techniques are utilized We retrospectively reviewed data from TKAs performed by three high volume surgeons between 2016 and 2019. We excluded bilateral TKA, revisions, hardware removal intraoperatively and other indications for TKA than primary OA. Power analysis determined 85 patients in both the cementless and cemented TKA groups. Patients were matched 1:1 for age, sex, BMI and surgeon. Use of TXA, intraoperative blood loss, differences in hemoglobin and hematocrit pre- and postoperatively days one, two, and three were recorded. Continuous variables were analyzed using T-tests and categorical variables were evaluated using Chi-squared tests.Introduction
Methods
Cementless TKA offers the potential for strong fixation through biologic fixation technology as compared to cemented TKA where fixation is achieved through mechanical integration of the cement. Few mid-term results are available for newer cementless TKA designs that have used additive manufacturing (3-D printing) for component design. The purpose of this study is to present minimum 5-year clinical outcomes and implant survivorship of a specific cementless TKA using a novel 3-D printed tibial baseplate. This is a single institution registry review of the prospectively obtained data on 296 cementless TKA using a novel 3-D printed tibial baseplate with minimum 5-year follow-up. Outcomes were determined by comparing pre- and post-operative Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) scores and pre- and post-operative 12 item Veterans RAND/Short Form Health Survey (VR/SF-12). Aseptic loosening as well as revision for any reason were the endpoints used to determine survivorship at 5 years.Introduction
Methods
In recent years, the use of modern cementless implants in total knee arthroplasty has been increasing in popularity. These implants take advantage of new technologies such as additive manufacturing and potentially provide a promising alternative to cemented implant designs. The purpose of this study was to compare implant migration and tibiofemoral contact kinematics of a cementless primary total knee arthroplasty (TKA) implanted using either a gap balancing (GB) or measured resection (MR) surgical technique. Thirty-nine patients undergoing unilateral TKA were recruited and assigned based on surgeon referral to an arthroplasty surgeon who utilizes either a GB (n = 19) or a MR (n = 20) surgical technique. All patients received an identical fixed-bearing, cruciate-retaining beaded peri-apatite coated cementless femoral component and a pegged highly porous cementless tibial baseplate with a condylar stabilizing tibial insert. Patients underwent a baseline radiostereometric analysis (RSA) exam at two weeks post-operation, with follow-up visits at six weeks, three months, six months, and one year post-operation. Migration including maximum total point motion (MTPM) of the femoral and tibial components was calculated over time. At the one year visit patients also underwent a kinematic exam using the RSA system.Background
Methods
Cementless total knees were historically associated with early failure. These failures, likely associated with implant design, made cemented total knee arthroplasty (TKA) the “gold standard”. Manufacturers have introduced newer uncemented technologies that provide good initial stability and utilize a highly-porous substrates for bony in-growth. Outcome data on these implants has been limited. In addition, these implants typically have a price premium which makes them difficult to use in the setting of cost containment and in at risk 90-day bundles. Our purpose was to compare 90-day outcomes of a new uncemented implant with those of a comparable cemented implant from the same manufacturer. We hypothesized that the implants would have equivalent 90-day clinical and economic outcomes. Ninety-day clinical and economic outcomes for 252 patients with prospectively collected data from the Michigan Arthroplasty Registry Collaborative Quality Initiative (MARCQI) database were reviewed. Ninety-day outcomes were compared between uncemented knees and an age-matched group of cemented knees (Triathlon cemented vs uncemented Triathalon-tritanium, Stryker Orthopedics, Mahwah, NJ, USA). Both cruciate retaining and posterior stabilized designs were included. MARCQI data: demographics, co-morbidities, length of stay, complications, emergency department visits, discharge disposition, and readmissions were reviewed. Financial data provided by the hospital's finance department was used for economic comparisons. Fischer's test was done to assess categorical data and a student's t-test was used to compare numerical data.Introduction
Methods
Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term migration with one- and two-year migration to evaluate the predictive ability of short-term migration data and to compare migration and inducible displacement between cemented and cementless (porous metal monoblock) components at least ten years postoperatively. Patients who had participated in RSA migration studies with two-year follow-up were recruited to return for a long-term follow-up, at least ten years from surgery. Two cemented tibial designs from two manufacturers and one porous metal monoblock cementless tibial design were studied. At the long-term follow-up, patients had supine RSA examinations to determine migration and loaded examinations (single leg stance) to determine inducible displacement. In total, 79 patients (54 female) returned, with mean time since surgery of 12 years (10 to 14). There were 58 cemented and 21 cementless tibial components.Aims
Patients and Methods
In a series of 1304 patients (1867 knees), the results of simultaneous and staged bilateral total knee arthroplasty were compared with each other and with unilateral total knee arthroplasty. The bilateral procedures had a significantly higher rate of complications than unilateral procedures, almost entirely because of thromboembolic problems. However, this did not correspond to an increase in mortality. If a bilateral procedure was indicated, then a simultaneous procedure had no increased risk over a staged procedure. There was no increase in cardiovascular complications, the rate of deep-vein thrombosis or pulmonary embolism or mortality. The rate of infection was lower with a bilateral procedure and the overall revision rate was less than 1% in all groups. The prosthesis functioned as well in all groups in the medium and longer term periods. We feel that simultaneous bilateral total knee arthroplasty is a safe and successful procedure when compared with a staged bilateral procedure. It also has the added benefit of single anaesthetic, reduced costs and decreased total recovery time when compared to a staged bilateral procedure. For these reasons it should be considered as an option in the presence of bilateral knee joint disease.
Aims. The use of
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in
Abstract. Introduction.
In this study we present our experience with
four generations of
We performed a prospective study of 54 patients (76 knees) who underwent Osteonics series 3000 cruciate-retaining
Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening. A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.Aims
Methods